5 MeV Mott PolarimeterProgress
Riad Suleiman
April 10, 2012
![Page 2: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/2.jpg)
Outline• Mott Scattering: Cross Section and Sherman Function
• Measuring Mott Asymmetry
• Mott Detectors
• Data Acquisition
• Analysis
• Charge Asymmetry
• 1% Polarization Measurement – Error Budget
![Page 3: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/3.jpg)
Mott Scattering
• Electron motion in the electric field of nucleus results in magnetic field in electron rest frame, , if is nucleus-electron separation, then and
• Interaction of this magnetic field with electron (spin) magnetic moment introduces a term in the scattering potential,
• Presence of spin-orbit term in scattering potential introduces spin dependence in scattering cross section which could be detected as a left/right count rate asymmetry
EVc
B
1 r
r
r
ZeE
3
Lmcr
Zevr
cr
ZeB
33
LV sso
SLrcm
Ze
r
ZeVVV soc
322
2
2
)(
![Page 4: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/4.jpg)
Note: • Parity-conserving: Measure spin-momentum correlation of the type:
Transverse (or Normal) Beam Asymmetry measured recently using the setup of parity-violating experiments at high energies (due to two-photon exchange) probes the same spin-momentum correlation as Mott Asymmetry at low energies (due to spin-orbit interaction of electron moving in a Coulomb field).
• Parity-violating: Measure spin-momentum correlation of the type:
21. kkS
1.kS
VVrr
LL LL VVrr
LL LL
![Page 5: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/5.jpg)
Mott Cross Section and Sherman Function• Mott cross section:
where, is the un-polarized cross section,
and is the analyzing power (Sherman Function),
• The Sherman Function is largest for high-Z (Gold, Z=79) targets and low-energy electrons
]ˆ)(1)[()( nPSI
)2/(cos
)(
)2/(sin
)(1)(
2
2
2
2
2
222
GF
c
Ze
p
cI
)()()()()()2/sin(
12)( *
222
GFGFIc
Ze
p
cS
)(I
)(S
Non-spin-flip Amplitude
Spin-flip Amplitude
![Page 6: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/6.jpg)
• Theoretical corrections to Sherman Function:I. Screening by atomic
electrons which is relevant for low energy electrons
II. Nuclear extended charge distribution which is relevant for high energy electrons
![Page 7: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/7.jpg)
Mott Polarimeter Optimization
• Statistical error of polarization measurement is proportional to inverse of Figure of Merit (fom),
The goal is to maximize fom
• The detector rate (R) is
2)()()( SIfom
e
I
M
NdIR beam
Au
AfoilAu )()(
![Page 8: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/8.jpg)
Measuring Mott Asymmetry
• How to measure the Mott Asymmetry ALR?– For one helicity state, measure the number of left and right E detector events, and – Flip the electron polarization, measure the number of events again, and – Calculate the cross-ratio (r),
– Then, the Mott Asymmetry (A),
• The same for AUD
• This cancels false asymmetries from detector efficiency, beam current, target thickness, and solid angle
• Dead time is caused by slow DAQ and is common to all detectors – cancels to all orders
LN
RL
RL
NN
NNr
RN
LN
RN
r
rALR
1
1
S
AP LR
![Page 9: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/9.jpg)
Statistical Uncertainty
• With the approximation,
Error simplifies to
or,
LRRL NNNNN
NA
4
1
2)(4
1
AN
• Time needed to measure beam polarization of to statistical error of is PP /
P
2222
4
22
)1()(
R
R
R
R
L
L
L
L
N
N
N
N
N
N
N
N
r
rA
fomPSPRARR
NT
222 2
1
))((2
1
)(2
12
![Page 10: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/10.jpg)
Sherman Function and Target Thickness• Single-Atom Sherman
