A biosensing method for detection of caffeine in coffee
Click here to load reader
Transcript of A biosensing method for detection of caffeine in coffee
A biosensing method for detection of caffeine incoffeeAndrea Pizzariello, Jozef Svorc, Miroslav Stred’ansky and Stanislav Miertus*POLYtech, Soc. Coop. r.l., Padriciano, 99, I-34012 Trieste, Italy
Abstract: A speci®c inhibition of 3',5'-cyclic phosphodiesterase (CPDE) from bovine heart by methyl-
xanthines was used in combination with a pH electrode to develop a new biosensing method for the
detection of caffeine in coffee. The potential response changes of the sensor were proportional to the
concentration of caffeine in the range 0±4mgmlÿ1. The response time was about 2±4min. The standard
deviation of ®ve measurements of a 0.2mgmlÿ1 caffeine solution was�7.1mgmlÿ1. The electrode gave a
detection limit of 0.6mglÿ1 caffeine. The concentration of caffeine in espresso coffee was analysed.
This model gave excellent correlation between observed and predicted caffeine values. This electrode
exhibits advantages such as fast response, short conditioning time and low cost of the instrumentation
used. We also expected to be able to perform the detection of caffeine in food and clinical analysis.
# 1999 Society of Chemical Industry
Keywords: sensor; caffeine; determination; cyclic nucleotide phosphodiesterase; coffee
INTRODUCTIONCaffeine, 1,3,7-trimethylxanthine, is a natural alkaloid
which is present, together with other trace purines, in
coffee1 and also in tea and soft drinks. Coffee is one of
the most popular drinks across the world and its
enormous commercial and social importance is ob-
vious. The total annual world production of green
coffee is about six million tons,2 with an estimated cost
of six millard US dollars. The roasting of the green
coffee beans yields the commercial coffee which is the
basis of the worldwide bitter and dark coffee bev-
erages, such as espresso. Caffeine contributes to a
particular proportion of the perceived bitterness of a
cup of coffee and also makes a small contribution to
espresso's strength and body.3 Moreover, the impor-
tance of the mild stimulating properties of caffeine
should also not be neglected: the preference between
higher and lower levels of stimulation from caffeine
may in practice dictate the blend consumed, therefore
its taste. Because of the relatively important role of
caffeine in determining the quality of coffee beverages,
the development of a sensitive, fast and cost-effective
method for monitoring caffeine is greatly needed.
Normally, high-performance liquid chromatogra-
phy separation4±8 and UV spectrophotometric detec-
tion9,10 methods are applied to both regular and
decaffeinated green and roasted coffees for caffeine
content determinations. Also, other methods such as
capillary electrophoresis,11±14 thin layer chromatogra-
phy15 and gas chromatography,16 are used for separa-
tion of caffeine in the analysis of mixtures, combined
with several other detection methods such as mass
spectroscopy16 and FTIR spectrophotometric
measurements.17±19 However, very costly instrumen-
tation, highly skilled technicians and complicated and
time-consuming procedures are required for such
methods.
Another possible technique is ¯ow injection
immunoassay using a solid phase reactor, which makes
the assay faster because no separation step is needed.20
However, the time and cost for monoclonal antibody
production and puri®cation, and the need for their
manipulation with extreme care, make this approach
not so easy to perform.
The development and application of new caffeine
detection methods remain an active area of investiga-
tion, particularly in food and clinical chemistry.
Signi®cant research and development activity has
been devoted to preparing compact analytical devices
comprising a bioactive sensing element integrated with
a suitable transducting system, also known as biosen-
sors,21 for determination of various inorganic, organic
and biological substances. The main advantages of
these devices are their speci®city, sensitivity and
simple preparation, and the fact that no other reagents
besides a buffer and a standard are usually required.
