Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

download Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

of 24

Transcript of Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    1/24

    ABSORPTION AND FLUORESCENCE SPECTROSCOPIES OF GREEN

    FLUORESCENT PROTEINAlgodon SML, Tamon LU and Tutor MV

    Absorption and fluorescence spectroscopies are both useful in characterizing

    protein samples. This experiment aimed to employ the Bradford assay coupledwith absorption spectroscopy to determine the concentration of recombinant

    green fluorescent protein (GFP) extract while fluorescent spectroscopy was done

    to determine the effect of extreme heat and pH to the fluorescence intensity.

    Absorbance readings of GFP, along with Bovine Serum Albumin protein standard,

    were measured at 595 nm. Using the BSA calibration curve, the mean

    concentration of the extract was calculated to be 1.658 mg/mL. The molar

    extinction coefficient was then determined to be 0.4604

    . In fluorescence

    spectroscopy, 0.02 M protein was optimal for monitoring fluorescence. The

    intensity (RFU) was observed at temperatures from 35C to 95C (increments of 10

    C), at pH 1.40 to 3.0 using HCl and at pH 11.0 to 12.30 using NaOH. Recovery of

    fluorescence was tested by re-exposing the protein to room temperature after heattreatment and by neutralizing the acid and base added. Results show that the

    melting temperature of the GFP extract was 55C. A decreasing trend of

    fluorescence was observed with increasing degree of all stress factors. The least

    fluorescence was observed at highly acidic conditions. For the recovery, the

    transition from 35C to room temperature (~23C) gave the highest recovery

    (71.48%) while that of 95C to room temperature gave the second lowest (1.042%).

    The basic condition had 10.86% recovery while the acidic condition gave the

    lowest, 0.01839%. Overall, fluorescence was most affected by the highly acidic

    environment.

    I. Introduction

    Spectroscopy deals with the interaction of radiated energy with matter resulting to the redirection

    of the radiation or transitions between the energy levels of the atoms or molecules. It came from the

    words spectrum, which pertains to a range of related qualities, and skopein, which means to

    look. Spectroscopic data, which is represented by a spectrum, is characterized by the response of the

    matter to electromagnetic radiation as a function of wavelength or frequency. Spectroscopic techniques

    often utilize the absorption, emission or scattering of electromagnetic radiation by matter to quantitatively

    and qualitatively assess the characteristics of the matter. The matter can be atoms, molecules, ions, or

    solids. Spectroscopy is also used to elucidate the components of a substance since the spectral lines

    obtained from conducting a spectroscopic technique are characteristic of elements. In the field of

    biochemistry, the device used for the analysis related to spectroscopy is a spectrophotometer consisting

    of electromagnetic radiation source, monochromator, sample holder and detector (Dinh, 2003).Due to the wide scope of spectroscopy, spectroscopic techniques have been classified into

    different categories based on the type of electromagnetic radiation used, type of interaction between the

    material and electromagnetic radiation and the type of material used. Two of the most common

    spectroscopic techniques are absorption and emission spectroscopies which are both useful in analyzing

    samples (Dinh, 2003).

    Absorption spectroscopy refers to the absorption of light through interactions with the sample.

    There are several types of absorption spectrometric techniques, such as ultraviolet-visible, infrared,

    atomic, and x-ray absorption spectrometry. These techniques vary mainly on the electromagnetic

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    2/24

    wavelength used. In absorption spectroscopy, the sample absorbs energy (photon) as a function of

    wavelength or frequency of light, generating an absorption spectrum. Absorption spectroscopy is

    particularly useful for detecting the presence of a substance in a sample and determining the amount of

    that substance in the sample. Analysis of data obtained by absorption spectroscopy of a sample to

    determine the concentration of a substance entails the application of Beers Law which states that the

    transmission (or transmissivity), T, of the light through a sample varies logarithmically with the product of

    absorbance coefficient, , for a certain substance and the distance travelled by the light or the path

    length, l(Hollas, 2004). From this, the equation for Beers Law isderived and is given by

    (1)On the other hand, the emission spectroscopy of a chemical substance pertains to the emission

    of electromagnetic radiation of a specific wavelength by an atom due to its transition from a higher energy

    level to a lower energy level. The energy emitted by an atom is approximately equal to the energy

    difference between the two energy levels. The occurrence of a wide variety of transitions and energy

    difference associated with them creates the electromagnetic spectrum characteristic of an atom. Thus,

    emission spectroscopy can be used to determine the presence of atoms in a substance of unknown

    composition and the atoms comprising a molecule can be elucidated. One particular type of emission

    spectroscopy is fluorescence spectroscopy, which deals with using light to excite molecules in a sample,

    causing these molecules to emit light, usually visible light, that can be detected (Lakowicz, 2006).In this experiment, absorbance and fluorescence spectroscopies were conducted on green

    fluorescent protein (GFP). The green fluorescent protein (GFP) is a fluorescent protein that can be

    isolated from the Pacific jellyfish Aequoria victoria and many other marine organisms. Its fluorophore

    originates from the internal Ser-Tyr-Gly sequence that is post-translationally modified to 4-(p-

    hydroxybenzylidene)-imidazolidin-5-one (See Figure 1). However, it should be noted that fluorescence is

    not an intrinsic property of the Ser-Tyr-Gly tripeptide. The cyclisation of the peptide and the oxidation of

    the tyrosine residue results to the formation of p-hydroxybenzylidene-imidazolidone structure which is

    responsible for the fluorescent phenotype (Yang et al., 1996). This fluorophore can exist in two resonant

    forms: (1) with a partial negative charge on the benzyl oxygen of the tyrosine and (2) with the charge on

    the carbonyl oxygen of the imidazolidone ring. These oxygen atoms form interactions with the basic

    residues His148 (with Tyr66) and Gln94 and Arg96 (with the imidazolidone) that stabilize and allow

    further delocalization of the charge on the fluorophore (Yang et al., 1996).

    Figure 1. Fluorescent structure of GFP. The formation of the 4-(p-hydroxybenzylidene)-imidazolidin-5-

    one structure of the Ser-Tyr-Gly tripeptide allows the protein to exhibit fluorescence (Yang et al., 1996).

    Aside from spectroscopy, Bradford reaction was used to verify the concentration of the samplethrough comparison with a standard. The change in color of Bradford reagent when it binds to proteins

    allows the sample to absorb light at 595 nm. The degree of absorption of light is directly proportional to

    the concentration of protein in the sample. Different factors, such as temperature, basic and acidic

    environments, that can affect the fluorescence of the sample, were also tested (Bradford, 1976).

    The experiment aimed to determine the concentration of the GFP sample through Bradford assay

    which involves absorption spectroscopy. Fluorescence spectroscopy was then employed to determine the

    effects of stress factors particularly heat and extreme pH to the structure, function and, consequently, the

    fluorescence of the protein sample.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    3/24

    II. Methodology

    Preparation of protein samples for absorption spectroscopy

    The 31st elution fraction containing the protein of interest and which has the brightest

    fluorescence under a UV transilluminator was subjected to absorption spectroscopy. The protein solution

    was diluted with 20 mM Tris-Cl (pH 8).

    Quantification of protein concentration through a Bradford reaction

    A total of 10 mL of 1X Bradford reagent was prepared. Then, 13 500 L aliquots of Bradford

    reagent was placed in 1.5 mL microcentrifuge tubes. A protein standard sample set (bovine serum

    albumin or BSA) was also prepared in concentrations of 0, 0.2, 0.4, 0.6, 0.8, 1.0 and 1.2 mg/mL in ddH2O.

    Undiluted, 1:10 dilution and 1:100 dilution in Tris-Cl pH 8.0 of the test protein sample was also prepared.

    Afterwards, 10 L of protein sample was added to the 500 L aliquots of Bradford reagent. The

    absorbance of each sample was measured at 595 nm. Consequently, a calibration curve was obtained

    from the observed absorbance of the protein standards and was used to determine the concentration of

    the test protein samples.

