Agilent Dissolution Seminar Series Welcome · Agilent Dissolution Seminar Series Welcome ... •...
Transcript of Agilent Dissolution Seminar Series Welcome · Agilent Dissolution Seminar Series Welcome ... •...
-
The Fundamentals of Dissolution
1 2015 All rights reserved
Agilent Dissolution Seminar Series
Welcome
Agilent Dissolution
Seminar Series
Terry Way BPharm MAPS
Dissolution Science Consultant
Agilent Technologies
2
-
The Fundamentals of Dissolution
2 2015 All rights reserved
Program Outline
1. The Dissolution Technique
2. Dissolution Automation - Sampling
3. Dissolution Automation - Direct Measurements
4. Dissolution Apparatus Qualification
5. Fundamental Principles of Drug Release
6. Current Trends in Dissolution Testing
3
AN OVERVIEW OF THE
DISSOLUTION TECHNIQUE
4
-
The Fundamentals of Dissolution
3 2015 All rights reserved
Introduction
The dissolution test has evolved to become a definitive tool used to characterize the performance characteristics of solid oral dosage forms.
As dosage forms have become more unique over the last fifty years, the dissolution apparatus has required continuous improvement and modification to provide suitable conditions for performance testing of a wide variety of products.
However, probably 99% of dissolution testing is performedon traditional tablets and capsules.
5
What do we test?
Dissolution is not just about orally ingested products such as tablets and capsules.
We also test :suspensions and powders coated beads and granulesointments, creams, gelstransdermal patches implants, stentsmedicated contact lenseswound care productsbone cementpowders for inhalationchewing gums, etc.
6
-
The Fundamentals of Dissolution
4 2015 All rights reserved
Dissolution - DefinitionGoogle search:
Search Results
Dissolution - Glossary page - UK Parliamentwww.parliament.uk Site information GlossaryCached - Similar
Dissolution is the official term for the end of a Parliament. After the Fixed Term Parliament Act was passed on 15 September 2011, the date of the...
[PDF] Dissolution and Removal from the Register of Charities - Charity ...www.charitycommission.gov.uk/library/guidance/csd1077a.pdf
File Format: PDF/Adobe Acrobat - Quick View
Egypt's dissolution of parliament is a counter-revolution in all but ...www.guardian.co.uk/.../egypt-dissolution-parliament-counter-revolut...Cached
Jun 15, 2012 David Hearst: Will the coup by the old regime, and its attempt to install Ahmed Shafiq as president reignite Egypt's revolutionary forces?
Early Dissolutionwww.insolvencydirect.bis.gov.uk/.../Dissolution/EarlyDissolution.ht...Cached
The dissolution of a company ends its legal existence. Its property then either belongs to the Crown, ...
Company Law: Voluntary dissolutionwww.companylawclub.co.uk/topics/voluntary_dissolution.shtml Cached
Voluntary dissolution. Introduction The application. Notifying interested parties. Objections Dissolution by Companies House and its effects.
The Dissolution of the Monasterieswww.historylearningsite.co.uk Tudor EnglandCached - Similar
The dissolution of the monasteries was a major event in the reign of Henry VIII. The dissolution effectively ended the power of the monasteries...
[PDF] D183 - About divorce/dissolution - Family Lawwww.familylaw.co.uk/system/uploads/attachments/0002/.../D183.pdf
File Format: PDF/Adobe Acrobat - Quick View What is the difference between a divorce and a dissolution? ...
[PDF] BR20 - Pension and divorce or dissolution - Gov.ukwww.direct.gov.uk/prod_consum_dg/groups/.../dg_180317.pdf Similar
File Format: PDF/Adobe Acrobat - Quick View Pensions and divorce or dissolution.
7
Dissolution - Definition
Academic definition:
Dissolution is the process by which a solid substance enters into a
solvent to form a solution.
Pharmaceutical definition:
Dissolution is a test used throughout the life cycle of a pharmaceutical
product to evaluate the rate of release of a drug substance from the
dosage form.
8
http://www.parliament.uk/site-information/glossary/dissolution/http://www.parliament.uk/http://www.parliament.uk/site-information/glossary/http://webcache.googleusercontent.com/search?q=cache:AYd3PdRoo34J:www.parliament.uk/site-information/glossary/dissolution/+dissolution&cd=4&hl=en&ct=clnk&gl=ukhttp://www.google.co.uk/search?hl=en&cr=countryUK|countryGB&tbo=1&rlz=1T4GGLL_en-GBGB358GB358&biw=1164&bih=792&q=related:www.parliament.uk/site-information/glossary/dissolution/+dissolution&sa=X&ei=jaH5UIqECeuZ0QWxkIGoAg&ved=0CE0QHzADhttp://www.charitycommission.gov.uk/library/guidance/csd1077a.pdfhttp://docs.google.com/viewer?a=v&q=cache:OHDeshayeD4J:www.charitycommission.gov.uk/library/guidance/csd1077a.pdf+dissolution&hl=en&gl=uk&pid=bl&srcid=ADGEESgbmngLFRNvJkzLNEgzASFZtwdRLy8m5vHheLgKtxhPbebPsWPvTFGjO_ETBt4y270AKOVOEfBPnkKT-UydlNp1buV3zLrSj7wMhNTIV-bXVhu6WXP43Iq44Xiwpd2rI_mIrp-V&sig=AHIEtbQHe3zg-bUNw7UezQBmIiX61njanQhttp://www.guardian.co.uk/commentisfree/2012/jun/15/egypt-dissolution-parliament-counter-revolutionhttp://webcache.googleusercontent.com/search?q=cache:bbx2Xy7L9BQJ:www.guardian.co.uk/commentisfree/2012/jun/15/egypt-dissolution-parliament-counter-revolution+dissolution&cd=11&hl=en&ct=clnk&gl=ukhttp://www.insolvencydirect.bis.gov.uk/freedomofinformation/technical/CaseHelpManual/D/Dissolution/EarlyDissolution.htmhttp://webcache.googleusercontent.com/search?q=cache:l4G7Kl1qfaQJ:www.insolvencydirect.bis.gov.uk/freedomofinformation/technical/CaseHelpManual/D/Dissolution/EarlyDissolution.htm+dissolution&cd=12&hl=en&ct=clnk&gl=ukhttp://www.companylawclub.co.uk/topics/voluntary_dissolution.shtmlhttp://webcache.googleusercontent.com/search?q=cache:tBE6XNVPkXMJ:www.companylawclub.co.uk/topics/voluntary_dissolution.shtml+dissolution&cd=14&hl=en&ct=clnk&gl=ukhttp://www.historylearningsite.co.uk/dissolution_monasteries.htmhttp://www.historylearningsite.co.uk/tudor_england.htmhttp://webcache.googleusercontent.com/search?q=cache:IYLqyjW-wmkJ:www.historylearningsite.co.uk/dissolution_monasteries.htm+dissolution&cd=16&hl=en&ct=clnk&gl=ukhttp://www.google.co.uk/search?hl=en&cr=countryUK|countryGB&tbo=1&rlz=1T4GGLL_en-GBGB358GB358&biw=1164&bih=792&q=related:www.historylearningsite.co.uk/dissolution_monasteries.htm+dissolution&sa=X&ei=zZ_5UL20OabZ0QXejIGYAg&ved=0CFIQHzAFOAohttp://www.familylaw.co.uk/system/uploads/attachments/0002/1227/D183.pdfhttp://docs.google.com/viewer?a=v&q=cache:JMJWUkUAlBcJ:www.familylaw.co.uk/system/uploads/attachments/0002/1227/D183.pdf+dissolution&hl=en&gl=uk&pid=bl&srcid=ADGEEShKrFb69qOAUH67cNSm2HfPKUIymnk3E2AcwdGqcdQ3bpkAP6ciPI86KpDQ5fFcW8eRc_UIbpCx5wcjN2ExMxJ0BLIVnpKYitF-MCYVBqgElbz54DJuUmIt0KpS6vyFKuxwHVOb&sig=AHIEtbQOIH7PHmFoxytvWlZi-SSlSQUWoQhttp://www.direct.gov.uk/prod_consum_dg/groups/dg_digitalassets/@dg/@en/@over50/documents/digitalasset/dg_180317.pdfhttp://www.google.co.uk/search?hl=en&cr=countryUK|countryGB&tbo=1&rlz=1T4GGLL_en-GBGB358GB358&biw=1164&bih=792&q=related:www.direct.gov.uk/prod_consum_dg/groups/dg_digitalassets/@dg/@en/@over50/documents/digitalasset/dg_180317.pdf+dissolution&sa=X&ei=zZ_5UL20OabZ0QXejIGYAg&ved=0CGkQHzAJOAohttp://docs.google.com/viewer?a=v&q=cache:vholP1oOqBYJ:www.direct.gov.uk/prod_consum_dg/groups/dg_digitalassets/@dg/@en/@over50/documents/digitalasset/dg_180317.pdf+dissolution&hl=en&gl=uk&pid=bl&srcid=ADGEESi0rzlIscmU89tONisEDZZzbX5xXt6NUMVjKWx10eO-9ZTQn4mUTAwWVLb_NvIQ2SjwoHfscSZt4hkiXbla5vyOElifq2KzvfkJi2_cSNSkYPV8v3a1A3IlWegUSzpIpEqIiHEO&sig=AHIEtbTxBf2nQj3qqiYUr04R34ye9LfTCg
-
The Fundamentals of Dissolution
5 2015 All rights reserved
An Overview of Dissolution
Generally, active pharmaceutical
ingredients (API) are mixed with
inactive excipient materials and
pressed into a tablet or filled into
a capsule.