Here we report the construction and properties of a
new, simple, speci®c, rapid and low-cost caffeine
biosensing system that consists of a glass membrane
potentiometric pH-sensitive electrode with phospho-
diesterase 3',5'-cyclic nucleotide and calmodulin
(CPDE±CM) as add-on reagents. The concept of
Journal of the Science of Food and Agriculture J Sci Food Agric 79:1136±1140 (1999)
* Correspondence to: S Miertus, ICS, Unido, Padriciano, 99, I-34012 Trieste, ItalyE-mail: [email protected]/grant sponsor: ILLYCAFFE’ SpA, Trieste(Received 24 December 1997; revised version 12 November 1998; accepted 26 January 1999)
# 1999 Society of Chemical Industry. J Sci Food Agric 0022±5142/99/$17.50 1136
the caffeine sensor is based on the hydrolysis of
adenosine 3',5'-cyclic monophosphate (c-AMP) to
adenosine 5'-monophosphate (AMP) and H3O�
through the effect of CPDE in the presence of the
protein activator (CM) and Ca2�:
c-AMP�H2Oÿÿÿÿÿÿÿÿ!CPDE
CM;Ca2�AMP�H3O�
When this enzymatic reaction is inhibited by
caffeine,22 the H3O� production, monitored by a
potentiometric pH probe, should be reduced in an
inversely proportional way with respect to the caffeine
concentration.
MATERIALS AND METHODSApparatusA glass membrane pH electrode (Radiometer model
PHI 2401) was used as the basis of the caffeine sensor.
The electrode was connected with a pH meter
(Radiometer model PHM 85). The sensor output
was registered on a Y±t paper recorder (AMEL model
868).
A Varian Cary-3E UV-vis spectrophotometer was
employed to obtain the UV-vis spectra. All measure-
ments were carried out at room temperature in 5 or
10ml beakers with magnetic stirring.
ReagentsAdenosine 3',5'-cyclic monophosphate (cod.n.
A9501), phosphodiesterase 3',5'-cyclic nucleotide
(EC 3.1.4.17) crude complex from bovine heart
(cod.n. P 0134), calmodulin (phosphodiesterase
3',5'-cyclic nucleotide activator), phosphodiesterase
I-type VII (EC 3.1.4.1) from Crotalus atrox venom
(cod.n. P 6761), caffeine, theophylline and theo-
bromine were obtained from Sigma Chemical Co, St
Louis, MO, USA; tris(hydroxymethyl)aminomethane
hydrochloride (TRIS-HCl) was obtained from Fluka
Chemie AG, Buchs, Switzerland; other analytical
grade reagents were commercially available. Milli-Q
(18.5Mcm) water was used throughout.
Phosphodiesterase assay and protein determinationCommercial CPDE titre was assayed by measuring the
release of inorganic phosphate with use of an excess of
5'-nucleotidase.23,24 The reaction mixture contained
0.36mmoles of cyclic-3',5'-AMP, 1.8mmoles of
MgSO4 and 36mmoles of TRIS-HCl buffer, pH 7.5,
with a suitable dilution of the phosphodiesterase
sample being tested in a total volume of 0.9ml. This
mixture was incubated at 30°C for 30min. After the
®rst 20min of incubation, 0.1ml of phosphodiesterase
I-type VII from Crotalus atrox venom solution was
added, containing 0.02U of 5'-exonuclease in 10mM
TRIS-HCl, pH 7.5. The entire reaction was termi-
nated by addition of cold 550gkgÿ1 trichloroacetic
acid. The resulting precipitate was removed by
centrifugation, and aliquots of the supernatant ¯uid
were analysed for inorganic phosphate by the following
method: 0.1ml of 25gkgÿ1 ammonium molybdate in
5N H2SO4 was added to a 0.5ml aliquot of the
supernatant solution and 0.35ml of distilled water.
The colour was read at 660nm. One unit of CPDE
enzyme was de®ned as the amount of the enzyme that
hydrolyses 1.0mmole of c-AMP to AMP per minute at
pH 7.5 at 30°C. Protein content was determined by
the Bradford Bio-Rad protein assay using serum
albumin as a standard.25
Determination of caffeine in standard solutionThe determination of caffeine by the sensor was
carried out at room temperature under a nitrogen
atmosphere in a jacketed reaction vessel equipped with
magnetic stirring. The glass membrane electrode was
immersed in stirred water solution (5ml) supple-
mented with CaCl2 (0.01mM) and calmodulin
(10Umlÿ1). The substrate (c-AMP) was added to
the system (®nal concentration 0.2 or 1mM) and the
pH was adjusted to 8.00 by addition of 0.2M TRIS-
HCl buffer, pH 9. After signal stabilisation the
hydrolysis was started by addition of CPDE (®nal
concentration 0.0424Umlÿ1). After the electrode
response became constant (3±5min), the sensor
calibrations were conducted in the open cell with
intensive magnetic stirring at room temperature. The
caffeine concentration was increased stepwise by
adding de®ned volumes of a 20mgmlÿ1 stock solu-
tion. Potential difference DE vs time curves were
recorded. The decrease in the recorded line slope was
proportional to the caffeine concentration.
