AC_PFT420_WR_02_0901.doc Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH. Misuse of texts, pictures or company logo will be prosecuted. Badger Meter Europa GmbH Nürtinger Str. 76 72639 Neuffen (Germany) Tel. +49-7025-9208-0 Fax +49-7025-9208-15 www.badgermeter.de [email protected] Flow transmitter Type PFT-420 Wiring analog output Calibration The meters are delivered calibrated. If the flow data change, the analog signal can be adjusted as follows: 1) Remove cover 2) Connect multimeter in series to the red wire 3) Connect pulse counter to pulse output (black, white) 4) Adjust max. flow 5) Adjust 20 mA at span potentiometer 6) Add pulses and calculate the K-factor Nullpunkt Reedschalter Eingang Zähler

description

manual tecnico flujometro

Transcript of AC_PFT420_WR_02_0479

Page 1: AC_PFT420_WR_02_0479

AC_PFT420_WR_02_0901.doc

Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH. Misuse of texts, pictures or company logo wil l be prosecuted.

Badger Meter Europa GmbH Nürtinger Str. 76 72639 Neuffen (Germany) Tel. +49-7025-9208-0 Fax +49-7025-9208-15 www.badgermeter.de [email protected]

Flow transmitter Type PFT-420 Wiring analog output

Calibration

The meters are delivered calibrated. If the flow data change, the analog signal can be adjusted as follows:

1) Remove cover 2) Connect multimeter in series to the red wire 3) Connect pulse counter to pulse output (black, white) 4) Adjust max. flow 5) Adjust 20 mA at span potentiometer 6) Add pulses and calculate the K-factor

Nullpunkt

Reedschalter

Eingang Zähler