9
An Overview of Dissolution
In the body, a pharmaceutical active ingredient must be in solution before it
can be absorbed by the blood and ultimately carried to the receptor site to
render a therapeutic effect.
Dissolution is the process by which that active ingredient enters into a
solvent to yield a solution.
10
-
The Fundamentals of Dissolution
6 2015 All rights reserved
An Overview of Dissolution
Solid oral dosage forms typically
begin to disintegrate and dissolve in
the stomach.
The resulting solution passes into the
small intestine where dissolution
continues.
Surface areas:
Stomach ~0.5m2
Small Intestine ~200m2
11
An Overview of Dissolution
The dissolved active ingredient is
absorbed into the blood stream
through the walls of the small
intestine.
The blood carries the active
ingredient to the site of therapeutic
effect.
12
-
The Fundamentals of Dissolution
7 2015 All rights reserved
An Overview of Dissolution
Basically, the dissolution test mimics the first few stages of this process
under very controlled laboratory conditions (in vitro).
For immediate release products:
Wetting in the stomach
Disintegration in the stomach
Deaggregation in the stomach
Dissolution in the stomach and intestine
Permeation through the intestinal wall
Absorption into the blood stream
Transit to the therapeutic site (via liver)
Decomposition and elimination
13
An Overview of Dissolution
Dissolution is an important tool for characterizing the biopharmaceutical
properties of a pharmaceutical product at different stages throughout its life
cycle.
Product Development
API characterisation, Formulation evaluation, Stability testing
Bioavailability / Bioequivalence
In Vitro / In Vivo Relationships
Quality Control
Pass / Fail product release
Scale-Up and Post-Approval Changes
Raw materials, Formulation, Process, Manufacturing site
14
-
The Fundamentals of Dissolution
8 2015 All rights reserved
An Overview of Dissolution
Dissolution is one the three primary tests used to release a finished drug
product:
Assay determines the overall potency of the batch and ensures the
accuracy of the finished drug product.
Dose Uniformity determines the consistency among the individual
dosage units and ensures the precision of the manufacturing process.
Dissolution ensures that the performance of the finished drug product is
consistent with the release rates of the API as determined in bioavailability
studies during the clinical trials.
15
An Overview of Dissolution
Dissolution assesses the performance of drug products
To be effective, the test should be:
Predictive
Comparative
Discriminatory
Reproducible
16
-
The Fundamentals of Dissolution
9 2015 All rights reserved
COMPENDIAL
DISSOLUTION TESTING
17
Compendial Requirements
The requirements for dissolution testing were harmonised through the ICH
Q4B Guidelines in 2006:
The pharmacopial texts are based on the original USP General Chapters
and but, despite harmonisation there are still various regional differences.
18/112
-
The Fundamentals of Dissolution
10 2015 All rights reserved
Apparatus
A general description of the assembly which is common to both Apparatus 1
and 2 is included in the description for Apparatus 1:
Vessels
glass or other inert, transparent material
cylindrical with hemispherical bottom
partially immersed in a water bath or heated by a heating jacket
1L, 2L or 4L nominal capacity
a fitted cover may be used to retard evaporation
Motor with speed-regulating device
An apparatus that permits observation of the specimen and stirring element
during the test is preferable.
19/112
Dissolution Vessels
Vessels
Apparatus 1 and 2 typically use a 1000mL hemispheric shaped vessel made
of glass or suitably inert material.
Media volume should be between 500 and 1000mL with 900mL used
historically.
The 1L vessel has dimensions of 98 -106mm i.d. and 160 - 210mm in height.
20
-
The Fundamentals of Dissolution
11 2015 All rights reserved
Large volume vessels have been required for testing APIs with low solubility and also larger veterinary formulations
21
2-Liter vessel has dimensions of 98-106mm id and 280-300mm in height.
4-Liter vessel has dimensions of 145-155mm id and 280-300mm in height.
Dissolution Vessels
Apparatus 1 - Baskets
Used for testing tablets and capsules where disintegrated granules or beads
would pass through the 10# of the disintegration test but would be retained
on the 40# of the dissolution basket - dissolution progressing inside the
basket.
The original acceptance criteria required
75% dissolution within a specified period.
Beyond this time, particles could be small
enough to pass through the basket mesh
and fall to the bottom of the vessel.
22/112
-
The Fundamentals of Dissolution
12 2015 All rights reserved
Apparatus 1 - Baskets
More commonly used today for non-disintegrating products.
The basket mesh must remain clear throughout the test but is prone to
occlusion by bubbles and undissolved materials such as gelatinous or waxy
excipients which inhibit the flow of media through the basket.
Often used for testing enteric products where the pH must be changed.
23/112
24
USP Apparatus 1
-
The Fundamentals of Dissolution
13 2015 All rights reserved
USP Apparatus 1
Baskets
The historical USP 40 mesh dissolution basket has 40 openings per linear
inch. Openings are equal in both directions producing a standard square
weave. USP specifies that 40 mesh (40 x 40) screen be manufactured with
wire having a nominal 0.25mm diameter.
Harmonized basket specifications are now referred to as 0.22-0.31 mm wire
diameter with wire openings of 0.36-0.44 mm.
25
USP vs. JP Basket
USP 40-mesh Basket JP 36-mesh Basket
Microscopic comparison in identical scale show the JP basket with
larger wire and fewer openings
Test results under identical conditions reveal lower results for
Prednisone calibrator tablets tested in the JP basket
26
-
The Fundamentals of Dissolution
14 2015 All rights reserved
USP Apparatus 1
Baskets
Dissolution baskets are fragile and require proper handling and care.
Attachment or removal from the basket shaft requires holding the upper rim.
When not in use, store in a protective case.
Carefully inspect for damage or excessive wear since defective or misshaped
baskets will affect test results.
A basket with gold coating 2.5 m thick (0.0001 inch) is an allowable
variation of the standard 40-mesh basket.
27
28
-
The Fundamentals of Dissolution
15 2015 All rights reserved
USP Apparatus 1
Starting a dissolution test with baskets:
The product is placed in a dry basket and attached to the shaft.
The shaft is lowered into the vessel to the correct height.
Then rotation is started.
Dissolution should occur within the basket.
29
Apparatus 2 - Paddles
Baskets are unsuitable for testing products that release small
particles e.g. tablets made by direct compression or powder-
filled capsules. Hence, paddles were introduced as an
alternative test.
Also used for testing powders and suspensions.
The dissolution process should occur throughout the
dissolution medium with good mixing provided by adequate
speed - usually 50-75rpm.
pH change can be achieved by addition of buffer concentrates.
With slower speeds (and for some hydrophobic products)
coning is a problem:
30/112
-
The Fundamentals of Dissolution
16 2015 All rights reserved
31
USP Apparatus 2
USP Apparatus 2
The paddle apparatus consists of a metallic or suitably inert, rigid blade and shaft comprising a single entity.
A suitable two-part detachable design may be used provided the assembly remains firmly engaged during the test.
The paddle blade and shaft may be coated with a suitable inert material.
32
-
The Fundamentals of Dissolution
17 2015 All rights reserved
USP Apparatus 2
Starting a dissolution test with paddles:
The dosage unit must be allowed to settle to the bottom of the vessel prior to
rotating the paddle.
Unfortunately some types of dosage forms may float to the surface,
especially capsules.
33
Apparatus 2 - Paddles
Sinkers are often used to prevent products floating:
A small, loose piece of nonreactive material,
such as not more than a few turns of wire helix,
may be attached to dosage units that would otherwise
float. (USP Sinker)
An alternative sinker device :
(Japanese Sinker)
Other validated sinker
devices may be used.
34/112
-
The Fundamentals of Dissolution
18 2015 All rights reserved
Special Apparatus
Stationary Basket assembly
35/112
Basket for USP FelodipineExtended Release Tablets
Peak Vessel
The peak vessel reduces the
inherent inconsistencies in the
hydrodynamics of standard
hemispherical dissolution vessels.
An inverted peak is incorporated
into the bottom of the vessel,
displacing the unstirred zone,
preventing cone formation
Especially useful for bead
formulations.
Non-Compendial Dissolution Apparatus
-
The Fundamentals of Dissolution
19 2015 All rights reserved
Non-Compendial Dissolution Apparatus
Mini-Paddle Apparatus
Mini-Basket Apparatus
Based on USP Apparatus 1&2
100 or 200 mL vessels
Minimum volume ~30 ml
Tablets, Capsules
37
Non-Compendial Dissolution Apparatus
38
-
The Fundamentals of Dissolution
20 2015 All rights reserved
THE DISSOLUTION TEST
39
The Dissolution Environment
A very sturdy surface should be used to support the dissolution
apparatus.
