4.3 0.88 0p9aie4-1gi/...4.3 9.0STD. LXX 0.88 0.50 3.9 TLXX This drawing is the property of Garner...
Transcript of 4.3 0.88 0p9aie4-1gi/...4.3 9.0STD. LXX 0.88 0.50 3.9 TLXX This drawing is the property of Garner...
-
4.3
9.0STD. LXX
0.88 0.50
3.9
TLXX
This drawing is the property of Garner Industries.It is issued in confidence on the condition that itis to be returned on request and that it is not tobe copied or reproduced in any way, or used in the manufacture of the subject matter thereof withoutthe written consent of Garner Industries. DIMENSIONS ARE IN INCHES UNLESS NOTED OTHERWISE
TITLE: Dimensional Drawing, Dust Detect NPT Mount Model No.: DD-3000 DWG No.: 418-0248 PAGE1 OF 2
Division of Garner Industries / 7201 North 98th Street / Lincoln, NE 68507(402) 434-9102 / fax (402) 434-9144 \ www.binmaster.com
SCALE: NONE REVISION-
ORDINARY/GENERAL PURPOSE ONLY, CE APPROVED (CSA/UL)HAZARDOUS LOCATION, CSA APPROVED, CLASS II, DIVISION II,
GROUPS E-G, ATEX, IECEx (APPROVAL PENDING)HAZARDOUS LOCATION, CSA APPROVAL, CLASS I, DIVISION I,
GROUPS A-D, ATEX, IECEx (APPROVAL PENDING)
IMPORTANT INSTALLATION NOTES1. HAZARDOUS AREA LOCATIONS: VERIFY REQUIRED CLASS AND DIVISON AGAINST SPECIFICAITON LISTED ON THE PROCESS DEVICE TAG2. DO NOT INSTALL ANY EQUIPMENT THAT IS NOT TAGGED SUITABLE FOR THE AREA CLASSIFICATION3. BEFORE INSTALLATION VERIFY THE DEVICE IS TAGGED AND SUITABLE FOR THE REQUIRED AMBIENT TEMPERATURE, PROCESS TEMPERATURE, AND WETTED MATERIALS4. THIS DEVICE IS NOT DESIGNED FOR USE AS A FUNCTIONAL SAFETY DEVICE. THE DEVICE MUST NOT USED AS PART OF, OR AS A AN INPUT TO A SAFETY SYSTEM5. ADHERE TO NEC, CEC, NEMA, NFPA, OSHA INSTALLATION AND MAINTENANCE GUIDELINES AND ALL APPLICABLE LOCAL CODES6. INSTALLATION AND SERVICE MUST BE PERFORMED BY QUALIFIED PERSONNEL WHILE POWER IS DISCONNECTED AND THE PROCESS IS OFF
ORDER CODES - EXAMPLE DD300.S2L12M1T11XN4G
DESCRIPTION CODE SPECIFICATION
DD-3000 PARTICULATE MONITOR DD300., DD305., DD310. 2 RELAYS 18 TO 60 VDC, 2 RELAYS 20 TO 250 VAC/VDC, 2-WIRE LOOP POWERED
AMBIENT TEMPERATURE S, L STANDARD AMBIENT TEMP: -13�$�F TO 158�$�F (-25
RESOLUTION 1, 2 HIGH RESOLUTION: 0.5pA, RANGE 0 TO 5000pA, STANDARD RESOLUTION : 5.0pA, RANGE 0 TO 5000pA
PROBE LENGTH L02, L03, L06, L12, L18, L24, L30, L36 1.5", 3.0", 6.0", 12.0"(STOCK SIZE), 18.0", 24.0"(STOCK SIZE), 30.0", 36.0"
TEFLON INSULATOR LENGTH TL0.5, TL2.0 SENSOR LENGTHS 3" & UNDER HAVE 0.5" TEFLON INSULATOR, SENSORS LENGTHS 6" & LONGER HAVE 2" TEFLON INSULATOR
MOUNTING M1, M3 0.5" NPT MALE, 1.5" TRI-CLAMP
PROCESS TEMPERATURE T1, T2, T3 STANDARD TEMP: -13�$�F TO 250�$�F MAX, HIGH TEMP: -13�$�F TO 450�$�F MAX, HIGH-HIGH TEMP.: -13�$�F TO 500�$�F
PROCESS PRESSURE 1 10 PSI MAX.
PROBE INSULATION C TEFLON COATED, (USE X IF NONE)
AREA CLASSIFICATION G ORDINARY/GENERAL PURPOSE
COVER REMOVED TO SHOW DETAIL
1/2" NPT TEFLON INSULATOR
3/4" NPTCONDUIT ENTRY
1/2" NPTCONDUIT ENTRY
1/2" NPTCONDUIT ENTRY
ELECTRONICS MODULE(PM 1-A SHOWN FOR REF.)
