ABG Equipment

download ABG Equipment

of 43

Transcript of ABG Equipment

  • 8/12/2019 ABG Equipment

    1/43

    Arterial Blood Gases

    - Sampling,Equipment, Calibration,and Quality Control

    RET 2414L

    Pulmonary Function Testing

    Module 6.0

  • 8/12/2019 ABG Equipment

    2/43

    Arterial Blood Gases

    Equipment

    Arterial Blood Gas (ABG) Kit

    Prepackaged and contains all necessaryequipment 3 5 cc syringe

    Pre-heparinized

    22ga x 2 needle

    Alcohol swap

    Gauze pad

    Biohazard bag

    Misc. items

  • 8/12/2019 ABG Equipment

    3/43

    ABG Specimen Collection/Handling

    Site Selection

  • 8/12/2019 ABG Equipment

    4/43

    ABG Specimen Collection/Handling

    Site Selection

    Radial Artery - 45insertion angle Requires modified Allens test for collateral circulation

    Brachial Artery - 60- 90insertion angle

    Femoral Artery - 90insertion angle

    Dorsalis Pedis Artery

    Site must be adequately compressed untilclotted Approximately 5 minutes Patients receiving anticoagulation therapy take longer

  • 8/12/2019 ABG Equipment

    5/43

    ABG Specimen Collection/Handling

    Hazards

    Hematoma

    Arterial laceration

    Hemorrhage

    Vasovagal reaction Sympathetic nervous system response to pain

    Loss of limb

  • 8/12/2019 ABG Equipment

    6/43

    ABG Specimen Collection/Handling

    Handling

    Blood gas specimen should collected

    anaerobically Expel air bubbles immediately

    In Vivo Values Air Contamination

    pH 7.40 7.45

    PCO2 40 30

    PO2 95 110

  • 8/12/2019 ABG Equipment

    7/43

    ABG Specimen Collection/Handling

    Handling

    Blood gas specimen must be adequately

    anticoagulated Sodium heparin

    Lithium heparin (electrolytes)

    Sample volume should be 1 2 ml

    Each laboratory has its own protocol

  • 8/12/2019 ABG Equipment

    8/43

    ABG Specimen Collection/Handling

    Handling

    Sample should be analyzed as soon as

    possible If iced sample can be stored

    Glass syringe 1 hour

    Plastic syringe 15 minutes

    Remember: Blood is living tissue that continues toconsume O2 and produce CO2

  • 8/12/2019 ABG Equipment

    9/43

    ABG Specimen Collection/Handling

    Handling

    Specimen should be adequately identified

    Patient name / ID number Date / Time

    Ordering physician

    Accession number

    Puncture site

    Oxygen adjunct and FiO2

    Ventilator settings (if applicable)

  • 8/12/2019 ABG Equipment

    10/43

    ABG Specimen Collection/Handling

    Handling

    Transport specimen to laboratory in a

    biohazard container

    Analyze specimen on an instrument thathas been recently calibrated

    Temperature correct specimen in analyzer Increase in patient temp: PO2, PCO2, pH

    Decrease in patient temp: PO2, PCO2, pH

  • 8/12/2019 ABG Equipment

    11/43

    Arterial Blood Gases

    Equipment

    Blood Gas Analyzer

    Electronic Circuitry Electrolyte Solution

    Electrodes

  • 8/12/2019 ABG Equipment

    12/43

    Arterial Blood Gases

    Equipment

    Electronic circuitry

    Takes electrical current changes produced in theelectrodes and provides a visual display

    Electrolyte Solution

    Helps to promote chemical reactions andelectrical current

  • 8/12/2019 ABG Equipment

    13/43

    Arterial Blood Gases

    Equipment

    Electrodes

    Utilized to measure values of ABG

    pH, PCO2, PO2

    All other blood gas values are calculated

  • 8/12/2019 ABG Equipment

    14/43

    Arterial Blood Gases

    Equipment

    pH Electrode

    Sanz Electrode

    Consists of two electrodes:

    sampling/measuring electrode

    reference electrode and electrolyte solution

  • 8/12/2019 ABG Equipment

    15/43

    Arterial Blood Gases

    The pH electrode is amicroelectrode, shown here withits plastic jacket. At the tip is asilver-silver chloridewire in a

    sealed-in buffer behind PH-sensitive quartz glass. Thereference electrode contains aplatinum wirein calomel pastethat rests in a 20% KCL solution.The blood sample is introduced in

    such a way that it contacts themeasuring electrode tip and theKCL. A voltmeter measure thepotential difference across thesample, which is proportional tothe pH

