    INSTALLATION AND OPERATING INSTRUCTIONSX-RAY FILM PROCESSOR DRR XR 04

  • 8/8/2019 60005_0307_gb

    2/142

  • 8/8/2019 60005_0307_gb

    3/143

    CONTENTS

    Important information

    1. Notes ......................................................... 41.1 Test of conformity .............................. 41.2 General Notes .................................... 41.3 General Safety Notes ........................ 41.4 Notes concerning Medical Appliances

    51.5 Electrical Safety Notes ...................... 51.6 Warnings and Symbols ..................... 5

    2. Product information ................................ 62.1 Use in accordance

    with regulations .................................. 6

    2.2 Use not in accordancewith regulations .................................. 6

    3. Model survey ........................................... 64. Delivery schedule .................................... 6

    4.2 Special extras .................................... 65. Technical data .......................................... 67. Function description ............................... 76. Function principle ................................... 7

    Installation

    8. InstallationRequirements .......................................... 88.1 Installation site ................................... 88.2 Installation possibilities of the

    Processor ........................................... 8

    Gebrauch

    9. Changing chemicalsand water .................................................. 99.1 Taking off daylight attachment........... 99.2 Cleaning film transport .................... 109.3 Removing heater ............................. 119.4 Disposing of used chemicals .......... 119.5 Mixing new chemicals .....................119.6 Assembling of the XR 04 .................12

    10. Operation ............................................... 1211. Replacing film transport cogs ............. 1312. Wiring diagram ......................................14

    Disposal

    13. Disposing of Appliance .........................14

  • 8/8/2019 60005_0307_gb

    4/144

    IMPORTANTINFORMATION

    1. NOTES1.1 Test of conformityThis product was tested for conformity to theGuidelines 93/42/EWG of the European Unionand has been found to satisfy all criteria ofthese guidelines.

    1.2 General Notes These Installation and Operating Instructions

    form an integral part of the unit. They must

    be kept close to the unit at all times. Preciseobservance of these instructions is a pre-condition for use of the unit for the intendedpurpose and for its correct operation. Newpersonnel must be made aware of thecontents, and they should be passed on tofuture operating staff.

    Safety for the operator as well as trouble-free operation of the unit are only ensured ifuse is made of original equipment parts.Moreover, use may only be made of those

    accessories that are specified in thetechnical documentation or that have beenexpressly approved and released by DrrDental for the intended purpose.

    Drr Dental cannot guarantee for the safetyor proper functioning of this unit in the casewhere parts or accessories are used whichare not supplied by Drr Dental.

    Drr Dental are only responsible for theequipment with regard to safety, reliability

    and proper functioning where assembly,resettings, changes or modifications,extensions and repairs have been carriedout by Drr Dental or an agency authorizedby Drr Dental and if the equipment is usedin conformity with the Installation andOperating Instructions.

    These Installation and Operating Instructionsconform to the relevant version of theequipment and the underlying safetystandards valid at the time of going to press.

    All switches, processes, trade marks,software programs and appliances named inthis document are registered names.

    Any reprinting of the technicaldocumentation, in whole or in part, is subjectto prior approval of Drr Dental being givenin writing.

    1.3 General Safety NotesThis appliance has been designed and

    constructed by Drr Dental so that correctusage of the appliance is virtually free of anypossible injury or danger. In spite of this, wefeel it is our duty to mention the followingsafety measures in order to prevent anypossible danger.

    When using this appliance all local andrelevant regulations must be observed!Converting or modifying the appliance inany way is strictly prohibited. In such cases,any and all guarantees immediately become

    invalid. The operation of modifiedappliances can be punishable by law. In theinterests of trouble-free operation theoperator is responsible for observing theseregulations.

    Retain the packaging for possible return ofthe product to the manufacturers. Ensurethat the packaging is kept out of the reach ofchildren. Only the original packagingprovides adequate protection duringtransport of the unit.

