A revolution in Analyzers with touch-screen display · 2020. 4. 11. · A revolution in Analyzers...

Click here to load reader

Transcript of A revolution in Analyzers with touch-screen display · 2020. 4. 11. · A revolution in Analyzers...

  • Bulletin 12A01A01-01E

    www.yokogawa.com/an/

    PH450G pH and Redox (ORP) analyzerSC450G Conductivity/Resistivity AnalyzerISC450G Inductive Conductivity Analyzer

    A revolution inAnalyzers with touch-screen display

  • Model PH450GpH and Redox (ORP) analyzer

    The PH450G offers the best accuracy in the industry by combining the pH measurement with advanced temperature compensation functionality, preloaded calibration standards and stability checks. The PH450G is a true multivariable analyzer that combines pH with Temperature and ORP (Redox) measurement and all these measurements can be utilised through the different output functions: two mA current outputs, four independent SPDT contact outputs and HART. Both DD and DTM files are available for direct connection to HART Handheld terminal, HIM monitor and Pactware PC configurator.

    Features:•AdvancedTemperaturecompensationmethodsinclude

    NEN6411 algorithm as best fit for ultra pure water analysis

    •Threepreloadedsetsofbuffertablescovermostcommonly used pH calibration practices

    •CapableofanalyzingpH,ORPorrHandTemperaturemeasurements at the same time with the same sensor

    •FullPIDcontrolpossibleforbothacidandalkalineaddition simultaneously

    •Supportsmanytemperaturecompensationelementsandsensor types

    The EXAxt 450 Series analyzer is another fine example of Yokogawa raising the benchmarks of innovation and quality. Yokogawa has refined its expertise in pH and conductivity measurement and control into this new and innovative product. It is packed with features designed to reduce commissioning time, minimize plant down time and simplify operation anad maintenance. It’s intuitive touch screen operation, presents process parameters and the advanced diagnostics in a clear and unambiguous way.

    Features of the EXAxt Series:• Touchscreen operation• Unique intuitive HMI menu structure in 8 languages• Intelligent step by step calibration routines• Unique problem solving guidance• Predefined buffer solutions• Various temperature compensation algorithms• Full P, PI, PID control on all outputs• Two mA-outputs and four SPDT relay contacts with display indicators• HART(R) Communications• Process data trending up to 2 weeks• Extensive logbooks for event storage

    SULFITE

    ACID

    NaOH

    Note: Illustration not actual product.

    The intuitive touch screen virtually eliminates the need for an instruction manual.

    The trend graph helps diagnose problems and view sensor performance over time.

    Its intuitive touch screen operation, presents process parameters and the advanced diagnostics in a clear and unambiguous way. Modern data processing methods enhance the ability to provide on-line monitoring, trending, logging of process events, and controlling to allow for rapid intervention in demand and changes in the process dynamics.

    Single and combination pH / ORP electrodes

    SC450GConductivity

    andResistivity

    SC450GConductivity

    andResistivity

    PH450GpH/ORPPH450GpH/ORP

    ISC450GInductive

    Conductivity

    ISC450GInductive

    Conductivity

  • Features:•Supports2-electrodeand4-electrodesensorsforhigh

    accuracy •Automaticcompensationofcableresistanceby4-wire

    measurement•Cellconditionmonitoringeliminatesriskoferrorsbycell

    fouling•Advancedtemperaturecompensationtechniquesforall

    pure water applications and concentration measurements in standard units

    •USPfunctionalitypresentinallconfigurationsincluding resistivity mode

    •0.5%ofreadingaccuracyoverthefullrangeof6decades

    POWER GENERATIONRegardless on how electricity is generated: water quality is essential.

    EXAxt 450 is especially suitable for this industry by the dedicated functionality.PH450G offers NEN6411 Temperature compensation methods for accurate pH analysis of the boiler feed water, the condensate and the steam quality.SC450G offers Cation, Morpholine and Ammonia temperature compensation functions for accurate Conductivity measurement of all water streams in the water-steam circuitry.

