ZETOM-CERT ISO 9001 INSTRUMENTATION FOR SOUND & VIBRATION MEASUREMENTS AND ANALYSIS SVAN 958 FOUR CHANNELS SOUND & VIBRATION LEVEL METER & ANALYSER (draft version) USER’S MANUAL USER’S MANUAL USER’S MANUAL USER’S MANUAL SVANTEK Sp. z o.o. WARSAW, April 2008

Transcript of Acelerômetro Svan 958 - Manual de Instruções

  • ZETOM-CERT

    ISO 9001

    INSTRUMENTATION FOR SOUND & VIBRATION

    MEASUREMENTS AND ANALYSIS

    SVAN 958

    FOUR CHANNELS SOUND & VIBRATION LEVEL METER & ANALYSER

    (draft version)

    USERS MANUALUSERS MANUALUSERS MANUALUSERS MANUAL

    SVANTEK Sp. z o.o. WARSAW, April 2008

  • SVAN 958 USER MANUAL _

    2

    Notice: This users manual presents the software revision named 3.01.0 / 3.05.1 (cf. the description of the UNIT LABEL path: MENU / DISPLAY / UNIT LABEL). The succeeding software revisions (marked with the bigger numbers) can slightly change the view of some displays presented in the text of the manual.

  • SVAN 958 USER MANUAL

    i

    CONTENTS

    1 INTRODUCTION 1-1 1.1 SVAN 958 as Sound Level Meter & Analyser 1-1 1.2 SVAN 958 as Vibration Meter & Analyser 1-2 1.3 General features of SVAN 958 1-2 1.4 Accessories included 1-3 1.5 Accessories available 1-3

    2 MANUAL CONTROL OF THE INSTRUMENT 2-1 2.1 Control push-buttons on the front panel 2-1 2.2 Input and output sockets of the instrument 2-6

    3 SETTING THE INSTRUMENT 3-1 3.1 Basis of the instruments control 3-1 3.2 Powering of the instrument 3-25 3.3 Initial setup of the instrument 3-27 3.4 Activation of optional functions 3-32 3.5 Memory organisation 3-33

    4 FUNCTIONS OF THE INSTRUMENT - FUNCTION 4-1 4.1 Measurement functions of the instrument - MEASUREMENT FUNCTION 4-1 4.2 Instruments calibration - CALIBRATION 4-2

    4.2.1 Calibration by the sensitivity introduction 4-2 4.2.2 Calibration by the measurement 4-5 4.2.3 View on the last calibration result - CALIBRATION HISTORY 4-8

    5 MEASUREMENT PARAMETERS SETTING - INPUT 5-1 5.1 Selection of measurement parameters - MEASUREMENT SETUP 5-2

    5.1.1 Setting time delay before the start of measurements - START DELAY 5-2 5.1.2 Setting integration period - INT. PERIOD 5-3 5.1.3 Setting the number of repetition of measurement cycles - CYCLES NUMBER 5-4 5.1.4 Setting time period between two writings to the loggers file - LOGGER STEP 5-4

    5.2 Setting parameters in channels - CHANNELS SETUP 5-6 5.2.1 Selection of measurement mode for sound or vibration - MODE 5-6 5.2.2 Measurement range setting - RANGE 5-7 5.2.3 Weighting filter selection in a profile - FILTER 5-7 5.2.4 RMS detector selection - DETECTOR 5-9 5.2.5 Setting the parameters of microphone (SM) - MICROPHONE CORRECTION 5-9

    5.3 Results selection for saving in a loggers file - LOGGER SETUP 5-10 5.3.1 Selecting results to be saved in instruments logger memory - LOGGER MODE 5-10 5.3.2 Selecting time-history results to be saved in the memory - LOGGER MODE: ON 5-10 5.3.3 Selecting parameters of low sampling rate time-domain signal recording (svn

    format) in memory of the instrument - LOGGER MODE: TIME 5-11

  • SVAN 958 USER MANUAL _

    ii

    5.4 Triggering mode and parameters selection - TRIGGER SETUP 5-12 5.4.1 Switching the triggering on and off - TRIGGER 5-12 5.4.2 Selection of the triggering signal - SOURCE 5-13 5.4.3 Selection of channel for triggering condition - CHANNEL 5-14 5.4.4 Setting the level of the triggering signal - LEVEL 5-14 5.4.5 Setting the speed of the triggering signal changes - GRADIENT 5-15 5.4.6 Setting the start of time-triggered measurement - RTC START 5-15 5.4.7 Setting the step for repetition of time-triggered measurement - RTC STEP 5-15

    5.5 Setting parameters of auxiliary functions - AUXILIARY SETUP 5-16 5.5.1 Setting the parameters of RPM function - RPM SETUP 5-16

    5.5.1.1 Enabling the RPM measurement - ENABLED 5-16 5.5.1.2 Selecting the number of pulses / rotations - PULSES / ROTATION 5-17 5.5.1.3 Selecting the unit of RPM measurement - UNIT 5-17 5.5.1.4 Activation of logger for RPM measurements - LOGGER 5-17

    5.5.2 Setting the parameters of attenuation measurements - SEAT SETUP 5-18 5.5.2.1 Enabling the attenuation measurements - ENABLED 5-18 5.5.2.2 Selection of the seat channel for attenuation measurements -

    SEAT CHANNEL 5-19 5.5.2.3 Selection of the base channel for attenuation measurements -

    BASE CHANNEL 5-19 5.5.3 Settings for vector calculations - VECTOR SETUP 5-19 5.5.4 Setting the parameters for dose measurements - HAV/WBV DOSE SETUP 5-21

    5.5.4.1 Enabling HAV/WBV settings - ENABLED 5-21 5.5.4.2 Setting the time of exposure - EXPOSURE TIME 5-22 5.5.4.3 Selection of channels for HAV/WBV DOSE calculation - X AXIS, Y AXIS,

    Z AXIS 5-22 5.5.4.4 Standards selection for HAV/WBV dose calculation - STANDARDS 5-22

    5.6 Selection of parameters for alarm triggering - ALARM SETUP 5-23 5.7 Selection of dosimeter parameters - DOSIMETER SETUP 5-26

    5.7.1 Setting the exposure time - EXPOSURE TIME 5-26 5.7.2 Setting criterion sound level - CRITERION LEVEL 5-26 5.7.3 Setting threshold level - THRESHOLD LEVEL 5-27 5.7.4 Setting exchange rate - EXCHANGE RATE 5-27

    5.8 Selection of FFT analysis parameters - FFT SETUP 5-27 5.8.1 Enabling the FFT analysis - ENABLED 5-28 5.8.2 Selecting the weighting filter during the FFT analysis - FILTER 5-28 5.8.3 Selecting the analysis band of the signal - BAND 5-28 5.8.4 Selecting the time window for the FFT analysis - WINDOW 5-29 5.8.5 Selecting the number of the lines in FFT analysis - LINES 5-29 5.8.6 Enabling the FFT spectra time-history logging - LOGGER 5-30

    5.9 Selection of 1/1 and 1/3 octave spectrum parameters - 1/1 OCTAVE SETUP and 1/3 OCTAVE SETUP 5-30

    5.9.1 Enabling 1/1 OCTAVE or 1/3 OCTAVE spectrum settings - ENABLED 5-30 5.9.2 Selecting the weighting filter during 1/1 OCTAVE or 1/3 OCTAVE analysis - FILTER5-31 5.9.3 Selecting the band during 1/1 OCTAVE or 1/3 OCTAVE analysis - BAND 5-31 5.9.4 Activation of logger for 1/1 OCTAVE or 1/3 OCTAVE analysis results - LOGGER 5-32

  • SVAN 958 USER MANUAL

    iii

    6 DATA AVAILABLE ON THE DISPLAY - DISPLAY 6-1 6.1 Selection of the modes of measurement results presentation - DISPLAY MODES 6-1 6.2 Selection of the parameters in graphical results presentations - DISPLAY SETUP 6-3

    6.2.1 Selection of the scale in graphical results presentations - DISPLAY SCALE 6-4 6.2.2 Scaling of the vertical axis of the graphical presentation - DYNAMIC 6-4 6.2.3 Scaling of the horizontal axis of the graphical presentation - XZOOM 6-5 6.2.4 Selection of the parameters of the SPECTRUM - SPECTRUM VIEW 6-5 6.2.5 Selection of the weighting filters - TOTAL VALUES 6-6

