Accuracy of ultrasound sonography in adenomyosis

39
Accuracy of TVUS in Adenomyosis

description

Can ultrasonography diagnose adenomyosis? this talk may present an answer to this

Transcript of Accuracy of ultrasound sonography in adenomyosis

Page 1: Accuracy of ultrasound sonography in adenomyosis

Accuracy of TVUS in Adenomyosis

Page 2: Accuracy of ultrasound sonography in adenomyosis

Introduction

• Adenomyosis is a frequent gynecologic pathology which affects women particularly in the fifth decade complaining of abnormal uterine bleeding.

Page 3: Accuracy of ultrasound sonography in adenomyosis

• Adenomyosis is thought to affect 1% of women.

• The aetiology is unclear.

Page 4: Accuracy of ultrasound sonography in adenomyosis

clinical presentation

• parous, perimenopausal patient with dysmenorrhea and menorrhagia.

• Physical examination may reveal an enlarged, tender uterus,

Page 5: Accuracy of ultrasound sonography in adenomyosis

Adenomyosis

• clinical presentation may suggest adenomyosis • imaging is often used prior to treatment

selected

Page 6: Accuracy of ultrasound sonography in adenomyosis

• but the diagnosis has been difficult to obtain without histologic confirmation.

Page 7: Accuracy of ultrasound sonography in adenomyosis

Invasive

Page 8: Accuracy of ultrasound sonography in adenomyosis

TVUS

• Available• Non invasive• Low cost• Operator dependent• ?? accurate

Page 9: Accuracy of ultrasound sonography in adenomyosis
Page 10: Accuracy of ultrasound sonography in adenomyosis

• asymmetric thickening of the anterior or posterior uterine wall or myometrial cysts which are small anechoic lakes of 2-6 mm in diameter which may represent cystic endometrial implants or focal myometrial hemorrhage, colour doppler can be used to differentiate it from myometrial vessels.

Page 11: Accuracy of ultrasound sonography in adenomyosis
Page 12: Accuracy of ultrasound sonography in adenomyosis
Page 13: Accuracy of ultrasound sonography in adenomyosis
Page 14: Accuracy of ultrasound sonography in adenomyosis

Aim of the study

• to measure the accuracy of the TVUS features in diagnosing adenomyosis

• to determine the diagnostic role of uterine artery Doppler in these cases.

Page 15: Accuracy of ultrasound sonography in adenomyosis

• Study design:a prospective comparative study.• Setting: Cairo University hospital

Page 16: Accuracy of ultrasound sonography in adenomyosis

363 Women

• scheduled for hysterectomy either abdominal, vaginal

• TVUS was done to detect adenomyosis • then compare the results with

histopathological findings• then evaluate if there is any relation of

adenomyosis with uterine artery Doppler indices

Page 17: Accuracy of ultrasound sonography in adenomyosis

352 women

• 11 patients were excluded from the study either because that they diagnosed as being unfit for surgery.

• 198 Patients were premenopausal (56.3 %) while 154 were postmenopausal (43.8 %).

Page 18: Accuracy of ultrasound sonography in adenomyosis

TVUS

• Ultrasound scanning was performed with Accuvix (Medison, Seoul, Korea) scanner using 4-7 MHZ endovaginal probe.

• Scanning was done in early morning to avoid fluctuations due to circadian rhythm of uterine artery blood flow.

Page 19: Accuracy of ultrasound sonography in adenomyosis

Moreover

• There are different ultrasonographic features which are suggested by many studies in order to reach a diagnosis of adenomyosis.

• This lead to different accuracy in transvaginal detection of adenomyosis

Page 20: Accuracy of ultrasound sonography in adenomyosis

Criteria examined

• uterine diameters, outline• myometrial echotexture,• myometrial linear striation,• myometrial cysts (its number if present),• endometrial myometrial junction,• uterine wall asymmetry and if there was any

associated gynecological pathology

Page 21: Accuracy of ultrasound sonography in adenomyosis

Criteria Details

• Myometrial cysts were defined as rounded anechoic area of 2-6 mm in diameter .

• Color Doppler was used to differentiate the myometrial cysts from myometrial blood vessels.

• Heterogenous myometrium was defined by the presence of an indistinctly marginated myometrial area with decreased or increased echogenicity

Page 22: Accuracy of ultrasound sonography in adenomyosis
Page 23: Accuracy of ultrasound sonography in adenomyosis

More details

• heterotopic endometrium extending into the inner myometrium can appear as echogenic linear striations. When these lines are small or indistinct, pseudo-widening of the endometrium or poor delineation of the endomyometrial junctional zone is seen.

• Globular and/or asymmetric uterus was defined as a regular enlarged uterus with possible myometrial asymmetry unrelated to leiomyoma.

Page 24: Accuracy of ultrasound sonography in adenomyosis

Histopathology

• the pathologist was blinded to the US and clinical findings.

• adenomyosis was diagnosed by migration of the endometrial glands from the stratum basale into the myometrium. The ectopic glands tend to be at least 2-3 mm below the endometrial-myometrial junction.

Page 25: Accuracy of ultrasound sonography in adenomyosis

• Adenomyosis was found in 37/352 (10.5 %) patients at the histopathological examination. Other pelvic pathology was associated in 10/37 of the adenomyotic cases (27.77%).

Page 26: Accuracy of ultrasound sonography in adenomyosis

• Leiomyomas were the most commonly associated gynecological pathology with adenomyosis (6/37, 16.2%) then endometrial hyperplasia (3/37, 8.3 %) and ovarian cyst (1/37, 2.7%).

