A 318 Lecture 2324 Ms Analyzers Hand Out
-
Upload
muhamad-ivan-abror -
Category
Documents
-
view
218 -
download
0
Transcript of A 318 Lecture 2324 Ms Analyzers Hand Out
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
1/10
Sample Exam Questions1. _____ Ions formed by ___ obtain multiple charges via proton transfer.
A. EI
B. FAB
C. ESI
D. MALDI
2. _____ Which source should be use to produce singly charged parent ionsfor myoglobin (16591 Da)?
A. CI
B. EI
C. ESI
D. MALDI
Mass AnalyzersContinuous (Filters)Quadrupole - Quad
Magnetic Sector - MS
Electrostatic Analyzer - ESA
Pulsed
Time-of-Flight (Reflectron) - TOF
Ion Trap (quadrupole or linear) IT (QIT or LIT)
Fourier Transform Ion Cyclotron Resonance FTICR
Orbitrap
Tandem
Triple-Quad, Ion trap, FTICR, Q-TOF, TOF-TOF, IT-FTICR, IT-Orbitrap
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
2/10
Mass Resolution / Mass Range
Instrument
11C,20C
Resolution ability of mass analyzer to separate ions w/ similar m/z values
Peaks are considered resolved if height of valley between them is less than10% of peak height
Resolution, R = m/m R = 27.5/1 = 27.5 to resolveAl+ from Si+
Al+ Si+
27 28
R = ~50
Al+ Si+
27 28
R = ~100
Al+ Si+
27 28
R = ~500
Nominal mass nearest whole number mass (28 for CO, 44 for CO2
)
Mass Range from smallest to largest m/z separated by mass analyzer
Quadrupole Mass Analyzer
Instrument
11B
Most commonly used bench-top mass analyzer inexpensive, rugged,compact, high scan rates (100 ms/spectrum)
+
+- -
four parallel cylindrical rods
opposite rodshave samepotential
stabletrajectory
unstable trajectory
One set of rods acts as low-pass filter, the other as high-pass filter
m/z
T
m/z
T
m/z
Tcombined
(Figures 11-4, 11-5, 11-6, 11-7)
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
3/10
Various Ion Guides Multi-poles
Figure 5. Radial pseudopotential contour in volts forthe multipoles and tripole cross-section. All graphsshow that the minimum point is in the center of eachmultipole and increases when it moves to the
electrodes. Calculation conditions: V0 = 10 V, = 1MHz, r0 = 3 mm, 1000 Da, ion charge = +1. (a)
quadrupole; (b) hexapole; (c) octapole; (d) tripole.
Figure 6. Ion guides AC voltage for stable ionguiding and tripole stable trajectory. 1 MHz,length 150 mm, r0 = 3 mm. Beam initialconditions: kinetic energy 10 eV, expansion
angle 3, and initial radius = 0.5 r0. (a) AC-mass stability diagrams; (b) cross-section view;(c) transversal view of the ion stable movementinside of the tripole rotational electric field.
Computer Simulations of a New Three Rods Ion Optic (TRIPOLE) with High Focusing and Mass Filtering Capabilities GaryAbdiel Salazar, Tsutomu Masujima J. American Society for Mass Spectrometry Vol. 18, Iss. 3 , March 2007, 413-421.
Einzel Lens Ion beam focusing
This shows the path of ions as they travel through a simulated Einzel lens. The ion mass is 200kDaaccelerated to 40kV, the middle plate potential is at 200V and the plate size is 1m these are verylarge settings used to show the focusing effect more easily and are not practical settings.
plates
OR
rings
Divergent ion beams arecollimated / focused afterpassing through 3-voltageelement
http://en.wikipedia.org/wiki/Einzel_lens
plates+V volts0 volts
0 volts 0 volts
0 volts
+V volts
ion path focus
Electric field lines
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
4/10
Magnetic Sector / Electrostatic Analyzer
Instrument
11B, 20C
Double-focusing mass spectrometer
Corrects for energy (in ESA) and directional (in MS) distributions of ions capable of very high (105) resolution - Single-focusing (MS only)instruments have lower (~2000) resolution
In MS: m/z = B2r2e/2V, scan by varying B, V, or r
Several arrangements for double-focusing instruments 2 common onesare Mattauch-Herzog (Fig. 11-9) with plane of double-focus & Nier-Johnson (Fig. 20-13) with point of double-focus
ESA
MS
focalplane
m1/zm2/z
3150
Mattauch-Herzog Nier-Johnson
point ofdoublefocus
MS
energy
focal planedirection
focal plane
ESA
ESA slit
+
+
-
-
Time-of-Flight Mass Analyzer
zm
eV
dt2
=
Instrument
11B, 20C
Brief pulses (few ns) of ions produced at 10 Hz to 50 kHz and acceleratedby 1 to 20 kV into ~1-m field-free drift region where ions separate asinverse of m/z lighter ions arrive at detector prior to heavier ions flighttimes of 1 to 30 s requires very fast (ns bins) electronics
0 V10 kV
large
electricfield
signal
detector1 m
field-free region
Simple, rugged, simultaneous detection, virtually unlimited mass range,rapid data acquisition (
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
5/10
Delayed Extraction TOF MS
Instrument
notes
Ions formed in source under field-free conditions (usually desorption from
surface), allowed to drift due to inherent velocity after brief delay, electricfield is applied, pushing ions toward detector such that ions w/ smallerinitial velocities receive more acceleration, reaching detector at same timeas ions of same m/z with larger initial velocities
initial desorption(0 s)
5 kV
5 kV
0 kV
delayed extraction(0.1 to 5 s)
0 kV
5 kV
10 kV
Length of delay and size of pulse determine the m/z that isoptimally focused
Reflectron Time-of-FlightMass Analyzer
amp comp
0 kV
L+M+H+
+20 kV
+22 kV
0 kV 0 kV
0 kV
1) Ions enter source region,accelerated toward reflectron.