The apparatus with full water bath and full vessels may weigh
up to 100 kg.
Benches should not impart vibration on the dissolution
apparatus.
Vibration has been shown to have a significant effect on dissolution rates.
Effects of both higher and lower dissolution rates have been seen.
40
-
The Fundamentals of Dissolution
21 2015 All rights reserved
The Dissolution Environment
Sources of Vibration
Fume hoods
Vacuum pumps
Construction
Mechanical shakers
Bench top centrifuges
Ultrasonic baths
Heater circulators or pumps
Worn parts or bearings
41
The Dissolution Environment
Lighting should be sufficient to perform visual observations.
Visual observations may offer clues to dissolution behavior.
Unusual observations should be documented
Considerations should be made for light sensitive products.
Local sinks and purified water sources for media prep and
disposal
Media handling and preparation equipment must not impart
vibration to the dissolution apparatus
42
-
The Fundamentals of Dissolution
22 2015 All rights reserved
The Dissolution Environment
Suitable media degassing equipment
Dedicated vessels and shafts (each apparatus)
Sufficient bench space for sample handling and analytical
measurement
43
The Dissolution Test
44
The analyst is responsible for verification of the physical
parameters.
- including that the equipment is clean.
-
The Fundamentals of Dissolution
23 2015 All rights reserved
The Dissolution Test
Prepare dissolution media
and properly deaerate.
USP Method: Heat media to
41C, vacuum filter through
0.45m filter, continue to pull
vacuum for 5 additional
minutes.
45
The Dissolution Test
Care must be taken in measuring the dissolution media to maintain the volumetric accuracy at 1%.
Volumetric accuracy is based on dissolution media measured at room temperature.
46
-
The Fundamentals of Dissolution
24 2015 All rights reserved
The Dissolution Test
Alternatively, dissolution media may be weighed.
47
The Dissolution Test
Measure and carefully
introduce dissolution
media to the vessel.
Have materials ready
for the test including
all sampling
equipment.
48
-
The Fundamentals of Dissolution
25 2015 All rights reserved
Allow media in each vessel to
reach 37 C 0.5 C and use
immediately.
49
The Dissolution Test
The Dissolution Test
Always handle dosage units with gloves (not cotton), forceps or tweezers which will not scratch or damage the surface of the dosage unit.
Examine the six dosage units. Do not use chipped, cracked or capped tablets
50
-
The Fundamentals of Dissolution
26 2015 All rights reserved
The Dissolution Test
Option: Record the dosage
unit weights? Weight is for
information and investigation
purposes only.
Dosage units are to be
chosen at random and may
not be selected or discarded
based on weight.
51
The Dissolution Test
Have analyst notebook or dissolution worksheets for recording information. All information must be recorded directly into analyst notebooks or official batch record worksheets.
52
-
The Fundamentals of Dissolution
27 2015 All rights reserved
The Dissolution Test
Prepare to drop tablets. Lower
paddles to proper height, or
suspend baskets until ready to
begin the test.
Note: Apparatus 1 baskets
should be tested immediately
after placing the tablet inside
and clipping to the shaft.
Exposure to humidity can alter
test results.
53
The Dissolution Test
Prepare calibrated timer.
Record times as tablets are dropped.
54
-
The Fundamentals of Dissolution
28 2015 All rights reserved
The Dissolution Test
For the paddle apparatus, drop the dosage units into non-rotating medium.
They must settle to the bottom of the vessel before rotation of the shaft
begins. Then start rotation.
Visually inspect the dosage forms for air bubbles immediately after dropping.
For the basket apparatus, lower the baskets into non-rotating medium.
When at the correct height, immediately start rotation.
Record any unusual observations.
55
The Dissolution Test
Most immediate release tablets disintegrate very rapidly and will reaggregate in the form of a cone on the bottom of the vessel.
As the active drug goes into solution during the dissolution process, some excipient material may be visible on the bottom of the vessel at the end of the test.
56
-
The Fundamentals of Dissolution
29 2015 All rights reserved
The Dissolution Test
Withdraw sample at the proper time 2% (a 30 minute sample must be pulled within 36 seconds) and filter immediately.
Temperature must be taken a second time at least, generally after the last sample is pulled and before the shaft has stopped
57
The samples are withdrawn at a depth halfway between the top of the paddle (or basket) and the top of the medium and not less than 1cm from the wall of the vessel.
The Dissolution Test
58
-
The Fundamentals of Dissolution
30 2015 All rights reserved
The Dissolution Test
Filtration stops the dissolution process
and defines the end of the first phase of
the test which is basically a sample
preparation period executed under strictly
controlled physical parameters.
59
The Dissolution Test
Once the sample has been filtered the second phase of the testing begins to
determine the analytical concentration of the sample.
Analytical concentration is generally determined through UV-Vis
spectroscopy or HPLC analysis. HPLC is primarily used for drug products
containing multiple active components or for stability testing.
The responses from the analytical finish will be used to calculate the amount
of sample released from the dosage form within the specific time interval.
60
-
The Fundamentals of Dissolution
31 2015 All rights reserved
USP Acceptance Table
Immediate Release Forms
Stage Number Tested Acceptance Criteria
S1 6 Each unit is Q + 5%.
S2 6 Average of 12 units (S1 + S2) is Q and no
unit is < Q 15%.
S3 12 Average of 24 units (S1 + S2 + S3) is Q,
not more than 2 units are < Q 15% and no
single unit is less than Q 25%.
If a sample fails either Stage S1 or S2, proceed to the next stage and test
the number of units indicated.
61
What is Q?
Q, as it relates to dissolution is commonly used in the USP for
immediate release and delayed release dosage forms.
The quantity of Q is the amount of dissolved active ingredient
specified in the individual monograph expressed as a
percentage of the labeled content.
62
-
The Fundamentals of Dissolution
32 2015 All rights reserved
CONCEPTS OF DISSOLUTION
AUTOMATION
63
The Concept of Automation
Where a sufficiently large number of similar units are to be
subjected routinely to the same type of examination, automated
methods of analysis may be far more efficient and precise than
manual methods.
USP Automated Methods
64
-
The Fundamentals of Dissolution
33 2015 All rights reserved
The Concept of Dissolution Automation
65
The Concept of Dissolution Automation
Dissolution consists of a series of unit operations.
Traditionally, these unit operations, when performed manually,
are quite technique-dependant which may cause
inconsistencies in test results.
66
-
The Fundamentals of Dissolution
34 2015 All rights reserved
The Concept of Dissolution Automation
The entire dissolution test is a series of unit operations, any
one of which may be manual or automatic:
Setup
The dissolution test
Sampling
Analysis and data reduction
Cleanup and changeover
67
Advantages and Limitations of Automation
Advantages of Automated Dissolution
Accuracy
Precision
Time saving buying time
Throughput cycle time reduction
Documentation of events
Data capture
68
-
The Fundamentals of Dissolution
35 2015 All rights reserved
Limitations of Dissolution Automation
Limitations of Automated Dissolution
Personnel
Synchronizing & sequential time demands
Evaporation
Altered hydrodynamics
Sample archiving
Validation
21 CFR Part 11
69
Future Challenges for Automation of Dissolution
Network control of dissolution systems
Video monitoring and recording
Remote data processing and report generation
Remote control of laboratory operations from home
70
-
The Fundamentals of Dissolution
36 2015 All rights reserved
Early Dissolution Automation
71
Simple Automation of Sample Collection
A fraction collector with peristaltic pump.
72
-
The Fundamentals of Dissolution
37 2015 All rights reserved
Multi-test Automation with Sample Collection
73
Advanced Automation of Sample Collection
New 850-DS Sampler
-
The Fundamentals of Dissolution
38 2015 All rights reserved
850-DS Dissolution Sampling StationOverview
Integrated syringe pump and fraction
collector
Streamlined design for ease of use and
reliable, repeatable operation
One sampler per dissolution apparatus
for ultimate flexibility
Built-in options and components for all
types of dissolution methods
Data export features ideal for paperless
environments
850-DS Dissolution Sampling StationOptions Printer, Filter Module
Standard:
Media replacement
Automated cleaning
Optional:
Built-in printer
Filter module (for dissolution
methods requiring 0.45m
filtration)
-
The Fundamentals of Dissolution
39 2015 All rights reserved
850-DS Dissolution Sampling StationOptions Sample Trays
Standard:
16 x 100 mm (up to 14 mL) 96 positions
Optional:
12 x 32 mm (2 mL vials) 96 positions
Agilent HPLC tray* 100 positions
Well plate (dual)* 192 positions
2 mL vials (dual)* 108 positions
*Requires needle block adjustment
850-DS Dissolution Sampling StationFeatures Color Touch Screen
Icons appear for key alerts Active tray format displayed prominently
Real-time display of method progress and remaining time points.
-
The Fundamentals of Dissolution
40 2015 All rights reserved
850-DS Dissolution Sampling StationFeatures Pump and Filtration
Easily disable filtration when not required
Improved sampling needle to prevent
clogging / coring of septa
Filter module incorporated to
save valuable bench space
Faster pump = more frequent time points
Stored volume calibration for repeatable precision
and accuracy
850-DS Dissolution Sampling StationFeatures Replacement Media and Cleaning
Automated cleaning and media replacement
is standard on all 850-DS sampling stations
An automated cleaning cycle can be added to the
end of any method to keep the system operating
properly and extend the life of internal components.
