• Variable field + 3T • High uniformity over the whole excitation range • Large 25 mm pole gap • Compact Low fringe magnetic field • Cryogen Free 3 T METROLOGY MAGNET APPLICATIONS • Magnetic field probe calibration. • Hall-effect studies NMR/NMRD • ESR • FMR • Transport studies PERFORMANCE • 3 T variable field Better than 50 ppm field uniformity (10 mm DSV) across the excitation range • Suitable for NMR tesla meter • Large fixed pole gap (25 mm) Large sample access slot: 25 x 150 mm Optional high stability unipolar power supply Optional electric shims [Z, Y, Z, Z2 ] to further increase uniformity (<1 ppm over 10 mm DSV) EASY TO USE Cryogen-free operation: no handling of liquid helium or nitrogen Magnet safety monitoring electronics included Reliable mechanical cooling (<4 kW input power) Low water cooling requirements, optional chiller EASY TO SITE • Compact size Very low fringe field, may be reduced further with optional enclosure Compressor can be sited remotely (up to 10 mm) Vertical or horizontal access slot orientation

Transcript of 3 T METROLOGY MAGNET · 2017-03-15 · 3 T METROLOGY MAGNET. APPLICATIONS •Magnetic field probe...

Page 1: 3 T METROLOGY MAGNET · 2017-03-15 · 3 T METROLOGY MAGNET. APPLICATIONS •Magnetic field probe calibration. •Hall-effect studies •NMR/NMRD •ESR •FMR •Transport studies.

• Variablefield+3T

• Highuniformityoverthewholeexcitationrange

• Large25mmpolegap

• Compact

Lowfringemagneticfield

• CryogenFree

3 T METROLOGY MAGNET

APPLICATIONS

• Magneticfieldprobecalibration.

• Hall-effectstudies

• NMR/NMRD

• ESR

• FMR

• Transportstudies

PERFORMANCE

• 3Tvariablefield

• Betterthan50ppmfielduniformity(10mmDSV)acrosstheexcitationrange

• SuitableforNMRteslameter

• Largefixedpolegap(25mm)

• Largesampleaccessslot:25x150mm

• Optionalhighstabilityunipolarpowersupply

• Optionalelectricshims[Z,Y,Z,Z2]tofurtherincreaseuniformity(<1ppmover10mmDSV)

EASY TO USE

• Cryogen-freeoperation:nohandlingofliquidheliumornitrogen

• Magnetsafetymonitoringelectronicsincluded

• Reliablemechanicalcooling(<4kWinputpower)

• Lowwatercoolingrequirements,optionalchiller

EASY TO SITE

• Compactsize

• Verylowfringefield,maybereducedfurtherwithoptionalenclosure

• Compressorcanbesited

r emotely(up to 10 mm)

• Verticalorhorizontalaccessslotorientation

Page 2: 3 T METROLOGY MAGNET · 2017-03-15 · 3 T METROLOGY MAGNET. APPLICATIONS •Magnetic field probe calibration. •Hall-effect studies •NMR/NMRD •ESR •FMR •Transport studies.

3 T METROLOGY MAGNET

hts-110.com

HTS-1101B Quadrant DriveWaiwhetuLower HuttNew Zealand

t +64 4 570 8880 e [email protected]

LinkedIn:nz.linkedin.com/company/hts-110

DistributedinNorthAmericabyGMWAssociates955IndustrialRoadSanCarlos,CA94070,USA

t +6508028292wgmw.com

SYSTEM SPECIFICATIONS:

• 3Tfullfieldat180A.

• Betterthan50ppmuniformity(10mmDSVacrosstheexcitationrange(<1ppmwithRTshims).

• Systemcooldown(roomtemperaturetooperatingtemperature):<30h(60Hz).

• Sampleaccess:25x150mm.

• Fringefield:5gausslineat<1.6m(axial)and1.3m(radial)frommagnetcentre.

• Magnetmass:200kg.

• Systemmasswithbipolarpowersupply,cryocoolerandcompressor:450kg.

STANDARD SYSTEM INCLUDES:

• Magnetsub-systemwithintegratedcold-head.

• 19”rack-mountedcompressorandpowersupply.

• Activemagnetprotectionelectronicslinkedtointegratedtemperaturesensorsandvoltagetaps.

• Magnetenergydumpsystem.

• Bipolarpowersupplyoroptionalhighstability.

(0.3ppm/h)unipolarsupply.

• 1yearwarranty.

SYSTEM OPTIONS:

• Fieldcontrolsystem.

• Electricalshimmingforimprovedfieldhomogeneity(X,Y,Z,Z2).

• Lowmaintenancevacuumpump.

• Magnetenclosureforreducedfringefield.

• Waterchiller.

• Extendedwarranty.

• RTshimset.

SITE REQUIREMENTS:

• <6litreswaterperminuteforcompressor

• 50/60Hz,~4kW.

• Scheduledmaintenanceoncold-headevery13,000h,compressor30,000h.

• Vacuumpumpforinitialinstallationandmaintenance;turbomolecularpumprecommendedwithminimumpumpingspeed30l/sandultimatepressure<1E–7mbar.

• Variable Field Relaxometry (NMRD NuclearMagnetic Relaxation Dispersion)

• Developed in association with Stelar (Italy)

• Used for example in characterisation of newMRI contrast agents optimised for newhigher-field MRI systems

APPLICATION EXAMPLE: