4172013 AdriaVidovic Thesis - Duke University
Transcript of 4172013 AdriaVidovic Thesis - Duke University
![Page 1: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/1.jpg)
i
v
On the Feasibility of a Novel In-‐‑Vivo Dosimeter for Brachytherapy
by
Adria Katarina Vidovic
Graduate Program in Medical Physics Duke University
Date:_______________________ Approved:
___________________________ Mark Oldham, Supervisor
___________________________
Oana Craciunescu
___________________________ Justus Adamson
___________________________
Robert Reiman
Thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in the
Graduate Program in Medical Physics in the Graduate School of
Duke University
2013
![Page 2: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/2.jpg)
i
v
ABSTRACT
On the Feasibility of a Novel In-‐‑Vivo Dosimeter for Brachytherapy
by
Adria Katarina Vidovic
Graduate Program in Medical Physics Duke University
Date:_______________________ Approved:
___________________________ Mark Oldham, Supervisor
___________________________
Oana Craciunescu
___________________________ Justus Adamson
___________________________
Robert Reiman
An abstract of a thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in the
Graduate Program in Medical Physics in the Graduate School of
Duke University
2013
![Page 3: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/3.jpg)
Copyright by Adria Katarina Vidovic
2013
![Page 4: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/4.jpg)
iv
Abstract Purpose: Clinical brachytherapy systems are capable of delivering very high doses with
high dose gradients. It is important therefore to be able to accurately verify the doses
calculated by brachytherapy treatment planning. Current dose verification methods are
limited by poor resolution, and in the presence of large dose gradients, may give non-‐‑
representative results [1]. This thesis aims to evaluate the feasibility of a novel
radiochromic dosimetry system for in-‐‑vivo dose verification in organs at risk (bladder
and rectum) in high dose rate (HDR) intracavitary gynecological brachytherapy through
a comparison with a gold standard.
Methods: A novel dosimeter PRESAGE®-‐‑IV designed for in-‐‑vivo dosimetry is
investigated. PRESAGE®-‐‑IV dosimeters are small cylinders 4mm in diameter by 20mm
in height. When irradiated, the dosimeters change color in proportion to the local
absorbed dose. The dosimeters were irradiated to doses between 1-‐‑15 Gy. Two methods
were investigated for readout of this radiochromic response: (i) a volume averaged
readout by conventional spectrophotometer, and (ii) a line profile readout by a novel 2D
projection imaging method utilizing a high-‐‑resolution (50 micron) telecentric optical
system. Method (i) is considered the gold standard, as it is has been extensively used
with PRESAGE® in well-‐‑defined optical-‐‑cuvettes. The feasibility of PRESAGE®-‐‑IV was
evaluated by comparison to standard PRESAGE® in optical-‐‑cuvettes. The feasibility of
the high-‐‑resolution readout (method ii) was evaluated by direct comparison against
![Page 5: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/5.jpg)
v
method (i). Dosimeters were also tested in-‐‑vivo on six patients undergoing Iridium-‐‑192
HDR intracavitary brachytherapy treatments and dose measurements were compared to
EclipseTreatment Planning System (Varian Medical Systems).
Results: When compared to the gold standard (optical-‐‑cuvettes), the sensitivity and
noise of PRESAGE®-‐‑IV shows a linear relationship in sensitivity between 1-‐‑15 Gy with a
95% confidence interval in the slope (0.8703 +/-‐‑ 0.0192). The feasibility of the high-‐‑
resolution readout (method ii) evaluated by direct comparison against method (i)
resulted in a sensitivity of 0.0136 ± 0.0002 and for the spectrophotometer 0.0135 ± 0.0002,
which is a 0.74% difference in sensitivity within the 95% confidence interval.
Examination of patient data showed large differences, and on average gave 19% and
22% differences in measured doses vs. Eclipse measurements in the bladder and rectum,
respectively.
Conclusions: Results show that a novel radiochromic dosimetry system for in-‐‑vivo dose
verification in organs at risk is feasible. The conventional spectrophotometer readout
method had the limitation that it averages the change in optical density over a 10 mm
area of the dosimeter. The novel, high-‐‑resolution 2D readout technique was found to
have the advantage of producing images that could be further analyzed through line
profiles in any area of the dosimeter. Due to the large differences in measured doses for
organs at risk, further work is needed to validate dosimeter-‐‑positioning technique.
![Page 6: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/6.jpg)
vi
Dedication
To my fiancé Adam, and my family for all the encouragement and support.
![Page 7: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/7.jpg)
vii
Contents
Abstract .......................................................................................................................................... iv
List of Tables ................................................................................................................................. ix
List of Figures ................................................................................................................................ x
List of Abbreviations ................................................................................................................. xiii
Acknowledgements ................................................................................................................... xiv
1. Introduction ............................................................................................................................... 1
1.1 Gynecological HDR Intracavitary brachytherapy ....................................................... 1
1.2 HDR Intracavitary brachytherapy treatment applicators and packing systems .... 2
1.3 Imaging in HDR intracavitary brachytherapy ............................................................. 5
1.4 Limitations in HDR intracavitary brachytherapy ........................................................ 5
1.5 Current in-‐‑vivo dose verification techniques. .............................................................. 6
1.6 PRESAGE® as an in-‐‑vivo dosimeter ............................................................................... 8
1.7 Scope and aims ................................................................................................................. 9
2. Methods .................................................................................................................................... 11
2.1 Characterization of PRESAGE®-‐‑IV dosimeters .......................................................... 11
2.1.1 Gold standard method using optical cuvettes ...................................................... 11
2.1.2 Dose response ............................................................................................................ 12
2.1.3 Energy response ........................................................................................................ 13
2.1.4 Temporal stability ..................................................................................................... 15
2.1.5 Temperature stability ................................................................................................ 15
2.1.6 Effect of cuvette material on PRESAGE®-‐‑IV .......................................................... 16
![Page 8: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/8.jpg)
viii
2.2 Investigating the feasibility of readout by comparison to gold standard .............. 18
2.2.1 Cuvette and dosimeter irradiation .......................................................................... 19
2.3 2D optical scanning and spectrophotometer reading ............................................... 20
2.3.1 Spectrophotometer technique .................................................................................. 21
2.3.2 2D optical scanning technique ................................................................................. 22
2.3.3 2D optical scanner image registration ...................................................................... 24
2.3.4 Quantitative comparison between the spectrophotometer and 2D optical scanner ................................................................................................................................... 25
2.4 Patient treatment ............................................................................................................ 26
2.4.1 Eclipse dose verification ........................................................................................... 28
3. Results and Discussion ........................................................................................................... 30
3.1 PRESAGE® Sensitivity and stabilities ......................................................................... 31
3.1.1 Dose response ............................................................................................................ 31
3.1.2 Energy Response ....................................................................................................... 32
3.1.3 Temporal stability ..................................................................................................... 35
3.1.4 Temperature stability ................................................................................................ 36
3.1.5 Effect of the cuvette material on PRESAGE®-‐‑IV ................................................... 37
3.2 Results of the feasibility of readout by comparison to gold standard .................... 38
3.3 Quantitative comparison between the 2D optical scanner and spectrophotometer ................................................................................................................................................. 39
3.4 Patient treatment ............................................................................................................ 40
4. Conclusions .............................................................................................................................. 49
References .................................................................................................................................... 51
![Page 9: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/9.jpg)
ix
