The 2470 High Voltage SourceMeter ® Source Measure Unit (SMU) Instrument brings advanced Touch, Test, Invent ® technology right to your fingertips. It combines an innovative graphical user interface (GUI) with capacitive touchscreen technology to make testing intuitive and minimize the learning curve to help engineers and scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized for characterizing and testing high voltage, low leakage devices, materials, and modules, such as silicon carbide (SiC), gallium nitride (GaN), power MOSFETs, transient suppression devices, circuit protection devices, power modules, batteries, and much more. These new capabilities, combined with Keithley’s decades of expertise in developing high precision, high-accuracy SMU instruments, make the 2470 a “go-to instrument” for high-voltage source and low-current measurement applications in the lab and in the test rack. Key Features Wide coverage up to 1100 V / 1 A DC 20 W max. • 10 fA measure resolution 0.012% basic measure accuracy with 6½-digit resolution Five-inch, high resolution capacitive touch screen GUI • Source and sink (4-quadrant) operation • SCPI and TSP ® scripting programming modes • TSP-Link for multi-channel I-V testing Front panel input banana jacks; rear panel high-voltage input triaxial connections • Built-in context-sensitive help Front-panel USB 2.0 memory I/O port for transferring data, test scripts, and test configurations 2470 Graphical SourceMeter ® SMU Instrument Datasheet A Tektronix Company

Transcript of 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent...

Page 1: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

The 2470 High Voltage SourceMeter® Source Measure Unit (SMU) Instrument brings advanced Touch, Test, Invent® technology right to your fingertips. It combines an innovative graphical user interface (GUI) with capacitive touchscreen technology to make testing intuitive and minimize the learning curve to help engineers and scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized for characterizing and testing high voltage, low leakage devices, materials, and modules, such as silicon carbide (SiC), gallium nitride (GaN), power MOSFETs, transient suppression devices, circuit protection devices, power modules, batteries, and much more. These new capabilities, combined with Keithley’s decades of expertise in developing high precision, high-accuracy SMU instruments, make the 2470 a “go-to instrument” for high-voltage source and low-current measurement applications in the lab and in the test rack.

Key Features• Wide coverage up to 1100 V / 1 A DC 20 W max.

• 10 fA measure resolution

• 0.012% basic measure accuracy with 6½-digit resolution

• Five-inch, high resolution capacitive touch screen GUI

• Source and sink (4-quadrant) operation

• SCPI and TSP® scripting programming modes

• TSP-Link for multi-channel I-V testing

• Front panel input banana jacks; rear panel high-voltage input triaxial connections

• Built-in context-sensitive help

• Front-panel USB 2.0 memory I/O port for transferring data, test scripts, and test configurations

2470 Graphical SourceMeter® SMU InstrumentDatasheet

A Tektronix Company

Page 2: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM2

Learn Faster; Work Smarter; Invent EasierThe 2470 features a five-inch, full-color, high resolution touchscreen that supports intuitive operation, helps operators become familiar with the instrument quickly, and optimizes overall speed and productivity. A simple icon-based menu structure reduces the number of steps required to configure a test by as much as 50 percent and eliminates the cumbersome multi-layer menu structures typically used on soft-key instruments. Built-in, context-sensitive help supports intuitive operation and minimizes the need to review a separate manual. These capabilities, combined with the 2470’s high versatility, simplify its operation in both basic and advanced measurement applications, regardless of the user’s previous experience in working with SMU instruments.

All-in-One SMU InstrumentThe 2470, Keithley’s fourth generation of SourceMeter SMUs, leverages the proven capabilities of the original 2410 High Voltage SourceMeter SMU Instrument. SMU instruments offer a highly flexible, four-quadrant voltage and current source/load coupled with precision voltage and current measurements. This all-in-one instrument has the capabilities of a:

• Precision power supply with V and I readback

• True current source

• Digital multimeter (DCV, DCI, ohms, and power with 6½-digit resolution)

• Precision electronic load

• Trigger controller

2470 main home screen.

2470 icon-based menu.

2470 power envelope.

