A 318 Lecture 2324 Ms Analyzers Hand Out

download A 318 Lecture 2324 Ms Analyzers Hand Out

of 10

Transcript of A 318 Lecture 2324 Ms Analyzers Hand Out

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    1/10

    Sample Exam Questions1. _____ Ions formed by ___ obtain multiple charges via proton transfer.

    A. EI

    B. FAB

    C. ESI

    D. MALDI

    2. _____ Which source should be use to produce singly charged parent ionsfor myoglobin (16591 Da)?

    A. CI

    B. EI

    C. ESI

    D. MALDI

    Mass AnalyzersContinuous (Filters)Quadrupole - Quad

    Magnetic Sector - MS

    Electrostatic Analyzer - ESA

    Pulsed

    Time-of-Flight (Reflectron) - TOF

    Ion Trap (quadrupole or linear) IT (QIT or LIT)

    Fourier Transform Ion Cyclotron Resonance FTICR

    Orbitrap

    Tandem

    Triple-Quad, Ion trap, FTICR, Q-TOF, TOF-TOF, IT-FTICR, IT-Orbitrap

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    2/10

    Mass Resolution / Mass Range

    Instrument

    11C,20C

    Resolution ability of mass analyzer to separate ions w/ similar m/z values

    Peaks are considered resolved if height of valley between them is less than10% of peak height

    Resolution, R = m/m R = 27.5/1 = 27.5 to resolveAl+ from Si+

    Al+ Si+

    27 28

    R = ~50

    Al+ Si+

    27 28

    R = ~100

    Al+ Si+

    27 28

    R = ~500

    Nominal mass nearest whole number mass (28 for CO, 44 for CO2

    )

    Mass Range from smallest to largest m/z separated by mass analyzer

    Quadrupole Mass Analyzer

    Instrument

    11B

    Most commonly used bench-top mass analyzer inexpensive, rugged,compact, high scan rates (100 ms/spectrum)

    +

    +- -

    four parallel cylindrical rods

    opposite rodshave samepotential

    stabletrajectory

    unstable trajectory

    One set of rods acts as low-pass filter, the other as high-pass filter

    m/z

    T

    m/z

    T

    m/z

    Tcombined

    (Figures 11-4, 11-5, 11-6, 11-7)

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    3/10

    Various Ion Guides Multi-poles

    Figure 5. Radial pseudopotential contour in volts forthe multipoles and tripole cross-section. All graphsshow that the minimum point is in the center of eachmultipole and increases when it moves to the

    electrodes. Calculation conditions: V0 = 10 V, = 1MHz, r0 = 3 mm, 1000 Da, ion charge = +1. (a)

    quadrupole; (b) hexapole; (c) octapole; (d) tripole.

    Figure 6. Ion guides AC voltage for stable ionguiding and tripole stable trajectory. 1 MHz,length 150 mm, r0 = 3 mm. Beam initialconditions: kinetic energy 10 eV, expansion

    angle 3, and initial radius = 0.5 r0. (a) AC-mass stability diagrams; (b) cross-section view;(c) transversal view of the ion stable movementinside of the tripole rotational electric field.

    Computer Simulations of a New Three Rods Ion Optic (TRIPOLE) with High Focusing and Mass Filtering Capabilities GaryAbdiel Salazar, Tsutomu Masujima J. American Society for Mass Spectrometry Vol. 18, Iss. 3 , March 2007, 413-421.

    Einzel Lens Ion beam focusing

    This shows the path of ions as they travel through a simulated Einzel lens. The ion mass is 200kDaaccelerated to 40kV, the middle plate potential is at 200V and the plate size is 1m these are verylarge settings used to show the focusing effect more easily and are not practical settings.

    plates

    OR

    rings

    Divergent ion beams arecollimated / focused afterpassing through 3-voltageelement

    http://en.wikipedia.org/wiki/Einzel_lens

    plates+V volts0 volts

    0 volts 0 volts

    0 volts

    +V volts

    ion path focus

    Electric field lines

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    4/10

    Magnetic Sector / Electrostatic Analyzer

    Instrument

    11B, 20C

    Double-focusing mass spectrometer

    Corrects for energy (in ESA) and directional (in MS) distributions of ions capable of very high (105) resolution - Single-focusing (MS only)instruments have lower (~2000) resolution

    In MS: m/z = B2r2e/2V, scan by varying B, V, or r

    Several arrangements for double-focusing instruments 2 common onesare Mattauch-Herzog (Fig. 11-9) with plane of double-focus & Nier-Johnson (Fig. 20-13) with point of double-focus

    ESA

    MS

    focalplane

    m1/zm2/z

    3150

    Mattauch-Herzog Nier-Johnson

    point ofdoublefocus

    MS

    energy

    focal planedirection

    focal plane

    ESA

    ESA slit

    +

    +

    -

    -

    Time-of-Flight Mass Analyzer

    zm

    eV

    dt2

    =

    Instrument

    11B, 20C

    Brief pulses (few ns) of ions produced at 10 Hz to 50 kHz and acceleratedby 1 to 20 kV into ~1-m field-free drift region where ions separate asinverse of m/z lighter ions arrive at detector prior to heavier ions flighttimes of 1 to 30 s requires very fast (ns bins) electronics

    0 V10 kV

    large

    electricfield

    signal

    detector1 m

    field-free region

    Simple, rugged, simultaneous detection, virtually unlimited mass range,rapid data acquisition (

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    5/10

    Delayed Extraction TOF MS

    Instrument

    notes

    Ions formed in source under field-free conditions (usually desorption from

    surface), allowed to drift due to inherent velocity after brief delay, electricfield is applied, pushing ions toward detector such that ions w/ smallerinitial velocities receive more acceleration, reaching detector at same timeas ions of same m/z with larger initial velocities

    initial desorption(0 s)

    5 kV

    5 kV

    0 kV

    delayed extraction(0.1 to 5 s)

    0 kV

    5 kV

    10 kV

    Length of delay and size of pulse determine the m/z that isoptimally focused

    Reflectron Time-of-FlightMass Analyzer

    amp comp

    0 kV

    L+M+H+

    +20 kV

    +22 kV

    0 kV 0 kV

    0 kV

    1) Ions enter source region,accelerated toward reflectron.