-
The Fundamentals of Dissolution
41 2015 All rights reserved
850-DS Dissolution Sampling StationFeatures User Levels and Data Export
Place instrument in specific Mode to restrict changes / access.
Three (3) data export options:
Printer, Serial Port, SD Card
Useful for method transfer (to another 850-DS) or LIMS integration.
Data export options:
SD Card or RS232
850-DS Dissolution Sampling StationSystem Configurations
708-DS Dissolution Apparatus
Cary 60 UV-Vis Spectrophotometer
8454 Diode-Array Spectrophotometer
850-DS Sampling Station (with Filter Module)
Bio-Dis Reciprocating Holder or
Apparatus 7 Reciprocating Cylinder
Dissolution Workstation Software
-
The Fundamentals of Dissolution
42 2015 All rights reserved
Automated Dissolution Analysis
UV-Visible Spectroscopy
HPLC
83
UV-Visible Spectroscopy
Pros
Fast
Reliable
Robust
Cost
Validation
84
-
The Fundamentals of Dissolution
43 2015 All rights reserved
UV-Visible Spectroscopy
Cons
Traditionally single component
Timing demands
Staggering starts
Excipient interference
Dilution
Cell pathlength
85
Fibre Optic Dissolution
86
-
The Fundamentals of Dissolution
44 2015 All rights reserved
Fibre-Optic Dissolution Testing
The use of fibre-optics for the determination of drugconcentrations during a dissolution test has
become a useful alternative to conventional
sampling techniques in research and development
as well as quality control environments.
First used in the late 1980s, several publicationsappeared in 1995 and full commercial systems
became available by 1999.
Fibre-Optic Dissolution Testing
The system allows continuous in-situ analysis ofdrug concentrations without the need for sample
withdrawal and preparation.
More comprehensive data can be obtained at a faster rate and the results can be interpreted
instantaneously.
Taking spectroscopy to the vessel: the flow-cell is immersed directly in the dissolution medium pump light not liquid
-
The Fundamentals of Dissolution
45 2015 All rights reserved
Advantages of Fibre Optics
No Sampling
Analysis in real time, igh speed collection
Increase in amount of collectable data
No Fluidics
pumps, valves
tubing
flow cells
No issues with liquid handling systems
carryover
leaking
adsorption / desorption
Procedure
Measure medium blanks
Measure Standards
Staggered start
Static or non-resident probes?
Optional post-run standards
-
The Fundamentals of Dissolution
46 2015 All rights reserved
Basics of Fibre-Optics
Quartz fibre transmits light by total internal reflection
Flexible (within limits)
Two fibres per probe
One coupled to source of mono-/poly-chromatic light
One coupled to detector
Direct Transmittance Transreflectance
Transreflectance Dip Probe
-
The Fundamentals of Dissolution
47 2015 All rights reserved
Dip Probe in-situ
Multiplexed Scanning Spectrophotometer
-
The Fundamentals of Dissolution
48 2015 All rights reserved
Multiplexed Scanning Spectrophotometer
One scanning spectrophotometer reading all fibres/vessels
Advantages
Only one detector/spectrometer to validate
Highest dynamic range
No stray light
Minimum solarization
Uses only low wavelengths when needed.
Use of a pulsed lamp allows light to be on only when measuring.
Disadvantages
Sequential data collection
Typical Dissolution
-
The Fundamentals of Dissolution
49 2015 All rights reserved
Considerations for Fibre-Optics
Excipient and bubble interference
Sticky residues
Cell path length
Optical fouling
Ease of blanking and standardization of each
probe
Filters (or lack thereof)
Validation
Significance of Probe Positioning
-
The Fundamentals of Dissolution
50 2015 All rights reserved
Problems
Interference in light path
Bubbles
degassing essential
No Filtration
Particles
Excipients
Typical Drug Spectrum
0
0.1
0.2
0.3
0.4
0.5
230 250 270 290 310 330 350 370 390
Wavelength (nm)
Ab
so
rban
ce
True drug absorption = measured - excipients
Baseline Correction
-
The Fundamentals of Dissolution
51 2015 All rights reserved
Baseline Correction
The spectrum of the interference should be flat to properly
perform the correction.
It only works as long as the total absorbance of the active, the
excipients, and the fibers is less than the maximum absorbance
limit of the spectrophotometer.
Active + Excipients + Fibers < Maximum Absorbance
0
0.5
1
1.5
2
2.5
3
AB
S (
AU
)
Active Excipients Fibers Total
Baseline Correction
-
The Fundamentals of Dissolution
52 2015 All rights reserved
Fibre-Optic Dissolution
Some products are suitable .. and some are not
Agilent Cary 60 with Fibre-Optics
Linear over 3.5 absorbance units.
Room light immune.
Split beam dual Si diode detectors.
High intensity Xenon flash lamp.
80 readings a second.
Scanning speed up to 24,000 nm/min
-
The Fundamentals of Dissolution
53 2015 All rights reserved
Agilent Cary 60 with Fibre-Optics
Very large dynamic working range, 0.0001 3.5 AU
Lamp only flashes when taking a reading giving it an extremely long life.
Very low baseline noise
Fast monochromotor, does not photobleach samples or fibres.
Agilent Cary 60 with Fibre-Optics
No warm up or stabilisation time for the lamp
Full UV-Vis 190 1100nm range.
Dedicated fibre optic dissolution software.
Full suite of stand-alone applications.
Monochromator
Sample
Detector
Xe Light
source
Entrance
slit
Exit slit
Fast Dispersion
element
Superior Optical Design.pdfSuperior Optical Design.pdf
-
The Fundamentals of Dissolution
54 2015 All rights reserved
Agilent 616 Fibre Optic Multiplexer
16 channel system for dual bath capability.
Purpose built for dissolution testing & Cary 60.
Each fibre is individually connected to the multiplexer.
Highest optical transmission of any system on the market allowing it to easily read in high turbidity solutions.
Agilent 616 Fibre Optic Multiplexer
High precision & high speed channel to channel movement.
Raw absorbances on a single channel @ 80 readings a second
Raw absorbances on multiple channels @ 6 readings every second
Full dissolution reports, graphs, and % dissolved data @ 6 readings every 30 seconds.
-
The Fundamentals of Dissolution
55 2015 All rights reserved
Agilent Fibre Optic Probes.
Specially designed for dissolution testing.
600 micron silica/silica fibers for optimised transmission down to 190 nm.
1, 2, 5, 10, 20 & 50 mm tips available.
Streamlined to minimize hydrodynamic influence.
Agilent UV Dissolution Software
110
A full suite of applications is included
such as:
Simple/Advanced Reads
Scan
Concentration
Scanning Kinetics
Validate
Align
UV Dissolution
Fibre Optic Multiplexer is fully compatible with all applications
-
The Fundamentals of Dissolution
56 2015 All rights reserved
UV Dissolution Hardware Configuration
111
Compatible equipment includes:
All Agilent dissolution apparatus
705, 708, 709, 7000, 7010, 7025 &
7030.
Non-Agilent dissolution apparatus
may also be used with a resident
probe setup.
UV Dissolution - Method Setup
112
Fully flexible timepoints
Actual times
(HHH:MM:SS) or intervals are
entered
No limit to number of stages
-
The Fundamentals of Dissolution
57 2015 All rights reserved
UV Dissolution - Method Setup
113
Tester Settings:
Apparatus type
Initial spin (RPM & time)
Spindle (RPM)
Infinity spin (RPM & time)
The Temperature page includes the
bath and vessel temperature
settings as well as tolerance
specifications
UV Dissolution - Method Setup
114
Standard Options:
Several online standard options
are available including: Pre &
Post Cycle Standard, Bridged
Mean, and Running Mean
Enhanced flexibility for
percentage and mg dissolved
calculations
Capsule Blank correction is also
available
-
The Fundamentals of Dissolution
58 2015 All rights reserved
UV Dissolution - Method Setup
115
Standard Options:
An option to measure a Check
Standard with a variable
tolerance field
Control Limits may be set up
to flag any out-of-tolerance
standards measured at any
time throughout the test
UV Dissolution - Method Setup
116
Analysis Options:
Analysis is possible using a single
wavelength, a scan range, or single
and scan
Scan analysis offers the choice
between mean or second derivative
Single and scan analysis allows you
to specify a wavelength for analysis,
and collect data at surrounding
wavelengths for possible
recalculation.
-
The Fundamentals of Dissolution
59 2015 All rights reserved
UV Dissolution - Method Setup
117
Analysis Options:
Full background Correction is
available using a single wavelength
or mean values of a scan range, to
correct for excipients in your solution.
UV Dissolution: Fiber Optic
118
The probe can be programmed to
dwell slightly under the vessel
media surface between time points.
This prevents drying of particles
and bubble formation while
minimizing hydro-dynamic
disturbance.
