45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

download 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

of 28

Transcript of 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    1/28

    Series 45 Frame F

    Open Circuit AxialPiston Pumps

    Service Manual

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    2/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Revisions

    2007 Sauer-Danoss. All rights reserved.

    Sauer-Danoss accepts no responsibility or possible errors in catalogs, brochures and other printed material.

    Sauer-Danoss reserves the right to alter its products without prior notice. This also applies to products

    already ordered provided that such alterations arent in conict with agreed specifcations. All trademarks in

    this material are properties o their respective owners. Sauer-Danoss and the Sauer-Danoss logotype are

    trademarks o the Sauer-Danoss Group.

    Front cover illustrations: F101 503, P105265E

    Table o Revisions

    D P Chd R.

    September 2007 28 revision to change per turn or displacement limiter AB

    November 2006 - First edition AA

    HistoRy o Revisions

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    3/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    intRoDuCtion

    teCHniCal

    sPeCiiCations

    eatuRes

    PRessuRe

    measuRement

    initial staRt-uP

    PRoCeDuRes

    luiD anD ilteR

    maintenanCe

    Overview ...........................................................................................................................................................5

    Saety precautions .........................................................................................................................................5

    Unintended machine movement .......................................................................................................5Flammable cleaning solvents ...............................................................................................................5

    Fluid under pressure ................................................................................................................................5

    Personal saety ...........................................................................................................................................5

    Symbols used in Sauer-Danoss literature.............................................................................................6

    General description ....................................................................................................................................... 7

    System circuit ...................................................................................................................................................8

    General specifcations ..................................................................................................................................9

    Type o mounting ..................................................................................................................................... 9

    Auxiliary mounting pad options ......................................................................................................... 9

    Control options ..........................................................................................................................................9Port options ................................................................................................................................................9

    Direction o rotation ................................................................................................................................9

    Installation position .................................................................................................................................9

    Technical specifcations ..........................................................................................................................9

    Hydraulic parameters..................................................................................................................................10

    Inlet pressure ............................................................................................................................................10

    Pressure compensator valve setting ................................................................................................10

    Case pressure ...........................................................................................................................................10

    Hydraulic uid .........................................................................................................................................10

    Temperature range.................................................................................................................................11

    Fluid viscosity ...........................................................................................................................................11

    Filtration ....................................................................................................................................................11

    Displacement limiter ...................................................................................................................................12

    Auxiliary mounting pads ...........................................................................................................................12

    Input shats .....................................................................................................................................................12

    Control options .............................................................................................................................................13

    Operation ...................................................................................................................................................13

    Control options .............................................................................................................................................14

    Required tools ...............................................................................................................................................15

    Port locations and gauge installation ...................................................................................................15

    General .............................................................................................................................................................16

    Start-up procedure ......................................................................................................................................16

    Recommendations.......................................................................................................................................17

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    4/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    tRoublesHooting

    aDjustments

    minoR RePaiR

    Excessive noise and / or vibration ..........................................................................................................18

    Actuator response is sluggish ..................................................................................................................18

    System operating hot .................................................................................................................................19Low pump output ow ..............................................................................................................................19

    Pressure or ow instability ........................................................................................................................19

    System pressure not reaching PC setting ............................................................................................20

    High inlet vacuum ........................................................................................................................................20

    PC control ........................................................................................................................................................21

    LS control ........................................................................................................................................................22

    Shat seal replacement ...............................................................................................................................23

    Removal ......................................................................................................................................................23Installation .................................................................................................................................................23

    Auxiliary pads ................................................................................................................................................24

    Removal ......................................................................................................................................................24

    Installation .................................................................................................................................................24

    Control..............................................................................................................................................................25

    Disassembly ..............................................................................................................................................25

    Inspection ..................................................................................................................................................25

    Reassembly ...............................................................................................................................................26

    Plug and ftting sizes and torques ..........................................................................................................27

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    5/28

    511005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Introduction

    This manual includes inormation or the installation, maintenance, and minor repair o

    the Series 45 Frame F open circuit axial piston pumps. The manual includes a description

    o the units and their individual components, troubleshooting inormation, and minorrepair procedures. Perorming installation, maintenance, and minor repair o Series 45

    Frame F axial piston pumps according to the procedures in this manual will not aect

    your warranty.

    Perorming minor repairs requires the unit to be removed rom the vehicle/machine.

    Thoroughly clean the unit beore beginning maintenance, or repair activities. Since dirt

    and contamination are the greatest enemies o any type o hydraulic equipment, ollow

    cleanliness requirements strictly. This is especially important when changing the system

    flter and when removing hoses or plumbing.

    A worldwide Global Service Partner Network is available or major repairs. Major repairs

    require the removal o the units endcap, which voids the warranty unless done by aGlobal Service Partner. Sauer-Danoss Global Service Partners are trained by the actory

    and certifed on a regular basis. You can locate your nearest Global Service Partner using

    the distributor locator at www.sauer-danoss.com

    Always consider saety precautions beore beginning a service procedure. Protect

    yoursel and others rom injury. Take these general precautions whenever servicing a

    hydraulic system.

    udd ch

    WarningUnintended movement o the machine or mechanism may cause injury to the technician

    or bystanders. To protect against unintended movement, secure the machine or

    disable / disconnect the mechanism while servicing.

    c

    WarningSome cleaning solvents are ammable. To avoid possible fre, do not use cleaning

    solvents in an area where a source o ignition may be present.

    d dr prr

    WarningEscaping hydraulic uid under pressure can have sufcient orce to penetrate your skin

    causing serious injury and/or inection. This uid may also be hot enough to cause burns.

    Use caution when dealing with hydraulic uid under pressure. Relieve pressure in thesystem beore removing hoses, fttings, gauges, or components. Never use your hand

    or any other body part to check or leaks in a pressurized line. Seek medical attention

    immediately i you are cut by hydraulic uid.

    Pr

    WarningProtect yoursel rom injury. Use proper saety equipment, including saety glasses, at all

    times.

    oveRview

    saety PReCautions

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    6/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Introduction

    symbols useD in

    saueR-Danoss

    liteRatuRe

    These symbols are in the illustrations and text o this manual. They communicate helpul

    inormation at the point where it is most useul to the reader.