    Should return of the product to themanufacturers be necessary during theguarantee period, Drr Dental accepts noresponsibility for damage occurring duringtransport where the original packaging wasnot used!

    Before every use the operator must checkthe functional safety and the condition of theappliance.

    The operator must be knowledgeable in the

    operation of the appliance. The product is not designed to be used in

    medical treatment areas where there existsthe danger of explosion. Areas whereexplosions could occur are those whereflammable anesthetic material, skincleansers, oxygen and skin disinfectants arepresent. This appliance is not to be used inareas where the atmosphere could causefire.

  • 8/8/2019 60005_0307_gb

    5/145

    1.4 Notes concerning MedicalAppliances

    This product is a technical medicalappliance and, as such, may only beoperated by trained personnel, or personswho, as a result of specialist knowledge, arefamiliar with this type of appliance.

    Do not lie multi-socket units on the floor.

    Other systems should not be plugged intothe same multi-socket unit.

    1.5 Electrical Safety Notes The appliance may only be connected to an

    earthed safety socket.

    Before connecting the appliance to thepower supply check that the electrical

    current and the frequency of the device asdescribed on the appliance are compatiblewith that of the power supply.

    Check the appliance and the power supplycables for possible damage beforeswitching on. Damaged cables, plugs andsockets must be replaced before use.

    Never touch open supply outlets andpatients simultaneously.

    1.6 Warnings and SymbolsIn the Installation and Operating Instructionsuse is made of the following terms or symbolsto denote information of special importance:

    Information and/or mandatory

    regulations or prohibitions for the

    prevention of personal injury orsubstantial property damage

    Warning of dangerous electrical

    voltage.

    Special information regarding theeconomical use of the equipment andother information

    CE-Labeling without Notified BodyNumber

    Observe Installation and OperatingInstructions!

    Date of manufacture

  • 8/8/2019 60005_0307_gb

    6/146

    4. DELIVERY SCHEDULE4.1 Optional extras

    1 X-ray filmProcessor XR 04.............................1740-01

    1740-02

    1740-031740-041 Securing set ........................... 1740-004-001 Chemical set...................... 1307-080-00/40

    1 Guarantee card ...................... 9000-464-861 Instruction manual ............. 9000-600-05/01

    1 Wall mounting ......................... 1740-011-00

    4.2 Special extras1 Heater, PC-board 110V ......... 1740-994+00

    1 Heater, PC-board 230V ......... 1740-993+001 Table trestle ............................ 1740-012-00

    5. TECHNICAL DATAModel 1740 - 01 02 03 04

    Voltage (V) 230 110 230 110

    Current

    Consumption

    (mA) 35 - 400 -Heater w/o w/o with with

    Frequency (Hz) ................................... 50 - 60Film run through time (adjustable min) 2 - 6Total output (W) .......................................... 85Heat output (W) ........................................... 40Protection rating Il, Protection type IP 21

    Heating up time(approx. min) (8 min / C) ................... 20 - 30

    Weight (kg) without chemicals ................ 7,5

    2. PRODUCT INFORMATION2.1 Use in accordance

    with regulationsThe Drr XR 04 X-ray developing machine isto be used exclusively for the automaticdevelopment of intra-oral X-ray films.

    Use in accordance with the regulationsincludes compliance with the instructions formounting and usage as well as compliancewith the provisions on setting up, installing andoperating the machine, and on warranty.Use in accordance with the regulations alsoincludes compliance with all legal provisionsin force in the place where the machine isbeing used, with regard to work safety and thedisposal of chemicals.

    2.2 Use not in accordancewith regulations

    If the machine is used for a purpose other thanthe one stated in the regulations, themanufacturer shall not be legally liable for anydamage arising as a result of same. In suchcases, the user shall incur the risk.

