Aespire View - GE Healthcare€¦ · Aespire* View shown with GE Tec 7* Vaporizer and CARESCAPE...
Transcript of Aespire View - GE Healthcare€¦ · Aespire* View shown with GE Tec 7* Vaporizer and CARESCAPE...
Aespire* View shown with GE Tec 7* Vaporizer and CARESCAPE Monitor B650
GE Healthcare
Aespire* ViewClarity of view…Clarity of decision
Features• Large12-inchcolorventilatordisplaywithcolor waveformandalarmmessageindicators
• Lightweightandcompactforeasymaneuverability
• OptionalTotalFreshGasmodulesforelectronicmeasurement anddisplayofthetotalfreshgasflowandindividual gasflow(O2, Air, N2O)
• OptionalintegratedauxiliaryO2 flowmeter andsuctioncontrol
• Enhancedmonitorintegrationcapabilitieswithour GE Healthcare Anesthesia Monitor and Compact Monitor
Advanced Ventilation: 7900 SmartVent*• Maximumversatilityforfullpatientrange
• VentilationModes:
-VCV(VolumeControl)
-PCVPressureControl(optional)
-PSVPro*(PressureSupportwithApneaBackup)(optional)
-SIMV/PSV(SynchronizedIntermittentMandatoryVentilation withpressuresupport)(optional)
-SIMV-PC(SynchronizedintermittentMandatoryVentilationwithpressurecontrol)(optional)
-PCV-VG(PressureControlledVentilation– VolumeGuaranteed)(optional)
• ElectronicPEEP
• Automaticfreshgasflow(tidalvolume)compensation
• Cardiacbypasscasemode
• Directaccesstoventilatorparametersettings
• Pressurewaveformforvisualreferenceona breath-by-breathbasis
• Smartalarmsdirectusertospecificproblems and affected parameters
• Inspiredoxygenmonitoring
Advanced Breathing System (ABS)• Easytoclean,autoclavable,latex-free
• Fastresponse–exceptionalforlowflowanesthesia
• Easyremoval–notoolsrequired
• Integrateddesign–lesspartsandconnections helpsreducepotentialforleaksandmisconnects
• Onestepbag/ventswitchturnsventilatoron/off
• OptionalCO2bypasswithelectronicdetection of“absorberoff”
• OptionalEZChangeandcondensor
Physical Specifications
Dimensions
Height: 136cm/53.5in
Width: 75cm/29.5in
Depth: 74cm/29.1in
Weight: approximately136kg/300lbs
Top shelf
Weightlimit: 34kg/75lbs
Width: 65cm/26in
Depth: 43cm/15.75in
Work surface
Height: 83.8cm/33in
Size: 2160cm2/334in2
Folding side shelf (optional)
Height: 87.5cm/34.5in
Width: 26.5cm/10.4in
Depth: 31.5cm/12.4in
Weightlimit: 11.3kg/25lbs
DIN rail
Sideofmachine:34.5cm/13.6in
Drawers (internal dimensions)
Height: 17.5cm/6.9in
Width: 33cm/13in
Depth: 26.5cm/10.4in
Absorber bag arm (optional)
Armlength: 30.5cm/12in
Bagarmheight 87cm/34.3in (adjustable): 104cm/40.9in
Casters
Diameter: 12.5cm/5in
Brakes: Individuallocking
Ventilator screen
Displaysize 31cm/12.1in(diagonal)
Ventilation operating modes
VCV
PCV
SIMV/PSV
PSVPro*withApneabackup