List of Tables Table 1: Monitor units delivered to 1.5% DEA formulation dosimeters and cuvettes ..... 20
Table 2: Overview of PRESAGE®-‐‑IV dosimeters irradiated in this work. .......................... 30
![Page 10: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/10.jpg)
x
List of Figures Figure 1: 45° T&R Applicator (Varian Medical Systems) [22] ................................................ 3
Figure 2: Stump Cylinder Applicator (Varian Medical Systems) [22] ................................... 3
Figure 3: Capri Applicator (Varian Medical Systems) [22] ..................................................... 4
Figure 4: PRESAGE®-‐‑IV dosimeters pre-‐‑irradiation (left) and post-‐‑irradiation 15 Gy (right). ............................................................................................................................................. 9
Figure 5: 1x1x4 cm3 cuvette (left) and Genesys® 20, ThermoSpectronic® spectrophotometer (right). ......................................................................................................... 12
Figure 6: PRESAGE® filled cuvettes irradiated with a 6MV photon beam to doses of 0-‐‑15 Gy. ................................................................................................................................................. 13
Figure 7: Axial Cross Section of Applicator with Dosimeters Visible with Beekley CT-‐‑ Spots® Localizers (left) and Sagittal Cross Section of Applicator with One Dosimeter Visible. .......................................................................................................................................... 14
Figure 8: Water Bath containing Source Guide Tube and Applicator with Dosimeters (left) and the Treatment Set Up (right). ................................................................................... 15
Figure 9: Example of the PRESAGE® molds in the shape of cuvettes. ................................ 17
Figure 10: PRESAGE®-‐‑Filled Cuvette (left) and PRESAGE® Mold (right). ......................... 18
Figure 11: PRESAGE®-‐‑Filled Cuvette (left) and Cuvette with PRESAGE®-‐‑IV (right) ....... 19
Figure 12: Side View of the Standard Treatment Set-‐‑Up ...................................................... 20
Figure 13: From left to right: Cuvette filled with mineral oil, Styrofoam holder and dosimeter, holder in cuvette, cuvette holder, spectrophotometer. ...................................... 22
Figure 14: Diagram of the DMicrOS System [15]. .................................................................. 23
Figure 15: Diagram of the jig (designed during this thesis work) that holds in-‐‑vivo dosimeters in tank. ...................................................................................................................... 23
Figure 16: An Example of a Line Profile of Two Dosimeters Irradiated to 10 Gy Using the Standard Treatment Setup Described in Section 2.2.1. The Line Profile Was Taken Along
![Page 11: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/11.jpg)
xi
the Center of Each Dosimeter to Measure the Change in Optical Density at Approximately the Same Length (10 mm) and Location (3 mm from the Bottom of the Dosimeter) as the Spectrophotometer. ..................................................................................... 25
Figure 17: PRESAGE®-‐‑IV dosimeter with and without Tegaderm™ and Beekley CT-‐‑ Spots® localizers. ......................................................................................................................... 26
Figure 18: Example of Dosimeter Placed on Alatus Balloon Packing System (top) and Dosimeters In-‐‑Vivo Near OARs in T&R Treatment (left and right) .................................... 27
Figure 19: Example of Dosimeters Placed on Cylinder Applicator Axial (left) and Sagittal (right) Views. ............................................................................................................................... 28
Figure 20: Example of Dosimeter Placed inside Capri Applicator Channel Axial (left) and Coronal View of Dosimeter with Point Doses (right). ........................................................... 28
Figure 21: Dose Response (Sensitivity) of All Five Formulations. ....................................... 32
Figure 22: The Sensitivity for the 1.5% O-‐‑MeO-‐‑DEA (2/14/13) Dosimeter Formulation. . 34
Figure 23: Dosimeter Dose Measurement of a KV Brachytherapy Source Using a MV Calibration Curve. ....................................................................................................................... 35
Figure 24: Dose Response (Sensitivity) Temporal Stability of the 1.7% O-‐‑MeO-‐‑DEA (11/27/12) Formulation. .............................................................................................................. 36
Figure 25: Change in OD with Temperature for the D21 Formulation. .............................. 37
Figure 26: The Change in Optical Density of the Molds vs. Cuvettes for the 1.5% O-‐‑MeO-‐‑DEA Formulation (1/31/13). ....................................................................................................... 38
Figure 27: The Change in Optical Density of the Dosimeters vs. Cuvettes for the 1.5% O-‐‑MeO-‐‑DEA Formulation (2/14/13). ............................................................................................ 39
Figure 28: The Comparison in Sensitivity of the DmicrOS system and the Spectrophotometer for the 1.5% O-‐‑MeO-‐‑DEA Formulation (2/14/13). ............................... 40
Figure 29: Dose Response (Sensitivity) of the D21 Formulation at Body Temperature. .. 41
Figure 30: T&R Patient 6 Treatment Dose Measurement Using the D21 Formulation. .... 42
Figure 31: Capri Patient 1 fx 1 Dose Measurement Using the D21 Formulation. .............. 43
![Page 12: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/12.jpg)
xii
Figure 32: Capri Patient 1 fx 1 Dose Measurement Using the D21 Formulation. .............. 43
Figure 33: Capri Patient 1 fx 3 Dose Measurement Using the D21 Formulation. ............. 44
Figure 34: Dose Response (Sensitivity) of the 2% O-‐‑MeO-‐‑DMA Formulation at Body Temperature. ................................................................................................................................ 45
Figure 35: T&R Patient 2 fx 1 Dose Measurement (2% O-‐‑MeO-‐‑DMA) ............................... 46
Figure 36: T&R Patient 2 fx 2 Dose Measurement (2% O-‐‑MeO-‐‑DMA) ............................... 46
Figure 37: Stump Patient 3 Dose Measurement (2% O-‐‑MeO-‐‑DMA) ................................... 47
Figure 38: Stump Patient 4 Dose Measurement (2% O-‐‑MeO-‐‑DMA) ................................... 48
Figure 39: Cylinder Patient 5 Dose Measurement (2% O-‐‑MeO-‐‑DMA) ............................... 48
![Page 13: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/13.jpg)
xiii
List of Abbreviations HDR-‐‑ High Dose Rate
CT-‐‑ Computed Tomography
MRI-‐‑ Magnetic Resonance Imaging
T&R-‐‑ Tandem and Ring
T&O-‐‑ Tandem and Ovoid
OAR-‐‑ Organ at Risk
DVH-‐‑ Dose Volume Histogram
TPS-‐‑ Treatment Planning System
TLD-‐‑ Thermoluminescent Dosimeter
MOSFET-‐‑ Metal oxide semiconductor field effect transistor detectors
O-‐‑MeO-‐‑DEA-‐‑ 2-‐‑methoxy-‐‑N, N-‐‑diethylamine
O-‐‑MeO-‐‑DMA-‐‑ 2-‐‑methoxy-‐‑N, N-‐‑dimethylamine
LED-‐‑Light Emitting Diode
CCD-‐‑ Charged-‐‑Coupled Device
DMicrOS-‐‑ Duke Micro Optical Scanner
DLOS-‐‑ Duke Large Field of View Optical-‐‑CT Scanner
![Page 14: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/14.jpg)
xiv
Acknowledgements I would like to acknowledge the support of my advisor Dr. Mark Oldham for his
endless advice and support. I would also like to give a special thanks to Titania Juang
for her help throughout my work and specifically in collecting patient data, performing
the temperature dependence experiment, and creating a jig for the dosimeters. I would
like to thank my committee members Dr. Oana Craciunescu, Dr. Justus Adamson, and
Dr. Robert Reiman, for their time and effort. I would also like to thank Dr. Junzo Chino,
Beverly Steffey, and Dr. Sheridan Meltsner who were involved in helping me collect
clinical data. Last, but not least I would like to thank Dr. John Adamovics and Heuris
Pharma LLC for providing our lab with PRESAGE®-‐‑IV dosimeters.
![Page 15: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/15.jpg)
1
1. Introduction
1.1 Gynecological HDR Intracavitary brachytherapy
Gynecological intracavitary brachytherapy is the clinical use of small,
encapsulated radioactive sources placed in the uterus, uterine cervix, or vagina directly
on a target volume for irradiation of malignant tumors or nonmalignant lesions [2]. The
principle advantage of brachytherapy is the superior localization of dose to the tumor
volume and the rapid dose fall-‐‑off in accordance to the inverse square law.
High Dose Rate (HDR) brachytherapy is a special method of delivering
brachytherapy at doses greater than 12 Gy/hour that employs computer controlled
remotely driven sources known as a remote afterloader. This eliminates the potential for
exposure to clinical staff, and has the ability to tailor radiation delivery to maximize
dose within a target volume and minimize dose in adjacent normal structures [3]. The
HDR brachytherapy program at Duke Medical Center uses a Varian® GammaMed™plus
iX remote afterloader for intracavitary brachytherapy treatments. This remote
afterloader delivers an Iridium-‐‑192 source and treatment plans are based on a 10 Curie
activity and an Air Kerma Rate of 0.063 Gy/hour (±5%) for 555 GBq at 1 meter.
The HDR delivery procedure involves adequate sedation of the patient that can
include general, spinal, or IV conscious sedation. The patient is kept under anesthesia
during the duration of applicator insertion and treatment delivery, which takes
approximately two to four hours [8]. The remote afterloader delivers sources through
![Page 16: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/16.jpg)
2
source guide tubes that are connected to treatment applicators that are inserted into the
body cavity. At Duke, these applicators (Section 1.2) are imaged by computed
tomography (CT) and magnetic resonance (MR) to ensure correct positioning. The
images of the patient with the applicators in situ are imported into treatment planning
software and the patient is brought into a dedicated shielded room for treatment. The
treatment plan is then optimized and delivered using the remote afterloader. On
completion of treatment, the applicators are carefully removed from the body.
