Page 1: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC CURRENT OSCILLOSCOPE PROBE SL261

E N G L I S H User Manual

Page 2: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

Serial #: ________________________________

Catalog #: 1201.51

Model #: SL261

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

®

I N S T R U M E N T S

Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

Page 3: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

Table of Contents

1. INTRODUCTION ............................................................................... 21.1 International Electrical Symbols ................................................31.2 DefinitionofMeasurementCategories .....................................31.3 ReceivingYourShipment ..........................................................31.4 OrderingInformation .................................................................4

1.4.1 AccessoriesandReplacementParts ............................4

2. PRODUCT FEATURES ...................................................................... 52.1 Description ................................................................................52.2 Compatibility .............................................................................52.3 Features ....................................................................................6

3. SPECIFICATIONS............................................................................. 73.1 ElectricalSpecifications ............................................................73.2 MechanicalSpecifications ........................................................83.3 EnvironmentalSpecifications ....................................................93.4 SafetySpecifications ................................................................93.5 TypicalResponseCurves .......................................................10

4. OPERATION .................................................................................. 124.1 ZeroAdjustment ......................................................................124.2 CurrentMeasurement .............................................................124.3 Battery Indication (Green LED) ...............................................134.4 Peak Overload (OL) Indication (Red LED) .............................13

5. MAINTENANCE ............................................................................. 145.1 Maintenance ...........................................................................145.2 Cleaning ..................................................................................145.3 ReplacingtheBatteries ...........................................................15

RepairandCalibration ...........................................................................16

TechnicalandSalesAssistance ............................................................16

Limited Warranty ...................................................................................17

WarrantyRepairs ...................................................................................17

Page 4: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

2 AC/DC Current Oscilloscope Probe Model SL261

CHAPTER 1

INTRODUCTION

WARNING The safety warnings are provided to ensure the safety of personnel and proper operation of the instrument. Read the instruction completely.

• Usecautiononanycircuit:potentiallyhighvoltagesandcurrentsmaybepresentandmayposeashockhazard.

• Donotusetheprobe ifdamaged.Alwaysconnect thecurrentprobetothemeasuringdevicebeforeitisconnectedaroundtheconductor

• Donotuseonnoninsulatedconductorwithapotentialtogroundgreaterthan600VCATIIIpollution2.Useextremecautionwhenclampingaroundbareconductorsorbusbars.

• Beforeeachuse,inspecttheprobe;lookforcracksinhousingoroutputcableinsulation.

• Donotuseclampinwetenvironmentorinlocationsthathazard-ousgasesexist.

• Donotusetheprobeanywherebeyondthetactilebarrier.

Page 5: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 3

1.1 International Electrical Symbols

This symbol signifies that the instrument is protected by double or rein-forced insulation.This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.This symbol refers to a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is per-mitted.Probe fitted with an electronic output limiter providing protection against voltage surges caused by the accidental opening of the probe secondary circuit to 30V max. peak.

In conformity with WEEE 2002/96/EC.

1.2 Definition of Measurement Categories

CAT II: Formeasurements performed on circuits directly connected totheelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.

CAT III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.

CAT IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.

1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsareconsistentwith thepacking list.Notify yourdistributor of anymissing items. If theequipmentappearstobedamaged,fileaclaimimmediatelywiththecar-rierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.

Page 6: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

4 AC/DC Current Oscilloscope Probe Model SL261

1.4 Ordering InformationAC/DCCurrentProbeModelSL261 (10A-100mV/A,100A-10mV/A,BNC) ..................................Cat. #1201.51Includes a user manual.

1.4.1 Accessories and Replacement Parts

Adapter-BNCAdapterforusewithAC/DCCurrentProbe ModelSL261&Meters8220,8333,8335&8336 ..............Cat. #2140.40

Page 7: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 5

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionThe SL261 AC/DC Current Oscilloscope Probe expands oscilloscopeapplicationsinindustrial,automotiveorpowerenvironments,andisidealfor analysis andmeasurement of distorted currentwaveformsandhar-monics.Theprobepermitsaccuratedisplayandmeasurementofcurrentsfrom100mAto100Arms,DCto100kHzwithoutbreakingintothecircuit.TheprobeusesHalleffecttechnologytomeasureACandDCsignals.Theprobeconnectsdirectlytoanoscilloscopethrougha2metercoaxialcablewithaninsulatedBNC.

2.2 CompatibilityTheModelSL261iscompatiblewithanyanalogordigitaloscilloscopeorothervoltage-measuringinstrumentwhichhasthefollowingfeatures:

• BNCinputconnector

• Rangecapableofdisplaying0.2to0.5Vperdivision

• Minimuminputimpedanceof1MΩ

Page 8: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

6 AC/DC Current Oscilloscope Probe Model SL261

2.3 Features

ZERO

OFF

I

BatteryCompartment

Screw

Range Selection Switch

Power Indication LED

Zero Adjust Knob

Current Flow for DC Measurementsor Supply to Load for AC Measurements

BatteryCompartment Cover

Ø = 0.46"(11.8 mm)

Peak OverloadIndication LED

Keep hands belowthis point

100 mV/A

10 mV/A

WARNING: Always keep hand on the handle part of the probe.

