alkaline glucose electrooxidation Synthesis and ... › suppdata › c5 › ra › c5ra06331k ›...
Transcript of alkaline glucose electrooxidation Synthesis and ... › suppdata › c5 › ra › c5ra06331k ›...
`
1
Electronic Supplementary Information
Synthesis and characterization of Pd-Ni core-shell nanocatalysts for
alkaline glucose electrooxidation
Cheng-Chuan Chen, and Lin-Chi Chen*
Department of Bio-Industrial Mechatronics Engineering,
National Taiwan University, No. 1, Sec. 4, Roosevelt Road, Taipei 10617, Taiwan*Corresponding author: E-mail: [email protected] / Fax: +886-2-2362-7620
(1) Pt-based and Pd-based bimetallic catalysts catalyze GORs
Table S1. GOR performances of carbon supported bimetallic catalysts (literature vs. this study).
Anodic peak for forward scan Anode Catalyst Onset potential
(V vs. Hg/HgO) Peak potential(V vs. Hg/HgO)
Peak current (mA cm-2)
Reference
Pd-Au/C -0.45 -0.15 2.5 [4]
Pt-Au/C -0.59 -0.22 9.5 [7]
Pt-Pb/C -0.56 -0.52 2.5 [9]
Pd-Rh/C -0.55 0.05 3.5 [13]
Pd-Ni/C -0.56 -0.03 34.2 This work
(2) SEM-EDS characterization of the Pd-Ni/C catalysts
In order to determine whether Ni elements exist in the Pd-Ni/C catalysts, the atomic
characterization was performed by the SEM-EDS measurement, as shown in Fig. S1.
Electronic Supplementary Material (ESI) for RSC Advances.This journal is © The Royal Society of Chemistry 2015
`
2
Figure S1. SEM-EDX micrographs of the Pd-Ni/C catalysts (A) Pd/C, (B) Pd-Ni/C (1:0.02), (C)
Pd-Ni/C (1:0.03), (D) Pd-Ni/C (1:0.06), (E) Pd-Ni/C (1:0.14), (F) Pd-Ni/C (1:0.33) and (G) Ni/C.
(3) TEM characterization of the Pd-Ni/C catalysts
The size and distribution of the PdshellNicore nanoparticles were analyzed by TEM characterization,
as shown in Fig. S2. The frequency of the metal particle size is estimated using a statistical analysis
of about 50 particles which were randomly selected from the TEM data. It can be seen that the
`
3
average particle size of Pd/C is 4.7 nm and the PdcoreNishell nanoparticles with an average diameter
of 4.6, 4.2, 4.1, 5.4 and 6.6 nm for Pd-Ni/C (1:0.02), Pd-Ni/C (1:0.03), Pd-Ni/C (1:0.06), Pd-Ni/C
(1:0.14) and Pd-Ni/C (1:0.33), respectively (Fig. S2A-F).
`
4
1 2 3 4 5 6 7 80
5
10
15
(A')Fr
eque
ncy
/ %
Particle diameter / nm
4.7 0.25 nm
1 2 3 4 5 6 7 80
5
10
15
(B')
Freq
uenc
y / %
Particle diameter / nm
4.6 0.29 nm
2 3 4 5 6 7 80
5
10
15
20
(C')
Freq
uenc
y / %
Particle diameter / nm
4.2 0.26 nm
1 2 3 4 5 6 7 8 9 100
5
10
15
20
(D')
Freq
uenc
y / %
Particle diameter / nm
4.1 0.21 nm
2 3 4 5 6 7 8 9 10 110
5
10
15(E')
Freq
uenc
y / %
Particle diameter / nm
5.4 0.35 nm
1 2 3 4 5 6 7 8 9 10 11 12 13 14 150
5
10
15(F')
Freq
uenc
y / %
Particle diameter / nm
6.6 0.7 nm
Figure S2. TEM images of the Pd-Ni/C catalysts (A) Pd/C, (B) Pd-Ni/C (1:0.02), (C) Pd-Ni/C
(1:0.03), (D) Pd-Ni/C (1:0.06), (E) Pd-Ni/C (1:0.14), (F) Pd-Ni/C (1:0.33) and (G) Ni/C and their
corresponding particle size distribution (A, B, C, D, Eand F).
