67132586
-
Upload
nthie-ungu -
Category
Documents
-
view
218 -
download
0
Transcript of 67132586
-
7/29/2019 67132586
1/8
ORIGINAL ARTICLE
Prevention of pressure ulcer:interaction of bodycharacteristics and differentmattressesTheodoros Moysidis, Wolfgang Niebel, Katharina Bartsch, Irene Maier,Nils Lehmann, Michael Nonnemacher, Knut Kroeger
Moysidis T, Niebel W, Bartsch K, Maier I, Lehmann N, Nonnemacher M, Kroeger K. Prevention of pressure ulcer:interaction of body characteristics and different mattresses. Int Wound J 2011; 8:578584
ABSTRACTWe analysed the effect of different body features on contact area, interface pressure and pressure distribution ofthree different mattresses. Thirty-eight volunteers (age ranged from 17 to 73 years, 23 females) were asked tolie on three different mattresses in a random order: I, standard hospital foam mattresses; II, higher specificationfoam mattresses (Viscorelax Sure); III, constant low pressure devices (CareMedx, AirSystems). Measurementswere performed in supine position and in a 90 left- and right-sided position, respectively, using a full-body mat(pressure mapping device Xsensor X2-Modell). Outcome variables were contact area (CA) in cm2, mean interfacepressure (IP) in mmHg and pressure distribution (PD) estimated as rate of low pressures between 5 and 33 mmHgon each mattress in percent. Mean CA was lowest in the standard hospital foam mattresses and increased in thehigher specification foam mattresses and was highest in the constant low pressure device (supine position: 491 86 cm2, 615 95 cm2, 685 116 cm2). Mean IP was highest in the standard hospital foam mattresses and lowerbut similar in the higher specification foam mattresses and the constant low pressure devices (supine position:
223 15 mmHg, 176 17 mmHg, 176 22 mmHg). Models were estimated for CA, IP and PD includingthe independent variables height, weight and waist-to-hip-ratio (WHR). They show that body morphology seems toplay a minor role for CA, IP and PD, but very thin and tall patients and very small and obese people might benefitfrom different mattresses. Our data show that CA increases with increasing specification of mattresses. Higherspecification foam mattresses and constant low pressure devices show similar IP, but constant low pressure devicesshow a wider pressuredistribution. Body morphology shouldbe considered to optimise preventionfor singlepatients.
Key words: Interface pressure Mattresses Pressure ulcer Waist-to-hip-ratio
Authors: T Moysidis, MD, Department of Angiology, HELIOS
Klinik Krefeld, Krefeld, Germany; W Niebel, MD, Department of
General and Transplant Surgery, University of Duisburg-Essen,
Germany; K Bartsch, Department of Angiology, HELIOS Klinik
Krefeld, Krefeld, Germany; I Maier, Clinical Nurse Manager,
University of Duisburg-Essen, Germany; N Lehmann, Insti-
tute for Medical Informatics, Biometry and Epidemiology, Med-
ical Faculty, University of Duisburg-Essen, Essen, Germany;
M Nonnemacher, Institute for Medical Informatics, Biometry
and Epidemiology, Medical Faculty, University of Duisburg-
Essen,Germany; K Kroger, Department of Angiology, HELIOS
Klinik Krefeld, Krefeld, Germany, Initiative Chronische Wunden.
e. V., Luenen, Germany
Address for correspondence: Professor K Kroeger, MD,
FASA, EFMA, Klinik f ur Angiologie, HELIOS Klinik Krefeld,
Lutherplatz 40, 47805 Krefeld, Germany
E-mail: [email protected]
INTRODUCTION
Pressure ulcer also known as bedsores, pres-Key Points
the awareness for prevention
of pressure ulcers and for early
detection increased in the past
decades and pressure relief is
the most mandatory concept to
prevent pressure ulcers
thus, pressure-relieving beds,
mattresses and seat cushions
are widely used as aids of
prevention in both institutional
and non institutional settings
sure sores or decubitus ulcer is a generally
preventable complication of immobility. The
awareness for prevention of pressure ulcersand for early detection increased in the past
decades and pressure relief is the most manda-
tory concept to prevent pressure ulcers (13).