Page 27: Accuracy of ultrasound sonography in adenomyosis

• Transvaginal ultrasound diagnosed adenomyosis in 48 cases out of 352 cases scheduled for hysterectomy, of whom, only 37 cases had a histopathological diagnosis of adenomyosis. (~10%)

Page 28: Accuracy of ultrasound sonography in adenomyosis

• The sensitivity, specificity, negative predictive value, positive predictive value and accuracy of transvaginal ultrasound in the diagnosis of adenomyosis were: 7 5.68%, 90.79%, 49.12%, 96.95% and 89.20 %respectively.

Page 29: Accuracy of ultrasound sonography in adenomyosis

Adenomyosis

(n = 37)

No adenomyosis

(n = 315)

P

Globular uterus

Yes

No

25(67.6)

12(32.4)

228(72.4)

87(27,6)

0.538

Asymmetrical anterior-posterior uterine wall

Yes

No

12(32.4)

25(67.6)

86(27.3)

229(72.7)

0.510

Heterogeneous myometrial echotexture

Yes

No

35(94.6)

2(5.4)

175(55.6)

140(44.4)

< 0.001

Poor Endometrial Delineation

Yes

No

28(75.7)

9(24.3)

144(45.7)

171(54.3)

< 0.001

Linear myometrial striation

Yes

No

18(48.6)

19(51.4)

29(9.2)

286(90.8)

< 0.001

The Presence of myometrial cysts

Yes

No

26(70.3)

11(29.7)

144(45.7)

171(54.3)

0.005

Page 30: Accuracy of ultrasound sonography in adenomyosis

Accuracy of different ultrasonographic features for diagnosing adenomyosis

Sensitivity Specificity PPV NPV accuracyHeterogeneous myometrial echotexture

94.59 44.44 16.67 98.59 49.72

Poor Endometrial Delineation 75.68 54.29 16.28 95.00 56.53

Linear myometrial striation 48.65 90.79 38.30 93.77 86.36

The Presence of myometrial cysts 70.27 54.29 15.29 93.96 55.97

Myometrial cysts > 2 96.15 100.00 100.00 99.31 99.41

Page 31: Accuracy of ultrasound sonography in adenomyosis

ultrasonographic features(myometrial cysts (arrows) and linear striations

Page 32: Accuracy of ultrasound sonography in adenomyosis

Doppler study of the uterine artery in a case documented to have adenomyosis.

Page 33: Accuracy of ultrasound sonography in adenomyosis

Comparison of uterine artery Doppler indices (RI and PI)

Adenomyotic cases

(n = 37)

Non adenomyotic cases

(n = 315)

P value

Uterine artery RI:

- Min- Max- Mean- SD

0.72

0.93

0.865

0.051

0.75

0.94

0.847

0.063

0.107

Uterine artery PI:

- Min- Max- Mean- SD

1.35

3.10

2.473

0.472

1.65

3.01

2.317

0.539

0.093

Page 34: Accuracy of ultrasound sonography in adenomyosis

• no previous study measure the uterine artery Doppler in cases of adenomyosis and compare them with those who are not documented histopathologicaly to have it ,

• Also there was no other study specify the number of myometrial cysts in cases of adenomyosis

Page 35: Accuracy of ultrasound sonography in adenomyosis

Other studies

Prevalence Sensitivity Specificity PPV NPV

Siedler et al. 8/80(10) 63 97 71 _

Fedele etal 22/43 (51) 80 74 73 81

Asher etal 17/20 (85) 86 50 90 20

Reinholdet al. 29/100 (29) 86 86 71 94

Brosen et al 28/56 (50) 53 75 86 77

Reinhold et al. 18/119 (`15) 89 89 71 96

Atzori et al. 15/175 (8.6) 86 96.2 68.4 98

Vercelini et al. 29/102(28) 82.7 67 50 90.7

Atri et al. 30/102(29.4) 81 71 54 90

Bazot et al. 40/120(33) 65 97.5 92.8 88.8

Bazot et al. 23/106(21.7) 80.9/38.4 100/97.5 100/83.3 40/82.9

Kepkek et al. 26/70(37.1) 80.8 61.4 55.3 84.4

Our study 37/352(10.5) 75.6 90.79 49.12 96.95

Page 36: Accuracy of ultrasound sonography in adenomyosis

• This study had the privilege of including a large sample volume but still there may be a need of further studies with larger sample volume to put a scoring system including both the clinical and ultrasonographic criteria for diagnosis of adenomyosis with a high accuracy.

Page 37: Accuracy of ultrasound sonography in adenomyosis

• We try to find if we can add uterine artery Doppler in the ultrasound features for adenomyosis but there was no statistical difference in either uterine artery Doppler indices (RI,PI) in both adenomyotic cases and non adenomyotic ones.

Page 38: Accuracy of ultrasound sonography in adenomyosis

Limitations

• different pathologists examine the hysterectomy specimens

• in cases where there is focal adenomyosis • the inclusion of only hysterectomized cases

Page 39: Accuracy of ultrasound sonography in adenomyosis

Conclusion

• subendometrial linear echogenic striations showed the highest specificity and accuracy .

• The myometrial cysts show a high negative predictive value and their number (if more than 2) represent the highest accuracy.

• Uterine artery Doppler study can't be used to diagnose adenomyosis.