Aacc Talk On Calibration Verification 2009

77
AACC Short Course 73214 How to Fulfill the CLIA Requirements for Calibration, Calibration Verification, and Reportable Range Anthony Killeen, MD, PhD William Castellani, MD (in conjunction with CAP)

Transcript of Aacc Talk On Calibration Verification 2009

Page 1: Aacc Talk On Calibration Verification 2009

AACC Short Course 73214How to Fulfill the CLIA Requirements

for Calibration, Calibration Verification, and Reportable Range

Anthony Killeen, MD, PhDWilliam Castellani, MD

(in conjunction with CAP)

Page 2: Aacc Talk On Calibration Verification 2009

This PowerPoint Slide Set

• http://tinyurl.com/kmq8vu

Page 3: Aacc Talk On Calibration Verification 2009

Outline

• Calibration Verification• Linearity• CVL Program at the CAP• Examples and Troubleshooting• Cases• Discussion of Selected CAP Checklist

Questions

Page 4: Aacc Talk On Calibration Verification 2009

Calibration Verification

• Periodic verification that the calibration of the analytical system remains valid

• Typically assessed by comparing test results with appropriate target values

• If the calibration changes, then patient test result values will also change

Page 5: Aacc Talk On Calibration Verification 2009

Linearity

• From CLSI document EP6-A (2003)– A quantitative analytical method is linear when

there exists a mathematically verified straight-line relationship between the observed values and the true concentrations or activities of the analyte.

– …if the true amount of an analyte in a sample were to double, clinicians expect that would be reflected in a doubling of the measured value.

– (Notice that there’s no reference to accuracy)

Page 6: Aacc Talk On Calibration Verification 2009

Calibration Verification vs. Linearity

• Calibration verification is the process of verifying agreement between calibrators (or other materials of known analyte concentrations) and measured values

• Calibrators should be traceable to a reference method to ensure accuracy

• Linearity evaluation does not require knowledge of the “true” analyte concentration

• “Linearity” does not appear in CLIA

Page 7: Aacc Talk On Calibration Verification 2009

Analytical Measurement Range vs. Linearity

• AMR is the range of analyte values that a method can directly measure without any dilution, concentration, or other pretreatment of the specimen that’s not part of the usual assay process

• A method should be linear across the AMR

Page 8: Aacc Talk On Calibration Verification 2009

Linearity as Defined in EP6-A vs. Instrument Response

• Linearity typically refers to the final analytical result, not the raw instrument output. A plot of analyte concentration vs. the raw instrument output may not be linear (e.g., competitive immunoassay)

• Linearity as used here is a straight-line relationship between “true” analyte concentrations and measured concentrations

Page 9: Aacc Talk On Calibration Verification 2009

Formal Calibration Verification/Linearity Program

• College of American Pathologists– CVL Surveys

• Other programs e.g., Maine Standards

Page 10: Aacc Talk On Calibration Verification 2009

Advantages to Participating in a CVL Program

• Supplements proficiency testing (PT)• Linearity testing often has smaller absolute

limits for error based on medically or analytically relevant criteria than does PT

• Challenges the full calibration range• Can detect problems earlier than QC or PT• Calibration verification is required by CLIA,

CVL program provides the materials and data analysis

Page 11: Aacc Talk On Calibration Verification 2009

Linearity Study Materials

• Generally 5 samples that span the AMR• Preparation– Individually spiked samples– Serial dilutions of a high concentration sample– Admixtures of low and high concentrations (CAP)– At least 2 replicates at each level to assess

random error

Page 12: Aacc Talk On Calibration Verification 2009

CAP Linearity Assessment

• Uses a polynomial method described in EP6-A• For statistical approach see – EP6-A CLSI/NCCLS Document, Wayne, PA (2003)– Kroll MH et al., Arch Pathol Lab Med 124:1331-8, 2000

Page 13: Aacc Talk On Calibration Verification 2009

Calibration Verification Evaluation

• Results reported as “verified” or “different”• If “verified”, then differences between your

means and the peer-group means are within the allowable error limits in the range specified

• If “different”, then at least one of your values exceeds the peer-group allowable error limits

Page 14: Aacc Talk On Calibration Verification 2009

Tabular Data

Graphical Data

Peer GroupSummary

Result

Calibration Verification Report Example

Page 15: Aacc Talk On Calibration Verification 2009

Difference = 100 * (Your Mean – Peer Mean) / Peer Mean

Allowable Error = larger of: One half of goal for total error OR minimum detectable difference

Page 16: Aacc Talk On Calibration Verification 2009
Page 17: Aacc Talk On Calibration Verification 2009

100% 100%

Page 18: Aacc Talk On Calibration Verification 2009

Note that CLIA Allowable Error for PT is ± 10% for albumin

Page 19: Aacc Talk On Calibration Verification 2009

Calibration not verified

Need to re-calibrate

High biases

Page 20: Aacc Talk On Calibration Verification 2009
Page 21: Aacc Talk On Calibration Verification 2009
Page 22: Aacc Talk On Calibration Verification 2009