    In most instances, the appearance o the symbol itsel denotes its meaning. The legend

    below defnes the symbol and explains its purpose.

    WARNING may result in injury

    CAUTION may result in damage toproduct or property

    Reusable part

    Non-reusable part, use a new part

    Non-removable item

    Option either part may exist

    Superseded parts are notinterchangeable

    Measurement required

    Flatness specifcation

    Parallelism specifcation

    External hex head

    Internal hex head

    Torx head

    O-ring boss port

    Tip, helpul suggestion

    Lubricate with hydraulic uid

    Apply grease / petroleum jelly

    Apply locking compound

    Inspect or wear or damage

    Clean area or part

    Be careul not to scratch or damage

    Note correct orientation

    Mark orientation or reinstallation

    Torque specifcation

    Press in press ft

    Pull out with tool press ft

    Cover splines with installationsleeve

    Pressure measurement / gaugelocation or specifcation

    waRRanty Perorming installation, maintenance, and minor repairs according to the procedures in

    this manual will not aect your warranty. Major repairs requiring the removal o a units

    endcap voids the warranty unless done by a Sauer-Danoss Global Service Partner.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    7/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Introduction

    Sauer-Danoss Series 45 Frame F open circuit piston pumps convert input torque into

    hydraulic power. The input shat transmits rotational orce rom the prime mover to

    the cylinder block. Tapered roller bearings support the shat at the ront and rear o thepump. The cylinder block connects to the input shat through splines on the shat. A

    lip-seal at the ront end o the pump prevents leakage where the shat exits the pump

    housing. The spinning cylinder block contains nine reciprocating pistons. Each piston

    has a brass slipper connected at one end by a ball joint. The slippers are held to the

    swashplate by the spring retainer and block spring. The block spring also holds the

    cylinder block to the valve plate. The reciprocating movement o the pistons occurs

    as the slippers slide against the inclined swashplate during rotation. The valve plate

    connects one hal o the cylinder block to pump inlet and the other hal to pump outlet.

    As each piston cycles in and out o its bore, uid is drawn rom the inlet and displaced to

    the outlet thereby imparting power into the system circuit. A small amount o uid leaks

    rom the cylinder block / valve plate and slipper / swashplate interaces or lubrication

    and cooling. Case drain ports return this uid to the reservoir.

    The angle o the swashplate controls the volume o uid displaced into the system

    circuit. The bias piston orces the swashplate into an inclined position (into stroke). The

    servo piston opposes the action o the bias piston and spring, orcing the swashplate out

    o stroke when hydraulic pressure in the control circuit rises above the spring orce.

    The pump control, by varying the pressure at the servo piston, controls the displacement

    o uid in the system circuit. Controls designed or Pressure Compensation (PC) or

    load sensing (LS) are available. For a detailed description o control operation, reer to

    Control op tions, operation, page 13.

    geneRal DesCRiPtion

    Pump and control sectional view

    Slipper

    LS adjustment

    PC adjustment

    LS spool

    PC spool

    LS control(attached to endcap)

    Servo piston

    Swashplate

    Taperedrollerbearing

    Shaft seal

    Tapered rollerbearing

    Input shaft

    Cylinderblock

    Valveplate

    Cylinderblockspring

    Piston

    Slipper retainer

    Bias spring

    P104 341EBias piston

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    8/28

    8 11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Introduction

    1 Full available ow is a unction o pump displacement, operating speed, and efciency. Reer to Series 45 Frame

    F Axial Piston Open Circuit Pumps Technical Inormation,50l098or details.

    system CiRCuit

    Pictorial circuit diagram

    System pressure

    Servo pressure

    Actuator pressure

    Load sense pressure

    Actuator return

    Suction / case drain /system return

    Series 45 F Frameopen circuit axialpiston pump withload sensing control PVG 32

    multi-sectionloadsensingcontrol

    valve

    P106 212E

    Reservoir Filter

    Heat exchanger

    Double-acting cylinder

    Bi-directionalgear motor

    The pump receives uid directly rom the reservoir through the inlet line. A screen in the

    inlet line protects the pump rom large contaminants. The pump outlet eeds a PVG-32

    multi-section, load sensing, directional control valve. The PVG valve directs pump ow tothe cylinder and gear motor. A heat exchanger cools the uid returning rom the valve. A

    flter cleans the uid beore it returns to the reservoir.

    Flow in the circuit determines the speed o the actuators. The position o the PVG valve

    determines the ow demand. A hydraulic pressure signal (LS signal) communicates

    demand to the pump control. The pump control monitors the pressure dierential

    between pump outlet and the LS signal, and regulates servo pressure to control the

    swashplate angle. Swashplate angle determines pump ow.

    Actuator load determines system pressure. The pump control monitors system pressure

    and will decrease the swashplate angle to reduce ow i system pressure reaches the

    PC setting. A system relie valve in the PVG valve acts as a back-up to control systempressure.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    9/28

    911005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Technical specifcations

    tp

    SAE-C mounting ange.

    ar pd p

    SAE-A, SAE-B, SAE-B-B, SAE-C, SAE C-C

    Cr p

    PC: Pressure Compensator

    LS: Load Sensing (with PC)

    Pr p

    Inlet and system ports: SAE anged ports, code 61 Inlet Code 62 outlet.

    Axial (end) ports or radial (side) ports.

    All other ports: SAE straight thread O-ring boss.