1.2 HDR Intracavitary brachytherapy treatment applicators and packing systems
Applicator selection is important in brachytherapy in order to customize
treatment delivery for each individual. Treatment applicators can serve many functions:
positioning the source with respect to the anatomy, shaping the anatomy, determining
the penetration of the dose, adding space between normal tissue and the sources, and
shielding normal structures [4]. The applicators used in this thesis are all CT/MRI
compatible Varian® Tandem and Ring (T&R), Stump Cylinder, Cylinder and the
multichannel Capri™.
The T&R applicator is intended for treatment of the cervix, cervix canal and
uterus (Figure 1). It consists of an intrauterine titanium tube known as a ‘tandem’ that
guides the remote afterloading sources, and a ring around the tandem shields in the
apparatus to reduce the radiation exposure to the bladder and rectum. The tandem can
![Page 17: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/17.jpg)
3
come in three different curvatures (15°, 30° or 45°) to allow for non-‐‑cylindrical dose
shaping relative to the vaginal axis.
Figure 1: 45° T&R Applicator (Varian Medical Systems) [22]
A Stump Cylinder applicator is used for post-‐‑operative treatment of the vaginal
cuff (Figure 2). A vaginal cylinder applicator is used to treat cancer of the vagina. It is a
smooth, plastic cylinder, measuring about one inch in diameter, with a single channel
where the radioactive source can travel and the distance between its tip and the source
remains constant ensuring a targeted, homogenous treatment dose.
Figure 2: Stump Cylinder Applicator (Varian Medical Systems) [22]
A Capri™ applicator is used to treat cancer of the vagina (Figure 3). It has a soft
![Page 18: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/18.jpg)
4
foam core with 13 channels, and offers flexible dosimetry designed to help provide
patient comfort. With its inflatable design, it offers flexible sizing to accommodate a
large range of patient sizes. Multichannel applicators like Capri™ facilitate better dose
optimization and reduction of the dose to the OARs [3].
Figure 3: Capri Applicator (Varian Medical Systems) [22]
Use of intracavitary applicators requires appropriate vaginal packing to secure
applicators and displace the rectum and bladder away from the radiation source to
minimize side effects and complications. Typically, this is done via gauze packing, rectal
retractor, and/or Foley balloon or packing balloon systems such as the Radiadyne®
Alatus™ Vaginal Balloon Packing System [17]. The Radiadyne® Alatus™ Vaginal Balloon
Packing System offers several benefits over traditional gauze packing, including
decreased risk of vaginal laceration during packing, greater reproducibility among
fractions, and the ability to easily reposition the applicators after the packing is in place
[18]. After treatment applicators have been inserted and secured, the patient-‐‑applicator
geometry is imaged using CT and MR.
![Page 19: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/19.jpg)
5
1.3 Imaging in HDR intracavitary brachytherapy
In the mid-‐‑1990’s, 3D brachytherapy planning systems started to emerge with the
use of CT and MR imaging. CT/MR-‐‑based planning provides the potential for rapid 3D
rendering and Dose Volume Histogram (DVH) generation, providing more realistic
calculation of dose within the tumor and adjacent OARs versus conventional orthogonal
film-‐‑based treatment planning [3]. CT is widely available, provides good spatial
resolution, the ability to discern objects on the basis of electron density, and has good
organ delineation [3]. MR has the ability to better depict target volumes than CT and
allows adequate depiction of size, location and paracervical involvement of the tumor
and its relations to an applicator [20].
The positive attributes of both imaging modalities can be displayed through the
use of image registration. CT is used for visualization of the treatment applicator, and
MR is optimal for the delineation process [3]. Imaging in brachytherapy is often
performed prior to each fraction, and all the information in the image studies acquired at
different points in time should ideally be combined to assess the total treatment
delivered to the patient. However, the use of MR prior to each fraction is costly, so a
compromise is usually made to image prior to the first fraction only, given that the
applicator type is consistent and that overall treatment duration is short [3].
1.4 Limitations in HDR intracavitary brachytherapy
The recent advances in brachytherapy planning software, MRI-‐‑compatible
![Page 20: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/20.jpg)
6
applicators, and image-‐‑guided brachytherapy have lead to more complex treatments
and a subsequent need for improved quality-‐‑assurance procedures [7]. One of the main
challenges in brachytherapy imaging is image registration of image sets taken at
different points in time [3].
Not only does organ motion increase or decrease the tumor volume, but
transportation of the patient from one location to another my influence the relative
positioning of applicators and organs. The localization, dose calculation, and the
treatment delivery for HDR intracavitary brachytherapy can take two to four hours, and
uncertainties in treatment can easily occur without detection [5]. Clinically related
uncertainties include target volume definition and contouring, applicator positioning,
organ motion, and inter-‐‑ and intra-‐‑fraction applicator movement [6]. These clinical
uncertainties can be verified, and minimized with the use of in-‐‑vivo dosimetry.
1.5 Current in-vivo dose verification techniques.
In-‐‑vivo dosimetry can help identify errors from applicators or afterloader
malfunctions during the first fraction of treatment to resolve discrepancies and can
provide dose verification to OARs, such as the bladder and the rectum. A number of
systems have been used and are being developed, such as lithium fluoride
thermoluminescent dosimeters (TLDs), diamond detectors, Metal oxide semiconductor
field effect transistor detectors (MOSFETS), and a novel scintillation detector
BrachyFOD™.
![Page 21: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/21.jpg)
7
TLDs have been frequently applied as an in-‐‑vivo dosimeter for HDR
brachytherapy because they are tissue equivalent, have a small sensitive volume, and
can be reused [8]. Diamond detectors have a small sensitive volume of a sensitive
volume of 6 mm3 and are sensitive enough for HDR brachytherapy dosimetry, but have
a much larger physical size than TLDs, MOSFETs, and scintillation dosimeters [21].
MOSFETS are attractive due to their small sensitive volume and direct readout [9]. The
RADPOS system (Best Medical, Canada) consists of a MOSFET dosimeter physically
coupled to a position-‐‑sensing probe, which deduces its 3D position in static magnetic
fields generated by a transmitter, shown to be accurate within 0.5 to 1.0mm [10]. A novel
scintillation detector, BrachyFOD™ has a sensitive measuring volume of 1 mm by 5 mm
(Bicron BC400 scintillator) coupled to a 0.98 mm core polymethyl methacrylate optical
fiber. It was found to be accurate to within 3% for distances of 10 to 100 mm from an
HDR Iridium-‐‑192 brachytherapy source in water, and it has an angular dependence of
less than 2% [21]. Only the BrachyFOD™ and the MOSFET are capable of real-‐‑time
measurements and are small enough for insertion into the urethra [21].
Although TLDs are most commonly used, they have the inability of real-‐‑time
readings, depth dependent sensitivity [6], and measure the dose only at a single point
[1]. Diamond detectors were found to have dose rate dependence whose relative
response decreases with increasing dose rate as well as its large rigid structure that
prevents it from being inserted in-‐‑vivo [21]. MOSFETS, including the RADPOS system
![Page 22: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/22.jpg)
8
that reduces positional uncertainty, still contain uncertainties in angular dependence
and are prone to calibration drift [9]. The ideal in-‐‑vivo dose verification method would
have the ability to be inserted into cavities without disruption to treatment, ability to
provide high resolution dose profiles along steep dose gradients and provide fast
acquisition read out time for a high patient load [12].
1.6 PRESAGE® as an in-vivo dosimeter
PRESAGE® is a three-‐‑dimensional dosimetry material that has the potential to be
an ideal dosimeter for in-‐‑vivo dose verification. PRESAGE® consists of an optically clear
polyurethane matrix containing a leuco dye that exhibits a linear radiochromic response
when exposed to ionizing radiation. A number of potential advantages accrue over other
gel dosimeters, including insensitivity to oxygen, radiation induced light absorption
contrast rather than scattering contrast, water equivalency (~1.05 g/cm3) and a solid
texture amenable to machining to a variety of shapes and sizes without the requirement
of an external container [11]. As the sensitivity, or optical density change per Gray, is
dependent on the concentration of leuco dye added to the polymer, each formulation of
PRESAGE® has a slightly different sensitivity [13], [14]. In addition, PRESAGE® has
been previously tested as a 3D verification dosimeter near a brachytherapy source (Ir-‐‑
192) and has shown to have no evidence of energy response [23].
