98_Pb_IRs_JAAS
Transcript of 98_Pb_IRs_JAAS
7/26/2019 98_Pb_IRs_JAAS
http://slidepdf.com/reader/full/98pbirsjaas 1/5
INTER-LABORATORY NOTE
Precise lead isotope ratios in Australian galena
samples by high resolution inductively coupled
plasma mass spectrometry
Ashley T. Townsend*a, Zongshou Yub, Peter McGoldrickb and James A. Huttona
aCentral Science L aboratory, University of T asmania, GPO Box 252-74, Hobart, T asmania, 7001, Australia
bCentre for Ore Deposits and Exploration Studies, University of T asmania, GPO Box 252-79, Hobart, T asmania, 7001, Australia
High resolution inductively coupled plasma mass spectrometry After careful consideration of all possible causes of instrumental
bias, the precision obtained from the analysis of NIST Standard(HR-ICP-MS) was used to measure Pb isotope ratios in
standard solutions and Pb mineral digests. The RSD values Reference Material (SRM ) 981 Natural Lead ranged from 0.04to 0.12%, depending on the ratio considered. The accuracyobtained for 208Pb/204Pb, 207Pb/204Pb, 206Pb/204Pb, 208Pb/206Pb
and 207Pb/206Pb were 0.13, 0.11, 0.11, 0.046 and 0.048%, and precision reported may be the best obtained from a
quadrupole instrument. However, the methodology and detailrespectively (values as 1s). These values were obtained from30 analyses of three diff erent standard sample types (multi- required to obtain such results could not be called routine.
High resolution ICP-MS (HR-ICP-MS) is a relatively newelement standard, NIST SRM 981 and a Broken Hill galena
digest). Based on 39 analyses of 11 galena samples from technique employing a magnetic sector mass analyser. To dateit has been applied in a variety of areas such as environmentaldiff erent locations around Australia, the diff erence between
HR-ICP-MS and conventional TIMS values for 208Pb/204Pb, and biological analysis and the monitoring of radionuclides(see the review by Becker and Dietze9). Only a limited number207Pb/204Pb, 206Pb/204Pb, 208Pb/206Pb and 207Pb/206Pb ratios
was generally better than±0.2%. This paper outlines a very of studies have been reported on measuring isotope ratios.10–14
With regard to Pb isotope ratios, Vanhaecke et al.10 obtainedsimple and rapid analytical method for the measurement of Pb
isotope ratios, and is one of the first studies to use a precision of 0.04% (n=10) for the 206Pb/ 207Pb ratio in astandard lead solution. Little has been reported about the PbHR-ICP-MS to measure Pb isotope ratios in galena and
galena-bearing ores. isotope ratio analysis of geological matrices.14 It should also
be noted, however, that double focusing multiple collectorKeywords: L ead isotope ratios; high resolution inductively
ICP-MS instruments have been used recently to measurecoupled plasma mass spectrometry; galena; ores
highly accurate isotope ratios (including Pb), and may seeincreased use in future work.15,16
In this paper, a fast and reliable method for the preciseThe study of the isotopic composition of Pb in rocks and
measurement of Pb isotope ratios in geological samples byminerals began over 80 years ago and was initially used as a
HR-ICP-MS is reported. The technique was applied to thegeochronological tool.1 Subsequent work showed that for
measurement of the Pb isotope composition of galena andmany types of geological materials, Pb isotope geochronology
galena-bearing ores from Broken Hill and Western Tasmania,was complicated by the presence of ‘common’ Pb ( i.e., non-
previously measured by TIMS.radiogenic Pb).2 Hence today, Pb isotope geochronology usesminerals that formed with very little Pb. 3 However, a second
EXPERIMENTALimportant practical application of Pb isotopes is in studyingand exploring for mineral deposits.4 The Pb isotope composi-
Reagentstion of rocks and Pb-bearing minerals can be used to trace
High purity HCl was prepared by doubly quartz distillingthe sources of metal in ores and to help discriminate barrenanalytical-reagent grade acid (Merck, Darmstadt, Germany),from mineralised ore systems.
