Additional Separation for Ambient Ionisation Mass ... data ultraFAIMS provides a practical means of...

1
Lauren Brown 1 ; Clara Feider 2 ; Ian Mesher 1 ; Alasdair Edge 1 ; Livia S. Eberlin 2 Ambient ionization mass spectrometry has been increasingly used to investigate the distribution of molecules in samples or objects, in situ, without the need for sample preparation. Disadvantages include: Complex biological samples have inherent matrix effects. No pre-separation means analysis cannot be selective. Highly abundant analytes may suppress the detection of lower abundance species. In tandem mass spectrometry, isobaric interferences can result in overlapping fragments which present a significant challenge for spectral interpretation. In-source separation techniques such as field asymmetric ion mobility spectrometry (FAIMS) can be used when chromatographic separation techniques cannot be deployed. FAIMS separates ions according to field-induced changes in their collision cross section (Figure 1). Hypothesized that this is due to the ability to change capillary wall thickness. Tubing wall thickness of 0.013" was used; thinner than the original design. Counteracts potential increased flow from narrower ID. DI2 also allowed the capillary length to be varied with ease. The MRFA signal intensity between DESI, DI1 and two DI2 capillaries (5” and 2”) was compared (Figure 11). The longer DESI-FAIMS capillaries showed little transmission of the doubly charged m/z 262 ion, but the signal was more abundant with the shorter capillary. Possibly due to greater path travelled by the ions resulting in charge reduction. Suggests capillary could be tailored to the application. As the field reverses polarity and changes magnitude, the ion changes direction and speed. Each ion has specific net sideways drift velocity. By applying a DC compensation field (CF), this drift can be corrected, focussing ions through the device. Analogous to atmospheric pressure quadrupole. Highly orthogonal to m/z separation offering a means to increase peak capacity for the analysis of complex samples by ambient ionization mass spectrometry. Can selectively transmit ion of interest prior to mass spectral detection. A recent study 1 demonstrated the use of FAIMS in combination with desorption electrospray ionization (DESI) (Figure 2). Inner diameter (ID) of 0.067” was used to best replicate the standard FAIMS inlet ID. Previously optimised for flow rate through the FAIMS chip for optimal full width at half maximum (FWHM) 2 . Different thickness walls were tested (Figure 6). Having the capillary attached to a Swagelok nut allows it to be rotated upon its axis. Allows adjusting of capillary angle towards stage. When nut is tightened it will lock its orientation. Thinnest walls gave best sensitivity Direction is fixed as a result of being welded Small changes in capillary direction could account for differences in signal intensity. Options: 1. Fix the capillary in the optimal location. 2. Make the direction of the bend adjustable (Figure 7). A modified FAIMS inlet (DI2) was designed to take a 1/16” Swagelok nut (Figure 8). DESI-FAIMS-MS resulted in enhanced mass spectral detection compared with DESI-MS alone. Increased s/n due to suppressed background. Increased absolute signal as trap was preferentially filled with FAIMS-transmitted low abundance ions, rather than naturally high abundance species. Here we describe optimisation of the mechanical integration of a desorption electrospray ionization (DESI) system, with FAIMS. Key goals of the design were to: Deliver a carrier gas velocity profile through the FAIMS chip optimized for high separation performance. Minimize transmission losses. In the first design iteration (DI1), an unheated extended transfer tube of 5” was welded on to the FAIMS inlet (Figure 5). 1. Introduction CF sweeps were acquired by rastering across a mouse kidney section, selected due to its high homogeneity, with both the DI1 and DI2 capillaries. Increased resolution between the doubly and singly charged cardiolipin peaks was achieved using the DI2 capillary due to narrower, more Gaussian FAIMS peaks (Figure 10). 4. DESI-FAIMS data ultraFAIMS provides a practical means of adding additional in-source separation for ambient ionisation mass spectrometry. The optimised DI2 inlet resulted in increased resolution and allowed sensitivity to be optimised via use of different length capillaries and wall thicknesses. Future experiments will look at increasing performance further by investigating different dimension inlets and capillaries, and different capillary lengths and materials. 5. Conclusions To integrate the ultraFAIMS device and Omni Spray source an extension to the transfer capillary is required (Figure 4). It was expected that the two different DESI-FAIMS interfaces may exhibit differences in flow through the FAIMS chip. 1/16” OD capillary will have slightly smaller ID compared to front insert (Figure 9). The narrower ID may produce higher transmission and wider FAIMS peak FWHM, due to an increase in flow rate through FAIMS device. 