Quadrant I

Quadrant IVQuadrant III

Quadrant II+100 mA

+1 A

-100 mA

-1 A

+10 mA

-10 mA

-750 mA

+750 mA

+20 V-20 V-200 V +200 V +1000 V–1000 V -8 V +8 V

Duty cycle limited

Page 3: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 3

2470 Graphical SourceMeter® SMU Instrument Data Sheet

OnlineHELP key

USB 2.0memory I/O

Front/rearinput selector

Rotarynavigation/control knob

5˝ color graphical touchscreen display

Ease of Use Beyond the TouchscreenIn addition to its five-inch color touchscreen, the 2470 front panel offers a variety of features that enhance its speed, user-friendliness, and learnability, including a USB 2.0 memory I/O port, a HELP key, a rotary navigation/control knob, a front/rear input terminal selector button, and banana jacks for basic bench applications. The USB 2.0 memory port simplifies storing test results and instrument configurations, uploading test scripts into the instrument, and installing system upgrades. All front-panel buttons are backlit to enhance visibility in low-light environments.

Four Quick Setup modes simplify instrument setup. With one touch, the instrument can be quickly configured for various operating modes without the need to configure the instrument indirectly for this operation.

Comprehensive Built-in ConnectivityRear panel access to rear-input connectors, remote control interfaces (GPIB, USB 2.0, and LXI/Ethernet), a D-sub 9-pin digital I/O port (for internal/external trigger signals and handler control), instrument interlock connector, and TSP-Link® connectors make it simple to configure multiple instrument test solutions and eliminate the need to invest in additional adapter accessories.

2470 front panel with high-resolution, capacitive touchscreen.

Rear panel connections are optimized for signal integrity and speed system setup.

One-touch Quickset modes speed measurement setups and minimize the time to measurements.

Ethernet

Digital I/O TSP-Link GPIB

Interlock

USB

Page 4: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM4

Convert Raw Data to InformationA full graphical plotting window converts raw data and displays it immediately as useful information, such as semiconductor I-V curves and waveforms. Using the 2470’s Sheet view, test data can also be displayed in tabular form. The instrument supports exporting data to a spreadsheet for further analysis, dramatically improving productivity for research, benchtop testing, device qualification, and debugging.

Built-in data display, charting, and spreadsheet export functions simplify converting test results into useful information.

TriggerFlow® Building Blocks for Instrument Control and ExecutionThe 2470 incorporates Keithley’s TriggerFlow triggering system, which provides user control of instrument execution. TriggerFlow diagrams are created in much the same way that flow charts are developed, using four building blocks:

• Wait – Waits for an event to occur before the flow continues

• Branch – Branches when a condition has been satisfied

• Action – Initiates an action in the instrument, for example, measure, source, delay, set digital I/O, etc.

• Notify – Notifies other equipment that an event has occurred

TriggerFlow building blocks allow creating triggering models that range from very simple to highly complex.

A TriggerFlow model using a combination of these building blocks can be created from the front panel or by sending remote commands. With the TriggerFlow system, users can build triggering models from very simple to complex with up to 255 block levels. The 2470 also includes basic triggering functions, including immediate, timer, and manual triggering.

Page 5: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 5

2470 Graphical SourceMeter® SMU Instrument Data Sheet

Unmatched System Integration and Programming FlexibilityWhen a 2470 is configured into a multi-channel I-V test system, its embedded Test Script Processor (TSP®) allows it to run test scripts, so users can create powerful measurement applications with significantly-reduced development times. TSP technology also offers channel expansion without a mainframe. Keithley’s TSP-Link® channel expansion bus can connect multiple 2470s and other TSP instruments, including Keithley’s other graphical SourceMeter instruments (2450, 2460, 2461), Keithley’s DMM7510, DMM6500, and DAQ6510 graphical multimeters, Series 2600B System SourceMeter SMU instruments, and Series 3700A Switch/Multimeter systems, in a master-subordinate configuration that operates as one integrated system. The TSP-Link expansion bus supports up to 32 units per GPIB or IP address, making it easy to scale a system to fit an application’s particular requirements.

Parallel Test CapabilityThe TSP technology in the 2470 supports testing multiple devices in parallel to meet the needs of device research, advanced semiconductor lab applications, and even high throughput production test. This parallel testing capability allows each instrument in the system to run its own complete test sequence, creating a fully multi-threaded test environment. The number of tests that can be run in parallel on a graphical SourceMeter can be as high as the number of instruments in the system.

Typical ApplicationsIdeal for current/voltage characterization and functional test of a wide range of today’s modern electronics and devices, including:

• Devices

– Power semiconductors and materials: SiC, GaN, power MOSFETs, power diodes

– IGBT

– Thyristors, SCRs

– Transient suppression devices

– Circuit protection devices: TVS, MOVs, fuses, etc.

– Connectors, switches, relays

– Power management modules

– Batteries

• Tests

– I-V characterization

– Forward voltage

– Breakdown voltage

– Leakage current

– Isolation

– Hi-Pot

– Dielectric withstanding

The 2470 is the ideal addition to your development bench for designing today’s high power devices and components.

Characterize transistors with multiple SMU instruments on wafer or packaged transistors

Page 6: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM6

Instrument Control Start-up SoftwareKickStart instrument control/start-up software enables users to start making measurements in minutes without programming. In most cases, users merely need to make some quick measurements, graph the data, and store the data to disk for later analysis in software environments such as Excel. KickStart offers:

• Instrument configuration control to perform I-V characterization

• Native X-Y graphing, panning, and zooming

• Spreadsheet/tabular viewing of data

• Saving and exporting of data for further analysis

• Saving of test setups

• Screenshot capturing of graphs

• Annotating of tests

• Command line dialog for sending and receiving data

• HTML help

• GPIB, USB 2.0, Ethernet compliance

Simplified Programming with Ready-to-Use Instrument DriversFor those who prefer to create their own customized application software, native National Instruments LabVIEW® drivers, and IVI-C and IVICOM drivers are available at tek.com/keithley.

KickStart start-up software lets users be ready to make measurements in minutes.

Page 7: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 7

2470 Graphical SourceMeter® SMU Instrument Data Sheet

Specifications

Voltage Specifications 1, 2

Source Measure 3

Range 4 Resolution

Accuracy 5 23 °C ± 5 °C, 1 Year ±(% setting + volts)

Noise (RMS) <10 Hz Resolution

Input Resistance

Accuracy 23 °C ± 5 °C, 1 Year ±(% reading + volts)

200.0000 mV 5 µV 0.015% + 200 µV 2 µV 100 nV > 10 GΩ 0.012% + 200 µV

2.000000 V 50 µV 0.020% + 300 µV 10 µV 1 µV > 10 GΩ 0.012% + 300 µV

20.00000 V 500 µV 0.015% + 2.4 mV 100 µV 10 µV > 10 GΩ 0.015% + 1 mV

200.0000 V 5 mV 0.015% + 24 mV 1 mV 100 µV > 10 GΩ 0.015% + 10 mV

1000.000 V 50 mV 0.02% + 100 mV 20 mV 10 mV > 10 GΩ 0.015% + 50 mV

Temperature Coefficient ± (0.15 × accuracy specification)/°C, 0 °C to 18 °C and 28 °C to 50 °C

Current Specifications 1, 2

Source Measure 3

Range 4 Resolution

Accuracy 5 23 °C ± 5 °C, 1 Year ±(% setting + amps)

Noise (RMS) <10 Hz Resolution

Voltage Burden

Accuracy 23 °C ± 5 °C, 1 Year

±(% reading + amps)

10.00000 nA 6 500 fA 0.100% + 200 pA 500 fA 10 fA < 100 µV 0.10% + 250 pA

100.0000 nA 6 5 pA 0.060% + 250 pA 500 fA 100 fA < 100 µV 0.060% + 300 pA

1.000000 μA 50 pA 0.025% + 400 pA 5 pA 1 pA < 100 µV 0.025% + 300 pA

10.00000 μA 500 pA 0.025% + 1.5 nA 40 pA 10 pA < 100 µV 0.025% + 700 pA

100.0000 μA 5 nA 0.020% + 15 nA 400 pA 100 pA < 100 µV 0.02% + 6 nA

1.000000 mA 50 nA 0.020% + 150 nA 5 nA 1 nA < 100 µV 0.02% + 60 nA

10.00000 mA 500 nA 0.020% + 1.5 μA 40 nA 10 nA < 100 µV 0.02% + 600 nA

100.0000 mA 5 μA 0.025% + 15 μA 100 nA 100 nA < 100 µV 0.025% + 6 μA

1.000000 A 50 μA 0.067% + 900 μA 10 μA 1 μA < 100 µV 0.03% + 500 μA

Temperature Coefficient ± (0.15 × accuracy specification)/°C, 0 °C to 18 °C and 28 °C to 50 °C