and high purity HF was prepared from analytical-reagentAt present, thermal ionisation mass spectrometry (TIMS) isgrade acid (Merck) that had been further purified via a Teflonthe technique of choice for all types of geological Pb isotopedistillation system. High purity HNO
3 (Mallinckrodt, St. Louis,measurements. However, the relatively high cost of TIMS
MO, USA) was used as received for sample digestion andinstrumentation and the (necessary) extensive chemical pre-solution acidification. Ultra-pure de-ionised water (≥18 MV)treatment of the sample prior to analysis have imposedwas prepared via a glass distillation unit followed by furtherserious limitations on the routine use of TIMS techniques inpurification with a Modulad water purification systemgeochemical exploration applications.(Continental Water System, Melbourne, Australia).Quadrupole based ICP-MS instruments have been used in
many studies to measure Pb isotope ratios.5–8 However, owingto the design and manner in which quadrupole mass analysers
Standards and sample preparationoperate, the technique has been found to be limited in both
precision and accuracy. The precision of isotope ratios meas- Instrument tuning and general Pb solutions were prepared bydilution from a Perkin-Elmer (Norwalk, CT, USA) 10 mg g−1ured on quadrupole ICP-MS instruments is typically 0.1%
RSD, whereas for ratios involving a low abundance isotope multi-element solution containing Pb (Cat. No. N930–0233).
A stock standard solution of NIST SRM 981 Natural Leadsuch as 2 04Pb, a precision of 0.2–1% RSD is usually obtained.8Recently, Begley and Sharp8 undertook an elegant study of was prepared by dissolving 0.05 g of Pb in 5 ml of dilute
HNO3
(1+4) with heating in a Savillex screw-topped TeflonPb isotope ratios using a quadrupole ICP-MS instrument.
Journal of Analytical Atomic Spectrometry, August 1998, Vol. 13 (809–813) 809
7/26/2019 98_Pb_IRs_JAAS
http://slidepdf.com/reader/full/98pbirsjaas 2/5
beaker.17 After cooling to room temperature, 1 ml of concen-Table 2 Method parameters for Pb isotope ratio measurements
trated HNO3
was added before dilution to 100 ml total volume(in polycarbonate). The stock standard solution was diluted Isotopes considered 201Hg or 202Hg, 204Pb, 206Pb, 207Pbto working concentrations (generally 10–50 ng g−1 ) before and 208Pb
Scan type Magnet fixed at 201Hg. Electric scansanalysis. This reference solution was used as the externalover other mass rangescalibration standard in this study.
Mass scanning window 3%Galena sample solutions were prepared as follows. SamplesMagnet settle time 0.1 s
were dried at 105 °C overnight and 0.1 g portions were addedDwell time per measurement 0.001 s
to Savillex screw-topped Teflon beakers. A 5 ml volume of Scan duration per sweep 5×0.025 s
HF-HNO3 –HCl (2+1+3) acid mixture was added to each Number of sweeps 2000Measurement time 250 ssample18 with overnight heating on a hot-plate at 130 °C. The
per replicateresulting solution was twice evaporated to incipient dryness,Detector mode Countingwith the addition of 1 ml of HNO
3 after each drying stage. All
Integration type Averagesolutions were stored in sealed polycarbonate containers.