    6.2.5.1 Selection of the weighting filters for the profiles - FILTER 6-6 6.3 Checking the state of the internal battery - POWER SUPPLY 6-8 6.4 Setting the parameters of the display - SCREEN SETUP 6-8

    6.4.1 Automatic switch off of the backlight - LIGHT TIMEOUT 6-9 6.4.2 Setting the brightness of the backlight - BRIGHTNESS 6-9 6.4.3 Setting the contrast of the display - CONTRAST 6-9

    6.5 Checking specification of the instrument - UNIT LABEL 6-10

    7 SAVING MEASUREMENT RESULTS - FILE 7-1 7.1 Saving files in the instruments memory - SAVE and AUTO NAME 7-3 7.2 Controlling the data storing in the instruments memory - SAVE OPTIONS 7-5

    7.2.1 Saving files in RAM memory - RAM FILE 7-6 7.2.2 Controlling of the measurement statistics savings - SAVE STATISTICS 7-6 7.2.3 Saving minimum values in the spectrum - MIN SPECTRUM 7-6 7.2.4 Saving maximum values in the spectrum - MAX SPECTRUM 7-7 7.2.5 Replacement of the existing files by the new ones - REPLACE 7-7 7.2.6 Controlling of the measurement results savings - AUTO SAVE 7-8 7.2.7 Direct access to SAVE / AUTO NAME function - DIRECT SAVE 7-8

    7.3 Loading the files with the measurement results - LOAD FILE 7-8 7.4 Checking the contents of the loaded file - LOGGER VIEW 7-10 7.5 Removing a file with the measurement results from memory - DELETE 7-11 7.6 Removing all files with measurement results from memory - DELETE ALL 7-13 7.7 Memory merging - DEFRAGMENTATION 7-14 7.8 Checking the contents of the memory - CATALOGUE 7-15 7.9 Checking the free space in the memory - FREE SPACE 7-16 7.10 Saving setup in the instruments memory - SAVE SETUP 7-17 7.11 Enabling of saving user filter option - SETUP OPTIONS 7-19 7.12 Loading the files with the configuration - LOAD SETUP 7-19

    8 SETUP MENU - SETUP 8-1 8.1 Setting the language of the user interface - LANGUAGE 8-2 8.2 Return to the factory made settings - CLEAR SETUP 8-2 8.3 Day time limits selection - DAY TIME LIMITS 8-3 8.4 Selection of the extended mode - EXT. I/O SETUP 8-4 8.5 Selection of few push-buttons modes - KEYBOARD SETUP 8-5

  • SVAN 958 USER MANUAL _

    iv

    8.5.1 Selection of the working mode of / push-buttons - SHIFT MODE 8-5 8.5.2 Selection of the working mode of push-button - START/STOP 8-5 8.5.3 Locking the keyboard - KEYLOCK 8-6

    8.6 Locking the MENU- MENU LOCK 8-6 8.7 Setting the reference signal in vibration measurements - REFERENCE LEVELS 8-7

    8.7.1 Setting the reference level of the acceleration signal - ACC 8-7 8.7.2 Setting the reference level of the velocity signal - VEL 8-7 8.7.3 Setting the reference level of the displacement signal - DIL 8-7

    8.8 Selection of detectors type in the LEQ (RMS) calculations - RMS INTEGRATION 8-8 8.9 Programming of the instruments internal Real Time Clock - RTC 8-9 8.10 Selection of statistics levels to be saved in a file - STATISTICAL LEVELS 8-9 8.11 Programming of the instruments internal timer - TIMER 8-10 8.12 Selection the USBHOST port functionality - USBHOST PORT 8-11 8.13 Introduction the filter coefficients for 1/1 OCTAVE and 1/3 OCTAVE analysis

    USER FILTERS SETUP 8-12 8.13.1 Selecting the mode for introduction of user filter coefficients - MODE 8-13 8.13.2 Selecting the filter to be viewed, edited or cleared - FILTER 8-13 8.13.3 Setting the coefficients of the user filters set - EDIT 8-13 8.13.4 Clearing the coefficients of the user filters - CLEAR 8-14

    8.14 Selection of the vibration units - VIBRATION UNITS 8-14 8.15 Warnings selection - WARNINGS 8-15

    8.15.1 Saving the measurement results in a file - RESULTS NOT SAVED 8-15 8.15.2 Excluding channel from vector calculations - VECTOR SETTINGS 8-16

    9 CALCULATION OF DOSE PARAMETERS - AUXILIARY FUNCTIONS 9-1 9.1 Selection of the calculation results - HAV CALCULATOR 9-1

    9.1.1 Selection of the file with result of measurement - SELECT RESULTS 9-2 9.1.2 Selection of the partial results - PARTIAL EAV/ELV 9-3 9.1.3 Selection of the partial results - PARTIAL RESULTS 9-4 9.1.4 Selection of the daily exposure - DAILY RESULTS 9-4

    9.2 Selection of the calculation results - WBV CALCULATOR 9-5 9.2.1 Selection of the file with result of measurement - SELECT RESULTS 9-5 9.2.2 Selection of the partial results - PARTIAL EAV/ELV 9-7 9.2.3 Selection of the partial results - PARTIAL RESULTS 9-7 9.2.4 Selection of the daily results - DAILY RESULTS 9-8

  • SVAN 958 USER MANUAL

    v

    LIST OF FIGURES

    SVAN 958 instrument with the seat accelerometer and the acoustic dosimeter microphone .....................................1-4 Control push-buttons of the SVAN 958 instrument ...................................................................................................2-1 Display with the MARKERS (after pressing and together).............................................................2-4 Displays with the activated markers ...........................................................................................................................2-4 Time-history plot with the indication of the active markers .......................................................................................2-5 Top cover of the SVAN 958 instrument in 1:1 scale .................................................................................................2-6 Bottom cover of the SVAN 958 instrument in 1:1 scale ............................................................................................2-6 Front panel of the SVAN 958 instrument in 1:1 scale................................................................................................2-7 Rear panel of the SVAN 958 instrument in 1:1 scale.................................................................................................2-8 Displays with the highlighted elements of the main list .............................................................................................3-1 Displays with the sub-lists which were lately accessed by the user (after double pressing of the push-

    button).............................................................................................................................................................3-1 Displays with the main list (a) and the elements of INPUT list (b)............................................................................3-2 MEASUREMENT SETUP window ..........................................................................................................................3-2 MEASUREMENT SETUP window with INT. PERIOD accessible..........................................................................3-2 Displays with the accessed INT. PERIOD after pressing the or push-buttons ..........................................3-2 Displays after three consecutive pressing of the push-button from MEASUREMENT SETUP window......3-2 Displays during and after the accessing FREE SPACE window................................................................................3-3 Edition of the text, which has to be a name of the file saved in instruments memory ................................................3-3 Control diagram of FUNCTION list ..........................................................................................................................3-4 Control diagram of INPUT list in 1/1 OCTAVE mode (vibration)............................................................................3-8 Control diagram of INPUT list in 1/3 OCTAVE mode (sound) ................................................................................3-8 Control diagram of INPUT list in the acoustic DOSIMETER mode .........................................................................3-8 Control diagram of INPUT list in FFT mode.............................................................................................................3-8 Control diagram of INPUT list in RT 60 mode..........................................................................................................3-9 Control diagram of INPUT list.................................................................................................................................3-11 Control diagram of DISPLAY list............................................................................................................................3-14 Control diagram of DISPLAY list in 1/1 OCTAVE and 1/3 OCTAVE analysis of vibration .................................3-14 Control diagram of DISPLAY SETUP sub-list in 1/1 OCTAVE and 1/3 OCTAVE analysis .................................3-15 Control diagram of FILE list ....................................................................................................................................3-18 Control diagram of SETUP list ................................................................................................................................3-22 Control diagram of AUXILIARY FUNCTION list .................................................................................................3-24 BATTERY windows with different sources powering the instrument: SA 15 external DC power adapter (a),

    SA 17A external battery pack (b), internal batteries (c) and USB power (d)................................................3-25 POWER SUPPLY window when the instrument is powered from external source .................................................3-25 Displays with the Battery icon (a) and in the POWER SUPPLY window (b) ......................................................3-26 Display with the Computer icon and in the POWER SUPPLY window...............................................................3-26 Displays during activation of LIGHT TIMEOUT option ........................................................................................3-26 Displays after switching on the instrument in VLM (a) and in SLM (b) ..................................................................3-27 Displays in one-channel (a) four channels (b) and 3 PROFILES display mode (c) .................................................3-27 Displays in one profile (a) and four channels (b) display mode with the vibration measurement (NON-