    Sanz Electrode(pH)

  • 8/12/2019 ABG Equipment

    16/43

    Arterial Blood Gases

    Equipment

    PCO2 Electrode

    Severinghaus Electrode

    May also be referred to as a modified Sanzelectrode

  • 8/12/2019 ABG Equipment

    17/43

    Arterial Blood Gases

    The PCO2 electrode is a modifiedpH electrode. The electrode hasa sealed-in buffer; an Ag-AgClreference band is the other half-cell. The entire electrode is

    encased in Lucite jacketfilledwith bicarbonate electrolyte. The

    jacket is capped with a Teflonmembranethat is permeable toCO2. A nylon meshcovers thepH-sensitive glass, acting as aspacerto maintain contact withthe electrolyte. CO2 diffusesthrough the Teflon membrane,combines with electrolyte, andalter the pH. The change in pH isdisplayed as partial pressure ofCO2.

    Severinghaus Electrode

    (PCO2)

  • 8/12/2019 ABG Equipment

    18/43

    Arterial Blood Gases

    Equipment

    PO2 Electrode

    Clark Electrode

    May also be referred to as a polarographicelectrode

    Periodic/routing cleaning of the tip with pumiceisrequired because polypropylene attracts protein

  • 8/12/2019 ABG Equipment

    19/43

    Arterial Blood Gases

    The PO2 electrode contains aplatinum cathodeand a silveranode. The electrode is polarizedby applying a slightly negativevoltage of approximately 630 mV.The tip is protected by apolypropylene membranethatallows O2 molecules to diffuse butprevents contamination of theplatinum wire. O2 migrates to thecathode and is reduced, picking upfree electrons that have come fromthe anode through a phosphate-

    potassium chloride electrolyte.Changes in the current flowingbetween the anode and cathoderesult from the amount of O2reduced in the electrolyte and areproportional to partial pressure ofO2.

    Clark Electrode (PO2)

  • 8/12/2019 ABG Equipment

    20/43

    Arterial Blood Gases

    Calibration Procedures

    To assure appropriate electronic function of

    the electrodes, calibration procedures areperformed

    Performed automatically every 30 minutes bythe ABG machine

    Performed on the pH, PCO2, PO2electrodes

    Specific procedure for each electrode

  • 8/12/2019 ABG Equipment

    21/43

    Arterial Blood Gases

    Calibration Procedures

    2-Point Calibration

    A low concentration and a highconcentration is used at both ends of thephysiological range to be measured

    Multiple-Point Calibration (3 or more points)

    Verifies whether the gas analyzer is linear ornot

  • 8/12/2019 ABG Equipment

    22/43

    Arterial Blood Gases

    Calibration Procedures

    pH Electrode

    Uses two specific buffers with approximatevalues of:

    6.840 buffer referred to as the zero point or low point buffer

    7.384 buffer high point or slope point buffer

  • 8/12/2019 ABG Equipment

    23/43

    Arterial Blood Gases

    Calibration Procedures

    pH Electrode

    Each buffer is injected into the sample chamber,one at a time

    The values of the buffer that is injected, shouldbe displayed on the ABG machine within aspecific SD (standard deviation)

  • 8/12/2019 ABG Equipment

    24/43

    Arterial Blood Gases

    Calibration Procedures

    pH Electrode

    Standard deviation for pH is + .005

    If value displayed is within the SD, machine iselectronically calibrated

    If value displayed is outside of the SD, machineneeds to be adjusted to assure electronic function

  • 8/12/2019 ABG Equipment

    25/43

    Arterial Blood Gases

    Calibration Procedures

    PO2& PCO2Electrode

    Uses two specific concentration of gases for eachelectrode with approximate concentrations of CO2and O2