Sample uptake and 2 minGalena solutions were diluted to a Pb concentration between equilibration time40 and 50 ng g−1 for HR-ICP-MS analysis. Rinse time bet ween 3 –5 min (with 5% HNO
3)
each sampleTotal time per sample ~15 min
Instrumentation Measured dead time In the range 20–50 ns, depending ondetector settings
Pb isotope ratios were measured on an Element HR-ICP-MSinstrument (Finnigan MAT, Bremen, Germany). This instru-
ment, utilising a magnetic sector mass analyser of reversedRES ULTS AND DIS CUS SIONNier–Johnson geometry, has pre-defined resolution settings
(m/Dm at 10% valley definition) of 300 (low), 3000 (medium) Mass discrimination correctionand 7500 (high). A resolution of 300 was used in this work,
Isotope ratio values obtained using ICP-MS may deviate fromproviding flat-topped peaks and maximum instrument sensi-the ‘true’ value as a function of the diff erence in mass betweentivity (typically greater than 106 counts s−1 for 10 ng g−1 115Inthe two isotopes measured. This eff ect is defined as mass biasor 7×105 counts s−1 for 10 ng g−1 2 08Pb). Instrument andand its value may be either positive or negative.23 The bias ormethod settings are outlined in Tables 1 and 2. A standarddiscrimination can only be determined experimentally and willMeinhard nebuliser and Scott double pass water cooled sprayvary with the configuration of the individual ICP-MS system. 22chamber were employed. Isotopes of interest were analysedIn this study, instrument mass discrimination was correctedusing electric scanning with the magnet held at fixed mass.by the use of a NIST SRM 981 external standard. UnknownThe secondary electron multiplier detector (with discretesamples to be measured were typically analysed betweendynodes) was operated in the counting mode. Instrumentalternate standard reference solutions. The use of this correc-tuning and optimisation were performed daily using a 10 ngtion method allows any measured diff erences in a Pb standardg−1 multi-element solution containing Pb. Further detailssample to correct the ratios found in unknown Pb solutionsconcerning this instrument have been reported elsewhere.19–21
(i.e., Pb standards correcting for Pb unknowns). Also, similarAll isotope ratio values were adjusted for instrument dead isotope abundances are typically measured for both standardstime.22 Each Pb isotope ratio was considered in dead time
and unknowns. However, on the negative side, there is a timedeterminations. The dead time was measured regularlydelay between each standard and sample in which changes inthroughout the study using three Pb solutions of diff erentplasma conditions can occur. The other option of using a Tlconcentrations, namely 10, 25 and 50 ng g−1 Pb. Dead timeinternal standard for mass discrimination correction was notvalues in the range 20–50 ns were typically found, dependinginvestigated in this study. This method involves measuringon the detector settings. Analyses were also blank corrected.another isotope (Tl) which may behave in a diff erent fashionto Pb whilst also increasing the analysis time for each scan(time which could be spent accumulating Pb data). The internalcorrection approach has an advantage, however, of continu-
Table 1 Typical instrumental configuration and settingsously monitoring the mass discrimination for each individualsample. Both methods have been used in previous Pb isotopeInstrument Finnigan MAT Element
Resolution (m/Dm) 300 ( low) studies6–8,11 and were outlined in detail by Begley and Sharp.8Rf power 1250 W
Nebuliser Meinhard concentricGeneral analytical and methodology considerationsCoolant argon flow rate ~12.5 l min−1*
Auxiliary argon flow rate ~1.1 l min−1*Method parameters were systematically varied and tested inNebuliser argon flow rate ~1.1 l min−1*order to obtain the most accurate and precise Pb isotopeSpray chamber Scott (double pass) type cooled
to 3.5–5 °C ratios. The first variable considered was the number of scansSampler cone Nickel, 1.1 mm aperture id for each sample. The ideal situation would be to acquire asSkimmer cone Nickel, 0.8 mm aperture id many scans as possible in the shortest time interval, henceInstrument tuning Performed using a 10 ng g−1
approaching the behaviour of a multiple collector system. Themulti-element solution
number of sweeps across the Pb isotopes was varied betweenIon transmission Typically 106 counts s−1 per 10 ng g−1500 ( total analysis time 62.5 s) and 3000 (375 s), with 2000115In or 7×105 counts s−1 for
10 ng g−1 208Pb scans (250 s) proving optimum considering Pb signal stabilities.Scan type Fixed magnet with electric scan over All scanning protocols were at a rate of 8 sweeps s−1, keeping
small mass range the dwell time for each mass segment constant. VanhaeckeIon sampling depth Adjusted daily*
et al.10 used 1200 scans in 2 min to measure 206Pb/ 207PbIon lens settings Adjusted daily†
isotope ratios in a standard Pb solution.* Adjusted in order to obtain maximum signal intensity. † AdjustedVanhaecke et al.11 also found that when considering both
in order to obtain maximum signal intensity and optimum resolution.Pb or Cu isotope ratios at resolution 3000 using HR-ICP-MS,
810 Journal of Analytical Atomic Spectrometry, August 1998, Vol. 13
7/26/2019 98_Pb_IRs_JAAS
http://slidepdf.com/reader/full/98pbirsjaas 3/5
concentration range 10–50 ng g−1. Work is continuing to
increase the concentration range for which accurate Pb ratioscan be obtained.