    Drc r

    Clockwise or counterclockwise.

    i p

    Installation position is discretionary. To satisy inlet pressure conditions, we recommend

    you locate the pump below the lowest level o hydraulic uid in the reservoir. The

    housing must always be flled with hydraulic uid.

    tchc pcc

    geneRal

    sPeCiiCations

    Features and options md

    r u b 90C

    Maximum displacement cm [in] 74 [4.52] 90 [5.49]

    Flow at rated speed (theoretical) l/min

    [US gal/min]

    177.6

    [46.9]

    198

    [52.3]

    Input torque at maximum displacement

    (theoretical)

    Nm/bar

    [lbin/1000 psi]

    1.178

    [719]

    1.432

    [874]

    Mass moment o inertia o internal

    rotating components

    kgm

    [slugt]

    0.00630

    [0.00465]

    0.00650

    [0.00480]

    Weight Axial ports kg [lb] 29 [63]

    Radial ports 31.8 [70.1]

    Rotation Clockwise, Counterclockwise

    Mounting SAE B 2 blot, SAE-C 4 bolt

    Auxiliary mounting SAE-A, SAE-B, SAE-BB, SAE-CSystem ports (type) 4-bolt split ange

    System ports (location) Axial, Radial

    Control types PC, Remote PC, LS, Electric

    Shats Splined 14 tooth, 17 tooth

    Straight 31.75 mm [1.25 in]

    Displacement limiters Optional, adjustable

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    10/28

    10 11005158 Rev. AB September 2007

    geneRal

    sPeCiiCations

    (cd)

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Technical specifcations

    i prr

    Minimum pressure, continuous = 0.8 bar absolute [6.7 inch Hg vacuum]

    (at reduced maximum pump speed)

    Minimum pressure, cold start = 0.5 bar absolute [15.1 inch Hg vacuum]

    Prr cpr

    HyDRauliC

    PaRameteRs

    C prr

    Maximum continuous: 0.5 bar [7 psi] above inlet

    Intermittent: 2 bar [29 psi] cold start

    Hdrc fdReer to Sauer-Danoss publication: Fluids and Filtration50l0. For inormation on

    biodegradable uids reer to Biodegradable Hydraulic Fluids,50l05. See Fluid and

    lter maintenance, page 17 or recommended uid and flter change intervals.

    Ratings md

    R u gb g90C

    Input speed1 minimum min-1 (rpm) 500 500

    continuous 2400 2200

    maximum 2800 2600

    Working pressure continuous bar [psi] 310 [4495] 260 [3770]

    maximum 400 [5800] 350 [5075]

    External shat

    loads

    External moment (Me) Nm [lbin] 300 [2655]

    Thrust in (Tin), out (Tout) N [lb ] 2900 [650]

    Bearing lie at 140 bar [2030 psi] B10 hours 31 300 17 400

    at 210 bar [3045 psi] 1110 5700

    at 260 bar [3770 psi] 4810 2490

    at 310 bar [4495 psi] 3080 1636

    Mounting ange

    load moments

    Vibratory (continuous) Nm [lbin] 3730 [33 000]

    Shock (maximum) 13 220 [117 000]

    1. Input speeds are valid at 1 bar absolute [0 in Hg vac] inlet pressure. See Inlet pressure vs. speedcharts.

    PC control setting range

    md r p

    b 100310 14504495

    90C 100260 14503770

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    11/28

    1111005158 Rev. AB September 2007

    HyDRauliC

    PaRameteRs

    (cd)

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Technical specifcations

    1 Hydraulic uid viscosity must be maintained within the prescribed limits.2 As measured at the hottest point in the system, e.g. drain line.

    r

    Required cleanliness level: ISO 4406 Class 18/13 or better. Reer to Sauer-Danoss

    publications Fluids and Filtration50l0 and Design Guidelines or Selecting and

    Maintaining the Required Hydraulic Fluid Cleanliness50l05. See Fluid and lter

    maintenance, page 17 or recommended uid and flter change intervals.

    tprr r

    d c

    Fluid viscosity limits

    Cd / (cs) sus

    min.continuous 9 58

    intermittent 6.4 47

    max.

    continuous 110 500

    intermittent(cold start)

    1000 4700

    Temperature limitsm

    (r, cd r)- 40 C [- 40 F]

    C 82 C [180 F]

    m 104 C [220 F]

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    12/28

    1 11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Features

    DisPlaCement limiteR

    auxiliaRy mounting

    PaDs

    inPut sHats Series 45 F Frame pumps are available with a variety o splined and straight keyed

    shats. For inormation on shats reer to Series 45 F Frame Axial Piston Open Circuit Pumps

    Technical Inormation50l098.

    Frame F Series 45 pumps are available

    with an optional adjustable maximum

    displacement limiter. The adjustablestop limits the pumps maximum

    displacement. The displacement change

    per turn is 6.8 cm/turn [0.47 in/turn] or

    the F90C and 6.0 cm/turn [6.1 in/turn]

    or the F74B

    Displacement limiter

    P104 345

    Auxiliary mounting pads are available or

    all radial ported Series 45 pumps. These

    pads are typically used or mountingauxiliary hydraulic pumps.

    Since the auxiliary pad operates under

    case pressure, you must use an O-ring to

    seal the auxiliary pump mounting ange

    to the pad. Oil rom the main pump case

    lubricates the drive coupling. For details

    reer to Series 45 F Frame Axial Piston

    Open Circuit Pumps Technical Inormation

    50l098.

    Auxiliary pad options

    P106 262E

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    13/28

    111005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Features

    The Series 45 Frame F pumps have two control options, a load sensing (LS) control with

    Pressure Compensator (PC) or a PC only control.

    b pr d r p hp

    p

    Cross-section pumpopr

    General

    The bias piston and spring (1&2) acts at

    all times to push the swashplate (4) to

    maximum angle causing the pump to

    stroke. The servo piston (3) acts against

    the bias piston and spring to reduce the

    swashplate angle causing the pump to

    destroke. Swashplate angle determines

    pump outlet ow. The pump control,

    depending on conditions in the system

    circuit, sets swashplate angle by meteringsystem pressure to the servo piston.

    PC control

    The PC control design maintains a

    constant pressure in the hydraulic circuit

    as ow varies. The PC control modulates

    pump ow accordingly to maintain

    system pressure at the PC setting as

    the PC adjusting plug (4) and spring (5)

    defne.

    When system pressure, acting on the

    non-spring end o the PC spool (6),

    overcomes the orce o the PC spring (5),

    the spool shits porting system pressure to the servo piston and the swashplate angle

    decreases. When system pressure drops below the PC setting, the PC spring shits the

    spool in the opposite direction connecting the servo piston to pump case and the

    swashplate angle increases. The swashplate is maintained at whatever angle is required

    to keep system pressure at the PC setting.