These qualities, and the pressing need for in-‐‑vivo dose verification in
brachytherapy, prompted the development of PRESAGE®-‐‑In-‐‑Vivo, or PRESAGE®-‐‑IV.
![Page 23: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/23.jpg)
9
Supplied by Heuris Pharma LLC, PRESAGE®-‐‑IV was specifically developed for the
purpose of in-‐‑vivo dose verification for organs at risk such as the bladder and rectum.
PRESAGE®-‐‑IV dosimeters are small cylinders 4mm in diameter by 20mm in height.
When irradiated, the dosimeters change color in proportion to the local absorbed dose as
seen in Figure 4 where the dosimeter on the left has not been exposed to radiation and
the dosimeter on the right has been exposed to 15 Gy.
Figure 4: PRESAGE®-‐‑IV dosimeters pre-‐‑irradiation (left) and post-‐‑irradiation 15 Gy (right).
1.7 Scope and aims
This thesis aims to evaluate the feasibility of a novel dosimetry readout system
for PRESAGE®-‐‑IV dosimeters and to demonstrate its utility as an in-‐‑vivo dosimeter for
brachytherapy treatment. The feasibility is investigated by delivering a range of doses to
dosimeters, measuring the change in optical density using spectrophotometry and a
novel 2D optical scanning technique, and comparing the change in optical density to a
gold standard. To demonstrate its utility as a dosimeter for in-‐‑vivo dose verification,
![Page 24: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/24.jpg)
10
dose measurements from PRESAGE®-‐‑IV dosimeters will be compared to dose line
profiles created in Eclipse® treatment planning system (TPS) for OARs in patient
treatments.
![Page 25: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/25.jpg)
11
2. Methods The characterization of PRESAGE®-‐‑IV dosimeters including a description of the
gold standard, dose response, energy response, temporal and temperature stabilities and
the effect of cuvette material on PRESAGE®-‐‑IV are introduced in Section 2.1. The
feasibility of a dosimetry system for PRESAGE®-‐‑IV dosimeters tested against the gold
standard is introduced in Section 2.2. The feasibility was verified by two dose readout
techniques (i) a volume averaged readout by spectrophotometer introduced in Section
2.3.1, and (ii) a 2D projection imaging in a high-‐‑resolution (50 micron) telecentric optical
system introduced in Sections 2.3.2-‐‑2.3.3. The quantitative comparison between the 2D
optical scanning technique and the spectrophotometer is described in Section 2.3.4. The
applicability of PRESAGE®-‐‑IV dosimeters in-‐‑vivo was tested in six HDR intracavitary
gynecological brachytherapy treatments described in Section 2.4 and compared to
Eclipse TPS dose estimations in Section 2.4.1.
2.1 Characterization of PRESAGE®-IV dosimeters
The characterization of PRESAGE®-‐‑IV dosimeters including a description of the
gold standard, dose response, energy response, temporal and temperature stabilities and
the effect of cuvette material on PRESAGE®-‐‑IV are introduced in this section.
2.1.1 Gold standard method using optical cuvettes
A cuvette is a small tube of square cross section, sealed at one end and made of
![Page 26: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/26.jpg)
12
clear plastic designed to hold samples for spectroscopic experiments. A
spectrophotometer measures intensity as a function of the light source wavelength and
is commonly used for the measurement of transmittance of solutions, and transparent or
opaque solids (Figure 5). The spectrophotometer’s simple, convenient, and reliable
readout over a precise and standardized distance and the ability to choose a specific
wavelength made it a perfect choice for measuring the change in optical density of
PRESAGE® material. PRESAGE® prepared in 1x1x4 cm3 cuvettes has been investigated
and provides reliable linear optical response to radiation dose, reproducibility, and
temporal stability [11].
Figure 5: 1x1x4 cm3 cuvette (left) and Genesys® 20, ThermoSpectronic® spectrophotometer (right).
2.1.2 Dose response
To measure the dose response, a standard calibration treatment (Section 2.2.1)
was delivered to three formulations of PRESAGE®. The sensitivity of each formulation
was determined by irradiating 1 x 1 cm plastic cuvettes filled with PRESAGE® with a
6MV photon beam, to doses in the range of dose per fraction used in HDR
![Page 27: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/27.jpg)
13
brachytherapy (1-‐‑15 Gy) (Figure 6) using a standard treatment set up. In gynecology, the
typical range of dose per fraction can be between 5.5-‐‑6 Gy [3]. A Genesys® 20,
ThermoSpectronic® spectrophotometer was used to measure the optical density of each
sample, using an empty cuvette as a zero. Cuvettes were scanned prior to the treatment
and irradiated to various known doses before being re-‐‑scanned. A linear calibration
curve was calculated by relating the delivered dose to the optical density change with
the slope being the sensitivity.
Figure 6: PRESAGE® filled cuvettes irradiated with a 6MV photon beam to doses of 0-‐‑15 Gy.
2.1.3 Energy response
To measure energy response from brachytherapy sources, three PRESAGE®-‐‑IV
dosimeters were taped onto a 3.5 cm diameter stump cylinder applicator and irradiated
to 7 Gy at a lateral distance of 1.95 cm from (2 mm to the center of PRESAGE®-‐‑IV
diameter) by a 7.046 Curie Ir-‐‑192 source (Figure 7). The PRESAGE®-‐‑IV dosimeters were
wrapped in two layers of Tegaderm™ film dressings. Two Beekley CT-‐‑ Spots® skin
markers 2.3mm and 4mm in diameter were placed on either end of the dosimeter for CT
localization. The applicator was placed in a water bath at room temperature 23°C and a
![Page 28: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/28.jpg)
14
source guide tube was attached to the applicator through the lid of the water bath,
securing the applicator position. The Ir-‐‑192 source was delivered by a Varian®
GammaMed™plus iX remote afterloader (Figure 8). In order to verify the dose in the
Eclipse Brachytherapy TPS, the dosimeters were read using the spectrophotometer
technique described in Section 2.3.1 and the 2D optical scanning technique described in
Section 2.3.2.
Figure 7: Axial Cross Section of Applicator with Dosimeters Visible with Beekley CT-‐‑ Spots® Localizers (left) and Sagittal Cross Section of Applicator with
One Dosimeter Visible.
![Page 29: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/29.jpg)
15
Figure 8: Water Bath containing Source Guide Tube and Applicator with Dosimeters (left) and the Treatment Set Up (right).
2.1.4 Temporal stability
Since previous formulations of PRESAGE® have been known to drift in optical
density over time, the optical density of each PRESAGE®-‐‑filled cuvette was tested for
temporal stability. Using a spectrophotometer, the optical density was measured
immediately after treatment delivery, a half hour, hour and every three days for 9 days
to test stability. Cuvettes were stored in a refrigerator (3 to 5°C) to improve stability by
minimizing kinetic effects that could lead to optical density drift. To avoid direct
temperature dependence effects while measuring optical density, cuvettes were taken
out of the refrigerator 1-‐‑2 hours prior to being scanned each day.
2.1.5 Temperature stability
Because the dosimeters will be placed in-‐‑vivo, it was important to test the
![Page 30: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/30.jpg)
16
dosimeters for temperature stability. To determine the change in optical density in
relation to temperature at the time of irradiation, dosimeters were placed in a water tank
at body temperature, 37°C and ±5°C and ±10°C from body temperature (27°C, 32°C,
42°C, 47°C). Two in-‐‑vivo dosimeters were placed side-‐‑by-‐‑side in a 1 cm thick sheet of
bolus material and treated with a 10x10 field at 100 cm SSD and a depth of 1.5 cm below
the water surface. They were placed on top of a 5 cm of solid water to allow for
backscatter. The treatment was delivered using 6 MV photons and a dose rate of 500
monitor units/min. With the field size and SSD set to calibration conditions, scatter
factors were 1.00 and only the percent depth dose table was necessary to calculate the
monitor units for the desired dose at 1.5 cm depth.
2.1.6 Effect of cuvette material on PRESAGE®-IV
In order to make sure there was no interaction between the cuvette material and
PRESAGE®-‐‑IV and its effects on the sensitivity of PRESAGE®-‐‑IV, a batch of PRESAGE®-‐‑
filled cuvettes and PRESAGE® molds in the shape of cuvettes had been created (Figure
9).
![Page 31: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/31.jpg)
17
Figure 9: Example of the PRESAGE® molds in the shape of cuvettes.
