Page 1:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

www.raytest.com

Page 2:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

ɣ-RADIATION Choose your application of

99mTc 0 – 200 keV

131I 200 – 700 keV

59Fe 700 – 2.600 keV

ɣ-HPLC ml / min

GABI

1x1" NaI(Tl)

pin hole

GABI

2x2" NaI(Tl)

pin hole

miniGITA

BGO

GITA BGO

ISOMED

MUCHA

2x2" NaI(Tl)

MUCHA

1x1" NaI(Tl)

GABI

3x3" NaI(Tl)

pin hole

MUCHA

3x3" NaI(Tl)Counting

Dose Calibrator

TLC 2x 50 x 200 mm

TLC 1x 50 x 200 mm

γ

Page 3:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com48 |

GABI Star is the latest technology µ- processor controlled counter with built in digital ratemeter.

GABI Star stand alone unit offers a keyboard and liquid crystal display for parameter dialog, entry and result presentation.

GABI Star has a high voltage unit for manual or preset selection of the required detector operation.

GABI Star has 2 simultaneous counting channels with each 2 individually selectable energy thresholds each for setting an individual counting window for a particular nuclide.

GABI Star is counting the radiation events in an energy chan-nel per selected interval time.

The information of counts per time can be transferred to a ra-temeter, which converts the digital format to an analog output format of 0 – 1V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard. The output si-gnal 0 – 1V of GABI Star can be easily adapted to the required chromatography conditions. GABI Star has an extremely high dynamic range from 0 – 600.000 c/s.

The information of counts per time is transferred to a PC run-ning GINA Star program. The chromatogram is displayed live on screen. No overriding of ranges can happen.

Peak integration, background subtraction, flow correction, absolute activity calculation etc can be performed manually or automatically. Limit-of-detection is determined for every ROI, displayed and printed. GINA Star program can control all modules of Agilent HPLC. All data and parameters are stored under one identity code. GINA Star can be installed as well in background of the original HPLC-operation program and the user can use the operation program of the HPLC supplier as well as the advantages of GINA Star.

125I Iodine radio-HPLC

99mTc Technetium

2 simultaneous counting channels

free threshold settings

extremely high dynamic range 0 – 600.000 c/s

LCD 4 lines, 20 characters each

keybord: 16 touch sensitive keys

output: 0 – 1 V

GABI Star radioactivity-HPLC-flow monitor, 1x1" NaI scintillation detector

GABI Star 1" pin-hole

energy calibration with 137Cs

Page 4:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

GABI Star

energy range: 30 – 1600 keV

high voltage range: 500 – 2000 V

count rate: 0 – 600.000 c/s

data input: 7 analog input channels, -0.5 to + 4.5V resolution 21 bit

data output: RS 232 C, 2 analog outputs 0 - 1 V resolution 20 bit

scintillation probe: 1"

Nal-scintillator: 1x1", pin-hole 8 mm diam. x 20 mm deep

housing: 39 mm ø stainless steel

window: 20 mg / cm2

shielding: lead 30 mm wall

flow cell: 100 µl coiled, thin plastic tube

01050100 GABI Star ɣ- HPLC-flow-detector

with keyboard and display analog output 0 – 1V

01052005 GABI Star scintillation detector 1 x 1" Nal scintillation low energy ɣ-probe window 200 mg / cm2

01052015 GABI Star 30 mm lead shielding for 1“ detector

option

PC + Windows program

01240012 GINA Star 4-channel program 2 radioactivity channel 2 other detectors input

01058007 ɣ- reference standard

01240074 installation

01240074 installation GABI Star stand alone

01240239 training GABI Star stand alone

01240218 installation GABI Star with GINA Star

01240220 training GABI Star with GINA Star

HPLC with 99mTc pin-hole scintillation detectorɣ-spectrum 99mTc

Integration N:\Helfer\GABI Star 1-Zoll Bohrloch (Mainz)\TC-99M Spektrum\140704_01

Start: 04.07.2014 08:44

100 200 300 400 500 600 700 keV

ChA

100 200 300 400 500 600 700 keV

0,0

2,0

4,0

6,0

8,0

10,0

12,0

14,0

16,0

18,0

20,0 CPS *1000

141 keV

Probenbeschreibung

Messung: 140704_01, Injektion : 04.07.2014 08:44

Methode: TC-99M Spektrum vom: 03.07.2014 13:11

Meßstation: 12

1" Bohrlochdetektor

Radioaktivitätsdetektor: raytest Gabi Star Serien Nr..: #27384 raytest GINA star 20.04.09