Proficiency Testing Data from Same Lab

Page 23: Aacc Talk On Calibration Verification 2009

CAP Linearity Evaluations

• Standard• Extended Range• Dilution• Results– Linear– Nonlinear– Imprecise

Page 24: Aacc Talk On Calibration Verification 2009

Result

Tabular Data

Graphical Data

Page 25: Aacc Talk On Calibration Verification 2009
Page 26: Aacc Talk On Calibration Verification 2009
Page 27: Aacc Talk On Calibration Verification 2009

Extended Range Linearity Evaluation

• Extended range evaluation is used if– Survey includes an additional high specimen that

is spiked and not part of the admixture set– If there is a large gap in the relative

concentrations of the admixture ratios such that a specimen(s) may have a large influence on the fitted regression line, the specimen(s) will be considered to be extended range

Page 28: Aacc Talk On Calibration Verification 2009

Scoring Extended Range Specimens

• Specimens evaluated for linearity excluding the extended range specimen

• If a range that includes the highest non-extended range specimen is linear, then the extended range specimen is evaluated

• If the mean of the extended range specimen value falls on the extrapolated line (within allowable error) then the linear range will include the extended range specimen

Page 29: Aacc Talk On Calibration Verification 2009

Scoring Extended Range Specimens (2)

• The extended range specimen is not used to calculate the best-fit line

Page 30: Aacc Talk On Calibration Verification 2009
Page 31: Aacc Talk On Calibration Verification 2009
Page 32: Aacc Talk On Calibration Verification 2009

Dilution Validation Linearity Evaluation

• Performed if– Lab reports that it diluted at least one specimen– Enough undiluted specimens (normally 5) are

reported for evaluation using the standard linearity evaluation

Page 33: Aacc Talk On Calibration Verification 2009
Page 34: Aacc Talk On Calibration Verification 2009
Page 35: Aacc Talk On Calibration Verification 2009

EXAMPLES AND TROUBLESHOOTING

Page 36: Aacc Talk On Calibration Verification 2009

Linearity demonstrated for LN-01 to LN-05 only

Linearity Example 1: Standard Evaluation, specimens excluded from the linear range

Page 37: Aacc Talk On Calibration Verification 2009

Linearity Example 1: Standard Evaluation, specimens excluded from the linear range

Page 38: Aacc Talk On Calibration Verification 2009

Example 1: Troubleshooting

• Does the linear range cover the AMR? If the high specimen is above the AMR, did you dilute? Was the dilution protocol followed?

• If the linear range does not cover the AMR, then there may be problems with reagents, specimen handling, or the test system

• Check QC, PT, calibration data• Address identified problems and re-run linearity• Consider adjusting AMR to cover the linear range

Page 39: Aacc Talk On Calibration Verification 2009

Example 2: Extended Range Evaluation, extended range sample not linear

Page 40: Aacc Talk On Calibration Verification 2009

Example 2: Extended Range Evaluation, extended range sample not linear

Page 41: Aacc Talk On Calibration Verification 2009

Example 2: Troubleshooting

• Confirm that the linear range covers the AMR• Check the dilution protocol. Did you dilute?

Dilution can introduce variability• If the linear range does not cover the AMR,

then there may be problems with reagents, specimen handling, or the test system

Page 42: Aacc Talk On Calibration Verification 2009

Example 3: Standard Evaluation, nonlinear data

(Best appreciated with graphical data)

Page 43: Aacc Talk On Calibration Verification 2009

Example 3: Standard Evaluation, nonlinear data

Page 44: Aacc Talk On Calibration Verification 2009

Example 3: Troubleshooting

• Is the peer group generally linear?• If the peer group is generally linear then there

may be problems with specimen handling or the test system

• Review QC, calibration, PT data• Eliminate specimen or reagent handling errors• Diagnose fix any identified test system failures• Re-run linearity study

Page 45: Aacc Talk On Calibration Verification 2009

Example 4: Dilution Validation Evaluation, nonlinear trend in non-diluted samples

Page 46: Aacc Talk On Calibration Verification 2009

Example 4: Troubleshooting• (Because non-diluted samples are nonlinear, no

evaluation of the diluted sample is performed• Is the peer group generally linear?• If the peer group is generally linear then there

may be problems with specimen handling or the test system

• Review QC, calibration, PT data• Eliminate specimen or reagent handling errors• Diagnose fix any identified test system failures• Re-run linearity study

Page 47: Aacc Talk On Calibration Verification 2009

Example 4: Standard Evaluation, large replicate imprecision

Page 48: Aacc Talk On Calibration Verification 2009

Example 5: Standard Evaluation, large replicate imprecision

Page 49: Aacc Talk On Calibration Verification 2009

Example 5: Troubleshooting

• Pattern suggests pipetting problems which should be carefully investigated

• Fix any identified test system failures• Re-run linearity study

Page 50: Aacc Talk On Calibration Verification 2009

Example 6: Standard Evaluation, only difference plot shows replicate imprecision (similar problem to Example 5)

Page 51: Aacc Talk On Calibration Verification 2009

Looks OK?