Since the path length through PRESAGE®-‐‑filled cuvettes is smaller (1 cm) than
PRESAGE® molds (1.2 cm) (Figure 10), a quantitative comparison needed to be made
between the optical densities of the molds and cuvettes. The sensitivity of PRESAGE®-‐‑
filled cuvettes and PRESAGE® molds was determined by irradiation to doses 1, 2.5, 5
and 8 Gy using a standard treatment set up (Section 2.2.1). The molds and cuvettes were
read using the spectrophotometer technique described in Section 2.3.1. A plot between
the change in optical density of the molds versus the cuvettes was created and the
change in optical density of the molds was multiplied by a factor of 1/1.2 cm to account
for the difference in path length between the molds and cuvettes. A linear slope would
indicate that the change in optical density with dose between the molds and cuvettes
was comparable, and that there was no effect on the sensitivity of PRESAGE® from the
cuvette material.
![Page 32: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/32.jpg)
18
Figure 10: PRESAGE®-‐‑Filled Cuvette (left) and PRESAGE® Mold (right).
2.2 Investigating the feasibility of readout by comparison to gold standard
Since the path length through PRESAGE®-‐‑filled cuvettes is larger than
PRESAGE®-‐‑IV dosimeters read in mineral oil, a quantitative comparison needed to be
made between the optical densities of the dosimeters and cuvettes (Figure 11). The
sensitivity of PRESAGE®-‐‑filled cuvettes and PRESAGE®-‐‑IV dosimeters was determined
by irradiation to different doses using a standard treatment set up (Section 2.2.1). The
dosimeters and cuvettes were read using the spectrophotometer technique described in
Section 2.3.1. A plot between the change in optical density of the dosimeters versus the
cuvettes was created and the change in optical density of the dosimeters was multiplied
by a factor of 1/.4 cm to account for the difference in path length between the dosimeters
and cuvettes. A linear slope would indicate that the change in optical density with dose
between the dosimeters and cuvettes was comparable, and that a novel dosimetry
system for PRESAGE®-‐‑IV dosimeters is feasible.
![Page 33: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/33.jpg)
19
Figure 11: PRESAGE®-‐‑Filled Cuvette (left) and Cuvette with PRESAGE®-‐‑IV (right)
2.2.1 Cuvette and dosimeter irradiation
Pairs of PRESAGE®-‐‑IV dosimeters and cuvettes were placed side-‐‑by-‐‑side and
treated with a 10 x 10 cm field at 100 cm SSD and a depth of 5.5 cm (Figure 12) between
doses of 1-‐‑15 Gy (Table 1). The treatment was delivered using 6 MV photons from a
Varian Clinac® 600C/D and a dose rate of 600 monitor units/min. The dosimeters and
cuvettes were placed in a 1 cm thick sheet of bolus material, with two 1 cm thick sheets
of bolus material below to reduce any air gap created by the difference in thicknesses of
the cuvettes and dosimeters. 10 cm of solid water was placed below to allow backscatter
and 5 cm of solid water above to achieve the 5.5 cm depth to the center of the 4mm
dosimeters and 1 cm cuvettes. With the field size and SSD set to calibration conditions,
scatter factors were 1.00 and only the percent depth dose table was necessary to calculate
the monitor units for the desired dose at 5.5 cm depth.
![Page 34: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/34.jpg)
20
Figure 12: Side View of the Standard Treatment Set-‐‑Up
Table 1: Monitor units delivered to 1.5% DEA formulation dosimeters and cuvettes
Cuvette and
Dosimeter
1 2 3 4 5 6 7 8
Dose (Gy) 0 1 2.5 5 8 10 12 15 MU 0 119 297 593 949 1186 1423 1779
2.3 2D optical scanning and spectrophotometer reading
The change in optical density of each dosimeter is proportional to dose [11]. To
determine the change in optical density, each dosimeter must be scanned before and
after irradiation. PRESAGE®-‐‑IV dosimeters were scanned to measure optical density
using a 2D projection telecentric optical system developed in house: the Duke Micro
Optical scanner (DMicrOS) and a spectrophotometer (Genesys® 20, ThermoSpectronic®).
The DMicrOS uses visible light and a CCD camera to take single projection images at 50-‐‑
![Page 35: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/35.jpg)
21
micron resolution. The spectrophotometer measures the absorbance at the peak
absorption wavelength of 633 nm.
2.3.1 Spectrophotometer technique
To avoid direct temperature dependence effects while measuring optical density,
cuvettes were taken out of the refrigerator 1-‐‑2 hours prior to being scanned each day
and the Genesys® 20, ThermoSpectronic® was turned on 1 hour before performing the
scan to let the light source warm up. The spectrophotometer was then set to the peak
absorption wavelength of 633 nm. An empty cuvette is used as a zero, and a cuvette
filled with mineral oil was read before and after each scan to ensure no fluctuation in
absorbance affected the readings. The dosimeter is placed in a holder and immersed in a
1x1x4 cm cuvette filled with mineral oil, which matches the refractive index of the
dosimeter to minimize bending of the light at the dosimeter-‐‑fluid interface. The cuvette
is placed in the spectrophotometer and the absorbance is read, which is directly
proportional to the thickness of the sample and to the concentration of the absorbing
material in the sample (Figure 13). A Kimwipe® with ethanol, following a dry Kimwipe®
is used to remove any residual oil that may appear outside of the cuvette between
readings.
![Page 36: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/36.jpg)
22
Figure 13: From left to right: Cuvette filled with mineral oil, Styrofoam holder and dosimeter, holder in cuvette, cuvette holder, spectrophotometer.
2.3.2 2D optical scanning technique
The DMicrOS system, a modified version of the Duke Large field of view
Optical-‐‑CT Scanner (DLOS) [15] illustrated in Figure 15, uses a 2W LED light source
behind a narrow band filter, giving the source a uniform flood field with wavelength of
633 nm. A 79 mm diameter telecentric lens provided a central region of parallel light
where the dosimeter can be imaged free of object magnification effects. The imaging lens
has a magnification of 0.25X and collimates any light with more than a 0.1-‐‑degree
deviation from the optical axis, effectively minimizing scattered light contributions. A
1608 x 1208 CCD based Basler camera is used for imaging.
![Page 37: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/37.jpg)
23
Figure 14: Diagram of the DMicrOS System [15].
A jig (Figure 15) was created during the time of this thesis work to hold two
PRESAGE®-‐‑IV dosimeters in a fluid bath with a matched refractive index to minimize
bending of the light. The jig fits onto the tank of the optical system in only one
orientation, guaranteeing registration between pre and post scans.
Figure 15: Diagram of the jig (designed during this thesis work) that holds in-‐‑vivo dosimeters in tank.
One projection was taken and to reduce noise, each projection was captured 100
times and averaged. After removing the dosimeter from the system, an image of the
flood field was captured with 100 averages. The light source was then blocked to acquire
![Page 38: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/38.jpg)
24
a dark field with 100 averages. The optical density of the scan was calculated by
subtracting out the background (dark) and normalizing by the flood to remove
inhomogeneity, as shown in the equation below:
2.3.3 2D optical scanner image registration
A graphical user interface (GUI) was created in MATLAB in order to ensure
registration between the pre and post scan images. The DMicrOS Image Registration
GUI reads in the pre and post scan images and converts them to the change in optical
density. The GUI displays the pre and post scan images and the change in optical
density between the two images. The GUI allows the user to move the post scan image
over the pre scan image through rotation and translation in order to visually observe the
image registration.
After the images are registered, a line profile is taken along the center of each
dosimeter to measure the change in optical density at approximately the same length (10
mm) and location (3 mm from the bottom of the dosimeter) as the spectrophotometer
(Figure 16). The mean of the line profile is calculated for each dosimeter and the average
of the two dosimeters is taken to get an average optical density reading.
![Page 39: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/39.jpg)
25
Figure 16: An Example of a Line Profile of Two Dosimeters Irradiated to 10 Gy Using the Standard Treatment Setup Described in Section 2.2.1. The Line Profile Was Taken Along the Center of Each Dosimeter to Measure the Change in Optical Density at Approximately the Same Length (10 mm) and Location (3 mm from the Bottom of
the Dosimeter) as the Spectrophotometer.
2.3.4 Quantitative comparison between the spectrophotometer and 2D optical scanner
To verify the 2D optical scanning technique (Section 2.3.2) with the
spectrophotometer technique (Section 2.3.1) a quantitative comparison needed to be
made between the optical densities of the dosimeters. The sensitivity of PRESAGE®-‐‑IV
dosimeters was determined by irradiation to doses between 1-‐‑12 Gy using a standard
treatment set up (Section 2.2.1). The dosimeters were read within minutes of each other
using the spectrophotometer and 2D optical scanning techniques. A linear calibration
![Page 40: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/40.jpg)
26
curve was calculated by relating the delivered dose to the optical density change with
the slope being the sensitivity.