3. DESI-FAIMS interface designs Figure 1. FAIMS separates ions according to field-induced changes in their collision cross section Figure 2. Schematic representation of selective FAIMS transmission of ion of interest using DESI-FAIMS-MS Figure 3. (a) ultraFAIMS T1 device and (b) installed on Q-Exactive with 2D Omni Spray www.owlstonemedical.com/ultraFAIMS Additional Separation for Ambient Ionisation Mass Spectrometry: The Development of a DESI-FAIMS-MS Interface References 1. Feider, C.L.; Elizondo, N.; Eberlin, L.S.; Anal. Chem. 2016 88 (23) 2. Brown et al.; Optimization and performance characterization of a microscale FAIMS chip coupled to an Orbitrap mass spectrometer. Poster presented at 62 nd ASMS Conference on Mass Spectrometry and Allied Topics, 2014, June 15 - 19, Baltimore, MD 1 Owlstone Medical Ltd., 162 Cambridge Science Park, Cambridge, CB4 OGH, UK, 2 University of Texas at Austin, Austin TX, 78712, USA For further information, email: [email protected] Experiments were carried out using an ultraFAIMS-T1 system (Owlstone Medical Ltd., Cambridge, UK) installed on Q-Exactive and Thermo Elite mass spectrometers (Thermo Scientific, San Jose, CA) with a 2D Omni Spray (Prosolia Inc., Indianapolis, IN) (Figure 3). ultraFAIMS is a miniaturised FAIMS device in which the electrodes are formed from a micro-manufactured chip. 2. Methods ION COLLISION CROSS SECTION VARIES AT HIGH VS LOW DISPERSION FIELDS, GIVING DIFFERENT MOBILITIES Compensation Field 1 Compensation Field 2 Compensation Field 3 Dispersion Field (DF) Electrode Electrode Electrode Electrode Carrier Gas Electrode Electrode 0 Time Low field High field Ion Ion Ions enter channel to detector Ions enter channel to detector Ions enter channel to detector 30.0° ±0.05 Ø 1.6 REAM 0.2 45.0° Capillary extention welded on to FAIMS inlet X 2 POS Ø 4.3 ±0.05 M8 x 1.0 Figure 5. (a) ultraFAIMS T1 device and (b) installed on Q-Exactive with 2D Omni Spray Figure 4. Standard FAIMS inlet design indicating part needing extension Figure 6. Comparison of thickest and thinnest walled capillaries with FAIMS voltages set to transmit cardiolipins Figure 7. Rotation of capillary towards stage via Swagelok nut Figure 8. 1/16” Swagelok nut DESI-FAIMS interface design (DI2) 0.020” 650 700 750 800 850 900 950 1000 m/z 0 20 40 60 80 100 0 20 40 60 80 100 Relative Abundance(%) 885.550 834.530 749.494 738.502 761.494 862.653 724.486 790.540 891.643 713.494 810.529 917.478 749.494 885.550 737.494 834.530 761.494 724.486 773.493 844.644 806.546 891.642 713.494 917.985 m/z 749.4 95 m/z 738.5 02 m/z 761.4 94 m/z 749.4 95 m/z 738.5 02 m/z 761.4 94 Thin (0.011 in.) NL=1.57E4 Thick (0.2 in.) NL=4.64 E3 0 % Relative Abundance 3mm 100 % (DF=220 Td, CF=2.20 Td) 1/16” OD Capillary 1/16” Swagelok nut 1/16” Swage cone Screws onto our piece – no solder required 1/16” OD Capillary will have slightly smaller ID compared to our 02-1519 part 1.6mm ID THRU M8 x 1.0 Ø 3.0 (52.0°) Ø 4.3 ±0.05 30.0° ±0.05° Ø 1.6 THRU 0.2 X 45° 3 POS 0.4 Ra Figure 9. Customised 1/16” Swagelok nut DESI-FAIMS inlet design Figure 11. MRFA signal intensity between DESI, DI1 and 5” and 2” capillary DI2 inlets Figure 10. DESI-FAIMS spectra of doubly and singly charged cardiolipin peaks from DI1 (left) and DI2 (right). CF Sweep Start CF Sweep End RT:0.00-4.27 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 Time (min) 0 10 20 30 40 50 60 70 80 90 100 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance NL: 2.25E2 m/z= 1449.57-1450.39 MS 07252017_oldFAIM S_Kidney1_DF210 CFneg1to3_240se c_IT NL: 1.31E3 724.17-724.65 MS 07252017_oldFAIM S_Kidney1_DF210 CFneg1to3_240se CF=1.38 Td CF=1.71 Td Doubly Charged CL Singly Charged CL CF Sweep Start CF Sweep End CF=1.06 Td RT:0.00-4.15 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 Time (min) 0 10 20 30 40 50 60 70 80 90 100 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance NL: 3.70E2 1449.52-1450.33 S_logcap_Kidney1_ 40sec_IT NL: 7.88E2 MS S_logcap_Kidney1_ 40sec_IT CF=1.94 Td Doubly Charged CL Singly Charged CL All data 210 Td DF, -1 to + Td CF 200 400 600 800 1000 1200 1400 1600 1800 2000 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance 524.3333 262.7500 Regular DESI Capillary NL=4.94E6 200 400 600 800 1000 1200 1400 1600 1800 2000 m/z 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance 262.7500 524.3333 Old DESI-FAIMS Capillary NL=3.59E5 200 400 600 800 1000 1200 1400 1600 1800 2000 m/z 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance 524.3333 262.7500 New DESI-FAIMS Capillary (5in) NL=4.24E5 m/z 200 400 600 800 1000 1200 1400 1600 1800 2000 m/z 0 10 20 30 40 50 60 70 80 90 100 Relative Abundance 524.4167 262.7500 New DESI-FAIMS Capillary (2in) NL=8.64E5 (a) (b) N 2 Moving Stage X Y Extended Capillary FAIMS Cone and Chip Mass Spectrometer m/z Desorbed Ions V Solvent