`
5
(4) ECSA measurement of the Pd-Ni/C catalysts
The ECSA value (a measure of electrochemical active surface area) was determined by the
voltammetric curves. During the forward scan of the Pd/C catalyst, the oxidation peak between -0.7
V and -0.5 V is ascribed to the absorption of hydrogen (Fig. S3A). In addition, the peak that occurs
in the range between +0.0 V and +0.4 V is attributed to the formation of Pd-OH. In the backward
scan, a sharp cathodic peak at -0.35 V indicates the reduction of Pd oxide. In Fig. S3B, the results
show that the cathodic reduction peak current densities gradually increase to a maximum value as
12 mA cm-2 on Pd-Ni/C (1:0.06) and then decline while the weight loadings of Ni increase. To
estimate the palladium electrochemical active surface area (ECSA) CV profiles were recorded
between-0.8 and 0.8 V vs. Hg/HgO with a scan rate of 50 mV s-1. Cyclic voltammograms (CVs)
were recorded in nitrogen saturated evaluated 0.5 mol L-1 NaOH electrolyte through bubbling with
nitrogen under nitrogen atmosphere at room temperature. The palladium ECSA was calculated
using the mean integral coulombic charge (Q) of the reduction area between -0.4 V and 0.1 V and
the reduction of palladium oxide monolayer is assumed as 405 μC cm-2 according to the following
equation (1):
mgcmC
μCgm
loading
H2
Pd40510Q Pdof ECSA
2
(1)
`
6
-0.8 -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8-10
-5
0
5
10
15 Pd/C
(A)
Curre
nt d
ensit
y / m
Acm
-2
Potential / V vs. Hg/HgO
(= 50 mV s-1)0.5 moL-1 NaOH
-0.8 -0.6 -0.4 -0.2 0.0 0.2-15
-10
-5
0
5
10(B)
(= 50 mV s-1)
(a) Pd/C(b) Pd-Ni/C (1:0.02)(c) Pd-Ni/C (1:0.03)(d) Pd-Ni/C (1:0.06)(e) Pd-Ni/C (1:0.14)( f ) Pd-Ni/C (1:0.33)(g) Ni/C
0.5 mol L-1 NaOH
Curre
nt d
ensit
y / m
Acm
-2
Potential / V vs. Hg/HgO
(a)
(b)(c)
(d)(e)
( f )
(g)
Figure S3. Cyclic voltammograms of (A) Pd/C and (B) the backward scan for the Pd-Ni/C
catalysts: (a) Pd/C, (b) Pd-Ni/C (1:0.02), (c) Pd-Ni/C (1:0.03), (d) Pd-Ni/C (1:0.06), (e) Pd-Ni/C
(1:0.14), (f) Pd-Ni/C (1:0.33) and (g) Ni/C in 0.5 mol L-1 NaOH solution at a scan rate of 50 mV s-1
at room temperature.
`
7
(5) Activation energy measurement of Pd-Ni/C (1:0.06) and Pd/C
The corresponding activation energy is determined using Arrhenius equation. Where I means the
corresponding current density at a specific potential, R is the universal gas constant (8.314 J K-1
mol-1), T is the absolute temperature in K and Ea is the activation energy value. The result shows
that the current densities of Pd/C and Pd-Ni/C (1:0.06) at an operating temperature of 318 K are
34.2 and 36.9 mA cm-2, respectively, as shown in Fig. S4.
/RTEaAeI
3.1 3.2 3.3 3.41.0
1.2
1.4
1.6
1.8
2.0
16.9 kJ mol-1
23.1 kJ mol-1
Pd/C Pd-Ni/C (1:0.06)
Log
I / m
Acm
-2
103 T-1 / K-1
I=Ae-Ea/RT
Figure S4. Arrhenius plots at different temperature (T = 298, 308 and 318 K) of the Pd/C and Pd-
Ni/C (1:0.06) in 0.5 mol L-1 NaOH solution containing 0.5 mol L-1 glucose at a scan rate of 50 mV
s-1.
`
8
(6) Long-term durability measurement of Pd-Ni/C (1:0.06) and Pd/C
For 500 continuous GOR cycles, the Pd/C and Pd-Ni/C (1:0.06) were studied by CV (100 mV s-
1).The retained activities is calculated using the ratio of the GOR peak current density at cycle number n (In) to the one at the first cycle (Io). The result shows a less activity loss of electrocatalytic activity for Pd-Ni/C (1:0.06) compared to Pd/C during the successive GOR duty cycles.
0 100 200 300 400 500
0
20
40
60
80
100
Pd/C Pd-Ni/C (1:0.06)
Activ
ity (I
fn /I f) /
%
Cycle number / n
Figure S5. GOR cycling stability test for the Pd/C and Pd-Ni/C (1:0.06) in 0.5 mol L-1 NaOH solution containing 0.5 mol L-1 glucose at a CV scan rate of 100 mV s-1 at room temperature.