Function must be corrected for plural scattering (a few large angle scattering) in the target:
• alpha = 0.3/um for 5 MeV electrons. Depends on electron energy and may depend on scattering angle
• Run with the thinnest target
)(SAS
d
SdS SA
eff
)(1
)(),(
)(effS
AP
![Page 11: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/11.jpg)
ddSI
dfom
eff
2),()(
),(
![Page 12: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/12.jpg)
Sherman Function Sensitivity toEnergy and Angle
![Page 13: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/13.jpg)
Three Targets (Z=29, 47,79)
![Page 14: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/14.jpg)
![Page 15: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/15.jpg)
I. BackgroundI. Shielding and Collimation
II. Coincidence, Time-of-flight
Or … Subtract background:
II. Target Thickness:o Single-Atom Sherman Function must be corrected for plural
scattering (a few large angle scattering) in target:
o SSA(0) = -0.5215, S(1.0 μm) = -0.4006
o If possible, run with the thinnest target
Corrections to Measured Asymmetry
d
SdS SA
1
)0()(
LrawLL brNN
2
2
)(
L
LrawL
L
L
N
brN
N
N
![Page 16: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/16.jpg)
5 MeV Mott Beamline
![Page 17: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/17.jpg)
Detector Assembly
I. Scattering Angle = 172.6°
II. Solid Angle = 0.18 msr
![Page 18: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/18.jpg)
New ΔE and E Detectors Installed
- H7415 (R6427) 1" PMT
- 1 mm x 1" x 1" EJ-212 Plastic Scintillator
- 0.125" x 1" x 2" Acrylic Light Guide
- H6559 (R6091) 3" PMT
- 3" diameter x 2.5" long EJ-200 Plastic Scintillator painted with EJ-510
![Page 19: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/19.jpg)
ΔE and E Signals
ΔE Detector
E Detector
![Page 20: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/20.jpg)
Old 5 MeV Mott DAQ- LeCroy CAMAC 4303 Time-to-FERA Converter (TFC)
- LeCroy CAMAC 4300B Fast Encoding and Readout ADC (FERA), 10 Bit
- ORTEC CAMAC HM 413 HISTO-MEMORY
![Page 21: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/21.jpg)
Detectors Spectra
- Beam Current = 0.5 μA
- Gold Target 1.0 μm
- Trigger Rate 1 kHz
- 5 minutes of data
![Page 22: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/22.jpg)
E Detectors Spectra
![Page 23: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/23.jpg)
Mott Asymmetries
- AUD = 33.98 ± 0.36 %
- ALR = -2.11 ± 0.44 %
- UD Instrumental (Detector Efficiency and Solid Angle) A1 = 12.91 ± 0.40 %
- LR Instrumental (Detector Efficiency and Solid Angle) A1 = -3.78 ± 0.44 %
- UD Instrumental (Beam Current and Target Thickness) A2 = 0.75 ± 0.41 %
- LR Instrumental (Beam Current and Target Thickness) A2 = -0.75 ± 0.44 %
![Page 24: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/24.jpg)
New DAQ for Mott Polarimeter
• Will record the pulse shape and timing of detected electrons• No dead time at rates <5 kHz, will be able to run at higher
currents• Can process delayed helicity reporting and measure time-of-
flight of detected electrons
• Consists of:– CODA (CEBAF Online Data Acquisition)– Hardware:
• VME64x Backplane 6U Crate
• Motorola MVME6100
• JLab Flash ADC: 16 channel, 12 bit, 250 MS/s
• SIS 3801 Scalers: Beam current and position
• CAEN V775 TDC: BFM
![Page 25: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/25.jpg)
DAQ Schematic Diagram
MVME6100MVME6100 FADCFADC ScalersScalersTDCTDC
∆E∆E
EE
BFMBFM
Helicity SignalsHelicity Signals
BPMBPM
BCMBCM
Helicity SignalsHelicity Signals
![Page 26: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/26.jpg)
