Achieving Peak Performance with Q Exactive Instrument Software 2.2 SP1 1 The world leader in serving science Thermo Fisher Scientific, Bremen, Germany

Transcript of Achieving Peak Performance with Q Exactive Instrument ... ›...

  • Achieving Peak Performance withQ Exactive Instrument Software 2.2 SP1

    1

    The world leader in serving science

    Thermo Fisher Scientific, Bremen, Germany

  • New Instrument Control SW 2.2 - Overview

    General• AGC improvements for UHPLC

    • Exact Mass Calculator

    • Centroid mode

    • Intelligent beam management

    Tune

    2

    • Calibration report

    • Direct LC control (Accela/Open AS)

    • Auto Source settings

    • Quadrupole Transmission Test

    • Sweep Gas is turned on in Standby Mode

    DIA• Advanced scan functions with Data Independent Analysis

  • General Improvements

    3

    The world leader in serving science

    1. AGC Improvements for Rapid Chromatography

    2. Exact Mass Calculator

    3. Data Acquisition: Centroid Mode

    4. Intelligent Beam Management

  • 1. AGC Improvements

    4

    The world leader in serving science

  • AGC Improvements

    Problem Statement:

    • In fast UHPLC-MS/(MS) runs AGC was sometimes too „old“ compared to change of ion population:

    � „saturation curve“ at high concentrations

    � Bad precision and accuracy at high concentrations

    5

  • Example: AGC (IC-SW 2.1)

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    Rela

    tive A

    bundance

    A

    B

    A

    In some cases of fast changes of the ion flux (high concentration & UHPLC)

    6

    2.80 2.85 2.90 2.95 3.00 3.05 3.10

    Time (min)

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    Rela

    tive A

    bundance

    0

    B

    ~ 15%Diff area (A vs B)

  • New AGC Concepts: What is Different for the User?

    7

    Upon the expected chromatographic peak width and the number of active targets taken from inclusion list information, the system is automatically choosing the AGC modes used for getting the best reference.

  • New AGC Results: t-HCD Alprazolam Product – MS2 Scan-to-Scan Mode

    RT: 2.4 - 2.6

    60

    80

    100

    Re

    lative A

    bun

    dance

    0

    20

    40

    60

    80

    100

    Re

    lative A

    bu

    ndance

    triplicates

    8

    2.40 2.45 2.50 2.55

    Time (min)

    0

    20

    40

    60

    80

    100

    Re

    lati

    ve A

    bun

    dance

    0

    20

    40

    Re

    lative A

    bun

    dance

    Much better reproducibility and “mapping” of the peak (~ 3% Diff)

  • 2. Exact Mass Calculator

    9

    The world leader in serving science

  • Exact Mass Calculator

    Q Exactive is purchased and sold more into „routine“applications than in research environments.

    People comming from nominal mass instrumentation often struggle with accurate mass calculation or find this tedious.

    �Currently, this creates a barrier for routine adoption of HR/AM

    10

    HR/AM

    Main requirements:

    � automated solution based on elemental composition

    � im-/export functionality from/to Excel

    � Availability in Tune and Method Editor

  • Exact Mass Calculator

    TUNE Method Editor

    11

  • Exact Mass Calculator

    12

    Lists can be exported to or imported from Excel:

  • 3. Centroid mode

    13

    The world leader in serving science

  • Centroid Mode

    Results in very large data files

    • Small molecule routine labs are used to centroids

    • post-processing not an option in regulated environments

    � Centroid data acquisition enabled in TUNE & ME

    14

  • Centroid Mode in the Method Setup

    15

  • Availability Through Method Editor

    16

  • Availability Through Tune

    Instrument Status Tree

    Advanced user Role

    17

    “right-click”

  • Data Volume Reduction

    0

    1000

    2000

    3000

    4000

    5000

    6000

    7000

    8000

    SIM 17k SIM 35k SIM 70k

    Centroid

    Profi le

    0

    20000

    40000

    60000

    80000

    100000

    120000

    140000

    160000

    180000

    FS 17k FS 35k FS 70k

    Centroid

    Profi le

    Da

    ta V

    olu

    me

    [kB

    ]

    Da

    ta V

    olu

    me

    [kB

    ]

    18

    • This data is from direct infusion of calmix for 5 min by using a FS and a tSIMexperiment.

    • As a rule of thumb, data volume is reduced by a factor of 3 for FS experiments.

  • Where is This Data Volume Reduction Coming From?

    In-spectrum stick plot:7 data points / peak

    In-spectrum point-to-point plot

    19

    In-spectrum point-to-point plot(default view)

    In-spectrum stick plot (default view):1 data point / peak

  • Why not Factor 7 Reduction?