2.4 Patient treatment
PRESAGE®-‐‑IV dosimeters were tested on their ability for in-‐‑vivo dose
verification for the bladder and rectum in six patients undergoing gynecologic
intracavitary HDR brachytherapy treatment. Dosimeters were taken out of the
refrigerator one hour before being delivered to the physician for use in patient
treatment. To ensure no direct contact with patient tissue, PRESAGE®-‐‑IV dosimeters
were wrapped with two layers of Tegaderm™ film dressings. Two Beekley CT-‐‑ Spots®
skin markers 2.3mm and 4mm in diameter were placed on either end of the dosimeter
for CT localization (Figure 17).
Figure 17: PRESAGE®-‐‑IV dosimeter with and without Tegaderm™ and Beekley CT-‐‑ Spots® localizers.
Dosimeter attachment is applicator dependent. For the T&R treatment, the
physician taped the dosimeters on a Radiadyne® Alatus™ Vaginal Balloon, one facing the
![Page 41: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/41.jpg)
27
bladder and the other facing the rectum and inserted the treatment applicator into the
patient (Figure 18). For cylinder and stump treatments the physician taped the
dosimeters directly on the applicator nearest to the OARs (Figure 19). For the
multichannel Capri™ applicator, the physician placed the dosimeters in the channels
facing the OARs (Figure 20). After treatment, the dosimeters were removed by the
physician and taken back to the lab and read on the spectrophotometer using the
method described in section 2.3.1. The time from end of treatment to the first reading
was approximately 10 minutes.
Figure 18: Example of Dosimeter Placed on Alatus Balloon Packing System (top) and Dosimeters In-‐‑Vivo Near OARs in T&R Treatment (left and right)
![Page 42: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/42.jpg)
28
Figure 19: Example of Dosimeters Placed on Cylinder Applicator Axial (left) and Sagittal (right) Views.
Figure 20: Example of Dosimeter Placed inside Capri Applicator Channel Axial (left) and Coronal View of Dosimeter with Point Doses (right).
2.4.1 Eclipse dose verification
In order to verify the dose measured by the dosimeter, a dose line profile was
taken along the length of the dosimeter in the Eclipse Brachytherapy TPS (Figures 18-‐‑
19). Since the spectrophotometer takes an average absorbance reading 3mm from the
bottom and 10mm along the height of the dosimeter, the dose profile in Eclipse was
![Page 43: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/43.jpg)
29
taken along the same dimensions to keep measurement consistency. The dose profile
from Eclipse was then averaged and compared to the measured dose.
![Page 44: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/44.jpg)
30
3. Results and Discussion Results reported in this section include the sensitivities of the five formulations
of PRESAGE® used in this experiment (Section 3.1) using the techniques described in
Sections 2.2 and 2.3. The dose response, energy response, temporal and temperature
stabilities, and the effect of the cuvette material on PRESAGE® are reported in Section 3.1
as well. Proof of feasibility of novel in-‐‑vivo dosimeters compared to the gold standard
cuvettes is reported in Section 3.2. Comparison between the 2D optical scanning and
spectrophotometer technique results are reported in Section 3.3. Finally, the results of
the comparison between the measured doses from the PRESAGE®-‐‑IV dosimeters and
Eclipse TPS in patient treatments is reported in Section 3.4. Table 2 reports the
formulations of the dosimeters irradiated in this thesis.
Table 2: Overview of PRESAGE®-‐‑IV dosimeters irradiated in this work.
Date Manufactured Formulation Dose (Gy)
Sensitivity Rank
Change in OD Over Time
10/28/2011 D21 2.36, 5, 7.50 2 Increase
05/04/2012 2% O-‐‑MeO-‐‑DMA
1.003, 2.504, 5, 8.002 1 Stable
10/01/2012 1.7% O-‐‑MeO-‐‑DMA 2.504, 8.002 6 Stable
11/27/2012 1.7% O-‐‑MeO-‐‑DEA
1.003, 2.504, 5, 8.002 5 Stable
01/31/2013 1.5% O-‐‑MeO-‐‑DEA
1.003, 2.504, 5, 8.002 3 Stable
02/14/2013 1.5% O-‐‑MeO-‐‑DEA
1, 2.5, 5, 8, 10, 12, 15 4 Stable
![Page 45: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/45.jpg)
31
3.1 PRESAGE® Sensitivity and stabilities
The D21 formulation and four formulations (2% O-‐‑MeO-‐‑DMA, 1.7% O-‐‑MeO-‐‑
DMA, 1.7% O-‐‑MeO-‐‑DEA and 1.5% O-‐‑MeO-‐‑DEA) [16] were irradiated as described in
Section 2.2.1. In Section 3.1.1, the dose response, or sensitivity, of each formulation are
reported. In Section 3.1.2, the KV energy response of the experiment described in Section
2.1.3 is reported. The temporal stability and temperature stability is reported in Sections
3.1.3 and 3.1.4, respectively. The results of the effect of the cuvette material on
PRESAGE®-‐‑IV is reported in Section 3.1.5.
3.1.1 Dose response
The sensitivity changes for all five formulations are presented in Figure 21. The
2% O-‐‑MeO-‐‑DMA formulation had the highest sensitivity with a slope of 0.0494 ± 0.0007
change in OD/Gy/cm. Next was the D21 formulation, which had a sensitivity of 0.0375 ±
0.001 change in OD/Gy/cm. The 1.5% O-‐‑MeO-‐‑DEA (1/31/13) formulation had a
sensitivity of 0.0356 ± 0.0015 change in OD/Gy/cm. The 1.5% O-MeO-DEA (2/14/13)
formulation had a sensitivity of 0.0299 ± 0.0005 change in OD/Gy/cm, and 83% of the
sensitivity of the 1.5% O-MeO-DEA (1/31/13) formulation. The 1.7% O-‐‑MeO-‐‑DEA
(11/27/12) formulation had a sensitivity of 0.0179 ± 0.0002 change in OD/Gy/cm and the
1.7% O-‐‑MeO-‐‑DMA (10/1/12) formulation had a sensitivity of 0.0073 ± 8.9E-‐‑05 change in
OD/Gy/cm and 14.8% of the sensitivity of the 2% O-‐‑MeO-‐‑DMA formulation. All were
within the 95% confidence level.
![Page 46: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/46.jpg)
32
Figure 21: Dose Response (Sensitivity) of All Five Formulations.
3.1.2 Energy Response
The sensitivity for the 1.5% O-‐‑MeO-‐‑DEA (2/14/13) dosimeter formulation is
presented in Figure 22 with a sensitivity slope of 0.0136 ± 0.0002 change in OD/Gy/cm.
This formulation was calibrated using a 6 MV photon beam using the technique
described in Section 2.2.1, and the slope of this sensitivity curve was used to calculate
the dose measured by the dosimeters irradiated by a KV Iridium-‐‑192 brachytherapy
source (Figure 23).
A dose of 7 Gy was delivered to the center of three dosimeters, with an 8.4 Gy
isodose line at the applicator/dosimeter wall indicating a 1.4 Gy dose drop off in 2 mm.
The average measurement using the spectrophotometer was 7.05 ± 0.11 Gy (0.66%
![Page 47: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/47.jpg)
33
difference), and the average measurement using a line profile in the 2D optical scanning
method read a dose of 7.18 ± 0.32 Gy (2.6% difference). The difference between the
Eclipse dose and the measured dose using the spectrophotometer reading technique
shows that the overestimation of dose is within the dose fall off range of 1.4 Gy with a
value of 1.35 Gy. The 2D Optical scanning technique has a larger difference of 2.6% and
the overestimation of dose is also within the dose fall off range of 1.4 Gy with a value of
1.22 Gy. The higher dose is indicative of its slightly (0.74%) greater sensitivity than the
spectrophotometer method, but the main reason could have been due to the reading
technique. The change in optical density line profile was taken in the center of the
dosimeter, but the positioning of the dosimeter in the jig could have allowed it to be
read on the side that had been exposed to the larger dose.
![Page 48: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/48.jpg)
34
Figure 22: The Sensitivity for the 1.5% O-‐‑MeO-‐‑DEA (2/14/13) Dosimeter Formulation.
![Page 49: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/49.jpg)
35
Figure 23: Dosimeter Dose Measurement of a KV Brachytherapy Source Using a MV Calibration Curve.