-
The Fundamentals of Dissolution
60 2015 All rights reserved
UV Dissolution: Fiber Optic
119
Multiplexer calibration and
probe performance is
monitored through the FO Mux
Calibration application.
Cary 60 Tests - Validate
120
Easy setup and execution of
routine calibration
Quick tests to monitor
instrument performance
and/or troubleshoot
Complete record of all tests
with time/date stamp
-
The Fundamentals of Dissolution
61 2015 All rights reserved
UV Dissolution: 21 CFR 11 Compliance
21 CFR Part 11 Compliance
121
UV Dissolution: 21 CFR 11 Compliance.
122
Spectroscopy Database
Administrator (SDA)
Spectroscopy Configuration
Manager (SCM)
Can be installed on a central
server or the local PC
The Win UV Dissolution software
runs off the local PC with
Windows 7 32- or 64- bit
-
The Fundamentals of Dissolution
62 2015 All rights reserved
Agilent Dissolution Seminar Series
Welcome
Replacement Media - 20 mL
-
The Fundamentals of Dissolution
63 2015 All rights reserved
No Replacement Media - 10 mL
QUALIFICATION OF DISSOLUTION
APPARATUS
126
-
The Fundamentals of Dissolution
64 2015 All rights reserved
Analytical Instrument Qualification
Good Manufacturing Practices (GMP) regulations
(e.g. FDAs 21 CFR 211.160 (b)(4)) require companies to
establish procedures ensuring the fitness for use of instruments
that generate data supporting regulated product testing.
No definitive guidance for the qualification of analytical
instruments.
USP proposed a General Chapter on Analytical
Instrument Qualification in 2005
127
Published in USP-NF in 2009
128
-
The Fundamentals of Dissolution
65 2015 All rights reserved
Analytical Instrument Qualification
USP provides a scientific approach to AIQ and
considers AIQ as one of the major components required for
generating reliable and consistent data.
AIQ is the collection of documented evidence showing that an instrument
performs suitably for its intended purpose.
Users of analytical equipment should:
validate their procedures
calibrate their instruments
perform additional instrument checks
System suitability tests
In-process control samples
129
USP definitions:
Design Qualification (DQ)
documented collection of activities that define the functional and
operational specifications of the instrument and criteria for selection of the
vendor, based on the intended purpose of the instrument.
Prior to purchase
Installation Qualification (IQ)
documented collection of activities necessary to establish that an
instrument is delivered as designed and specified, and is properly installed
in the selected environment, and that this environment is suitable for the
instrument.
performed at time of installation
130
-
The Fundamentals of Dissolution
66 2015 All rights reserved
USP definitions:
Operational Qualification (OQ)
documented collection of activities necessary to demonstrate that an
instrument will function according to its operational specification in the
selected environment.
After installation or major repair
Performance Qualification (PQ)
documented collection of activities necessary to demonstrate that an
instrument consistently performs according to the specifications defined by
the user, and is appropriate for the intended use
Performed after completion of OQ and periodically at specified intervals for
each instrument
131
USP definitions:
Performance Qualification (PQ):
Usually based on the instruments application and may consist
of analyzing known components or standards
PQ tests may be modular or holistic
PQ tests should be based on good science and reflect the
general intended use of the instrument
When an instrument fails to meet PQ test specifications, it
requires maintenance or repair
132
-
The Fundamentals of Dissolution
67 2015 All rights reserved
Requirements for dissolution equipment
OQ
after installation or repair
conformance to mechanical specifications
PQ
after installation, repair and at regular intervals
conformance to mechanical specifications
Performance Verification Test
133
Qualification of Dissolution Apparatus
USP requirements for pharmacopial dissolution tests were first introduced in 1970 for 6 monographs.
The basic concepts of the dissolution apparatus were established by empirical means rather than sound scientific and engineering considerations.
It very quickly became obvious that tests with different apparatus produced different results
apparatus was often custom-made
control of mechanical parameters was inadequate
vibration was thought to be a key factor but couldnt be conveniently quantified at the time.
Use of a calibrator was suggested
134
-
The Fundamentals of Dissolution
68 2015 All rights reserved
Mechanical Qualification
Specifications and tolerances for the apparatus itself are given in the
General Chapters
Tolerances for operating parameters are also provided,
some more specific than others, e.g.
speed 4%
temperature 0.5C
rotation without significant wobble
no significant vibration
These specifications and tolerances are the minimum standards
allowable and have since been supplemented by tighter standards
published by various authorities including USP, FDA, ASTM, FIP
135
Mechanical Qualification
FDA published The Use of Mechanical Calibration of
Dissolution Apparatus 1 and 2 in 2006:
Enhanced Mechanical Calibration (MC or MQ) of dissolution apparatus
may be used as an alternative to the current Apparatus Suitability
procedure for Dissolution Apparatus 1 and 2 described in USP
The Mechanical Calibration procedure should specify the frequency at
which each calibration step is performed
Either the USP procedure or an appropriate Mechanical Calibration
method executed according to a written procedure will satisfy the CGMP
requirement for calibration of the laboratory apparatus
136
-
The Fundamentals of Dissolution
69 2015 All rights reserved
Mechanical Qualification
In collaboration with FDA, in 2007 ASTM published Standard
Practice for Qualification of Basket and Paddle Dissolution Apparatus
outlines the procedures and tolerances for enhanced mechanical calibration
Moore et al published an Implementation Guidance document in 2010
The Open Drug Delivery Journal, 2010, 4, 14-20
2.2 Once a unit meets all of the mechanical specifications included in
this practice, it is considered calibrated and further calibration with
dissolution calibrator tablets is not required.
Abstract: This guidance is intended to serve as a companion document for ASTM
Standard E 2503-07, Standard Practice for Qualification of Basket and Paddle
Dissolution Apparatus, by providing practical information useful for the
implementation of mechanical calibration. Particular focus is placed on use of the
available tools to make the required measurements.
137
Mechanical Qualification
In 2007, USP published the Dissolution Toolkit - Procedures for
Mechanical Calibration and Performance Verification Test
This provides
...a description of best practices associated with the mechanical
calibration and performance verification test for the USP basket and paddle
dissolution apparatuses and test assemblies.
very detailed mechanical calibration procedure along with measuring
techniques, tools required, and frequency of measurement
detailed procedures for the use and interpretation of the USP Performance
Verification Test (PVT)
138
-
The Fundamentals of Dissolution
70 2015 All rights reserved
Mechanical Qualification
139
Specifications and Tolerances
Parameter ICH
(USP, JP, EP)
FDA
DPA-LOP.002
ASTM
E2503-07
USP
Toolkit Ver 2.0
Basket and
Paddle Depth
25 2 mm 25 2 mm 25 2 mm
(or
-
The Fundamentals of Dissolution
71 2015 All rights reserved
Specifications and Tolerances
Parameter ICH
(USP, JP, EP)
FDA
DPA-LOP.002
ASTM
E2503-07
USP
Toolkit Ver 2.0
Vessel/Shaft
Centering
NMT 2 mm
from center
axis
1.0 mm from
center line
1.0 mm from
center line
NMT 2.0 mm
difference (4 -
90 positions)
Vessel
Verticality
Not Measured 1.0 from
vertical (2 - 90
positions)
1.0 from
vertical (2 -
90 positions)
NMT 0.5 from
vertical
Vessel Plate
Level
Not Measured Not Measured Not Measured NMT 0.5 from
horizontal
Performance
Verification
Test (PVT)
USP
Prednisone
Tablets RS
Not Measured Not Measured USP
Prednisone
Tablets RS
141
Apparatus Suitability
Initially, the primary purpose of the USP Apparatus Suitability
test with Prednisone and Salicylic Acid tablets was to indicate
the influence of environmental factors and vibration on the
apparatus since most other parameters could be controlled by
mechanical measurements
The original test became known as Calibration although this
was not a true indication of the test being performed; later
changed to Performance Verification Test (PVT)
The PVT, has been responsible for detecting problems
associated with dissolution apparatus that are found to be within
mechanical tolerances
142
-
The Fundamentals of Dissolution
72 2015 All rights reserved
Apparatus Suitability
The Apparatus Suitability Test became a USP requirement for all
dissolution equipment in 1978
Determine the acceptable performance of the dissolution test assembly
periodically. The suitability for the individual apparatus is demonstrated
by the Apparatus Suitability Test.
Apparatus Suitability Test, Apparatus 1 and 2 Individually test 1
tablet of the USP Dissolution Calibrator, Disintegrating Type and 1 tablet
of USP Dissolution Calibrator, Nondisintegrating Type, according to the
operating conditions specified. The apparatus is suitable if the results
obtained are within the acceptable range stated in the certificate for that
calibrator in the apparatus tested.
143
Apparatus Suitability
Initially, 8 tests required:
50mg Prednisone Calibrator Tablets - disintegrating
baskets and paddles both at 50 and 100rpm
tested in 900ml water for 30 minutes
300mg Salicylic Acid Calibrator Tablets - non-disintegrating
baskets and paddles both at 50 and 100rpm
tested in 900ml phosphate buffer for 30 minutes
Acceptance criteria (upper and lower limits ) applied to each individual
position.