    ContRol oPtions

    PC spool shifts to port system pressure to

    servo piston

    Cross-section PC control

    P104 049

    4

    56

    P106 261E1

    234

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    14/28

    1 11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Features

    ContRol oPtions

    (cd)

    LS control

    The LS control design matches pump

    ow with system demand. The LS controlsenses the ow demand o the system

    as a pressure drop across the external

    Control valve (ECV). As the ECV opens

    and closes, the pressure delta across

    the valve changes. When opening, the

    delta decreases. When closing, the delta

    increases. The LS control then increases

    or decreases pump ow to the system

    until the pressure delta becomes equal

    to the LS setting as defned by the LS

    adjusting plug (7) and spring (8).

    Load

    pressure

    Load sensepressure

    Systempressure

    Tank

    Return

    pressure

    P101 665E

    Typical load-sensing control valve

    Prr drp cr r cr

    d dd

    The LS control consists o two spool

    valves that connect the servo pistoneither to pump case or system pressure.

    The PC spool (6) controls the pressure-

    compensating unction o the control

    as previously described. The LS spool (9)

    controls the load-sensing unction. The

    PC spool has priority over the LS spool.

    Through internal porting, system pressure

    (upstream o ECV) is applied to the non-

    spring end o the LS spool, and through

    hydraulic line connected at port X, LS

    pressure (downstream o ECV) is applied

    to the spring end. This arrangement allows the LS spool to act on the delta between system

    pressure and LS pressure. The LS spring sets the threshold o operation (LS setting).

    Because the swashplate is biased to maximum angle, the pump attempts to deliver

    ull ow to the hydraulic system. When the ow being delivered exceeds demand, the

    pressure delta across the ECV is great enough to overcome spring orce and shit the

    LS spool porting system pressure to the servo piston. The pump de-strokes reducing

    ow until the delta across the ECV becomes equal to the LS setting. When ow being

    delivered is less than demand, the delta across the ECV drops below the LS setting and

    the LS spring shits the spool connecting the servo piston to pump case. The pump

    strokes increasing ow until the delta across the ECV becomes equal to the LS setting.

    When the external control valve is placed in neutral, it connects the LS signal line todrain. With no LS pressure acting on the non-spring end o the LS spool, the pump

    adjusts stroke to whatever position necessary to maintain system pressure at the LS

    setting. The pump is now in standby mode.

    Because o the series arrangement o the LS and PC spools, the PC spool will override

    the LS spool. I at any time system pressure reaches the PC setting, the PC spool will shit

    blocking the passage that connects the LS spool with the servo piston, thus porting

    system pressure to the servo piston causing the pump to destroke.

    ls p h pr prr r

    p

    Cross-section LS control

    Spool

    Spring

    guide

    Spring

    7LS adjustingplug

    LS spring

    PC adjusting

    plugPlug

    Plug

    Spool Spring

    guide

    Spring

    (PC heavy)

    P104 051

    6

    89

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    15/28

    1511005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Pressure measurement

    The service procedures described in this manual can be perormed using common

    mechanics hand tools. Special tools, i required are shown. Calibrate pressure gauges

    requently to ensure accuracy. Use snubbers to protect gauges.

    The illustration below shows gauge port locations. Recommended pressure gauges and

    fttings are in the table.

    Gauge port locations

    P106 263E

    LegendB = Main pressure lineS = Suction lineL1, L2 = Case drain linesX = Load sensing pressure portM2 = Gauge port for port BM4 = Gauge port servo pressure

    X

    M2

    L1

    L2

    SB

    M4

    RequiReD tools

    PoRt loCations anD

    gauge installation

    Gauge and port inormation

    Pr Prp R

    M2 System pressure 0-300 bar [0-5000 psi] 9/16 - 18 O-ring ftting

    M4 Servo pressure 0-300 bar [0-5000 psi] 9/16 - 18 O-ring ftting

    L1, L2 Case pressure 0-10 bar [0-100 psi] 7/8 - 14 O-ring ftting

    X1 LS signal 0-300 bar [0-5000 psi]7/16 - 20 O-ring ftting

    (tee into LS signal line)

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    16/28

    1 11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Initial start-up procedures

    CautionIncorrect shat alignment

    may result in damage

    to drive shat, bearings,

    or seal which can cause

    external oil leakage.

    geneRal

    staRt-uP PRoCeDuRe 1. Connect the pump to the prime mover. Ensure that pump shat is properly aligned

    with the shat o the prime mover. 2. Fill the main pump housing with clean fltered hydraulic uid. Pour fltered oil

    directly into the upper most case drain port.

    3. To ensure the pump stays flled with oil, install the case drain line in the upper most

    case drain port.

    4. Fill the inlet line leading rom the pump to the reservoir. Check the inlet line or

    properly tightened fttings. Be certain it is ree o restrictions and air leaks.

    5. Fill the reservoir with recommended hydraulic uid. Always flter uid through a 10

    micron flter pouring into the reservoir. Never reuse hydraulic uid.

    6. Install a gauge at port or M2 to monitor system pressure during start up.

    Follow recommendations in the vehicle / machine operators manual or prime mover

    start up procedures.

    7. While watching the pressure gauge installed at M2, jog the prime mover or run at

    the lowest possible speed until system pressure builds to normal levels (minimum

    11 bar [160 psi]). Once system pressure is established, increase to ull operating

    speed. I system pressure is not maintained, shut down the prime mover, determine

    cause, and take corrective action. Reer to Troubleshooting, page 18.

    8. Operate the hydraulic system or at least fteen minutes under light load conditions.

    9. Check and adjust control settings as necessary ater installation. Reer to

    Adjustments, page 21.

    10. Shut down the prime mover and remove the pressure gauge. Replace plug at port M2.

    11. Check the uid level in the reservoir; add clean fltered uid i necessary.

    The pump is now ready or operation.

    Follow this procedure when starting-up a new Series 45 installation or when restarting

    an installation in which the pump has been removed.

    WarningUnintended movement o the machine or mechanism may cause injury to the technician

    or bystanders. To protect against unintended movement, secure the machine or disable/

    disconnect the mechanism while servicing.

    Prior to installing the pump, inspect or damage incurred during shipping. Make certain

    all system components (reservoir, hoses, valves, fttings, heat exchanger, etc.) are clean

    prior to flling with uid.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    17/28

    111005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Fluid and flter maintenance

    To ensure optimum lie o Series 45 products, perorm regular maintenance o the uid

    and flter. Contaminated uid is the main cause o unit ailure. Take care to maintain uid

    cleanliness when servicing.