3.1.3 Temporal stability
In Figure 24, the results show that the 1.7% O-‐‑MeO-‐‑DEA formulation is stable
over time. This formulation had an initial sensitivity of 0.0179 ± 0.0002 change in
OD/Gy/cm in the 95% confidence level. The sensitivity was tracked daily over nine days
and decreased to 0.0178 change in OD/Gy/cm nine days after irradiation. The difference
between the initial sensitivity and the sensitivity nine days after irradiation was found to
be within the 95% confidence level. The differences could have been attributed to a small
error in the reading technique, which could have been caused by residual mineral oil on
the outside of the cuvette.
![Page 50: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/50.jpg)
36
Figure 24: Dose Response (Sensitivity) Temporal Stability of the 1.7% O-‐‑MeO-‐‑DEA (11/27/12) Formulation.
3.1.4 Temperature stability
The change in optical density with temperature of the D21 formulation at the
time of irradiation is illustrated in Figure 25. There is a 35% difference between the
optical density of the dosimeter at 27°C and 47°C and a 5% increase in the change in
optical density from body temperature 37°C to 47°C. Extrapolating the line to room
temperature 22°C, there is a 30% increase in optical density from room temperature to
body temperature 37°C indicating a temperature dependence that makes the dosimeters
more sensitive at higher doses.
![Page 51: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/51.jpg)
37
Figure 25: Change in OD with Temperature for the D21 Formulation.
3.1.5 Effect of the cuvette material on PRESAGE®-IV
The change in the optical density between the molds and the cuvettes for the
1.5% O-‐‑MeO-‐‑DEA formulation had a slope of 0.9108 ± 0.05 in the 95% confidence
interval. The results shown in Figure 26 shows that there is a maximum difference of
10% between the sensitivities of the molds and the cuvettes, with the difference getting
larger as the dose increases from 1-‐‑8 Gy. This 10% difference could be attributed to the
aberrations (tacky surface) on the surface of the molds compared to the smooth surface
of the cuvette.
![Page 52: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/52.jpg)
38
Figure 26: The Change in Optical Density of the Molds vs. Cuvettes for the 1.5% O-‐‑MeO-‐‑DEA Formulation (1/31/13).
3.2 Results of the feasibility of readout by comparison to gold standard
The results of the change in optical density between the gold standard 1x1x4cm
optical cuvettes filled with PRESAGE® and PRESAGE®-‐‑IV dosimeters read in mineral
oil are reported in Figure 27. The 1.5% O-‐‑MeO-‐‑DEA (2/14/13) formulation was used for
the proof of feasibility for its temporal stability and good sensitivity. In Figure 27, the
slope 0.8703 ± 0.0192 indicates that there is a linear relationship between the cuvettes
and in-‐‑vivo dosimeters within the 95% confidence interval and a dosimetry system for
PRESAGE®-‐‑IV dosimeters is feasible.
![Page 53: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/53.jpg)
39
Figure 27: The Change in Optical Density of the Dosimeters vs. Cuvettes for the 1.5% O-‐‑MeO-‐‑DEA Formulation (2/14/13).
3.3 Quantitative comparison between the 2D optical scanner and spectrophotometer
The results of the sensitivity difference between the 2D optical scanner and the
spectrophotometer for PRESAGE®-‐‑IV dosimeters read in mineral oil for dosimeters
irradiated between the doses of 1-‐‑12 Gy are reported in Figure 28. The 1.5% O-‐‑MeO-‐‑
DEA (2/14/13) formulation was used for this comparison for its temporal stability and
good sensitivity. In Figure 28, the sensitivity, or change in OD/Gy/cm slope of the
DmicrOS technique was found to be 0.0136 ± 0.0002 and for the spectrophotometer
0.0135 ± 0.0002, which is a 0.74% difference in sensitivity within the 95% confidence
interval. The results show that the 2D optical scanning method is slightly more sensitive
![Page 54: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/54.jpg)
40
than the spectrophotometer method as predicted and that the DmicrOS system is a
feasible method of reading PRESAGE®-‐‑IV dosimeters.
Figure 28: The Comparison in Sensitivity of the DmicrOS system and the Spectrophotometer for the 1.5% O-‐‑MeO-‐‑DEA Formulation (2/14/13).
3.4 Patient treatment
Figure 29 shows the dose response curve of the D21 formulation taken at body
temperature (37°C) with a control at room temperature using the technique described in
Section 2.1.4. This dose response curve was used to estimate the dose in the following
patient treatment plots using the D21 formulation (Figures 30-‐‑33).
![Page 55: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/55.jpg)
41
Figure 29: Dose Response (Sensitivity) of the D21 Formulation at Body Temperature.
The results of T&R patient 6 treatment are displayed in Figure 30. The T&R
patient measured dose shows a 0% and 2.8% difference between the measured doses
and the Eclipse doses for the bladder and rectum, respectfully. These results are the best
results received from using PRESAGE®-‐‑IV dosimeters for in-‐‑vivo dose verification. In
the following patient results, the differences between measured doses and Eclipse doses
for OARs are quite large. A number of reasons for these large differences could be due to
the positional stability of the dosimeters, organ motion, and patient motion for the
duration of treatment (2-‐‑4 hours on average).
![Page 56: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/56.jpg)
42
Figure 30: T&R Patient 6 Treatment Dose Measurement Using the D21 Formulation.
The results of the Capri patient 1 fraction (fx) 1 treatment are displayed in Figure
31. The Capri patient measured dose shows a 12% and 20% difference between the
measured doses and the Eclipse doses for the bladder and rectum, respectfully. The
results of the Capri patient 1 fraction 2 treatment are displayed in Figure 32. The Capri
patient measured dose shows a 4% and 7% difference between the measured doses and
the Eclipse doses for the bladder and rectum, respectfully. The results of the Capri
patient 1 fraction 3 treatment are displayed in Figure 33. The Capri patient measured
dose shows a 9% and 19% difference between the measured doses and the Eclipse doses
for the bladder and rectum, respectfully.
![Page 57: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/57.jpg)
43
Figure 31: Capri Patient 1 fx 1 Dose Measurement Using the D21 Formulation.
Figure 32: Capri Patient 1 fx 1 Dose Measurement Using the D21 Formulation.
![Page 58: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/58.jpg)
44
Figure 33: Capri Patient 1 fx 3 Dose Measurement Using the D21 Formulation.
The variation in dose verification over the course of three fractions shows that
patient organ motion can be a big factor in in-‐‑vivo dose verification. The rectal dose
differences remained the larger of the two OARs indicating that rectal filling could be a
cause for these differences.
Figure 34 shows the dose response curve of the 2% O-‐‑MeO-‐‑DMA formulation
taken at body temperature (36.8°C) with a control at room temperature using the
technique described in Section 2.1.4. This dose response curve was used to estimate the
dose in the following patient treatment plots using the 2% O-‐‑MeO-‐‑DMA formulation
(Figures 35-‐‑39).
![Page 59: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/59.jpg)
45
Figure 34: Dose Response (Sensitivity) of the 2% O-‐‑MeO-‐‑DMA Formulation at Body Temperature.
The results of the T&R patient 2 fraction 1 treatment are displayed in Figure 35.
The T&R patient measured dose shows a 38% and 19% difference between the measured
doses and the Eclipse doses for the bladder and rectum, respectfully. The results of the
T&R patient 2 fraction 2 treatment are displayed in Figure 36. The T&R patient
measured dose shows a 34% and 73% difference between the measured doses and the
Eclipse doses for the bladder and rectum, respectfully.
![Page 60: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/60.jpg)
46
Figure 35: T&R Patient 2 fx 1 Dose Measurement (2% O-‐‑MeO-‐‑DMA)
Figure 36: T&R Patient 2 fx 2 Dose Measurement (2% O-‐‑MeO-‐‑DMA)
![Page 61: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/61.jpg)
47
The results of the stump patient 3 treatment are displayed in Figure 37. The
stump patient measured dose shows a 13% and 11% difference between the measured
doses and the Eclipse doses for the bladder and rectum, respectfully. The results of the
stump patient 4 treatment are displayed in Figure 38. The stump patient measured dose
shows a 39% and 14% difference between the measured doses and the Eclipse doses for
the bladder and rectum, respectfully. The results of the cylinder patient 5 treatment are
displayed in Figure 39. The stump patient measured dose shows a 19% and 30%
difference between the measured doses and the Eclipse doses for the bladder and
rectum, respectfully.
