Achieving Peak Performance withQ Exactive Instrument Software 2.2 SP1
1
The world leader in serving science
Thermo Fisher Scientific, Bremen, Germany
-
New Instrument Control SW 2.2 - Overview
General• AGC improvements for UHPLC
• Exact Mass Calculator
• Centroid mode
• Intelligent beam management
Tune
2
• Calibration report
• Direct LC control (Accela/Open AS)
• Auto Source settings
• Quadrupole Transmission Test
• Sweep Gas is turned on in Standby Mode
DIA• Advanced scan functions with Data Independent Analysis
-
General Improvements
3
The world leader in serving science
1. AGC Improvements for Rapid Chromatography
2. Exact Mass Calculator
3. Data Acquisition: Centroid Mode
4. Intelligent Beam Management
-
1. AGC Improvements
4
The world leader in serving science
-
AGC Improvements
Problem Statement:
• In fast UHPLC-MS/(MS) runs AGC was sometimes too „old“ compared to change of ion population:
� „saturation curve“ at high concentrations
� Bad precision and accuracy at high concentrations
5
-
Example: AGC (IC-SW 2.1)
10
20
30
40
50
60
70
80
90
100
Rela
tive A
bundance
A
B
A
In some cases of fast changes of the ion flux (high concentration & UHPLC)
6
2.80 2.85 2.90 2.95 3.00 3.05 3.10
Time (min)
0
10
20
30
40
50
60
70
80
90
100
Rela
tive A
bundance
0
B
~ 15%Diff area (A vs B)
-
New AGC Concepts: What is Different for the User?
7
Upon the expected chromatographic peak width and the number of active targets taken from inclusion list information, the system is automatically choosing the AGC modes used for getting the best reference.
-
New AGC Results: t-HCD Alprazolam Product – MS2 Scan-to-Scan Mode
RT: 2.4 - 2.6
60
80
100
Re
lative A
bun
dance
0
20
40
60
80
100
Re
lative A
bu
ndance
triplicates
8
2.40 2.45 2.50 2.55
Time (min)
0
20
40
60
80
100
Re
lati
ve A
bun
dance
0
20
40
Re
lative A
bun
dance
Much better reproducibility and “mapping” of the peak (~ 3% Diff)
-
2. Exact Mass Calculator
9
The world leader in serving science
-
Exact Mass Calculator
Q Exactive is purchased and sold more into „routine“applications than in research environments.
People comming from nominal mass instrumentation often struggle with accurate mass calculation or find this tedious.
�Currently, this creates a barrier for routine adoption of HR/AM
10
HR/AM
Main requirements:
� automated solution based on elemental composition
� im-/export functionality from/to Excel
� Availability in Tune and Method Editor
-
Exact Mass Calculator
TUNE Method Editor
11
-
Exact Mass Calculator
12
Lists can be exported to or imported from Excel:
-
3. Centroid mode
13
The world leader in serving science
-
Centroid Mode
Results in very large data files
• Small molecule routine labs are used to centroids
• post-processing not an option in regulated environments
� Centroid data acquisition enabled in TUNE & ME
14
-
Centroid Mode in the Method Setup
15
-
Availability Through Method Editor
16
-
Availability Through Tune
Instrument Status Tree
Advanced user Role
17
“right-click”
-
Data Volume Reduction
0
1000
2000
3000
4000
5000
6000
7000
8000
SIM 17k SIM 35k SIM 70k
Centroid
Profi le
0
20000
40000
60000
80000
100000
120000
140000
160000
180000
FS 17k FS 35k FS 70k
Centroid
Profi le
Da
ta V
olu
me
[kB
]
Da
ta V
olu
me
[kB
]
18
• This data is from direct infusion of calmix for 5 min by using a FS and a tSIMexperiment.
• As a rule of thumb, data volume is reduced by a factor of 3 for FS experiments.
-
Where is This Data Volume Reduction Coming From?
In-spectrum stick plot:7 data points / peak
In-spectrum point-to-point plot
19
In-spectrum point-to-point plot(default view)
In-spectrum stick plot (default view):1 data point / peak
-
Why not Factor 7 Reduction?
