Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications...

12
ISO 9001 Manufacturer pressure Specification Sheet SS-IPI Pressure range IPI100A 0 to 7 bar absolute / 100 psi IPI030C Vacuum to 2 bar / 30 psi IPI100C Vacuum to 7 bar / 100 psi IPI300C Vacuum to 21 bar / 300 psi IPI500C Vacuum to 35 bar / 500 psi IPI015G 0 to 1 bar / 15 psi IPI01KG 0 to 70 bar / 1,000 psi IPI02KG 0 to 140 bar / 2,000 psi IPI03KG 0 to 200 bar / 3,000 psi IPI05KG 0 to 350 bar / 5,000 psi IPI10KG 0 to 700 bar / 10,000 psi High accuracy ±0.05% of F.S. for positive pressure Vacuum ranges to 500 psi (35 bar) High flexibility Easily disassemble the test system and use the indicator for other tasks True field indicator Lightweight and portable with full tempera- ture-compensation, long battery life and large display for easy visibility Record min and max readings Capture the min and max pressure readings for safety valve applications Complete marine program Part of a complete program of marine approved temperature, pressure and signal calibrators; including temperature sensors ATEX and CSA certification The IPI Indicator is ATEX and CSA certified and designed for use in potentially explosive environments ATEX JOFRA IPI indicators bring together the ease of an analog gauge with the accuracy and easy-to-read display of a digital calibrator. Use the JOFRA IPI in applications from safety valve checks to system pressure verification. Furthermore the IPI indicator is ATEX and CSA certified for use in poten- tially explosive environments such as oil refineries, chemical plants and offshore platform, where there is a risk of inflammable gases. This series is designed to meet your pressure measurement applications and make the tasks easier. The IPI offers 18 different pressure units, long battery life, high accuracy, and even serial communications. The accuracy of the IPI rivals that of a pressure calibrator and is temperature compensated for shop or in process applications. This versatile unit is available as a stand alone indicator or in a complete test system. You can perform a calibration locally without returning the IPI unit to the manufacturer. All you need is an accurate pressure reference. If you do require factory calibration, the pump and indicator are independent and only the indicator needs to be returned. The JOFRA IPI is available as an indicator or in one of 6 test ready sys- tems that are complete and equipped to meet your pressure measuring or testing needs. Industrial Pressure Calibrator IPI Indicators CSA Certified for use in potentially explosive environments

Transcript of Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications...

Page 1: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

ISO 9001 Manufacturer

p r e s s u r e

Specification SheetSS-IPI

Pressure rangeIPI100A 0to7barabsolute/100psiIPI030C Vacuumto2bar/30psiIPI100C Vacuumto7bar/100psiIPI300C Vacuumto21bar/300psiIPI500C Vacuumto35bar/500psiIPI015G 0to1bar/15psiIPI01KG 0to70bar/1,000psiIPI02KG 0to140bar/2,000psiIPI03KG 0to200bar/3,000psiIPI05KG 0to350bar/5,000psiIPI10KG 0to700bar/10,000psi

High accuracy±0.05% of F.S. for positive pressureVacuum ranges to 500 psi (35 bar)

High flexibilityEasily disassemble the test system and use the indicator for other tasks

True field indicatorLightweight and portable with full tempera-ture-compensation, long battery life and large display for easy visibility

Record min and max readingsCapture the min and max pressure readings for safety valve applications

Complete marine programPart of a complete program of marine approved temperature, pressure and signal calibrators; including temperature sensors

ATEX and CSA certificationThe IPI Indicator is ATEX and CSA certified and designed for use in potentially explosive environments

ATEX

JOFRAIPIindicatorsbringtogethertheeaseofananaloggaugewiththeaccuracyandeasy-to-readdisplayofadigitalcalibrator.

UsetheJOFRAIPIinapplicationsfromsafetyvalvecheckstosystempressureverification.

FurthermoretheIPIindicatorisATEXandCSAcertifiedforuseinpoten-tiallyexplosiveenvironmentssuchasoilrefineries,chemicalplantsandoffshoreplatform,wherethereisariskofinflammablegases.

Thisseriesisdesignedtomeetyourpressuremeasurementapplicationsandmakethetaskseasier.TheIPIoffers18differentpressureunits,longbatterylife,highaccuracy,andevenserialcommunications.TheaccuracyoftheIPIrivalsthatofapressurecalibratorandistemperaturecompensatedforshoporinprocessapplications.Thisversatileunitisavailableasastandaloneindicatororinacompletetestsystem.

YoucanperformacalibrationlocallywithoutreturningtheIPIunittothemanufacturer.Allyouneedisanaccuratepressurereference.Ifyoudorequirefactorycalibration,thepumpandindicatorareindependentandonlytheindicatorneedstobereturned.

TheJOFRAIPIisavailableasanindicatororinoneof6testreadysys-temsthatarecompleteandequippedtomeetyourpressuremeasuringortestingneeds.

Industrial Pressure Calibrator

IPI Indicators

CSA

Certified for use

in potentially explosive

environments

Page 2: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

2 Industrial Pressure Calibrator • IPI

Clear dual line displayTheIPIdisplaydoesmorethan

justshowapressurereading.Theusercanchecktemperatureand

exactbatteryvoltagethroughtheuseofthekeypad.The

large,51/2digit,0.65in(1.65cm)characterdisplayislargeandeasy-to-read;evenfromadistance.Iconsarealsoincludedtoindicatebatterylife,engineeringunits,andthereisabargraphthatshowsthepercentageofscalereading.Allofthisinformationcanbereadinlowlightwiththeuseofthebacklight.