Notes1. Speed = 1 PLC.2. All specifications are guaranteed with output ON.3. Accuracies apply to 2-wire and 4-wire modes when properly zeroed. For the 200 mV and 1 A ranges, the voltage burden may exceed the specification in 2-wire mode.4. Maximum display and programmming ranges are 5% overrange for voltage, except for the 1000 V range, which is 10% overrange (1100 V), and 5% overrange for current (for example,

1.05 A on the 1 A range).5. For sink mode, accuracy is ±(0.15% + offset × 4) except for 1 A range, accuracy is: ± (1.5% + offset × 8).6. Rear-panel triaxial connections only.

Resistance Measurement Accuracy (Local Or Remote Sense) 7, 8, 9

Range Default Resolution 10 Default Test Current

Normal Accuracy 23 °C ± 5 °C, 1 Year

± (% reading + ohms)

Enhanced Accuracy 11 23 °C ± 5 °C, 1 Year

± (% reading + ohms)

< 2.000000 Ω 12 1 µΩ User-defined Source IACC + Meas VACC Meas IACC + Meas VACC

20.00000 Ω 10 µΩ 100 mA 0.098% + 0.003 Ω 0.073% + 0.001 Ω

200.0000 Ω 100 µΩ 10 mA 0.077% + 0.03 Ω 0.053% + 0.01 Ω

2.000000 kΩ 1 mΩ 1 mA 0.066% + 0.3 Ω 0.045% + 0.1 Ω

20.00000 kΩ 10 mΩ 100 µA 0.063% + 3 Ω 0.043% + 1 Ω

200.0000 kΩ 100 mΩ 10 µA 0.065% + 30 Ω 0.046% + 10 Ω

2.000000 MΩ 1 Ω 1 µA 0.11% + 300 Ω 0.049% + 100 Ω

20.00000 MΩ 10 Ω 1 µA 0.11% + 1 kΩ 0.052% + 500 Ω

200.0000 MΩ 13 100 Ω 100 nA 0.655% + 10 kΩ 0.349% + 5 kΩ

> 200.0000 MΩ 12, 13 — User-defined Source IACC + Meas VACC Meas IACC + Meas VACC

Page 8: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM8

Temperature Coefficient ± (0.15 × accuracy specification)/°C 0 °C to 18 °C and 28 °C to 50 °C

Source Current, Measure Resistance Mode Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense)

Source Voltage, Measure Resistance Mode Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense)

Guard Output Impedance ≥ 300 Ω typical

Notes7. Speed = 1 PLC.8. All specifications are guaranteed with output ON.9. Accuracies apply to 2-wire and 4-wire modes when properly zeroed.10. Measure resolution 6.5 digits.11. Source readback enabled; offset compensation on.12. Source current, measure resistance or source voltage, measure resistance only.13. Rear-panel triaxial connections only.

Supplemental SpecificationsOverrange 105% of range for 200 mV to 200 V, source and measure ranges; 110% of range for 1000 V, source and

measure ranges

Regulation

Voltage Line: 0.01% of range Load: 0.01% of range + 100 µV

Current Line: 0.01% of range Load: 0.01% of range + 100 pA

Source Limits

Voltage Source Current Limit Bipolar current limit set with a single value Minimum value is 10% of range

Current Source Voltage Limit Bipolar voltage limit set with a single value Minimum value is 10% of range

Voltage Limit/Current Limit Accuracy Add 0.3% of range and ±0.02% of reading to base specification

Overshoot

Voltage Source < 0.1% typical Step size = Full scale, resistive load, 20 V range, 10 mA current limit

Current Source < 0.1% typical Step size = Full scale, resistive load of 10 kΩ, 1 mA range, 20 V voltage limit

Range Change Overshoot Overshoot into a fully resistive 100 kΩ load, 10 Hz to 20 MHz bandwidth, adjacent ranges: 250 mV typical

Output Settling Time Time required to reach within 0.1% of final value: 20 V range, 100 mA current limit: < 200 µs typical

Maximum Slew Rate 14 0.2 V/µs, 200 V range, 100 mA limit into a 2 kΩ load (typical) 0.5 V/µs, 1000 V range, 10 mA limit into a 100 kΩ load (typical)

Overvoltage Protection User-selectable values, 10% tolerance; factory default = none

Voltage Source Noise 10 Hz to 20 MHz (RMS): 4 mV typical into a resistive load

Common Mode Voltage 250 V DC

Common Mode Isolation > 1 GΩ, < 1000 pF

Notes14. High capacitance mode off.