Firmware V4.8

Gabi Star Quadratisch-Durchflußzelle #0

Software Version: 5.9, Service Pack 9, Build 5079

Integration ChA

Substanz Ret Typ Fläche %Fläche

keV CPS*keV %

141 keV 121,55 BB(M) 356764,6 100,00

Summe in ROI 356764,6 100,00

Fläche (total) 381433,4

Ext. BKG. 0,00 CPS

Page 5:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com50 |

radioactivity HPLC

100 – 700 keV, 2x2" NaI(TL)

GABI Starradioactivity-HPLC-flow-monitor, 2x2" NaI scintillation detector

automatic energy calibration

2 simultaneous counting channels

free threshold settings

extremely high dynamic range 0 – 600.000 c/s

output: 0 – 1 V analog output

digital communication to PC

including GINA Star program

live record on PC screen

no range overriding

peak integration, manual or automatic

limit-of-detection calculation

statistical accurcy for every integral

GABI Star is the latest technology µ- processor controlled counter with built in digital ratemeter.

GABI Star stand alone unit offers a keyboard and liquid crystal display for parameter dialog, entry and result presentation.

GABI Star has a high voltage unit for manual or preset selection of the required detector operation.

GABI Star has 2 simultaneous counting channels with each 2 individually selectable energy thresholds each for setting an individual counting window for a particular nuclide.

GABI Star is counting the radiation events in an energy chan-nel per selected interval time.

The information of counts per time is transferred to a rateme-ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard. The output signal 0 – 1 V of GABI Star can be easily adapted to the required chromatography conditions. GABI Star has an extremely high dynamic range from 0 – 600.000 c/s.

Alternatively:

The information of counts per time is transferred to a PC run-ning GINA Star program. The chromatogram is displayed live on screen. No overriding of ranges can happen.

Peak integration, background subtraction, flow correction, absolute activity calculation etc can be performed manually or automatically. Limit-of-detection is determined for every ROI, displayed and printed. GINA Star program can control all modules of Agilent HPLC. All data and parameters are stored under one identity code. GINA Star can be installed as well in background of the original HPLC-operation program and the user can use the operation program of the HPLC supplier as well as the advantages of GINA Star.

GABI Star 2" pin-hole

Integration H:\GD\Mucha 2 Zoll Pinhole Mainz\Tc-99m\140402_03 Start: 02.04.2014 12:46

0 100 200 300 400 500 600 700 keV

Tr1

0 100 200 300 400 500 600 700 keV

0

100

200

300

400

500

600

700

800

CPS/keV

140,keV

Probenbeschreibung

Messung: 140402_03, gestartet: 02.04.2014 12:46

Methode: Tc-99m vom: 02.04.2014 12:45

Aktivität: ,017 MBq zum Referenz Zeitpunkt: 02.04.2014 11.46

Aktivität in 1 mL Lösung; gemessen im GC-Gläschen

Software Version: 5.9, Service Pack 9, Build 5079

Integration Tr1

Substanz Ret Typ Fläche %Fläche

keV cps %

140,keV 154,00 BB(M) 23743,16 100,00

Summe in ROI 23743,16 100,00

Fläche (total) 26441,64

Ext. BKG. 0,00 CPS/keV

 ɣ-spectrum of 99mTc

Page 6:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

01050100 GABI Star ɣ- HPLC-flow-detector

with keyboard and display

0 – 1 V analog output

digital output to PC

01059025 GABI Star scintillation probe 2x2" pin-hole, 16 x 30 mm

01059024 GABI Star 50 mm lead shielding

01059023 GABI Star flow cell 2x2" pin-hole

01058004 ɣy- reference source, 137Cs for 2x2", 3,7 kBq

option

PC + Windows program

0124012 GINA 4-channel programm incl. manual evaluation

01058007 ɣ- reference standard

01240074 installation

01240074 installation GABI Star stand alone

01240239 training GABI Star stand alone

01240218 installation GABI Star with GINA Star

01240220 training GABI Star with GINA Star

ɣ-spectrum 137Cs, reference standard ɣ-spectrum 18F

GABI*

energy range: 30 – 1600 keV high voltage range: 500 – 2000 V count rate: 0 – 500.000 c/s

data input: 7 analog input channels, -0.5 to + 4.5 V resolution 21 bit

data output: RS 232 C, 2 analog outputs 0 - 1 V resolution 20 bit

SCINTILLATION PROBES

NaI(Tl) size: 2 x 2"

pin hole size: 15 mm diameter 30 mm deep

energy range: 100 – 700 keV

typical nuclide: 18F, 99mTc

shielding: 50 mm lead

Integration H:\GD\Mucha 2 Zoll Pinhole Mainz\I-131\140402_05 Start: 02.04.2014 14:20