Pressure connections1/4”NPTmalelowerconnec-tionisstandard.A1/4”BSPmaleadapterisincluded.TheIPImaybeorderedwithabackmountingconfiguration.Thisdesignhasthe1/4in.NPTmaleconnec-tiononthebackofthecasevicethelowermanifoldconnection.Thisallowsforuseinpanelmountedappli-cationsorinapplicationswhereflatmountingisnecessarysuchasinanoverhead.Thisconfigurationisnotdeliveredwithaprotectiveboot.

MAX / MINAMIN/MAXfunctionreadsthemaximumandminimumpressurethathavebeenauto-maticallystored.Monitorasafetyvalveorlookforpressurespikesintheprocess.TheIPIalsohavetheabilitytochecktheminimumandmaximumreadingsduringtheransient.

ZEROKeepthehighaccuracy:ZEROtheinstrumentbeforeeverytest.

ON / OFFTheIPIisdeliveredwiththeauto-shutoffactiveandsetto30minutes:likemanyofourotherinstruments.Thisfeaturecanbereducedfrom30minutesdownto1minutein1minuteincrements.Incaseswhereaconstantreadingisneces-sary,thiscanalsobeturnedcompletelyofftoallowforcontinuousoperation.

CONFIGThisisthekeytoaccessalluser-settablefunctions.TheCONFIGkeyisusedinconjunctionwiththe▼and ▲ functionsontheZEROandMAX/MINkeystoselectandchangedifferentfunctions.

Youcansetoneof20engineeringunits,changetheauto-shutofffunctionsetting,displaytheactualbatteryvoltage,displaytheactualtemperature(in°Cor°F),turnthedamp-eningonoroff,changethesamplerate,andsettheTarevalue.Inadditiontothe20availableengineeringunitsontheIPI,youcancreateyourownunittomeetyourmeasurementneed.Forexample,ifyouneedtotakealevelmeasurement,simplydeterminethepressurethatequatestooneunitofthatmeasurementandsetthatasyourengineeringunit.Theapplicationmayrequiregallons,feet,inches,meters,liters,orsimilar;withtheIPIyoucantakethemeasurement.

ENTERSaveselectionsandturnthebacklightonandoff.

External powerSometimesitismoreconvenienttotakeadvantageofanexistingpowersourceratherthanusingbatteries.TheIPImaybeorderedwitha24VDCpowersupplyoption.Theunithasterminalstoacceptthepositiveandnegativeleadstoallowforeasyconnectiontothepowersupply.Theseunitswillnotbesuppliedwithbatteriesandbatteriesshouldnotbeinstalledwhentheunitisconnectedto24VDCpower.

TheJOFRAIPIdigitalpressureindicatortakestheconceptofananalogtestgauge,andbringsittoanewlevel.TheIPIcombinestheaccuracyofdigitaltechnologywiththesimplicityofananaloggauge,andachievesperformance,ease-of-use,andafeaturesetunmatchedinthepressuremeasurementworld.SetupoftheIPIisfastandstraight-forward,throughamenu-drivendisplay,withminimaltext,andintuitivefunctions,thatissimpleenoughtoallowtheindicatortobeusedanywhereintheworld,withouttheneedformultilingualdisplays.

Protective bootStandardconfigurationIPIunitsandthosesuppliedwithsystemsaredeliveredwithaprotectiveboot.Unitswithbackmountingandthe24VDCpowersupplyarenotdeliv-eredwiththeprotectivebootduetodesignlimitations.

Serial communicationsTheIPIhasaserialportthatisaccessiblethoughthebackofthecase.ThisfeatureusesstandardASCIIcommandstoallowforextractionofdatafromtheIPIwhiletakingreadings.Thisrequirestheoptionalserialcable.

Page 3: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

Hazard location information / approvalsAnexplosiveatmosphereisdefinedasa”mixturewithair,underatmosphericconditions,offlammablesubstancesintheformofgases,vapours,mistsordustsinwhich,afterignitionhasoccurred,combustionspreadstotheentireunburnedmixture”.

ThestandardintheEuropeanUnionhasbeensetwiththe9/94/ECDirective,commonlycalledATEX(“AtmosphèresExplosibles,”Frenchforexplosiveatmospheres).

TheJOFRAIPIIndustrialPressureIndicatorisATEXapprovedbyKEMAascomplyingwiththeEssentialHealthandSafetyRequirementsrelatedtothedesignandcon-structionofequipmentandprotectivesystemsintendedforuseinpotentiallyexplosiveatmospheresgiveninAnnexIIinthedirective,andwiththefollowingrating:II3GEExnAIIBT6(Ta=–10°C...+55°C).

TheJOFRAIPIIndustrialPressureIndicatorisalsocertifiedbyCSAasconformingtorelevantCanadianandUSAstandardswiththefollowingrating:Class1,Div.2,GroupsA-D.

SeethedefinitionsregardinghazardouslocationsinNFPA70,Article500orCSAC22.1Section18.NFPA70,Article500andCSAC22.1Section18.

Field recalibrationTheIPIdoesnotneedtobereturnedtothefactoryforcalibration.Ifyouhaveareliableandaccuratepressurereferenceoralocallaboratory,youmayre-calibratetheIPIlocally.Thisfeatureispasswordprotected.

Sampling rateTheIPIsamplingrateisuserselectable.Ifyouwanttocapturefastsystemtransients,theIPIcantakeareading10timespersecond.Conversely,youmaywanttocon-servebatterylifeandonlyneedperiodicsamples.Then,youcanselect1sampleperevery2seconds.Thisworkswellforinprocessandpanelmountedapplications.Youcanalsochoosetheacceptedinstrumentsamplingrateof3samplespersecondfornormaluseortoallowforfilter-ing,youcanselect1samplepersecond.