All ratio measurements were dead time corrected. Even withthis adjustment, it was found that the accuracy of eachmeasurement could be further improved when the Pb concen-
tration in the NIST SRM 981 external calibration standardand the unknown sample were closely matched. As such theconcentration of Pb in each sample was often adjusted prior
to ratio determinations.Along with the four Pb isotopes of interest, 202Hg and 201Hg
were also measured as 204Pb (1.4% abundant) has an isobaricinterference from 204Hg (6.85% abundant) in resolution 300.For geological samples that were previously ground in tungsten
Fig. 1 Measured 208Pb/ 204Pb ratios versus Pb concentration in crushing mills, 201Hg was used to correct 204Pb, as 202Hgstandard solutions. Ratios were measured over two consecutive days. suff ers interference by 186W16O.7 For other samples 202Hg wasNote that the values shown have not been corrected for any mass used. It was expected that during sample digestion any volatilediscrimination.
Hg would be driven off during the multiple drying stages,while the Hg5Pb ratio in rocks and sulfides is generally low(1510–100).24 No samples analysed in this study were foundthe precision was improved when the mass scanning windowto have appreciable levels of Hg or W. However, it waswas reduced from 20 to 10% (the percentage window value isimportant to include these isotopes in method developmentbased on the expected mass width of a peak in the respectivework for future samples when more appreciable Hg or Wresolution). It was suggested that when smaller mass windows
concentrations may be encountered.were used, only the centre of each peak was scanned, minimis-ing the influence of any mass calibration instabilities. Althoughmore crucial at resolution 3000 away from the flat-toppedpeaks of resolution 300, a similar systematic study was under-
Precisiontaken in this work, considering Pb ratios at low resolution.Mass windows between 1 and 10% were considered, and it Lead isotope ratio precision data for 10 consecutive measure-
ments of three diff erent standard solutions are given in Table 3.was found that a mass scanning window of 3% gave isotopesignals with minimum variation over the total number of scans. Precisions obtained for ratios involving the low abundance
204Pb isotope were 0.10–0.13% for 208Pb/ 204Pb, 0.10–0.11%The working Pb concentration range for which accurate Pb
isotope ratios could be obtained was also investigated. At Pb for 207Pb/ 204Pb and 0.10–0.11% for 2 06Pb/ 204Pb (values as 1s).Improved precisions were found for those ratios involving theconcentrations below 10 ng g−1, ratios with 204Pb as the basis
were found to be limited by counting statistics, resulting in more abundant isotopes, namely 0.041–0.046% for 20 8Pb/ 206Pband 0.044–0.048% for 207Pb/ 206Pb. No diff erence in measure-less accurate results (a concentration of 10 ng g−1 corresponded
to approximately 15 000 counts s−1 of 204Pb). This is shown ment precision was apparent between standard Pb solutions
and more ‘complex’ galena digest samples. Theoreticalmore clearly in Fig. 1 for the 208
Pb/ 204
Pb ratio measured inPb standard solutions of various concentrations analysed over precisions based on Poisson counting statistics were alsocalculated and are shown in Table 3 (based on 208Pb signalstwo consecutive days. Similar results were found for the other