0 200 400 600 800 keV

Tr1

0 200 400 600 800 keV

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

2122 CPS/keV

Reg #1

Probenbeschreibung

Messung: 140402_05, gestartet: 02.04.2014 14:20

Methode: I-131 vom: 02.04.2014 14:20

Aktivität: ,005 MBq zum Referenz Zeitpunkt: 02.04.2014 11.46

Aktivität in 1 mL Lösung; gemessen im GC-Gläschen

Software Version: 5.9, Service Pack 9, Build 5079

Integration Tr1

Substanz Ret Typ Fläche %Fläche

keV cps %

Reg #1 388,00 BB(M) 862,7509 100,00

Summe in ROI 862,7509 100,00

Fläche (total) 1824,8634

Ext. BKG. 0,00 CPS/keV

Page 7:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com52 |

GABI Star is the latest technology µ- processor controlled counter with built in digital ratemeter.

GABI Star stand alone unit offers a keyboard and liquid crystal display for parameter dialog, entry and result presentation.

GABI Star has a high voltage unit for manual or preset selection of the required detector operation.

GABI Star has 2 simultaneous counting channels with each 2 individually selectable energy thresholds each for setting an individual counting window for a particular nuclide.

GABI Star is counting the radiation events in an energy chan-nel per selected interval time.

The information of counts per time is transferred to a rateme-ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard. The output signal 0 – 1 V of GABI Star can be easily adapted to the required chromatography conditions. GABI Star has an extremely high dynamic range from 0 – 600.000 c/s.

Alternatively:

The information of counts per time is transferred to a PC run-ning GINA Star program. The chromatogram is displayed live on screen. No overriding of ranges can happen.

Peak integration, background subtraction, flow correction, absolute activity calculation etc can be performed manually or automatically. Limit-of-detection is determined for every ROI, displayed and printed. GINA Star program can control all modules of Agilent HPLC. All data and parameters are stored under one identity code. GINA Star can be installed as well in background of the original HPLC-operation program and the user can use the operation program of the HPLC supplier as well as the advantages of GINA Star.

ɣ radioactivity HPLC radiochemical purity test

20 – 2000 keV multiple nuclides

automatic energy calibration

2 simultaneous counting channels

free threshold settings

extremely high dynamic range 0 – 600.000 c/s

LCD 4 lines, 20 characters each

keybord: 16 touch sensitive keys

output: 0 – 1 V

digital communication to PC

including GINA Star program

live record on PC screen

no range overriding

peak integration, manual or automatic

limit-of-detection calculation

statistical accurcy for every integral

GABI Star radioactivity-HPLC-flow-monitor, 3x3" NaI scintillation detector

GABI Star 3" pin-hole

Page 8:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

GABI Star

energy range: 10 – 2000 keV high voltage range: 500 – 2000 V count rate: 0 – 600.000 c/s

data input: 7 analog input channels, -0.5 to + 4.5V resolution 21 bit

data output: RS 232 C, 2 analog outputs 0 - 1V resolution 20 bit

SCINTILLATION PROBES

NaI(Tl) size: 3x3"

pin hole size: 16 mm diameter 40 mm deep

energy range: 600 – 2000 keV

typical nuclide: 59Fe

shielding: 50 mm lead

01050100 GABI Star ɣ-HPLC-flow-detector

with keyboard and display

analog output 0 – 1V

digital output to PC

01052005 GABI Star scintillation probe 3x3" pin-hole, 16 x 40 mm

01052015 GABI Star 50 mm lead shielding

01059027 GABI Star flow cell 3x3” pin-hole

01058004 ɣɣ- reference source, 137Cs for 2x2", 3x3"

option

PC + Windows program

0124012 GINA 4-channel programm incl. manual evaluation

01058007 ɣ- reference standard

01240074 installation

01240074 installation GABI Star stand alone

01240239 training GABI Star stand alone

01240218 installation GABI Star with GINA Star

01240220 training GABI Star with GINA Star

ɣ-spectrum reference standard 137Cs in 3x3" pin-hole scintillation detector ɣ-spectrum of 18F in 3x3" pin-hole scintillation detector

0 500 1000 1500 keV

Tr1

0 500 1000 1500 keV

0

20

40

60

80

100

120

140

160

180 CPS/keV

Ba 32keV

Cs137 662keV

Integraton N:\walker\3x3 pinhole cs137 spectrum\3x3 pinhole cs137 spectrum, IQ OQ 3_7_09