-
4.3
6.0 3.0STD. LXX
3.0STD.
0.80 0.50
3.9
TLXX
This drawing is the property of Garner Industries.It is issued in confidence on the condition that itis to be returned on request and that it is not tobe copied or reproduced in any way, or used in the manufacture of the subject matter thereof withoutthe written consent of Garner Industries. DIMENSIONS ARE IN INCHES UNLESS NOTED OTHERWISE
TITLE: Dimensional Drawing, Dust Detect, Tri-Clamp Mount Model No.: DD-3000 DWG No.: 418-0248 PAGE2 OF 2
Division of Garner Industries / 7201 North 98th Street / Lincoln, NE 68507(402) 434-9102 / fax (402) 434-9144 \ www.binmaster.com
SCALE: NONE REVISION-
ORDINARY/GENERAL PURPOSE ONLY, CE APPROVED (CSA/UL)HAZARDOUS LOCATION, CSA APPROVED, CLASS II, DIVISION II,
GROUPS E-G, ATEX, IECEx (APPROVAL PENDING)HAZARDOUS LOCATION, CSA APPROVAL, CLASS I, DIVISION I,
GROUPS A-D, ATEX, IECEx (APPROVAL PENDING)
IMPORTANT INSTALLATION NOTES1. HAZARDOUS AREA LOCATIONS: VERIFY REQUIRED CLASS AND DIVISON AGAINST SPECIFICAITON LISTED ON THE PROCESS DEVICE TAG2. DO NOT INSTALL ANY EQUIPMENT THAT IS NOT TAGGED SUITABLE FOR THE AREA CLASSIFICATION3. BEFORE INSTALLATION VERIFY THE DEVICE IS TAGGED AND SUITABLE FOR THE REQUIRED AMBIENT TEMPERATURE, PROCESS TEMPERATURE, AND WETTED MATERIALS4. THIS DEVICE IS NOT DESIGNED FOR USE AS A FUNCTIONAL SAFETY DEVICE. THE DEVICE MUST NOT USED AS PART OF, OR AS A AN INPUT TO A SAFETY SYSTEM5. ADHERE TO NEC, CEC, NEMA, NFPA, OSHA INSTALLATION AND MAINTENANCE GUIDELINES AND ALL APPLICABLE LOCAL CODES6. INSTALLATION AND SERVICE MUST BE PERFORMED BY QUALIFIED PERSONNEL WHILE POWER IS DISCONNECTED AND THE PROCESS IS OFF
ORDER CODES - EXAMPLE DD300.S2L12M1T11XN4G
DESCRIPTION CODE SPECIFICATION
DD-3000 PARTICULATE MONITOR DD300., DD305., DD310. 2 RELAYS 18 TO 60 VDC, 2 RELAYS 20 TO 250 VAC/VDC, 2-WIRE LOOP POWERED
AMBIENT TEMPERATURE S, L STANDARD AMBIENT TEMP: -13�$�F TO 158�$�F (-25
RESOLUTION 1, 2 HIGH RESOLUTION: 0.5pA, RANGE 0 TO 5000pA, STANDARD RESOLUTION : 5.0pA, RANGE 0 TO 5000pA
PROBE LENGTH L02, L03, L06, L12, L18, L24, L30, L36 1.5", 3.0", 6.0", 12.0"(STOCK SIZE), 18.0", 24.0"(STOCK SIZE), 30.0", 36.0"
TEFLON INSULATOR LENGTH TL0.5, TL2.0 SENSOR LENGTHS 3" & UNDER HAVE 0.5" TEFLON INSULATOR, SENSORS LENGTHS 6" & LONGER HAVE 2" TEFLON INSULATOR
MOUNTING M1, M3 0.5" NPT MALE, 1.5" TRI-CLAMP
PROCESS TEMPERATURE T1, T2, T3 STANDARD TEMP: -13�$�F TO 250�$�F MAX, HIGH TEMP: -13�$�F TO 450�$�F MAX, HIGH-HIGH TEMP.: -13�$�F TO 500�$�F
PROCESS PRESSURE 1 10 PSI MAX.
PROBE INSULATION C TEFLON COATED, (USE X IF NONE)
AREA CLASSIFICATION G ORDINARY/GENERAL PURPOSE
COVER REMOVED TO SHOW DETAIL
MOUNT
TRI-CLAMP
GASKET
1.5" QC
TEFLONINSULATOR
3/4" NPTCONDUIT ENTRY
1/2" NPTCONDUIT ENTRY
1/2" NPTCONDUIT ENTRY
ELECTRONICS MODULE(PM 1-A SHOWN FOR REF.)
Sheet 1Sheet 2