Page 9: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 9

2470 Graphical SourceMeter® SMU Instrument Data Sheet

Noise Rejection (typical)

NPLC NMRR CMRR

0.01 — 60 dB

0.1 — 60 dB

1 60 dB 100 dB*

* Except 10 nA and 100 nA current ranges ~90 dB

Load Impedance

Normal Mode 20 nF typical

High-Capacitance Mode Stable into 1 μF typical (specification only valid for ranges ≥100 μA)

Maximum Voltage Drop Between Force and Sense Terminals 5 V

Maximum Sense Lead Resistance 1 MΩ for rated accuracy

Sense Input Impedance > 10 GΩ

Guard Offset Voltage < 300 µV typical

System Measurement Speeds 15

Reading rates (readings per second) typical for 60 Hz (50 Hz), script (TSP) programmed

NPLCTrigger Origin

Measure to Memory

Measure to GPIB

Measure to USB

Measure to LAN

Source Measure Sweep to Memory

Source measure sweep to

GPIB

Source Measure Sweep to

USB

Source Measure Sweep to

LAN

0.01 Internal 3150 (2800) 2760 (2570) 2825 (2570) 2740 (2530) 1710 (1620) 1620 (1540) 1630 (1540) 1620 (1540)

0.01 External 2170 (2050) 2120 (2003) 2170 (2010) 2100 (1990) 1670 (1590) 1580 (1500) 1590 (1510) 1580 (1510)

0.10 Internal 540 (460) 530 (450) 530 (450) 530 (450) 470 (410) 460 (400) 470 (400) 470 (400)

0.10 External 500 (430) 490 (420) 500 (425) 480 (420) 450 (400) 460 (390) 460 (390) 410 (350)

1.00 Internal 59 (49) 58 (49) 59 (49) 59 (49) 58 (48) 58 (48) 58 (48) 57 (48)

1.00 External 58 (48) 57 (48) 58 (48) 58 (48) 57 (48) 57 (48) 57 (48) 55 (48)

Reading rates (readings per second) typical for 60 Hz (50 Hz), SCPI programmed 16

NPLCTrigger Origin

Measure to Memory

Measure to GPIB

Measure to USB

Measure to LAN

Source Measure Sweep to memory

Source Measure Sweep to

GPIB

Source Measure Sweep to

USB

Source Measure Sweep to

LAN

0.01 Internal 3040 (2800) 3000 (2760) 3000 (2760) 3010 (2710) 1710 (1630) 1610 (1544) 1440 (1380) 1690 (1590)

0.01 External 2320 (2165) 2290 (2140) 2340 (2150) 2290 (2130) 1680 (1590) 1560 (1525) 1410 (1360) 1660 (1560)

0.10 Internal 540 (460) 540 (450) 540 (460) 540 (450) 470 (410) 470 (410) 450 (390) 470 (410)

0.10 External 510 (440) 510 (430) 510 (440) 510 (430) 470 (400) 470 (400) 450 (390) 470 (400)

1.00 Internal 59 (49) 59 (49) 59 (49) 59 (49) 58 (48) 58 (48) 57 (48) 58 (48)

1.00 External 58 (49) 58 (49) 58 (49) 58 (49) 58 (48) 58 (48) 57 (47) 58 (48)

Notes15. Reading rates applicable for voltage or current measurements, autozero off, autorange off, filter off, binary reading format, and source readback off.16. SCPI programming mode.

Page 10: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM10

General Characteristics(Default mode unless specified)

Factory Default Standard Power-Up Setting SCPI mode

Source Output Modes Fixed DC level Memory/configuration list (mixed function) Stair (linear and logarithmic)

Memory Buffer >5,000,000 readings with selected measured values and timestamp

Real-Time Clock Lithium battery backup (more than 3 years of battery life)

Remote Interfaces GPIB: IEEE Std 488.1 compliant; supports IEEE Std 488.2 common commands and status model topology

USB device (rear panel, type B): 2.0 full-speed USBTMC

USB host (front panel, type A): USB 2.0, support for flash drives, FAT32

Ethernet: RJ-45 connector, 10/100 BT

IP Configuration Static or DHCP

Expansion Interface The TSP-Link® expansion interface allows TSP-enabled instruments to trigger and communicate with each other