    METRIC units) .............................................................................................................................................3-27 Displays in one profile (a) and 3 PROFILES display mode (b) 4 CHANNEL display mode (c).............................3-28 Display with all available icons................................................................................................................................3-28 Display with Paper sheet icon...............................................................................................................................3-28 Display with Battery icon .....................................................................................................................................3-29 Display with Computer icon .................................................................................................................................3-29 Display with Antenna (Tree) icon .....................................................................................................................3-29 Display with Loudspeaker icon.............................................................................................................................3-29 Display with Headphone icon...............................................................................................................................3-29 Display with Envelope icon ..................................................................................................................................3-30 Display with Bell icon ..........................................................................................................................................3-30 Display with Timer icon .......................................................................................................................................3-30 Display with Arrows icon .....................................................................................................................................3-30 Displays when the level of the signal is too low during the measurements of the sound -UNDERRANGE text

    on the bottom of the display (a) and the arrows directed down in 4 CHANNELS mode (b) ........................3-31

  • SVAN 958 USER MANUAL _

    vi

    FUNCTION list opened, MEASUREMENT FUNCTION selected (a) and MEASUR. FUNCTION sub-list opened (b)..................................................................................................................................................... 3-32

    Displays during the entering of the access code to a function.................................................................................. 3-32 Display after the unsuccessful verification of the access code................................................................................. 3-33 Scheme of the instrument memory organisation with the USBHOST and memory stick connected ..................... 3-34 Displays with the main list with FUNCTION text selected (a) and FUNCTION window with

    MEASUREMENT FUNCTION selected (b) and CALIBRATION selected (c) ........................................... 4-1 MEASUREMENT FUNCTION window with all available functions ...................................................................... 4-1 Displays during the activation of the optional function ............................................................................................. 4-1 CALIBRATION windows in CHANNEL 1 .............................................................................................................. 4-2 CALIBRATION windows with the selected calibration mode and after entering this mode..................................... 4-3 CALIBRATION window with SENSITIVITY positions not accessible ................................................................... 4-3 CALIBRATION window with the sensitivity greater than 50.0 mV / Pa and calculated calibration factor .............. 4-3 CALIBRATION window with the sensitivity smaller than 50.0 mV / Pa and calculated calibration factor.............. 4-3 CALIBRATION windows with the lowest possible sensitivity and the greatest calibration factor (a) and the

    greatest sensitivity and the lowest calibration factor (b)................................................................................. 4-4 CALIBRATION windows with the selected calibration mode and after entering this mode..................................... 4-4 CALIBRATION windows with SENSITIVITY positions not accessible.................................................................. 4-4 CALIBRATION window during setting the sensitivity greater than 10.0 mV / ms-2................................................. 4-4 CALIBRATION window during setting the sensitivity smaller than 10.0 mV / ms-2 ................................................ 4-5 CALIBRATION windows with the lowest possible sensitivity and the greatest calibration factor (a)...................... 4-5 and the greatest sensitivity and the lowest calibration factor (b) ............................................................................... 4-5 CALIBRATION windows during the calibration measurement ................................................................................ 4-6 CALIBRATION windows after the measurements and after the acceptance of the calibration factor value............. 4-6 CALIBRATION windows with the selected calibration mode and after entering this mode..................................... 4-7 CALIBRATION windows during and after the calibration measurements................................................................ 4-7 Confirmation of calibration result by pressing the push-button............................................................... 4-7 CALIBRATION window with CALIBRATION HISTORY text selected (a) CALIBRATION HISTORY

    window opened with the last calibration result (b) ......................................................................................... 4-8 CALIBRATION HISTORY windows with the last calibration records .................................................................... 4-8 CALIBRATION HISTORY windows for the first (1) and fourth (4) channel .......................................................... 4-8 Main list with INPUT text selected............................................................................................................................ 5-1 INPUT list in LEVEL METER (a), in 1/1 OCTAVE analyser (b), 1/3 OCTAVE analyser (c), in

    DOSIMETER (d) and in FFT analyser (e) ..................................................................................................... 5-1 Displays with not active sub-lists of INPUT list during the measurement................................................................. 5-1 MEASUREMENT SETUP window.......................................................................................................................... 5-2 MEASUREMENT SETUP windows; the setting of START DELAY with 1-second step ....................................... 5-2 MEASUREMENT SETUP windows; the setting of START DELAY with 10-seconds step.................................... 5-3 MEASUREMENT SETUP windows; the setting of INT. PERIOD with 1-second step ........................................... 5-3 MEASUREMENT SETUP windows; the setting of INT. PERIOD with 10-seconds step........................................ 5-3 MEASUREMENT SETUP windows; the setting of INT. PERIOD with 1-minute and 10-minutes step .................. 5-3 MEASUREMENT SETUP windows; the setting of INT. PERIOD with 1-hour and 10-hours step ......................... 5-3 Display, when the INTEGRATION PERIOD is too short for AUTO SAVE option................................................. 5-4 CYCLES NUMBER setting with the step equal to one ............................................................................................. 5-4 CYCLES NUMBER setting with the step equal to 20............................................................................................... 5-4 MEASUR. SETUP windows, LOGGER STEP selection.......................................................................................... 5-5 Relations between INTEGR. PERIOD and LOGGER STEP .................................................................................... 5-5 INPUT list with CHANNELS SETUP selected and channel selection...................................................................... 5-6 CHANNEL 1 SETUP window in vibration mode ..................................................................................................... 5-6 CHANNEL 1 SETUP windows with MODE selection ............................................................................................. 5-6 CHANNEL 1 SETUP windows in VIBRATION MODE with RANGE selection.................................................... 5-7 CHANNEL 1 SETUP windows in SOUND MODE with RANGE selection ............................................................ 5-7 CHANNEL 1 SETUP with PROFILE 1 text highlighted .......................................................................................... 5-7 CH x PROFILE y windows with the selection of the weighting filter in SM............................................................. 5-7 CHANNEL 1 SETUP windows with the selection of the weighting filter in acceleration measurements ................. 5-8 CHANNEL 1 SETUP windows with the selection of the weighting filter in velocity measurements........................ 5-8 CHANNEL 1 SETUP windows with the selection of the weighting filter in displacement measurements ............... 5-8 CHANNEL x PROFILE y windows (SM) with the selection of the RMS detector................................................... 5-9 CHANNEL x PROFILE y windows (VM) with the selection of the RMS detector .................................................. 5-9

  • SVAN 958 USER MANUAL

    vii

    CHANNEL 4 SETUP, MICROPHONE CORRECTION text selected .....................................................................5-9 MICROPHONE CORRECTION window, DIFFUSE FIELD selection and OUTDOOR selection..........................5-9 LOGGER SETUP option selected and available LOGGER MODES .....................................................................5-10 LOGGER SETUP windows (VM) with the selection of the results to be saved or saved in a loggers file.............5-10 LOGGER SETUP windows (SM) with the selection of the results to be saved in a loggers file............................5-10 AUXILIARY LOGGER window with the selection of VECTOR and RPM result to be saved in a loggers file ...5-11 LOGGER SETUP window, SAMPLING RATE selection ......................................................................................5-11 LOGGER SETUP window, RPM selection .............................................................................................................5-11 LOGGER SETUP window, CHANNEL 4 deactivation ..........................................................................................5-12 LOGGER VIEW window with the file with time-domain signal recorded and message on the display after