SIMV-PC
PCV-VG
Ventilator (VT) parameter ranges
Tidalvolumerange: 20to1500mL(Volume ControlSIMV/PSVand
PCV-VGmodes)
5to1500mL(Pressure ControlMode)
Incrementalsettings: 20to50mL (incrementsof1mL)
50to100mL (incrementsof5mL)
100to300mL (incrementsof10mL)
300to1000mL (incrementsof25mL)
1000to1500mL (incrementsof50mL)
Minutevolumerange: 0to99.9L/min
Pressure(PInspired)range: 5to60cmH2O (incrementsof1cmH2O)
Pressure(Plimit)range: 12to100cmH2O (incrementsof1cmH2O)
Pressure(Psupport)range: Off,2to40cmH2O (incrementsof1cmH2O)
Rate: 4to100breathsperminute forPCV-VG,VolumeControl andPressureControlvent
modes
2to60breathsperminute forSIMV/PSV,PSVProandSIMV
vent modes (increments of1breathperminute)
Inspiratory/expiratoryratio: 2:1to1:8(incrementsof0.5)
Inspiratorytime: 0.2to5.0seconds (incrementsof0.1seconds) (SIMVandPSVProventmodes)
Triggerwindow: 0to80%(incrementsof5%)
Flowtrigger: 0.2to1.0L/min (incrementsof0.2L/min)
1to10L/min(increments of0.5L/min)
Inspirationterminationlevel: 5to75%(incrementsof5%)
Backupmodedelay: 10to30seconds (incrementsof5seconds)
Positive End Expiratory Pressure (PEEP)
Type: Integrated,electronically controlled
Range: OFF,4to30cmH2O (incrementsof1cmH2O)
Ventilator performance
Pressurerangeatinlet: 240kPato700kPa/ 35psigto100psig
Peakgasflow: 120L/min+freshgasflow
Flowvalverange: 1to120L/min
Freshgasflowcompensation
Ventilator monitoring
Expiratoryminutevolumerange: 0to99.9/min
Expiratorytidalvolumerange: 0to>1500mL
O2%: <5to110%
Peakpressure: –20to120cmH2O
Meanpressure: –20to120cmH2O
Plateaupressure: 0to120cmH2O
Pressurewaveformsweepspeed:4to25breathsperminute (0to16seconds)
26to75breathsperminute (0to8seconds)
75breathsperminute (0to4seconds)
Ventilator Operating Specifications
Tec6PlusVaporizer Tec 7 Vaporizer
Ventilator accuracy
Delivery/monitoring accuracy
Volumedelivery: >210mL=betterthan7% <210mL=betterthan15mL <60mL=betterthan10mL
Pressuredelivery: ±10%or±3cmH2O
PEEPdelivery: ±1.5cmH2O
Volumemonitoring: >210mL=betterthan9% <210mL=betterthan18mL <60mL=betterthan10mL
Pressuremonitoring: ±5%or±2cmH2O
Alarm settings
Tidalvolume(TVexp): Low:OFF,5to1500mL High:20to1600mL,OFF
Minutevolume(Mvexp): Low:OFF,0.1to10L/min High:0to30L/min,OFF
Inspiredoxygen(FiO2): Low:18to99% High:21to99%,OFF
Apneaalarm: MechanicalventilationON: <5mLbreathmeasuredin 30seconds
MechanicalventilationOFF: <5mLbreathmeasuredin 30seconds
Lowairwaypressure: 4cmH2OabovePEEP
Highpressure: 12to100cmH2O (incrementsof1cmH2O)
Sustainedairwaypressure: Mechanical ventilation ON: Plimit <30cmH2O, thesustainedlimitis6cmH2O Plimit30to60cmH2O, thesustainedlimitis20%ofPlimit
Plimit >