2) Ions separate in space based ontheir relative mass-to-charge (m/z).
3) Ions reverse path in reflectron.
4) Ions impact detector.
Flight time
Signal
Instrument
notes
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
6/10
Ion Trap Mass Analyzer
Instrument
20C
Quadrupole (3-D) ion trap consists of a donut-shaped ring electrode & two
end cap electrodes (Fig. 20-15) ions enter through one end cap, arestabilized by radio-frequency voltage applied to ring electrode while endcaps are at ground ions of individual m/z are sent through other end capto detector during scan rugged, compact, inexpensive
todetector
fromsource
endcaps
ringelectrode
Linear (2-D) ion trap resembles quadrupole mass analyzer
3D ion trap
http://aemc.jpl.nasa.gov/activities/mms.cfm
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
7/10
2D (linear) Ion Trap MS
Simion simulation of the resonanceexcitation field comparing (a) a threesection device and (b) a single section
device with end plates for axial trapping.
(a) Simion simulation of 100 eV ions
exiting the 2-D trap along the length ofthe exit slot and being focussed ontothe 15kV conversion dynode. (b) 10keV ions simulated.
A two-dimensional quadrupole ion trap massspectrometer Jae C. Schwartz, Michael W. Senko,and John E. P. Syka J. American Society for Mass
Spectrometry Vol. 13, Iss. 6 , June 2002, 659-669.
Linear Ion Trap Instrument
The overall instrument configuration along with typical operating voltages and pressures.
A two-dimensional quadrupole ion trap mass spectrometer Jae C. Schwartz, Michael W. Senko, and
John E. P. Syka J. American Society for Mass Spectrometry Vol. 13, Iss. 6 , June 2002, 659-669.
ion transmissionmass
analysis
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
8/10
FT Ion Cyclotron Resonance
Instrument
20C
Ions trapped in cubic cell inside bore of (usually superconducting) magnet
excited from center by brief RF pulse into circular path in planeperpendicular to magnetic field, B (Fig. 20-16, 20-17)
transmitterplate
receiverplate
receiverplate
B
Electronsdrawn throughcircuit as ionsapproach /recede fromreceiver plates
Image current measured through circuit depends on the Free-InductionDecay (FID) of ions in their characteristic cyclotron frequency,c
zeB
mc=
FTICR Signal Processing
Instrument
20C
Excitation RF
FID signal
time
Frequency (kHz)
zeB
mc=
m/z (amu)
FT
Capable of very high (>106) resolution,improves with longer FID signal; veryexpensive (>$400k) due tosuperconducting magnet (>3 T)
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
9/10
Orbitrap MS
http://www.thermo.com/eThermo/CMA/PDFs/Various/File_31081.pdf
http://www.thermo.com/eThermo/CMA/PDFs
/Articles/articlesFile_30850.pdf
High Resolving Power example
http://www.thermo.com/eThermo/CMA/PDFs/Articles/articlesFile_30850.pdf
Phosphatidylethanolamines
PE 18:0/18:0 and
PE O-16:1p/22:6
(isobaric)
Larger resolvingpower requires moretime per scan.
-
7/27/2019 A 318 Lecture 2324 Ms Analyzers Hand Out
10/10
Mass Analyzer Figures of Merit
Instrument
20C
Resolution m/z range scan time
Quadrupole Quad ~2000 < 2000 100 ms
Magnetic Sector ~2500 < 2000 ~1 s
Double-focusing (ESA/MS) 50,000 < 2000 ~1 s
Time-of-Flight ~4000 unlimited 100 s
Reflectron TOF 20,000 unlimited 100 s
Ion Trap (3D) ~2000 < 2000 100 ms
Linear (2D) Ion Trap ~2000 < 2000 20 ms
FTICR > 200,000 50,000+ 0.1-10 s
Orbitrap > 50,000 50,000+ 0.1-10 s
Tandem Mass Analyzers
Instrument
notes
Perform tandem mass spectrometry single m/z is isolated, fragmented,and daughter ions are scanned
Triple-Quad collision-induced dissociation (CID) in 2nd quad
Ion trap isolate parent ion in trap, CID, scan daughter ions
FTICR excite unwanted ions from trap, collide ions w/ surface, recordFID of fragments
Q-TOF select precursor in quad, CID, send fragments into TOF
TOF-TOF ion gate selects precursor from TOF1, CID or PID to makefragments
IT-FTICR/IT-Orbitrap select precursor in ion trap, fragment, analyzefragments in FTICR/Orbitrap OR mass analyze precursors inFTICR/Orbitrap while isolating/fragmenting/mass analyzing fragments inIT (usually LIT)