    Uses two different tanks of gas to accomplish this

  • 8/12/2019 ABG Equipment

    26/43

    Arterial Blood Gases

    Calibration Procedures

    PO2 & PCO2 Electrode

    Tank One Low CO2 (5%) - balance High O2 (12% or 20%) Balance Nitrogen

    Tank Two High CO2 (10%) slope O2 (0%) Balance Nitrogen

  • 8/12/2019 ABG Equipment

    27/43

    Arterial Blood Gases

    Calibration Procedures

    PO2 & PCO2 Electrode

    Must convert tank concentration from % tomm Hg

    (PBPH2O) x tank concentration = mm Hg

    (760 47) x 0.12 = 85.65 mm Hg

  • 8/12/2019 ABG Equipment

    28/43

    Arterial Blood Gases

    Calibration Procedures

    PO2& PCO2Electrode

    The values calculated for the CO2and O2concentration should be displayed on the ABGanalyzer within a specific SD (standard deviation)

  • 8/12/2019 ABG Equipment

    29/43

    Arterial Blood Gases

    Calibration Procedures

    PO2 & PCO2 Electrode

    Standard deviation for PO2and CO2is + 0.5

    If values displayed are within the SD, machine is

    electronically calibrated

    If value displayed is outside of the SD, machineneeds to be adjusted to assure electronic function

  • 8/12/2019 ABG Equipment

    30/43

    Arterial Blood Gases

    Calibration Procedures

    Troubleshooting

    If the ABG machine will not calibrate, check:

    The buffers

    The mixed gases

    The electrodes membrane

    The electrode itself

  • 8/12/2019 ABG Equipment

    31/43

    Arterial Blood Gases

    Quality Control

    Calibration vs. Quality Control

    Calibration is when the equipment is adjustedor correctedto match the control standards

    Quality Control testing must be performed on a

    regular basis to determine the accuracyandprecisionof the equipment against a knownstandard

  • 8/12/2019 ABG Equipment

    32/43

    Arterial Blood Gases

    Quality Control

    Accuracy vs. Precision

    Accuracy refers to the mean (average) value ofseveral measurements

    Precision refers to how consistently the same

    measurement will produce the same results

  • 8/12/2019 ABG Equipment

    33/43

    Arterial Blood Gases

    Quality Control

    Must be run every shift

    Utilize a knownconcentration of gasesand buffers in a vial ofliquid

    Run three levels of QC

    Level 1 Acidosis

    Level 2 Normal

    Level 3 - Alkalosis

  • 8/12/2019 ABG Equipment

    34/43

    Arterial Blood Gases

    Quality Control

    KNOWN VALUE IN MUST EQUAL

    KNOW VALUE OUT!

  • 8/12/2019 ABG Equipment

    35/43

    Arterial Blood Gases

    Quality Control

    When QC is run it must be recorded

    and maintained onsite; in the ABGlaboratory

    Must be available for review by State

    agencies on demand

  • 8/12/2019 ABG Equipment

    36/43

    Arterial Blood Gases

    Quality Control Plotting

    In control

    Trend

    Random Error

    Out of Control

  • 8/12/2019 ABG Equipment

    37/43

    Arterial Blood Gases

    Quality Control

    In control

    All QC runs are within the acceptable +SD

  • 8/12/2019 ABG Equipment

    38/43

    Arterial Blood Gases

    Quality Control

    Trend

    All QC runs within the acceptable +SD buttrending towards one side of the SD

  • 8/12/2019 ABG Equipment

    39/43

    Arterial Blood Gases

    Quality Control

    Random Error

    All QC runs within the acceptable +SD exceptfor one run

  • 8/12/2019 ABG Equipment

    40/43

    Arterial Blood Gases

    Quality Control

    Out of control

    Two or more QC runs are out of theacceptable +SD

  • 8/12/2019 ABG Equipment

    41/43

    Capnography

    Capnography is the continuous, noninvasivemonitoring of expired CO2 and analysis of thesingle-breath CO2 waveform

  • 8/12/2019 ABG Equipment

    42/43

    Capnography

    Capnography is performed utilizing:

    Infrared analyzer

    CO and CO2 absorb infrared radiation

  • 8/12/2019 ABG Equipment

    43/43

    Capnography

    Capnography is performed utilizing:

    Infrared analyzer

    Requires accurate calibration

    2 gas concentrations used

    Room air 5% CO2 mixture

    Inaccurate reading can occur when:

    Condensation of water in sample tubing,connectors, or sample chamber

    Flow changes after calibration

    Saturation of a desiccant column

    Long sampling lines can cause waveformdamping