60 cm H2O, thesustainedlimitis12cmH2O
PEEP and mechanical ventilation ON: Sustainedlimitincreasesby PEEPminus2cmH2O
Mechanical ventilation OFF: Plimit ≤ 60 cm H2O, thesustainedlimitis50%ofPlimit Plimit > 60 cm H2O, thesustainedlimitis30cmH2O
Subatmosphericpressure: Paw<-10cmH2O
Alarm silence countdowntimer: 120to0seconds
Ventilator components
Flow transducer
Type: Variableorificeflowsensor
Dimensions: 22mmODand15mmID
Location: Inspiratoryoutletand expiratoryoutlet
(optionalautoclavablesensoravailable)
Oxygen Sensor
Type: Galvanicfuelcell
LifeCycle Approximately12months (Dependentonusage)
Anesthetic agent delivery
Delivery
Vaporizers: Tec6Plus,Tec7
Numberofpositions: 2
Mounting: Tool-freeinstallation Selectatec® manifold interlocksandisolates vaporizers
Electrical specifications
Current leakage
100/120V: <300µA
220/240V: <500µA
Power and battery backup
Powerinput: 100-120Vac,50/60Hz 220-240Vac,50/60Hz
Backuppower: Demonstratedbatterybackup timeundertypicaloperating conditionsis90minuteswhen fullycharged
Batterytype: Internalrechargeablesealed lead acid
Powercord: Length:5m/16.4ftor3m/9.8ft Rating:10A@220Vacor15A @120Vac
Communication port
Serialinterface: RS-232compatibleport
Inlet/outlet modules
120 V
Systemcircuitbreakers: 15A
Outlets: 4outletsonback,3-2A, 1-3Aindividualbreakers with isolation transformer
Pneumatic specifications
Auxiliary common gas outlet
Connector: ISO22mmODand15mmID
Gas supply
Pipelineinputrange: 280kPato600kPa/41psi to87psi
Pipelineconnections: DISS-male,DISS-female, AS4059,FS90-116,SS8752430, orNIST(ISO5359).Allfittings
availableforO2, N2O,andAir, and contain pipeline filter and checkvalve.
Cylinderinput: Pinindexedinaccordancewith CGA-V-1orDIN(nutandgland); containsinputfilterandcheck valve
Note:Maximum3cylinders;two inboardmounted,oneoutboard mounted.
Primaryregulator diaphragmminimum burstpressure: 2758kPa/400psig
Primaryregulator nominaloutput: <338kPa/49psig Pinindexedcylinder connections
<407kPa/59psig DINcylinder connections
O2 controls
Method: Proportionatedecreaseof N2OwithreductioninO2
pressure
Supplyfailurealarm: Range:193kPato221kPa/ 28psigto32psig
Soundsatmaximumvolume every10seconds
O2flush: Range:35to75L/min
Flowmeters
O2ranges: 0.05to0.95L/min and1.0to15.0L/min; MinimumO2flow: 50mL/min±25mL
N2Oranges: 0to0.95L/minand 1.0to10.0L/min
Airrange: 0to0.95and1.0to15.0L/min
Calibration: Percentof Accuracy fullscaleflow (%offlowrate)
100 ±2.5% 90 ±2.5% 80 ±2.6% 70 ±2.7% 60 ±2.9% 50 ±3.1% 40 ±3.4% 30 ±4.0% 20 ±5.0% 10 ±8.1%
Calibrationconditions:*20°C/68°F,101.3kPa/760mmHg
*Differentbreathingcircuitpressures,barometricpressures ortemperatureschangeflowtubeaccuracy.