Page 9: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 7

CHAPTER 3

SPECIFICATIONS

3.1 Electrical Specifications*Referenceconditions:23°C±5°C,20to75%RH,DCto1kHz,probezeroed,1minutewarmup,batteriesat9V±0.1V,externalmagneticfield<40A/m,noDCcomponent,noexternalcurrentcarryingconductor,1MΩ/100pFload,conductorcenteredinjaw.

Current Range: 100mV/A:100mAto10Apeak10mV/A:1to100Apeak

Output Signal:1000mVpeakmax

AC Current Accuracy:Aftercalibrationandforoneyear(zeroprobebeforemakingmeasurement)

Range Accuracy Phase Shift100mV/A

(50mA to 10AAC/DCpeak) ±3%ofreading±50mA <1.5°fromDCto65Hz

10mV/A (500mA to 40AAC/DCpeak) ±4%ofreading±50mA

<1°fromDCto65Hz10mV/A (40A to 80ADCpeak) 12%ofreading±50mA

10mV/A (80A to 100AACpeak) 15%ofreading

Frequency Range: DCto100kHz(-3dBwithcurrentderating)

Noise:Range10mV/A:480µVRange100mV/A:3mV

Slew Rate: Range10mV/A:20mV/µsRange100mV/A:0.3V/µs

Load Impedance: >1MΩ/100pF

Page 10: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

8 AC/DC Current Oscilloscope Probe Model SL261

Insertion Impedance (50/60 Hz): 100mV/A:3µs 10mV/A:0.01Ω

Rise or Fall Time:Range100mV/A:3µs Range10mV/A:4µs

Working Voltage: 600Vrmsmax

Common Mode Voltage: 600Vrmsmax

Influence of Adjacent Conductor: < 0.2mA/A AC

Influence of Conductor Position in Jaw: 0.5%ofreadingatkHz

Battery: 9Valkaline(NEDA1604A,IEC6LR61)

Low Battery: GreenLEDwhen≥6.5V

Overload Indication:RedLEDindicatesinputgreaterthantheselectedrange

Typical Consumption: 8.6mA

Battery Life: 55hourstypical

3.2 Mechanical SpecificationsZero Adjustment: 20turnpotentiometerMaximum Cable Diameter: 0.46" (11.8 mm)

Case Protection: IP20perIEC529Drop Test: 1.0mon38mmofoakonconcrete;testaccordingtoIEC1010

Mechanical Shock: 100G;testperIEC68-2-27Vibration: TestperIEC68-2-6,frequencyrange10Hzto55Hz,amplitude0.15mm

Handle: Lexan®920A,UL94V2

Dimensions: 9.09x1.42x2.64"(231x36x67mm)

Weight: 11.6oz(330g)withbatteryColor: LightgrayOutput Lead: InsulatedcoaxialcablewithinsulatedBNCconnectorCable Length: 6.5foot(2m)

Page 11: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 9

3.3 Environmental SpecificationsOperating Temperature:0°to50°C(32°to122°F)Storage Temperature:-30°to80°C(-22°to176°F)Operating Relative Humidity:10°to30°C:85±5%RH(withoutcondensation)40°to50°C:45±5%RH(withoutcondensation)

Influence of Temperature: <0.2%per°C

Altitude:Non-operating:0to12,000m Operating:0to2000m

3.4 Safety Specifications

E192383

Electrical:DoubleinsulationorreinforcedinsulationbetweentheprimaryorsecondaryandtheoutercaseofthehandleperEN61010-2-032.

600V,CATIII,PollutionDegree:2

Electromagnetic Compatibility: EN 50081-1 Class BEN50082-2ElectrostaticdischargeIEC1000-4-2 Radiated Field IEC 1000-4-3 Fast Transients IEC 1000-4-4 MagneticFieldat50/60HzIEC1000-4-8

*Specifications are subject to change without notice

Page 12: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

10 AC/DC Current Oscilloscope Probe Model SL261

3.5 Typical Response Curves

0

-1

-2

-3

-4

-5

Frequency

Frequency in Hz

range 100 mV/A

range 10 mV/A

frequency response

measurement current: 1A sine

100 1000 10000 100000gain in dB

0

-10

-50

-100

-150

Phase Shift

Frequency in Hz

range 100 mV/A

range 10 mV/A

phase graph

measurement current: 1A sine

100 10000 100000phase in °

Page 13: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 11

0%

-2%

-4%

-6%

-8%

-10%

-12%

-14%

-16%

0A 20A 40A 60A 80A 100AAccuracy

Unc

erta

inty

Typical

Specification

Page 14: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

12 AC/DC Current Oscilloscope Probe Model SL261

CHAPTER 4

OPERATION

4.1 Zero AdjustmentTheprobehasazeroadjustmentthatshouldbeadjustedbeforemeasure-ment.Alternatively,youmay“zero”withtheoscilloscopeinstead.