    ISC450G offers the wide range ability and high accuracy of the ISC450G and the corresponding sensors. Naturally dedicated algorithms for concentration control of the regeneration chemicals are standard. The efficient use of regeneration chemicals and rinsing water requires the accuracy offered by these analyzers.

    PULP AND PAPERThe Pulp and Paper industry requires hassle free measurements. The analyzers must be able to operate trouble free for extended periods in a rough environment with a minimum of maintenance.

    PH450G is especially suitable for the many pH and ORP control loops in the pulp and paper processing.

    SC450G is the best assurance for reliable monitoring of the condensate lines to protect the steam boilers.

    ISC450G is the ideal instrument to monitor the many effluent streams and the concentration control of the various liquor streams.

    Model SC450G Conductivity / Resistivity Analyzer

    The EXAxt SC450G is designed to combine the superior functionality of the Yokogawa EXA series with the ease of useofferedinpocketcomputers(PDA).TrulyuniqueistheEXAxt 450 series Human Machine Interface. The high resolution graphical display and the touch screen operation make all information visible to the operator. Configuration with the touch screen is as easy as operating a PDA. Simply choose the language of choice and on screen instructions assure that the best configuration for the application is obtained.

    2-electrode and 4-electrode conductivity sensors

    PHARMACEUTICAL The pharmaceutical industry demands high product reliability and stability. Any drift in the electrochemical analysis may indicate a deviation from the strict specifications of reagents or products.Accurate and traceable calibration of the electrochemical sensors and instruments is mandatory.

    The combination of Yokogawa instrumentation and Yokogawa/Hamilton sensors meet the expectations of this industry.All Conductivity analyzers meet the requirements that are specified in USP first issued with USP23.

    BREWERYIn modern breweries the repeatability of the product is the most important characte-ristic of quality. Strict process control is mandatory to assure this constant quality.

    Protection against contamination by microorganism is one of the major concerns in the Biotechnological process. This results in clean processing and well-controlled CIP processes. A good control of the CIP (Cleaning in Place) is necessary to combine effective cleaning with effective use of chemicals.

    With in-line cleaning there is an inherent risk of contamination of the product by possible malfunctioning of the cleaning processes and therefore accurate process analyzers are a must for quality monitoring.

    EXAxt 450 meets all the requirements from this industry for their needs for electrochemical analysis.

    Model ISC450G Inductive Conductivity Analyzer

    The new EXAxt ISC450G has been designed to meet thehighestmarketrequirementsoftoday'sindustry.Thisanalyzerisuniqueasitwillnolongerbetheuncertaintyfactorofyourqualitydemands,butinsteadwill bring you to a higher level. TheEXAxtwillhelpyouandguideyouastheuniqueHuman Machine Interface (HMI) will not only present you reliable process data, but diagnostics, trends, logbooks and step by step calibration routines to become your friend in analytical measurement and helpyougainingthathigherlevelofquality.Theinstrument is easy to set up, very intuitive and can be used without the need of an instruction manual.

    Features:•PreloadedcalibrationmethodsaccordingtoOIML

    recommendations•Preloadedmatricesformostcommonelectrolytes•Accuracyof0.5%plus1µS/cmfrom0to2000mS/cmwithonecellconstant

    •Tworangeswithtwotemperaturecompensationmethods can be used for different fluids measured with the same sensor.