Transcript of Additional Separation for Ambient Ionisation Mass ... data ultraFAIMS provides a practical means of...

Page 1: Additional Separation for Ambient Ionisation Mass ... data ultraFAIMS provides a practical means of adding additional in-source separation for ambient ionisation mass spectrometry.

Lauren Brown1; Clara Feider2; Ian Mesher1; Alasdair Edge1; Livia S. Eberlin2

Ambient ionization mass spectrometry has been increasingly used to investigate the distribution of molecules in samples or objects, in situ, without the need for sample preparation. Disadvantages include:• Complex biological samples have inherent matrix e�ects.• No pre-separation means analysis cannot be selective.• Highly abundant analytes may suppress the detection of lower

abundance species.• In tandem mass spectrometry, isobaric interferences can result in

overlapping fragments which present a significant challenge for spectral interpretation.

In-source separation techniques such as field asymmetric ion mobility spectrometry (FAIMS) can be used when chromatographic separation techniques cannot be deployed.• FAIMS separates ions according to field-induced changes in their

collision cross section (Figure 1).

Hypothesized that this is due to the ability to change capillary wall thickness. • Tubing wall thickness of 0.013" was used; thinner than the original

design.• Counteracts potential increased flow from narrower ID.DI2 also allowed the capillary length to be varied with ease.• The MRFA signal intensity between DESI, DI1 and two DI2

capillaries (5” and 2”) was compared (Figure 11).

The longer DESI-FAIMS capillaries showed little transmission of the doubly charged m/z 262 ion, but the signal was more abundant with the shorter capillary. • Possibly due to greater path travelled by the ions resulting in

charge reduction. • Suggests capillary could be tailored to the application.

• As the field reverses polarity and changes magnitude, the ion changes direction and speed.

• Each ion has specific net sideways drift velocity.• By applying a DC compensation field (CF), this drift can be

corrected, focussing ions through the device.• Analogous to atmospheric pressure quadrupole.• Highly orthogonal to m/z separation o�ering a means to increase

peak capacity for the analysis of complex samples by ambient ionization mass spectrometry.

• Can selectively transmit ion of interest prior to mass spectral detection.

A recent study1 demonstrated the use of FAIMS in combination with desorption electrospray ionization (DESI) (Figure 2).

• Inner diameter (ID) of 0.067” was used to best replicate the standard FAIMS inlet ID.