    Fluorescence intensity testThe fluorescence from 525-570 nm with excitation at 460 nm of 1 mL sterile ddH2O placed in a

    plastic cuvette was measured. Subsequent additions of sufficient amount of protein to the 1 mL sterile

    ddH2O were conducted to produce a 1 M, 5M and 10 M concentration of protein solution, the

    absorbance values of which was also obtained at the same wavelength of light.

    Thermostability test

    For the thermostability test, 4 mL of the protein solution of unknown concentration was prepared.

    Afterwards, 1 mL of the protein solution was placed into two identical cuvettes; the rest was stored for

    other experiments. The fluorescence of the two samples (test and control) at 525-570 nm with excitation

    at 460 nm was obtained. The control sample was kept at room temperature while the other samples were

    subjected to 2 minute incubations at 35C, 45C, 55C, 65C, 75C, and 85C. The fluorescence of the

    two samples at 525-570 nm with excitation at 460 nm was again measured. The test sample was

    incubated at room temperature for 2 minutes after each reading. The fluorescence of the two samples at

    525-570 nm with excitation at 460 nm was then measured to test for signal recovery. The temperature at

    which 50% of fluorescence was lost and the temperature at which fluorescence is lost completely was

    determined.

    Resistance to acidic conditions

    To test the resistance of the protein to acidic conditions, 1 mL of the protein solution was placed

    into another cuvette. The fluorescence of the two samples (test and control) at 525-570 nm with excitation

    at 460 nm was then measured. Subsequent additions of 1M HCl to the test sample were conducted to

    obtain the following HCl concentrations: 1 mM, 2 mM, 5 mM, 10 mM, 20 mM and 40 mM. The

    fluorescence of the two samples (test and control) at 525-570 nm (excitation at 460 nm) was thenmeasured after each treatment. The buffer HCl concentrations at which 50% fluorescence was lost and at

    which fluorescence was lost completely was also determined. Finally, the acidic buffer conditions were

    neutralized with a series of addition of NaOH (1 mM, 2 mM, 5 mM, 10 mM, 20 mM, and 40 mM) and

    recovery of the fluorescent signal was tested after each treatment.

    Resistance to basic conditions

    For the resistance to basic conditions test, 1 mL of the protein solution was placed into another

    cuvette. The fluorescence of the two samples (test and control) at 525-570 nm (excitation at 460 nm) was

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    4/24

    obtained. The control sample was kept at pH 7.5 while sufficient 1M NaOH was added in succession to

    the test sample to get the following NaOH concentrations: 1 mM, 2 mM, 5 mM, 10 mM, 20 mM and 40

    mM. The fluorescence of the two samples (test and control) at 525-570 nm (excitation at 460 nm) was

    then measured after each treatment. The buffer NaOH concentrations at which 50% fluorescence was

    lost and at which fluorescence was lost completely was also determined. Finally, the basic buffer

    conditions were neutralized with the subsequent additions of HCl (1 mM, 2 mM, 5 mM, 10 mM, 20 mM,

    and 40 mM) and recovery of the fluorescent signal was tested after each treatment.

    III. Results

    Absorbance Spectroscopy

    Table 1. Mean absorbance values at 595 nm of

    Bovine Serum Albumin (BSA) protein standard

    added with Bradford reagent

    BSA

    Concentration

    mg/mL

    Mean Absorbance (at 595nm)

    ActualA Corrected

    B

    0 0.511 0

    0.2 0.558 0.047

    0.4 0.711 0.200

    0.6 0.827 0.316

    0.8 0.923 0.412

    1.0 0.997 0.486

    1.2 1.108 0.597A

    Individual absorbance readings are given in Appendix Supplementary Table 1B

    Corrected values were obtained by subtracting the actual mean absorbance (0.511) of 0 mg/mL BSA inBradford reagent to the actual values of the different concentrations of BSA.

    Table 1 shows the mean absorbance of the different concentrations of BSA in Bradford Reagent

    at 595 nm. Notably, the absorbance increased with increasing concentration. The absorbance values of

    those solutions were used to construct a standard curve for the Bradford reaction (See Figure 2 below)

    and the equation of the line is also given. The curve had a positive slope and a regression coefficient (R2)

    close to 1.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    5/24

    Figure 2. Standard Curve for Bradford Reaction with BSA as the protein standard

    The same relationship, which is the direct proportionality of absorbance to concentration, was

    observed for the mean absorbance values of the GFP positive control and test sample given in Table 2.

    As the GFP concentration is decreased by diluting the sample, absorbance also decreased. Unlike most

    of the values, there are absorbance values that are greater than one for both GFP positive and test

    sample and these values are italicized in Table 2 below. For instance, the actual mean absorbance of the

    undiluted GFP positive (1.982) is greater than 1 even after correction (1.471). Another observation is that

    the actual value for the 1:10 dilution of the GFP positive which is 1.056 became less than 1 after

    correction (0.545). The same is true for the undiluted GFP test sample in which the 1.445 absorbance

    value became 0.934. A negative corrected absorbance was obtained for the 1:100 dilution of the GFP test

    sample after correction (-0.046) with a corresponding negative value for protein concentration. The valuesare underlined below.

    Table 2. GFP concentrations of positive and test samples calculated using their mean absorbance

    and the BSA protein standard curve

    GFP Positive Control Test Sample

    Mean Absorbance

    (at 595 nm)

    Calculate

    d

    Concentr

    ation

    (1x)C

    e

    E

    Mean Absorbance

    (at 595 nm)

    Calculate

    d

    Concentr

    ation

    (1x)C

    e

    E

    ActualA

    CorrectedB

    ActualA

    CorrectedB

    Undiluted

    1.982 1.471 2.887 1.445 0.934 1.843 0.5066

    1:10 1.056 0.545 10.87 0.5347 0.572 0.061 1.473 0.4141

    1:100 0.546 0.035 9.677 0.3617 0.465 -0.046 -.6.128

    Mean 10.27D 0.4482 1.658

    D 0.4604

    AIndividual absorbance readings are given in Appendix A Supplementary Table 1

    BCorrected values were obtained by subtracting the actual mean absorbance (0.511) of 0 mg/mL BSA in

    Bradford reagent to the actual values of the different concentrations of BSA.

    y = 0.5146x - 0.0148R = 0.9904

    -0.1

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0 0.2 0.4 0.6 0.8 1 1.2 1.4

    Absorbanceat595nm

    BSA Concentration (mg/mL)

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    6/24

    CThe corrected absorbance values of the GFP solutions were used to derive the concentration.

    DThe extreme values (marked with strikethrough) were not included in the calculation.

    EIn calculating the molar extinction coefficients, the actual concentrations of the diluted solutions were

    used, not the derived value for the stock (or the 1x). Also, the extreme concentration values were not

    used to solve for e.

    The standard curve was used to calculate the protein concentrations of the GFP control and test

    samples shown in Table 2 (See Appendix A for the calculation). The mean concentration for the GFP

    positive control and test sample were calculated to be 10.27 mg/mL and 1.658 mg/mL, respectively.

    Table 2 also shows the calculated values for the molar extinction coefficient (e). The mean values for the

    control and test sample, 0.4482

    and 0.4604

    respectively, are close to each other.

    Fluorescence Spectroscopy

    A. Intensity Test

    Figure 3. Fluorescence Intensity of GFP at different concentrations. The fluorescence intensity was

    measured using a fluorometer at a wavelength range of 525 nm to 570 nm. Three readings at each

    concentration were obtained and the mean value was used to construct the graph. (See Appendix B for

    the actual values).

    The behavior of the graph in Figure 3 shows that fluorescence intensity increases with protein

    concentration. The sample with a protein concentration of 0.02 uM highlighted red in Figure 2 gave anintensity that is intermediate of the values given by other samples.