    3. MODEL SURVEYX-ray film Processor XR 04

    Model:Model:Model:Model:Model: 1740-01 (230 V)1740-02 (110 V )1740-03 (230 V) with heater1740-04 (110 V) with heater

  • 8/8/2019 60005_0307_gb

    7/147

    6. FUNCTION PRINCIPLE

    E Developer bathF Fixer bath

    W Water bath1 Processing time regulator2 Unit switch ON/OFF3 Film transporter4 Heater5 Transporter drive6 Perforated collector

    1

    2

    3

    4 5

    6

    7. FUNCTION DESCRIPTIONSwitch on main switch (2).Heater for Developer/Fixer baths is started

    (8 min/C to 25C bath temperature).Switch on regulator (1) for the film

    transport and set the processing time

    (2-6 min.).

    The films are carried via the film transporterthrough the Developer-/Fixer baths. Theprocessed film remains in the perforatedcollector and should be taken out afterapprox. 5 mins.

  • 8/8/2019 60005_0307_gb

    8/148

    INSTALLATION

    8. INSTALLATIONREQUIREMENTS

    8.1 Installation site The Processor XR 04 must be assembled in a

    dry, well ventilated room.

    Never expose the Processor to direct

    sunlight!

    With too strong sunlight pre-exposure offilms and overheating of chemicals canoccur.

    8.2 Installation possibilities of the

    ProcessorTo obtain perfect development results, the

    installation area must be stable, flat and level.

    Assemble at or bring up to working height.

    Near a sink.

    Near a mains socket.At a distance of approx.1.5 m from theinstallation site of the XR 04, there should be amains socket for the power cable of the XR 04.The mains socket should be sited so that thepower plug of the machine is easily visibleand accessible without danger.

    (DIN 15990, para. 3.3.2 and 3.4)

    For safety technical reasons the

    power connection should only bemade shortly before putting the

    machine into operation.

    Wall mounting see. pict. 1.

    Mark the position for drilling the holes for theXR 04 (10) on the wall.

    Drill the holes, insert wall plugs and fit screws. Hang XR 04 on the screws, adjust as

    necessary, then tighten screws.

    Table mounting see pict. 2.

    Fit the upper part of the XR 04 onto the edgeof the table and fasten it (11).

    Table installation see pict. 3.

    For this assembly it is necessary to have aseparate table trestle (12), (not included in thedelivery schedule)

    Order no.1740-012-00.

    1

    2

    3

    10

    11

    12

  • 8/8/2019 60005_0307_gb

    9/149

    4

    5

    GEBRAUCH

    9. CHANGING CHEMICALSAND WATER

    Before working on opened

    equipment, pull out the mains plug.

    Change chemicals after approx. 200 films orafter 3 weeks at the latest.

    Change water twice a week.

    The information for changing thechemicals refer only when using"DRR Chemicals"."DRR Chemicals"."DRR Chemicals"."DRR Chemicals"."DRR Chemicals".In Germany, as well as some othercountries, used x-ray chemicals areclassified as special waste and mustbe collected separately and disposedof.

    9.1 Taking off daylight attachmentBefore taking off the daylight

    attachment: Pull out mains plug.

    Push frame support (14) inwards with both

    thumbs. Swing daylight attachment (13)upwards and remove, see pict. 4.

    13

    14

    3

  • 8/8/2019 60005_0307_gb

    10/1410

    7b

    6

    7

    7a

    9.2 Cleaning film transport Lift the film transport (3, Fig. 5) upwards to

    remove and rinse with warm water atapprox. 40C.

    If the film transport is very dirty it can beseparated into two pieces and cleaned.Hereby :

    Pull the 4 snap fasteners of the film transport(16) out until they click into place.

    Using even pressure pull the top half fromthe bottom half (Fig. 6).

    Rinse both halves and, if necessary, cleanwith a brush.

    Please note that when cleaning theopen film transport the toothed beltcan get out of position. Therefore,before reassembling the film transport

    check the position of the cogs (19)and the toothed belt (20) as follows:

    The transport arm (17) should be turnedturnedturnedturnedturnedclockwise (21)clockwise (21)clockwise (21)clockwise (21)clockwise (21) until it aligns with marking(A).