Thus, pressure-relieving beds, mattresses and
seat cushions arewidely used as aids of preven-
tion in both institutional and non institutional
settings. Technical analysis concentrates on the
quantification of the applied pressure at the
patientsupport interface and the limitations
2011 The Authors
578 2011 Blackwell Publishing Ltd and Medicalhelplines.com Inc International Wound Journal Vol 8 No 6
-
7/29/2019 67132586
2/8
Characteristics of mattresses
and variability of current techniques, mea-
surement systems and data presentation (47).
A recent Cochrane analysis concluded that
higher specification foam mattresses rather
than standard hospital foam mattresses can
reduce the incidence of pressure ulcers in
people at risk but the relative merits of alter-nating and constant low pressure devices
are unclear (8,9). Many risk factors predis-
posing patients to develop pressure ulcer are
described and implemented in a variety of
risk scores (3,10,11). However, there are less
data regarding the interaction of body shape
and specific mattresses. A study of individu-
als with spinal cord injuries had shown that
peak seat-interface pressure was highest in the
thin elderly group, which had the lowest BMI
(body mass index) levels. Differences in the
peak seat-interface pressures were less as BMI
increased (12).
Key Points
we hypothesised that body
morphologyhas differenteffects
on interface pressure depending
on the specification of different
mattresses
thus, we analysed the effect ofdifferent body features on con-
tact area (CA), interface pres-
sure (IP) and pressure distri-
bution (PD) of three different
mattresses
38 volunteers (age ranged from
17to73years,23females)were
asked to lie on three different
mattresses in a random order
the participants were asked
to stay on each mattress for
5 minutes until the pressure
measurement was started
We hypothesised that body morphology has
different effects on interface pressure depend-
ing on the specification of different mattresses.
Thus, we analysed the effect of different body
features on contact area (CA), interface pres-
sure (IP) andpressuredistribution (PD) of three
different mattresses.
METHODSAt the University Hospital of Essen, Germany,
healthy volunteers (students, visitors andemployees) were asked to participate in this
study. People were selected to cover a wide
range of age, weight and height (Table 1).
Finally, 38 volunteers (age ranged from 17 to
Table 1 Characteristics of body morphology of the 23 females
and 15 males enrolled in the project
Variable Mean SD Range
Age (years) 419 160 1773
Height (cm) 1714 112 150199
Weight (kg) 749 151 47130
BMI (kg/cm2) 255 49 1745
Body surface area (m2) 19 02 1425
Chest circumference (cm) 998 102 79133
Waist circumference (cm) 923 155 67134
Hip circumference (cm) 1034 116 88149
Waist-to-hip-ratio 089 009 071106
Leg length (cm) 944 94 74114
Shoe size (German size) 408 31 3549
BMI, body mass index.
73 years, 23 females) were asked to lie on three
different mattresses in a random order:
I. A standard hospital foam mattress, which
was the standard mattress in the hospital till
2005 (weight 12 kg, thickness 14 cm, density
35 kg/cbm).
II. A higher specification foam mattress(Viscorelax Sure, Hapeka, Buhl, Germany),
which was bought for some departments, that
is, the Department of Oncology, with a higher
frequency of patients a risk. The mattress is in
the superstructure of viscoelastic foam and the
base of high-quality open-pore polyether foam.
Sinking into the viscoelastic foam reduces the
hammock effect and provides an optimum
pressure relief and functional support of the
spine. Mattress forms at body temperature and
thus adapts itself to the body contours (weight
17 kg, thickness 17 cm, density 80/40 kg/cbm).
III. A constant low pressure device (Care
Medx, AirSystems), which may be ordered
for special risk patients. The modular mattress
system has20 airchambers. It hasan individual
pressure profile control for four body regions
(head, chest, pelvis and foot). The upper is
made of nylon.
All three systems were placed near each
other in the same room. The participants were
asked to stay on each mattress for 5 minutes
until the pressure measurement was started.
Measurements were performed in supine posi-
tion and in a 90
left- and right-sided posi-tion, respectively. The participants were asked
to take off thick or constricting clothing but
remained dressed. Most of them wear a shirt
and pants.
A full-body mat [pressure mapping device
Xsensor X2-Modell (Fa: FSA, type Ultra-Thin,
Software Version 40.)] was used to record the
pressures over the complete body (7). Outcome
variable were CA in cm2, IP in mmHg and PD
estimated as rate of low pressures between 5
and 33 mmHg on each mattress in percent.