    Check the reservoir daily or proper uid level, the presence o water, and rancid uid

    odor. Water in the uid may be noted by a cloudy or milky appearance or ree water

    in the bottom o the reservoir. Rancid odor indicates the uid has been exposed to

    excessive heat. Change the uid immediately i these conditions occur. Correct the

    problem immediately.

    ReCommenDations

    Change the uid and flter per the

    vehicle / machine manuacturers

    recommendations or at these intervals:

    Change the uid more requently i itbecomes contaminated with oreign

    Fluid and lter change interval

    Rrr p m ch r

    Sealed 2000 hours

    Breather 500 hours

    matter (dirt, water, grease, etc.) or i the uid is subjected to temperature levels greater

    than the recommended maximum.

    Dispose o used hydraulic uid properly. Never reuse hydraulic uid.

    Change flters whenever the uid is changed or when the flter indicator shows that it is

    necessary to change the flter. Replace all uid lost during flter change.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    18/28

    18 11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Troubleshooting

    exCessive noise anD /

    oR vibRation

    aCtuatoR ResPonse is

    sluggisH

    i Dcrp ac

    Check uid level in reservoir.Insufcient hydraulic uid causes

    cavitation.Fill the reservoir to proper level.

    Check or air in system.Air in system causes noisy, erratic

    control.

    Purge air and tighten fttings.

    Check inlet or leaks.

    Check pump inlet pressure /

    vacuum.

    Improper inlet conditions cause

    erratic behavior and low output

    ow.

    Correct pump inlet pressure /

    vacuum conditions. Reer to

    Hydraulic parameters, page 10.

    Inspect shat couplings.

    A loose or incorrect shat coupling

    causes excessive noise and / or

    vibration.

    Repair or replace coupling and

    ensure that correct coupling is

    used.

    Check shat alignment.Misaligned shats create excessive

    noise and / or vibration.Correct shat misalignment.

    Hydraulic uid viscosity above

    acceptable limits.

    Hydraulic uid viscosity above

    acceptable limits or low uid

    temperature will not allow the

    pump to fll or control to operate

    properly.

    Allow system to warm up beore

    operating, or use uid with the

    appropriate viscosity grade or

    expected operating temperatures.

    See Hydraulic Fluids, Series 45

    F Frame Technical Inormation

    Manual,50l098.

    i Dcrp ac

    Check external system relie valve

    setting.

    Low external relie valve setting

    slows down system.

    Adjust external relie valve

    setting ollowing manuacturers

    recommendations. External relie

    setting must be above PC setting

    to operate properly.

    Check PC and LS control setting.

    Low PC setting prevents the pump

    rom achieving ull stroke. Low LS

    setting limits output ow.

    Adjust PC and LS setting. Reer to

    Adjustments, pages 21 and 22.

    Check LS control signal pressures.Incorrect LS signal will not allow

    pump to operate correctly.

    Inspect system to ensure that

    proper LS signal transmit to pump.

    Internal system leaks.Worn internal parts dont allow the

    pump to operate properly.

    Reer to Sauer-Danoss Global

    Service Partner or required repair.

    Hydraulic uid viscosity above

    acceptable limits.

    Hydraulic uid viscosity above

    acceptable limits or low uid

    temperature will not allow the

    pump to fll or control to operateproperly.

    Allow system to warm up beore

    operation or sue uid with the

    appropriate viscosity grade or

    expected operating temperatures.

    See Hydraulic Fluids, Series 45F Frame Technical Inormation

    Manual,50l098.

    Check external system valves.Malunctioning valves may not

    allow system to respond properly.

    Repair or replace system valving as

    required.

    Check pump case pressure.High case pressure causes the

    system to be sluggish.Correct case drain line restrictions.

    Check pump inlet pressure /

    vacuum.

    High inlet vacuum causes low

    output ow.Correct inlet pressure conditions.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    19/28

    1911005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Troubleshooting

    system oPeRating Hot

    low PumP outPut

    low

    PRessuRe oR low

    instability

    i Dcrp ac

    Check uid level in reservoir.

    Insufcient volume o hydraulic

    uid will not meet cooling

    demands o system.

    Fill reservoir to proper level. Veriy

    proper size o reservoir.

    Inspect heat exchanger. Check air

    ow and input air temperature or

    the heat exchanger.

    Insufcient air ow, high input air

    temperature, or undersized heat

    exchangers will not meet cooling

    demands o the system.

    Clean, repair, or replace heat

    exchanger as required. Veriy

    proper size o heat exchanger.

    Check external system relie valve

    setting.

    Fluid passing through relie valve

    adds heat to system.

    Adjust external system relie valve

    setting ollowing manuacturers

    recommendations. External relie

    valve setting must be above PC

    setting or proper operation.

    Check pump inlet pressure /

    vacuum.

    High inlet vacuum adds heat to

    system.

    Correct inlet pressure / vacuum

    conditions.

    i Dcrp ac

    Check uid level in reservoir.

    Insufcient hydraulic uid will limit

    output ow and cause internal

    damage to pump.

    Fill the reservoir to proper level.

    Hydraulic uid viscosity above

    acceptable limits.

    Fluid viscosity above acceptable

    limits or low uid temperature

    will not allow the pump to fll or

    control to operate properly.

    Allow system to warm up beore

    operating, or use uid with the

    appropriate viscosity grade or

    expected operating temperatures.

    See Hydraulic fuids, Series 45

    F Frame Technical Inormation

    Manual,50l098.

    Check external system relie valve

    setting.

    Eternal relie valve set below PC

    setting causes low output ow.

    Adjust external relie valve

    ollowing manuacturers

    recommendation. External relie

    valve setting must be above PC

    setting to operate properly.

    Check PC and LS control setting.Low PC setting prevents the pump

    rom achieving ull stroke.

    Adjust PC and LS setting. Reer to

    Adjustment, page 21.

    Check pump inlet pressure /

    vacuum.

    High inlet vacuum causes low

    output ow.Correct inlet pressure conditions.

    Check input speed. Low input speeds decrease ow. Adjust input speed.

    Check pump rotation.Incorrect rotational confguration

    causes low ow.