Pb ratios considered. This has not been noted as a problem of the order of 1×106 –3×106 counts s−1 ). In agreement withother workers,13 external precisions approximately 2–3 timesin other HR-ICP-MS studies as the ratios investigated have
typically involved 206Pb, 207Pb and 208Pb isotopes only.10,14 greater than that predicted from counting statistics alone werefound, suggesting noise contributions from other sources suchAt the other extreme, Pb concentrations above 50 ng g−1 were
not considered as the signal from 208Pb required measurement as the sample introduction system and plasma flicker. Theprecisions obtained for ratios referenced to 206Pb are similarin the analogue detector mode to avoid detector saturation
(under normal operating conditions and detector settings, to that found by Vanhaecke et al.10 using a similar HR-ICP-MS
instrument to measure the 206Pb/ 207Pb ratio of a standard Pb50 ng g−1 of 208Pb corresponded to 3×106 –3.5×106 countss−1). Rather than limit the accuracy of the ratio determinations solution. The external precisions given in Table 3 are also
similar to the best obtained using a quadrupole ICP-MS byit was decided to make all measurements in the same detectormode (counting). Hence ratios were only determined in the Pb Begley and Sharp.8
Table 3 Pb isotope ratio precision and accuracy data for a standard multi-element solution, NIST SRM 981 and a Broken Hill galena digest.Samples were analysed at a Pb concentration of 45 ng g−1
208Pb/ 204Pb 207Pb/ 204Pb 206Pb/ 204Pb 208Pb/ 206Pb 207Pb/ 206Pb
Multi-element standard solution— Precision for 10 measurements, RSD (%) 0.11 0.11 0.10 0.041 0.047Theoretical RSD (%)* 0.054 0.055 0.055 0.015 0.016
NIST SRM 981 Natural L ead— Precision for 10 measurements, RSD (%) 0.10 0.11 0.11 0.046 0.048Theoretical RSD (%)* 0.049 0.050 0.050 0.014 0.015
Broken Hill galena— Measured ratio 35.520 15.353 15.985 2.222 0.961Reference value† 35.599 15.370 15.994 2.226 0.961ICP-MS bias from TIMS value (%) −0.22 −0.11 −0.06 −0.18 0Precision for 10 measurements, RSD (%) 0.13 0.10 0.11 0.043 0.044Theoretical RSD (%)* 0.052 0.053 0.053 0.015 0.016
* From Poisson counting statistics. † From TIMS analysis.
Journal of Analytical Atomic Spectrometry, August 1998, Vol. 13 811
7/26/2019 98_Pb_IRs_JAAS
http://slidepdf.com/reader/full/98pbirsjaas 4/5
T a b l e 4
M e a s u r e d P b i s o t o p e r a t i o s i n s e l e c t e d A u s t r a l i a n g a l e n a s a m p l e s . R a t i o s f r o m H R - I C P - M S a n d T I M S a r e c o m p a r e d
2 0 8 / 2 0 4
2 0 7 / 2 0 4
2 0 6 / 2 0 4
2 0 8 / 2 0 6
2 0 7 / 2 0 6
S a m
p l e
D e p o s i t
D †
D †
D †
D †
D †
c o
d e
l o c a t i o n
n *
I C P - M S
T I M S
( % )
I C P - M S
T I M S
( % )
I C P - M S
T I M S
( % )
I C P - M S
T I M S
( % )
I C P
- M S
T I M S
( % )
Z 5 1 9 9
O c e a n a
2
3 8 . 2
2 2
3 8 . 1
2 1
+ 0 . 2
6
1 5 . 6
0 3
1 5 . 6
0 7
− 0 . 0
3
1 8 . 3
1 5
1 8 . 2
8 1
+ 0 . 1
9
2 . 0
8 6 8
2 . 0
8 5 3
+ 0 . 0
7
0 . 8
5 1 9
0 . 8
5 3 7
− 0 . 2
1
Z 5 2 2 1
A r g e n t 2
3
3 8 . 4
0 2
3 8 . 4
1 7
− 0 . 0
4
1 5 . 6
5 1
1 5 . 6
2 2
+ 0 . 1
9
1 8 . 4
9 2
1 8 . 5
3 1
− 0 . 2
1
2 . 0
7 5 3
2 . 0
7 3 1
+ 0 . 1
1
0 . 8
4 5 8
0 . 8
4 3 0
+ 0 . 3
3
Z 5 2 3 2
S w a n s e a
3
3 8 . 2
4 0
3 8 . 2
4 1
− 0 . 0
0 3
1 5 . 6
4 8
1 5 . 6
3 3
+ 0 . 1
0
1 8 . 3
3 1
1 8 . 3
6 8
− 0 . 2
0
2 . 0
8 9 4
2 . 0
8 1 9
+ 0 . 3
6
0 . 8
5 5 0
0 . 8
5 1 1
+ 0 . 4
6
Z 5 2 3 7
M o n t a n a
3
3 8 . 4
7 4
3 8 . 4
7 2
+ 0 . 0
0 5
1 5 . 6
4 4
1 5 . 6
2 5
+ 0 . 1
2
1 8 . 5
5 0
1 8 . 5
6 1
− 0 . 0
6
2 . 0
7 5 3
2 . 0
7 2 7
+ 0 . 1
3
0 . 8
4 3 9
0 . 8
4 1 8
+ 0 . 2
4
Z 5 2 7 7
N .