    0

    20000

    40000

    60000

    80000

    100000

    120000

    0 0.02 0.04 0.06 0.08 0.1 0.12

    FluorescenceIntensity(RFU)

    GFP Concentration (uM)

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    7/24

    B. Thermostability Test

    Figure 4. Effect of varying temperature to the fluorescence of GFP sample (0.2 uM). The same

    fluorometer was used and the wavelength range is 525 nm to 570 nm. Three readings at each

    temperature were obtained and the mean value was used to construct the graph. (See Appendix B for the

    actual values).

    ^ The initial intensity represented by the leftmost data points (orange and blue diamonds) were measured

    at room temperature (RT, 23C).

    *The control was not subjected at varying temperatures. It was constantly at room temperature; however,

    its fluorescence was also measured each time the test sample was read so it has readings for recovery.

    **After subjecting the test sample at higher temperatures, it was allowed to recover at RT for 2 minutes

    and its fluorescence was again measured. For the control, the same process was done but it was not

    subjected to a different temperature; it remained at RT.

    In Figure 4, the fluorescence intensity of the test sample decreased as the incubation

    temperature was increased. Abrupt decrease in fluorescence was observed from 35C to 45C but it

    became relatively uniform downstream of the graph. When GFP was allowed to recover at room

    temperature (23C) every after exposure to higher temperature, fluorescence was not reset to the initial

    value; in fact, the fluorescence readings after recovery also showed a decreasing trend. However, in

    general, the fluorescence after recovery was slightly stronger than that with heat treatment. An exception

    would be at 35C.

    Notably, approximately 50% of the initial fluorescence of the test sample was lost at 55C;however, complete loss of fluorescence was not observed. The lowest intensity value was 73.54 RFU.

    Relative to the test sample, the control, which was kept at room temperature, gave fluorescence

    values that did not change drastically. The values ranged from 8000 to 11000 RFU while the test sample

    reached below 100 RFU. Consistently, the control which was subjected at a lower temperature than all

    the heat treatments had stronger fluorescence than the test sample.

    Initial intensity(RT)^

    0

    2000

    4000

    6000

    8000

    10000

    12000

    0 20 40 60 80 100

    FluorescenceIntensity(RF

    U)

    Temperature (C)

    Test Sample

    Test Sample (Recovery)**

    Control*

    Control (Recovery)**

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    8/24

    Figure 5. Effect of varying HCl concentration to the fluorescence of GFP sample (0.2 uM). The

    same fluorometer was used and the wavelength range is 525 nm to 570 nm.Three readings at each HCl

    concentration were obtained and the mean value was used to construct the graph. (See Appendix B for

    the actual values).

    ^After reaching the 40mM HCl concentration, NaOH was added to successively lower the acid

    concentration again to 20, 10, 5, 2, 1 and 0 mM.

    *GFP was suspended in sterile distilled water as well as the test samples.

    Based on Figure 5, a 1mM HCl concentration (pH 3) already caused a sudden decrease in

    fluorescence intensity. As acid concentration was further increased up to 40 mM (pH 1.40), intensitycontinued to decrease at smaller increments. In contrast, the control showed a curved plot that is

    gradually decreasing. Negative intensity values of the test sample, -2.4067, -3.1467 and -3.3933 RFU,

    were obtained for 10, 20 and 40 mM HCl, respectively. When NaOH was added, original fluorescence at

    10 to 40 mM HCl were attained; but, the sample did not recover its initial intensity upon complete

    neutralization of the added acid. Before addition of the acid, the sample gave a fluorescence intensity

    value of 11078 RFU and upon complete neutralization, the sample gave a fluorescence intensity value of

    2.037 RFU. Reduction of fluorescence to 50% should be determined in between 0 to 1 mM HCl; however

    this cannot be observed due to drastic changes in the fluorescence intensity. Complete loss of

    fluorescence was not observed at all.

    -2000

    0

    2000

    4000

    6000

    8000

    10000

    12000

    0 10 20 30 40 50

    FluorescenceIntensity

    (RFU)

    HCl concentration (mM)

    HCl added

    Recovery (neutralizedby NaOH)^

    Control (pH 7)*

    HCl pH

    1 mM = 3.0

    2 mM = 2.70

    5 mM = 2.30

    10 mM = 2.0

    20 mM = 1.70

    40 mM = 1.40

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    9/24

    Figure 6. Effect of varying NaOH concentration to the fluorescence of GFP sample (0.2 uM). The

    same fluorometer was used and the wavelength range is 525 nm to 570 nm. Three readings at each

    NaOH concentration were obtained and the mean value was used to construct the graph. (See Appendix

    B for the actual values).

    ^After reaching the 40mM NaOH concentration, HCl was added to successively lower the acid

    concentration again to 20, 10, 5, 2, 1 and 0 mM.

    *GFP was suspended in sterile distilled water as well as the test samples.

    Similar to the observation under acidic conditions, the increasing base concentration greatly

    decreased the exhibited fluorescence intensity of the sample as shown in Figure 6. However, the basic

    condition, analogous to the control, gradually decreased the fluorescence intensity while the acidic

    condition, as mentioned, immediately decreased the samples fluorescence intensity at 1 mM HCl (pH 3).Yet again, neutralization of the sample did not completely recover its fluorescence intensity. The sample

    initially exhibited fluorescence intensity value of 11003 RFU and then exhibited 1195 RFU fluorescence

    intensity upon complete neutralization. Also, original fluorescence was not attained at the other

    concentrations unlike in the case of the acidic environment. Fluorescence reduction to 50% was

    observed at 2 mM NaOH (pH 11.30). Complete loss of fluorescence was again not observed.

    Table 3 summarizes the behaviour of the samples fluorescence at different conditions. Among

    the three parameters, the highest percent recovery was obtained in the thermostability test specifically

    after reverting the temperature from 35C to room temperature (23 C) while only 1.042% was recovered

    at RT after exposure to the highest temperature used, 95C. The lowest percent recovery was obtained

    when 40 mM HCl was neutralized with the same concentration of NaOH. The fluorescence recovery from

    basic conditions was higher than from acidic pH.

    0

    2000

    4000

    6000

    8000

    10000

    12000

    0 10 20 30 40 50

    FluorescenceIntensity(RFU)

    NaOH concentration (mM)

    NaOH added

    NaOH neutralized byHCl^

    Control (pH 7)*

    NaOH pH

    1 mM = 11.0

    2 mM = 11.30

    5 mM = 11.70

    10 mM = 12.0

    20 mM = 12.30

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    10/24

    Table 3. Loss and Recovery of GFP (0.2 uM) fluorescence under different environments

    ConditionFluorescence Intensity

    Percent RecoveryA

    50% lost Initial (RFU) Recovered (RFU)

    Temperature at ~55C

    71.48% for 35C

    (highest)

    1.042% for 95C

    (lowest)

    Acid Cannot be determined 11078 2.037 0.01839 %

    Base at ~2 mM (pH 11.30) 11003 1195 10.86 %

    AIn calculating the percent recovery for temperature, the fluorescence intensity reading at RT after

    exposure to heat treatment was compared with the initial intensity obtained at RT which is 11170 RFU.

    IV. Discussion

    Spectroscopy is the study of spectra. A spectrum (pl. spectra) is defined in physics as a

    distribution of a characteristic of a physical system or a phenomenon. In the field of physical

    biochemistry, various types of spectroscopic techniques are being used to study biological molecules,such as absorption, emission, mass, and nuclear magnetic resonance spectroscopy (van Holde et al,

    1998). In this experiment, the concentration of the protein sample was determined through absorption

    spectroscopy and the effect of different stress factors, specifically extreme heat and pH, were determined

    through fluorescence spectroscopy.

    Absorption Spectroscopy

    Absorption spectroscopy deals with the transfer of electrons from the ground to excited

    state. This transfer, which allows the absorption of energy at a particular wavelength, occurs due to the

    presence of different molecular orbitals (See Figure 7). Electrons transfer from the ground to excited

    state when the molecules of a sample are exposed to a light bearing an energy that allows an electronic

    transition to occur. When this happens, some of the energy is absorbed. Electronic transitions occur

    from the highest occupied molecular orbital (HOMO) to the lowest unoccupied molecular orbital (LUMO),

    and of the six possible transitions shown in Figure 7, only the leftmost two transitions are achievable

    within 200-800 nm wavelength range (Laqua, 1988).