Damping adjustmentThedampingfunctioncanbeturnedonoroffontheIPI.Thisallowsforreadingstobeintegrated,whichaccountsformomentarychangessuchasthosefrompulsingsources.

TareBeyondzeroingtheIPI,youmayhavetoaccountforresidualpressure.TheTarefeatureallowsyoutotakecareofthaterrorandpreventthemanualcalculationofthedif-ference.Thiscanbeusedincombinationwiththecustomengineeringunitstomakethelevelmeasurementseasier.

Temperature display and compensation BecausetheIPIisdesignedforin-processtasks,temperaturecompensationisincludedtomakethejobeasier.ThisallowstheIPItomaintainaccuracyoverthemeasurementrange.Themeasuredtemperaturemaybecheckedfromthekeypad.

REQUIREMENTS JOFRACALMinimum hardware requirements: • Intel®Pentium®lI1.4GHzprocessor.

• 64MBRAM(128MBrecommended)

• 80MBfreediskspaceonharddisk(120MBrecommended)priortoinstallation

• StandardVGA(800x600,256colours).1024x768recommended.

• CD-ROMdriveforinstallationofprogram

• 1or2freeRS-232serialports,dependingonconfiguration

Minimum software equirements:• MicrosoftWindows®98,MicrosoftWindows®NT4.0,

MicrosoftWindows®2000,MicrosoftWindows®ME,MicrosoftWindows®XP,Vista.

• Systemfonts:MSSansSerifandArial

JOFRACAL CALIBRATION SOFTWARE

JOFRACALcalibrationsoftwareensureseasycalibrationofRTD´s,thermocouples,transmitters,thermoswithes,pressuregaugesandpressureswitches.JOFRACALcanbeusedwithJOFRADPC-500,APC,CPCandIPIpres-surecalibrators,allJOFRAtemperaturecalibrators,aswellasJOFRAAMC910,ASC300multisignalcalibratorandASM-800signalmultiscanner.WhenusedwithJOFRAASM-800signalmultiscanner,JOFRACALcanperformasimultaneoussemiautomaticcalibrationonupto24pres-sureand/ortemperaturedevicesundertestinanycombi-nation.

JOFRACALsoftwarecontrolsthecompletecalibrationprocedure,storestheresultsandprovidesacalibrationaudittrailthroughhard-copycertificates.Allcalibrationdataarestoredforeachsensortomonitordriftandopti-miserecalibrationintervals.Aschedulerfeatureallowsplanningoffuturecalibrations.

Page 4: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

4 Industrial Pressure Calibrator • IPI

IPI

111 mm36 mm

38 mm

25,4 mm

99,3 mm

125 mm

Pressure; gauge / compound ranges

psi...........................................vacuumto30,100,300,500bar....................................................vacuumto2,7,21,35psi.................... 0to15,1,000,2,000,3,000,5,000,10,000bar........................................ 0to1,70,140,200,350,700

Pressure; absolute ranges

psi............................................................................0to100bar...............................................................................0to7

Engineering units

Userdefined................................... Oneuser-definableunitUserselectable........................................................ 20units

(PSI,Bar,kg/cm2,inH2O(4°C,20°Cor60°F),ftH2O(4°C,20°Cor60°F),mmH2O(4°Cand20°C),cmH2O(4°Cand20°C),mH2O(4°Cand20°C),KPa,mBAR,inHg,mmHg,Torr)

Notallunitsareavailibaleinallranges.

Pressure accuracy

Pressure........................................................... ±0.05%F.S.Fulltemperaturecompensation......0to50°C/32to122°FVacuum(100,300,500psiindicator/7,21,35barindicator)...................... ±0.25%F.S.Vacuum(30psiindicator/2barindicator).......... ±0.1%F.S.

Includingnon-linearity,hysteresis,repeatability,andtemperatureeffect.F.S.(fullscale)=maximumgaugepressureinpsi

Serial communication interface

Connector.......................................................... StereojackSerial......... 0-3VDC,9600baud,8data,noparity,1stopProtocol..................................... ASCllcommandlanguage

Media compatibility

Liquidsandgassescompatiblewith316stainlesssteel.

Environmetal

Storagetemperature....................-20to70°C/-4to158°FOperatingtemperature...............-10to55°C/-14to131°F

Pressure connection

Allranges.....................................................1/4”NPTmale

Adaptersto1/4”BSPmaleareincludedasstandard

Pressure overload

1to35bar/15to500psi……..............................3Xrange70to350bar/1,000to5,000psi........................2Xrange700bar/10,000psi…………….........................1.5XrangeOverloadalarm…………………………………OLondisplayOverloadalarmrange……………………………1.2Xrange

Display

Display.......................................................... Backlight,blueDisplayresolution.......................... 5digitfloatingdecimalBargraph........................................ 20segment,0to100%Displayindicators..............................Engineeringunitsicon...................................................Lowbatteryindicationicon.............................. Batterylifeindicationutilizingbargraph.............................................MeasuredtemperaturedisplayDisplayupdate......................................... twicepersecond

Power supply

Battery.......................................................... (3)AAAlkalineBatterylife......... 1500operationalhourswithoutbacklightBatterylife......... 2000operationalhourswithoutbacklight..............................................................atlowsamplingrateBatterylife.................150operationalhourswithbacklightLowbatteryindicator.............................................at3VDCOptionalinputport..................... 24VDCconnection1)2)3)

1)Unitsdeliveredwith24VDCpowersupplyconnectionarenotsuppliedwithbatteries

2)Batteriesshouldnotbeinstalledwhenusing24VDCpowersupplyasdamagetotheIPImayoccur

3)Unitswith24VDCpowersupplyconnectioncannotbeorderedwiththeprotectivebootoption

Instrument dimensions

IndicatorLxWxH..... 125x111x38mm/4.9x4.4x1.5inIndicatorweight(includingbattery)................1.6lb/0.7kgInputport..........................1/4”NPTMalelowerconnectionOptionalinputport........1/4”NPTMalebackconnection1)

3)Unitswithbackconnectioncannotbeorderedwiththeprotec-tivebootoption

Shipping dimensions

IndicatorLxWxH... 250x160x100mm/9.8x6.3x3.9inIndicatorweight..............................................1.8lb/0.8kgInstrument case

Rating..........................................................NEMA4/IP651)

1)Doesnotapplyiftheserialconnectionisinuseoriftheexternal24VDCpoweroptionisused.