Variation between positions was not considered.
144
-
The Fundamentals of Dissolution
73 2015 All rights reserved
Apparatus Suitability
USP rationalised the requirements in 1997:
With experience of the tests, it was found that just testing at a single speed
gave sufficient information.
If the equipment was only used with one apparatus (baskets or paddles)
then the tests were only required with that apparatus.
In 1999, a new 10mg Prednisone tablet formulation was
introduced
tested in 500ml water
this formulation was found to be extremely sensitive to mechanical
variations in the apparatus and also to experimental procedures such as
degassing of the medium.
145
Apparatus Suitability
With these new tablets, industry started to have problems meeting the requirements and the credibility of the test was doubted.
After many experiments, it was found that there was no significant problem attributable to the tablets themselves and the procedure was shown to be a very sensitive, holistic test for qualification of dissolution equipment.
It was also concluded that sufficient information was obtained from the Prednisone tablet tests and additional testing with the Salicylic Acid tablets was no longer necessary
the requirement to test with Salicylic Acid tablets was withdrawn in December 2009.
146
-
The Fundamentals of Dissolution
74 2015 All rights reserved
Apparatus Suitability
The problem with the original requirement to evaluate individual
positions:
- Is this acceptable?
25.0
30.0
35.0
40.0
45.0
50.0
55.0
60.0
65.0
0 1 2
1
147
Apparatus Suitability
The procedure and requirements for the Apparatus Suitability
Test have been the subject of extensive review over several
years.
USP revised the concept of the test from a test on each
individual position in a dissolution tester to the ISO approach of
instrument proficiency testing based on tests on the whole
instrument.
This introduced Performance Verification Testing (PVT)
148
-
The Fundamentals of Dissolution
75 2015 All rights reserved
Apparatus Suitability
Performance Verification Testing (PVT)
Statistical analysis of the results of tests on all positions of an apparatus
now required.
Both the Mean of the results and the % Coefficient of Variation must pass
the acceptance criteria.
The Prednisone tablets were renamed as Reference Standard Tablets to
remove the implication of calibration.
149
2009 revision to USP
APPARATUS SUITABILITY
The determination of suitability of a test assembly to perform
dissolution testing must include conformance to the
dimensions and tolerances of the apparatus as given above.
Determine the acceptable performance of the dissolution test
assembly periodically. The suitability for the individual
apparatus is demonstrated by the Performance Verification
Test.
Performance Verification Test, Apparatus 1 and 2
Test USP Prednisone Tablets RS according to the operating
conditions specified. The apparatus is suitable if the results
obtained are within the acceptable range stated in the
technical data sheet specific to the lot used and the apparatus
tested.
150
-
The Fundamentals of Dissolution
76 2015 All rights reserved
Ph. Eur. 2.9.3
Apparatus suitability. The determination of suitability of a test assembly to
perform dissolution testing must include conformance to the dimensions
and tolerances of the apparatus as given above.
Determine the acceptable performance of the dissolution test assembly
periodically.
Apparatus suitability The determination of suitability of a test assembly
to perform dissolution testing must include conformance to the dimensions
and tolerances of the apparatus as given above.
Determine the acceptable performance of the dissolution test assembly
periodically.
J.P. 6.10
151
Apparatus Suitability
USP has always regarded mechanical calibration as a pre-
requisite for the Apparatus Suitability Test
not an alternative as implied by FDA
The new approach with the statistical evaluation of the
Performance Verification Test has shown that much dissolution
apparatus has been inadequately controlled in the past.
152
-
The Fundamentals of Dissolution
77 2015 All rights reserved
USP Performance Verification Test
PVT provides experimental proof that the dissolution test
assembly is suitable for dissolution testing
Apparatus 1 and Apparatus 2
Prednisone RS tablets
Apparatus 3
Chlorpheniramine Maleate Extended Release RS tablets
(requirement withdrawn from end of February 2012)
Apparatus 4
currently not available (no monographs used App. 4 until 2013)
153
USP Performance Verification Test
154
-
The Fundamentals of Dissolution
78 2015 All rights reserved
The USP ISO-based PVT Test
PVT now designed as a Single-Stage testing approach i.e. two
consecutive runs using all positions of an instrument
Collect the results as % prednisone dissolved at 30 min of two consecutive
runs on the dissolution assembly
Calculate the overall geometric mean (GM) and the mean of the variances
of both runs
Compare the obtained results with the applicable acceptance limits
PVT passes if both values meet acceptance criteria
155
Optional Two-Stage testing
Possibility of stopping after the first step
Step 1:
Perform one test with all positions of the dissolution assembly and
calculate % prednisone dissolved at 30 min
Determine the geometric mean (GM) and percent coefficient of variation
(%CV)
Compare the GM and %CV to the acceptance criteria.
If both values meet acceptance criteria, test finished
If not, go to Step 2
156
-
The Fundamentals of Dissolution
79 2015 All rights reserved
Optional Two-Stage testing
Step 2:
Perform an additional test and record the percent dissolved value
Determine the geometric mean (GM) and percent coefficient of variation (%CV) for this second test
Compute a pooled (from Step 1 and 2) GM and the mean of the variances of both runs
Compare to the acceptance criteria. If both values meet acceptance criteria, PVT passed
If GM and/or %CV do not meet the acceptance criteria the test has failed and further work is needed.
USP does not offer guidance as to which approach to use.
157
Acceptance Limits for the PVT with Prednisone
tablets (July 2015):
158
-
The Fundamentals of Dissolution
80 2015 All rights reserved
From experience...
The most common causes of failure are:
procedural
improper degassing
inaccurate dispensing
resident sampling probes and/or thermometers
sampling technique
equipment (this is the purpose of the test)
poor quality vessels
apparatus alignment
vibrations
159
Causes of failure
The most common cause of PVT failures due to the procedure
is degassing.
The recommended degassing technique is simple and efficient
Heat media to 41-45C
Vacuum filter (
-
The Fundamentals of Dissolution
81 2015 All rights reserved
De-aerated vs. Non-de-aerated Medium
161
The Prednisone Test
Cone formation is an essential and critical characteristic of the
test
dissolution occurs at the surface of the cone
saturated solution in the centre
any disturbance of the cone changes the results
162
-
The Fundamentals of Dissolution
82 2015 All rights reserved
Dispensing
The medium should be used immediately after degassing
temperature should not fall below 37C
dispense gravimetrically
163
Sampling
Samples should be taken at the correct time from the correct
position
30 minutes 2% (36 seconds)
mid-point between surface of
medium and top of paddle or
basket, at least 1cm from vessel wall
take a large sample (20-30ml)
filter immediately
164
-
The Fundamentals of Dissolution
83 2015 All rights reserved
Vessels
The most common cause of PVT failures due to the apparatus is
the vessels.
Vessels usually made by traditional glass-blowing techniques
Each vessel is more or less unique since they are essentially hand-made,
individual flaws add to hydrodynamic variability
Modern manufacture of vessels is better controlled and individual vessels
may be certified
165
Vessels
The compendial specification is not well-defined:
made of glass or other inert, transparent material
cylindrical with a hemispherical bottom
sides are flanged at the top
for a nominal capacity of 1 liter, the height is 160mm to 210mm
and its inside diameter is 98mm to 106mm
166
-
The Fundamentals of Dissolution
84 2015 All rights reserved
Vessels
Even with perfect manufacture, within specification differences are significant
Such differences can change dissolution results
Surface +17% Area
Volume of -10%Cylinder
Volume of +27%Hemisphere
98mm ID for 900ml 106mm ID
167
Vessels
Vessel Dimensions Inner Diameter
Sets of 6 vessels from 10 different manufacturers
99
100
101
102
103
104
105
0 2 4 6 8 10 12
Source
Avera
ge In
ner
Dia
mete
r (m
m)
168
-
The Fundamentals of Dissolution
85 2015 All rights reserved
MECHANICAL QUALIFICATION STANDARDS
169
Dissolution Mechanical Qualification Standard
Requirements
General MQ Requirements:
Check vessel, basket and paddle dimensions on receipt
Perform maintenance procedures recommended by
manufacturer
Perform mechanical qualification:
After apparatus is moved
After apparatus is repaired
After 6-months from previous calibration
Perform Operational Checks at each time of use
170
-
The Fundamentals of Dissolution
86 2015 All rights reserved
Dissolution Mechanical Qualification Standard
Requirements
Check Dimensions Upon Receipt:
Paddle, Basket/Shaft, and Vessel apparatus components must be verified that they conform with the harmonized standard USP Dissolution (USP, JP, EP)
Individual measurements for each dimension of each component must be documented
Certificates of Conformance (COC) may be obtained from Agilent to document conformance.
Otherwise, measurements must be documented by the end user.
