![Page 1: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/1.jpg)
1
Adaptive Acquisition in Real-Time Cardiac MRI
Partly funded by NIH HLBI Grant No. R21HL62336-02
and NSF Grant BES 02-01876
Yoram BreslerECE, BioEng, Coordinated Science Lab
University of Illinois at Urbana-Champaign
Work with Students:
Nitin Aggarwal, Qi Zhao, Saptarshi Bondyapadhyay
IPAM Workshop HM2006IPAM Workshop HM2006
![Page 2: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/2.jpg)
2
Cardiac MRI : Applications and Challenges
Applications :
Needs :
Grand Challenge : 4D (5Grand Challenge : 4D (5--6D?!) cardiac imaging6D?!) cardiac imaging
• Higher spatial resolution in 3D
• Higher temporal resolution
• Imaging without breathholding
• Real-time
• Spectroscopic, diffusion imaging
1. Vascular disease and tissue characterization
a. Visualization of coronary arteries
b. Coronary plaque characterization
2. Functional assessment of the ventricles
3. Myocardial dynamics, perfusion, and viability
4. Valvular function
![Page 3: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/3.jpg)
3
Cardiac MRI : Problems & “Solutions”
Problems :• Insufficient MR data acquisition speed for Nyquist sampling
• Difficult to account for two distinct motion processes :
Heartbeat
Approaches:
• Faster data acquisition sequences limited speed-ups:
– Hardware (gradients, field strength)– Energy deposition and peripheral nervous stimulation– MR physics
• Cardio-respiratory gating time-averaged snapshots; long
• Breathholding patient cooperation; limited time-window
• Reduced data acquisition methods (Keyhole, Rigr, Dime, etc.)
Respiration
![Page 4: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/4.jpg)
4
Overview
Aim:
• Produce a real-time, time-resolved cardiac cine
No breath holding
No cardio-respiratory gating
Approach:
• Model-based, minimum-redundancy
![Page 5: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/5.jpg)
5
To Model or Not to Model?To Model or Not to Model?
The truth, only the truth, and all the truth.But, a finite number of k-t space samples tells nothing at all!
Dynamic Imaging is an ill-posed inverse problem
Modeling in dynamic imaging is unavoidable
Everyone models -- whether they admit it or not
![Page 6: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/6.jpg)
6
Rules for SafeRules for Safe®® ModelingModeling
Don’t overdoBe explicitAnalyze robustness to deviation from modeling assumptionsLearn model from the imaged subject data
![Page 7: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/7.jpg)
7
Overview
Aim:
Produce a time-resolved cardiac cine without breathholding or cardio-respiratory gating
Approach:A. Use spatio-temporal (2D+t, 3D+t) formulation
B. Model both respiratory and cardiac motion
C. Adapt model to imaged subject
D. Adapt MR data acquisition to model
E. Tailor reconstruction to A - D
![Page 8: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/8.jpg)
8
MR Imaging (Cartesian) and TSS(Cartesian) and TSS
Imaging Equation : ( )( ( ) ) ( , ) −= ∫ i t
F O V
s t I t e k rk r d r
MR Data ObjectSpatial Fourier Transform
Time-Sequential (TS) Constraint : Only 1 k-space line can be acquired at a time
k- space sampling MR image (static object)
kx
ky
t=t0
t=t1
.
.
.
.
t=t2 Spatial Fourier Transform
![Page 9: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/9.jpg)
9
MR Imaging : Dynamic Object
Desired image size = 128*128Max. temporal frequency = 10 Hz
Nyquist sampling rate = 2560 lines/si.e. 0.4 msRT ≈
Conventional Imaging (TR=2ms)
Artifacts are due to temporal undersampling
Correct mathematical formulation is in the spatio-temporal domain
– Objects are functions in (r,t) domain– Sampling occurs in (k,t) domain– Sampling can be analyzed in the dual (r,f)
domain
For unique reconstruction, object I(r,t) must be restricted to a model class, M
![Page 10: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/10.jpg)
10
Dynamic Imaging: DefinitionDynamic Imaging: DefinitionGiven a class of spatio-temporal signals with :
Spatial ROI : & Temporal ROI :
( , )g y t ∈MM
{ }y S∈ { }t∈T
Find a sampling schedule so that is recoverable from the corresponding samples.