Detector Signals to fADC(Parasitic to old DAQ)
E Detectors (50 ns)
ΔE Detectors (50 ns)
Data Trigger (50 ns)
ADC GATE
![Page 27: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/27.jpg)
Pair Sync
Delayed Helicity
T- Settle
Pattern Sync
Helicity Signals
![Page 28: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/28.jpg)
FADC Chan
Signal
0 E LEFT
1 E RIGHT
2 E UP
3 E DOWN
4 ΔE LEFT
5 ΔE RIGHT
6 ΔE UP
7 ΔE DOWN
8 BFM
9
10 Mott Trigger
11
12 Delayed Helicity
13 T_Settle
14 Pattern-Sync
15 Pair-Sync
Mott Trigger
Left E
Left ∆E
FADC Signals Signals on ScopeSignals on Scope
Signals in FADC DataSignals in FADC Data
![Page 29: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/29.jpg)
PedestalADCE
ADCPedestal
sample
sample
38
5
1
97
60
64
60
Calculate pedestal and Energy:Calculate pedestal and Energy:
Few FADC Events in LEFT DetectorFew FADC Events in LEFT Detector
![Page 30: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/30.jpg)
![Page 31: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/31.jpg)
![Page 32: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/32.jpg)
Energy Resolution=2.7%
(same as old DAQ)
Energy Resolution=2.7%
(same as old DAQ)
![Page 33: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/33.jpg)
Pedestal ResolutionPedestal Resolution
Energy Resolution=2.4% (after Pedestal
Correction)
Energy Resolution=2.4% (after Pedestal
Correction)
![Page 34: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/34.jpg)
![Page 35: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/35.jpg)
Mott Asymmetry
IHWP Run # P_V P_X
OUT 2652 86.4 +/- 1.1 (stat) 0.6 +/- 1.2 (stat)
IN 2653 -85.6 +/- 1.0 (stat) -3.4 +/- 1.1 (stat)
OUT 2654 85.2 +/- 1.1 (stat) 2.4 +/- 1.2 (stat)
IN 2655 -85.6 +/- 1.0 (stat) -3.0 +/- 1.0 (stat)
AVERAGE 85.7 +/- 0.5 (stat) 2.4 +/- 0.6 (stat)
IHWP FILE P_V P_X
OUT 16Feb12 08:30:41 86.9 +/- 1.1 (stat) 1.6 +/- 1.2 (stat)
IN 16Feb12 08:36:18 -87.3 +/- 1.1 (stat) -2.7 +/- 1.1 (stat)
OUT 16Feb12 08:54:56 86.3 +/- 1.1 (stat) 2.7 +/- 1.2 (stat)
IN 16Feb12 08:54:56 -85.7 +/- 1.1 (stat) -4.1 +/- 1.1 (stat)
AVERAGE 86.6 +/- 0.6 (stat) 2.8 +/- 0.6 (stat)
Data with Old DAQData with Old DAQ
Simultaneous Data with New DAQSimultaneous Data with New DAQ
Difference due to different energy cuts
in analysis
Difference due to different energy cuts
in analysis
![Page 36: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/36.jpg)
BFM
BFM after-140 mV offset
BFM after-145 mV discrimination (TDC)
BFM in FADC
BFM Signal
![Page 37: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/37.jpg)
Hall C 499 MHz Beam
![Page 38: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/38.jpg)
499 MHz Beam
Hall B and Hall C beams at 499 MHzHall B and Hall C beams at 499 MHz
![Page 39: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/39.jpg)
5 MeV Mott Beam-line
Beam
Dump12 ns
![Page 40: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/40.jpg)
31 MHz Beam
Here, “E” is the peak of Energy signal (sample=74)Here, “E” is the peak of Energy signal (sample=74)
![Page 41: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/41.jpg)
Mott Sweep Magnet at -5A
![Page 42: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/42.jpg)
![Page 43: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/43.jpg)
![Page 44: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/44.jpg)
Mott Sweep Magnet at 0A
![Page 45: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/45.jpg)
![Page 46: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/46.jpg)
![Page 47: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/47.jpg)
Mott Sweep Magnet at +5A