    • Raw data file size is dependent on all data which is written to the file (e.g. scan header information)

    • The ratio of all this additional information is relatively high for QE data

    20

  • 4. Intelligent Beam Management

    21

    The world leader in serving science

  • Intelligent Ion Beam Management

    • The Q-Exactive, like all other Q-TOF/TRAP devices uses the quadrupole to rapdily select ions of interest for further mass selection

    • When there are large amounts of matrix material injected into the mass spectrometer, the quadrupole needs to

    HCD cell C-TrapQuadrupoleMass Filter

    OrbitrapMass Analyzer

    S-lensIon Source

    22

    spectrometer, the quadrupole needs to filter the vast majority of these ions away.

    • As part of this filtering processing, a large percentage of the filtered ions will be deposited on the rod set. This is true for ALL quadrupole designs.

  • Intelligent Ion Beam Management

    • The Exactive Series 2.2SP1 software works to minimize the amount of time that the quadrupole needs to work filtering ions

    • Becuase the Q-Exactive is a pulsed device, once the C-Trap is filled and waiting for injection of ions into the

    HCD cell C-TrapQuadrupoleMass Filter

    OrbitrapMass Analyzer

    S-lensIon Source

    23

    waiting for injection of ions into the orbitrap, there is no need to continue to filter ions.

    • The new software built into 2.2SP1 intellegently optimizes the filling/filtering routines to give maximum signal and minimize filtering times.

  • Early Software – Constant Filter Mode

    Full MSHCD

    1

    HCD

    2

    HCD

    3

    HCD

    4

    HCD

    5

    HCD

    6

    HCD

    7

    HCD

    8

    HCD

    9

    HCD

    10

    265 ms 64 ms

    Full MS

    split lens:open according to injection times

    24

    Quad in fullscan =wide isolation range

    • With the early software design, the quadrupole started filtering as soon as the previous scan was complete

    • Filtered ions were almost always hitting the rods.

    Quad operating in isolation mode

    Quad

  • New in SW 2.2: RF-only Mode

    Full MSHCD

    1

    HCD

    2

    HCD

    3

    HCD

    4

    HCD

    5

    HCD

    6

    HCD

    7

    HCD

    8

    HCD

    9

    HCD

    10Full MS

    split lens:open according to injection times

    25

    Quad operating in isolation mode

    • Quad is operating in isolation mode only during injection times. • Once the C-Trap is filled, the quadrupole switches to RF-only. • Only during injection times excluded peptides hit the rods. In RF-only the complete

    ion beam is forwarded through the quad and deffered at the split lens

    Quad operating in RF-only mode

    Quad in fullscan =wide isolation range

  • New in SW 2.2: RF-only Mode

    Full MSHCD

    1

    HCD

    2

    HCD

    3

    HCD

    4

    HCD

    5

    HCD

    6

    HCD

    7

    HCD

    8

    HCD

    9

    HCD

    10Full MS

    split lens:open according to injection times

    26

    Quad operating in isolation mode

    • NOTE: When the target value is set ultra-high, the ratio of filling to scanning time is such that the quads will be in filter mode for a very high percentage of the time.

    • Do high target values really help? See next slide

    Quad operating in RF-only mode

    Quad in fullscan =wide isolation range

  • Hela Digest (1µg, 1% FDR, 75 min), Average (Triplicates)

    1500

    2000

    2500

    3000

    3500

    Ax

    is T

    itle

    Protein Groups

    1500

    2000

    2500

    3000

    3500

    Ax

    is T

    itle

    Protein Groups

    5e4 1e5 1e6

    Num

    ber

    of

    Pro

    tein

    Gro

    ups

    27

    50000 100000 1000000

    Protein Groups 3082 3138 3089

    0

    500

    1000

    50000 100000 1000000

    Protein Groups 3082 3138 3089

    0

    500

    1000

    For many high-throughput, high load proteomics workflows, a larger target value provides

    no extra IDs.

    Always use the lowest target value possible. This is the easiest way to increase uptime.

    Num

    ber

    of

    Pro

    tein

    Gro

    ups

  • Intelligent Ion Beam Management

    • Of course, even with the improved beam management and lower target values, there may be cases of customers running “extreme” sample conditions.

    • The new software now includes a test to determine if the Q1 has become

    HCD cell C-TrapQuadrupoleMass Filter

    OrbitrapMass Analyzer

    S-lensIon Source

    28

    to determine if the Q1 has become contaminated (see next slide).

    • For these types of workflows, the Thermo Scientific service organization offers a number of very affordable options for frequent Q1 cleaning.