4.2 Current Measurement

WARNING: Always connect the probe to the instrument before clamping onto the circuit under test.

• ConnecttheModelSL261totheproperinputchannelontheoscil-loscope.

• Beginwiththeleastsensitiverangeonthecurrentprobe(10mV/A)

• Selectthe0.5V/Divisionrangeontheoscilloscope.

• Clamptheprobeontheconductortobemeasuredandreadthecurrentflowingdirectlyonyouroscilloscope.

(Remember to unclamp the probe from the conductor before discon-necting it from your meter or instrument.)

Youmayalsouseyouroscilloscopetoamplifythesignalwhileusingthe100mV/Aprobe range (whichoffers thebestaccuracyand leastphaseshift).

NOTE: It is possible to change the range on the current probe without remov-ing the probe from the current carrying conductor, but it is important to remember not to exceed the permissible peak ratings of 1000mV peak or 2000mV peak to peak maximum. The peak ratings by range are: 10A peak on the 100mV/A range, 100A peak on the 10mV/A range.

Page 15: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 13

4.3 Battery Indication (Green LED)TheprobehasabatteryconditionLED.Toensureproperreadingswithyourcurrentprobe,besurethatthegreenLEDislitduringmeasurement.Ifnot,replacethe9Vbattery.

4.4 Peak Overload (OL) Indication (Red LED)TheSL261offersanoverloadindicator.IftheredLEDilluminatesduringmeasurement,thisindicatesthatthepeakvalueexceedstheinstrumentresponselevelandthattheoutputisdistorted.Switchtheprobetoahigherrangeifpossible.

Page 16: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

14 AC/DC Current Oscilloscope Probe Model SL261

CHAPTER 5

MAINTENANCE

5.1 Maintenance

WARNING• Toensureoptimumperformance,itisimportanttokeeptheprobe

jawmatingsurfacescleanatalltimes.Failuretodosomayresultinerrorinreadings.

• Formaintenanceuseonlyspecifiedfactoryreplacementparts.

• Toavoidelectricalshock,donotattempttoperformanyservicingunlessyouarequalifiedtodoso.

• Toavoidelectricalshockand/ordamagetotheinstrument,donotallowwaterorotherforeignsubstancesintothecase.

• Disconnecttheunitfromallcircuitsandtestcablesbeforeopen-ingthecase.

5.2 Cleaning• Tocleantheprobebody,useasoftclothdampenedinasolution

ofmilddetergentandwater.Tocleanthecore,openthejawandcleantheexposedcoresurfaceswithacottonswabdampenedwith isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol orethanol).Lubricatethejawsmatingsurfaceswithalightoil.

• Do not use chemicals containing benzine, benzene, toluene,xylene,acetone,orsimilarsolvents.

• Donotimmersetheprobeinliquidsoruseabrasivecleaners.

Page 17: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 15

5.3 Replacing the Batteries

CAUTION: Risk of electric shock. Disconnect all input(s) from the unit or remove the probe from any conductor before opening the battery cover to change the batteries.

If thegreenbattery indicationLEDdoesnot lightupwhen theprobe isturnedon,thebatterieswillneedtobereplaced.

Donotreplacethebatterywhileprobeisinuse.

Toreplacethebattery:

• Disconnecttheprobefromthecircuitandtheoscilloscope.

• Turntheprobe“OFF”.

• Unscrewthebatterycompartmentscrewandpulloutthebatterycom-partmentcover.

• Replacethebatteryandputthecoverbackon.

Page 18: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

16 AC/DC Current Oscilloscope Probe Model SL261

Repair and Calibration

Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508) 698-2115 Fax: (508)698-2118 E-mail:[email protected] www.aemc.com

NOTE: Do not ship Instruments to our Foxborough, MA address.

Page 19: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

AC/DC Current Oscilloscope Probe Model SL261 17

Limited Warranty

TheAC/DCCurrentOscilloscopeProbeModelSL261iswarrantedtotheown-erforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.

Please print the online Warranty Coverage Information for your records.

What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

REGISTER ONLINE AT:www.aemc.com

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.

NOTE: You must obtain a CSA# before returning any instrument.

Page 20: AC/DC CURRENT OSCILLOSCOPE PROBEg... · 2019-08-08 · AC/DC Current Oscilloscope Probe Model SL261 5 CHAPTER 2 PRODUCT FEATURES 2.1 Description The SL261 AC/DC Current Oscilloscope

11/15

99-MAN100042v15

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346

www.aemc.com