• Raw data file size is dependent on all data which is written to the file (e.g. scan header information)
• The ratio of all this additional information is relatively high for QE data
20
-
4. Intelligent Beam Management
21
The world leader in serving science
-
Intelligent Ion Beam Management
• The Q-Exactive, like all other Q-TOF/TRAP devices uses the quadrupole to rapdily select ions of interest for further mass selection
• When there are large amounts of matrix material injected into the mass spectrometer, the quadrupole needs to
HCD cell C-TrapQuadrupoleMass Filter
OrbitrapMass Analyzer
S-lensIon Source
22
spectrometer, the quadrupole needs to filter the vast majority of these ions away.
• As part of this filtering processing, a large percentage of the filtered ions will be deposited on the rod set. This is true for ALL quadrupole designs.
-
Intelligent Ion Beam Management
• The Exactive Series 2.2SP1 software works to minimize the amount of time that the quadrupole needs to work filtering ions
• Becuase the Q-Exactive is a pulsed device, once the C-Trap is filled and waiting for injection of ions into the
HCD cell C-TrapQuadrupoleMass Filter
OrbitrapMass Analyzer
S-lensIon Source
23
waiting for injection of ions into the orbitrap, there is no need to continue to filter ions.
• The new software built into 2.2SP1 intellegently optimizes the filling/filtering routines to give maximum signal and minimize filtering times.
-
Early Software – Constant Filter Mode
Full MSHCD
1
HCD
2
HCD
3
HCD
4
HCD
5
HCD
6
HCD
7
HCD
8
HCD
9
HCD
10
265 ms 64 ms
Full MS
split lens:open according to injection times
24
Quad in fullscan =wide isolation range
• With the early software design, the quadrupole started filtering as soon as the previous scan was complete
• Filtered ions were almost always hitting the rods.
Quad operating in isolation mode
Quad
-
New in SW 2.2: RF-only Mode
Full MSHCD
1
HCD
2
HCD
3
HCD
4
HCD
5
HCD
6
HCD
7
HCD
8
HCD
9
HCD
10Full MS
split lens:open according to injection times
25
Quad operating in isolation mode
• Quad is operating in isolation mode only during injection times. • Once the C-Trap is filled, the quadrupole switches to RF-only. • Only during injection times excluded peptides hit the rods. In RF-only the complete
ion beam is forwarded through the quad and deffered at the split lens
Quad operating in RF-only mode
Quad in fullscan =wide isolation range
-
New in SW 2.2: RF-only Mode
Full MSHCD
1
HCD
2
HCD
3
HCD
4
HCD
5
HCD
6
HCD
7
HCD
8
HCD
9
HCD
10Full MS
split lens:open according to injection times
26
Quad operating in isolation mode
• NOTE: When the target value is set ultra-high, the ratio of filling to scanning time is such that the quads will be in filter mode for a very high percentage of the time.
• Do high target values really help? See next slide
Quad operating in RF-only mode
Quad in fullscan =wide isolation range
-
Hela Digest (1µg, 1% FDR, 75 min), Average (Triplicates)
1500
2000
2500
3000
3500
Ax
is T
itle
Protein Groups
1500
2000
2500
3000
3500
Ax
is T
itle
Protein Groups
5e4 1e5 1e6
Num
ber
of
Pro
tein
Gro
ups
27
50000 100000 1000000
Protein Groups 3082 3138 3089
0
500
1000
50000 100000 1000000
Protein Groups 3082 3138 3089
0
500
1000
For many high-throughput, high load proteomics workflows, a larger target value provides
no extra IDs.
Always use the lowest target value possible. This is the easiest way to increase uptime.
Num
ber
of
Pro
tein
Gro
ups
-
Intelligent Ion Beam Management
• Of course, even with the improved beam management and lower target values, there may be cases of customers running “extreme” sample conditions.
• The new software now includes a test to determine if the Q1 has become
HCD cell C-TrapQuadrupoleMass Filter
OrbitrapMass Analyzer
S-lensIon Source
28
to determine if the Q1 has become contaminated (see next slide).
• For these types of workflows, the Thermo Scientific service organization offers a number of very affordable options for frequent Q1 cleaning.