{ }( , ), ,g y t y t∈ ∈S T{ },n nk tΨ =
Fourier Transform
f
B yk
t
S
y
Dual k-t (D-k-t) Support (essential) :
2 ( )supp ( , ) yj k y ftg y t e dydtπ− +⎧ ⎫⎨ ⎬⎩ ⎭∫
S T
B ==
![Page 11: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/11.jpg)
11
Cardiac Image models (Phenomenological)
Characteristics :• The (highly) dynamic portion of the object (i.e heart) is
spatially localized within the field-of-view.
• The cardiac motion is quasi-periodic
• Even with breathhold thorax is not static
Model :
Object to be imaged :
Spatio-temporal spectrum :
Dual k-t (D-k-t) support :
( , )I tr2( , ) ( , ) −= ∫ i f tI f I t e dtπr r
supp{ ( , )}I frB
y
k
(cm)x(cm)
t(Hz)f (Hz)
==
![Page 12: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/12.jpg)
12
TSS Problem
Given model class (= D-k-t support) find a time-sequential
sampling schedule from which I(r,t) can be reconstructed to
prescribed resolution.
tΔ
N
tΔ
N
tΔ
N
k
(a)
t
N
k
(c)
t
k
(b)
t
N
Instantaneous Progressive TS Scrambled TS
ky ky ky
![Page 13: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/13.jpg)
13
TSS Problem
Questions :• Design of sampling schedule Ψ?
• Conditions on sampling rate ?
• Reconstruction from acquired samples ?
Properties :
• Aliasing in TSS determined not only by density of points (sampling rate) in Ψ, but also by the temporal order in which points are visited [Allebach,1987]
• Order of acquisition need not be determined by the adjacency of k-space sample locations
• Optimization of Ψ is a very difficult combinatorial problem. Need to explore 256! possible orderings for 256 sample locations.
• Solution: Unified TSS theory ( Willis & Bresler, IEEE Trans. IT,1997)
![Page 14: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/14.jpg)
14
Time-Sequential Sampling Theory
![Page 15: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/15.jpg)
15
Key Idea : Consider TSS schedules that lie on a lattice
Results:
Sampling pattern design through constrained geometric packing of B, S.
Reconstruction of g(y,t) through linear filtering
Bounds on achievable TS sampling rates
In practice, sampling rate reduced by large factor compared to conventional sampling schemes
TSS Theory (Willis & Bresler, 1992, 1997)
![Page 16: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/16.jpg)
16
Lattice TheoryLattice Theory
Lattice : A lattice ΛA is the set :
where ai (in Rn) are linearly independent
Basis Matrix : A basis matrix of ΛA is A = [a1,a2, …]
Fact : Every (rational) lattice has a basis matrix of the form
Polar Lattice : The polar lattice of a lattice ΛA is a latticeΛA* with basis matrix A* = A-T
1{ : }n
i i iim m
=
Λ = ∈∑A a Z
•a1
a2
0 R
D sT
⎛ ⎞= ⎜ ⎟⎝ ⎠
A
•
•
•
••
•
•
![Page 17: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/17.jpg)
17
Fact : If signal g is sampled on lattice ΛA then it’s spectrum is replicated in the frequency domain on the polar lattice ΛA*
Signal g can be recovered from it’s samples iff the replicas do not overlap
Lattice ΛA* packs the spectral support B of g
Notation : ΛA ε R(B)
Multidimensional Sampling TheoryMultidimensional Sampling Theory
![Page 18: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/18.jpg)
18
TimeTime--Sequential Sampling on a Sequential Sampling on a LatticeLattice
If sample points (kn ,tn ) are chosen to lie on a lattice ΛA, in general, the time-sequential constraint will not be met.