    attempt to open the file .................................................................................................................................5-12 INPUT window with TRIGGER SETUP text selected ............................................................................................5-12 TRIGGER SETUP window with TRIGGER mode selection...................................................................................5-12 Displays during the measurements while the triggering condition is not fulfilled....................................................5-13 Displays during the measurements while the triggering condition is not fulfilled....................................................5-13 TRIGGER SETUP windows with SOURCE selection ............................................................................................5-14 TRIGGER SETUP windows with CHANNEL selection .........................................................................................5-14 TRIGGER SETUP windows with LEVEL selection in VM ....................................................................................5-14 TRIGGER SETUP windows with GRADIENT selection (1 dB/ms and 10 dB/ms step up) ...................................5-15 TRIGGER SETUP windows with RTC START selection with 1-second step ........................................................5-15 TRIGGER SETUP windows with RTC START selection with 1-minute step ........................................................5-15 TRIGGER SETUP windows, RTC STEP selection with 1-second step ..................................................................5-15 TRIGGER SETUP windows, RTC STEP selection with 1-minute step ..................................................................5-16 INPUT list with AUXILIARY SETUP text highlighted (displayed inversely)........................................................5-16 AUXILIARY SETUP windows with all available positions....................................................................................5-16 RPM SETUP window with enabling RPM function ................................................................................................5-16 RPM SETUP window with PULSES/ROTATION selection with 1-unit step .........................................................5-17 RPM SETUP window with PULSES/ROTATION selection with 10-units step .....................................................5-17 RPM SETUP window; the UNIT selection..............................................................................................................5-17 RPM SETUP windows with LOGGER activation ...................................................................................................5-18 AUXILIARY SETUP window with SEAT SETUP text highlighted ...................................................................5-18 Selection of SEAT results presentation....................................................................................................................5-18 SEAT SETUP windows with the enabling of the function ...................................................................................5-19 SEAT SETUP windows with SEAT CHANNEL selection ..................................................................................5-19 SEAT SETUP windows with BASE CHANNEL selection..................................................................................5-19 AUXILIARY SETUP window with VECTOR SETUP text highlighted .................................................................5-19 VECTOR SETUP windows with the coefficient selection with 1-unit step .............................................................5-20 VECTOR SETUP windows with the coefficient selection with 10-units step .........................................................5-20 VECTOR SETUP windows with the selection of results from channels for vector calculation...............................5-20 Display in the starting measurements when user selects incorrect filters for HAV/WBV calculations....................5-20 AUXILIARY SETUP window with HAV/WBV DOSE SETUP text selected........................................................5-21 Display after starting measurements when user selected incorrect filters for HAV/WBV DOSE calculations........5-21 HAV/WBV DOSE windows with the enabling of the option...................................................................................5-22 HAV/WBV DOSE windows with EXPOSURE TIME selection with 30-minutes step...........................................5-22 HAV/WBV DOSE windows with EXPOSURE TIME selection with 1-minute step ..............................................5-22 HAV/WBV DOSE windows with the selection of the channel for X AXIS ............................................................5-22 HAV/WBV DOSE windows with STANDARDS text highlighted..........................................................................5-23 STANDARDS windows with STANDARD selection .............................................................................................5-23 INPUT list, ALARM SETUP text highlighted.........................................................................................................5-23 ALARM SETUP windows, CHANNEL 2 or VECTOR selection...........................................................................5-24 Vibration channel opened in LM (a), sound channel opened in LM (b) VECTOR ALARM window opened in

    LM (c) vibration channel opened in 1/1 OCTAVE (d) and 1/3 OCTAVE (e) analysis mode ......................5-24 Setting ALARM conditions, TRIGGER mode selection..........................................................................................5-24 Setting ALARM conditions, INTEGRATION period selection ..............................................................................5-24 Source selection in the case of vibration mode ........................................................................................................5-25 Source selection in the case of sound mode .............................................................................................................5-25 SOURCE selection for 1/3 OCTAVE option...........................................................................................................5-25 LEVEL selection with 0.1dB step and with 1dB step in the vibration channel/mode ..............................................5-25 LEVEL selection with 0.1dB step and with 1dB step in sound channel/mode.........................................................5-25

  • SVAN 958 USER MANUAL _

    viii

    DOSIMETER SETUP selected in INPUT list and DOSIMETER SETUP window................................................ 5-26 DOSIMETER SETUP windows with EXPOSURE TIME selection....................................................................... 5-26 DOSIMETER SETUP windows with CRITERION LEVEL selection ................................................................... 5-26 DOSIMETER SETUP windows with THRESHOLD LEVEL selection ................................................................. 5-27 DOSIMETER SETUP windows with EXCHANGE RATE selection ..................................................................... 5-27 FFT SETUP selected in INPUT list (a), FFT SETUP window (b) and opened FFT SETUP (1) window (c) ......... 5-28 FFT SETUP window in channel 1, enabling FFT analysis ...................................................................................... 5-28 FFT SETUP (4) window with FILTER selection in sound mode ............................................................................ 5-28 FFT SETUP (1) window with FILTER position not accessible in vibration mode ................................................. 5-28 FFT SETUP (1) window with BAND selection....................................................................................................... 5-29 FFT SETUP (1) window with WINDOW selection ................................................................................................ 5-29 FFT SETUP (1) window with LINES selection....................................................................................................... 5-29 FFT SETUP (x) window with LOGGER activation ................................................................................................ 5-30 INPUT windows in the case of 1/1 OCTAVE mode and in 1/3 OCTAVE mode.................................................... 5-30 1/1 OCTAVE SETUP windows with CHANNEL selection.................................................................................... 5-30 SPECTRUM window opened for channel 1 ............................................................................................................ 5-31 SPECTRUM (4) windows with FILTER selection in sound mode.......................................................................... 5-31 SPECTRUM (1) window with FILTER position not accessible in vibration mode................................................. 5-31 SPECTRUM windows with BAND not accessible in VM (a) and BAND selection in SM (b)............................... 5-31 SPECTRUM (4) windows with LOGGER selection ............................................................................................... 5-32 Main list with DISPLAY text highlighted (displayed inversely) ............................................................................... 6-1 DISPLAY window..................................................................................................................................................... 6-1 One-profile presentation mode .................................................................................................................................. 6-2 DISPLAY MODES widows and available presentation modes................................................................................. 6-2 LOGGER mode when there is nothing in the logger to be displayed (after setting LOGGER as active) .................. 6-2 MENU, DISPLAY, DISPLAY SETUP and DISPLAY SETUP (1) windows .......................................................... 6-3 DISPLAY SETUP windows with DISPLAY SCALE text highlighted ..................................................................... 6-3 DISPLAY SETUP window, AUXILIARY text selected, and scale selection in the opened window ....................... 6-4 DISPLAY SETUP (1) windows with DISPLAY SCALE text highlighted................................................................ 6-4 Displays with the possible options of the SCALE in vibration mode ........................................................................ 6-4 Displays with the possible values of the vertical axis in LOGGER and SPECTRUM presentations......................... 6-5 DISPLAY SCALE (1) windows with XZOOM selection........................................................................................ 6-5 MENU, DISPLAY, DISPLAY SETUP and DISPLAY SETUP (1) windows .......................................................... 6-5 SPECTRUM VIEW windows with VIEW selection ................................................................................................. 6-5 SPECTRUM VIEW windows with TYPE for vibration measurements (a) and for sound measurements (b) ........... 6-6 SPECTRUM VIEW windows with FILTER for vibration measurements (a) and for sound measurements (b)........ 6-6 SPECTRUM VIEW windows with MINIMUM / MAXIMUM spectrum selection.................................................. 6-6 DISPLAY SETUP windows with TOTAL VALUES text highlighted and CHANNEL x TOTALS entered ........... 6-6 CHANNEL 1 TOTALS windows in the case of sound measurements with filter selection ...................................... 6-7 CHANNEL 1 TOTALS windows in the case of vibration measurements with FILTER selection............................ 6-7 CHANNEL 1 TOTALS windows in the case of vibration measurements with TYPE selection ............................... 6-7 CHANNEL 1 TOTALS windows in the case of vibration measurements with calibration factor setting.................. 6-8 DISPLAY window with POWER SUPPLY text highlighted .................................................................................... 6-8 POWER SUPPLY windows for different sources powering the instrument.............................................................. 6-8 DISPLAY window with SCREEN SETUP text highlighted...................................................................................... 6-9 SCREEN SETUP windows with LIGHT TIMEOUT active (a), and not active (b) .................................................. 6-9 SCREEN SETUP windows with BRIGHTNESS position active .............................................................................. 6-9 SCREEN SETUP windows with CONTRAST setting ............................................................................................ 6-10 DISPLAY window with UNIT LABEL text highlighted......................................................................................... 6-10 UNIT LABEL windows opened and after scrolling with the , push-buttons .......................................... 6-10 Main list with FILE text highlighted (displayed inversely)........................................................................................ 7-2 FILE list..................................................................................................................................................................... 7-2 FILE list with SAVE text highlighted (displayed inversely)...................................................................................... 7-3 SAVE window opened............................................................................................................................................... 7-3 Displays during the files name edition...................................................................................................................... 7-3 Displays in FILE NAME edition after pressing and push-buttons ....................................................... 7-3 Displays in FILE NAME edition after pressing and push-buttons ....................................................... 7-4 SAVE window with AUTO NAME function selection ............................................................................................. 7-4