LXI Compliance LXI version 1.4 Core 2011

TSP Mode Embedded Test Script Processor (TSP) accessible from any host interface

Display Five-inch capacitive touch, color TFT WVGA (800 × 480) with LED backlight

Input Signal Connections Front: Banana. Rear: High-voltage triaxial

Programmability SCPI or TSP command sets

Interlock Active high-input

Digital I/O Lines: Six input/output, user-defined, for digital I/O or triggering

Connector: 9-pin female D

Input Signal Levels: 0.7 V (maximum logic low), 3.7 V (minimum logic high)

Input Voltage Limits: –0.25 V (absolute minimum), +5.25 V (absolute maximum)

Maximum Source Current: +2.0 mA at > 2.7 V (per pin)

Maximum Sink Current: –50 mA at 0.7 V (per pin, solid-state fuse protected)

5 V Power Supply Pin: Limited to 500 mA at > 4 V (solid-state fuse protected)

Handler: User-definable start of test, end of test, four category bits

Cooling Forced air, variable speed

Overtemperature Protection Internally sensed temperature overload puts instrument in standby mode

Power Supply 100 VRMS to 240 VRMS, 50 Hz or 60 Hz (automatically detected at power up)

VA Rating 220 VA maximum

Altitude Maximum 2000 meters (6562 feet) above sea level

EMC Conforms to European Union EMC Directive

Safety NRTL listed to UL61010-1 and UL61010-2-30; conforms to European Union Low Voltage Directive

RoHS Conforms to European Union Restriction on Hazardous Substances Directive

Vibration MIL-PRF-28800F Class 3 Random

Warm up One hour to rated accuracies

Page 11: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 11

2470 Graphical SourceMeter® SMU Instrument Data Sheet

Dimensions With handle and bumpers: 106 mm × 255 mm × 425 mm (4.18 in. high × 10.05 in. wide × 16.75 in. deep)

Without handle and bumpers: 88 mm × 213 mm × 403 mm (3.46 in. high × 8.39 in. wide × 15.87 in. deep)

Weight With handle and bumpers: 4.54 kg (10 lb)

Without handle and bumpers: 4.08 kg (9.0 lb)

Environment Operating: 0 °C to 50 °C, 70% relative humidity up to 35 °C; derate 3% relative humidity/°C, 35 °C to 50 °C

Storage: –25 °C to 65 °C

Pollution Category: 2

SourceMeter® is a registered trademark of Keithley Instruments LLC. All other trademarks are the property of their respective owners.

Supplied Accessories8608 High Performance Test Leads

USB-B-1 USB Cable, Type A to Type B, 1 m (3.3 ft)

CS-1616-3 Safety Interlock Mating Connector

17469460X TSP-Link/Ethernet Cable

2470 QuickStart Guide

Test Script Builder Software (available at www.tektronix.com)

LabVIEW and IVI Drivers (available at www.tektronix.com)

Available Accessories

Test Leads and Probes8605 High Performance Modular Test Leads

8606 High Performance Modular Probe Kit

8608 High Performance Test Leads

Cables, Connectors, AdaptersTRX-1100V-BAN HV Triax to Banana Adapter. Converts the 4 triax connectors on the rear panel to 5 banana jacks

TRX-1100V-* 3-slot HV Low Noise Triax Cable (0.5 m, 1 m, 2 m, 3 m lengths available)

8607 2-wire, 1000 V Banana Cables, 1 m (3.3 ft)

CS-1616-3 Safety Interlock Mating Connector

Communication Interfaces & Cables7007-1 Shielded GPIB Cable, 1 m (3.3 ft)

7007-2 Shielded GPIB Cable, 2 m (6.6 ft)

17469460X TSP-Link Ethernet Cable

KPCI-488LPA IEEE-488 Interface for PCI Bus

KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter

USB-B-1 USB Cable, Type A to Type B, 1 m (3.3 ft)

Page 12: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Datasheet

TEK.COM12

Triggering and Control 2450-TLINK DB-9 to Trigger Link Connector Adapter

8501-1 Trigger Link Cable, DIN-to-DIN, 1 m (3.3 ft.)