However there may be only one sample point in S at any given sampling time instant, i.e ΛA is time-sequential w.r.t. S
Notation : ΛA ε T(S)
Result : ΛA ε T(S) if lattice spanned by D (i.e. ΛD), packs S
DTR
k
t0 R
D sT
⎛ ⎞= ⎜ ⎟⎝ ⎠
A
S2
S1
•S1
ΛD + S1
•
•
•
•
••
•
•
• •
k
![Page 19: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/19.jpg)
19
Find : Lattice sampling schedule ΛA ={kn,tn}n such that :
1. Any signal in M can be recovered from the samples
2. Schedule is time-sequential and temporally uniform w.r.t S (with time period TR),
3. TR is maximized
Solution:
Solution computed by searching for lattices subject to the “dual” packing constraints :
1.ΛA* packs the dual k-t support B2. (ΛD , S) tiles R
Sampling Schedule DesignSampling Schedule Design
; ( ) ( ) ( )0
arg max
R
RD s
T
T⎛ ⎞
= Λ ∈ ∩ ∩⎜ ⎟⎝ ⎠
AA R B T S U S
![Page 20: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/20.jpg)
20
Reconstruction and ResultsReconstruction and Results
Reconstruction method :
Filter the samples with filter with frequency response
Performance bounds for TSS :
Asymptotically and in practice – for optimum schedule :
• Bounds achievable
• No penalty for restriction to lattice patterns !
• No penalty for time-sequential constraint !
( ) ( ), ,tH y f y fχ= B
1 1( , )( ( )) ( ) ( ) ( )R
crit
Td volume volume volume
Λ ≤ ≤Λ ⋅ ⋅
SB S B S
opt
prog
T ( ( ))Speed Gain factor = GT ( )
volume bounding boxvolume
≤B
B==
![Page 21: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/21.jpg)
21
Adaptive Imaging SchemeAdaptive Imaging Scheme
Pilot-scan Acquisition
Spectral Analysis
Dual-k-t support Adaptation of TS Sampling
Pattern
TS Sampling pattern
Imaging Acquisition
Image Reconstructioncine
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
Pre-Imaging
y
k
(cm)x(cm)
(Hz)t
![Page 22: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/22.jpg)
22
Support of I(kx,y,kt)
kx – phase-encodeky - readout
f (Hz)
y (cm)
t (ms)
ky (1/cm)
I(kx,ky,t)
Spectral Analysis
A. PreA. Pre--imaging Acquisitionimaging Acquisition
Aim : estimate the D-k-t support
Method : For each kx …
{ ( , , )}x
x tk
supp I k y k=UB
•• Acquisition of DAcquisition of D--kk--tt support is easy!support is easy!
![Page 23: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/23.jpg)
23
B. Adptation of TS Imaging
Given :•D-k-t support •Desired spatial resolution (kmax)
Find : Minimum rate TS sampling pattern Ψ = {k(n),nTR}n from which I(r,t) can be reconstructed
f (Hz)
y (cm)
ky (1/cm)
Lattice TS Sampling Pattern
Lattice Packing of Subject to TSS Constraint
t (ms)
B
B
![Page 24: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/24.jpg)
24
MRI ExperimentMRI ExperimentObject: Gel phantom moved by a stepping motor.
Motion:
• Driven by phase of ECG. HRV = 63-68 BPM
• Fundamental frequency scaled to
about 0.2Hz
STS Model:
• 10 harmonics, bandwidth 1/20 f0• Nyquist sampling period = 1ms
Acquisition: VARIAN/SISCO 4.7T imaging spectrometer
• 146 sec acquisition interval (29 “heart beats”)
• TSS Acquisition: Tr= 17.8 ms
• Progressive Acquisition: Tr=8 ms
0 5 10 15 20 25 30 35 40 45 5060
61
62
63
64
65
66
67
68
Heart−beat
Inst
anta
neou
s H
eart
Rat
e (b
pm)
![Page 25: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/25.jpg)
25
Experimental ResultExperimental Result
Progressive sampling
(Tr=8 ms)
TS Lattice Sampling
(Tr=18 ms)
Resolution: 256 x 256 (0.4x0.4 mm )Resolution: 256 x 256 (0.4x0.4 mm )
![Page 26: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/26.jpg)
26
In Vivo MR Experiment In Vivo MR Experiment (Preliminary)(Preliminary)
≈Nyq( 0.7ms)RT
Time-Sequential Imaging(TR=3.5ms)
Conventional Imaging (TR=3.5ms)
UngatedUngated AcquisitionAcquisition
![Page 27: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/27.jpg)
27
In Vivo MR Experiment In Vivo MR Experiment (Preliminary)(Preliminary)
Time-Sequential Imaging(TR=3.5ms)
Conventional Imaging (TR=3.5ms) ≈Nyq( 0.7ms)RT
••UngatedUngated AcquisitionAcquisition
