AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik 1 , Rajni Natesan 1,2 , Sanghyeb Lee 1 , and James Wiskin 1 1 QT Ultrasound Labs, Novato, CA 2 MD Anderson Cancer Center, Houston, TX

Transcript of AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast...

Page 1: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

AI‐based quantitative breast density assessment using transmission ultrasound

Bilal Malik1, Rajni Natesan1,2, Sanghyeb Lee1, and James Wiskin11QT Ultrasound Labs, Novato, CA2MD Anderson Cancer Center, Houston, TX

Page 2: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Disclosures

Principal Scientist, QT Ultrasound LLC

Grant funding from National Institutes of Health (NIH) 

Page 3: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Purpose

Growing body of evidence indicates that breast density is one of the most important independent risk factors of breast cancer

Currently, mammography is the only FDA‐cleared means to evaluate breast density in a general screening population.

We present 3D transmission ultrasound as a method to visualize and differentiate fibroglandular tissue within the breast and use a fully automated segmentation method machine learning‐based method to quantitatively assess the breast density

Page 4: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

QT Scanner – transmission and reflection ultrasound

Page 5: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

QT speed of sound and reflection images

Page 6: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

3D image volume of speed of sound and reflection

Speed of Sound Reflection

Page 7: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Transmission & Reflection: normal breast anatomy

Transmission >

Reflection >

Page 8: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Tissue segmentation algorithm

Segment breast from surrounding water using 

attenuation images

Determine ‘border’ pixels based on proximity

Calculate breast density

Segmentation of high‐speed breast tissue from the total breast volume

• Fuzzy clustering into two categories

• Membership map generation

• Thresholding

Page 9: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Testing on tissue phantoms

Density based on theoretical volume = 7.1%  Density based on QBD = 7.6% 

Page 10: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Testing on clinical images

• Application of algorithm on 100 unilateral breast scans

• Mammography performed within 90 days of transmission imaging

• Both QBD and VolparaDensityTM(v3.1) scores were available.

• Correlation quantified using Spearman coefficient

Page 11: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Segmentation of fibroglandular tissue

QBD= 10.9%  QBD= 29.5%  QBD= 62.4% Wiskin et al., Medical Physics, 2019, in press

Page 12: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Correlation of QBD with VolparaDensity

• Spearman r = 0.94 (95% CI: 0.91‐0.96); p<0.0001)

• Deming linear regression shows a relationship of VolparaDensity = 0.53(QBD) ‐ 0.87

Page 13: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

QBD relationship with VolparaDensity similar to MRI

MRI %FGV

Volp

ara

%FG

V

Ref: Wang et al., PLoS One, 8(12), 2013

Page 14: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Validation of segmentation algorithm using large format histology

TU fibroglandular volume = 45.1 %  H&E fibroglandular volume = 42.3 %

Page 15: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

QT speed of sound image – QBD= 34.7% MUSE image – equivalent breast density = 37.9%

Validation of segmentation algorithm using UV microscopy

Page 16: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Precision of QBD measurement

• Scanned a single breast/patient ten times• Calculated QBD for individual scans• Mean QBD value = 9.4 %; Standard deviation = 0.2 %

Page 17: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Volumetric rendering of segmented breast tissue

Page 18: AI‐based quantitative breast density assessment using ...€¦ · AI‐based quantitative breast density assessment using transmission ultrasound Bilal Malik1, Rajni Natesan1,2,

Conclusions

The presented segmentation method can accurately identify the fibroglandular tissue volume within the whole breast.

The results indicate that breast density as assessed by fully automated means using TU can be of significant clinical value and play an important role in breast cancer risk assessment.