Sensitivity of the Xsensor X2-Modell is
-
7/29/2019 67132586
3/8
Characteristics of mattresses
Table 2 Contact areas and pressures measured while lying on the rear and in the left and right side, respectively, on the three
different mattresses
Mattresses I II III
Supine position
Contact area (cm2) 491 86 615 95
-
7/29/2019 67132586
4/8
-
7/29/2019 67132586
5/8
Characteristics of mattresses
Table 4 Results of the final linear regression analysis in supine position
Weight Size Waist/hip ratio Intercept R2
Surface area
I 507583
40236129
-
7/29/2019 67132586
6/8
Characteristics of mattresses
with the Cochrane analysis (8). Thus, our
study concept can be assumed to be valid.
The repeatability of a technique for measur-
ing interface pressures has been assessed by
Allen et al. Pressure measured using a Tal-
ley SA500 Pressure Evaluator (Omega Medical
Systems, Pottstown, PA) under six anatomi-cal sites was found to vary significantly more
between days than between repeats on the
same day (P < 002) (14).
Key Points
ourdatashow that CAincreases
with increasing specification of
mattresses
constant low pressure devices
differ from higher specification
foam mattresses by a widerpressure distribution
body morphology should be
considered to optimise preven-
tion for single patientsSimilar studies have been carried out before,
but they focused on IP and did not analyse the
influence of body morphology in detail (1418).
Hockersen et al. compared IP in the Ortho-
derm Convertible II (Bio-Clinic Inc., Toronto,
Ontario, Canada) low air-loss mattress, in a
Stryker PMS(StrykerInc., Kalamazoo, MI)bed;
the FluidAir Elite (KCI Inc., San Antonio, TX)
air-fluidised bed; the Pegasus Airwave mat-
tress (Pegasus Airwave Inc., Boca Raton, FL) in
a Stryker PMSbed anda standard hospital mat-
tress in a Stryker PMS bed (17). They showed
that the Pegasus Airwave mattress total IP val-
ues of the anatomical regions (total body, torso
and hips) of the body are lower than those
of the other three tested mattresses. The total
body and hip pressure resulting from pres-
sure maturation was significantly less in the
Pegasus Airwave mattress than in the hospital
bed, the air-fluidised bed and the low air-loss
bed. Allen et al. measured IP interface pressure
in four continuous airflow mattress overlays[Clinirest (SSI Medical Services Inc, Yonkers,
NY), FirstStep (KCI)] and alternating pressure
air mattresses [Airwave (Pegasus) and Nimbus
(Huntleigh Healthcare Limited, Bedfordshire,
UK)] (14). Their data suggest a clinical benefitat
the occiput and heel using an alternating pres-
sure air mattress and a benefit in using a con-
tinuous airflow mattress overlay at other sites.
In contrast to the literature we focus on body
morphology. Thicker people have relatively
smaller BSA than thinner people presenting
with the same weight. BSA is the theoretical
limitation of CA. Thus, almost spherical-
shaped bodies should have higher IP than
thin cuboid bodies of the same weight.
These assumptions retrieve themselves in our
final models estimating the CA and IP on
the different mattresses. Weight is the most
important variable of course, but size and
WHR play a role too as they describe the
spherical shape of the body. The last two
variables become more important in the higher
Table 5 Estimated CA, IP and pressure distribution in supine
position for two different body shapes based on the models in
Table 4
CA (cm2) IP (mmHg)
Pressure distribution (%
between 5 and 35 mmHg)
Patient A: weight 90 kg, size 160 cm, WHR 1 5I 589 227 781
II 660 184 885
III 870 171 962
Patient B: weight 90 kg, size 190 cm, WHR 091
I 567 227 783
II 715 178 890
III 779 185 898
CA, contact area; IP, interface pressure; WHR, waist-to-hip-ratio.
specification foam mattresses and constant low
pressure devices, because they allow the body
to cave in. As we showed in Table 5 very thin
and tall patients should benefit from higherspecification foam mattresses, whereas small
and obese people should benefit from constant
low pressure devices. This influence of body
morphology on IP and pressure distribution
might be considered in future developments of
pressure-relieving mattresses.
LIMITATIONSAlthough this study has a clear design and
deals with objective measurements, some lim-
itations have to be mentioned. The number of
participants is rather small and does not neces-sarily cover the whole range of variables with
the same frequency. Repetitive measurements
on the same mattresses could have shown
variability in the data. Longer resting periods
on the different mattresses could have given
insight into temporary changes of pressure
distribution.