8501-2 Trigger Link Cable, DIN-to-DIN, 2 m (6.6 ft.)

Rack Mount Kits 4299-8 Single Fixed Rack Mount Kit

4299-9 Dual Fixed Rack Mount Kit

4299-10 Dual Fixed Rack Mount Kit. Mount one 2450 and one Series 26xxB

4299-11 Dual Fixed Rack Mount Kit. Mount one 2450 and one Series 2400, Series 2000, etc.

Available Services2470-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

2470-5Y-EW 1-year factory warranty extended to 5 years from date of shipment

C/2470-3Y-17025 KeithleyCare® 3 Year ISO 17025 Calibration Plan

C/2470-3Y-DATA KeithleyCare 3 Year Calibration w/Data Plan

C/2470-3Y-STD KeithleyCare 3 Year STD Calibration Plan

C/2470-5Y-17025 KeithleyCare 5 Year ISO 17025 Calibration Plan

C/2470-5Y-DATA KeithleyCare 5 Year Calibration w/Data Plan

C/2470-5Y-STD KeithleyCare 5 Year STD Calibration Plan

C/NEW DATA Calibration Data for New Units

C/NEW DATA ISO ISO-17025 Calibration Data for New Units

Ordering Information2470 1000 V, 1 A, 20 W SourceMeter SMU Instrument

Warranty InformationWarranty Summary This section summarizes the warranties of the 2470. For complete warranty information, refer to the

Tektronix warranty page at https://www.tek.com/service/warranties/warranty-2. Any portion of the product that is not manufactured by Keithley is not covered by this warranty and Keithley will have no duty to enforce any other manufacturer’s warranties.

Hardware Warranty Keithley warrants the Keithley manufactured portion of the hardware for a period of one year from defects in materials or workmanship; provided that such defect has not been caused by use of the Keithley hardware which is not in accordance with the hardware instructions. The warranty does not apply upon any modification of Keithley hardware made by the customer or operation of the hardware outside the environmental specifications.

Software Warranty Keithley warrants for the Keithley produced portion of the software or firmware will conform in all material respects with the published specifications for a period of ninety (90) days; provided the software is used on the product for which it is intended in accordance with the software instructions. Keithley does not warrant that operation of the software will be uninterrupted or error-free, or that the software will be adequate for the customer’s intended application. The warranty does not apply upon any modification of the software made by the customer.

Page 13: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

TEK.COM 13

2470 Graphical SourceMeter® SMU Instrument Data Sheet

Page 14: 2470 Graphical SourceMeter SMU Instrument · scientists learn faster, work smarter, and invent easier. With its 1100 V and 10 fA capability, the 2470 is optimized ... and optimizes

Contact Information: Australia* 1 800 709 465

Austria 00800 2255 4835

Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777

Belgium* 00800 2255 4835

Brazil +55 (11) 3759 7627

Canada 1 800 833 9200

Central East Europe / Baltics +41 52 675 3777

Central Europe / Greece +41 52 675 3777

Denmark +45 80 88 1401

Finland +41 52 675 3777

France* 00800 2255 4835

Germany* 00800 2255 4835

Hong Kong 400 820 5835

India 000 800 650 1835

Indonesia 007 803 601 5249

Italy 00800 2255 4835

Japan 81 (3) 6714 3010

Luxembourg +41 52 675 3777

Malaysia 1 800 22 55835

Mexico, Central/South America and Caribbean 52 (55) 56 04 50 90

Middle East, Asia, and North Africa +41 52 675 3777

The Netherlands* 00800 2255 4835

New Zealand 0800 800 238

Norway 800 16098

People’s Republic of China 400 820 5835

Philippines 1 800 1601 0077

Poland +41 52 675 3777

Portugal 80 08 12370

Republic of Korea +82 2 565 1455

Russia / CIS +7 (495) 6647564

Singapore 800 6011 473

South Africa +41 52 675 3777

Spain* 00800 2255 4835

Sweden* 00800 2255 4835

Switzerland* 00800 2255 4835

Taiwan 886 (2) 2656 6688

Thailand 1 800 011 931

United Kingdom / Ireland* 00800 2255 4835

USA 1 800 833 9200

Vietnam 12060128

* European toll-free number. If not accessible, call: +41 52 675 3777

Find more valuable resources at TEK.COM

Copyright © Tektronix. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies. 060719.SBG 1KW-61523-0