    Figure 7. Different molecular orbitals and the transitions electrons can undergo. The transfer of

    electrons from HOMO to LUMO due to excitation at a particular wavelength causes the absorption of

    energy. In the figure, only the two leftmost transitions can be achieved in excitations using lights with

    wavelength from 200-800 nm.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    11/24

    Absorption spectroscopy is done through an absorbance spectrophotometer, an instrument that

    measures the amount of light transmitted through the sample. The sample to be analyzed is placed in a

    sample holder, called a cuvette, and then positioned in the instrument. A monochromatic light is then

    allowed to pass through the solution and the amount of light that passed through the sample

    (transmittance) or absorbed (absorbance) is measured by a light meter. In this type of spectroscopy, the

    absorbance data is the one that is obtained (Hollas, 2004). The wavelength at which the absorption

    occurs, as well as the degree of absorption, is recorded by an optical spectrometer. The spectrum

    generated from the data is presented as a graph of absorbance (A) versus wavelength and is called the

    absorbance spectrum. The absorbance values that can be obtained are between the range of 0 to 2, with

    the zero value implying that no absorption occurred and a value of 2 implies that 99% absorption

    occurred (Laqua, 1988). Since the other compounds in the sample, such as the solvent itself, can

    interfere with the data by absorbing at the same wavelength as the sample to be analyzed, the

    absorbance of the test solution is compared to a blank solution. The blank solution contains everything

    that can be found on the test solution except the sample to be analyzed and the absorbance of the blank

    solution is set at zero to correct the errors from the possible interference. Using the absorbance spectrum

    obtained, the optimal wavelength, which is the wavelength that is most absorbed by the sample, can then

    be determined. It is at the optimal wavelength that subsequent absorption spectroscopy of samples is

    measured (Hollas, 2004).In order to analyze data from an absorption spectroscopy, it is important that the relationships

    between the components of the technique must first be established. The relationship between the degree

    of absorption of light (UV or visible light) and the properties of the material in a sample solution through

    which light is passed through can be derived from Beer-Lambert Law, also known as Beers Law. The

    equation for the relationship between transmission, absorbance coefficient and the path length is given by

    (2)

    However, since the absorbance coefficient of the sample can be obtained from the product of

    molar absorptivity or extinction coefficient and the concentration of the sample, Equation 2 can be

    rewritten as

    (3)

    Since the absorption is related to transmissivity by the following equation,

    ( ) (4)a linear relationship of absorbance with respect to the concentration of the sample can then be observed

    and this is given by the equation:

    (5)Since the absorbance of the sample can be measured and the concentration of the standard

    protein and path length is known, a calibration curve can be obtained and the value of absorbance

    coefficient can be calculated. Consequently, the unknown concentration of the test protein can be

    calculated by substituting the values of the path length of the sample container and the absorbance

    reading of the sample in conjunction with the calculated absorbance coefficient.

    Before conducting the Bradford reaction on the protein sample, the sample was first diluted with

    20 mM Tris-Cl pH 8.0 since absorption spectroscopy only applies to dilute solutions as can be seen later

    on the limitations of Beers Law. The Tris-Cl solution was set at pH 8.0 since the physiological pH of the

    green fluorescent protein is at this level. To quantify the protein concentration through a Bradford

    reaction, a protein standard set of bovine serum albumin (BSA) in concentrations of 0, 0.2, 0.4, 0.6, 0.8,

    1.0, and 1.2 mg/mL in ddH2O was prepared in order for a calibration curve to be constructed which will be

    the basis for calculating the concentration of the protein in the test protein sample. BSA was used

    because of its ability to increase intensity signal in total protein assays, and its stability and lack of

    interaction with components of biochemical reactions. Aside from that, BSA is readily available to

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    12/24

    researchers because of its low cost and can be easily obtained at a high concentration by purifying

    bovine blood, a product of the cattle industry. Technically, if the protein concentrations are the same, the

    same absorbance values should be obtained for proteins provided that the same dilution buffer and same

    stock solution of Bradford reagent was used for the assay. However, due to differences in concentration

    of the protein, different absorbance values are obtained, but an equation showing the definite relationship

    between concentration of the protein and the corresponding absorbance value can be obtained since the

    concentration of the BSA is known and in turn, this can be used to determine the unknown concentration

    of the protein sample (Doumas, 1975). A test protein sample set consisting of undiluted, 1:10 dilution, and

    1:100 dilution with the Tris-Cl pH 8.0 was also prepared since, as have been mentioned, absorption

    spectroscopy only applies to dilute solutions. The positive control used was asFP504, another fluorescent

    protein which has an excitation maxima at 471 nm and 494 nm and emission maximum at 504 nm

    (Concepcion, 2008). The positive control was necessary to confirm that the instrument used in reading

    the absorbance of the samples is indeed functional. Following preparation of the protein standard set and

    test protein sample, a Bradford reaction was conducted wherein 10 L of the protein sample and 500 L

    aliquots of the Bradford reagent were mixed. It is necessary that the Bradford reagent is always mixed

    with the protein sample at a larger amount since it is not known how much Bradford reagent will be

    needed to react with the protein sample due to the fact that its concentration is unknown. If the Bradford

    reagent is small in amount that it cannot bind all of the proteins present in the sample, then a less intensechange in color from the Bradford reaction will be observed and the concentration of the protein that will

    be calculated will be smaller. On the other hand, if there is an excess of Bradford reagent, then the

    Bradford reagent will be able to bind all of the proteins present in the sample and since it is the interaction

    between the Bradford reagent and the protein in the sample that is detected by the spectrophotometer,

    the absorbance obtained will still be accurate. The mixture of the Bradford reagent and protein test

    sample was placed in a plastic cuvette for absorbance value reading since binding of the protein-dye

    complex has been observed only with quartz cuvettes and may be eliminated by using either glass or

    plastic cuvettes (Bradford, 1976).

    In the experiment, Bradford assay was performed to determine the concentration of GFP samples

    specifically by treating them with Bradford reagent and measuring their absorbance using a

    spectrophotometer. This protein determination method is based on the shift in the absorbance maximum

    of Coomassie Brilliant Blue G-250 in the reagent. Under acidic conditions, this dye exists in doubly

    protonated tan to brown cationic form with absorbance maximum at 470 nm. Upon introduction to protein

    samples, the dye binds with basic amino acids particularly to arginine through a combination of

    hydrophobic interactions and heteropolar bonding (electrostatic interactions) (Georgiou et al., 2008). Van

    der Waals forces and hydrophobic interactions may also be observed between the dye and the aromatic

    amino acid residues (Trp, Tyr and Phe) (Compton and Jones, 1985). This binding then converts the dye

    into its stable unprotonated blue form which absorbs maximally at 595 nm (Bradford, 1976).

    A study has shown that the molar extinction coefficient of a dye-protein complex is constant over

    a 10-fold concentration range (Spector, 1978). With only two unknown quantities left, Beers law can be

    used to calculate concentration in terms of absorbance, the value obtained from the spectrophotometer.

