3.3.1.b. Cypher instrument manual (1).pdf

download 3.3.1.b. Cypher instrument manual (1).pdf

of 8

Transcript of 3.3.1.b. Cypher instrument manual (1).pdf

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    1/19

     

    3.3.1. Instrument Manual v 1.0 Page 1 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    !"#$%& ()*

    +,-.&/0%,. *1,/12

    "#$%# &'($#() *)%+,)-+

    3$4,%5 6789 :78;

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    2/19

     

    3.3.1. Instrument Manual v 1.0 Page 2 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    "#$%&'()#%* ++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++

    ,$)& -.()* ++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++

    •  /.(01) 0&)0.&.%23#

    •  /.(01) $)% '0 3# (24&3$430)

    •  51.4) 4.#%21)6)& 3# 7318)&

    •  934'$ 3#*

    o  :.#%21)6)&

    o  /.(01)

    •  51.4) 1.$)&

    •  ;).>)

    •  /32#> 2# %.002#>

    o  @'#2#>

    •  A.$24 .#.1B$2$*

    o  C3'>7#)$$

    o  /)4%23#

    o  D;

    •  ;.%. %&.#$#.%'&)* +++++++++++++++++++++++++++++++

    !"#$%# '()# )*+% ,(-. '##/% )( 0# 1(.2"#)#/ 3+)*+' 4 .(')*% ,-(. )*# %)$-)

    /$)#5

    6, '()7 -#8)-$+'+'9 .+9*) 0# -#:;+-#/5 

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    3/19

     

    3.3.1. Instrument Manual v 1.0 Page 3 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    4/19

     

    3.3.1. Instrument Manual v 1.0 Page 4 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    CONT CT MODE

     

    1. Set up the system:

    •  Turn on controller

    •  Turn on computer

    2. Set up the computer program:

    •  Click on Asylum Research icon on the desktop.

    When you open Asylum Research program the initial window is:

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    5/19

     

    3.3.1. Instrument Manual v 1.0 Page 5 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    In order to continue you must click ‘yes’ to the following question that will appear in thescreen:

    3. Mount probe and place the sample:

    Select the right holder to mount the probe. In this chapter we are explaining theprocess for analysing a sample in air, not liquid; therefore pick the holder with the word

     AIR written on it.

    Slide the cantilever holder in.

    Push down the holder to mount the tip. The little metal arm that holds the cantilever willbe pushed up when you push the holder down. Place cantilever carefully usingtweezers.

    Prepare your sample by mounting it on a metal disk using double sided tape.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    6/19

     

    3.3.1. Instrument Manual v 1.0 Page 6 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    Pull up the handle on the right side of the microscope in order to unlock the door andthen open it.

    Next, move the stage out by pulling up the black handle located on the right side of thestage (see image below).

    Once you have mounted your sample on a metal disk with a sticky-tape. Place the diskon the sample holder showed in the picture below.

    Use the screwdriver that is located on the left and loosen the little screw that is shown inthe image below, so the cantilever holder can be held in place.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    7/19

     

    3.3.1. Instrument Manual v 1.0 Page 7 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    Once you have placed the cantilever holder in the right place, tighten the screw again.

    Once the cantilever holder and the sample are in the right place, slide the stage backand push the black handle down.Close the door and push the handle on the side to lock the door.

    4. Focus on tip:

    Once you press ‘yes’ a new window will appear:

    Bring the sample and the tip closer by using the roulette at the front of the microscopeto put the tip closer.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    8/19

     

    3.3.1. Instrument Manual v 1.0 Page 8 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    On the software at the bottom tool bar there is a camera sign. Click on it so the videowindow comes up.

    If the screen is too dark, the illumination can be adjusted by dragging the cursor at thebottom of the video window (see image above).

    Once you have answered yes to the question above, a new window will appear (EngagePanel). Click on: Focus on Tip.

    In order to focus on the tip you have to use the arrows in the Engage Panel (see pictureabove) to move the focus closer or further.

    Once you are happy with the focus, click on SET (next to Focus on Tip).

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    9/19

     

    3.3.1. Instrument Manual v 1.0 Page 9 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    5. Focus on sample

     At the Engage Panel click on Focus On Sample:

    To focus on the sample you have to use the arrows in the Engage Panel (see pictureabove) to move the focus closer or further.

    Once you are happy with the focus, click on SET (next to Focus on Sample).

    6. Align the laser

    Turn on the laser.

     At the Engage Panel click on Focus On Tip.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    10/19

     

    3.3.1. Instrument Manual v 1.0 Page 10 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    In order to align the laser, click spot on on the video screen. Focus on the tip and thenclick on the location of the tip where you want the laser to be:

    7. Adjust photodiode signal:

     At Engage Panel select the Detector tab:

    Press the button showed above in the green circle. The red dot will move to the centre:Deflection=0 and Lateral=0 (as seen in image).

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    11/19

     

    3.3.1. Instrument Manual v 1.0 Page 11 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    Because we are using contact mode it’s important to place the red dot at -2V for thedeflection (vertical axis). Use the arrows next to the Deflection axis to bring thephotodiode to -2 position.