![Page 48: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/48.jpg)
![Page 49: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/49.jpg)
![Page 50: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/50.jpg)
Spectra from Broken Gold Mott TargetMainly Dump EventsMainly Dump Events
![Page 51: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/51.jpg)
Mott Asymmetry vs. Energy
Select events coming from target with a cut on TOFSelect events coming from target with a cut on TOF
![Page 52: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/52.jpg)
Mott Asymmetry vs. TOF
Asymmetry of Dump Event = 0Asymmetry of Dump Event = 0
Asymmetry of Chamber Event = 0Asymmetry of Chamber Event = 0
![Page 53: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/53.jpg)
S1 Chan Signal
0 BCM0L02 OUTPUT 2
1 40 MHz Clock
2 Mott Trigger
3
4
5
6
7
8
9
10
11
12 BCM0L02 OUTPUT 3
13 Delayed Helicity
14 Battery
15
S1 Chan
BPM Signal
16 0L01 X+
17 0L01 X-
18 0L01 Y+
19 0L01 Y-
20 0L02 X+
21 0L02 X-
22 0L02 Y+
23 0L02 Y-
24 5D00 X+
25 5D00 X-
26 5D00 Y+
27 5D00 Y-
28 5D01 X+
29 5D01 X-
30 5D01 Y+
31 5D01 Y-
BPM 0L01
BPM 0L02
BPM 5D00
BPM 5D01
Scaler Readout
![Page 54: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/54.jpg)
BCM0L02 Readout
I. BCM0L02 Receiver output:
1. OUTPUT 2: 0.0 – 1.0 µA → 0 – 10 V
II. Connected to VtoF (1 MHz, 10 V)
III. Charge Asymmetry Test:
o PITA (Polarization Induced Transport Asymmetry): charge asymmetry depends on Pockels Cell HV
o Experiments use PITA to implement charge feedbacko Measure PITA slope at PITA = -2000, 0, +2000 DAC (Nominal 40000 –
about 2.9 kV on Pockels Cell).
o Measure PITA Slope for beam currents: 1.0 µA, 0.108 µA, 0.040 µA, and 0.008 µA
![Page 55: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/55.jpg)
Charge Asymmetry Test
30 Hz 8-window Delay Quartet
![Page 56: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/56.jpg)
PITA Scan at 1.0 µA
Mott DAQI. PITA Slope = -1.02 ppm/DAC
QWeak DAQI. PITA Slope = -1.06 ppm/DAC
![Page 57: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/57.jpg)
PITA at 0.108 µA
Mott DAQI. PITA Slope = -1.05 ppm/DAC
![Page 58: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/58.jpg)
PITA at 0.040 µA
Mott DAQI. PITA Slope = -1.1 ppm/DAC
![Page 59: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/59.jpg)
PITA at 0.008 µA
Overwhelmed by noise
Now, reliable charge asymmetry measurement to about 40 nA
Next, …
I. Calibrate Receiver to 100 nA FSII. Measure charge asymmetry at 10 nAIII. For even lower beam currents:
I. Measure charge asymmetry at 1 µA and at 10 nA → We know charge asymmetry at all currents
![Page 60: 5 MeV Mott Polarimeter Progress](https://reader035.fdocuments.in/reader035/viewer/2022062309/568151a2550346895dbfd2e4/html5/thumbnails/60.jpg)
Precision Mott Electron Polarimeter• Goal: Measure and publish electron polarization over range of 2 to 8 MeV
with absolute uncertainty better than 1%.
– Fast statistical measurements of experiment asymmetry (A) <0.5% vs. target thickness– Demonstrate instrumental false asymmetry <0.3%
• Eliminate background events with Time-of-Flight (ToF) cut• Super-ratio calculation (double arm + helicity reversal)
– New Data Acquisition system based on Hall D 250MHz FADC– Calculate the effective Sherman Function (S) < 0.8%
S
AP
Effective Sherman Function Error (∆S/S)
Nuclear Corrections 0.5%
Target Thickness Extrapolation 0.6%
Scattering angle (0.1°) 0.1%
Electron Energy (0.5%) 0.1%
Plan
I. Theoretical support for Sherman function calculation
II. Measurements are collected “for free” parasitic to PEPPo in May/June, 2012
III. GEANT4 Simulation – No Extrapolation