    ISC40 Inductive Conductivity sensors

  • DIAGNOSTICS•NaturallyEXAxt450featuresself-diagnostics.Thewatchdogtimer

    checkscontinuouslytheintegrityofthesignalprocessingandfirmwareroutines

    •Howeverthereliabilityandstabilityoftheelectronicsarestillhighcomparedwiththesensorcharacteristics

    • Yokogawamonitorsthesensorsin-linetopreventprocesscontrolproblemswhensensorfailureoccursorwhensensorfoulingmightresultinmeasuringerrors

    DIA

    GN

    OST

    ICS

    INSTRUCTIONS• TheconceptofEXAxt450isbasedontheassumptionthattheuser

    hasnotimeorinteresttostudyInstructionManualswhenservicingtheelectrochemicalanalyzers

    • Thereforeallinformationaboutconfiguration,calibration,andtroubleshootingissharedwiththeuseron-lineinplainEnglish

    • Thisavoidswastingtimefortheoperatorormaintenancetechnician;italsoguaranteeserror-freemaintenanceandthusbetteraccuracyandreliabilityoftheanalysis

    • Bytheway:InadditiontoplainEnglishtheusercanchooseGerman,French,ItalianorSpanish

    INST

    RU

    CTI

    ON

    S

    MAINTENANCE• EXAxtmakesmaintenanceoftheelectrochemicalanalyzervery

    easywiththeon-lineinstructionsandthetouchscreenoperation•Hownicewoulditbetopredicttheneedformaintenanceor

    calibrationinthefuturetoallowscheduledmaintenancewithouttheriskofdowntimebyunexpectedfailure

    • EXAxtstoresthecalibrationresultofthelastfivecalibrationstopredictwhenthesensordriftexceedstheselectedlimits

    • Inadditiontothis,thediagnosticvariablesarelogged,andbasedonthisinformationapredictionismadeastowhenthesensorneedsmaintenancetopreventmeasuringerrorsoccuring

    MA

    INTE

    NA

    NCE

    FUNCTIONALITY• EXAxt450comesasacompleteelectrochemicalanalyzerwithfull

    Indication-Alarm-Controllingcapabilities• ThegraphicalLCDshowsuptothreedynamicvariablesandtrendgraphof

    theprimaryvariable• Thetwocurrentoutputscanbeusedtotransmitprocessvariablesor

    controlleroutput• ThefourSPDTcontactscanbeconfiguredasHI/LOcontactoutput,butalso

    ascontroloutputtosolenoidsormeteringpumps• Theoutputcanbepulselengthorpulsefrequencycontrolledbythepowerful

    PIDcontrolfunction

    FUN

    CTI

    ON

    ALI

    TYH

    ARD

    WA

    RE

    HARDWARE• TheelectronicsarehousedinaruggedcastAluminumhousing.• TheenvironmentalprotectionmeetsIP66/NEMA4Xandiscertifiedby

    KEMA•Accesstothespacywiringcompartmentiseasythroughthehingedfrontdoor• TouchscreenandDisplayarewellprotectedbythepolycarbonateflexible

    window• TheunitisideallysuitedforFieldandPanelmounting

    CO

    MPEN

    SATI

    ON

    COMPENSATION• Electrochemicalanalysisresultsareofteninfluencedbyotherprocess

    variables.Theseuncontrolledvariablesresultinpoorprocesscontrol,especiallywhentheuserisunawareoftheinfluence

    • EXAxt450deliversaccuratecompensationofthesevariablestoassureaccurateprocesscontrol

    Examples - CompensationoftemperatureonpHorORPvalues:NEN6411forstrong

    electrolytes;MATRIXorTCforotherapplications - Compensationofnon-linearityonconductivitymeasurements - MatrixcompensationoftemperatureforConductivitymeasurement

  • STANDARD SPECIFICATION (Common)

    Display GraphicalQuarterVGA(320×240pixels)LCDwithLEDbacklightandtouchscreen.