-
New Tool to Check for Q1 Charging Status
• Tune: Advance Mode
• Evaluation/Extra Evaluation/ Isolation Transmission Endurance Test
• Start evaluation with:
• Evaluating with Postive Mode
29
• Evaluating with Postive Mode Calibration solution takes 16mins
• Stay spray TIC variation < 10%
• Fresh clean Calibration solution
-
New Tool to Check for Q1 Charging Status (cont.)
• How to read the Evaluation Result
• The evaluation result are presented as a Transmission Score
• A Transmission Score close to 1 indicates a clean Q
30
• A Transmission Score less than 1 indicates contamination of the quad
• When the score is < 0.5 combined with a 20% decrease of protein IDs from a QC standard, the quad should be cleaned!
-
Changes in the Tune Window
31
The world leader in serving science
1. Calibration Reports
2. Direct LC Control
3. Auto Source Settings
-
1. Calibration Reports
Direct link to pdf document
32
• All reports are stored automatically • The customer can choose between
• whole calibration report (includes all calibration procedures which have been updated
on a selected date)
• custom spectral mass calibration
• spectral mass calibration (pos or neg)
• list of all reports
-
Mass Calibration Report
Calibration History
Calibration
Header
33
Calibration Values
Calibration Plot
-
2. LC Direct Control from Tune
• Access via „LC Direct Control“ icon or „Menu/Windows“ if an Accela LC instrument was configured
34
-
LC Direct Control from Tune: Accela Systems
• Set toggle „Take pump under control“ for Accela Pump Direct Control. Check Status in Xcalibur Sequence Editor „Direct Control“.
35
Tune Instrument Direct Control Xcalibur Sequence Editor
-
Direct Control for LC Systems
• Supported Thermo systems:
☺Accela Pump 1250
☺Accela Autosampler Open AS
☺CTC PAL Autosampler
• Thermo systems not supported yet for TUNE direct control:
36
�Proxeon EASY-nLC 1000
�Dionex Ultimate 3000 systems
• Third party instruments are not supported yet
-
3. Source Gases - Autodefaults
• Default source voltage and gas settings are applied according to flow rate
37
-
Example: HESI Default Settings
38
Default settings available for APCI as well.Default values can be looked up in the online help.
-
Data Independent Analysis
39
The world leader in serving science
1. Data Independent Acquisition (DIA)
2. MSX-DIA
-
DIA − High Throughput Comprehensive Quantification Acquisition Methods with Qualitative Confirmation
• In DIA (data-independent acquisition) experiments, a set mass range is pre-defined that corresponds to predominant precursor m/z range for enzymatic peptides. MS/MS data are collected repeatedly until all precursor ions in the defined mass range are selected for fragmentation, yielding MS/MS spectra on all precursor ions.
• By acquiring MS/MS data for all precursor ions in each sample, DIA seeks to increase reproducibility and comprehensiveness of data collection within different samples. No detailed sample knowledge is required prior to the DIA-based analysis.
40
• Data collection produces a complete record of quantitative data and a targeted data analysis strategy can be employed to mine additional analytes, retrospectively relying on MS/Ms spectral library.
• The new instrument control software 2.2 on Q Exactive offers two different DIA approaches: DIA and MSX-DIA (multiplexed DIA).
-
Targeted MS/MS – „Classic“ Approach for Quantification
60
70
80
90
100R
ela
tive
Ab
un
da
nce
684.340
☺ highly specific☺ very fast� Misses most of the mass range� Quan-Query for prior defined compounds
☺ highly specific☺ very fast� Misses most of the mass range� Quan-Query for prior defined compounds
41
300 400 500 600 700 800 900 1000 1100 1200
m/z
0
10
20
30
40
50
Re
lative
Ab
un
da
nce
488.279
877.897640.797
760.329599.837
786.879470.735
541.275
360.705
1001.935 1105.478
945.397
-
AIF – All-in Approach for Quantification
60
70
80
90
100R
ela
tive
Ab
un
da
nce
684.340
☺ Covers entire mass range☺ Fast duty cycle☺ Quan-Query for every compound
in mass range� limited specificity and selectivity� limited dynamic range
☺ Covers entire mass range☺ Fast duty cycle☺ Quan-Query for every compound
in mass range� limited specificity and selectivity� limited dynamic range
42
300 400 500 600 700 800 900 1000 1100 1200
m/z
0
10
20
30
40
50
Re
lative
Ab
un
da
nce
488.279
877.897640.797
760.329599.837
786.879470.735
541.275
360.705
1001.935 1105.478
945.397
-
DIA on Q Exactive
• Thermo – Data Independent Acquisition (DIA) on Q Exactive
• DIA method with wider isolation window (up to 50Da) sequentially scanning through the entire mass range
• multiplexed DIA (msxDIA) method using narrow isolation window width and multiplexing MS2 in a random fashion
• Advantages of DIA on Q Exactive
• Capability to use higher resolution set up for more accurate quantitative results by
43
• Capability to use higher resolution set up for more accurate quantitative results by resolving analytes from interferences.
