2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

download 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

of 9

Transcript of 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    1/9

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    2/9

    660 X. Yang et al. / Sensorsand ActuatorsB 177 (2013) 659667

    have been applied since the 1970s by the US Air Force [9]. Over

    several decades, 3 wt% TCP has been added into commercial jet

    engine oils, such as Mobil Jet Oil [10]. TCP could be mixed with the

    breathable air entering the airliner cabin during engine oil leakage

    [11]. Although the most toxic ortho-isomer of TCP is intentionally

    excluded from these fluids [12], alternatively used meta- and

    para-isomers can have a potentially severe effect on human health

    mainly through inhalation of aerosols and dermal adsorption.

    The most commonly used methods for detecting TCP include:

    gas chromatography and mass spectrometry GCMS [13], highperformance liquid chromatography (HPLC) [14] or thin layer

    chromatography (TLC) [15]. Gas chromatography measure-

    ment can be based on flame photometric detector (GC/FPD),

    nitrogenphosphorus sensitive detector (GC/NPD) [16]. Although

    they are able to detect TCP below ppm level and with high selectiv-

    ity, these devices have several drawbacks, such as large size, high

    cost, complexity, and a need for highly trained operators, which

    make their use on aircraft impractical. Alternatively, electrochemi-

    cal sensors have the advantages oflow price, highsensitivity, small

    size, ease of operation, and rapid response. A portable real-time

    electrochemical sensor with linear response for ppb level ofTCP

    in gas has been recently reported by our group [17,18]. Since TCP

    itself is electrochemically inactive, it was converted to electro-

    active cresol for detection. However, the sensor system suffered

    from electrode fouling which limited its repetitive usage in long

    term applications. A long lasting electrode is needed for on-site

    continuous monitoring, so electrode surfaces must be modified to

    minimize or eliminate fouling.

    Electrode fouling is a common problem during electrochem-

    ical analysis of phenolic compounds, such as cresol, and causes

    decay in signal response during repetitive measurements [1927].

    The fouling is caused by the formation ofa passive polymeric film

    on the electrode surface [23,25,28,29]. Upon anodic oxidation of

    cresol, phenoxy radicals form, which then couple to form dimers

    or oligomers, and finally, a polymeric film deposits on electrode

    surface [30]. This polymeric film is tough, thermally stable, and

    so chemically inert that little oxidation or hydrolysis will occur in

    either acidic or basic media [26]. The films are characterized by

    low permeability [31] and strong adhesion to the electrode, whichblocks the surface, and yields electrode fouling [1922,24]. Various

    surface treatments and modifications have been used to reduce or

    even avoid electrode fouling by preventing polymeric substrates

    from absorbing onto the surface of electrode [32,33]. For instance,

    a special compound called sodium 3,5-dibromo-4-nitroso benzene

    sulfonate (DBNBS) wasused as an anti-foulingagent against the for-

    mationofthecresolpolymericfilm[33]. TheDBNBSmoleculereacts

    with the oxidized radical ofcresol to form a compound which does

    not adhere to the surface ofthe electrode and consequently pre-

    ventsthe fouling effect. However, thisapproachis difficult to realize

    in practical applications. Overall, the DBNBS should be present in

    the buffer solution in stoichiometric concentration with the ana-

    lyte of interest, which makes its application for real world samples

    problematic.

    Having advantages ofhigh electronic conductivity and porosity[34], conducting polymers have attracted considerable interests in

    recent years. Modification ofconductive polymer for preventing

    electrode fouling has been reported by many researchers [3538].

    Poly(3,4-ethylenedioxythiophene) (PEDOT) is one of the widely

    used conductive polymers. Patra and Munichandraiah [34], Heras

    et al. [39], and Lupu et al. [40] applied PEDOT on electrodes

    through electropolymerization for detection of phenolic com-

    pounds. The surfactant poly(sodium-4-styrenesulfonate) (NaPSS)

    was additionally used during PEDOT electropolymerization to

    avoid the problems of[41]: (1) the low solubility of thiophene

    structures in water, (2) the oxidation potentials higher than that of

    water, and (3) and the water-catalyzed formation of thienyl cation

    radicals which can activate concomitant reactions and prevent the

    formation of the main polymer [42]. Moreover, the hydrophobic

    hydrocarbon residues ofPSS exhibit strongaffinityfor PEDOT, while

    the hydrophilic sulphonic groups are oriented toward or even pro-

    trude into the solution and may hence induce poor adhesion of

    the fouling polymeric film [39,43]. NaPSS also induces the forma-

    tionof a permeable and less compact polymer network, and yields

    high (ionic) conductivity andpermeability to thefouling polymeric

    film. This makes the permeation ofphenolic compounds through

    the film and oxidation inside possible, ensuring electroneutrality[39,43]. In the present work, PEDOT:PSS has been used to mod-

    ify the commonly used glassy carbon electrode [44,45] to develop

    a sensor for continuous monitoring of TCP in gas and has been

    investigated.