Approvals - IPI system

CEConformity..............EN61326:1998,EN60079-0:2006,................................................................. EN60079-15:2005

TheJOFRAIPIsystemsaretypeapprovedbyDetNorskeVeritas.Findthecertificateatwww.jofra.comApproval,Certificateno..........................................A-10549

Ex approvals - IPI indicator only

CSA......................................Class1,Div.2,GroupsA-D1)

ATEX................ II3GEExnAIIBT6(Ta=–10°C...+55°C)1)

1)The24voltversionisnotCSAorATEXapproved

FUNCTIONAL SPECIFICATIONS

Page 5: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

JOFRA IPI PRESSURE RANGES

Resolution obtained by the IPI indicator

IPI015G0 to 15 psi0 to 1 bar

IPI01KG0 to 1,000 psi0 to 70 bar

IPI02KG0 to 2,000 psi0 to 140 bar

IPI03KG0 to 3,000 psi0 to 200 bar

IPI05KG0 to 5,000 psi0 to 350 bar

IPI10KG0 to 10,000 psi 0 to 700 bar

Imperial ranges

psi 15.000 1000.0 2000.0 3000.0 5000.0 10000

inH2O@4°C 415.21 27681 55361 83042 N/A N/A

inH2O@20°C 415.95 27730 55459 83189 N/A N/A

inH2O@60°F 415.62 27708 55416 83124 N/A N/A

ftH2O@4°C 34.601 2306.7 4613.5 6920.2 11534 23067

ftH2O@20°C 34.662 2310.8 4621.6 6932.4 11554 23108

ftH2O@60°C 34.635 2309.0 4618.0 6927.0 11545 23090

inHg@0°C 30.540 2036.0 4072.1 6108.1 10180 20360

Torr 775.73 51715 N/A N/A N/A N/A

Metric ranges

bar 1.0000 70.000 140.00 200.00 350.00 700.00

mbar 1000.0 70000 N/A N/A N/A N/A

kPa 100.00 7000.0 14000 20000 35000 70000

kg/cm2 1.0197 71.380 142.76 203.94 356.90 713.80

cmH2O@4°C 1019.7 71382 N/A N/A N/A N/A

cmH2O@20°C 1021.5 71508 N/A N/A N/A N/A

mH2O@4°C 10.197 713.82 1427.6 2039.5 3569.1 7138.2

mH2O@20°C 10.215 715.08 1430.2 2043.1 3575.4 7150.8

mmHg@0°C 750.06 52504 N/A N/A N/A N/A

Resolution obtained by the IPI indicator

IPI30CVacuum to 30 psiVacuum to 2 bar

IPI100A / IPI100CVacuum to 100 psiVacuum to 7 bar

IPI300CVacuum to 300 psiVacuum to 21 bar

IPI500CVacuum to 500 psiVacuum to 35 bar

Imperial ranges

psi -12.000 30.000 -12.000 100.00 -12.000 300.00 -12.00 500.00

inH2O@4°C -332.17 830.42 -332.2 2768.1 -332.2 8304.2 -332 13840

inH2O@20°C -332.76 831.89 -332.7 2773.0 -332.7 8318.9 -333 13865

inH2O@60°F -332.50 831.24 -332.5 2770.8 -332.5 8312.4 -332 13854

ftH2O@4°C -27.681 69.202 -27.68 230.67 -27.68 692.02 -27.7 1153.4

ftH2O@20°C -27.730 69.324 -27.73 231.08 -27.73 693.24 -27.7 1155.4

ftH2O@60°C -27.708 69.270 -27.71 230.90 -27.71 692.70 -27.7 1154.5

inHg@0°C -24.432 61.081 -24.43 203.60 -24.43 610.81 -24.4 1018.0

Torr -620.6 1551.5 -620.6 5171.5 -620 15514 -621 25858

Metric ranges

bar -0.8300 2.0000 -0.8300 7.0000 -0.8300 21.000 -0.8300 35.000

mbar -830.0 2000.0 -830.0 7000.0 -830.00 21000 -830 35000

kPa -83.00 200.00 -83.00 700.00 -83.0 2100.0 -83.0 3500.0

kg/cm2 -0.8464 2.0394 -0.8464 7.1380 -0.846 21.414 -0.846 35.690

cmH2O@4°C -846.4 2039.5 -846.4 7138.2 -846 21415 -846 35691

cmH2O@20°C -847.9 2043.1 -847.9 7150.8 -847 21452 -848 35754

mH2O@4°C -8.464 20.395 -8.464 71.382 -8.46 214.15 -8.46 356.91

mH2O@20°C -8.479 20.431 -8.479 71.508 -8.48 214.52 -8.48 357.54

mmHg@0°C -622.6 1500.1 -622.6 5250.4 -622 15751 -623 26252

ThistableshowstheresolutionsthatcanbeobtainedbytheIPIindicatorsthroughoutallengineeringunits.