• Previously optimised for flow rate through the FAIMS chip for optimal full width at half maximum (FWHM)2.

• Di�erent thickness walls were tested (Figure 6).

• Having the capillary attached to a Swagelok nut allows it to be rotated upon its axis.

• Allows adjusting of capillary angle towards stage.

• When nut is tightened it will lock its orientation.

• Thinnest walls gave best sensitivity• Direction is fixed as a result of being welded• Small changes in capillary direction could account for di�erences

in signal intensity.Options:1. Fix the capillary in the optimal location.2. Make the direction of the bend adjustable (Figure 7).

A modified FAIMS inlet (DI2) was designed to take a 1/16” Swagelok nut (Figure 8).

• DESI-FAIMS-MS resulted in enhanced mass spectral detection compared with DESI-MS alone.

• Increased s/n due to suppressed background.• Increased absolute signal as trap was preferentially filled with

FAIMS-transmitted low abundance ions, rather than naturally high abundance species.

Here we describe optimisation of the mechanical integration of a desorption electrospray ionization (DESI) system, with FAIMS.

Key goals of the design were to:• Deliver a carrier gas velocity profile through the FAIMS chip

optimized for high separation performance. • Minimize transmission losses.In the first design iteration (DI1), an unheated extended transfer tube of 5” was welded on to the FAIMS inlet (Figure 5).

1. Introduction

CF sweeps were acquired by rastering across a mouse kidney section, selected due to its high homogeneity, with both the DI1 and DI2 capillaries.Increased resolution between the doubly and singly charged cardiolipin peaks was achieved using the DI2 capillary due to narrower, more Gaussian FAIMS peaks (Figure 10).

4. DESI-FAIMS data

ultraFAIMS provides a practical means of adding additional in-source separation for ambient ionisation mass spectrometry.The optimised DI2 inlet resulted in increased resolution and allowed sensitivity to be optimised via use of different length capillaries and wall thicknesses. Future experiments will look at increasing performance further by investigating different dimension inlets and capillaries, and different capillary lengths and materials.

5. Conclusions

To integrate the ultraFAIMS device and Omni Spray source an extension to the transfer capillary is required (Figure 4).

It was expected that the two di�erent DESI-FAIMS interfaces may exhibit di�erences in flow through the FAIMS chip.• 1/16” OD capillary will have slightly smaller ID compared to front

insert (Figure 9).

• The narrower ID may produce higher transmission and wider FAIMS peak FWHM, due to an increase in flow rate through FAIMS device.

3. DESI-FAIMS interface designs

Figure 1. FAIMS separates ions according to field-induced changes in their collision cross section

Figure 2. Schematic representation of selective FAIMS transmission of ion of interest using DESI-FAIMS-MS

Figure 3. (a) ultraFAIMS T1 device and (b) installed on Q-Exactive with 2D Omni Spray

www.owlstonemedical.com/ultraFAIMS

Additional Separation for Ambient Ionisation Mass Spectrometry:The Development of a DESI-FAIMS-MS Interface

References1. Feider, C.L.; Elizondo, N.; Eberlin, L.S.; Anal. Chem. 2016 88 (23)2. Brown et al.; Optimization and performance characterization of a

microscale FAIMS chip coupled to an Orbitrap mass spectrometer. Poster presented at 62nd ASMS Conference on Mass Spectrometry and Allied Topics, 2014, June 15 - 19, Baltimore, MD

1Owlstone Medical Ltd., 162 Cambridge Science Park, Cambridge, CB4 OGH, UK, 2University of Texas at Austin, Austin TX, 78712, USAFor further information, email: [email protected]

Experiments were carried out using an ultraFAIMS-T1 system (Owlstone Medical Ltd., Cambridge, UK) installed on Q-Exactive and Thermo Elite mass spectrometers (Thermo Scientific, San Jose, CA) with a 2D Omni Spray (Prosolia Inc., Indianapolis, IN) (Figure 3).ultraFAIMS is a miniaturised FAIMS device in which the electrodes are formed from a micro-manufactured chip.