Determination of caffeine in espresso coffeeTo measure the caffeine content of an espresso coffee,
5ml of water supplemented with CaCl2 (0.01mM),
calmodulin (10Umlÿ1) and c-AMP (®nal concentra-
tion 0.2 or 1mM) was placed in a 10ml beaker and the
pH was adjusted to 8.00 by addition of 0.2M TRIS-
HCl buffer, pH 9, then 10ml of ¯occulated espresso
coffee was added. Typical response times ranged from
60 to 120s. Complete re-establishment of the original
electrode potential was accomplished in less than
1min by replacing the electrode in caffeine-free water
supplemented as above for a new coffee sample
analysis. Caffeine concentrations were derived from
the calibration plot.
The ¯occulation procedure was based on the
treatment of 5ml of crude espresso coffee with 500ml
of a solution of potassium hexacyanoferrate(II) tri-
hydrate (0.150gmlÿ1 K4[Fe(CN)6].3H2O) ®rst, then
500ml of a solution of zinc chloride anhydrate
(0.1422gmlÿ1 ZnCl2) was added. The suspension
was mixed and ®nally ®ltered.
RESULTS AND DISCUSSIONpH dependence and buffer capacityThe hydrolysis of c-AMP to AMP at pH values
between 7 and 8 produces a change in pH during the
J Sci Food Agric 79:1136±1140 (1999) 1137
Detection of caffeine in coffee
reaction. In this context we have evaluated the effect of
CPDE-substrate concentration (c-AMP) on the rate
of hydrolysis. We have found that with increasing
c-AMP concentration the rate of hydrolysis decreases.
Since in our system the detected species are H3O�
ions, their sensitivity depends strongly on buffer
capacity.
To adjust the pH to the optimal value for CPDE
(pH 8.0), it is necessary to add an alkaline buffer
solution (0.2M TRIS-HCl pH 9.0). If the concentra-
tion of c-AMP increases, more buffer is necessary and
the buffer capacity increases as well. Therefore a lower
sensitivity at higher substrate concentration is ob-
served.
Amount of enzyme and substrate (c-AMP)concentrationIn Fig 1 the effect of CPDE activity on the rate of
hydrolysis at constant concentration of c-AMP (1mM)
is shown. In this case the rate of hydrolysis (expressed
as the change in potential, DEdtÿ1) is a linear function
of CPDE activity.
The in¯uence of c-AMP concentration on the
sensitivity of the caffeine sensor is also due to the fact
that CPDE inhibition by methylxanthine derivatives is
apparently competitive.23 Thus a c-AMP concentra-
tion increase should cause of an additional lowering of
the sensitivity.
Effect of calmodulin (CM)The CPDE activator CM played a most important role
in the performance of the sensor. The addition of
10Umlÿ1 of CM to the assay solution resulted in a
more than threefold increase in the hydrolytic activity
of CPDE, although the enzyme used contained
`saturated activity' (declared by the supplier) of the
activator.
Analytical characteristics of the sensorThe potentiometric response of the caffeine sensor
with CPDE and CM in solution was examined using
caffeine standard solutions or directly with the solid
substance dissolved in the assay solution. The poten-
tial (E) was recorded as a function of time (t) and the
slope of the experimental curve is proportional to the
rate of c-AMP hydrolysis.
Fig 2 shows the calibration graph which was
constructed by plotting the steady state values of
DEdtÿ1 against the caffeine concentration. In addition,
Fig 3 shows another calibration plot obtained for a
higher concentration of c-AMP (1mM). While at a
c-AMP concentration of 1mM the dependence is
linear up to 4mgmlÿ1 of caffeine, at a c-AMP
concentration of 0.2mM the dependence is not linear
but the sensitivity is signi®cantly higher. Although at
the c-AMP concentration of 1mM the sensitivity is
decidedly lower, the range of linearity is doubled,
making this sensor useful for determining caffeine in
the range of 0±4mgmlÿ1.