    Note: After all the alignment and focusing is done, press Move To Safety Height at theEngage Panel so the tip is at a safe distance of the surface and it doesn’t crash the

    sample.

    8. Set initial Scan Parameters:

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    12/19

     

    3.3.1. Instrument Manual v 1.0 Page 12 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    9. Engage and start taking images:

     At the engage panel (back to Approach tab), click on “Start Tip Approach”:

     After the tip has approached, press engage:

    Then in the Master Panel click “Do Scan” to start.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    13/19

     

    3.3.1. Instrument Manual v 1.0 Page 13 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    10. Adjust Scan Parameters:

    If the feedback parameters are adjusted optimally, the trace and retrace scan linesshould look very similar, but may not look completely identical.(Trace and Retrace scan lines are just below the image)

    If the graphs don’t look very similar change the values of the Velocity Tip, Integral Gainand Setpoint in order to optimise them.

    Remember that in Contact mode, increasing the Setpoint increases the force on thesample. Minimizing the imaging force is recommended for most applications.

    11. Save images:

    By default when opening the Asylum Research program, all the pictures will be savedautomatically.

    You can see that in the Master Panel:

    Give a name to the image in Base Name. If more than one image of the same sample istaken, it is not necessary to change the name every time. The Base suffix will keepchanging (starting from 0).

    If you want to change the folder where the data and images are stored press Path:

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    14/19

     

    3.3.1. Instrument Manual v 1.0 Page 14 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

     And follow the instructions:

    12. Stop the scan and turn off the microscope:

    If you want to stop scanning the sample press the button Stop!!! on the Master Panel :

     At the Engage Panel press (Un)Load Sample:

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    15/19

     

    3.3.1. Instrument Manual v 1.0 Page 15 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    The optical head will go up, so it is safe to remove the sample and the cantilever.

    •  Turn off the laser

    •  Turn the controller off

    •  Open the microscope door

    •  Slide the stage out•  Remove the cantilever holder and the sample

    •  Slide the stage back in

    •  Close the microscope door

    •  Close Asylum Research program

    •  Turn off the computer.

    .

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    16/19

     

    3.3.1. Instrument Manual v 1.0 Page 16 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    T PPING MODE

    Follow the same steps as for contact mode.

    In Step 7: Adjust Photodiode Signal, there is a small change. For tapping mode theDeflection and Lateral have to be 0. Therefore once you click on the button:

    It will automatically align the Deflection and Lateral to (0,0) so you don't need to doanything else.

    Continue with Step 8: Set Initial Scan Parameters

    Note that you need to change the imaging mode to AC Mode (Tapping).

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    17/19

     

    3.3.1. Instrument Manual v 1.0 Page 17 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    Before engaging (step 9) for tapping mode it is necessary to tune the cantilever 

    because each tip has a different frequency so this Auto Tune tool adjusts it in order to

    carry out a good experiment.

    At the master panel click on the Tune thumb. Then in the Tune Panel click on AutoTune. The default Target Amplitude is 1V but depending on the sample this will need to

    be changed (check at the end of this manual for suggestions).

    Once you have finished tuning the cantilever, you are ready to continue with Step 9 toStep 12, which are exactly the same as the Contact Mode steps.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    18/19

     

    3.3.1. Instrument Manual v 1.0 Page 18 of 19

    Last revised: May 2015 Responsible Officer: Marta Redrado Notivoli Next Review: May 2017

    APPENDIX: TIPS AND SUGGESTIONS

    Turn on the isolation table when you have finished setting up the sample and thecantilever holder.

    When you turn it on, it will first do a height adjustment.

    Once adjustment is finished the screen of the table will show “Isolation disabled”, thenpress button “E”.

    It will enable the isolation, the screen will now show “Isolation enabled” and the red lightwill be on.

  • 8/19/2019 3.3.1.b. Cypher instrument manual (1).pdf

    19/19

     

    3.3.1. Instrument Manual v 1.0 Page 19 of 19

    Contact Mode:

    In contact mode when you increase the Set Point, you are increasing the force of the tipagainst the surface and when you decrease the Set Point, you are decreasing the forceof the tip against the surface.

    Tapping Mode:

    When you increase the Set Point, you are decreasing the force of the tip against thesurface and when you decrease the Set Point, you are increasing the force of the tipagainst the surface. Note it is exactly the opposite in contact mode.

    Usually when doing Auto Tune the target amplitude by default is 1V and in the MasterPanel the Set Point is 800mv (80% of the target Amplitude).

    Depending on the sample, we should set different amplitudes. If the surface is very flat,it is recommended to set the Target Amplitude to 0.5V and the Set Point to 400mV(80% of the Target Amplitude).If the surface is very rough it is recommended to set the Target Amplitude to 2v and theSet Point to 1600mV (80% of the Target Amplitude).

    The Drive Amplitude is usually not changed. If you decide to change it, be careful andchange it slowly while imaging.

    Image Analysis

    For more information check the manual: Basics for image analysis using Igor Pro.