Figure 37: Stump Patient 3 Dose Measurement (2% O-‐‑MeO-‐‑DMA)
![Page 62: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/62.jpg)
48
Figure 38: Stump Patient 4 Dose Measurement (2% O-‐‑MeO-‐‑DMA)
Figure 39: Cylinder Patient 5 Dose Measurement (2% O-‐‑MeO-‐‑DMA)
![Page 63: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/63.jpg)
49
4. Conclusions The aims of the project were to evaluate the feasibility of a dosimetry system for
PRESAGE®-‐‑IV dosimeters by comparing it to the gold standard using PRESAGE®-‐‑filled
optical cuvettes in spectrophotometry and validating the DMicrOS 2D optical scanning
system, and to demonstrate its utility as an in-‐‑vivo dosimeter for brachytherapy
treatment. The aim of evaluating the feasibility of a dosimetry system was met, but there
is substantial room for future investigation of PRESAGE®-‐‑IV dosimeters in patient
treatment.
The spectrophotometer reading of the change in optical densities of the 1.5% O-‐‑
MeO-‐‑DEA formulation of the gold standard 1x1x4cm optical cuvettes filled with
PRESAGE® and PRESAGE®-‐‑IV dosimeters in mineral oil had a linear slope 0.8703 ±
0.0192 when plotted against each other (Figure 27), and indicated that a dosimetry
system for in-‐‑vivo dosimeters is feasible. The results provide verification of the potential
that this dosimetry system has for easily and quickly reading PRESAGE®-‐‑IV dosimeters.
In Figure 28, the sensitivity, or change in OD/Gy/cm slope of the DMicrOS technique
was found to be 0.0136 ± 0.0002 and for the spectrophotometer 0.0135 ± 0.0002, which is
a 0.74% difference in sensitivity within the 95% confidence interval, which showed that
the two techniques are comparable.
With the positive results of the feasibility of the dosimetry system, came the
negative results of patient treatment. The estimated doses for patient 6 treatment using a
![Page 64: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/64.jpg)
50
T&R (Figure 30) showed a 0% and 2.8% difference between the estimated doses and the
Eclipse doses for the bladder and rectum, respectfully. These results were the best
results received from using PRESAGE®-‐‑IV dosimeters for in-‐‑vivo dose verification. In
the following patient results, the differences between measured doses and Eclipse doses
for OARs were quite large. A number of reasons for these large differences could be due
to the positional stability of the dosimeters, organ motion, and patient motion for the
duration of treatment (2-‐‑4 hours on average). More patient trials need to be investigated
to show proof of concept for the dosimeter as a reliable method of dose verification for
OARs in HDR brachytherapy treatments.
There are several future directions to expand on the results of this work. Since
PRESAGE®-‐‑IV has the capability of 3D readings, the 2D optical scanning system can be
turned into an optical-‐‑CT system capable of reading the dose gradients that are surely
present in the dosimeters. Also, creating a standard procedure of application of in-‐‑vivo
dosimeters on intracavitary treatment applicators to eliminate unknowns is a high
priority. Once a standard procedure is achieved, dose estimates can become more
reliable and the differences from Eclipse measurements will decrease.
![Page 65: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/65.jpg)
51
References 1. McJury, M., Oldham, M., Cosgrove, V., et al. 2000. Review article: Radiation
dosimetry using polymer gels: methods and applications. The British Journal of Radiology 73: 919-‐‑929.
2. Nath, Ravinder, Lowell L Anderson, Jerome A Meli, Arthur J Olch, Judith Anne
Stitt, and Jeffrey F Williamson. 1997. Code of Practice for Brachytherapy Physics : Report of the AAPM Radiation Therapy Committee Task Group No . 56. 24 (56): 1557–1598.
3. Hendee, W. et al. 2013. Comprehensive Brachytherapy: Physical and Clinical Aspects.
CRC Press. p. 370-‐‑71 4. Thomadson, B.R., (2000). Achieving Quality in Brachytherapy. London, England:
IOP Publishing. 5. Nag, S, B Erickson, B Thomadsen, C Orton, J D Demanes, and D Petereit. 2000.
The American Brachytherapy Society Recommendations for High-‐‑dose-‐‑rate Brachytherapy for Carcinoma of the Cervix. International Journal of Radiation Oncology, Biology, Physics 48 (1) (August 1): 201–11.
6. Palmer, Antony, David Bradley, and Andrew Nisbet. 2012. Physics-‐‑aspects of Dose
Accuracy in High Dose Rate (HDR) Brachytherapy: Source Dosimetry, Treatment Planning, Equipment Performance and in Vivo Verification Techniques. Journal of Contemporary Brachytherapy 4 (2) (July): 81–91.
7. Williamson, J. 2008. Current brachytherapy quality assurance guidance: does it meet
the challenges of emerging image-‐‑guided technologies? International Journal of Radiation Oncology, Biology, and Physics 71 (1): 518-‐‑522.
8. Devlin, P.M 2007. Brachytherapy: Applications and Techniques. Lippincott Williams
& Wilkins p. 233-‐‑35. 9. Haughey A, Coalter G, Mugabe K. Evaluation of linear array MOSFET detectors for
in vivo dosimetry to measure rectal dose in HDR brachytherapy. Australas Phys Eng Sci Med 2011; 34: 361-‐‑366.
![Page 66: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/66.jpg)
52
10. Reniers B, Landry G, Eichner R, Hallil A, Verhaegen F. (2012). In vivo dosimetry for gynaecological brachytherapy using a novel position sensitive radiation detector: feasibility study. Med Phys. 2012 Apr. 39(4):1925-‐‑35.
11. Guo, P Y, J a Adamovics, and M Oldham. 2006. Characterization of a New
Radiochromic Three-‐‑dimensional Dosimeter. Medical Physics 33 (5): 1338–1345. 12. Guo, P., J. Adamovics, and M. Oldham, A practical three-‐‑dimensional dosimetry
system for radiation therapy. Med Phys, 2006. 33(10): p. 3962-‐‑72. 13. Pierquet, M., et al., An investigation into a new re-‐‑useable 3D radiochromic dosimetry
material, Presage. J Phys Conf Ser, 2010. 250(1): p. 1-‐‑4. 14. Brown, S., et al., Radiological properties of the PRESAGE and PAGAT polymer
dosimeters. Appl Radiat Isot, 2008. 66(12): p. 1970-‐‑4. 15. Thomas, A., et al., Commissioning and benchmarking a 3D dosimetry system for
clinical use. Med Phys, 2011. 38(8): p. 4846-‐‑57. 16. Juang, T., et al., Customising PRESAGE for Diverse Applications. IC3D Dose 2012.
Shangri La Hotel, Sydney, Australia. 5 November 2012. 17. Ahmed, M.I., et al., Initial Experience with Alatus Vaginal Balloon Packing (VBP)
Compared to Standard Vaginal Gauze Packing (SGP) for Tandem and Ovoid Brachytherapy Treatment of Carcinoma of the Cervix. International Journal of Radiation Oncology, Biology, and Physics 78 (3): S409.
18. Rockey, W.M., et al., The Dosimetric Impact of Vaginal Balloon Packing on
Brachytherapy for Gynecological Cancer. International Journal of Radiation Oncology, Biology, and Physics 81 (18): S457.
19. Kirisits, C., et al., 2011 “Physics for Image-‐‑Guided Brachytherapy” Gynecologic
Radiation Therapy: Novel approaches to Image-‐‑Guidance and Management. Springer Heidelberg Dordrecht. London. New York. p.143.
20. Petric, P. et al., 2011 “Physics for Image-‐‑Guided Brachytherapy” Adaptive
Contouring of the Target Volume and Organs at Risk. Springer Heidelberg Dordrecht. London. New York. p.99.
21. Lambert, J. et al. 2007 “In vivo dosimeters for HDR brachytherapy: A comparison of a
![Page 67: 4172013 AdriaVidovic Thesis - Duke University](https://reader033.fdocuments.in/reader033/viewer/2022061607/629f9321f633f016032a85e7/html5/thumbnails/67.jpg)
53
diamond detector, MOSFET, TLD, and scintillation detector” Med Phys, 2007. 34(5): 1759-‐‑65.
22. Varian Medical Systems, 2011. “Varian BrachyTherapy Applicators and Accessories Catalog” Varian Medical Systems, Inc.
23. Pierquet, M., et al., On the Feasibility of Verification of 3D Dosimetry Near
Brachytherapy Sources Using PRESAGE/Optical-‐‑CT. J Phys Conf Ser, 2010. 250(1): 012091-‐‑1-‐‑012091-‐‑5