    The film carrier (18) should now be alignedwith marking (B). If not, then correct theposition of the toothed belt (20):

    Fig. 7a: Hold the film transport, so that thelarge, white cogs cannot turn.

    Fig. 7b: Lift the toothed belt (20) from thecogs (19), align the film carrier (18) withmarking (B).Replace the toothed belt in the correctposition and press onto the teeth of the cogwheel.

    Here by take care that the toothed beltsits correctly on the cog with nooverlap, otherwise there is the dangerthat the belt comes loose from thecogs during operation.

    Place the two halves of the film transportagainst each other and replace the snapfasteners.

    16

    17

    18

    19

    20

    21

    18

    19

    20

  • 8/8/2019 60005_0307_gb

    11/1411

    8

    W

    E

    F

    9

    9.3 Removing heaterOnly necessary with table-top model,otherwise the Developer-/Fixercontainers cannot be removed.

    Lift up Heater (4) and hang on frame.

    9.4 Disposing of used chemicals Remove by rotating the 3 containers

    Developer (E)/Fixer (F)/Water (W) and verycarefully take off downwards.

    Empty out chemical and water containers.

    Dispose of chemicals.

    Rince containers with warm water and cleanthoroughly.

    9.5 Mixing new chemicalsFill containers only up to the

    markings.

    Danger of "spillage" during

    insertion of containers.

    Pour developer up to black marking incontainer (E) and then fill up with fresh waterto the upper marking. Insert container androtate to fasten.

    Pour fixer up to the red marking in container

    (F) and then fill up with fresh water to theupper marking.Insert container and rotate to fasten.

    Fill container (W) with white markings up tothe upper mark with fresh water.Insert container and rotate to fasten.

  • 8/8/2019 60005_0307_gb

    12/1412

    10

    F WE

    11

    12

    Pos. 1

    Pos. 2

    9.6 Assembling of the XR 04 Replace heater (4) and film transporter (3),

    see pict. 5.

    Replace daylight attachment (13).

    Put the perforated collector (6) in the watercontainer (6), see pict. 11 and right fold-inpage.

    10. OPERATION Machine switch (2) ON.

    Developer-/Fixer bath are being warmed up.

    Processor regulating timer (1) ON.Set processing time, see pict. 12.without heater:

    Regulator on pos. 13 1/2 minswith heater:

    Regulator on pos. 22 1/2 mins.

    Place the wrapped exposed film in thedaylight attachment.

    Insert hands into rubber cuffs and unwrapfilm.

    Enter film into the transporter system.

    Remove processed film from the water bath.Washing: A min. of 5 mins in the perforatedcollector (6) (max. 1 hour)

    Processor regulating timer (1)OFF.

    13

    4

    6

  • 8/8/2019 60005_0307_gb

    13/1413

    11. REPLACING FILMTRANSPORT COGS

    Remove the locking ring of the defective cogand pull the cog from the roller.

    Align the transport arm (A) with marking (B),press new cog into position and secure withthe locking ring.

    Place the toothed belt cog (X) onto the filmcarrier roller (C) so that this appears as a D .

    The transport arms (A) must be aligned withthe markings (B) and the film carrier (C)shows a D shape.

    Place the toothed belt in position andsynchronise, see Point 9.2.

    X B A

    B

    A

    C

  • 8/8/2019 60005_0307_gb

    14/14

    DISPOSAL

    13. DISPOSING OF APPLIANCEThe built-in electronic plate and componentsmust be disposed of in the same way aselectronic scrap metal.The remaining structural components can bedisposed in accordance with the relevant localregulations.

    X1

    X2

    N

    L

    NTC

    X4

    X2

    X1

    2 1

    4 3

    X3

    X3

    30 31 32

    36 35 34 33

    12. WIRING DIAGRAM

    30 Mains31 Heater32 Processing time regulator33 Drive motor34 PC board35 Unit switch36 Fuse 100 mA T