    Housing Case:CastAluminumhousingwithchemicallyresistantcoatingCover,Window:PolycarbonateProtection:IP66/NEMA4X/CSAType3SColor:SilvergreyMountinghardware:Pipe,PanelorWallmounting

    Dimension 144(W)×144(H)×Approx.144(D)mmPowersupply Ratings:100-240VACAcceptablerange;90to264VAC

    Ratings:50/60HzAcceptablerange;50Hz±5%,60Hz±5%Ratings:12-24VDCAcceptablerange;10.8to26.4VDC

    Environmentandoperationalconditions

    Ambienttemperature:-20to+55°CStoragetemperature:-30to+70°CHumidity:10to90%RHat40°C(100°F)(non-condensing)

    Safety,EMCandRoHSconformingstandards

    Safety:EN61010-1,EN61010-2-030,EN61010-2-201,CAN/CSAC22.2No.61010-1,CAN/CSAC22.2No.61010-2-030,CAN/CSAIEC61010-2-201,UL61010-1,UL61010-2-030,UL61010-2-201

    EMC:EN61326-1ClassA,Table2,EN61326-2-3,EN61000-3-2ClassA,EN61000-3-3,RCM:EN61326-1ClassA,KoreaElectromagneticConformityStandard

    RoHS:EN50581

    PH450G SPECIFICATION

    Measuringrages pH;-2to16pHORP;-1500to1500mVrH;0to100rH

    Transmissionsignals Twoisolatedoutputsof4-20mADC.Max.load600Ω.Bi-directionalHART®digitalcommunication,superimposedonmA1(4-20mA)signal.SelectablefrompH,temperature,ORPorrH.

    Contactoutputs FourSPDTrelaycontactswithdisplayindicators.High/Lowprocessalarms,controlfunction,wash,hold,fail.

    Contactinput Remotewashcyclestart.Temperaturecompensation

    Automaticormanual.CompensationtoNernstequation.Processcompensationbyconfigurabletemperaturecoefficient,NEN6411forwaterorstrongacids/basesorprogrammablematrix.

    Calibration Semi-automatic1or2pointcalibrationusingpre-configuredNIST,US,DINbuffertables4,7&9,orwithuserdefinedbuffertables,withautomaticstabilitycheck.Manualadjustmenttograbsample.

    Accuracy pHinput;≤0.01pHORPinput;≤1mVTemperature;≤0.3°C(≤0.4°CforPt100)Stepresponse;≤4secfor90%(pH7-pH4)

    RefertotheGS12B07C05-01Eformoredetails.

    SC450G SPECIFICATION

    Measuringranges Conductivity:0μS/cmto200mS/cm×C(Max.1999mS/cm)Resistivity:0.05kΩ/Cto999MΩ×cmTemperature:Pt1000:-20to250°C,Pt100&Ni100:-20to

    200°C,NTC8k55:-10to120°C,Pb36(JISNTC6k)-20to120°C

    Transmissionsignals Twoisolatedoutputsof4-20mADC.Max.load600Ω.Bi-directionalHART®digitalcommunication,superimposedonmA1(4-20mA)signal.Selectablefromconductivity,resistivity,concentrationandtemperature.

    Contactoutputs FourSPDTrelaycontactswithdisplayindicators.High/Lowprocessalarms,USP,controlfunction,hold,fail

    Contactinput Remoterangeswitching.Temperaturecompensation

    Automaticormanual,fortemperatureranges.Referencetemperaturecanbeprogrammablefrom0to100°C.

    Calibration Semi-automaticcalibrationusingpre-configuredOIML(KCl)buffer.Manualadjustmenttograbsample

    Accuracy Conductivity/resistivity;≤0.5%ofreadingTemperature;≤0.3°C(≤0.4°CforPt100)Stepresponse;≤4secfor90%(fora2decadestep)

    RefertotheGS12D08N05-01Eformoredetails.

    ISC450G SPECIFICATION

    Measuringranges Conductivity:0μS/cmto1999mS/cmTemperature:-20to140°C

    Transmissionsignals Twoisolatedoutputsof4-20mADC.Max.load600Ω.Bi-directionalHART®digitalcommunication,superimposedonmA1(4-20mA)signal.Selectablefromconductivity,concentrationandtemperature.