T a s . M i n e
3
3 8 . 2
8 8
3 8 . 2
9 6
− 0 . 0
2
1 5 . 6
2 0
1 5 . 6
2 4
− 0 . 0
3
1 8 . 4
2 3
1 8 . 4
2 8
− 0 . 0
3
2 . 0
8 4 1
2 . 0
7 8 1
+ 0 . 2
9
0 . 8
5 0 2
0 . 8
4 7 8
+ 0 . 2
8
H 1 5
H e l l y e r
2
3 8 . 2
4 1
3 8 . 1
7 8
+ 0 . 1
7
1 5 . 6
3 4
1 5 . 6
1 2
+ 0 . 1
4
1 8 . 3
8 0
1 8 . 3
6 1
+ 0 . 1
0
2 . 0
7 8 1
2 . 0
7 9 3
− 0 . 0
6
0 . 8
5 0 2
0 . 8
5 0 3
− 0 . 0
1
E B 1
0 0 7
E l l i o t B a y
2
3 7 . 8
6 8
3 7 . 9
1
− 0 . 1
1
1 5 . 5
5 1
1 5 . 5
8
− 0 . 1
9
1 8 . 0
9 2
1 8 . 1
1 9
− 0 . 1
5
2 . 0
8 7 7
2 . 0
9 2 3
− 0 . 2
2
0 . 8
6 0 9
0 . 8
5 9 9
+ 0 . 1
2
E B 7
2 0 8 6
E l l i o t B a y
2
3 7 . 8
8 7
3 7 . 9
1
− 0 . 0
6
1 5 . 5
8 3
1 5 . 5
8
+ 0 . 0
2
1 8 . 0
6 3
1 8 . 0
7 5
− 0 . 0
7
2 . 0
9 2 9
2 . 0
9 7 4
− 0 . 2
1
0 . 8
6 2 7
0 . 8
6 2 0
+ 0 . 0
8
E B 7
2 0 8 7
E l l i o t B a y
2
3 7 . 9
0 0
3 7 . 8
7 6
+ 0 . 0
6
1 5 . 5
8 7
1 5 . 5
7
+ 0 . 1
1
1 8 . 0
9 5
1 8 . 0
9 8
− 0 . 0
2
2 . 0
9 4 6
2 . 0
9 2 8
+ 0 . 0
9
0 . 8
6 1 3
0 . 8
6 0 3
+ 0 . 1
2
E B 7
2 0 9 6
E l l i o t B a y
2
3 8 . 0
8 3
3 8 . 0
4 1
+ 0 . 1
1
1 5 . 6
2 8
1 5 . 6
0 5
+ 0 . 1
5
1 8 . 1
9 8
1 8 . 1
9 1
+ 0 . 0
4
2 . 0
9 5 3
2 . 0
9 1 2
+ 0 . 2
0
0 . 8
5 8 8
0 . 8
5 7 8
+ 0 . 1
2
B H 2
B r o k e n H i l l
1 5
3 5 . 5
3 0
3 5 . 5
9 9
− 0 . 1
9
1 5 . 3
8 4
1 5 . 3
7
+ 0 . 0
9
1 6 . 0
0 4
1 5 . 9
9 4
+ 0 . 0
6
2 . 2
2 0 8
2 . 2
2 5 8
− 0 . 2
3
0 . 9
6 1 3
0 . 9
6 1 0
+ 0 . 0
3
*
n = N u m b e r o f a n a l y s e s p e r f o r m e d o n e a c h
s a m p l e ( a v e r a g e v a l u e s s h o w n ) .