![Page 28: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/28.jpg)
28
DiscussionDiscussionImaging scheme compatible with:– Cartesian (Zhao et al 2001, Bresler 2002, Aggarwal
et al 2002), radial k-space sampling (Willis & Bresler, 1995)
– 3DReconstruction method : – Filtering with interpolating LSI filter (ibid)– Fitting of a parametric spatio-temporal model
(Zhao et al 2002)Lattice optimality criterion can be modified to consider :– TR,– SNR or CNR,– Hardware or physiological constraints
![Page 29: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/29.jpg)
29
Respiratory Motion
![Page 30: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/30.jpg)
30
Effect of Respiratory Motion: AnalysisEffect of Respiratory Motion: AnalysisModel as smooth spatial co-ordinate warp:
Δ(·,t) lowpass (respiratory drift in breathhold) :broadening of temporal harmonic bands
around nf0
Δ(·,t) periodic (free-breathing)creation of new (temporal) spectral bands
around nf0+mfm
∂= ≈ +
∂( , )
( , ) (Δ( , ), ) ( , ) Δ( , )I y t
G y t I y t t I y t y ty
∂≈ + ⊗
∂( , )
( , ) ( , ) Δ( , )f
I y fG y f I y f y f
y
f0: fundamental frequency of cardiac motion ; fm: fundamental frequency of respiratory motion
![Page 31: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/31.jpg)
31
for periodic Δ(f)for low-pass Δ(f)
Effect of Respiratory Motion: Confirmation
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 50
0.5
1
1.5
Hz
I(k,f)G(k,f)
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 50
0.5
1
1.5
Hz
I(k,f)G(k,f)
0 0an ( , )( , ) d yy G kI k ff0 0an ( , )( , ) d yy G kI k ff
![Page 32: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/32.jpg)
32
Effect on TS Imaging
Respiratory motion increases the D-k-t support
Required sampling rate increases
-- OR --
Respiratory motion artifacts created
Ground Truth Uncorrected TS Imaging
![Page 33: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/33.jpg)
33
G(r,t) : Actual cardiac time-varying image
I(r,t) : Respiration-free cardiac image Modeled by D-k-t Model
: Time-varying scaling and rotation matrix
: Time-varying displacement vector
R3 3P( )t ×∈3 1( )q t R ×∈
Affine Respiratory Motion Model
(r, ) [P( )r ( ), ]G t I t q t t= +
Properties :
• Dominates respiratory motion of heart walls, arteries and adjoining tissue
• Subsumes translation, expansion and rotation
![Page 34: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/34.jpg)
34
Affine Corrected Sampling Pattern
2 ( )( , ) | ( )| [ ( ) , ]Tj πk q t TI k t P t e G P t k t−= i i
• Measurement Equation :
Ψ { ( ), }TSR nk n nT=• I(r,t) can be reconstructed from samples at
Γ { ( ) ( ), }T TSR R nP nT k n nT=
I(r,t) can be reconstructed from samples of G(k,t) at
2 ( )1( , ) [ ( ) , ]
| ( )|Tj πk q t TG k t e I P t k t
P t−=
• Reconstruction Equation :
kxkx
Amplitude Correction
Phase Correction
![Page 35: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/35.jpg)
35
Imaging SchemeImaging Scheme
STS Support
Estimation
STS Support TS Sampling Pattern Design
TS Sampling pattern Ψ
Imaging Acquisition
Image Reconstruction
(LSI Filter; IFFT)
cine
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
20 40 60 80 100 120
20
40
60
80
100
120
Pre-Imaging
y
k
(cm)x(cm)
(Hz)t
Respiratory Adaptation
P, q
P, q
ACTS Sampling pattern Γ
![Page 36: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/36.jpg)
36
Estimation of Affine Motion
Respiratory Model Calibration [Manke:2002]:
Estimate the subject-specific relationship between affine motion parameters (P,q) and the respiratory phase η
Respiratory Model
Calibration
Affine Parameter Estimation
Respiratory Phase η
P, qP(η) q(η)
Diaphragmatic navigator/Respiratory bellows
Pre-Imaging
![Page 37: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/37.jpg)
37
Simulation : Affine Respiratory Motion
Affine Respiratory Motion
Simulation Parameters :
• Respiratory Motion : Affine†
• Respiratory rate = 16 breaths/min