Page 6: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

6 Industrial Pressure Calibrator • IPI

AccessoriesPart No. Description75P014 ServicekitforT-960pump75P016 ServicekitforT-970pumpT-733-2 PressurehoseforT-960/970,0.6meterstd.T-733-3 PressurehoseforT-960/970,0.5meterT-733-4 PressurehoseforT-960/970,1.0meterT-733-5 PressurehoseforT-960/970,2.0meterT-733-6 PressurehoseforT-960/970,5.0meter65R191 SetofBSPfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x3/8”,1x1/2”65R192 SetofNPTfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x1/4”,1x3/8”1x1/2”60R191 1setofpackingandsealsfor65R191and65R192123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACSPK-HPC-005 QuickconnectorsetforIPIsystemA/B

Ordering information System AOrder no. Description Pressure indicator IPI IndustrialPressureIndicator

Indicator pressure range 015G 0to1bar(15psi) 100A 0to7barabsolut(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number AXX IPISystemA(pumpT-960) AHX IPISystemA(pumpT-970)

Certification G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI030CBXXAXXG Sample order numberJOFRAIPI2bar(30psi)indicatorwithbootandinpres-sureSystemAwithpneumaticpressurehandpumpandNISTtraceablecalibrationcertificate.

• 0to1bar(15psi)• Vacuumto2bar(30psi)• Vacuumto7bar(100psi)• Vacuumto21bar(300psi)

ThissystemincludestheJOFRAIPIcalibratortogetherwithoneofthefollowingpneu-matichandpumps:T-960andT-970,dependingonpressurerange.SystemAisaneasy-to-usesinglehandoperatedcalibrationsystem.Noneedforaflatsurfacetooperatethesystem-justholditinyourhand.

TheSystemAcomesinanaluminiumcarryingcasewithcutoutsforfittings,hose,andthecompleteassembledcali-brationunit-notimeisrequiredtoassembletheuniteverytimeyouneedit.Thecalibrator(indicator)canatanytimebeusedseparatelyforotherpressuretesttasks.Aspecialquickconnectorbetweenthepumpandtheindi-catormakesitpossibletoseperatethesysteminsecondsandtoswiveltheindicatorforeasyviewing.

TheJOFRAIPIdeliveryalwaysincludesaprotectivebootfortheindicator.AllIPIindicatorsusedinSystemAaredeliveredcalibratedinbothpressureandvacuum.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

JOFRA IPI System A

T-960 and T-970 pumps for system AModelT-960pneumaticpressurepumpfeaturesanextendedrangevernieradjustmentforpreciselycontrollingpumppressureandcomfortablepistolgriphandle.

ModelT-970pneumaticpressurepumpissimilartotheT-960,butgeneratepressureupto40bar(580psi).

Pressurerange,T-960.....................0to2bar/0to30psiPressurerange,T-970..................0to40bar/0to580psiType..................................................................... PheumaticTestmedium.....................................................................AirOperation.................................................................. Scissor“O”-rings..................................................................Buna-NWettedparts......................Aluminium,brass,stainlesssteel,...................................................................... nylon,NylatronGSConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize.........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb

Ordering Part No. DescriptionT-960 Pump:0to2bar(30psi)T-970 Pump:0to40bar(580psi)

Included with System A delivery• Standarddelivery(seepage12)• HandpumpT-960orT-970• 3/8”BSPfemaleadapterwith1/4”BSPandNPT

femaleterminations

Shipping InformationSize...................570x180x350mm/22x7,1x13,8inchWeight........................................................5,5kg/12.13lb

• 0.6m/24inhosewith1/4”BSPandNPTfemaleterminations

• 1rollofTeflontape• Protectivecarryingcase• QuickConnectorbetweenpumpandcalibrator

Page 7: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

AccessoriesPart No. Description75P017 ServicekitforT-975pumpT-733-2 PressurehoseforT-975,0.6meterstd.T-733-3 PressurehoseforT-975,0.5meterT-733-4 PressurehoseforT-975,1.0meterT-733-5 PressurehoseforT-975,2.0meterT-733-6 PressurehoseforT-975,5.0meter65R191 SetofBSPfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x3/8”,1x1/2”65R192 SetofNPTfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x1/4”,1x3/8”1x1/2”60R191 1setofpackingandsealsfor65R191and65R192123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACSPK-HPC-005 QuickconnectorsetforIPIsystemA/B

Included with System B delivery• Standarddelivery(seepage12)• HandpumpT-975forallranges• 3/8”BSPfemaleadapterwith1/4”BSPandNPT

femaleterminations• 0.6m/24inhosewith1/4”BSPandNPTfemale

terminations• 1rollofTeflontape• Protectivecarryingcase• QuickConnectorbetweenpumpandcalibrator

Ordering information System BOrder No. Description Pressure indicator IPI IndustrialPressureIndicator

Indicator pressure range 100A 0to7barabsolute(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) 500C Vacuumto35bar(500psi) NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number BXX IPISystemB(pumpT-975)

Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI300CBXXBXXG Sample order number JOFRAIPI21bar(300psi)indicationwithbootandinpressureSystemBkitwithpneumaticcombinationhand-pumpandNISTtraceablecalibrationcertificate.

T-975 hand pump for system BModelT-975pneumaticpressurepumpfeaturesbothvacuumandpressuregeneration.Abutton(valve)makesiteasytoswitchfromvacuumtopressuremeasurements.Built-inreleasevalve,verniervalveforfineadjustmentanddualpressureoutputenablesafeandsimpleoperation.Theunitoffersbothmetricandimperialthreadsontheref-erenceconnectionandthehose.