2. Methods

ION COLLISION CROSS SECTION VARIES AT HIGH VSLOW DISPERSION FIELDS, GIVING DIFFERENT MOBILITIES

CompensationField 1

CompensationField 2

CompensationField 3

DispersionField (DF)

Electrode

Electrode

Electrode

Electrode

Carrier Gas

Electrode

Electrode

0 Time

Low field

High field

Ion

Ion

Ions enterchannel

todetector

Ions enterchannel

todetector

Ions enterchannel

todetector

30.0° ±0.05

Ø 1

.6R

EA

M

0.2 45.0°

Capillary extention weldedon to FAIMS inlet

X 2 POS

Ø 4

.3 ±

0.0

5

M8 x 1.0

Figure 5. (a) ultraFAIMS T1 device and (b) installed on Q-Exactive with 2D Omni Spray

Figure 4. Standard FAIMS inlet design indicating part needing extension

Figure 6. Comparison of thickest and thinnest walled capillaries with FAIMS voltages set to transmit cardiolipins

Figure 7. Rotation of capillary towards stage via Swagelok nut

Figure 8. 1/16” Swagelok nut DESI-FAIMS interface design (DI2)

0.020”

650 700 750 800 850 900 950 1000m/z

0

20

40

60

80

100

0

20

40

60

80

100

Rel

ativ

e A

bun

dan

ce(%

)

885.550

834.530

749.494

738.502761.494

862.653724.486 790.540

891.643

713.494

810.529

917.478

749.494

885.550737.494 834.530

761.494

724.486

773.493 844.644806.546

891.642

713.494917.985

m/z749.4

95

m/z738.5

02

m/z761.4

94

m/z749.4

95

m/z738.5

02

m/z761.494

Thin (0.011 in.) NL=1.57E 4

Thick (0.2 in.)

NL=4.64E 3

0% Relative

Abundance 3 mm

100%

(DF=220 Td, CF=2.20 Td)

1/16” OD Capillary1/16” Swagelok nut

1/16” Swage cone

Screws onto ourpiece – no solderrequired

1/16” OD Capillary willhave slightly smaller IDcompared to our02-1519 part

1.6mm ID THRU

M8 x 1.0

Ø

3.0

(52

.0°)

Ø 4

.3 ±

0.0

5

30.0° ±0.05°

Ø 1

.6 T

HR

U 0.2 X 45°

3 POS

0.4 Ra

Figure 9. Customised 1/16” Swagelok nut DESI-FAIMS inlet design

Figure 11. MRFA signal intensity between DESI, DI1 and 5” and 2” capillary DI2 inlets

Figure 10. DESI-FAIMS spectra of doubly and singly charged cardiolipin peaks from DI1 (left) and DI2 (right).

CF Sweep Start

CF Sweep End

RT:0.00-4.27

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2

Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

NL: 2.25E2m/z= 1449.57-1450.39 MS 07252017_oldFAIMS_Kidney1_DF210CFneg1to3_240sec_IT

NL: 1.31E3724.17-724.65 MS 07252017_oldFAIMS_Kidney1_DF210CFneg1to3_240se

CF=1.38 Td

CF=1.71 Td

Doubly Charged

CL

Singly Charged

CL

CF Sweep Start

CF Sweep End

CF=1.06 TdRT:0.00-4.15

0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0

Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

NL: 3.70E21449.52-1450.33 S_logcap_Kidney1_40sec_IT

NL: 7.88E2MS S_logcap_Kidney1_40sec_IT

CF=1.94 Td

Doubly Charged

CL

Singly Charged

CL

All data 210 Td DF, -1 to + Td CF

200 400 600 800 1000 1200 1400 1600 1800 20000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bun

dan

ce 524.3333

262.7500Regular DESI CapillaryNL=4.94E6

200 400 600 800 1000 1200 1400 1600 1800 2000

m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bun

dan

ce

262.7500

524.3333Old DESI-FAIMS CapillaryNL=3.59E5

200 400 600 800 1000 1200 1400 1600 1800 2000

m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bun

dan

ce

524.3333

262.7500

New DESI-FAIMS Capillary (5in)NL=4.24E5

m/z

200 400 600 800 1000 1200 1400 1600 1800 2000m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bun

dan

ce

524.4167

262.7500

New DESI-FAIMS Capillary (2in)NL=8.64E5

(a) (b)

N2

MovingStage

X

Y

Extended Capillary

FAIMS Coneand Chip Mass Spectrometer

m/z

DesorbedIons

VSolvent