• Cardiac rate = 60 beats/min
• Optimal TR = 7.1 ms
• Total acquisition time = 30s
† : Parameters adapted from Manke:2002
![Page 38: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/38.jpg)
38
Simulation : Cardiac + Affine Respiratory Motion
Affine-Corrected TS Imaging
Ground Truth
Uncorrected TS Imaging
![Page 39: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/39.jpg)
39
Modeling Robustness: Non-Affine Respiratory Motion
Non-Affine Respiratory
Motion
Simulation Parameters :
• Respiratory Motion :
• Spine is stationary
• Expansion motion for rib-cage
• Affine motion for region surrounding the heart
• Respiratory rate = 16 breaths/min
• Cardiac rate = 60 beats/min
• Optimal TR = 7.1 ms
• Total acquisition time = 30s
![Page 40: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/40.jpg)
40
Simulation : Cardiac + Non-Affine Respiratory Motion
Affine-Corrected TS Imaging
Ground Truth
Uncorrected TS Imaging
![Page 41: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/41.jpg)
41
Simulation : Cardiac + Non-Affine Respiratory Motion
Affine-Corrected TS Imaging
Uncorrected TS Imaging
![Page 42: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/42.jpg)
42
Heart Rate Variation:Cardiac Aperiodicity
![Page 43: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/43.jpg)
43
SpatioSpatio--Temporal Spectrum Model and AperiodicityTemporal Spectrum Model and Aperiodicity
-Regions with significant temporal variation spatially localized
-Cardiac motion is quasi-periodic
Spectral bands
-Spectral bands broadened due to aperiodicity of dynamics
Higher sampling rate required
y
k
(cm)x(cm)
t(Hz)
![Page 44: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/44.jpg)
44
TimeTime--WarpWarp & HRV& HRV
Time-Warp Model :
I(x,y,t) : Time-varying cardiac image
G(x,y,t) : Idealized time-varying cardiac image with narrow-bandedD-k-t support
Φ(t) : Time-warp; assumed monotonic, slowly varying
( , , ) ( , ,Φ( ))I x y t G x y t=
y
k
(cm)x(cm)
(Hz)t
Explicitly accounting for timeExplicitly accounting for time--warp narrows Dwarp narrows D--kk--tt support support
Lowers sampling rate requiredLowers sampling rate required
![Page 45: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/45.jpg)
45
TimeTime--WarpWarp in Actionin Action
Temporal Spectrum
(before de-warping)
Temporal Spectrum (after de-warping)
ECGECG
••DewarpDewarp spectral analysisspectral analysis
••Adapt sampling to Adapt sampling to dewarpeddewarped signalsignal
••DewarpDewarp acquired signalacquired signal
•• Reconstruct from nonuniform samplesReconstruct from nonuniform samples
•• DewarpingDewarping uses uses dynamic dynamic programmingprogramming
![Page 46: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/46.jpg)
46
Discussion
• Real-time free-breathing cardiac MR imaging schemewithout ECG or respiratory gating (Aggarwal et al 2004.)
• Method robust to respiratory motion modeling inaccuracies.
• Adaptive scheme is tolerant of aperiodicity & heart rate variability (Aggarwal et al 2002, Zhao et al 2002.)
• About 5x-20x reduction in sampling rate vs. Nyquist
• Improvements in sampling rate can be traded for increased spatial/temporal resolution, 3D, or SNR/CNR
• Generalizable to higher-dimensions, alternate k-space sampling trajectories
• Complementary to fast acquisition sequences or parallel MR imaging
• Similar techniques for other dynamic imaging applications
![Page 47: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/47.jpg)
47
ConclusionConclusion
Study sampling theoryModel the dataBe adaptive in– Modeling– Acquisition– Reconstruction
Analyze & predict performance
![Page 48: Adaptive Acquisition in Real-Time Cardiac MRIhelper.ipam.ucla.edu/publications/hm2006/hm2006_6111.pdf · Cardiac MRI : Applications and Challenges Applications : ÎNeeds : Grand Challenge](https://reader036.fdocuments.in/reader036/viewer/2022080720/5f79a486c423d03c24357593/html5/thumbnails/48.jpg)
48
Thank you for your attention!