  • New Tool to Check for Q1 Charging Status

    • Tune: Advance Mode

    • Evaluation/Extra Evaluation/ Isolation Transmission Endurance Test

    • Start evaluation with:

    • Evaluating with Postive Mode

    29

    • Evaluating with Postive Mode Calibration solution takes 16mins

    • Stay spray TIC variation < 10%

    • Fresh clean Calibration solution

  • New Tool to Check for Q1 Charging Status (cont.)

    • How to read the Evaluation Result

    • The evaluation result are presented as a Transmission Score

    • A Transmission Score close to 1 indicates a clean Q

    30

    • A Transmission Score less than 1 indicates contamination of the quad

    • When the score is < 0.5 combined with a 20% decrease of protein IDs from a QC standard, the quad should be cleaned!

  • Changes in the Tune Window

    31

    The world leader in serving science

    1. Calibration Reports

    2. Direct LC Control

    3. Auto Source Settings

  • 1. Calibration Reports

    Direct link to pdf document

    32

    • All reports are stored automatically • The customer can choose between

    • whole calibration report (includes all calibration procedures which have been updated

    on a selected date)

    • custom spectral mass calibration

    • spectral mass calibration (pos or neg)

    • list of all reports

  • Mass Calibration Report

    Calibration History

    Calibration

    Header

    33

    Calibration Values

    Calibration Plot

  • 2. LC Direct Control from Tune

    • Access via „LC Direct Control“ icon or „Menu/Windows“ if an Accela LC instrument was configured

    34

  • LC Direct Control from Tune: Accela Systems

    • Set toggle „Take pump under control“ for Accela Pump Direct Control. Check Status in Xcalibur Sequence Editor „Direct Control“.

    35

    Tune Instrument Direct Control Xcalibur Sequence Editor

  • Direct Control for LC Systems

    • Supported Thermo systems:

    ☺Accela Pump 1250

    ☺Accela Autosampler Open AS

    ☺CTC PAL Autosampler

    • Thermo systems not supported yet for TUNE direct control:

    36

    �Proxeon EASY-nLC 1000

    �Dionex Ultimate 3000 systems

    • Third party instruments are not supported yet

  • 3. Source Gases - Autodefaults

    • Default source voltage and gas settings are applied according to flow rate

    37

  • Example: HESI Default Settings

    38

    Default settings available for APCI as well.Default values can be looked up in the online help.

  • Data Independent Analysis

    39

    The world leader in serving science

    1. Data Independent Acquisition (DIA)

    2. MSX-DIA

  • DIA − High Throughput Comprehensive Quantification Acquisition Methods with Qualitative Confirmation

    • In DIA (data-independent acquisition) experiments, a set mass range is pre-defined that corresponds to predominant precursor m/z range for enzymatic peptides. MS/MS data are collected repeatedly until all precursor ions in the defined mass range are selected for fragmentation, yielding MS/MS spectra on all precursor ions.

    • By acquiring MS/MS data for all precursor ions in each sample, DIA seeks to increase reproducibility and comprehensiveness of data collection within different samples. No detailed sample knowledge is required prior to the DIA-based analysis.

    40

    • Data collection produces a complete record of quantitative data and a targeted data analysis strategy can be employed to mine additional analytes, retrospectively relying on MS/Ms spectral library.

    • The new instrument control software 2.2 on Q Exactive offers two different DIA approaches: DIA and MSX-DIA (multiplexed DIA).

  • Targeted MS/MS – „Classic“ Approach for Quantification

    60

    70

    80

    90

    100R

    ela

    tive

    Ab

    un

    da

    nce

    684.340

    ☺ highly specific☺ very fast� Misses most of the mass range� Quan-Query for prior defined compounds

    ☺ highly specific☺ very fast� Misses most of the mass range� Quan-Query for prior defined compounds

    41

    300 400 500 600 700 800 900 1000 1100 1200

    m/z

    0

    10

    20

    30

    40

    50

    Re

    lative

    Ab

    un

    da

    nce

    488.279

    877.897640.797

    760.329599.837

    786.879470.735

    541.275

    360.705

    1001.935 1105.478

    945.397

  • AIF – All-in Approach for Quantification

    60

    70

    80

    90

    100R

    ela

    tive

    Ab

    un

    da

    nce

    684.340

    ☺ Covers entire mass range☺ Fast duty cycle☺ Quan-Query for every compound

    in mass range� limited specificity and selectivity� limited dynamic range

    ☺ Covers entire mass range☺ Fast duty cycle☺ Quan-Query for every compound

    in mass range� limited specificity and selectivity� limited dynamic range

    42

    300 400 500 600 700 800 900 1000 1100 1200

    m/z

    0

    10

    20

    30

    40

    50

    Re

    lative

    Ab

    un

    da

    nce

    488.279

    877.897640.797

    760.329599.837

    786.879470.735

    541.275

    360.705

    1001.935 1105.478

    945.397

  • DIA on Q Exactive

    • Thermo – Data Independent Acquisition (DIA) on Q Exactive

    • DIA method with wider isolation window (up to 50Da) sequentially scanning through the entire mass range

    • multiplexed DIA (msxDIA) method using narrow isolation window width and multiplexing MS2 in a random fashion

    • Advantages of DIA on Q Exactive

    • Capability to use higher resolution set up for more accurate quantitative results by

    43

    • Capability to use higher resolution set up for more accurate quantitative results by resolving analytes from interferences.