    Use pump with appropriate

    rotational confguration.

    i Dcrp ac

    Check or air in system.Air in system causes erratic

    operation.

    Activate PC allowing system

    to bleed air. Check inlet line or

    leaks and eliminate source o air

    ingression.

    Check control spools.Sticking control spools cause

    erratic operation.

    Inspect spools or ree movement

    in bore. Clean or replace.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    20/28

    0 11005158 Rev. AB September 2007

    PRessuRe oR low

    instability

    (cd)

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Troubleshooting

    system PRessuRe not

    ReaCHing PC setting

    HigH inlet vaCuum

    CautionHigh inlet vacuum causes

    cavitation which candamage internal pump

    components.

    i Dcrp ac

    Check LS setting.Low LS setting may cause

    instability.

    Adjust LS setting to proper level.

    SeeAdjustments, page 21.

    Check LS signal line.Blocked LS signal line intereres

    with proper LS operation.Remove blockage.

    Check external relie valve and PC

    setting.

    Insufcient pressure dierential

    between PC setting and external

    relie valve.

    Adjust external relie valve or PC

    control settings to appropriate

    level. Relie valve setting must

    be above PC setting to operate

    properly.

    Check external relie valve.

    Chattering external relie valve

    may cause unstable eedback to

    pump control.

    Adjust or replace relie valve.

    i Dcrp ac

    Check PC control setting.System pressure will not rise above

    PC setting.

    Adjust PC to appropriate setting.

    Reer toAdjustments, page 21.

    Check external relie valve.

    External relie valve setting below

    PC setting prevents pressure

    compensation.

    Adjust external relie valve

    according to manuacturers

    recommendations. External relie

    valve must be set above PC setting

    to operate properly.

    Inspect PC control spring.Broken, damaged, or missing

    spring will cause erratic operation.Replace the spring as required.

    Inspect PC spool or wear.Wear o PC spool causes internal

    leakage in the control.Replace the spool as required.

    Inspect PC spool or proper

    orientation.

    Improper orientation results in

    poor operation.Correct orientation o spool.

    Check PC control or

    contamination.

    Contamination may interere with

    movement o the PC spool.

    Clean PC control components, take

    appropriate action to eliminate

    contamination.

    i Dcrp ac

    Check uid temperature.

    Low temperature increases

    viscosity. High uid viscosity

    causes high inlet vacuum.

    Allow system to warm up beore

    operating.

    Inspect inlet screen. Blocked or restricted inlet screencauses high inlet vacuum.

    Clean screen / remove blockage.

    Check inlet piping.Too many fttings, bends, or long

    piping causes high inlet vacuum.

    Eliminate fttings to make path

    more direct.

    Hydraulic uid viscosity above

    acceptable limits.

    High uid viscosity causes high

    inlet vacuum.

    Select uid with appropriate

    viscosity or expected operating

    temperature. See Hydraulic

    fuids, Series 45 F Frame Technical

    Inormation Manual, 50l098.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    21/28

    111005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Adjustments

    PC ContRol PC setting is indicated in the pump model code. Reer to the Series 45 Frame F Open

    Circuit Axial Piston Pumps Technical Inormation Manual,50l098, or more inormation.

    Beore perorming adjustments, read page 15, Pressure measurement.WarningEscaping hydraulic uid

    under pressure can

    have sufcient orce

    to penetrate your skin

    causing serious injury

    and/or inection. Relieve

    pressure in the system

    beore removing hoses,

    fttings, gauges, or

    components.

    Unintended movement o

    the machine or mechanism

    may cause injury to the

    technician or bystanders.

    To protect against

    unintended movement,

    secure the machine or

    disable / disconnect

    the mechanism while

    servicing.

    1. Install a pressure gauge in port M2

    to measure system pressure. Install a

    pressure gauge in case drain port L1

    or L2 to measure case pressure.C

    2. Start the prime mover and allow

    uid to reach normal operating

    temperature. Operate a hydraulic

    unction to its ull extension, loadingthe pump to maximum pressure and

    zero ow.

    3. Loosen the PC set screw and turn the

    PC adjusting plug until the gauge

    at port M2 indicates the desired

    setting. Clockwise rotation increases

    pressure, counterclockwise rotation

    decreases; approximate gain 42 bar

    [610 psi] per turn.

    I the pressure does not increase, an

    external system relie valve may require

    adjustment. External system relie valve

    must be set above the PC setting or

    proper operation.

    4. While holding the position o the

    PC adjusting plug, torque the PC set

    screw to 9 Nm [7 lbt].

    1 PC setting is reerenced to case pressure. Subtract case pressure rom system pressure to compute the actual

    setting.

    PC control adjustment

    Gauge port M2

    0 - 300 bar [0 - 4351 psi]

    3/16 in

    7/16-20

    50 Nm

    [37 lbfft]

    Case drain port L1

    0 - 10 bar [0 - 145 psi]

    3/8 in

    7/8-14

    90 Nm

    [67 lbfft]LS set screw

    4 mm9 Nm

    [7 lbfft]

    P106 264E

    X Port

    0-300 bar

    [0-5000 psi]

    3/16 in

    7/16-20

    20 Nm

    [51 lbfft]

    CautionContamination can

    damage internal

    components and void the

    manuacturers warranty.

    Take precautions to ensure

    system cleanliness when

    removing and reinstalling

    system lines.

    5. Stop the prime mover, remove the pressure gauges, and return the system to its

    normal operating confguration.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    22/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Adjustments

    ls ContRol

    1. Install a pressure gauge in port M2

    to measure system pressure. Install

    a pressure gauge in drain port L1 or

    L2 to measure case pressure. Tee-in

    a gauge to the LS / remote PC signal

    line (port X).C

    2. Start the prime mover and allow uid

    to reach normal operating temperature.

    Slowly operate a hydraulic unctionthat will demand approximately hal

    ow rom the pump, but keep system

    pressure below the PC set point.

    3. Loosen the LS set screw. While

    watching the pressure gauges,

    turn the LS adjusting plug until the

    desired pressure dierential between

    port M2 and port X is achieved.

    Clockwise rotation increases

    the setting, counterclockwise

    rotation decreases it; approximate

    gain = 17 bar [250 psi] per turn.