    2) Ions separate in space based ontheir relative mass-to-charge (m/z).

    3) Ions reverse path in reflectron.

    4) Ions impact detector.

    Flight time

    Signal

    Instrument

    notes

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    6/10

    Ion Trap Mass Analyzer

    Instrument

    20C

    Quadrupole (3-D) ion trap consists of a donut-shaped ring electrode & two

    end cap electrodes (Fig. 20-15) ions enter through one end cap, arestabilized by radio-frequency voltage applied to ring electrode while endcaps are at ground ions of individual m/z are sent through other end capto detector during scan rugged, compact, inexpensive

    todetector

    fromsource

    endcaps

    ringelectrode

    Linear (2-D) ion trap resembles quadrupole mass analyzer

    3D ion trap

    http://aemc.jpl.nasa.gov/activities/mms.cfm

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    7/10

    2D (linear) Ion Trap MS

    Simion simulation of the resonanceexcitation field comparing (a) a threesection device and (b) a single section

    device with end plates for axial trapping.

    (a) Simion simulation of 100 eV ions

    exiting the 2-D trap along the length ofthe exit slot and being focussed ontothe 15kV conversion dynode. (b) 10keV ions simulated.

    A two-dimensional quadrupole ion trap massspectrometer Jae C. Schwartz, Michael W. Senko,and John E. P. Syka J. American Society for Mass

    Spectrometry Vol. 13, Iss. 6 , June 2002, 659-669.

    Linear Ion Trap Instrument

    The overall instrument configuration along with typical operating voltages and pressures.

    A two-dimensional quadrupole ion trap mass spectrometer Jae C. Schwartz, Michael W. Senko, and

    John E. P. Syka J. American Society for Mass Spectrometry Vol. 13, Iss. 6 , June 2002, 659-669.

    ion transmissionmass

    analysis

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    8/10

    FT Ion Cyclotron Resonance

    Instrument

    20C

    Ions trapped in cubic cell inside bore of (usually superconducting) magnet

    excited from center by brief RF pulse into circular path in planeperpendicular to magnetic field, B (Fig. 20-16, 20-17)

    transmitterplate

    receiverplate

    receiverplate

    B

    Electronsdrawn throughcircuit as ionsapproach /recede fromreceiver plates

    Image current measured through circuit depends on the Free-InductionDecay (FID) of ions in their characteristic cyclotron frequency,c

    zeB

    mc=

    FTICR Signal Processing

    Instrument

    20C

    Excitation RF

    FID signal

    time

    Frequency (kHz)

    zeB

    mc=

    m/z (amu)

    FT

    Capable of very high (>106) resolution,improves with longer FID signal; veryexpensive (>$400k) due tosuperconducting magnet (>3 T)

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    9/10

    Orbitrap MS

    http://www.thermo.com/eThermo/CMA/PDFs/Various/File_31081.pdf

    http://www.thermo.com/eThermo/CMA/PDFs

    /Articles/articlesFile_30850.pdf

    High Resolving Power example

    http://www.thermo.com/eThermo/CMA/PDFs/Articles/articlesFile_30850.pdf

    Phosphatidylethanolamines

    PE 18:0/18:0 and

    PE O-16:1p/22:6

    (isobaric)

    Larger resolvingpower requires moretime per scan.

  • 7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out

    10/10

    Mass Analyzer Figures of Merit

    Instrument

    20C

    Resolution m/z range scan time

    Quadrupole Quad ~2000 < 2000 100 ms

    Magnetic Sector ~2500 < 2000 ~1 s

    Double-focusing (ESA/MS) 50,000 < 2000 ~1 s

    Time-of-Flight ~4000 unlimited 100 s

    Reflectron TOF 20,000 unlimited 100 s

    Ion Trap (3D) ~2000 < 2000 100 ms

    Linear (2D) Ion Trap ~2000 < 2000 20 ms

    FTICR > 200,000 50,000+ 0.1-10 s

    Orbitrap > 50,000 50,000+ 0.1-10 s

    Tandem Mass Analyzers

    Instrument

    notes

    Perform tandem mass spectrometry single m/z is isolated, fragmented,and daughter ions are scanned

    Triple-Quad collision-induced dissociation (CID) in 2nd quad

    Ion trap isolate parent ion in trap, CID, scan daughter ions

    FTICR excite unwanted ions from trap, collide ions w/ surface, recordFID of fragments

    Q-TOF select precursor in quad, CID, send fragments into TOF

    TOF-TOF ion gate selects precursor from TOF1, CID or PID to makefragments

    IT-FTICR/IT-Orbitrap select precursor in ion trap, fragment, analyzefragments in FTICR/Orbitrap OR mass analyze precursors inFTICR/Orbitrap while isolating/fragmenting/mass analyzing fragments inIT (usually LIT)