    • Flexibility to use different combination of mass range and resolution to adapt different research needs.

    • Easy methods set up with new Method Editor.

  • DIA – Exhaustive Approach for Quantification

    60

    70

    80

    90

    100R

    ela

    tive

    Ab

    un

    da

    nce

    684.340

    ☺ covers entire mass range☺ Quan-Query for every

    compounds within mass range� less specific� Medium dynamic range� Low duty cycle

    ☺ covers entire mass range☺ Quan-Query for every

    compounds within mass range� less specific� Medium dynamic range� Low duty cycle

    44

    300 400 500 600 700 800 900 1000 1100 1200

    m/z

    0

    10

    20

    30

    40

    50

    Re

    lative

    Ab

    un

    da

    nce

    488.279

    877.897640.797

    760.329599.837

    786.879470.735

    541.275

    360.705

    1001.935 1105.478

    945.397

  • MSX-DIA– New Approach to Combine the Advantages*

    60

    70

    80

    90

    100R

    ela

    tive

    Ab

    un

    da

    nce

    684.340

    ☺ highly specific☺ covers entire mass range☺ Quan-Query for every

    compounds within mass range☺ deconvolution

    ☺ highly specific☺ covers entire mass range☺ Quan-Query for every

    compounds within mass range☺ deconvolution

    45

    300 400 500 600 700 800 900 1000 1100 1200

    m/z

    0

    10

    20

    30

    40

    50

    Re

    lative

    Ab

    un

    da

    nce

    488.279

    877.897640.797

    760.329599.837

    786.879470.735

    541.275

    360.705

    1001.935 1105.478

    945.397

    ☺ deconvolution� dyn. range limited by inj. time� low duty cycle (2,5-4 sec)

    ☺ deconvolution� dyn. range limited by inj. time� low duty cycle (2,5-4 sec)

    *Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1

  • HR/AM DIA Method

    Software 2.2

    HR/AM DIA MethodInstrument Control

    Software 2.2

    Protein Digest

    MS/MS Spectral Library

    DevelopmentUsing Pinpoint

    DIA and msxDIA Workflow on Q Exactive

    DIA

    46

    Targeted QuanData Extraction Using PinpointMultiplexed DIA

    (msxDIA)

    Predefined include lists and data processing is not part of Method Editor. This will be provided by third party software packages like Skyline.

  • Setup DIA in ME

    47

  • MSX-DIA Setup

    48

  • Appendix

    49

    The world leader in serving science

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    50

    Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1,1University of Washington, Seattle, WA; 2Thermo Fisher Scientific, Bremen, GERMANY; 3University of Pittsburgh, Pittsburgh, PA; 4ThermoFisher Scientific, San Jose, CA

    Poster: ThP26 ASMS 2012 Jarrett Egertson1; Andreas Kuehn2; Gennifer Merrihew1; Nicholas Bateman3; Brendan Maclean1; Jesse D. Canterbury4; Markus Kellmann2; Vlad Zabrouskov4; Christine Wu 3; Michael J. Maccoss1,1University of Washington, Seattle, WA; 2Thermo Fisher Scientific, Bremen, GERMANY; 3University of Pittsburgh, Pittsburgh, PA; 4ThermoFisher Scientific, San Jose, CA

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    51

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    52

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    53

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    54

    Scan 3

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    55

    Scan 3

    Scan 20

    . . .

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    56

    Scan 3

    Scan 20

    . . .

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    57

    Scan 3

    Scan 20

    . . .

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    58

    Scan 3

    Scan 20

    . . .

    Scan 21

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    59

    Scan 3

    Scan 20

    . . .

    Scan 21

  • Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    60

    Scan 3

    Scan 20

    . . .

    Scan 21

  • Multiplexed DIA (Example Mass Range 500-900, iso. 4Da, Multiplexing Count 5)

    Scan 1

    100 4 m/z-wide windows = 400 m/z

    m/z500 900

    Scan 2

    Scan 3

    61

    Scan 3

    Scan 20

    . . .

    Scan 21

  • Inte

    nsit

    y

    Deconvolution

    62

    m/z

    Inte

    nsit

    y

  • MSX – DIA

    63

    Demultiplexing improves results !