    4. While holding the position o the

    LS adjusting plug, torque the LS set

    screw to 9 Nm [7 lbt].

    5. Operate a hydraulic unction to its

    ull extension loading the pump at

    maximum pressure and zero ow.

    6. Loosen the PC set screw and turn the

    PC adjusting plug until the pressure

    WarningEscaping hydraulic uid

    under pressure can

    have sufcient orce

    to penetrate your skin

    causing serious injury

    and/or inection. Relieve

    pressure in the system

    beore removing hoses,

    fttings, gauges, or

    components.

    Unintended movement o

    the machine or mechanism

    may cause injury to the

    technician or bystanders.

    To protect against

    unintended movement,

    secure the machine or

    disable / disconnect

    the mechanism while

    servicing.

    LS control adjustment

    ad p, cr, d

    c h

    CautionContamination can

    damage internal

    components and void the

    manuacturers warranty.

    Take precautions to ensure

    system cleanliness when

    removing and reinstalling

    system lines.

    gauge at port M1 or M2 indicates the desired setting . Clockwise rotation increasespressure, counterclockwise rotation decreases it; approximate gain = 42 bar [610 psi]

    per turn.

    7. While holding the position o the PC adjusting plug, torque the PC set screw to

    9 Nm [7 lbt].

    8. Stop the prime mover, remove the pressure gauges, and return the system to its

    normal operating confguration.

    The LS setting is indicated in the pump model code. Reer to the Series 45 Frame F Open

    Circuit Axial Piston Pumps Technical Inormation Manual,50l098, or more inormation.

    Beore perorming adjustments, read page 15, Pressure measurement.

    I the pressure does not

    increase, an externalsystem relie valve may

    require adjustment.

    External system relie

    valve must be set above

    the PC setting or proper

    operation.

    1 The LS setting is a dierential pressure. Subtract pilot pressure at port X rom system pressure at port M2 to

    compute the actual setting. PC setting is reerenced to case pressure. Subtract case pressure rom system pressure to compute the actual

    setting.

    Gauge port M2

    0 - 300 bar [0 - 4351 psi]

    3/16 in

    7/16-20

    50 Nm

    [37 lbfft]

    Case drain port L1

    0 - 10 bar [0 - 145 psi]

    3/8 in

    7/8-14

    90 Nm

    [67 lbfft]LS set screw

    4 mm9 Nm

    [7 lbfft]

    P106 264E

    X Port

    0-300 bar

    [0-5000 psi]

    3/16 in

    7/16-20

    20 Nm

    [51 lbfft]

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    23/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Minor repair

    sHat seal

    RePlaCement

    The Series 45 open circuit variable pumps use a lip-type shat seal. You can replace this

    seal without major disassembly o the unit. Replacing the shat seal requires removing

    the pump rom the machine.

    CautionPremature bearing ailure

    can result i the shat seal

    contacts the shat bearing.

    Press the seal into the

    housing only ar enough

    to clear the retaining ring

    groove.

    R

    1. Using the appropriate snap-ring

    pliers, remove the retaining ring

    (K010) rom the housing.

    2. Remove the shat seal (K020) rom

    the bore in the pump housing and

    discard. C

    Puncture the ace o the seal with a

    packing hook, or use a slide-hammertype puller to remove the seal.

    i

    3. Inspect the pump housing and new

    seal or damage. Inspect the sealing

    area on the shat or rust, wear, or

    contamination. Polish the sealing

    area on the shat i necessary.

    4. Lubricate the lip o the new shat

    seal with clean hydraulic uid.

    Shat seal and retaining ring

    Place a protective sleeve over the shat end to prevent damage to the seal during

    installation.

    5. Keeping the seal perpendicular to the shat, press the new seal into the housing just

    ar enough to clear the retaining ring groove. Install seal with the cupped sidetoward the shat bearing. Do not damage the seal during installation.

    6. Using the appropriate snap ring pliers, install the seal retaining ring.

    7. Remove the installation sleeve.

    K020

    K010

    E101 254

    C CautionDont damage the pump

    housing or shat.

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    24/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Minor repair

    auxiliaRy PaDs You may install auxiliary mounting pads on pumps equipped with through-drive radial

    ported end caps. Follow these steps to either remove, replace, or exchange auxiliary

    mounting pads.

    R

    1. Remove the screws (J130), retaining the cover plate (J110) or auxiliary pump (not

    shown). Remove the shipping cover or auxiliary pump and its seal (J120).

    2. Remove the drive coupling (J140) i present.

    3. Remove the 4 screws (J100) retaining the pad adapter (J080) to the endcap. Discard

    the pad adapter O-ring (J090) i present.

    i

    4. Lubricate new O-ring (J090) with petroleum jelly. Install the pad adapter to theendcap.

    5. Install the 4 screws (J100 ) and torque to 55 Nm [40 lbt].

    6. Install the drive coupling (J140) i present.

    7. Install shipping cover or auxiliary pump with seal (J120).

    8. Install the screws (J130) and torque to 105 Nm [77 lbt].

    CautionShipping cover is intended

    only to retain coupling

    during shipment and

    storage. Do not operate

    pump with coupling and

    shipping cover installed. Auxiliary mounting pads

    J095

    J140

    J080J120

    J110

    J080

    J140

    J095

    J110

    J120

    P106 267E

    J080

    J100

    J120

    J110

    J130

    J140

    J095

    J130

    J100

    105 Nm

    [77 lbfft]

    55 Nm[40 lbfft]

    105 Nm

    [77 lbfft]

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    25/28

    511005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Minor repair

    ContRol Control assembly

    ls cr h; pr C10 hrh C10 d

    C11 hrh C118 r d PC cr

    D

    1. Remove the 4 screws (C300) holding

    the control housing onto the endcap.

    2. Remove the control and discard the

    4 interace O-rings (C200).

    3. Remove the PC set screw (C102),

    PC adjusting plug (C138), O-ring

    (C138A), springs (C134, C135), and

    seat (C133). Discard the O-ring.

    4. Remove the plug (C103), O-ring

    (C103A), and PC spool (C132) rom

    the control housing. Discard theO-ring. Note orientation o the spool

    or reassembly.