The potentiometric response of the sensor was
reproducible. The standard deviation for ®ve measure-
Figure 1. Influence of CPDE activity on rate of c-AMP hydrolysis (DEdtÿ1)for 1mM c-AMP (points are means of five replications for eachconcentration; related SDs are in range �3.3%).
Figure 2. Influence of caffeine concentration on rate of c-AMP hydrolysis(DEdtÿ1) for 0.2mM c-AMP and 0.0424Umlÿ1 CPDE activity (points aremeans of five replications for each concentration; related SDs are in range�3.7%).
Figure 3. Influence of caffeine concentration on rate of c-AMP hydrolysis(DEdtÿ1) for 1.0mM c-AMP and 0.0424Umlÿ1 CPDE activity (points aremeans of five replications for each concentration; related SDs are in range�3.5%).
1138 J Sci Food Agric 79:1136±1140 (1999)
A Pizzariello et al
ments of a 0.2mgmlÿ1 caffeine solution (at pH 8.00)
was �7.1mgmlÿ1. The estimated detection limit was
0.6mglÿ1.
InterferencesThe inhibition of CPDE by theophylline, theobromine
and caffeine has been described in the literature.26±29
We tested the effect of theophylline and theobromine
on the rate of c-AMP hydrolysis. We have found
inhibition of c-AMP hydrolysis by both methyl-
xanthines. Fig 4 shows the relative inhibition of CPDE
hydrolytic activity as a function of theophylline,
caffeine and theobromine concentrations. At 50%
inhibition, theophylline was 16.7- and 10-fold more
potent than theobromine and caffeine respectively. In
any case, theophylline and theobromine are present
only in trace amounts, in comparison with caffeine, in
coffee products,30 so the present bioanalytical proce-
dure is applicable for caffeine content determination in
coffee beverages without signi®cant interferences.
Measurements of real samplesThe sensor was also applied for the direct determina-
tion of caffeine in real samples. Typical response
curves are shown in Fig 5, directly obtained with
espresso coffee samples.
Espresso coffee is served in a small, heavy china cup
with a capacity of 30±50ml, half-full with a dark brew
topped by a thick, clean, light reddish-brown foam of
tiny bubbles. The liquid part of the espresso is a very
complex matrix. It is in fact a concentrated solution of
salts, acids, sugars, caffeine and many other complex
substances, forming the matrix wherein three dis-
persed phases coexist (emulsion, suspension and gas
bubbles).
In our enzyme pH-based caffeine sensor, biospeci-
®city is obviously de®ned by use of CPDE, while the
electrochemical selectivity to caffeine and non-speci®c
buffering agents has to be dealt with by removal of
interfering components by gravity sedimentation,
aided by chemical ¯occulating agents. Thus, in the
sensor system presented in this paper, the major
contribution of the endogenous interferences can be
preliminarily eliminated by a simple and fast ¯occula-
tion procedure of the coffee samples. The ¯occulation
procedure leads to aggregation of large molecules or
colloidal particles, suspended in the coffee sample,
into clusters or clumps. When this occurs, by addition
of the appropriate coagulant agent, the aggregates
precipitate or separate from the dissolved state, with
complete removal of total particulate, colour and
turbidity, and thus major interfering components are
eliminated.31
The time of a single analysis in this way is as short as
6±8min, which can be considered the minimum time
attainable with a method involving also ¯occulation
pre-treatment. It is possible to perform about 10 assays
per hour. Table 1 reports a series of sample analyses
from three different kinds of coffee. The same samples
were also analysed by an HPLC method (De Palo,
unpublished). The results give a correlation plot y(caffeine sensor determination)=1.159x (HPLC
determination)ÿ0.167, with a coef®cient of correla-
tion of 0.999. This indicates good agreement between
the two methods.
CONCLUSIONSThese preliminary results demonstrate that a potentio-
Figure 4. Relative potencies of methylxanthine inhibition of c-AMPhydrolysis for 0.2mM c-AMP and 0.0424Umlÿ1 CPDE activity.