171
Dissolution Mechanical Qualification Standard
Requirements
Perform Maintenance Procedures:
Lubricate moving parts
Check belt for wear and proper
tension
Check power cords and cable
connections for wear
Check alignment of belt, pulleys and
spindle housing
Clean water bath, replace tubing if
necessary and use algaecide
approved for heater/circulators
172
-
The Fundamentals of Dissolution
87 2015 All rights reserved
Dissolution Mechanical Qualification Standard
Requirements
Mechanical Qualification Requirements:
Shaft Wobble
Paddle/Basket Shaft Verticality
Basket Wobble
Vessel Centering
Vessel Verticality
Paddle/Basket Depth
Rotational Speed
173
Dissolution Mechanical Qualification Standard
Requirements
Operational Checks (document each time of use):
Basket/shaft examination
Paddle examination
Vessel examination
Vessel Temperature
Vibration
174
-
The Fundamentals of Dissolution
88 2015 All rights reserved
Dissolution Mechanical Qualification Standard
Requirements
Basket and Basket Shaft, must be free from:
Defects
Rusting
Corrosion
Loose wires
Clogged mesh openings
Dented sides or bottom
Knicks, dents or misshapen appearance
O-ring are not compliant; three clips required
175
Dissolution Mechanical Qualification Standard
Requirements
Paddle Examination, must be free from:
Defects
Rusting
Corrosion
Peeling or loose coating
Knicks, dents or misshapen appearance
176
-
The Fundamentals of Dissolution
89 2015 All rights reserved
Dissolution Mechanical Qualification Standard
Requirements
Vessel Examination, must be free from:
Scratches
Cracks
Pits
Residue
Surface irregularities
177
TOOLS FOR MECHANICAL QUALIFICATION
178
-
The Fundamentals of Dissolution
90 2015 All rights reserved
Mechanical Qualification
Height:
Wobble:
179
Mechanical Qualification
Temperature Level
180
-
The Fundamentals of Dissolution
91 2015 All rights reserved
The Agilent 280-DS
Fast less than 35 minutes
Simple only two devices
Easy to Use - software prompts every measurement
Data Capture - standardized reporting, historical tracking and trending by serial number (instrument, vessel, paddle, basket, etc)
21 CFR Part 11 Compliant -Certificates of Conformance - storage and tracking
Control - Agilent dissolution apparatus during measurement
181
280-DS Mechanical Qualification System
Components
Vessel Module
Instrument Module
Temperature ProbePersonal Computer
(Optionally Supplied)
Software
182
-
The Fundamentals of Dissolution
92 2015 All rights reserved
280-DS Mechanical Qualification SystemVessel Module (VM)
Once placed in the vessel
this will measure:
Spindle Speed (RPM)
Shaft Wobble
Basket Wobble
Shaft Verticality
Vessel Verticality
Vessel Centering
Basket/Paddle height
280-DS Mechanical Qualification System
Instrument Module (IM)
The Instrument Module (IM) is placed on
the vessel plate to measure:
Vessel Plate Level
Vibration
Temperature
184
-
The Fundamentals of Dissolution
93 2015 All rights reserved
280-DS Mechanical Qualification System
How it Works
185
280-DS Mechanical Qualification SystemSoftware Capabilities
Configuration files maintain all relevant
apparatus data including serial numbers of
critical components
Method files may be chosen from a pre-loaded
list of regulated procedures or customized based
on specific internal requirements
All necessary controls available for use in a 21
CFR Part 11 environment including User
Groups, Audit Trails, etc.
Text directions as well as graphics instruct the
user how and when to perform each
measurement
All results are stored within a database that may
be created locally or on a secure network
-
The Fundamentals of Dissolution
94 2015 All rights reserved
280-DS Mechanical Qualification SystemSoftware Capabilities
Select specific positions to be tested for quick investigations
Record exact orientation of vessel within the data file
Filter past reports using various criteria including dates, operator,
apparatus, etc.
Conveniently monitor system and/or method changes with built-in audit
trail reports
Inspect individual position measurements
using the data trending capability
Certificates for all components can be
stored and linked within Apparatus file.
280-DS Mechanical Qualification SystemSoftware Capabilities
Comprehensive Reporting:
-
The Fundamentals of Dissolution
95 2015 All rights reserved
280-DS Mechanical Qualification SystemVibration Monitoring
PVT OR MQ
WHAT IS RIGHT FOR YOUR LABORATORY?
190
-
The Fundamentals of Dissolution
96 2015 All rights reserved
PVT or MQ
What Is Right For Your Laboratory?
PVT Advantages:
Holistic test with actual controlled tablets
Conforms closely to USP Analytical Instrument Qualification, requirement for Performance Qualification
New USP Acceptance Criteria provide statistical interpretation of results for an instrument as a whole.
Detects issues with Vessel Quality
Detects environmental issues
PhRMA studies showed PVT should be maintained until a definitive vibration specification is developed
191
PVT or MQ
What Is Right For Your Laboratory?
When a PVT failure is
observed it is a situation that
requires investigation.
The PVT does not tell us
exactly what is wrong but if we
proceed without resolution, the
consequences may be
catastrophic!
192
-
The Fundamentals of Dissolution
97 2015 All rights reserved
PVT or MQ
What Is Right For Your Laboratory?
PVT Disadvantages:
Time takes significant time to perform
Expense:
Cost of USP prednisone tablets
Cost of USP reference standards
Analysts time away from production
Expertise - requires training and experience
FDA claims of USP tablet variability
193
PVT or MQ
What Is Right For Your Laboratory?
Time Savings (35-min vs a few hours with PVT)
Ease of use allows shorter qualification intervals
Data that helps you pinpoint possible problems
Now have the ability to trend parameter variation over time
No guesswork associated with manual
and analog gauges
No standards or tablets to purchase
Provides instant failure investigation
information
Reports exact information required by
Enhanced MQ Standards
194
-
The Fundamentals of Dissolution
98 2015 All rights reserved
PVT or MQ
What Is Right For Your Laboratory?
MQ Disadvantages:
Misinterpretation of Certificates of Conformance for components These are not to be confused with manufacturing conformance certificates which state parts are manufactured according to USP specifications
Subjectivity of Component Examination at time of use
Does not account for:
Vessel Quality
Vibration
Deaeration
Other Laboratory Environment Issues
196
PVT or MQ
What Is Right For Your Laboratory?
Primary reason customers transition to MQ:
Anticipated savings for time and money
Anticipated reduction in frequency and intensity of
investigations
Common misconceptions about transition from PVT to MQ:
Perceived variability in USP Prednisone PVT cited by FDA
Requirement for individual component Certificates of
Conformance is not necessary
MQ will reduce apparatus variability, however, vessel quality
and effects of vibration will not be challenged
197
-
The Fundamentals of Dissolution
99 2015 All rights reserved
Current Situation with PVT vs EMC
PVT is only a USP requirement
Conformance to USP is only required for products with USP monographs
marketed in the USA
USP General Notices 6.30 allows alternative validated procedures
Federal law states that FDA should enforce USP requirements
FDA inspectors have issued 483s to companies who have just abandoned
PVT with no documented justification of EMC as an alternative procedure
198
Current Situation with PVT vs EMC
There has recently been debate around the industry about the
necessity of PVT using the Prednisone Tablets.
FDA and FIP have indicated that they do not consider PVT
essential if there is an adequate enhanced mechanical
calibration regime.
PVT is required by the USP but is only suggested by other
pharmacopias.
USP states that enhanced mechanical calibration is an essential pre-
requisite to PVT.
199
-
The Fundamentals of Dissolution
100 2015 All rights reserved
FUNDAMENTALS OF DRUG
RELEASE
- BIOPHARMACEUTICS
Bioavailability and Bioequivalence
FDA Definitions
Pharmaceutical Equivalence
Same active drug ingredient; same strength; same dosage form and route of administration; comparable labeling; meets compendial or other applicable standards of strength, quality, purity, and identity.
Bioavailability
The rate and extent to which the active ingredient or active moiety is absorbed from a drug product and becomes available at the site of action.
Bioequivalence
Two pharmaceutical equivalent drug products are bioequivalent if after drug administration, the bioavailabilities (rate and extent of drug availability) provide similar effects with respect to efficacy and safety.
201
-
The Fundamentals of Dissolution
101 2015 All rights reserved
Bioavailability and Bioequivalence
WHO Definitions
Pharmaceutical Alternative Products
They contain the same molar amount of the same pharmaceutical
ingredient but differ in dosage form and/or chemical form and are
administered by the same route.
Therapeutic Equivalence
They are pharmaceutically equivalent or alternatives and their safety and
efficacy are the same when administered by the same route..
Interchangeable Pharmaceutical Product
Therapeutically equivalent to a comparator product.
202
Biopharmaceutics Classification System (BCS)
The BCS is a scientific framework for classifying drug
substances based on their aqueous solubility and intestinal
permeability
When combined with the dissolution of the drug product, the
BCS takes into account three major factors that govern the rate
and extent of drug absorption from IR solid oral dosage forms:
solubility, intestinal permeability and dissolution
203
-
The Fundamentals of Dissolution
102 2015 All rights reserved
Biopharmaceutics Classification System - BCS
The BCS is effectively a risk tool adopted by regulatory agencies which:
characterises the absorption properties of a drug
characterises the likely clinical performance (PharmacoKinetics) of new and old formulations
BCS terminology is a useful way to describe the biopharmaceutical properties of a compound.