    The protein sample with the greatest concentration will exhibit the greatest absorption at 595 nm. This is

    because the Coomassie Brilliant Blue G-250 dye binds proportionally to the proteins. The Bradford assayis a reproducible, easy and fast method to use for protein quantification. The dye binding process is

    completed at approximately 2 minutes and the color can be maintained for an hour; thus, critical timing for

    assay is not required. It is not prone to significant interference by cations (e.g. sodium or potassium) and

    carbohydrates such as sucrose. Interfering color is only caused by significant amounts of detergents that

    may remain on the glassware as well as strong alkaline buffering reagents (Bradford, 1976). Furthermore,

    the protein-dye complex has a high extinction coefficient which corresponds to high-intensity absorption

    of light (Singh et al., 2001). Despite these advantages, the Bradford assay also has limitations to be

    considered. It cannot be used to quantify all proteins types since there are events wherein the protein

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    13/24

    sample assays may deviate chemically from Beers Law due to the proteins associa tion, dissociation and

    interaction with the Bradford reagent and, sometimes, the solvent of the sample solution. In addition,

    proteins that do not have Tyrosine and Tryptophan will not react properly with the Bradford reagent, and

    hence, will not achieve the optimal tan colored solution that is required for the assay since the Bradford

    reagent reacts strongly with these amino acids. Another limitation of this assay is the 2-10 mg protein

    concentration limit. The Bradford assay is not valid for protein concentrations below or above this limit

    because the plot that can be obtained from these data do not exhibit a linear graph anymore. This

    therefore means that the Beers law cannot be applied anymore and possible errors, such as

    underestimation for samples below 2 mg and overestimation for samples above 10 mg, can occur. As for

    experiments where sample recovery is important, the Bradford assay is not recommended to be used

    because the Bradford reagent has the capacity to denature the proteins due to its acidic property. In

    relation this, the acidity of the Bradford reagent can also cause the proteins to aggregate, which can lead

    to the sample being less quantified due to the unavailability of the residues that are supposed to react

    with the reagent (Ninfa and Ballou, 1998).

    At all cases, the absorbance readings of the control are greater than those of the test sample.

    Expectedly, the mean stock concentration of the positive control which is 10.27 mg/mL is higher than that

    of the test sample, which is only 1.658 mg/mL. This can be expected since the concentrations of the two

    were not equalized.When the stock concentrations were calculated from both diluted and undiluted GFP and

    asFP504 solutions, extreme values were obtained. For the positive control, the undiluted solution yielded

    a value (2.887 mg/mL) that is way lower than the stock concentrations derived from 1:10 and 1:100

    dilutions which are 10.87 mg/mL and 9.677 mg/mL. It was not included in the calculation of the mean

    concentration so as not to incur a large error. This anomalous value may be attributed to the relatively

    high corrected absorbance of the undiluted solution (1.471) which is probably out of the range wherein

    absorbance is linearly related to concentration. In fact, it was the only corrected absorbance greater than

    1. The range depends on the Bradford reagent and equipment but often, it does not include readings

    greater than 1. The corrected absorbance values for the diluted solutions of asFP504 were only 0.545

    (1:10) and 0.035 (1:100). In another case, the GFP test sample diluted a hundred fold (1:100) yielded a

    negative concentration value which is -6.128 because the corrected absorbance is negative (-0.046).

    Theoretically, obtaining a negative absorbance value is not feasible because it will lead to a negative

    concentration which is not possible. However, there are instances that the absorbance readings will be

    negative, simply because of the errors in the blank sample or due to very low concentration of the analyte

    in the solution that it is beyond the detection limit of the Bradford reagent. The latter most likely explains

    the negative value because a 100-fold dilution of the GFP test sample with a mean concentration of 1.658

    mg/mL would lower the concentration to approximately 0.02 mg/mL which is way lower than the minimum

    value that can be detected by the reagent (2 mg) given that 1 mL of the solution is in the cuvette

    (equivalent to 0.02 mg protein in cuvette).

    Proteins absorb a particular wavelength differently and so they differ in the molar extinction or

    absorption coefficient. This value is also affected by factors such as solvent, solution composition, and

    temperature and so environmental parameters were maintained all throughout the experiment to keep

    constant and make the relationship of absorbance and concentration linear (Skoog et al., 2004). TheasFP504 and GFP have similar absorption capabilities and this is evident in their comparable calculated

    molar extinction coefficients, 0.4482

    and 0.4604

    , respectively. This similarity in the degree of

    absorbing light as well as in the degree of binding of the reagent to both proteins makes asFP504 a valid

    positive control. A commercially available GFP is not necessary. Percent deviation was not calculated

    because the molar extinction coefficients are expected to be slightly different since they come from

    different proteins.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    14/24

    To avoid anomalous data such as negative values for absorbance and overestimated or

    underestimated values for concentration, the protein solution subjected to Bradford assay should be

    diluted appropriately depending on the detection limit of the equipment and Bradford reagent used.

    Beers Law only describes the absorbance of dilute solutions. At highly concentrated solutions, the

    absorptivity is not constant and independent of concentration. It is already dependent on a quantity called

    refractive index which may considerably vary at high concentrations. At high concentrations, the plot of

    absorbance against concentration fails to follow a straight line because it bends toward the concentration

    axis (x-axis) (Willard et al., 1981).

    Fluorescence Spectroscopy

    Fluorescence is an excitation and emission process in that molecules absorb energy in the form

    of electromagnetic radiation of a certain wavelength and frequency. This electromagnetic radiation then

    allows an electron to go from its ground state to excited state. The emission of a different wavelength and

    frequency of electromagnetic radiation as the electrons return to its ground state is then detected and

    processed as the fluorescence intensity. The wavelength of the emitted electromagnetic radiation is

    different because energy is lost during the transitions, and as can be deduced from Plancks equation and

    the relationship between the frequency and wavelength, the wavelength increases as the energy

    decreases. Various factors such as the protein structure and rigidity, solution pH, temperature andconcentration of the proteins can affect the degree of fluorescence exhibited by the protein (Lakowicz,

    2006).

    In the experiment, 1 mL of sterile ddH2O placed in a plastic cuvette served as the blank. A blank

    was necessary to eliminate the contributions of other interfering substances present in the sample that

    may also fluoresce. The absorbance was measured at 525-570 nm because the fluorescent protein

    exhibits maximum emission at that wavelength range (Yang, et al., 1996).

    The increasing trend of fluorescence intensity as the protein concentration is increased is

    expected because more protein molecules contribute to the signal detected. Usually, a linear correlation

    between fluorescent protein concentration and fluorescence intensity is observed; but it should be noted

    that the relationship between fluorescence and protein concentration is not really linear. In some cases, a

    linear relationship between fluorescence intensity and protein concentration is observed at a certain

    range of concentration specified for the protein used. More often than not, fluorescence is reduced at

    higher protein concentrations due to reabsorption of the fluorescence photon by neighboring molecules,

    limiting the linearity of fluorescence intensity against protein concentration to only that of dilute protein

    sample solutions. In addition, at higher protein concentrations, self-quenching, inner-filtering and other

    artifacts may occur, further reducing the intensity of fluorescence detected (Brandt, 2010). With these, an

    intermediate concentration value was chosen to further monitor the fluorescence of GFP at different

    environmental parameters (heat and pH). In this case, 0.02 uM GFP, which gave a fluorescence of 11657

    RFU, was used. The highest intensity was 95313 (0.1 uM GFP) RFU while the lowest is 4561 RFU (0.01

    uM GFP).

    A. Thermostability Test

    As for the thermostability test, it was observed that the fluorescence intensity of GFP decreasedas the incubation temperature increased. Although the -can structure of GFP is stable enough for it to

    maintain fluorescence in the presence of detergents such as SDS and the fluorophore is highly protected

    within the central helix near the geometric center, prolonged exposure to high temperature can

    completely denature it and quench its fluorescence (Yang et al., 1996). When heat is added to the

    system, proteins undergo denaturation, accounting for the observations made for the test sample. The

    kinetic energy of the molecules within the protein increases as heat increases, and the increase in kinetic

    energy leads to higher vibrations of the molecules, disrupting hydrogen bonds and non-polar hydrophobic

    interactions within the protein. Hydrogen bonds and hydrophobic interactions are responsible for holding

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    15/24

    the protein together and disruption of these bonds may lead to loss of the secondary structure of the

    protein which may lead to the eventual loss of biological functions of the protein such as fluorescence

    (Lewis, 1926).