    Contactoutputs FourSPDTrelaycontactswithdisplayindicators.High/Lowprocessalarms,controlfunction,hold,fail

    Contactinput Remoterangeswitching.Temperaturecompensation

    Automaticormanual,fortemperatureranges.Referencetemperaturecanbeprogrammablefrom0to100°C.

    Calibration Semi-automaticcalibrationusingpre-configuredOIML(KCl)buffer.Manualadjustmenttograbsample.

    Accuracy Conductivity;≤0.5%ofreading±1.0μS/cmTemperature;≤0.3°CTemp.compensation;≤1%forNaCl,≤1%,formatrixStepresponse;≤4secfor90%(fora2decadestep)

    RefertotheGS12D06D05-01Eformoredetails.

    CO

    MM

    UN

    ICA

    TIO

    NS

    OPEN

    ARCH

    ITEC

    TURE

    COMMUNICATIONS•AlldynamicvariablesaretransmittedthroughHARTtotheHOST

    computerortoanyPC•ConfigurationscanbedownloadedoruploadedwiththeEXAxt

    ConfiguratorthatissuppliedfreeofchargethroughInternet• Remoteconfigurationofthemanyfunctionsandinterrogation

    ofthemanydiagnosticdatacanbedonethroughPactware.DownloadedfreeofchargebyYokogawacustomers.

    OPEN ARCHITECTURE•CompatibilityisthekeywordfortheEXAxt450series•AllunitsarecompatiblewithYokogawasensors,stockedbycustomersas

    spareparts.• TheEXAxt450seriesarealsocompatiblewithmostcommerciallyavailable

    sensorsmakingthemsuitableasplantstandardelectrochemicalanalyzers• Forexampleupto6differenttemperaturecompensationelementsare

    supported:alreadyaccuratelycalibrated

  • Bulletin 12A01A01-01E

    www.yokogawa.com/an/

    [Style: S2]

    Model SuffixCodeOptionCode Description

    PH450GSG450GISC450GPower

    Type

    Mounting Hardware

    HoodConduit adapter

    Tag Plate

    ---------------------------

    -A-D

    -A-U

    ---------------------------------------------------------------

    /UM/U/PM/H5/AFTG/ANSI/AM20/SCT

    pH/ORP ConverterConductivity/Resistivity ConverterInductive Conductivity ConverterAC Version (100-240VAC)DC Version (12-24VDC)General purpose versionFM version (*3)Universal mounting kit (panel, pipe, wall)Pipe and wall mounting hardware (*2)Panel mounting hardware (*2)Awning hood (stainless steel) (*2)G1/2 (*2)1/2NPT (*2)M20 (*2)Stainless steel tag plate (*1)

    Notes:*1 If the tag number is predefined with the purchase, Yokogawa will inscript the tag plate with the specified tag number, and program the tag number in the converter.*2 Option codes /U, /PM, /H5, /AFTG, /ANSI and /AM20 are not specified for FM version (-U).*3 Nonincendive for FM only.

    A

    D E F

    B C

    184(7.2")

    144(5.67")

    72(2.8")

    144(

    5.67

    ") 20(0.79")

    24.5

    (1")

    M20 cable gland(When shipped, not installed)

    220(8.66")

    A : For output signalB : For contact inputC : For sensor cableD : For contact output (S1 and S2)E : For contact output (S3 and S4)F : For power supply

    20(0.79")

    121.

    5(4.

    78")

    Grounding terminal (M4 screw)

    Adapter49

    (1.93")Approx.55(2.2")

    138(5.43")

    138(5.43")

    M6

    M6

    M5

    Unit: mm (inch)min.185 (7.25")

    min

    .195

    (7.7

    5")

    138(5.43")

    138

    (5.4

    3")

    +1 0

    +1 0

    Hood (Option code: /H5, some cutout on the left side cover)

    Adapter for Coduit Work (optional)(option code : / AFTG, / ANSI, / AM20)

    G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)M20 screw (/AM20)

    Dimensions

    Model and Suffix Codes