Pressurerange...........-0.91to40bar/-27inHgto580psiType..................................................................... PheumaticTestmedium.....................................................................AirOperation.................................................................. Scissor“O”-rings..................................................................Buna-NWettedparts.......Anodizedaluminium,nickel-platedbrass,............................................................ stainlesssteel,NylonConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize.........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb

OrderingPart No. DescriptionT-975 Pump:-0.91to40bar/-27inHgto580psi

• Vacuumto2bar(30psi)• Vacuumto7bar(100psi)• Vacuumto21bar(300psi)• Vacuumto35bar(500psi)

Thissystemincludesthedualfunctionpneumatichandpump.Withthispumpyoucancali-bratebothvacuumandpres-sureapplications.Ittakesjustafewsecondstoswitchbetweenvacuumandpressure,justpushonebutton(valve)ontheT-975handpump.

TheSystemBcomesinanaluminiumcarryingcasewithcutoutsforfittings,hoseandthecompleteassembledcali-brationunit-notimerequiredtoassembletheuniteverytime.

Theindicatorandthepumparefittedwithaspecialquickconnectorthatmakesitpossibletoseparatethesysteminsecondswithoutanytool.Thisquickconnectoralsomakesitpossibletousetheassembledsystemwiththeindicatortwistedtoanyangle.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

JOFRA IPI System B

Shipping InformationSize...................570x180x350mm/22x7,1x13,8inchWeight........................................................5,5kg/12.13lb

Page 8: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

8 Industrial Pressure Calibrator • IPI

AccessoriesPart No. DescriptionT-656 ServicekitforT-620andT-620HpumpsT-649 PressurehoseforT-620/T-620Hpumps65P175 1mhosewith1/4”BSPfemaletermination Max.pressure350bar(5,000psi)65P180 2mhosewith1/4”BSPfemaletermination Max.pressure350bar(5,000psi)123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACMGAAA/GL OilforT-620&T-1pumpAAAOILin 1-GALLONCANMGAAA/QT OilforT-620&T-1pumpAAAOILin1- QUARTCAN

Shipping InformationSize...................560x188x350mm/22x7,4x13,8inchWeight........................................................5,5kg/12.13lb

• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)

ThissystemconsistsofanIPIindicatortogetherwithahydraulic,highpressurehandpumpT-620orT-620H,whichfeaturesanoilreservoirtoprimethesystem.SystemCisaneasy-to-usesingle-handoperatedtestsystem.Noneedforaflatsurface(table)tooperatethesystem.Thesystemincludesreleasevalveandverniervalveforfineadjustmenttoenablesafeandsimpleoperationofthepump.

TheSystemCcomesinaprotectivecarryingcasewithcutoutsforfittings,hose,andthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressureindicatorseparatelyforotherpressuretesttasks.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

Included with System C delivery• Standarddelivery(seepage12)• HandpumpT-620orT-620H• 0.6m(24in)hosewith1/4”NPTfemaleand1/4”BSP

terminations• 1rollofTeflontape• Protectivecarryingcase

AMETEK hydraulic handpumpThiseasy-to-usehandpumpisspeciallydesignedforhighpressureapplications.Thepumphasabuilt-inreservoir,ventvalve,andverniervalveforfineadjustment.Thepumphasadualpressureoutput-oneforthetestobjectandonefortheIPIindicator.

Pressurerange,T-620.................... 0to200bar/3,000psiPressurerange,T-620H................. 0to350bar/5,000psiType....................................................................... HydraulicTestmedium.....................................................................OilReservoircapacity.......................................200cl/0.5pintOperation..............................................................Pistol-grip“O”-rings..................................................................Buna-NWettedparts..........Aluminum,brass,stainlesssteel,LexanConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb

OrderingPart No. DescriptionT-620 Pump:0to200bar(3,000psi)T-620H Pump:0to350bar(5,000psi)

Ordering information IPI System COrder No. Description Pressure indicatorIPI IndustrialPressureIndicator

Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar3,000psi) 05KG 0to350bar5,000psi) NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number CXX IPISystemC(pumpT-620/T-620H)

Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI01KGBXXCXXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootandinpressureSystemCwithhydraulicpressurehandpumpandNISTtraceablecalibrationcertificate.

JOFRA IPI System C

Page 9: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

AccessoriesPart No. Description60P013 Verniervalvefineadjustmentfor65-P016/01760P016 Packingsetforscrewpumpoilpartno.65-P01660P017 Packingsetforscrewpumpwaterpartno.65-P01765-R975 Reservoirsetwithseals,valveandbondedseal60R135 Valveforoilreservoir65P175 1m.Pressurehose,(Max.350bar/5,000psi) *1/4”BSPfemaleto1/4”BSPmale65P180 2m.Pressurehose,(Max.350bar/5,000psi) *1/4”BSPfemaleto1/4”BSPmale60I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale65N000 Extensiontube,120mm(Max.350bar/5.000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VAC50-REP615 Quickconnectorset,female1/4”BSP to1/4”BSPSPK-HPC-004BanjoConnectorSetforpump

Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight...........................................................12kg/26.5lb

• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)

ThissystemconsistsofanIPIindicatortogetherwitharug-ged,hydraulic,highpressurescrewpump.SystemDisaneasy-to-usetestsystem.Placethepumponaflatsurfaceandturnthehandletogeneratepressure.Youcanadjusttheindicatortofitthebestviewingangle.Thesystemincludesamanifoldforconnectionofthetestdevice,afineadjustmentverniervalve(optional),andareservoirforextraoilorwater(optional).

