PRODUCT INFORMATION
Applications:Features:
TGS 4160 - for the detection of Carbon Dioxide
The gure below represents typical sensitivity characteristicsof TGS4160. The Y-axis is indicated as EMF which is denedas follows:
EMF=EMF1 - EMF2where
EMF1=EMF in 350 ppm CO 2 EMF2=EMF in listed gas concentration
The gure below shows typical humidity dependency for anenergized sensor. Again, the Y-axis is indicated as EMFwhich is dened as follows:
EMF=EMF1 - EMF2where
EMF1=EMF in 350 ppm CO 2 EMF2=EMF in 1000ppm CO 2
* Air quality control* CO 2 control in agricultural applications* CO 2 monitoring
The TGS4160 is a hybrid sensor unit composed of a carbon dioxide sensitiveelement and a thermistor. A wide range of 350~50,000ppm of carbon dioxide canbe detected by TGS4160, making it ideal for usage in a variety of applications.
The CO 2 sensitive element consists of a solid electrolyte formed between twoelectrodes, together with a printed heater (Pt) substrate. By monitoring thechange in electromotive force (EMF) generated between the two electrodes,
it is possible to measure CO 2 gas concentration.Adsorbent (zeolite) is lled between the internal cover and the outer cover forthe purpose of reducing the inuence of interference gases.
TGS4160 exhibits a linear relationship between EMF and CO 2 gas concentrationon a logarithmic scale. The sensor displays good long term stability and showsexcellent durability against the effects of high humidity.
* High selectivity to CO 2* Low dependency on humidity* Long life
Humidity Dependency:Sensitivity Characteristics:
0
50
100
150
100 1000 10000 100000
CO2COEtOH
Gas Concentration (ppm)
220
250
280
310
340
370
400
430
0
10
20
30
40
50
60
70
0 20 40 60 80 100
Relative Humidity (%)
EMF (350ppm CO2)
EMF (350ppm CO2 - 1000ppm CO2)
IMPORTANT NOTE: OPERATING CONDITIONS IN WHICH FIGARO SENSORS ARE USED WILL VARY WITH EACH CUSTOMERS SPECIFIC APPLICATIONS. FIGARO STRONGLYRECOMMENDS CONSULTING OUR TECHNICAL STAFF BEFORE DEPLOYING FIGARO SENSORS IN YOUR APPLICATION AND, IN PARTICULAR, WHEN CUSTOMERS TARGETGASES ARE NOT LISTED HEREIN. FIGARO CANNOT ASSUME ANY RESPONSIBILITY FOR ANY USE OF ITS SENSORS IN A PRODUCT OR APPLICATION FOR WHICH SENSOR HASNOT BEEN SPECIFICALLY TESTED BY FIGARO.
-
7/31/2019 4160 PDF
2/2
F I G R O A
T G S 4 1 6 0
1
2
34
5
6
17.7 1.0
24.1 1.0
Gr
24.3 1.0
6.0 0.5
6.0 0.5
5.8 0.3
White Dot
45 130'
45 130'
1.0 0.1
9.5 0.2
rebmunledoM 0614SGT
epyttnemelegnisneS etylortcelediloS
sesagtegraT edixoidnobraC
egnarnoitcetedlacipyT mpp000,05~053
lacirtcelEscitsiretcarahc
tsetdradnatsrednusnoitidnoc
ecnatsiserretaeH RH 5.11 1.1 .pmetmoorta
tnerrucretaeH IH Am052.xorppa
rewopretaeHnoitpmusnoc PH W52.1.xorppa
egatlovretaeH VH 0.5 )CD(V2.0
ecrofevitomortcelE FME OCmpp053nivm094~022 2
ytivitisneS FME Vm27~44 OCmpp053FME -2OCmpp0053FME 2
scitsiretcarahcrosneSemitesnopseR )eulavlaniffo%09ot(.nim2.xorppa
ycaruccatnemerusaeM .xorppa OCmpp000,1ta%02 2
snoitidnocgnitarepO HR%59~5,C05~01-
snoitidnocegarortS HR%09~5,C06~02-)legacilishtiwgabfoorperutsiomnierots(
tsetdradnatSsnoitidnoc
snoitidnocsagtseT OC 2 riani02ta 56,C2 HR%5
snoitidnoctiucriC VH 0.5= CDV50.0
doirepgninoitidnoCtseterofeb syad7
Structure and Dimensions:
Basic Measuring Circuit:
REV: 01/06
The TGS4160 sensor requires heatervoltage (V H) input. The heater voltage isapplied to the integrated heater in order
to maintain the sensing element at aspecic temperature which is optimal forsensing. Electromotive force (EMF) ofthe sensor should be measured using ahigh impedance (> 100 G) operationalamplier with bias current < 1pA (e.g.Texas Instruments' model #TLC271).Since the solid electrolyte type sensor
Specications:
FIGARO USA, INC.121 S. Wilke Rd. Suite 300Arlington Heights, IL 60005 USAPhone: (1)-847-832-1701Fax: (1)-847-832-1705email: [email protected]
FIGARO ENGINEERING INC.1-5-11 Senba-nishiMino, Osaka 562-8505 JAPANPhone: (81)-72-728-2561Fax: (81)-72-728-0467www.garo.co.jpemail: [email protected]
functions as a kind of battery, theEMF value itself would drift using thisbasic measuring circuit. However, thechange of EMF value (EMF) shows a
stable relationship with the change ofCO 2 concentration. Therefore, in orderto obtain an accurate measurementof CO 2, a special microprocessor forsignal processing should be used withTGS4160. Figaro can provide a specialevaluation sensor module (AM-4) forTGS4160.
Sensing Element Structure:
Top view
Bottom view
(u/m = mm)
DF
C
A
Bottom View (Sensor Element)
A: Lead wires
Side view (Sensor Element)
B : Pt Heater
F : Sealing GlassE : Counter Electrode(Anode)
C : Solid Electrolyte
D : Sensing Electrode(Cathode)
For information on warranty, please referto Standard Terms and Conditions of Saleof Figaro USA Inc.
EMFVH
TGS4160
1 3
6 4
IBIAS < typ. 1pA
GND
2
5
Thermistor
Thermistor
Operational amplifier
+
-