    Complete heat denaturation of GFP is attained by exposing the protein at temperatures as high

    as 90C in denaturing agents such as guanidium chloride (Yang et al., 1996). In general, the temperature

    range of the treatment was not high enough to completely denature the protein and so complete loss of

    fluorescence did not happen; only a decrease in fluorescence was observed which is expected at

    temperatures greater than 30C. Fluorescence was not lost completely at 95C, which is possibly due to

    the absence of other denaturing agents. However, a 150-fold decrease to 73.54 RFU (from 11170 RFU)

    is a significant decrease. After exposure to higher temperature, the protein was allowed to recover for 2

    minutes at room temperature (~23C). Failure to attain the initial intensity value proved that heat

    denaturation is indeed irreversible (Yang et al., 1996).

    The control sample, on the other hand, only exhibited a gradual decrease in fluorescence.

    Theoretically, this should not be observed since control was maintained at room temperature. Most likely,

    the cause is fluorescence quenching, which refers to any process that decreases the fluorescence

    intensity of the sample. A common quencher is the oxygen molecule which is ubiquitously present in the

    environment. Upon contact with the quencher, the fluorophore reverts to its ground state without emitting

    a photon. Fluorescence intensity readings were acquired a couple of times and chances of oxygen havingcontact with the fluorophore increased with time (Lakowicz, 2006). Quenching might have also

    contributed to decrease in fluorescence of the test samples but only causing minor effects.

    The melting temperature (Tm) was determined to be 55C. According to Sauer et al. (2011), GFP

    has a melting temperature above 65C. Despite this, the value from this experiment cannot be dismissed

    because a protein may have several melting temperatures depending on the experimental conditions and

    the analytical technique used in determining it. In addition, the value is within the usual melting

    temperature of proteins which is 40C to 80C. Proteins having Tm higher than the range are already

    thermophiles (Rajni and Mattiasson, 2003).

    The range of decrease of the controls fluorescence intensity (11000 RFU to 8000 RFU) is

    narrower than that of the test sample (11,000 RFU to 100 RFU). This shows that heat can drastically alter

    the fluorescence of the protein.

    B. pH-stability Test

    Aside from protein concentration and temperature, pH also affects fluorescence, particularly the

    spectral properties of the protein. It can alter the interactions between residues involved in fluorescence,

    in particular the salt bridges, by changing the ionization state of those amino acids. The protein maintains

    fluorescence over a wide pH range, specifically from 6 to 10 for Campbell and Choy (2001) and 8 to 11

    for Haupts et al. (1998). However, it decreases at a pH lower than 6. This explains the decreasing trend

    of fluorescence when it was subjected to acidic conditions. The pH range used was 1.40 to 3.0 and

    relative to the range wherein fluorescence can be stably observed, the least acidic pH (1 mM HCl, pH 3.0)

    is already low enough to cause the abrupt decrease in fluorescence intensity observed. In fact, at this pH

    more than 50% of fluorescence was lost. From 11007 RFU, fluorescence lowered to 208 RFU when the

    HCl concentration became 1 mM. Additionally, it has been observed that GFP fluorescence is highlysensitive to proton concentration such that it decreases to zero below pH 4. Negative intensity values

    were even obtained in the experiment particularly for 10, 20 and 40 mM HCl (pH 2, 1.70 and 1.40,

    respectively) proving that fluorescence had become very lowbeyond the detection limit of the

    instrument. At low pH, the chromophore is observed to decrease absorption at 488 nm excitation

    wavelength which is near the value that was used in this experiment that is 460 nm. Decrease in

    absorption would result to a decrease in intensity of the fluorescence emitted. The decrease in

    absorption is due to the protonation of the hydroxyl groups of a member of the active site, Tyr66

    resulting

    to a non-fluorescent species. This eradicates fluorescence quenching as the reason.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    16/24

    According to Haupts et al. (1998), fluorescence of the protein also decreases at high pH values.

    The basic pH values used as treatment were all greater than pH 11 and expectedly the fluorescence

    intensity was observed to decrease. However, it should be noted that the decrease in fluorescence

    intensity of the protein with increasing basicity was far more gradual than the decrease in acidic

    conditions. From 11002 RFU, the lowest intensity attained was 3030 RFU at the highest NaOH

    concentration which is 40 mM or pH 12.60. Fifty percent fluorescence loss needed a higher concentration

    of NaOH, specifically at 2 mM NaOH or at pH 11.30. Also, complete loss of fluorescence was not

    observed. Most likely, this subtle effect of basic pH on fluorescence is because of the chromophore being

    protonated in a slightly different mechanism than what happens in acidic conditions. According to Haupts

    et al. (1998), the protonated species is a result of the pH-independent internal protonation process

    meaning that within the fluorescent protein, a proton is fluctuating or transferring between the hydroxyl

    groups of Tyr66

    and an internal proton binding site. The proton is not coming from the bulk solution. At low

    pH, this internal protonation may occur at the same time, another proton from the bulk solution can again

    cause protonation owing to the drastic decrease in fluorescence observed and the higher percent

    recovery at basic condtions (10.86%) than at acidic environment (0.01839%).

    At both pH fractions, recovery of the initial fluorescence intensity was not observed. This is

    because treatment at extreme pH values causes mostly irreversible protonation of the active site and

    alterations in the secondary structure of the protein.Other properties of the buffer that may be taken into account are salt concentrations, presence of

    detergents and bovine serum albumin (BSA). High salt concentrations may induce aggregation of protein

    samples. On the other hand, presence of detergents and BSA can interfere with fluorescence signal

    intensity and other interactions within the protein. If some other substance is needed to be present in the

    buffer, it is important that this substance is measured to take into account any background fluorescence

    contributed by the added substance (Ugwu & Apte, 2004).

    The highest percent recovery (71.48%) was obtained during the thermostability test; when the

    protein was re-exposed to RT after heat treatment at 37C. Additionally, percent recovery due to removal

    of 45C-, 55 C-, 65C- and 75C-heat treatments were all higher than the value due to removal of acid

    and base (See Appendix Supplementary Table 7 for the actual values). This further proves the

    robustness of the protein at temperatures lower than 80C. The recovery after exposure to 85C was

    11.42% which is comparable with the basic environment, 10.86%. Recovery due to temperature reset

    from 95C to room temperature and removal of 40 mM HCl yielded the low percent recovery values,

    1.041 % and 0.01839% respectively. This indicates that exposure of protein to high temperature and

    extremely acidic pH (~1.40) causes nearly irreversible denaturation or loss in fluorescence.

    The degree of fluorescence recovery may vary among stress factors but in all cases the

    percentage never reached 100%. Given that the fluorescence is primarily dependent on the primary

    sequence of the protein rather than to its three-dimensional structure, partial recovery means that the

    stress factors were able to alter the primary sequence of most of the protein molecules particularly in the

    active sites. The alterations in the amino acid sequence hinder the protein from reverting to its native

    conformation which means that the denaturation is irreversible. This claim is based on Anfinsens

    theorem which states that in a given environment, the proteins native conformation is determined by the

    primary structure, its amino acid sequence, at least for small globular proteins (Anfinsen, 1973).Irreversible heat denaturation occurs often at extreme values when protein molecules become

    aggregated. A heat treatment that only causes unfolding of the native protein by weakening interactions

    is reversible but once the protein molecules become fully denatured and their hydrophobic groups are

    exposed, they tend to aggregate to hide those groups from the polar environment. To revert to the native

    state, the aggregated protein should overcome an energy barrier. In the case of extreme pH, degradation

    of acid-labile residues can cause cleavage of peptide bonds as well as alkaline hydrolysis at highly basic

    environment.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    17/24

    Quenching was observed in the control which meant that it could have contributed also to the

    decrease in fluorescence of the test samples. To eliminate this error and to make sure that the stress

    factor alone contributes to the recorded fluorescence intensity, exposure to common quenchers such as

    oxygen may be avoided by covering the sample before and after taking the readings. Additionally, since

    the extract was determined to be not pure, other quenchers might be present in the solution. Using a pure

    sample will make the data more reliable.