TheSystemDcomesinaprotectivecarryingcasewithcutoutsforfittingsandthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

AMETEK screwpumpThe65-P016and65-P017screwpumpsaredesignedforveryeasygenerationofhighpressure.Thescrew-pumpshasa4connectionmanifold.Fineadjustment,verniervalve,liquidreservoir,andfittingsareoptional.TheAMETEKscrewpumpmaybedeliveredfor2differenttestmedia:hydraulicoilorwater.

Pressurerange............................... 350bar/0to5,000psiType....................................................................... HydraulicTestmedia............................................. Hydraulicoil,waterOperation..........................................................Screwpump“O”-rings..................................................................Buna-NWettedparts..................... Aluminum,brass,stainlesssteelConnectionstotestobjects................ 4x1/4”BSPfemaleSize....................36.6x19.5x10.5cm/14.4x7.7x4.1inWeight............................................................3.7kg/8.2lb

OrderingPart No. Description65-P016 Screwpumpoil:0to350bar(5000psi)1)65-P017 Screwpumpwater:0to350bar(5000psi)1)

1)25ml(0.85oz)reservoir

Included with System D delivery• Standarddelivery(seepage12)• Screwpump65-P016forhydraulicoilor65-P017for

water• 1rollofTeflontape• 10bondedseals• 6mmAllenkey• Protectivecarryingcase• BanjoConnectorSet

Ordering information IPI System DOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator

Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number DOX IPISystemD(pump65-P016) DWX IPISystemD(pump65-P017)

Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI01KGBXXDOXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootandinpressureSystemDwithhydraulicoilscrewpumpandNISTtraceablecalibrationcertificate.

JOFRA IPI System D

Page 10: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

10 Industrial Pressure Calibrator • IPI

AMETEK high pressure hydraulic oil pump The65-P014hydraulicpumpisdesignedforhighpres-sureapplications.Thepumphasa4connectionoutputmanifold.Fineadjustmentverniervalveandfittingsareoptional.

Pressurerange............................. 0to700bar/10,000psiType....................................................................... HydraulicTestmedia........................................................................OilOperation............................................................ Jackpump“O”-rings..................................................................Buna-NWettedparts..................... Aluminum,brass,stainlesssteelConnectionstotestobjects................ 4x1/4”BSPfemaleSize....................36.5x20.0x14.2cm/14.4x7.9x5,6inWeight........................................................ 5.3kg/11.6lbs

Ordering Part No. Description65-P014 Hydraulicpumpforoil:0to700bar(10,000psi)1)

1)450ml(15.22oz)reservoir

• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)• 0to700bar(10,000psi)

ThissystemconsistsofanIPIindicatortogetherwitharug-ged,hydraulic,highpressurepump.

SystemEisaneasy-to-usetestsystem.Thehydraulicpumpmakesitveryeasytoprimethesystem.Youcanadjusttheindicatortofitthebestviewingangle.Thesys-temincludesamanifoldforconnectionoftwotestdevicesandthefineadjustmentverniervalveisstandard.

TheSystemEcomesinaprotectivecarryingcasewithcutoutsforfittingsandthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

AccessoriesPart No. Description60P013 Verniervalveforfineadjustment60P014 PackingsetforJackpumppartno.65-P01460I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VAC50-REP615 Quickconnectorset,female1/4”BSP to1/4”BSP

Included with System E delivery• Standarddelivery(seepage12)• Hydraulicpump65-P014forhydraulicoil• Verniervalve• 1rollofTeflontape• 10bondedseals• 6mmAllenkey• Protectivecarryingcase

Ordering information IPI System EOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator

Indicator pressure range01KG 0to70bar(1,000psi)02KG 0to140bar(2,000psi)03KG 0to200bar(3,000psi)05KG 0to350bar(5,000psi)10KG 0to700bar(10,000psi)NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number EXX IPISystemE(pump65-P014)

Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI01KGBXXEXXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootaninpressuresystemSystemEwithhighpressurepumpforhydraulicoilandNISTtraceablecalibrationcertificate.

JOFRA IPI System E

Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight...........................................................13,5kg/30lb

Page 11: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

Included with System F delivery• Standarddelivery(seepage12)• TypeTHydraulicpump• 1rollofTeflontape• Protectivecarryingcase

• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)• 0to700bar(10,000psi)

TheSystemFhydraulicpumpmakesitveryeasytoprimethesystem.Itcontains75in3/1.23lofliquid.Thesystemincludesamanifoldforcon-nectionoftwotestdevicesandthefineadjustmentverniervalveisstandard.TheSystemFcomesinaprotectivecar-ryingcasewithcutoutsforfittings.Easyandfastconnec-tionbetweenpumpandcalibratormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.

OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.

AccessoriesPart No. DescriptionT-250 ServicekitforTpumpwithBunaNO-ringsT-250 ServicekitforTpumpwithVITONNO-ringsT-250 ServicekitforTpumpwithEPTNO-rings60I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACMGAAA/GL OilforT-620&T-1pumpAAAOILin 1-GALLONCANMGAAA/QT OilforT-620&T-1pumpAAAOILin1- QUARTCAN

Ordering information IPI System FOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator

Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) 10KG 0to700bar(10,000psi) NONE Pumpsystemonly(nocalibrator)

Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)

Power supply X Batteries(standard)

Mounting X Lowermount(standard)

Base model number FWB IPISystemF(pumpT-1) FWV IPISystemF(pumpT-1/VITON) FWE IPISystemF(pumpT-1/EPT) FOB IPISystemF(pumpT-1/OIL) FOV IPISystemF(pumpT-1/OIL/VITON) FOE IPISystemF(pumpT-1/OIL/EPT)

Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)

IPI10KGBXXFOEG Sample order number JOFRAIPI700bar(10,000psi)indicatorwithbootandinpressureSystemFwithTypeTpumpforoil,EPTsealsandNISTtraceablecalibrationcertificate.