† D = D i ff e r e n c e b e t w e e n I C P - M S a n d T I M S v a l u e s .
812 Journal of Analytical Atomic Spectrometry, August 1998, Vol. 13
7/26/2019 98_Pb_IRs_JAAS
http://slidepdf.com/reader/full/98pbirsjaas 5/5
measured in Pb mineral concentrates using HR-ICP-MS.
Based on the analysis of a series of standard reference solutions,this comparatively fast analytical method provided Pb ratioswith precisions of 0.13% RSD or better for ratios to 204Pb.For ratios to 206Pb, the precisions obtained were of the orderof 0.045%. TIMS and HR-ICP-MS were found to be in
agreement to an accuracy of ±0.2%, irrespective of the ratioconsidered. This level of accuracy is adequate for many geoch-emical applications. The relative ease and speed with which
these values were obtained suggest that HR-ICP-MS may seeincreased use for geochemical Pb isotope ratio measurements.
The authors acknowledge the facilities and support providedFig. 2 Comparison between average 206Pb/ 204Pb ratio values by the Central Science Laboratory and Centre for Ore Depositsobtained via HR-ICP-MS and TIMS techniques (for purposes of scale
and Exploration Studies ( CODES), University of Tasmania.the Broken Hill galena sample has been omitted).
Ross Large, Paul Kitto, Bruce Gemmel and David Steele are
thanked for their support, samples and related information.Comparison with TIMS values and accuracy of the analytical
method
REFERENCESIn order to study the accuracy of the present method, the Pbisotope ratios of 11 galena and galena-rich ore samples were 1 Holmes, A., T he Age of the Earth, Harper, New York, 1913, p. 194.
2 Faure, G., Principles of Isotope Geology, Wiley, New York,measured by TIMS at CSIRO North Ryde and by HR-ICP-MS1977, p. 464.at the University of Tasmania. Samples were typically analysed
3 Schaltegger, U., Schneider, J. L., Maurin, J. C., and Corfu, F.,
2–3 times by HR-ICP-MS, with the Pb concentration of the Earth Planet. Sci. L ett., 1996, 144 , 403.NIST SRM 981 external calibration standard and the unknown 4 Gulson, B. L., L ead Isotopes in Mineral Exploration, Elsevier,Pb sample being closely matched. Comparative data are given Amsterdam, 1986, p. 245.in Table 4. There is very good agreement between the two 5 Smith, R. G., Brooker, E. J., Douglas, D. J., Quan, E. S. K., and
Rosenblatt, G., J. Geochem. Explor., 1984, 21 , 385.techniques and this is highlighted in Fig. 2, showing6 Date, A. R., and Cheung, Y. Y., Analyst, 1987, 112 , 1531.HR-ICP-MS values for the 206Pb/ 204Pb ratio plotted against7 Longerich, H. P., Wilton, D. H. C., and Fryer, B. J., J. Geochem.
