ACE - HPLC · Comparative data may not be representative of all applications. ACE UltraCore...

4
ACE ACE HPLC and UHPLC Columns Independent Column Comparisons UHPLC and HPLC Columns UHPLC and HPLC columns Porous and solid-core particles Asymmetry and efficiency comparisons Independently tested for impartial results

Transcript of ACE - HPLC · Comparative data may not be representative of all applications. ACE UltraCore...

ACE

ACEHPLC and UHPLC Columns

Independent Column Comparisons

UHPLC and HPLC Columns

• UHPLC and HPLC columns

• Porous and solid-core particles

• Asymmetry and efficiency comparisons

• Independently tested for impartial results

www.ace-hplc.com ACE UHPLC and HPLC Columns - Independent Comparisons

“Independent Testing Shows ACE HPLC Columns Deliver Outstanding Peak Shape”

Independent Testing Comparison #1: HPLC Columns

Column Dimensions: 150 x 4.6mm, 5µm Mobile Phase: 80:20 MeOH/5mM potassium phosphate buffer (pH 7.0) Flow Rate: 2.0ml/min Temperature: 24°C

The above data was obtained from the National Institute of Standards and Technology (NIST), Certificate of Analysis for Standard Reference Material 870 - “Column Performance Test Mixture for Liquid Chromatography” at the NIST internet site http:/ois.nist.gov/srmcatalog/certificates/870.pdf in September 2002. The NIST test mixture, which is designed to characterize general aspects of HPLC was revised in December 2002. Comparative data may not be representative of all applications.

Peak Asymmetry Comparison

Importance of Peak Asymmetry (As)

Lower Tailing

Higher Tailing

As (Amitriptyline)

Si-OH

ACE® Stationary Phases Virtually Eliminate the Negative Effects of Silanols on UHPLC & HPLC Separations

Comparison of Leading 5µm C18 Columns- Data obtained from the National Institute of Standards and Technology (NIST), USA

■ Leading 5µm C18 column brands

■ Basic molecule testing

■ Peak asymmetry investigation

“Elution of organic bases (eg amitriptyline) with severe peak tailing is often associated with high silanol activity; however, the elution of such compounds with symmetrical peak shape is considered indicative of column deactivation.”

www.ace-hplc.com ACE UHPLC and HPLC Columns - Independent Comparisons

Comparison of Column Inertness at Intermediate pH- Reproduced with kind permission of The Open University, UK

■ Leading column brands in 50 x 2.1mm LC/MS compatible dimensions at pH 5.8

■ Silica, Hybrid and Superficially Porous particle technologies compared

■ Comparison of column efficiency for pyridine – a basic molecule

■ Efficiency measured at 5% peak height to account for peak tailing effects

Column Dimensions: 50 x 2.1mm (a50 x 2.0mm) Sample: 1) uracil 2) pyridine 3) phenol Mobile Phase: 30:70 MeOH/10mM NH4OAc in H2O (pH 5.8) Flow Rate: 0.20ml/min Temperature: 22°C Wavelength: 254nm Comparative data may not be representative of all applications.

Time - Minutes

Peak Efficiency Comparison

68,900pl/m

42,400pl/m

41,900pl/m

30,800pl/m

30,700pl/m

26,800pl/m

25,600pl/m

21,900pl/m

21,200pl/m

20,700pl/m

10,300pl/m

7,300pl/m

ACE Excel 2 SuperC18

ZORBAX Eclipse 1.8 XDB C18

Agilent Poroshell 2.7 120 EC C18

Waters XBridge 2.5 BEH C18 XP

Waters ACQUITY 1.7 BEH C18

Phenomenex Kinetex 2.6 C18

Waters XBridge 2.5 C18

Phenomenex Luna 2.5 C18(2) HSTa

NUCLEOSHELL 2.7 RP18a

Phenomenex Kinetex 1.7 C18

ZORBAX Extend 1.8 C18

Thermo Accucore 2.6 C18

0 20,000 40,000 60,000 80,000 100,000

N0.05 (pyridine) (plates/m)

Application # 1513

www.ace-hplc.com ACE UHPLC and HPLC Columns - Independent Comparisons

Independent Testing Comparison #2: UHPLC Columns

Higher Efficiency

Lower Efficiency

ACE Excel 2 SuperC18(fully porous ultra-inert silica)N0.05 (pyr) = 68,900pl/m

1

2

3

0 1 2 3 4 5

Waters ACQUITY 1.7 BEH C18(hybrid particle)

N0.05 (pyr) = 30,700pl/m

1

2

3

Time - Minutes0 1 2 3 4 5

Phenomenex Kinetex 1.7 C18(core-shell particle)

N0.05 (pyr) = 20,700pl/m

1

2

3

Time - Minutes

0 1 2 3 4 5

“Independent Testing Shows ACE UHPLC Columns Deliver Outstanding Efficiency”

Independent Testing Comparison #3: Solid-Core Particles

Column Dimensions: 50 x 2.1mm Sample: 1) uracil 2) pyridine 3) phenol Mobile Phase: 30:70 (v/v) MeOH/10mM NH

4OAc in H

2O (pH 5.8) Flow Rate: 0.20ml/min Temperature: 22°C Wavelength: 254nm

Comparative data may not be representative of all applications.

ACE UltraCore SuperC18 Provides Exceptional Efficiency- Reproduced with kind permission of The Open University, UK.

■ Leading column brands from major manufacturers investigated

■ Comparison of column efficiency for pyridine – a basic molecule

ACE UltraCore 2.5 SuperC18

Phenomenex Kinetex 1.3 C18

Phenomenex Kinetex 1.7 C18

Waters Cortecs 1.6 C18

Phenomenex Kinetex 2.6 C18

Thermo Accucore 2.6 C18

117 bar

450 bar

235 bar

285 bar

157 bar

117 bar

Pressure Efficiency

0 20,000 40,000 60,000 80,000

N0.5 (pyridine) (plates/m)

49,100pl/m

44,400pl/m

35,600pl/m

35,200pl/m

34,800pl/m

19,200pl/mLower Efficiency

Higher Efficiency

Peak Efficiency Comparison Application # 1802

■ Improved efficiency and peak shape has a direct effect upon signal response

Reduced Tailing - Increased Peak HeightIncreased

Tailing - Reduced

Peak Height

IncreasedResponse

Improve your MS signal response by selecting

high efficiency, low tailing ACE UHPLC and

HPLC Columns

ACE® UHPLC and HPLC columns are available through our international distributor network:

UHPLC and HPLC Columns

www.ace-hplc.com

Advanced Chromatography Technologies Limited, 1 Berry Street, Aberdeen, AB25 1HF, ScotlandTel: +44 (0) 1224 704 554 • Fax: +44 (0) 1224 841 301 • email: [email protected] • www.ace-hplc.com

ACE®

“Independent Testing Shows ACE Solid-Core Columns Deliver Outstanding Performance”

Reduce Peak Tailing to Improve MS Signal Response