JOFRA IPI System F

1)Customermustverifycompatibilityofsealsandpumpmaterialswithpressuremedium.Vitonisaregisteredtrademark.2)BunaNarethestandard”O”ringmaterialforthehydraulictesters.Theyareusedwithwater/alcoholornoncorrosiveoils.3)Viton”O”ringmaterialisusedforallcorrosiveoilsotherthanSkydrol.4)EPT”O”ringmaterialisspecifi-callyforSkydrolOilusedforhydraulicsystemsinairplanes,itisverycorrosive.

AMETEK Type T hydraulic oil or water pumpThetypeTpumpisdesignedforhighpressureapplica-tionsupto15,000psi(1,000bar).Thepumpandsystemmaybeorderedwitheitheroilorawater/alcoholmixtureasthepressuremedium.Thepumpisconstructedof300seriesstainlesssteelandMonelallowingfortheuseofotherhydraulicmedia.Therearethreeavailablesealpackagesforthesystem:Buna-N,Viton®,andEPT.1)Thepumpfeaturesadualpressureoutputmanifold,fineadjust-mentverniervalve,reliefvalve,anddualvolumecontrolforrapidpressureincreaseatlowerpressuresandeasierpumpingathigherpressures.

Pressurerange.......................... 0to1,000bar/15,000psiType....................................................................... HydraulicTestmedia............................................. Hydraulicoil,waterReservoircapacity...........................................1.23l/75in3

Operation............................................................ Jackpump“O”-rings......... Buna-N(standard)orEPT/Viton(optional)Wettedpart.......................................Stainlesssteel,MonelConnectionstotestobjects................... 1/4”and1/2”BSP............ terminations;1/4”and1/2”PTfemaleterminationsSize................................ 79.4x22.9x50.8cm/31.3x9x20inWeight........................................................... 9.1kg/20lbs

OrderingPart No. Description Seals T-1 0to1,000bar(15,000psi)water-BunaN2)

T-1/VITON 0to1,000bar(15,000psi)water-Viton3)

T-1/EPT 0to1,000bar(15,000psi)water-EPT4)

T-1/OIL 0to1,000bar(15,000psi)oil-BunaN2)

T-1/OIL/VITON 0to1,000bar(15,000psi)oil-Viton3)

T-1/OIL/EPT 0to1,000bar(15,000psi)oil-EPT4)

Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight..............................................................18kg/40lb

Page 12: Ametek - Jofra Industrial Pressure Calibrator IPI Indicator · Use the JOFRA IPI in applications from safety valve checks to system ... or testing needs. ... JOFRACAL software controls

www.jofra.comInformation in this document is subject to change without notice. ©2009, by AMETEK, Inc., www.ametek.com. All rights reserved.

Pub code SS-IPI Issue 0912

AMETEK Calibration Instrumentsis one of the world’s leading manufacturers and developers of calibration instruments for temperature, pressure and process signals as well as for temperature sensors both from a commercial and a technological point of view.

JOFRA Temperature InstrumentsPortable precision thermometers. Dry-block and liquid bath calibrators: 5 series, with more than 25 models and temperature ranges from -90° to 1205°C / -130° to 2200°F. All featuring speed, portability, accuracy and advanced documenting functions with JOFRACAL calibration software.

JOFRA Pressure InstrumentsConvenient electronic systems ranging from -25 mbar to 1000 bar (0.4 to 15,000 psi) - multiple choices of pressure ranges, pumps and accuracies, fully temperature-compensated for problem-free and accurate field use.

JOFRA Signal InstrumentsProcess signal measurement and simulation for easy control loop calibration and measurement tasks - from handheld field instruments to laboratory reference level bench top instruments.

JOFRA / JF Marine InstrumentsA complete range of calibration equipment for temperature, pressure and signal, approved for marine use.

FP Temperature SensorsA complete range of temperature sensors for industrial and marine use.

M&G Pressure TestersPneumatic floating-ball or hydraulic piston dead weight testers with accuracies to 0.015% of reading. M&G PumpsPressure generators from small pneumatic “bicycle” style pumps to hydraulic pumps generating up to 1,000 bar (15,000 psi).

...because calibration is a matter of confidence

Order No. Description Pressure indicatorIPI IndustrialPressureIndicator

Indicator pressure range 100A 0to7batabsolute(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) 500C Vacuumto35bar(500psi) 015G 0to1bar(15psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) 10KG 0to700bar10,000psi()

Boot B Withboot(standardforindicatorwithbatterypower andlowermounting-andonallsystemkits) X Noboot(standardforindicatorswithbackmounting and/or24VDCpowerssupplyconnection)

Power supply X Batteries(standard) E Connectionforexternal24VDCpowersupply1)

Mounting X Lowermount(standard) B Backmount

Base model number IND IPIindicatorwithstandardaccessoriesforconfiguration

Certification G NISTtraceablecertificate(standard) H Accreditedcertificate(optional–onquotationbasis)

IPI030CBXXINDG Sample order numberJOFRAIPI2bar(30psi)indicatorwithboot,standardaccessories,andNISTtraceablecalibrationcertificate.

ORDERING INFORMATION - JOFRA IPI INDICATOR

• IPIindicator

• CalibrationcertificationperformancetraceabletoNIST

• (3)AAbatteries

• Adapterto1/4”BSPmale

• Usermanual

• Protectiveboot

STANDARD DELIVERY

1)The24voltversionisnotCSAorATEXapproved

Kyle
Rectangle