Figure 5. Profiles for potential response of caffeine electrode exposed toflocculated espresso coffee samples: (b1) reference baseline; (a) 10mlflocculated water; (b) 10ml decaffeinated coffee; (c) 10ml arabica coffee;(d) 10ml robusta coffee. Solution: 0.2mM c-AMP, 10Umlÿ1 calmodulin,0.32mM TRIS-HCl, pH 8.0.
Table 1. Analysis of espresso coffee samples by an HPLCmethod and by the caffeine sensor (mean�SD, n=3)
Caffeine (mgmlÿ1)
Coffee HPLC method a Caffeine sensor
Decaffeinated 0.35�0.05 0.27�0.04
Arabica 1.41�0.07 1.0�0.06
Robusta 2.32�0.15 2.56�0.12
a Performed at the Standard Chemical Laboratory for the
Technology of Commerce, Trieste, Italy.
J Sci Food Agric 79:1136±1140 (1999) 1139
Detection of caffeine in coffee
metric enzyme sensor can be used successfully for the
direct detection of caffeine in coffee beverage samples.
The system described here does not require any
sophisticated equipment, being based on a pH meter
with a glass membrane electrode. Up to 10 samples
can be analysed per hour. Thus the sensor system is
obtained with minimum instrumental cost and maxi-
mum speed of analysis. The manufacture of a sensor
with immobilised enzyme on the electrode surface and
the evaluation of its performance are now in progress.
ACKNOWLEDGEMENTSThis work has been carried out within a project
supported by ILLYCAFFE' SpA. (Trieste, Italy). We
also thank E Illy and M Petracco from ILLYCAFFE'
SpA for valuable discussion and scienti®c support.
REFERENCES1 Kappler AW and Baumann TW, Purine alkaloid pattern in coffee
beans. Proc 11th ASIC Colloq, pp 173±279 (1985).
2 Martin MJ, Pablos F and Gonzalez AG, Application of pattern
recognition to the discrimination of roasted coffees. Anal Chim
Acta 320:191±197 (1996).
3 Illy A and Viani R, De®nition of quality. In Espresso Coffee: the
Chemistry of Quality, Ed by Illy A and Viani R, Academic Press,
London, pp 24±26 (1995).
4 Maier HG, Tillack B and Holler U, Determination of caffeine in
roasted coffee. D-Lebensm Rundschau 92:283±285 (1996).
5 Mayer A, Ngiruwonsanga T and Henze G, Determination of
adenine, caffeine, theophylline and theobromine by HPLC
with amperometric detection. Fres J Anal Chem 356:284±287
(1996).
6 Budvaribarany Z, Szasz G and Gyimesiforras K, Optimized and
validated HPLC methods for compendial quality assessment
methylxanthine derivatives. J Liq Chromatogr Relat Technol
20:1233±1242 (1997).
7 Moriyasu T, Saito K, Nakazato M, Ishikawa F, Fujinuma K,
Nishima T and Tamura Y, Simultaneous determination of
caffeine, theobromine and theophylline in foods by HPLC. J
Food Hyg Soc Jpn 37:14±19 (1996).
8 Stadler RH and Fay LB, Antioxidative reactions of caffeine:
formation of 8 oxocaffeine (1,3,7 trimethyluric acid) in coffee
subjected to oxidative stress. J Agric Food Chem 43:1332±1338
(1995).
9 Atay O and Oztop F, Quantitative determination by using HPLC
and GLC methods for cocaine-HCl in synthetic binary
mixtures with procaine-HCl, lidocaine-HCl and caffeine. Anal
Lett 30:565±584 (1997).
10 Martin MJ, Pablos F, Bello MA and Gonzalez AG, Determina-
tion of trigonelline in green and roasted coffee from single
column ionic chromatography. Fres J Anal Chem 357:357±358
(1997).
11 Matsuda R, Hayashi Y, Sasaki K and Saito Y, Deductive
prediction of measurement precision and optimization of
integration time and wavelength in capillary electrophoresis.
Chromatographia 41:707±714 (1995).