In August 2000 the FDA introduced this into guidance, which:
Sets solubility, permeability and dissolution limits
Allowed biowaivers (removal of bioequivalence studies) for Class 1 compounds/products due to their lower level of biopharmaceutical risk
204/112
Biopharmaceutics Classification System - BCS
205/112
1
3
2
4
SolubilityLow
Dose NOT soluble in
250 mL pH 1 to 7.5
Permeability
High
Extent of
absorption
>90%
For drugs in the Class 1 category, if dissolution >85% in 30mins, the FDA may waive the requirements for BA/BE studies.
HighDose soluble in
250 mL pH 1 to 7.5
Low
Extent of
absorption
-
The Fundamentals of Dissolution
103 2015 All rights reserved
Biopharmaceutics Classification System - BCS
Examples
206/112
Class 1: High Solubility High Permeability
Verapamil
Propranolol
Metoprolol
Class 3: High Solubility Low Permeability
Ranitidine
Cimetidine
Atenolol
Class 2: Low SolubilityHigh Permeability
Ketoprofen
Naproxen
Carbamazepine
Class 4: Low Solubility
Low Permeability
Furosemide
Hydrochlorthiazide
207
BCS - Solubility
The solubility class boundary is based on the highest dose strength of
an IR product that is the subject of a biowaiver request
A drug substance is considered highly soluble when the highest dose
strength is soluble in 250 ml or less of aqueous media over the pH
range of 1-7.5
The volume estimate of 250 ml is derived from typical BE study
protocols that prescribe administration of a drug product to fasting
human volunteers with a glass
(about 8 ounces) of water
-
The Fundamentals of Dissolution
104 2015 All rights reserved
208
BCS - Permeability
The permeability class boundary is based
directly on the rate of mass transfer of a drug substance in humans
across human intestinal membrane
Alternatively, non-human systems capable of predicting the extent
of drug absorption in humans can be used (e.g.; in vitro epithelial
cell culture methods, CaCo2 cells)
In the absence of evidence suggesting instability in the
gastrointestinal tract, a drug substance is considered to be highly
permeable when the extent of absorption in humans is determined
to be 90% or more of an administered dose based on a mass
balance determination or in comparison to an intravenous reference
dose
209
BCS - Dissolution
An IR drug product is considered rapidly dissolving when no less than
85% of the labeled amount of the drug substance dissolves within 30
minutes using
USP Apparatus I (basket) at 100 rpm,
or Apparatus II (paddle) at 50 rpm
A volume of 900 ml or less in each of the following media:
0.1 N HCl or Simulated Gastric Fluid USP without enzymes
pH 4.5 buffer
pH 6.8 buffer or Simulated Intestinal Fluid USP without
enzymes
-
The Fundamentals of Dissolution
105 2015 All rights reserved
In Vitro In Vivo Correlation?
In Vitro In Vivo
210/112
In Vitro In Vivo Correlation
Early practice was to develop dissolution requirements based on the in
vitro performance of clinically successful formulations.
Similarity in dissolution behavior has long been sought from the
perspectives of both bioavailability and quality control considerations.
The goal of the pharmaceutical scientist is to find a relationship between an
in vitro characteristic of a dosage form and its in vivo performance.
In vitroin vivo correlation, IVIVC, refers to the establishment of a rational
relationship between a biological property, or a parameter derived from a
biological property produced by a dosage form, and a physicochemical
property or characteristic of the same dosage form.
USP In Vitro And In Vivo Evaluation Of Dosage Forms
211
-
The Fundamentals of Dissolution
106 2015 All rights reserved
212
Parameters to be Correlated
In Vivo data is taken from human pharmacokinetic studies
Overlapping of kinetic subprocesses
Apparent in vivo dissolution needs to be deconvoluted
at least distribution and elimination taken into account
Prerequisite is that in vivo dissolution is rate limiting step
In Vitro data is taken from dissolution experiments
Experimental design determines primary data
closed models (paddle) provide cumulative profiles
Open models (flow-through cell) provide fractionated profiles
No preference for model, profiles a convertible
213
In Vivo
Transformation of plasma levels to in vivo dissolution
-
The Fundamentals of Dissolution
107 2015 All rights reserved
214
In Vitro In Vivo Correlation
The biological properties most commonly used are one or more
pharmacokinetic parameters obtained following the administration of the
dosage form
AUC
Cmax
The physicochemical property most commonly used is a dosage form's
in vitro dissolution behavior
% drug released under a given set of conditions
The relationship between the two properties, biological and physicochemical,
is then expressed quantitatively
215
IVIVC Correlation Levels according to USP
Level A
Highest category of correlation
Represents a point-to-point relationship between in vitro dissolution
and the in vivo input rate of the drug from the dosage form
Compares the % drug released versus % drug absorbed
-
The Fundamentals of Dissolution
108 2015 All rights reserved
216
IVIVC Correlation Levels
Level A Advantages:
A point-to-point correlation is developed
All in vitro dissolution data and all the in vivo plasma drug concentration profile data are used
An in vitro dissolution curve can serve as a surrogate for in vivo performance
A change in manufacturing site, method of manufacture, raw material supplies, minor formulation modification, and even product strength using the same formulation can be justified without the need for additional human studies
A truly meaningful (in vivo indicating) quality control procedure, which is predictive of a dosage form's performance, is defined
Can justify quality control limits
217
Level A IVIVC Example
0
2
4
6
8
0 2 4 6 8 10 12
Time (hr)
Pla
sma
Conce
ntr
atio
n (
ng/m
L)
Formulation 1
Formulation 2
0
20
40
60
80
100
0 1 2 3 4
Time (hr)
Perc
en
t d
rug
rele
ased
(%
)
Formulation 1
Formulation 2
Dissolution profiles for two different tablet formulations
Bioavailability from two different tablet formulations
-
The Fundamentals of Dissolution
109 2015 All rights reserved
218
Level B
Utilizes the principles of statistical moment analysis
Compares the mean in vitro dissolution time, MDT, to either the
mean residence time, MRT (mean time that the drug molecules
stay in the body), or the mean in vivo dissolution time
Utilizes all of the in vitro and in vivo data
Not considered to be a point-to-point correlation because it does
not reflect the actual in vivo plasma level curve
Cannot rely upon a Level B correlation alone to justify formulation
modification, manufacturing site change, excipient source change,
batch-to-batch quality, etc.
Cannot justify quality control limits
IVIVC Correlation Levels
219
Level B Correlation
In Vitro - In Vivo Correlation of Mean Dissolution Times
MDTvivo, tablet = MRTtablet - MRTsolution
-0.1 0 0.1 0.2 0.3 0.4 0.5 0.6MDTvitro [h]
-0.5
0
0.5
1
1.5
2
2.5
MD
Tviv
o [
h]
RR
T1T1
T2T2
T3T3
-
The Fundamentals of Dissolution
110 2015 All rights reserved
220
IVIVC Correlation Levels
Level C
Relates one dissolution time point (t50%, t90%, etc.) to one
pharmacokinetic parameter such as AUC, Cmax, or Tmax
Single point correlation
Does not reflect the complete shape of the plasma level, which
is the critical factor that defines the performance of finished
drug products
Has limited usefulness in predicting in vivo drug performance
Cannot be used to justify formulation modification,
manufacturing site change, excipient source change, batch-to-
batch quality, etc. or quality control limits
221
Level C Correlation
Tmax vs. 30 minute dissolution (different formulation
variables)
-
The Fundamentals of Dissolution
111 2015 All rights reserved
222
Level C Correlation
Cmax vs. 30 and 60 minute dissolution (different
formulation variables)
223
In Vitro In Vivo Correlation
Immediate release and rapidly dissolving drug products
Difficult to obtain an IVIVC due to relationship of dissolution and
absorption in vivo
Rapid dissolution - may only obtain one or two dissolution
samples within short time period (e.g.;
-
The Fundamentals of Dissolution
112 2015 All rights reserved
224
In Vitro In Vivo Correlation
Extended release drug products
IVIVC works best for ER drug products where in vivo drug release is rate limiting step by design
Need to compare at least two formulation variables to set specification ranges
More rapid drug release
Less rapid drug release
Alternatively target vs. upper/lower side batches
Need to consider different USP Apparatus and/or different dissolution conditions
225
In Vitro In Vivo Correlation
IVIVC is drug formulation dependent under specific dissolution
conditions
Generic drug products containing same pharmaceutical
ingredients may not have the same IVIVC
-
The Fundamentals of Dissolution
113 2015 All rights reserved
226
Lack of In Vitro-In Vivo Correlation
May not be able to obtain an IVIVC for your drug product where :
Systemic drug absorption is the rate-limiting step for absorption
Variations in drug dissolution/release are not reflected in
variations in drug absorption
Dissolution is not rate limiting step
Drug dissolution test is not discriminating
May need to modify dissolution conditions
227
Lack of In Vitro-In Vivo Correlation
Bioindicative dissolution method needed
Dissolution media do not reflect physiological conditions in the GI tract
pH in different regions of GI tract
Contents of GI tract
Liquids and solids
Fed or fasted state
Normal digestive enzymes
Water content
Other factors affecting systemic drug absorption
GI transit time
Pre-systemic drug elimination (first pass eff