    V. Conclusion

    The concentration of the GFP extract was successfully determined to be 1.658 mg/mL through

    the Bradford assay. Results of fluorescence spectroscopy showed that an increase in temperature

    starting from 35 C to 95C decreases the fluorescence intensity of the protein. The same effect was

    observed as the environment became more highly acidic and more highly basic in particular at values

    outside the stable range of the protein which is around pH 8 to 11. Removal of the stress factors resulted

    to minimal recovery of fluorescence suggesting that extreme heat and pH conditions cause nearly

    irreversible denaturation with the acidic environment having the most profound effect.

    VI. References

    Anfinsen, C. B.. "Principles That Govern The Folding Of Protein Chains." Science181.4096 (1973): 223-

    230. Print.

    Bradford, Marion M. "A Rapid and Sensitive Method for the Quantitation of Microgram Quantities of

    Protein Utilizing the Principle of Protein-Dye Binding."Analyical Biochemistry72 (1976): 248-254.

    Web. 13 Oct. 2013.

    Brandt, Mark. "Fluorescence Spectroscopy." Rose-Hulman Institute of Technology. N.p., 2010. Web. 9

    Sept. 2013.

    Campbell, Tessa N., and Choy, Francis Y.M. "The Effect of pH on Green Fluorescent Protein: a Brief

    Review." Molecular Biology Today 2.1 (2001): 1-4. Web. 13 Oct. 2013.

    http://www.horizonpress.com/backlist/mbt/v/v2/01.pdf.

    Compton, S. J., and Jones, C. G. "Mechanism of dye response and interference in the Bradford protein

    assay." Analytical Biochemistry151.2 (1985): 369-74. US National Library of Medicine National

    Institutes of Health. Web. 13 Oct. 2013.http://www.ncbi.nlm.nih.gov/pubmed/4096375

    Concepcion, Carla. "Cloning of the asFP504 gene in a mammalian expression vector and heterologous

    expression of the monemerric cyan fluorescent protein (mCFP) variant in HEK293 and 4T1 cell

    lines." Unpublished thesis (2008).

    Dinh, Tuan. Handbook of spectroscopy. Weinheim: Wiley-VCH, 2003. Print.

    Doumas, B T. "Standards for total serum protein assays--a collaborative study." Clinical Chemistry21.8

    (1975): 1159-66. Print.

    Georgiou, C.D., Grintzalis, K., Zervoudakis, G. and Papapostolou, I. "Mechanism of Coomassie brilliant

    blue G-250 binding to proteins: a hydrophobic assay for nanogram quantities of proteins."

    http://hoffman.cm.utexas.edu/courses/bradford_assay.pdfhttp://www.rose-hulman.edu/~brandt/Fluorescence/Fluorescence_Introduction.pdfhttp://www.horizonpress.com/backlist/mbt/v/v2/01.pdf.http://www.ncbi.nlm.nih.gov/pubmed/4096375http://www.ncbi.nlm.nih.gov/pubmed/4096375http://www.horizonpress.com/backlist/mbt/v/v2/01.pdf.http://www.rose-hulman.edu/~brandt/Fluorescence/Fluorescence_Introduction.pdfhttp://hoffman.cm.utexas.edu/courses/bradford_assay.pdf
  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    18/24

    Analytical and Bioanalytical Chemistry 391.1 (2008): 391-403. US National Library of Medicine

    National Institutes of Health. Web. 13 Oct. 2013.http://www.ncbi.nlm.nih.gov/pubmed/18327568

    Haupts, Ulrich, Maiti, Sudipta, Schwille, Petra and Webb, Watt W.. "Dynamics of fluorescence fluctuations

    in green fluorescent protein observed by fluorescence correlation spectroscopy." Proceedings of

    the National Academy of Sciences of the United States of America 95.23 (1998): 13573-8. Print.

    Hollas, J. Michael. Modern spectroscopy. 4th ed. Chichester: Wiley, 2004. Print.

    Kaul, Rajni, and Mattiasson, Bo. Isolation and purification of proteins. New York: Marcel Dekker, 2003.

    Print.

    Lakowicz, Joseph R. Principles of fluorescence spectroscopy. 3rd ed. New York: Springer, 2006. Web. 13

    Oct 2013. < http://link.springer.com/chapter/10.1007%2F978-0-387-46312-4_8#page-2>

    Laqua, K. "Molecular Absorption Spectroscopy, Ultraviolet and Visible (UV/Vis)." Pure and Applied

    Chemistry 60.9 (1088): n. pag. Internation Union of Pure and Applied Chemistry - IUPAC. Web.

    14 Oct. 2013.

    Lewis, P. S. The Kinetics of Protein Denaturation: Part III.The Influence of Neutral Salts on the Velocity

    of the Heat Denaturation of Oxyhaemoglobin. Biochem J. 1926; 20(5):984992. Print.

    Ninfa, Alexander J., and Ballou, David P. Fundamental laboratory approaches for biochemistry and

    biotechnology. Bethesda, Md.: Fitzgerald Science Press, 1998. Print.

    Sauer, Markus, Hofkens, Johan and Enderlein, J. Handbook of fluorescence spectroscopy and imaging

    from single molecules to ensembles. Weinheim: Wiley-VCH, 2011. Print.

    Spector, T. "Refinement of the Coomassie blue method of protein quantitation. A simple and linear

    spectrophotometric assay for less than or equal to 0.5 to 50 micrograms of protein." Analytical

    Biochemistry86 (1978): 142-146. Print.

    Singh, K., Sandhu, G. K., Lark, B. S. and Sud, S. P. "Molar extinction coefficients of some carbohydrates

    in aqueous solutions." Journal of Physics 58.3 (2002): 521-528. Indian Academy of Sciences.

    Web. 13 Oct. 2014.

    Skoog, D.A., West, D.M., James Holler, F. and Crouch, S.R. Fundamentals of Analytical Chemistry. 8th

    edition. Brooks/Cole. 2004. p. 716, 723, 729.

    Ugwu, Sydney O., and Apte, Shireesh P. The Effect of Buffers on Protein Conformational Stability. USA:

    Pharmaceutical Technology, 2004. Print.

    Weijers, Mireille, Barneveld, Peter A., Cohen Stuart, Martien A., and Visschers, Ronald W. "Heat-induced

    denaturation and aggregation of ovalbumin at neutral pH described by irreversible first-order

    kinetics." Protein Science 12 (2003): 2693-2703. US National Library of Medicine National

    Institutes of Health. Web. 13 Oct. 2013.

    http://www.ncbi.nlm.nih.gov/pubmed/18327568http://www.ias.ac.in/pramana/march2002/P5435.PDFhttp://www.ias.ac.in/pramana/march2002/P5435.PDFhttp://www.ncbi.nlm.nih.gov/pubmed/18327568
  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    19/24

    Willard, H.H., Merritt, L.L. Jr., Dean, J.A. and Settle, F.A. Instrumental Methods of Analysis. 6th edition.

    Litton Educational Publishing, Inc. 1981. p.19, 66-69, 72-73, 75.

    Yang, Fan, Moss, Larry G. and Phillips, George N. "The Molecular Structure Of Green Fluorescent

    Protein." Nature Biotechnology14.10 (1996): 1246-1251. Print.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    20/24

    VII. Appendix

    A. Absorption Spectroscopy

    1. Raw Data

    Supplementary Table 1. Absorbance readings of BSA standard and GFP positive

    control and test samples all added with Bradford reagent

    Absorbance

    BSA concentration

    mg/mLReading 1 Reading 2 Reading 3 Mean

    0 0.51 0.511 0.512 0.511

    0.2 0.558 0.558 0.559 0.558333

    0.4 0.709 0.711 0.712 0.710667

    0.6 0.826 0.827 0.827 0.826667

    0.8 0.922 0.923 0.924 0.923

    1 0.997 0.997 0.997 0.997

    1.2 1.105 1.109 1.111 1.108333

    GFP Positive control

    Undiluted 1.982 1.982 1.982 1.982

    10^-1 1.055 1.056 1.056 1.055667

    10^-2 0.54 0.547 0.551 0.546

    GFP Test sample

    Undiluted 1.443 1.444 1.447 1.444667

    10^-1 0.572 0.572 0.572 0.572

    10^-2 0.463 0.465 0.466 0.464667

    2. Sample Calculations

    Determination of Protein Concentration of GFP samples

    wherein c is concentration (1x) in mg/mL, a is the corrected absorbance at 595 nm and df is the dilution

    factor.