    For PC only controls, skip steps 5

    through 7

    5. Remove the plug (C105) and O-ring

    (C105A), or the plug (C106) and

    O-ring (C106A). Discard the O-ring

    (C105A or C106A).

    6. Remove the LS set screw (C102),

    LS adjusting plug (C118), O-ring

    (C118A), back-up rings (C118B),

    springs (C114, C115), and seat (C113).

    Discard the O-ring and backup rings.

    7. Remove the plug (C104), O-ring

    (C104A), and LS spool (C112) rom the control housing; discard the O-ring. Note

    orientation o the spool or reassembly.

    ipc

    8. Inspect the adjusting plugs or wear at the tips and where they contact the seat;

    replace as necessary.

    9. Inspect the springs and spring guides or wear or damage; replace as necessary.

    10. Careully inspect the spools. Ensure the sealing lands are ree o nicks and scratches.

    Check the ends that contact the spring guides or wear. Replace spools as necessary.

    11. Inspect the control housing or damage. Check the spool bores or excessive wear.

    12. Clean all parts and lubricate spools, springs, guides, and new O-rings with clean

    hydraulic uid.

    12 Nm

    [9 lbfft]

    12 Nm

    [9 lbfft]

    12 Nm

    [9 lbfft]

    6 Nm

    [5 lbfft]

    12 Nm

    [9 lbfft]

    9 Nm

    [7 lbfft]

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    26/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Minor repair

    12 Nm

    [9 lbfft]

    12 Nm

    [9 lbfft]

    12 Nm

    [9 lbfft]

    6 Nm

    [5 lbfft]

    12 Nm

    [9 lbfft]

    9 Nm

    [7 lbfft]

    Control assembly

    ls cr h; pr C10 hrh C10 d

    C11 hrh C118 r d PC cr

    R

    13. Install the PC spool (C132), spherical

    end frst, into the PC bore. Using anew O-ring (C103A), install the plug

    (C103). Torque the plug (C103) to

    12 Nm [9 lbt].

    14. Place the two PC springs (C134,

    C135) onto the spring guide (C133)

    and install into the PC bore. Place

    a new O-ring (C138A) onto the PC

    adjusting screw and thread it into

    the PC bore until ush, then make

    another ull turn. Install and torque

    the PC set screw (C102) to 9 Nm[7 lbt].

    For PC only controls, skip steps 15

    through 17.

    15. Install the LS spool (C112), spherical

    end frst, into the LS bore. Using a

    new O-ring (C105A or C106A), install

    the plug (C105 or C106). Torque

    the plug (C105 or C106) to 12 Nm

    [9 lbt].

    16. Using a new O-ring (C104A), install

    the plug (C104). Torque the plug to

    12 Nm [9 lbt].

    17. Place the two LS springs (C114, C115)

    onto the spring guide (C113) and

    install into the LS bore. Place a new

    O-ring (C118A) and back-up rings

    (C118B) onto the LS adjusting screw

    and thread it into the LS bore until ush, then make another ull turn. Install and

    torque the LS set screw (C102) to 9 Nm [7 lbt].

    18. Using petroleum jelly to retain them, install 4 new interace O-rings (C200) in therecesses on the control housing.

    19. Install the control assembly onto the endcap using the 4 screws (C300). Torque the

    screws to 6 Nm [5 lbt]. Torque screws in a criss-cross pattern and re-torque the frst

    screw to ensure proper torque retention.

    20. Check and adjust the control setting. SeeAdjustments, page 21.

    ContRol

    (cd)

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    27/28

    11005158 Rev. AB September 2007

    Series 45 Frame F Open Circuit Axial Piston Pumps

    Service Manual

    Minor repair

    Plug anD itting sizes

    anD toRques

    I any plugs or fttings are removed rom the unit during service, install and torque as

    indicated here. This drawing is a composite. Individual pump confgurations may be

    dierent. The illustration below shows the appropriate wrench size and torque or plugsthat may be installed in the pump:

    120 Nm

    [155 lbfft]

    12 Nm

    [9 lbfft]

    Plug locations, sizes, and torques

  • 7/27/2019 45 Series F Frame Service Manual (11005158 Rev AB Sept 2007)

    28/28

    ouR PRoDuCts

    Hydrostatic transmissions

    Hydraulic power steering

    Electric power steering

    Electrohydraulic power steering

    Closed and open circuit axial pistonpumps and motors

    Gear pumps and motors

    Bent axis motors

    Orbital motors

    Transit mixer drives

    Planetary compact gears

    Proportional valves

    Directional spool valves

    Cartridge valves

    Hydraulic integrated circuits

    Hydrostatic transaxles

    Integrated systems

    Fan drive systems

    Electrohydraulics

    Microcontrollers and sotware

    Electric motors and inverters

    Joysticks and control handlesDisplays

    Sensors

    sr-D m Pr d Cr s

    mrk ldr wrdd

    Sauer-Danoss is a comprehensive supplier providing complete

    systems to the global mobile market.

    Sauer-Danoss serves markets such as agriculture, construction, road

    building, material handling, municipal, orestry, tur care, and manyothers.

    We oer our customers optimum solutions or their needs and

    develop new products and systems in close cooperation and

    partnership with them.

    Sauer-Danoss specializes in integrating a ull range o system

    components to provide vehicle designers with the most advanced

    total system design.

    Sauer-Danoss provides comprehensive worldwide service or its

    products through an extensive network o Global Service Partners

    strategically located in all parts o the world.

    Local address:

    Sauer-Danoss (US) Company

    2800 East 13th Street

    Ames, IA 50010, USA

    Phone: +1 515 239-6000

    Fax: +1 515 239-6618

    Sauer-Danoss ApS

    DK-6430 Nordborg, Denmark

    Phone: +45 7488 4444

    Fax: +45 7488 4400

    Sauer-Danoss-Daikin LTD

    Sannomiya Grand Bldg. 8F

    2-2-21 Isogami-dori, Chuo-ku

    Kobe, Hyogo 651-0086, Japan

    Phone: +81 78 231 5001

    Fax: +81 78 231 5004

    Sauer-Danoss GmbH & Co. OHG

    Postach 2460, D-24531 Neumnster

    Krokamp 35, D-24539 Neumnster, Germany

    Phone: +49 4321 871-0

    Fax: +49 4321 871 122