12 Walker JC, Zaugg SE and Walker EB, Analysis of beverages by
capillary electrophoresis. J Chromatogr A 781:481±485 (1997).
13 Zhao YP and Lunte CE, Detemination of caffeine and its
metabolites by micellar electrokinetic capillary electrophoresis.
J Chromatogr B 688:265±274 (1997).
14 Vogt C, Conradi S and Rohde E, Determination of caffeine and
other purine compounds in food and pharmaceuticals by
micellar electrokinetic chromatography. J Chem Educ 74:1126±
1130 (1997).
15 Kunugi A and Tabei K, Simultaneous determination of purine
alkaloids in daily foods by high performance thin layer
chromatography. HCR J High Resol Chromatogr 20:456±458
(1997).
16 Statheropoulos M, Smaragdi E, Tzamtzis N and Georgakopou-
los C, Principal component analysis for resolving coeluting
substances in gas chromatography mass spectrometry doping
control analysis. Anal Chim Acta 331:53±61 (1996).
17 Norton KL and Grif®ths PR, Performance characteristics of a
real time direct deposition supercritical ¯uid chromatography
fourier transform infrared spectrophotometry system. J Chro-
matogr A 703:503±522 (1995).
18 Daghbouche Y, Garrigues S, Vidal MT and Delaguardia M,
Flow injection Fourier transform infrared determination of
caffeine in soft drinks. Anal Chem 69:1086±1091 (1997).
19 Bouhsain Z, Garrigues S and Delaguardia M, Clean method for
simultaneous determination of propyphenazone and caffeine
in pharmaceuticals by ¯ow injection Fourier transform infrared
spectrometry. Analyst 122:441±445 (1997).
20 Rico CM, Fernandez MD, Gutierrez AM, Conception M,
Conde P and Camara C, Development of a ¯ow ¯uoro-
immunosensor for determination of theophylline. Analyst
120:2589±2591 (1995).
21 Frew JE and Hill HAO, Electrochemical biosensors. Anal Chem
59:933A±944A (1987).
22 Beavo JA and Reifsnyder DH, Primary sequence of cylic
nucleotide phosphodiesterase isoenzymes and the design of
selective inhibitors. Trends Pharmacol Sci 11:150-±155 (1990).
23 Butcher RW and Sutherland EW, Adenosine 3', 5' phosphate in
biological materials. Puri®cation and properties of cyclic 3',5' -nucleotide phosphodiesterase and use of this enzyme to
characterize adenosine 3',5' phosphate in human urine. J Biol
Chem 237:1244±1250 (1962).
24 Ho HC, Teo TS, Desai R and Wang JH, Catalytic and regulatory
properties of two forms of bovine heart cyclic nucleotide
phosphodiesterase. Biochim Biophys Acta 429:461±473 (1976).
25 Bradford MM, A rapid and sensitive method for the quanti®ca-
tion of microgram quantities of proteins utilizing the principle
of protein±dye binding. Anal Biochem 72:248±254 (1976).
26 Leblanc J, Richard D and Racotta IS, Metabolic and hormone
related response to caffeine in rats. Pharmacol Res 32:129±133
(1995).
27 Howell LL, Cof®n VL and Spealman RD, Behavioral and
physiological effects of xanthines in non human primates.
Psycopharmacology 129:1±14 (1997).
28 Zatz M and Heath JR, Calcium and photoentrainment in chick
pineal cells revisited: effects of caffeine, thapsigargin, EGTA
and light on the melatonin rhythm. J Neurochem 65:1332±
13341 (1995).
29 Schmidt J, Hatzelmann A, Fleissner S, Heimannwetshat I,
Lindstaedt R and Szeleyi I, Effect of phosphodiesterase
inhibition on IL4 and IL5 production of the murine T(H)2
type T cell clone D10.G4.1. Immunopharmacology 30:191±198
(1995).
30 Viani R, The composition of coffee. In Caffeine and Health, Ed by
Grattini S, Raven Press, New York pp 17±41 (1993).
31 Stephenson RJ and Duff SJB, Coagulation and precipitation of a
mechanical pulping ef¯uent removal of carbon, colour and
turbidity. Water Res 30:781±792 (1996).
1140 J Sci Food Agric 79:1136±1140 (1999)
A Pizzariello et al