Start: 03.07.2009 16:35

Sample descripton

Measurement:ऀ IQ OQ 3_7_09, started:ऀ 03.07.2009 16:35

Method:ऀ Mucha from:ऀ 03.07.2009 16:30

Zählrate ( 27.083) überschreitet die Gerätespezifkaton ( 20.000)

3x3" NaI pinhole Ser.Nr. 185 Crydet WT76S/UA78

Mucha Star #1276

Cs137 #37195 9,7 kBq

Measurement tme: 1 min

Integration N:\Helfer\Messungen\MUCHA 3 Bohrloch\F-18, 140605_01 Start: 05.06.2014

11:39

0 500 1000 1500 keV

SumTrace

0 500 1000 1500 keV

0

50

100

150

200

250

300CPS/keV

Vernichtungspeak

Summenpeak

Probenbeschreibung

Messung: 140605_01, gestartet: 05.06.2014 11:39

Methode: F-18 vom: 05.06.2014 11:37

Zählrate ( 81.820) exceeds the insturment specification ( 50.000)

Messung mit 3" Bohrlochsonde

Software Version: 5.9, Service Pack 9, Build 5079

Integration SumTrace

Substanz Ret Typ Fläche %Fläche

keV cps %

Vernichtungspea

k

508,00 BB(M) 18758,64 68,38

Summenpeak 1024,00 BB(M) 8675,73 31,62

Summe in ROI 27434,37 100,00

Fläche (total) 81663,66

Page 9:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com54 |

ɣ- miniGITA Star ɣ- radioactivity-TLC-scanner, BGO scintillation crystal

ɣ radioactivity TLC

30 – 1600 KeV, BGO

radiochemical purity

single trace TLC

one plate 200 x 200 mm

multiple nuclide

one trace scan 25 x 200 mm

automatic energy calibration

extremely high counting rate

dead time correction

automatic decay correction

live display on screen

peak integration, TLC evaluation

limit-of detection calculation

miniGITA Star is a scanning device, which moves a radioactivi-ty detector along one trace from start to front.

The repeated, fast, continuous detection from start to front and back compensates automatically the radioactive decay of the compound over the scan length.

For ɣ- nuclides, miniGITA Star is using a scintillation probe with a BGO crystal. Due to its density, BGO has a quite high stopping power for gamma radiation and a reasonable energy resoluti-on. BGO is mechanically stable and non hydroscopic. Relative small size and special shape enable a sophisticated design of the scintillation probe.

There are 5 mechanical collimators designed for the energy ranges of 0 - 60 keV, 60 - 150 keV, 150 - 250 keV, 250 - 450 keV, > 450 keV. Depending on the radiation energy of the radioac-tive compound, the suitable collimator is inserted into the de-tector. Simple tools help to keep the distance exactly the same between the sample surface and detector entry window.

miniGITA Star offers a calibration and sensitivity self check.

Using a suitable reference standard and running the calibration program will result in an energy spectrum scan and calibration.

After a scan the TLC-plate can be moved manually to the fol-lowing trace and the next TLC can be examined.

The chromatogram is displayed live on the screen of the

connected PC. Peak integration and evaluation can be perfor-med manually or automatically.

The measurement and data handing is digital (single event counting) and limit of detection can be determined for every small peak.

Page 10:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

ɣ-miniGITA Star

Scan area: 25 x 200 mm

Scan speed: selectable

Traces: 1

Detector: BGO - scintillation probe

Nuclides: gamma

Energy: 50 – 1600 keV

Activity: 10 Bq – 100 MBq

Decay: corrected

Maximum countrate: 200.000 cps

COLLIMATORS

0-60 keV stainless steel, typical nuclide I-125 3 mm high

60 – 150 keV tungsten, typical nuclide 99mTc 5mm high

150 – 250 keV tungsten, typical nuclide In-111 10 mm high

250 – 450 keV tungsten typical nuclide I-131 15 mm high

> 450 keV tungsten typical nuclide 18F 20 mm high 123I bkg 0.7 cps (20 - 100 keV) sensitivity 20 Bq in 10 min. resolution 2 - 3 mm depending on collimator