• Flexibility to use different combination of mass range and resolution to adapt different research needs.
• Easy methods set up with new Method Editor.
-
DIA – Exhaustive Approach for Quantification
60
70
80
90
100R
ela
tive
Ab
un
da
nce
684.340
☺ covers entire mass range☺ Quan-Query for every
compounds within mass range� less specific� Medium dynamic range� Low duty cycle
☺ covers entire mass range☺ Quan-Query for every
compounds within mass range� less specific� Medium dynamic range� Low duty cycle
44
300 400 500 600 700 800 900 1000 1100 1200
m/z
0
10
20
30
40
50
Re
lative
Ab
un
da
nce
488.279
877.897640.797
760.329599.837
786.879470.735
541.275
360.705
1001.935 1105.478
945.397
-
MSX-DIA– New Approach to Combine the Advantages*
60
70
80
90
100R
ela
tive
Ab
un
da
nce
684.340
☺ highly specific☺ covers entire mass range☺ Quan-Query for every
compounds within mass range☺ deconvolution
☺ highly specific☺ covers entire mass range☺ Quan-Query for every
compounds within mass range☺ deconvolution
45
300 400 500 600 700 800 900 1000 1100 1200
m/z
0
10
20
30
40
50
Re
lative
Ab
un
da
nce
488.279
877.897640.797
760.329599.837
786.879470.735
541.275
360.705
1001.935 1105.478
945.397
☺ deconvolution� dyn. range limited by inj. time� low duty cycle (2,5-4 sec)
☺ deconvolution� dyn. range limited by inj. time� low duty cycle (2,5-4 sec)
*Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1
-
HR/AM DIA Method
Software 2.2
HR/AM DIA MethodInstrument Control
Software 2.2
Protein Digest
MS/MS Spectral Library
DevelopmentUsing Pinpoint
DIA and msxDIA Workflow on Q Exactive
DIA
46
Targeted QuanData Extraction Using PinpointMultiplexed DIA
(msxDIA)
Predefined include lists and data processing is not part of Method Editor. This will be provided by third party software packages like Skyline.
-
Setup DIA in ME
47
-
MSX-DIA Setup
48
-
Appendix
49
The world leader in serving science
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
50
Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1,1University of Washington, Seattle, WA; 2Thermo Fisher Scientific, Bremen, GERMANY; 3University of Pittsburgh, Pittsburgh, PA; 4ThermoFisher Scientific, San Jose, CA
Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1,1University of Washington, Seattle, WA; 2Thermo Fisher Scientific, Bremen, GERMANY; 3University of Pittsburgh, Pittsburgh, PA; 4ThermoFisher Scientific, San Jose, CA
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
51
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
52
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
53
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
54
Scan 3
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
55
Scan 3
Scan 20
. . .
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
56
Scan 3
Scan 20
. . .
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
57
Scan 3
Scan 20
. . .
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
58
Scan 3
Scan 20
. . .
Scan 21
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
59
Scan 3
Scan 20
. . .
Scan 21
-
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
60
Scan 3
Scan 20
. . .
Scan 21
-
Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)
Scan 1
100 4 m/z-wide windows = 400 m/z
m/z500 900
Scan 2
Scan 3
61
Scan 3
Scan 20
. . .
Scan 21
-
Inte
nsit
y
Deconvolution
62
m/z
Inte
nsit
y
-
MSX – DIA
63
Demultiplexing improves results !