TIMS data (206Pb/ 204PbHR-ICP-MS
=0.98×206Pb/ 204PbTIMS+
Explor., 1992, 43 , 111.0.45, r2=0.983 for n=10). The ratio precision from 8 Begley, I. S., and Sharp, B. L., J. Anal. At. Spectrom., 1997, 12, 395.HR-ICP-MS was about 0.05% and that from TIMS was about 9 Becker, J. S., and Dietze, H.-J., J. Anal. At. Spectrom., 1997, 12, 881.0.02–0.05%.15 A similar figure was presented in work by Date 10 Vanhaecke, F., Moens, L., and Dams, R., Anal. Chem., 1996,
68, 567.and Cheung6 using an early model quadrupole ICP-MS. The11 Vanhaecke, F., Moens, L., Dams, R., Papadakis, I., and Taylor, P.,diff erence in Pb ratio values between the two analytical tech-
Anal. Chem., 1997, 69 , 268.niques was generally less than ±0.2%, and was in the range
12 Cornec, F. L., and Correge, T., J. Anal. At. Spectrom., 1997, 12, 969.0.001–0.1 for 208Pb/ 204Pb, 0.004–0.03 for 207Pb/ 204Pb, 13 Sturup, S., Hansen, M., and Mølgaard, C., J. Anal. At. Spectrom.,0.003–0.04 for 206Pb/ 204Pb, 0.001–0.008 for 208Pb/ 206Pb and
1997, 12 , 919.0.0001–0.004 for 207Pb/ 206Pb. 14 Shuttleworth, S., Morris, J., Crombie, K., Ridings, R., andArvidson, R., in Proceedings of the 3rd International Conferenceon the Analysis of Geological and Environmental Materials,
Geological implications Colorado, USA, June 1997, p. 100.15 Walder, A. J., Platzner, I., and Freedman, P. A., J. Anal. At.
Lead isotopes are powerful tools in exploration for mineral Spectrom., 1993, 8 , 19.deposits and the investigation of problems of ore genesis.4 16 Walder, A. J., and Furuta, N., Anal. Sci., 1993, 9 , 675.
17 Potts, P. J. , A Handbook of Silicate Rock Analysis, Blackie,However, unlike multi-element geochemical analyses, they areGlasgow, 1987, p. 622.not commonly used when exploring for new deposits owing to
18 Chao, T. T, and Sanzolone, R. F., J. Geochem. Explor., 1992, 44, 65.the expense of TIMS analysis. From an earlier study,4 the19 Feldmann, I., Tittes, W., Jakubowski, N., Stuewer, D., and
Cambrian volcanogenic massive sulfide (VMS) and post-Giessmann, U., J. Anal. At. Spectrom., 1994, 9, 1007.
Cambrian (Post C) mineralisation in Western Tasmania can 20 Moens, L., Verrept, P., Dams, R., Greb, U., Jung, G., and Laser,be clearly separated when the Pb isotope ratio precision is less B., J. Anal. At. Spectrom., 1994, 9, 1075.
21 Moens, L., Vanhaecke, F., Riondato, J., and Dams, R., J. Anal.than ±0.1 for 208Pb/ 204Pb and ±0.05 for 207Pb/ 204Pb andAt. Spectrom., 1995, 10 , 569.206Pb/ 204Pb (in terms of absolute ratio units). Our work has
22 Price Russ, G., in Applications of Inductively Coupled Plasmashown that Pb isotope ratios can be measured relativelyMass Spectrometry, ed. Date, A. R., and Gray, A. L., Blackie,
quickly by HR-ICP-MS to a precision approaching thatGlasgow, 1989, pp. 90–110.
obtained by TIMS on the same mixed acid digestion solutions 23 Jarvis, K. E., Gray, A. L., and Houk, R. S., Handbook of normally used in commercial geochemical analyses. This results Inductively Coupled Plasma Mass Spectrometry, Blackie, Glasgow,
1992, p. 380.in a reduction in cost when compared with TIMS measure-24 Sewell, D. M., and Wheatley, C. J. V., J. Geochem. Explor., 1994,ments, and opens up the possibility of the routine use of Pb
50, 351.isotopes in geochemical exploration surveys for new mineral
deposits.Paper 8/ 01397G
Received February 18, 1998CONCLUSION Accepted May 7, 1998
This work has clearly shown that under routine operating
conditions, precise and accurate Pb isotope ratios can be
Journal of Analytical Atomic Spectrometry, August 1998, Vol. 13 813