AMTI Force Plate System. AMTI Force Plate & Amplifier. AMTI Amplifier. Zero Reset Button. Press before each trial. Incorrect Amplifier Settings Clips Force. Over amplification causes clipping of force signals. - PowerPoint PPT Presentation

Transcript of AMTI Force Plate System

Page 1: AMTI Force Plate System

AMTI Force Plate AMTI Force Plate SystemSystem

Page 2: AMTI Force Plate System

AMTI Force Plate & AmplifierAMTI Force Plate & Amplifier

Page 3: AMTI Force Plate System

AMTI AmplifierAMTI Amplifier

Zero Reset Button. Press before each trial.

Page 4: AMTI Force Plate System

Incorrect Amplifier Settings Clips Incorrect Amplifier Settings Clips ForceForce

0

200

400

600

800

1000

1200

1400

1600

0 50 100 150 200 250 300 350

Time (ms)

Ve

rtic

al F

orc

e (

N)

0

200

400

600

800

1000

1200

1400

0 50 100 150 200 250 300 350

Time (ms)

Ve

rtic

al F

orc

e (

N)

Notice the SQUARE tops on the force curve on the right. The force values are being clipped. Reduce the Fz AMTI Amplifier Gain from 4000 to 2000, or from 2000 to 1000 to eliminate clipping of force.

Over amplification causes clipping of force signals.

Page 5: AMTI Force Plate System

Software Amplifier Settings Must Software Amplifier Settings Must Match AMTI Amplifier Gain Settings.Match AMTI Amplifier Gain Settings.

These settings (4000, 2000, 1000) must match.