    GFP test sample (1:10 dilution)

    Determination of mean molar extinction coefficient (e) for GFP

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    21/24

    wherein eis the molar extinction coefficient in mL*mg-1

    cm-1

    , c is the concentration (1x) in mg/mL, a is

    the corrected absorbance at 595 nm and lis the path length which is in this case 1 cm.

    Positive control (asFP504)

    GFP Test Sample

    B. Fluorescence Spectroscopy

    1. Raw Data

    Intensity Test

    Supplementary Table 2. Fluorescence intensity readings at varying concentrations of the protein

    Protein Concentration (uM) Reading 1 Reading 2 Reading 3 Mean CorrectedA

    0 144.54 144.57 144.89 144.6666667 0

    0.01 4562.87 4561.92 4558 4560.93 4416.263

    0.05 37336.83 37294.26 37289.56 37306.88333 11656.69

    0.1 95383.35 95475.84 95514.23 95457.80667 37162.22

    0.02 11844.37 11793.78 11765.91 11801.35333 95313.14A

    Corrected values were obtained by subtracting the mean absorbance (144.67) of 0 mg/mL protein to the

    mean values of the different concentrations.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    22/24

    Thermostability Test

    Supplementary Table 3. Fluorescence intensity readings of control at varying temperatures

    Temperature

    (OC)

    A

    Control

    Reading

    1

    Reading

    2

    Reading

    3Mean

    Corrected

    Initial 10990.12 10996.55 11005.14 10997.27 10852.6

    35 10693.29 10675.37 10647.45 10672.03667 10527.37

    RT 10365.38 10341.64 10313.8 10340.27333 10195.61

    45 10060.99 10050.88 10045.03 10052.3 9907.633

    RT 9938.79 9951.25 9959.11 9949.716667 9805.05

    55 9675.74 9682.59 9683.28 9680.536667 9535.87

    RT 9618.36 9613.82 9614.19 9615.456667 9470.79

    65 9454.24 9452.51 9442.99 9449.913333 9305.247

    RT 9203.29 9194.41 9184.43 9194.043333 9049.377

    75 9198.63 9197.4 9200.08 9198.703333 9054.037

    RT 9157.66 9149.64 9149.95 9152.416667 9007.75

    85 8886.46 8876.53 8869.87 8877.62 8732.953

    RT 8611.8 8614.87 8616.5 8614.39 8469.723

    95 8661.83 8654.62 8633.15 8649.866667 8505.2

    RT 8418.48 8410.7 8405.22 8411.466667 8266.8A

    RT stands for room temperature and is approximately 23C.BCorrected values were obtained by subtracting the mean absorbance (144.67) of 0 mg/mL protein to the

    actual fluorescence intensity means.

    Supplementary Table 4. Fluorescence intensity readings of test sample at varying temperatures

    Temperature

    (OC)

    Test

    Reading

    1

    Reading

    2

    Reading

    3Means

    Corrected

    Initial 11343.26 11346.35 11253.34 11314.31667 11169.6535 9135.38 9120.95 9096.9 9117.743333 8973.077

    RT 8117.59 8129.65 8140.4 8129.213333 7984.547

    45 6631.49 6686.6 6730.55 6682.88 6538.213

    RT 6909.53 6908.49 6904.72 6907.58 6762.913

    55 5129.47 5256.22 5318.98 5234.89 5090.223

    RT 5635.38 5644.05 5661.25 5646.893333 5502.227

    65 3849.42 3915.11 3999.11 3921.213333 3776.547

    RT 4447.06 4476.73 4498.45 4474.08 4329.413

    75 2513.55 2664.44 2740.2 2639.396667 2494.73

    RT 3168.85 3186.91 3204.52 3186.76 3042.093

    85 1008.81 1097.07 1143.26 1083.046667 938.38RT 1415.63 1422.15 1424.1 1420.626667 1275.96

    95 213.56 223.71 217.35 218.2066667 73.54

    RT 259.25 261.54 262.31 261.0333333 116.3667A

    Corrected values were obtained by subtracting the mean absorbance (144.67) of 0 mg/mL protein to the

    actual fluorescence intensity means.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    23/24

    Resistance to Acidic Conditions

    Supplementary Table 5. Fluorescence intensity readings of test sample at varying acidic pH

    HCl

    Concentration

    (mM)

    Reading

    1

    Reading

    2

    Reading

    3Mean

    CorrectedB

    0 11241.26 11224.04 11201.6 11222.3 11077.63

    1 373.07 349.82 336.82 353.2366667 208.57

    2 225.62 224.47 223.76 224.6166667 79.95

    5 148 146.93 146 146.9766667 2.31

    10 142.33 142.25 142.2 142.26 -2.40667

    20 141.53 141.49 141.54 141.52 -3.14667

    40 141.28 141.28 141.26 141.2733333 -3.39333

    40 141.28 141.28 141.26 141.2733333 -3.39333

    20A

    142.56 142.5 142.39 142.4833333 -2.18333

    10

    146.02 145.81 145.87 145.9 1.233333

    5

    146.6 146.26 146.11 146.3233333 1.656667

    2

    146.03 145.74 145.65 145.8066667 1.14

    1A 145.82 145.62 145.55 145.6633333 0.996667

    0

    146.77 146.7 146.64 146.7033333 2.036667A

    Sample is neutralized with NaOH to attain desired acid concentration.BCorrected values were obtained by subtracting the mean absorbance (144.67) of 0 mg/mL protein to the

    actual fluorescence intensity means.

    Resistance to Basic Conditions

    Supplementary Table 6. Fluorescence intensity readings of test sample at varying basic pH

    NaOH Concentration

    (mM)Reading 1 Reading 2 Reading 3 Mean Corrected

    B

    0 11154.17 11149.5 11138.41 11147.36 11002.691 9612.89 9470.2 9093.94 9392.343333 9247.677

    2 5561.27 5149.61 4888.01 5199.63 5054.963

    5 4135.68 4116.89 4106.98 4119.85 3975.183

    10 3963.4 3962.4 3961.19 3962.33 3817.663

    20 3766.92 3768.76 3769.84 3768.506667 3623.84

    40 3193.14 3173.74 3158.83 3175.236667 3030.57

    40

    3193.14 3173.74 3158.83 3175.236667 3030.57

    20A

    2707.65 2702.87 2697.97 2702.83 2558.163

    10

    2225.17 2138.25 2085.01 2149.476667 2004.81

    5

    1507.55 1505.67 1501.48 1504.9 1360.233

    2

    A

    1389.4 1392.23 1394.7 1392.11 1247.4431

    A 1386.96 1384.98 1382.04 1384.66 1239.993

    0

    1339.47 1339.49 1339.46 1339.473333 1194.807A

    Sample is neutralized with HCl to attain desired base concentration.BCorrected values were obtained by subtracting the mean absorbance (144.67) of 0 mg/mL protein to the

    actual fluorescence intensity means.

  • 8/14/2019 Absorbance and Fluorescence Spectroscopies of Green Fluorescent Protein.docx

    24/24

    2. Percent Recovery

    hermostability test

    Supplementary Table 7. Percent recovery after removal of heat treatment

    Temperature

    (C)

    Recovered

    RFU

    %

    RecoveryA

    35 7984.547 71.4843

    45 6762.913 60.54723

    55 5502.227 49.26051

    65 4329.413 38.76051

    75 3042.093 27.23535

    85 1275.96 11.42346

    95 116.3667 1.041811A

    The initial fluorescence intensity used to calculate is 11169.65 RFU.

    Stability test in acidic conditions

    Stability test in basic conditions