  • SVAN 958 USER MANUAL

    ix

    Displays after SAVE operation when there is no results to save (a) and the file with the selected name already exists in the instruments memory (b) .............................................................................................................7-4

    Display after the execution of SAVE operation .........................................................................................................7-4 SAVE window AUTO NAME option selected (a), after saving the file with the increased name (b) and after

    repeated enter to SAVE window (c) ...............................................................................................................7-5 Displays after the files name selection (a) and with the message if REPLACE option is not active (b) ...................7-5 FILE list with SAVE OPTIONS text highlighted (displayed inversely) ....................................................................7-5 SAVE OPTIONS windows with the selection of RAM FILE parameters..................................................................7-6 SAVE OPTIONS windows with the selection of SAVE STATISTICS.....................................................................7-6 SAVE OPTIONS windows with the selection of MIN. SPECTRUM saving ............................................................7-6 SAVE OPTIONS windows with the selection of MAX. SPECTRUM saving...........................................................7-7 SAVE OPTIONS windows with the selection of REPLACE.....................................................................................7-7 SAVE OPTIONS windows with the selection of AUTO SAVE parameters..............................................................7-8 Displays after entering FILE NAME edition mode for the files saved with AUTO SAVE option active (after

    pressing push-button in SAVE OPTIONS window, AUTO SAVE switched on).........................7-8 SAVE OPTIONS windows with the selection of DIRECT SAVE.............................................................................7-8 FILE list with LOAD FILE text highlighted (displayed inversely) ............................................................................7-9 Displays with the message stating the reason for unfeasibility of the required operation ..........................................7-9 Displays during the execution of LOAD FILE operation...........................................................................................7-9 Displays during the overview of the file list...............................................................................................................7-9 Execution of the LOAD FILE operation ..................................................................................................................7-10 FILE list with LOGGER VIEW text highlighted .....................................................................................................7-10 LOGGER VIEW windows with the selection of the logger file to be viewed..........................................................7-10 Selection of the logger plot from different channels ................................................................................................7-11 FILE list with DELETE text highlighted (displayed inversely) ...............................................................................7-11 DELETE windows opened with RESULT FILES and SETUP FILES selected ......................................................7-11 Displays with the message stating the reason for unfeasibility of the required operation ........................................7-11 DELETE window when the memory of the instrument is empty .............................................................................7-12 DELETE windows with the selection of the file to be deleted.................................................................................7-12 Execution of DELETE operation .............................................................................................................................7-12 Execution of the 02JAN5 file deletion and the influence of this process on the memory space ..............................7-12 FILE list with DELETE ALL text highlighted (displayed inversely).......................................................................7-13 DELETE ALL windows and the selection of the files to be deleted ........................................................................7-13 Displays during the execution of DELETE ALL operation .....................................................................................7-13 Displays with the not possible confirmation of DELETE ALL order during the measurement ...............................7-13 Displays during the execution of DELETE ALL operation and after this execution ...............................................7-14 FILE list with DEFRAGMENTATION text highlighted .........................................................................................7-14 Displays before the execution of DEFRAGMENTATION operation......................................................................7-14 Display in the case when DEFRAGMENTATION operation is not required..........................................................7-14 Displays during the execution of DEFRAGMENTATION operation......................................................................7-15 Displays with the state of the files memory before DELETE operation (a), after DELETE operation (b) after

    the execution of DEFRAGMENTATION operation (c) and the state of the memory after the execution of the operation (d) .......................................................................................................................................7-15

    FILE list with CATALOGUE text highlighted ........................................................................................................7-15 Displays with the contest of CATALOGUE operation when the memory is empty.................................................7-16 Exemplary contents of CATALOGUE window.......................................................................................................7-16 FILE list with FREE SPACE text highlighted (displayed inversely) .......................................................................7-16 FREE SPACE window after the execution of the DELETE ALL operation............................................................7-16 FREE SPACE window with the number depending on the measurements and operations performed.....................7-17 FILE list with SAVE SETUP text highlighted (displayed inversely).......................................................................7-17 SAVE SETUP window ............................................................................................................................................7-17 Displays during the process of setting the character in the edited name ..................................................................7-17 Displays in FILE NAME edition after pressing and push-buttons .....................................................7-18 Displays in FILE NAME edition after pressing and push-buttons .....................................................7-18 Displays with AUTO NAME function switched on and after saving the file with the increased name....................7-18 Displays after SAVE operation when the file with the selected name already exists in the instruments

    memory (with the message if REPLACE is not active) ................................................................................7-18 FILE list with SETUP OPTIONS text highlighted...................................................................................................7-19 SETUP OPTIONS window with SAVE USER FILTER selection ..........................................................................7-19

  • SVAN 958 USER MANUAL _

    x

    FILE list with LOAD SETUP text highlighted (displayed inversely)...................................................................... 7-19 Displays with the message stating the reason for unfeasibility of the required operation ........................................ 7-20 Displays during the execution of LOAD SETUP operation .................................................................................... 7-20 Displays during the overview of the file list ............................................................................................................ 7-20 Display during and after the execution of LOAD SETUP operation ....................................................................... 7-20 Main list with SETUP text highlighted (displayed inversely).................................................................................... 8-1 Displays with SETUP list .......................................................................................................................................... 8-2 Displays with LANGUAGE list................................................................................................................................. 8-2 SETUP list with CLEAR SETUP text highlighted (displayed inversely) .................................................................. 8-2 Displays with the request for the confirmation for CLEAR SETUP execution ......................................................... 8-3 Display after the execution of CLEAR SETUP function........................................................................................... 8-3 SETUP list with DAY TIME LIMITS text highlighted (displayed inversely)........................................................... 8-3 Displays with the available DAY TIME LIMITS...................................................................................................... 8-3 SETUP list in sound measurements with EXT. I/O SETUP text highlighted (displayed inversely) .......................... 8-4 EXT. I/O SETUP windows with the different I/O devices selection ......................................................................... 8-4 EXT. I/O SETUP windows with the channel selection for output signal................................................................... 8-4 SETUP list with KEYBOARD SETUP text highlighted (displayed inversely) ......................................................... 8-5 KEYBOARD SETUP windows with the available settings in SHIFT MODE.......................................................... 8-5 KEYBOARD SETUP windows with the available settings in START/STOP .......................................................... 8-6 KEYBOARD SETUP window with the activation of KEYLOCK option................................................................. 8-6 SETUP list with MENU LOCK text highlighted (displayed inversely)..................................................................... 8-6 MENU LOCK windows with available options......................................................................................................... 8-6 SETUP list with REFERENCE LEVELS text highlighted (displayed inversely)...................................................... 8-7 REFERENCE LEVELS windows with the reference level setting of acceleration signal ......................................... 8-7 REFERENCE LEVELS windows with the reference level setting of velocity signal................................................ 8-7 REFERENCE LEVELS windows with the reference level setting of displacement signal........................................ 8-8 REFERENCE LEVELS window with the reference level of the acoustic signal....................................................... 8-8 SETUP list with RMS INTEGRATION text highlighted (displayed inversely)........................................................ 8-8 Displays with the available options of RMS INTEGRATION .................................................................................. 8-8 SETUP list with RTC text highlighted (displayed inversely) .................................................................................... 8-9 RTC window.............................................................................................................................................................. 8-9 SETUP list with STATISTICAL LEVELS text highlighted (displayed inversely) ................................................... 8-9 STATISTICAL LEVELS windows ......................................................................................................................... 8-10 SETUP list with TIMER text highlighted (displayed inversely).............................................................................. 8-10 TIMER window....................................................................................................................................................... 8-10 SETUP list with USBHOST PORT text highlighted (displayed inversely)........................................................... 8-11 USBHOST SETUP window .................................................................................................................................. 8-11 Displays during the activation of USB hosts functions........................................................................................... 8-11 SETUP list with USER FILTERS SETUP text highlighted (displayed inversely) .................................................. 8-12 SPECTRUM BASED windows with MODE selection ........................................................................................... 8-13 SPECTRUM BASED windows with the filter selection for vibration (a) and for sound (b)................................... 8-13 SPECTRUM BASED windows with EDIT selected ............................................................................................... 8-13 EDIT windows with the setting of the filters coefficient ........................................................................................ 8-13 EDIT windows with the setting of the filters coefficient (cont.)............................................................................. 8-14 SPECTRUM BASED windows with CLEAR selected ........................................................................................... 8-14 Displays with the request for the confirmation for CLEAR FILTER execution ...................................................... 8-14 SETUP list with VIBRATION UNITS text highlighted (displayed inversely)........................................................ 8-14 VIBRATION UNITS windows ............................................................................................................................... 8-15 SETUP list with WARNINGS text highlighted (displayed inversely)..................................................................... 8-15 WARNINGS windows with RESULT NOT SAVE selected................................................................................... 8-15 Displays with the warning that the previous results were not saved and the confirmation....................................... 8-15 WARNINGS windows with the selection of VECTOR SETTINGS....................................................................... 8-16 Display with the warning that CHANNEL x has been excluded from vector calculation........................................ 8-16 Main list with AUXILIARY FUNCTIONS text highlighted (displayed inversely)................................................... 9-1 AUXILIARY FUNCTIONS windows....................................................................................................................... 9-1 HAV CALCULATOR window ................................................................................................................................. 9-2 HAV CALCULATOR window with SELECT RESULTS text highlighted (displayed inversely) ............................ 9-2 SELECT RESULTS list............................................................................................................................................. 9-2 Displays during the selection of the files ................................................................................................................... 9-2