CONCLUSIONOur data show that CA increases with
increasing specification of mattresses. Con-
stant low pressure devices differ from
higher specification foam mattresses by a
wider pressure distribution. Body morphology
should be considered to optimise prevention
for single patients.
REFERENCES1 Hardin J, Cronin S, Cahill K. Comparison of the
effectiveness of two pressure relieving surfaces:
low-air-loss versus static fluid. Ostomy Wound
Manage 2000;46:506.
2011 The Authors
2011 Blackwell Publishing Ltd and Medicalhelplines.com Inc 583
-
7/29/2019 67132586
7/8
Characteristics of mattresses
2 Whitehead SJ, Trueman P. To what extent can
pressure relieving surfaces help reduce the costs
of pressure ulcers? Nurs Times 2010;106:1012.
3 Wilborn D, Grittner U, Dassen T, Kottner J. The
national expert standard pressure ulcer preven-
tion in nursing and pressure ulcer prevalence in
German health care facilities: a multilevel analy-
sis. J Clin Nurs 2010;19:336471.
4 Ferguson-Pell M, Cardi M. Pressure mapping sys-
tems. Team Rehab Rep 1992;3:2832.
5 Harstall, C. The effectiveness of interface pressure
measurements. Report der Alberta Heritage
Foundation for Medical Research, Edmonton,
Alberta, Canada, 1996. Report No 1.
6 Jay R. Pressure and shear: their effect on sup-
port surface choice. Ostomy Wound Manage
1995;41:3645.
7 Swain ID, Bader DL. The measurement of interface
pressure and its role in soft tissue breakdown. J
Tissue Viability 2002;12:1324.
8 McInnes E, Bell-Syer SE, Dumville JC, Legood R,
Cullum NA. Support surfaces for pressure
ulcer prevention. Cochrane Database Syst Rev
2008;CD001735.
9 Moore ZE, Cowman S. Repositioning for treating
pressure ulcers. Cochrane Database Syst Rev
2009;CD006898.
10 Pancorbo-Hidalgo PL, Garcia-Fernandez FP, Lopez-
Medina IM, Alvarez-Nieto C. Risk assessment
scales for pressure ulcerprevention: a systematic
review. J Adv Nurs 2006;54:94110.
11 Wilborn D, Halfens R, Dassen T. Pressure ulcer:
prevention protocols and prevalence. J Eval Clin
Pract 2006;12:6308.
12 Kernozek TW, Wilder PA, Amundson A, Hum-
mer J. The effects of body mass index on
peak seat-interface pressure of institutionalized
elderly. Arch Phys Med Rehabil 2002;83:86871.
13 Reger SI, Ranganathan VK, Sahgal V. Support sur-
face interface pressure, microenvironment, and
the prevalence of pressure ulcers: an analy-
sis of the literature. Ostomy Wound Manage
2007;53:508.
14 Allen V, Ryan DW, Murray A. Measurements of
interface pressure between body sites and the
surfaces of four specialised air mattresses. Br
J Clin Pract 1994;48:1259.
15 Allen V, Ryan DW, Murray A. Repeatability of
subject/bed interface pressure measurements. J
Biomed Eng 1993;15:32932.
16 Gil-Agudo A, De la Pena-Gonzalez A, Del Ama-
Espinosa A, Perez-Rizo E, Daz-Domnguez E,
Sanchez-Ramos A. Comparative study of pres-
sure distribution at the user-cushion interface
with different cushions in a population with
spinal cord injury. Clin Biomech (Bristol, Avon)
2009;24:55863.
17 Hickerson WL,Slugocki GM,Thaker RL,Dunkan R,
Bishop JF, Parks JK. Comparison of total body
tissue interface pressure of specialized pressure-
relieving mattresses. J Long Term Eff Med
Implants 2004;14:8194.
18 Goossens RH, Rithalia SV. Physiological response
of the heel tissue on pressure relief between
threealternating pressure air mattresses. J Tissue
Viability 2008;17:1014.
2011 The Authors
584 2011 Blackwell Publishing Ltd and Medicalhelplines.com Inc
-
7/29/2019 67132586
8/8
Copyright of International Wound Journal is the property of Wiley-Blackwell and its content may not be copied
or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission.
However, users may print, download, or email articles for individual use.