02900012 minGITA Star -TLC-scanner all programs included

02900005 miniGITA Star collimator 60 - 150 keV

02900011 ɣ- reference source with holder

option

02900004 miniGITA Star collimator 0 - 60 keV

02900006 miniGITA Star collimator 150 - 250 keV

02900007 miniGITA Star collimator 250 - 450 keV

02900008 miniGITA Star collimator > 450 keV

01240206 installation

01240212 training

complete installation requires PC and WINDOWS

ɣ-spectrum 137Cs, 27kBq, energy calibration, BGO scintillationradioactive TLC, 99mTc

Integration N:\Geber\miniGita\Spektrum\13G1037_Cs.rsp Start: 19.11.2013 13:53

100 200 300 400 500 600 700 keV

TLC

100 200 300 400 500 600 700 keV

0

5

10

15

20

25

30

35

40

45

50

55

60

65C/keV

Reg #1

Probenbeschreibung

Messung: 13G1037_Cs.rsp, gestartet: 19.11.2013 13:53

Messzeit: 0:04:49 Auflösung: 6,27 keV

Hochspannung: 745,0 V

HV= 745 V

Radioaktivitätsdetektor: raytest MiniGITA Serien Nr..: 1301077

Andere Quadratisch-Durchflußzelle #0

Software Version: 5.9, Service Pack 9, Build 5079

Integration TLC

Substanz Ret Typ Fläche %Fläche

keV Counts %

Reg #1 665,10 DD(M) 6272,000 100,00

Summe in ROI 6272,000 100,00

Fläche (total) 13205,000

Ext. BKG. 0,00 C/keV

Page 11:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com56 |

GITA Star ɣ- radioactivity-multi-TLC-scanner, BGO scintillation detector

ɣ radioactivity TLC

radiochemical purity

multiple trace TLC

2 plates 200 x 200 mm

multiple nuclides

80 trace scans on 2 TLC plates 200 x 200 mm

automatic energy calibration

extremely high counting rate

dead time correction

automatic decay correction

live display on screen

peak integration,

TLC evaluation

limit-of detection calculation

GITA Star is a scanning device, which moves the radioactivity detector along one trace from start to front and goes to the next pre-programmed trace position and scans that trace with individual nuclide settings.

For ɣɣ- nuclides, GITA Star is using a scintillation probe with a BGO crystal. Due to its density, BGO has a quite high stopping power for radiation and a resonable energy resolution. BGO is mechanically stable and non-hydroscopic. Relative small size and special shape enable a sophisticated design of the scintil-lation probe.

There are 5 mechanical collimators designed for the energy ranges of 0 - 60 keV, 60 - 150 keV, 150 - 250 keV, 250 - 450 keV, >450 keV. Depending on the radiation energy of the radioactive compound, the suitable collimator is inserted into the detector. Simple tools help to keep the distance exactly the same bet-ween the sample surface and detector entry window.

GITA Star offers a calibration and sensitivity check by a refe-rence source.

After all traces of one TLC plate 200 x 200 mm are ready, the second TLC-plate 200 x 200 mm can be measured.

A single chromatogram can be displayed live on the screen of the connected PC. Multiple traces can be displayed 3- dimen-sional.

Peak integration and evaluation can be performed manually or automatically. The measurement and data handling is digital (single event counting) and limit of detection can be deter-mined for every small peak.

scintillation probe 18F linearity on TLC

Page 12:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

GITA Star

Scan area: 400 x 200 mm

Scan speed: selectable

Traces: 80

Detector: BGO - scintillation probe

Nuclides: gamma

Energy: 30 – 1600 keV

Activity: 10 Bq – 100 MBq

Decay: corrected

Maximum countrate: 200.000 cps

COLLIMATORS

0-60 keV stainless steel, typical nuclide I-125 3 mm high

60 – 150 keV tungsten, typical nuclide Tc-99m 5mm high

150 – 250 keV tungsten, typical nuclide In-111 10 mm high

250 – 450 keV tungsten typical nuclide I-131 15 mm high

> 450 keV tungsten typical nuclide F-18 20 mm high 123I bkg 0.7 cps (20 - 100 keV) sensitivity 20 Bq in 10 min. resolution 2-3 mm depending on collimator

07000010 GITA Starɣmultiple-TLC-scanner all programs included

02900005 GITA Star collimator 60 - 150 keV

02900011 ɣy- reference source with holder

option

02900004 miniGITA Star collimator 0 - 60 keV

02900005 miniGITA Star collimator 60 - 150 keV

02900006 miniGITA Star collimator 150 - 250 keV

02900007 miniGITA Star collimator 250 - 450 keV

02900008 miniGITA Star collimator > 450 keV

01240242 installation

1240243 training

complete installation requires PC and WINDOWS

ɣ-spectrum 137Cs, 27 kBq, energy calibration „BGO scintillator“radioactive TLC, 123I