  • SVAN 958 USER MANUAL

    xi

    Display with the selecting file and after the execution of the operation .....................................................................9-3 Display after checking the contents of the file without HANDARM data................................................................9-3 SELECT FILE windows during the execution of the setting EXPOSURE TIME operation .....................................9-3 HAV CALCULATOR window with PARTIAL EAV/ELV text highlighted (displayed inversely) ..........................9-3 Displays with the empty partial result list (a), the selecting file in SELECT RESULTS window (b) and

    PARTIAL EAV/ELV results for selected files (c)..........................................................................................9-4 HAV CALCULATOR window with PARTIAL RESULTS text highlighted (displayed inversely) ..........................9-4 EXPOSURE results for selected files ........................................................................................................................9-4 HAV CALCULATOR window with DAILY RESULTS text highlighted (displayed inversely)...............................9-4 DAILY RESULTS window with the result for selected files.....................................................................................9-5 WBV CALCULATOR window.................................................................................................................................9-5 WBV CALCULATOR window with SELECT RESULTS text highlighted (displayed inversely)............................9-6 Displays in the SELECT RESULTS list ....................................................................................................................9-6 SELECT FILE windows with the selection of the file ...............................................................................................9-6 Displays with the selected file (a) and after the execution of the operation (b)..........................................................9-6 Display after the execution of loading a file without WBV data................................................................................9-6 Displays during the execution of the setting EXPOSURE TIME operation ..............................................................9-7 WBV CALCULATOR window with PARTIAL EAV/ELV text highlighted (displayed inversely)..........................9-7 Displays with the empty partial result list (a), the selecting file in SELECT RESULT sub-list (b) and

    PARTIAL EAV/ELV results for selected files (c)..........................................................................................9-7 WBV CALCULATOR window with PARTIAL RESULTS text highlighted (displayed inversely) .........................9-8 PARTIAL RESULTS with EXPOSURE DOSE for selected files.............................................................................9-8 WBV CALCULATOR window with DAILY RESULTS text highlighted (displayed inversely)..............................9-8 DAILY EXPOSURE and DAILY DOSE results for selected file .............................................................................9-8

  • SVAN 958 USER MANUAL

    1-1

    1 INTRODUCTION

    The SVAN 958 is digital, four channels 0.5 Hz to 20 kHz signal analyser including Type 1 sound level meter (meeting IEC 61672-1:2002) and vibration meter (meeting ISO 8041:2005). It is an ideal choice for the Human Vibration (according to the ISO 2631-1, 2 & 5 and ISO 5349-1 & 2 standards) and noise measurements in the occupational health and safety monitoring tasks. All required weighting filters, transducers and adapters for triaxial Whole-Body and triaxial Hand-Arm vibration measurements (VM) are available with this instrument.

    Each of four channels can work simultaneously with independently configured input (transducer type), filters and RMS detector time constants (e.g. simultaneous three-axis measurement of the Whole-Body vibration and noise dose).

    Four channels allow parallel measurements with independently defined filters and RMS detector time constants in every channel. Additionally, in the case of sound measurements (SM), each channel calculates simultaneously the results in three independent profiles. Each channel (and profile for SM) provides "multidimensional" analysis of measured signal (like LEQ, LMax, LMin, LPeak, SPL, SEL in the case of SM or RMS, PEAK, PEAKPEAK (PP), VDV, MTVV in the case of VM). Advanced time-history logging, in non-volatile 32 MB internal memory, provides very powerful measurement capability. The external USB Memory Stick extends this facility almost unlimitedly. Results can be easy downloaded to any PC using standard USB (or optional RS 232 and IrDA) interface and SvanPC software.

    Using computational power of its digital signal processor the SVAN 958 instrument can perform advanced frequency analysis simultaneously to the meter mode: - real time 1/1 or 1/3 octave analysis including statistical calculations - FFT analysis including cross spectra.

    Reverberation Time measurements, noise dosimeter and rotation speed measurements are available as options for the SVAN 958 instrument.

    The time-domain signal recording on the external USB memory stick is also available as an exceptional option.

    Fast USB 1.1 interface (12 MHz) creates real time link for the PC "front-end" application of the SVAN 958. The instrument can be fully remotely controlled.

    The instrument is powered by four AA standard or rechargeable batteries (separate charger is required). The powering of the instrument from the external DC power source or the USB interface is also provided.

    Robust case and light weight design accomplish the exceptional features of this new generation instrument.

    1.1 SVAN 958 as Sound Level Meter & Analyser

    Noise measurements (SPL, LEQ, SEL, Lden, Ltm3, Ltm5 and statistics) with Type 1 accuracy (IEC 61672-1:2002) in the frequency range 10 Hz 20 kHz (with the SV 22 microphone)

    Simultaneous measurements in three profiles of any channel with the independent set of IMPULSE, FAST and SLOW detectors with standard A, C, LIN and G filters

    Digital True RMS detector with Peak detection, resolution 0.1 dB, Time Constants: SLOW, FAST, IMPULSE

    1/1 octave and 1/3 octave real time analysis (optional) - 15 filters with the centre frequencies from 1 Hz to 16 kHz, Type 1 - IEC 1260 and 45 filters with the centre frequencies from 0.8 Hz to 20 kHz, Type 1 - IEC 1260

    Measurement range: 17 dBA RMS 140 dBA Peak

  • SVAN 958 USER MANUAL _

    1-2

    FFT real time analysis with up to 1920 lines in 22.4 kHz band with HANNING, RECTANGLE, KAISER-BESSEL or FLAT TOP window (option)

    Reverberation Time analysis (RT 60) in 1/3 octave bands (option) Internal noise level less than 17 dBA RMS

    1.2 SVAN 958 as Vibration Meter & Analyser

    Vibration measurements according to ISO 2631-1, 2 & 5 and ISO 5349-1 & 2 with Type 1 accuracy (ISO 8041:2005) in the frequency range 0.5 Hz3 kHz (with SV 39A/L accelerometer) or 2 Hz 10 kHz (with SV 3023 M2 accelerometer)

    Simultaneous RMS, VDV, MTVV or MAX, PEAK, PP measurements in four channels with independent set of filters and detector constants

    Digital True RMS & RMQ detectors with Peak detection, resolution 0.1 dB, Time Constants: from 100 ms to 10 s

    Wd, Wk, Wc, Wj, Wm, Wb, Wg (ISO 2631), Wh (ISO 5439), HP1, HP3, HP10, Vel1, Vel3, Vel10, VelMF, Dil1, Dil3, Dil10, KB weighting filters

    1/1 octave and 1/3 octave real time analysis (optional) - 15 filters with the centre frequencies from 1 Hz to 16 kHz, Type 1 - IEC 1260 and 45 filters with the centre frequencies from 0.8 Hz to 20 kHz, Type 1 - IEC 1260

    FFT real time analysis with up to 1920 lines in 22.4 kHz band with HANNING, RECTANGLE, KAISER-BESSEL or FLAT TOP window (option)

    RPM rotation speed measurements parallel to the vibration measurement (1 99999) (option) Measurement range: 0.003 ms-2 RMS 500 ms-2 Peak (for the accelerometer with the sensitivity equal

    to 100 mV/g)

    1.3 General features of SVAN 958

    Internal logger function for logging more than two weeks of 1-second PEAK / MAX / MIN / RMS results in the case of SM and PEAK / PP / MAX (or MTVV) / RMS / VDV results in the case of VM (32 MB of non-volatile memory, optional USB memory stick)