Total flow sensor
Accuracy: Greaterof±25sccm(smlpm) or ±6.0%ofmeasuredvalue
Calibration 23ºC±2ºC,69.7± 2psiconditions:
Note:Differentbreathingcircuitpressures,barometricpressuresortemperaturechangestotalflowsensoraccuracy
Hypoxic guard system
Type: MechanicalLink-25
Range: Providesanominalminimum25% concentrationofoxygeninO2/N2Omixture
Environmental specifications
System operation
Temperature: 10°to40°C/50°to104°F
Humidity: 15to95%relativehumidity
Altitude: –440to3565m/500to800mmHg
System storage
Temperature: –25°to65°C/–13°to149°F
Humidity: 15to95%relativehumidity
Altitude: –440to5860m/375to800mmHg
Oxygencellstorage: –15°to50°C/5°to122°F
Electromagnetic compatibility
Immunity: Complieswithallrequirements ofEN60601-1-2
Emissions: CISPR11group1classB
Approvals: UL60601-1, CSAC22.2#601.1 EN/IEC60601-1 CE
Breathing circuit specifications
Operational modes
Breathingcircuitiscirclemodeonly
Carbon dioxide absorbent canister
Absorbentcapacity: 800g
Integratedexpiratorylimbwaterreservoír
Ports and connectors
Exhalation: 22mmODISO15mmIDtaper
Inhalation: 22mmODISO15mmIDtaper
Bagport: 22mmOD
Pressure gauge
Scalerange: -2to10kPa/–20to100cmH20
Bag-to-Ventilator switch
Type: Bi-stable
Control: Controlsventilatoranddirectionof breathinggaswithinthecircuit
Integrated Adjustable Pressure Limiting (APL) valve
Range: 0.5to70cmH2O
Tactileknob indicationat: 30cmH2Oandabove
Adjustmentrangeof rotation: 0.5to30cmH2O(0to230°) 30to70cmH2O(230to330°)
Materials
Allmaterialsincontactwithexhaledpatientgasesareautoclavable,exceptdisposableflowsensorsandO2cell.(Autoclavableflowsensorsoptional).
Allmaterialsincontactwithpatientgasarefreeofnaturalrubberlatex.
Breathing circuit parameters
Compliance: Bagmode: 1.82mL/cmH2O
Mechanical mode: Automatically compensates for compression losses within theabsorberandbellows assembly
Circuitvolume: 2.7LVentMode (includingabsorber)
1.2LBagMode
Expiratory PexpBagMode PexpVentMode resistance: Flowrate Pressuredrop Pressuredrop
10L/min 0.78cmH2O 0.77cmH2O
30L/min 1.59cmH2O 1.71cmH2O
60L/min 3.48cmH2O 3.88cmH2O
Note:Withpatientcircuitandwyepieceadd+0.89cmH2O
Anesthetic gas scavenging
System Type Outlet Connector Hospital waste gas disposal system requirements
ActiveAdjustable DISSEVAC 12inHg(305mmHg)minimumvacuum@36LPMflowFlow,HighVacuum ExtractFlowlimitedto30LPM@12inHgvacuum
ActiveLowFlow, DISSEVAC 12inHg(305mmHg)minimumvacuum@36LPMflowHighVacuum
ActiveLowFlow, 12.7mmBarb 36LPMminimumflowLowVacuum
©2011GeneralElectricCompany–AllRightsReserved.
GeneralElectricCompanyreservestherighttomakechangesinspecificationandfeaturesshownherein, ordiscontinuetheproductdescribedatanytimewithout noticeorobligation.ContactyourGErepresentativefor themostcurrentinformation.
GEandGEMonogramaretrademarks ofGeneralElectricCompany.
*Aespire,Tec,andPSVProaretrademarks ofGeneralElectricCompany.
Datex-Ohmeda,Inc.,aGeneralElectriccompany,doingbusinessasGEHealthcare.
About GE Healthcare GE Healthcare provides transformational medical technologiesandservicesthatareshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingandinformationtechnologies,medicaldiagnostics,patientmonitoringsystems,drugdiscovery,biopharmaceuticalmanufacturingtechnologies,performanceimprovementandperformancesolutionsserviceshelpourcustomerstodeliverbettercaretomorepeoplearoundtheworldatalowercost.Inaddition,wepartnerwithhealthcareleaders,strivingtoleveragetheglobalpolicychangenecessarytoimplementasuccessfulshifttosustainablehealthcaresystems.
Our“healthymagination”visionforthefutureinvites theworldtojoinusonourjourneyaswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccessandimprovingqualityaroundtheworld. HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployeesarecommittedtoservinghealthcare professionals and their patients in more than 100countries.FormoreinformationaboutGEHealthcare,visitourwebsiteatwww.gehealthcare.com.
GE Healthcare 3030OhmedaDrive P.O.Box7550 Madison,WI53707-7550 U.S.A.
www.gehealthcare.com
AN-0235-08.11-EN-US DOC0000000