Example 6: Standard Evaluation, only difference plot shows replicate imprecision (similar problem to Example 5)

Page 52: Aacc Talk On Calibration Verification 2009

Example 6: Troubleshooting

• Pattern suggests pipetting problems which should be carefully investigated

• Fix any identified test system failures• Re-run linearity study

Page 53: Aacc Talk On Calibration Verification 2009

Example 7: Standard Evaluation, Imprecision from poor fit

Page 54: Aacc Talk On Calibration Verification 2009

Example 7: Standard Evaluation, Imprecision from poor fit

Here, the term “imprecision” refers to the inability to find a subset of data displayinglinearity with adequate precision. Points zigzag around the best-fit line.

Page 55: Aacc Talk On Calibration Verification 2009

Example 7:Troubleshooting

• Unusual pattern, may be difficult to diagnose• Is the peer group linear? If not, call Customer

Service• Highest value may be diluted and a dilution

error is masking loss of recovery at the high end of the AMR

• Specimen handling problems may cause this

Page 56: Aacc Talk On Calibration Verification 2009

Example 8: Extended Range Evaluation, Imprecise in non-extended range, extended range sample(s) not evaluated

Page 57: Aacc Talk On Calibration Verification 2009

Example 8: Extended Range Evaluation, Imprecise in non-extended range, extended range sample(s) not evaluated

Large gap indicates extended range

Imprecise. E.R. specimennot evaluated

Page 58: Aacc Talk On Calibration Verification 2009

LN-24

• Creatinine linearity survey• Uses human serum spiked with pure

crystalline creatinine• Accuracy assigned values (NIST) allow for

accuracy based grading in addition to calibration verification and linearity checks

Page 59: Aacc Talk On Calibration Verification 2009

Case 1. Calibration Verification (AST)

Page 60: Aacc Talk On Calibration Verification 2009

Linearity study show the data were linear

Page 61: Aacc Talk On Calibration Verification 2009

Case 2. Calibration Verification (calcium)

Page 62: Aacc Talk On Calibration Verification 2009
Page 63: Aacc Talk On Calibration Verification 2009

Case 3. Calibration Verification

Page 64: Aacc Talk On Calibration Verification 2009
Page 65: Aacc Talk On Calibration Verification 2009

Case 4. Linearity (testosterone)

Page 66: Aacc Talk On Calibration Verification 2009
Page 67: Aacc Talk On Calibration Verification 2009

Executive Summary Page from CVL Survey

Page 68: Aacc Talk On Calibration Verification 2009

SOME CAP CHECKLIST QUESTIONS

Page 69: Aacc Talk On Calibration Verification 2009

CHM. 13000

• Are calibration procedures for each method adequate, and are the calibration results documented?

Page 70: Aacc Talk On Calibration Verification 2009

CHM.13100

• Are high quality materials with method- and matrix-appropriate target values used for calibration and calibration verification whenever possible?

Page 71: Aacc Talk On Calibration Verification 2009

CHM.13125

• Are all calibration materials used for non-FDA cleared assays documented as to quality?

• NOTE: Standards used to prepare calibrators for non-FDA-cleared assays require certificates of purity from the vendor, or a check on purity as part of the initial assay validation process. The laboratory should document the accuracy of a new lot of calibrators by checking the new lot against the current lot.

Page 72: Aacc Talk On Calibration Verification 2009

CHM.13400• Are criteria established for frequency of recalibration or

calibration verification, and the acceptability of results?• • NOTE: Criteria typically include: • At changes of reagent lots for chemically or physically active or

critical components, unless the laboratory can demonstrate that the use of different lots does not affect the accuracy of patient/client test results and the range used to report patient/client test data

• QC fails to meet established criteria• After major maintenance or service• When recommended by the manufacturer• At least every 6 months

Page 73: Aacc Talk On Calibration Verification 2009

CHM.13500

• Is the method system recalibrated when calibration verification fails to meet the established criteria of the laboratory?

Page 74: Aacc Talk On Calibration Verification 2009

CHM.13600

• Is validation of the analytical measurement range (AMR) performed with matrix-appropriate materials which include the low, mid and high range of the AMR, and is the process documented?

Page 75: Aacc Talk On Calibration Verification 2009

CHM.13700

• Are criteria established for validating the analytical measurement range, and is compliance documented?

Page 76: Aacc Talk On Calibration Verification 2009

GEN.42085• Is the reportable range verified/established for each

analytic procedure before implementation?• NOTE: The reportable range includes all results that may be

reliably reported, and embraces two types of ranges: • The ANALYTICAL MEASUREMENT RANGE (AMR) is the

range of analyte values that a method can directly measure on the specimen without any dilution, concentration, or other pretreatment not part of the usual assay process

• The CLINICALLY REPORTABLE RANGE (CRR) is the range of analyte values that a method can measure, allowing for specimen dilution, concentration, or other pretreatment used to extend the direct analytical measurement range

Page 77: Aacc Talk On Calibration Verification 2009

Summary of Topics Covered

• Calibration Verification• Linearity• CVL Program at the CAP• Examples and Troubleshooting• Cases• Discussion of Selected CAP Checklist

Questions