    USB 1.1 Client, USB Host, RS 232 (option, SV 55 required) and IrDA (option) interfaces Powered by four AA standard batteries (operation time >10 hours) or four AA rechargeable

    batteries (e.g. NiMH - operation time > 16 hours), SA 17A external battery pack (operation time > 14 hours), external DC power source (6 V 15 V) or USB interface (500 mA)

    Acoustic dosimeter function (option) Integration time programmable up to 24 hours Time-domain signal recording on USB memory stick (option) Handheld and robust case Light weight (only 510 grams including batteries)

  • SVAN 958 USER MANUAL

    1-3

    1.4 Accessories included

    SC 09 - I/O cable SC 16 - USB 1.1 cable SC 61 - integrated connector (TNC to BNC) four AA standard (alkaline) batteries SvanPC for Windows 95/98/NT/2000/XP

    1.5 Accessories available

    SV 22 - (Type 1) prepolarised condenser microphone , sensitivity 50 mV/Pa SV 22L - SV 22 microphone selected for the 1 Hz (-1 dB) low frequency response SV 12L - microphone preamplifier SA 08 - gooseneck for the microphone preamplifier SC 26 - extension cable for the microphone preamplifier SV 25 - (Type 2), dosimeter, ceramic microphone with integrated preamplifier SV 39A/L - seat accelerometer with the nominal sensitivity 10 mV / ms-2 (100 mV / g) for Whole Body measurements SV 3143M1 - IEPE type triaxial accelerometer from DYTRAN with the nominal sensitivity 10 mV / ms-2 (100 mV / g), SC 38 cable required SV 3233A - IEPE type triaxial accelerometer from DYTRAN with the nominal sensitivity 100 mV / ms-2 (1000 mV / g); SC 38 cable required SV 3023M2 - IEPE type triaxial accelerometer from DYTRAN with the nominal sensitivity 1 mV / ms-2 (10 mV / g); SC 38 cable required SC 50Z - car cigarette plug to external power supply plug SC 27 - TNC (plug) to TNC (plug) coil cable SC 38 - 4-pins Microtech to LEMO 4-pins cable (2.7m) (for SV 3023M2, SV 3233A, SV 3143M1) SC 39P - LEMO 4-pins (plug) to 3 * BNC sockets cable SC 49 - LEMO 4-pins (plug) to 3 * TNC sockets (0.7 m) SC 242 - cable for SV 207 accelerometer in hermetic mounting box and SVAN 958 (3 m) SV 48 - voltage to IEPE (ICP) converter SV 50 - set for Hand-Arm measurements (SV 3023M2 accelerometer with 10 mV/g sensitivity) SV50_48 - Hand-Arm Vib. measurements set (SV 3023M2, SC 38, SA 50, SA 51, SA 52, small box for SA 48) SV 207_SO - Building Vibration Measurement set (SV 3233A accelerometer with hermetic mounting box and SC 242 cable) SV 55 - RS 232 interface SV 56 - IrDA interface SV 60 - sound measurements set (SV 22 microphone, SV 12L preamplifier SA 08 goseneck and SA 22 windscreen)

  • SVAN 958 USER MANUAL _

    1-4

    SA 17A - external battery pack SA 15 - power supply unit SA 22 - windscreen SA 201A - outdoor microphone kit SA 46 - carrying belt-bag for SVAN 94x and SVAN 95x (leather) SA 47- carrying bag (fabric material) SA 48 - carrying case (waterproof) SA 50 - Hand-Arm measurements adapter, shaped base (for SV 3023M2 acceler.) SA 51 - Hand-Arm measurements adapter, flat base (for the SV 3023M2 acceler.) SA 52 - Hand-Arm measurements , adapter, direct (for the SV 3023M2 acceler.) SvanPC+ for Windows 98/2000/XP

    SVAN 958 instrument with the seat accelerometer and the acoustic dosimeter microphone

  • SVAN 958 USER MANUAL

    2-1

    2 MANUAL CONTROL OF THE INSTRUMENT

    The control of the instrument is developed in the fully "conversational" way. The user can "programme" the operation of the instrument selecting the proper option from the MENU. Thanks to it, the number of the control push-buttons of the instrument is reduced to nine.

    2.1 Control push-buttons on the front panel

    On the front panel of the instrument, there are located the following control push-buttons: 1. , (), [] 2. , (), [] 3. , [Markers] 4. , [Markers] 5. 6. 7. 8. 9.

    The name given in brackets (...) denotes the second push-button function which is available after pressing it in conjunction (or in sequence) with the push-button. The name given in square brackets [] denotes the push-button function which is available after pressing it in conjunction (or in a sequence) with the push-button.

    Control push-buttons of the SVAN 958 instrument

    The second function of a push-button (written in red colour on a push-button) can be used when

    the push-button is pressed. This push-button can be used in two different ways: as SHIFT in the keyboard; both push-button must be pressed in parallel as 2nd Fun; this push-button can be pressed and released before pressing the second one

    pressed in conjunction with enables the user to enter the MARKERS on the logger plots during the measurement.

  • SVAN 958 USER MANUAL .

    2-2

    Notice: The operation of this push button can be set as the Shift mode or the 2nd Fun. mode in the SHIFT MODE (path: MENU / SETUP / KEYBOARD SETUP / SHIFT MODE), see Chapter with the SETUP list description.

    This push-button enables one to choose the third push-button function in the case of [] and [] push-buttons. In order to select the third function user must press the and the second push-button simultaneously.

    This push-button enables one to start the measurement process, when the instrument is not

    measuring or to stop it, when the instrument is measuring. It is also possible to set such mode of this push-button, in which in order to start or stop the measurements the user has to press it simultaneously with the one.

    Notice: The changing of the push-button mode is performed in the START/STOP (path: MENU / SETUP / KEYBOARD SETUP / START/STOP), see Chapter with the SETUP list description.

    Notice: The simultaneous pressing of the and push-buttons switches the instrument on and off.

    This push-button enables one to enter the selected operation mode or to confirm control options. Some additional functions of this push-button will be described in the following chapters.

    () This push-button (pressed together with the one) enables the user to enter the main list

    containing six sub-lists: FUNCTION, INPUT, DISPLAY, FILE, SETUP and AUXILIARY FUNCTIONS. Each of the mentioned above sub-lists consists of the sub-lists, elements and data windows. These main sub-lists will be detailed described in the following chapters of the manual. Double pressed push-button enters the HISTORY list containing eight last opened sub-lists. It often speeds up the control of the instrument.

    [] This push-button (pressed together with the push-button) enables the user to save

    measurement results as a file in the internal instruments memory or on the USB memory stick. There are two ways of saving a file: with AUTO NAME option - saving a file with the name automatically increased by one (e.g. JAN0, JAN1, JAN2) or save a file by editing its name in the FILE NAME position (see Chapter with the FILE list description).

  • SVAN 958 USER MANUAL

    2-3

    This push-button closes the control lists, sub-lists or windows. It acts in opposite to the

    push-button. When the window is closed pressing the push-button, any changes made in it are (in almost all cases) ignored.

    [] This push-button enables one to break temporary the measurement process. The subsequent

    pressing of the push-button deletes the measurement result from the last one second. Up to fifteen last seconds of the measurement can be cancelled in this way.

    , These push-buttons enable one, in particular, to:

    select the options in an active position in the "horizontal direction" (e.g. filter: LIN, A, C or G, Integration period: 1s, 2s, 3s, etc.)

    select the measurement result to be displayed (e.g. PEAK, MAX, MIN, etc.) in one-channel and multi-channel modes of results presentation)

    control the cursor in SPECTRUM, LOGGER and STATISTICS modes of results presentation select the position of the character in the text edition (i.e. in the FILE NAME menu) switch on/off markers 2 and 3 switch on/off the BACKLIGHT of the display ( + pressed together)

    (, ) The , push-buttons pressed in conjunction (or in sequence) with the enable

    the user, in particular, to: speed up the changing of the numerical values of the parameters (i.e. the step is increased from 1

    to 10 in the setting of START DELAY - path: MENU / INPUT / MEASUREMENT SETUP / START DELAY)

    insert or delete a character in the text edition modes change the statistics class (the number displayed after the letter L) in one-channel and multi-channel

    modes of results presentation Some other possible reactions of the instrument on the pressing of these push-buttons

    will be described in details in the following chapters.