Integration N:\Geber\miniGita\Spektrum\13G1037_Cs.rsp Start: 19.11.2013 13:53

100 200 300 400 500 600 700 keV

TLC

100 200 300 400 500 600 700 keV

0

5

10

15

20

25

30

35

40

45

50

55

60

65C/keV

Reg #1

Probenbeschreibung

Messung: 13G1037_Cs.rsp, gestartet: 19.11.2013 13:53

Messzeit: 0:04:49 Auflösung: 6,27 keV

Hochspannung: 745,0 V

HV= 745 V

Radioaktivitätsdetektor: raytest MiniGITA Serien Nr..: 1301077

Andere Quadratisch-Durchflußzelle #0

Software Version: 5.9, Service Pack 9, Build 5079

Integration TLC

Substanz Ret Typ Fläche %Fläche

keV Counts %

Reg #1 665,10 DD(M) 6272,000 100,00

Summe in ROI 6272,000 100,00

Fläche (total) 13205,000

Ext. BKG. 0,00 C/keV

Page 13:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com58 |

ɣ total activity meter

nuclear medicine

radiopharmaceutical QC

PET quality control

The well-type ionisation chamber is designed for all potenti-al applications in nuclear medicine and radiopharmaceutical production.

The sensitive electronics is integrated in the foot of the upright standing chamber.

The accuracy over a very large range of activity is very high. The long term stability is due to the sealing technology of the chamber excellent.

Tools for samples and various positioning are supplied.

The measurement geometry is almost 4 Pi.

Background is automatically measured and compensated.

Absolute activity is automatically calculated for more than 30 nuclides.

Minimization of components, integration of all components into the flat housing, requiring only little set up space.

State-of-the-art, powerful computer technology according to the industry standard.

Menu guided Windows based software.

Integration into network and combination with database sys-tem possible for radiopharmaceutical balancing and measure-ment.

Integrated quality control according to EN / DIN standard with protocol printout and data monitoring.

Software includes program for radiopharmaceutical identity checks, e.g. during production of PET nuclides.

User specific result display through adjustable user interface.

Container (syringe size, ampoule, bottle, capsule) and contents are taken into account.

(important especially withɣβ- sources)

Measurement of the activity of all nuclides common in PET; measuring range up to 300 GBq, rapid determination of the bottled activity (T= 1 – 1.5 s).

Background measurement and compensation.

Calculation of measured value for freely selectable application times.

Allocation of measured value to patient possible.

Network connection with patient database.

Reliable chute ionisation chamber in various design.

High measurement accuracy, low position dependency of sam-ple.

Connection of several measurement chambers to one central electronics (multi-chamber-concept).

Upgrade of measurement electronics with additional I/O ports e.g. to control syringe filling machines.

Combinable with:

- laser scanner for reading barcodes.

- transponder readers for control of access right - network systems

Expandable by user-specific software e.g.

- depot management

- nuclide management

- balancing

Certified in accordance with Medical Product Act CE 0123

ISOMEDdose calibrator

Dose calibrator ISOMED

BENEFITS

CONTINUED BENEFITS

Page 14:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

ISOMED

measuring range: 99mTc, 40 kB 9 – 50 GBq at 200 GBq 7% add. error 18F 60 kBq – 70 GBq at 300 GBq 7 % add. error

range setting: automatically alternative fixed for PET filling

energy range: 25 keV – 3 MeV gamma

measurement time:

incl. range change: 2 – 15 s

no rang change: 1-3 s

basic error: < 5 %

linearity error: < 2 %

result display: on screen

stored isotope table: 11C, 13N, 15O, 18F, 32P, 51Cr, 54Mn, 57Co, 58Co, 59Fe, 60Co, 67Ga, 75Se, 89Sr, 90Y, 99Mo, 99mTc, 111In, 113In, 123I, 125I, 131I, 133Xe, 137Cs, 140Ba, 153Sm, 169Er , 169Yb, 186Re, 188Re, 197Hg, 201Tl, 224Ra

containers:

injector syringe: 1, 2, 3, 5, 10, 20 ml

bottles: 5, 10, (P6) 15, 20 ml

ampoules: 5 ml

contents: 0.1- 99.9 ml

measurement cham.: 125/184 mm diameter chute diameter 47 mm

height: 387 mm, chute depth 205 mm

shielding: 4 mm basic Pb shielding

additional: 16 or 50 mm Pb shielding

Dose calibrator ISOMED

07161025 Activimeter 2010 dose calibrator PD version

07161022 Activimeter calibration source, 137Cs, 5 MBq

alternative:

07161023 shielding 16 mm wall

alternative:

07161024 shielding 50 mm wall

option:

01240074 installation dose calibrator

FEATURES

high energy range

more than 30 nuclides

very high accuracy

long term stability

PC-version

Page 15:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com60 |

ɣ spectroscopy

nuclide impurity, low energy

food analysis, low energy

water analysis, low energy

waste analysis, low energy

easy operation

life spectrum display

selectable nuclide ROI`s

calibration to Bq

storage of measurements

management of files

MUCHA Nova The gamma radiation measuring system is designed for analysis of environmental samples or for determi-nation of very low activities in food, water waste samples, etc.

The use of the multi-channel-analyser is simple and self explai-ning. The user is guided through a dialog on the screen.

Display of the measured spectrum is live.

Automatic energy calibration is obtained by using a suitable

137Cs calibration source.

Selected energy regions are coloured.

Background spectra can be accumulated and automatically subtracted.

In order to offer a suitable spectrum display, a low, medium and high energy range can be selected.

MUCHA Novamulti-channel-analyser for ɣ- spectroscopy, 1x1" NaI(TL)

1024 channel MUCHA Nova, 1“ NaI(TI)

Integration H:\GD\S25S_UA50_Cs137_1 Start: 24.06.2014 12:36

0 500 1000 1500 keV

Tr1

0 500 1000 1500 keV

0

1

2

3

4

5

6

7

8

9

10 CPS/keV

32 keV

662 KeV,!37 Cs

Sample description

Messung: S25S_UA50_Cs137_1, started: 24.06.2014 12:36

Methode: Mucha from: 24.06.2014 12:39

3.3 MBq 137Cs (LR620). NaI 1" cristall distance to source 25cm.

Software Version: 5.9, service pack 9, Build 5079

Integration Tr1

Substanc

e

Net R/T Type Area %Area R/T Start Stop Half life

t.

keV cps % keV keV keV min

32 keV 32,00 BB(M) 77,9131 33,62 32,00 14,00 52,00 -

662

KeV,!37

Cs

738,00 BB(M) 153,846

0

66,38 738,00 646,00 812,00 -

Sum in

ROI

231,759

1

100,00

Area

(total)

1179,92

00

Ext.

BKG.

0,00

CPS/ke

V

ɣ-spectrum 1x1" NaI(Tl) full crystal, 137Cs standard

Page 16:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

MUCHA Nova

range: 1024 channels

conversion time: free running ADC

signal input: positive

offset: adjustable

lower discriminator: adjustable

high voltage: 0 – 1300 V

preamplifier: automatic

amplifier: automatic

display: on PC

power: 90 – 260 VAC, 50 – 60 Hz

temperature: 5 – 40°C

ɣ-spectrum, 18F, 1x1" NaI(TI) ɣ-spectrum 1x1", full crystal, 99mTc

12000302 MUCHA Nova 1024 channel analyser I/0 to PC incl. program

01052000 Scitillation Detector NaI(TI) 1” full crystal

01052015 leadshielding for 1" scintillation detector

01058007 ɣ- refernece standard 137Cs, 9 kBq, disk source

01240099 installation

01240211 training

Integration N:\Helfer\MUCHA 1-Zoll Vollkristall (Mainz)\F-18\140704_01 Start:

04.07.2014 10:40

0 100 200 300 400 500 600 700 keV

Tr1

0 100 200 300 400 500 600 700 keV

0

2

4

6

8

10

12

14

16

18

20

22

CPS/keV

18-F

Probenbeschreibung

Messung: 140704_01, gestartet: 04.07.2014 10:40

Methode: F-18 vom: 04.07.2014 10:41

1" Vollkristalldetektor

Software Version: 5.9, Service Pack 9, Build 5079

Integration Tr1

Substanz Ret Typ Fläche %Fläche

keV cps %

18-F 516,00 BB(M) 1169,077 100,00

Summe in ROI 1169,077 100,00

Fläche (total) 3767,275

Ext. BKG. 0,00 CPS/keV

Integration N:\Helfer\MUCHA 1-Zoll Vollkristall (Mainz)\Tc-99m\140704_01 Start:

04.07.2014 10:20

0 100 200 300 400 500 600 700 keV

Tr1

0 100 200 300 400 500 600 700 keV

0

50

100

150

200

250

CPS/keV

Reg #1

Probenbeschreibung

Messung: 140704_01, gestartet: 04.07.2014 10:20

Methode: Tc-99m vom: 04.07.2014 10:18

1" Vollkristalldetektor

Software Version: 5.9, Service Pack 9, Build 5079

Integration Tr1

Substanz Ret Typ Fläche %Fläche

keV cps %

Reg #1 146,00 BB(M) 5890,372 100,00

Summe in ROI 5890,372 100,00

Fläche (total) 8134,717

Ext. BKG. 0,00 CPS/keV

Page 17:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com62 |

ɣ spectrometry

nuclide impurity

food analysis

water analysis

waste analysis

easy to use

life spectrum display

selectable nuclide ROI`s

calibration to Bq

quality control for PET

MUCHA Nova the gamma radiation measuring system is desi-gned for applications, for analysis of environmental samples or for determination of very low activities in food, water and waste samples.

The use of the multi-channel-analyser is simple and self explai-ning. The user is guided through a dialog on the screen.

Every screen offers all operational functions as needed for the analysis to be performed.

Display of the measured spectrum is live.

Automatic energy calibration is obtained by using a suitable 137Cs calibration source.

Selected energy regions are coloured.

Background spectra can be accumulated and automatically subtracted.

In order to offer a suitable spectrum display, a low, medium and high energy range can be selected.

MUCHA Novamulti-channel-analyser for ɣ- spectroscopy, 2x2" NaI(TI)

1024 channel MUCHA Star, 2" NaI(TI)

Page 18:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

MUCHA Nova

range: 1024 channels

conversion time: free running ADC

signal input: positive

offset: adjustable

lower discriminator: adjustable

high voltage: 0 – 1300 V

preamplifier: automatic

amplifier: automatic

display: on PC

power: 90 – 260 VAC, 50 – 60 Hz

temperature: 5 – 40°C

137Cs, 9 kBq energy calibrationMarinelli 1 l, 2x2" NaI(TI), mixed calibration source

01240238 MUCHA Nova 1024 channel analyser I/0 to PC incl. program

09400530 MUCHA Nova 2x2" NaI(TI)

09400534 MUCHA Nova lead shielding, 50 mm Copper 2 mm

01058007 ɣ- refernece standard 137Cs, 9 kBq, disk source

PC + Windows program

options:

01240099 installation

01240211 training

Page 19:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

APPLICATION

GENERAL DESCRIPTION FEATURES

www.raytest.com64 |

ɣ spectrometry

nuclide determination

nuclide impurity

food analysis

water analysis

waste analysis

easy to use

life spectrum display

selectable nuclide ROI`s

calibration to Bq

nuclide impurity in %

quality control for PET

MUCHA Star the gamma radiation measuring system is desi-gned for applications, for analysis of environmental samples or for determination of very low activities in food, water and waste samples.

The use of the multi-channel-analyser is simple and self explai-ning. The user is guided through a dialog on the screen.

Every screen offers all operational functions as needed for the analysis to be performed.

Display of the measured spectrum is live.

Automatic energy calibration is obtained by using a suitable 137Cs calibration source.

Selected energy regions are coloured.

Background spectra can be accumulated and automatically subtracted for the selected measurement time.

In order to offer a suitable spectrum display, a low, medium and high energy range can be selected.

At the end of any measurement the spectrum can be printed or transferred for further analysis or storage to a PC.

MUCHA Nova multi-channel-analyser for ɣ- spectroscopy, 3x3" NaI(TI)

1024 channel UCHA Star, 3“ NaI(TI)

Page 20:  · 2015-01-22 · ter, which converts the digital format to an analog output for-mat of 0 – 1 V. Individual parameters of the ratemeter can be selected and entered by LCD and keyboard.

TECHNICAL DATA ORDERING INFORMATION

γ

MUCHA*

range: 1024 channels

conversion time: free running ADC

signal input: positive

offset: adjustable

lower discriminator: adjustable

high voltage: 0 – 1300 V

preamplifier: automatic

amplifier: automatic

display: on PC

power: 90 – 260 VAC, 50 – 60 Hz

temperature: 5 – 40°C

calibration 137Cs, 3x3" NaI (TI)multi element standard, 3x3" NaI (TI)

01240238 MUCHA Star multi channel analyser incl. PC program

09400515 MUCHA Star 3x3" NaI(TI) pin-hole, 16 x 40 mm

09400534 MUCHA Star lead shielding, 50 mm Copper, 2 mm

01058007 ɣ- refernece standard 137Cs, 9 kBq, disk source

094005 MUCHA Star Marinelli beaker 1 liter

PC + Windows program

options:

01240099 installation

01240211 training