    ,

    The , push-buttons enable one, in particular, to: change the mode of results presentation select the proper character from the list in the text edition mode switch the active sub-list in a list programme the Real Time Clock (RTC) and TIMER switch on/off markers 1 and 4

    Some other possible reactions of the instrument on the pressing of these push-buttons will be described in details in the following chapters.

  • SVAN 958 USER MANUAL .

    2-4

    (, ) The , push-buttons pressed in conjunction (or in sequence) with the enable one,

    in particular, to: change the relation between the Y-axis and X-axis of all plots presented on the screen switch the channels and profiles in one-channel and STATISTICS modes of results presentation switch the active channel in multi-channels mode of results presentation

    Some other possible reactions of the instrument on the pressing of these push-buttons will be described in details in the following chapters.

    [Markers] The Markers enable the user to mark the special events, which occurred during the performed

    measurements (i.e. the airplane flight, the dogs barking, the trains drive etc.). In order to enter the markers the logger has to be switched on (path: MENU / INPUT / LOGGER SETUP / LOGGER MODE: ON) and one or more logger options (PEAK, MAX, MIN, RMS in the case of sound measurements and PEAK, PP, MAX, RMS, VDV in the case of vibration measurements) in channels have to be chosen (path: MENU / INPUT / LOGGER SETUP / CHANNEL x). In order to enter the marker the user must press and push-buttons simultaneously during the measurement. The ENTER MARKER window opens and there are four available marker numbers. To choose marker number 1 the user must press push button (number 2 - , number - 3 and number 4 - ).

    The ENTER MARKER window closes automatically and chosen marker is activated (after pressing + again active marker number will be highlighted). In order to switch off the marker, the user has to open the ENTER MARKER window and press this push-button, which refers to the marker to be switched off. Up to four markers can be switched on at the same time.

    The current state of the markers is indicated in the loggers file and can be used to show them using dedicated presentation software.

    Display with the MARKERS (after pressing and together)

    Displays with the activated markers

    The exemplary presentation of the markers on the time-history plot is shown below (to view a plot with the markers the user has to transfer data to the proper software).

  • SVAN 958 USER MANUAL

    2-5

    Time-history plot with the indication of the active markers

    20

    30

    40

    50

    60

    70

    80

    13:30:00 13:30:09 13:30:17 13:30:26 13:30:35 13:30:43 13:30:52

    LEQ Marker 1 Marker 2 Marker 3 Marker 4

  • SVAN 958 USER MANUAL .

    2-6

    2.2 Input and output sockets of the instrument

    The measurement inputs, called Channels, are placed on the top cover of the instrument. There are 4-pins Lemo compatible socket type ENB.0B.304 for Channels 13 and TNC for Channel 4, all with IEPE power supply for the accelerometers or microphone preamplifiers.

    The microphone preamplifier SV 12L has the proper plug-in with the screw for direct connection with the instrument to the TNC connector (Channel 4) but it is recommended to use the preamplifier with any of the extension cables (i.e. SC 26) or the SA 08 gooseneck. The same type of the connector should be used to attach one-channel accelerometer to Channel 4. The SC 27 coiled cable is recommended in this case. In order to connect the SV 12L microphone preamplifier to Channels 13 one has to use the SC 49 cable (LEMO 4-pins plug to 3 * TNC sockets, 0.7 meters long). The SC 49 or SC 39P (LEMO 4-pins plug to 3 * BNC sockets, 0.7 meters long) cables should be used to connect one-channel accelerometer to any of the Channels 13. The triaxial accelerometers can be easy connected to Channels 13 by means of the SC 38 cable (4-pins Microtech to LEMO 4-pins, 2.7 meters long). It is recommended to attach the SV 25 dosimeter microphone with the integrated preamplifier and a cable to Channel 4.

    Notice: Pay attention that the TNC connector should be always twisted to the light resistance but the LEMO connector is a push-pull only.

    The full description of the signals connected to the sockets is given in the Appendix C.

    Top cover of the SVAN 958 instrument in 1:1 scale

    In the bottom cover there are four sockets, placed from the right to the left as follows: Ext. Pow., USB Host, USB Device and I/O.

    Bottom cover of the SVAN 958 instrument in 1:1 scale

    The USB 1.1 Client interface (the USB Device socket) is the serial interface working with 12 MHz clock. Thanks to its speed, this interface is widely used in all PCs. In the instrument, the standard 4-pins socket is used described in details in Appendix C.

    The USB Host 1.1 interface can be used to connect the external storage, enabling the device to register virtually infinite sequence of measurement results.

  • SVAN 958 USER MANUAL

    2-7

    The Ext. Pow. socket located on the bottom cover of the instrument is Marushin MJ-14 compatible socket, dedicated for the standard 5.5 / 2.1 mm plug (the right one in the Fig. above). The user can connect the external mains adapter (110 V / 230 V) which furnishes the proper DC level. The instrument can be charged from the external DC source (6 V / 500 mA DC 15 V / 250 mA DC). The current consumption depends on the voltage of the power supplier.

    The additional input / output socket, called I/O, is 1-pin LEMO compatible socket type ERN.00.250 (the left one in the Fig. above). The function of this socket can be selected from menu (path: MENU / SETUP / EXT. I/O SETUP / MODE). The socket can be used as: analogue output with the signal from the input of the analogue / digital converter (before

    the correction); this signal can be registered using magnetic recorder or observed on the oscilloscope (the ANALOG setting)

    digital input for external interrupt (the DIGITAL IN setting) digital output for external trigger (the DIGITAL OUT setting)

    Notice: Switch the power off before connecting the instrument to any other device (e.g. a printer or a Personal Computer).

    Front panel of the SVAN 958 instrument in 1:1 scale

  • SVAN 958 USER MANUAL .

    2-8

    Rear panel of the SVAN 958 instrument in 1:1 scale

  • SVAN 958 USER MANUAL

    3-1

    3 SETTING THE INSTRUMENT

    In order to perform the measurements using the instrument the user has only to plug-in the preamplifier with the microphone or the proper accelerometer and to switch the power on.

    Notice: The user has to press the and push-buttons in parallel in order to switch the power On/Off.

    3.1 Basis of the instruments control

    The instrument is controlled by means of nine push-buttons of the keyboard. Using these push-buttons one can access all available functions. The functions are placed in the system of lists and sub-lists. The main list contains the headers of six lists, which also contain sub-lists or positions (elements). The main list is opened after pressing the push-button. This list contains the following elements: FUNCTION, INPUT, DISPLAY, FILE, SETUP and AUXILIARY FUNCTIONS. The elements of each list are described in details in the consecutive Chapters. Only one list can be accessed in a time, this one which name is highlighted (displayed inversely). The change of the highlighted line is done after pressing the , (or , ) push-buttons.

    Displays with the highlighted elements of the main list

    After double pressing of the push-button the HISTORY list, containing eight sub-lists lately accessed by the user, appears on the display. The example of this list is presented below. Such solution enables one to access very fast eight, the most frequently used, lists without the necessity of passing the whole path.

    Displays with the sub-lists which were lately accessed by the user (after double pressing of the push-button)

    After the selection of the desired list (the or push-buttons), the user has to press the push-button in order to enter it. After this new sub-lists, positions (elements) or various data specification appear on the display.

  • SVAN 958 USER MANUAL .

    3-2

    a) b) Displays with the main list (a) and the elements of INPUT list (b)

    Next pressing of the push-button enables one to access mentioned above sub-lists.

    MEASUREMENT SETUP window

    The desired position of a list is accessed after pressing the or push-button.

    MEASUREMENT SETUP window with INT. PERIOD accessible

    The change of the value in a selected position is performed by pressing the or push-buttons.

    Displays with the accessed INT. PERIOD after pressing the or push-buttons

    The push-button is used for the confirmation of the selection in a position and for closing the opened sub-list. The sub-list is closed ignoring any changes made in a window by pressing the push-button.

    Displays after three consecutive pressing of the push-button from MEASUREMENT SETUP window

    As it was mentioned, some of the sub-lists end with the windows informing the user about the state of the instrument, available memory, not existing files or loggers, standards fulfilled by the unit, etc.

  • SVAN 958 USER MANUAL

    3-3

    Displays during and after the accessing FREE SPACE window

    In order to close such window the user has to press the push-button. In the instrument, there are also windows, which are used for entering text (i.e. the name of the file).

    Edition of the text, which has to be a name of the file saved in instruments memory

    Below the structure of the elements of the main list is presented. The more detailed description of the FUNCTION, INPUT, DISPLAY, FILE, SETUP