    2. Materials and methods

    2.1. Reagents, solutions, and instruments

    All aqueous solutions were prepared using de-ionized

    Milli-Q water (18m cm). 3,4-Ethylenedioxythiophene (EDOT)

    (SigmaAldrich, MO), poly(sodium-4-styrenesulfonate) (NaPSS)

    (SigmaAldrich, MO, MW70,000), and lithium perchlorate

    (LiClO4) (SigmaAldrich, MO) were used for electro-synthesis of

    PEDOT:PSS [39]. The stock 20mM tri-p-cresyl phosphate (p-TCP)

    (SigmaAldrich, MO) solution was prepared in methanol (BDH-

    VWR) and diluted to obtain TCP samples with concentrations of

    interest. All TCP samples were converted to cresol with the aid of

    alkaline catalyst. Alkaline catalyst was made from NaOH (Fisher,

    NJ) and neutral aluminum oxide (Al2O3) (SigmaAldrich, MO) and

    packed in Pipet filter tips (USA Scientific, Inc., 20L leveled filter

    tips). p-Cresol (Acros Organics, NJ, 99+%) was dissolved in 0.2M

    Na2HPO4/0.2 M KH2PO4/10m M NaCl buffer (0.4 M phosphate

    buffer,pH = 6.67). Na2HPO4, KH2PO4, and NaCl were obtained from

    SigmaAldrich, MO.

    2.2. Electrodes and electrochemical sensor

    All amperometric experiments were performed with CH 93Instruments (CH1910B) Bi-Potentiostat. A desktop computer was

    used to collect the data. Flow injection analysis (FIA) was carried

    out using a unicell electrode set (BASi, IN) and a switch injec-

    tion unit (Valco Instruments Co. Inc.) with a 50Lsample loading

    loop. The flow rate was maintained at 20mL/min by using a single

    syringe pump (KD Scientific, MA). The unicell electrode set con-

    sists of one glassy carbon working electrode cell (2mm ), one

    electrode set including the stainless steel counter electrode and

    Ag/AgCl reference electrode, and one circular gasket (BASi, IN). A

    coating solution was applied on the Ag/AgCl reference electrode

    surface before detection as per instruction from the company. The

    working electrode was polished with alumina powder (BUEHLER,

    IL, 1, 0.3, and 0.05m in order). 10mM NaCl was contained in all

    solutions to maintain the potential ofreference electrode. 0.64 V

    vs. Ag/AgCl/10mM NaCl was applied in amperometry. In batchmode, a glassy carbon electrode (3mm ), Pt counter electrode,

    and Ag/AgCl/3M KCl reference electrode (BASi, IN) were used.

    2.3. Electropolymerization of PEDOT:PSS onglassy carbon

    electrode

    The electropolymerization ofmonomer of EDOT was performed

    under amperometric conditions in aqueous solution. The solution

    contained 5 mM EDOT (0.7108g/L), 0.1M NaPSS (20.62g/L), and

    0.1M LiClO4 (10.64 g/L). The unicell glassy carbon electrode was

    polished with alumina as mentionedin Section 2.2, rinsedand soni-

    cated with de-ionized water in an ultrasonic bath for 5 min, flushed

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    3/9

    X. Yang et al. / Sensors and Actuators B 177 (2013) 659667 661

    Fig. 1. Schematicrepresentativeof TCPsamplingsystem. TCPwas heated to 230 C inoil bathand gasified along withN 2 bubbling, and entered thealkalinecatalyst column

    set insidean automatic TCP conversion box. In theelectronicallycontrolled TCPconversion box, as shown in thedottedsquarein above figure, TCPwas hydrolyzed for later

    detection. The control of the automatic TCPconversion box is described in Ref. [17].

    with ethanol and water, and dried with N2 gas. 0.5mC of charge

    was applied on the glassy carbon electrode with 2mm diameter

    in order to obtain the thickness ofinterest ofelectropolymeriza-

    tion layer, unlessotherwisestated. The potential was maintainedat

    0.95Vvs.Ag/AgCl/3MKCl. Thecolorofelectrodesurfaceturnedyel-lowafter modification. For comparison, different amount ofcharges

    have been applied including 0.3, 1, 1.3, 2, 4, and 20mC on the same

    electrode.

    2.4. Automatic sampling system forhydrolysis of TCP samples

    and TCP in engine oils

    Since p-TCP has a very low saturated vapor pressure at room

    temperature (1104 mmHg0.01Paat20 C,100Paat220 C,

    1000Pa at 260 C, [4446]), the TCP samples in gas phase are not

    readily available. A system was built in our lab for TCP sampling

    and is shown schematically in Fig. 1. 0.5mL of TCP methanol solu-

    tions with concentrations of10, 20, 30, 50, 100, 200, and 300M

    were prepared from the20 mM TCPstock solution.N2was bubbledat a flow rate of 1.1L/min through the container with TCP for

    5min at room temperature and then 5 min at 230 C (temperature

    controlled with an oil bath) to make sure all TCP was evaporated.

    The total N2 volume in 10min was 11 L 0.5mol (considered at

    room temperature), therefore, TCP samples with 10, 20, 30, 50,

    100, 200, and 300 ppb in gas phase were realized. Each sample was

    flown along with N2 through the alkaline catalyst column where

    it was hydrolyzed, following the dotted (red) arrow path in Fig. 1.

    The alkaline catalyst column was prepared by packing a 100mg of

    mixture of NaOHand Al2O3(1:10wt) inside.The preparation ofthe

    catalyst is described in more detail in Ref. [17]. In the electronically

    controlled TCP conversion box (dotted square in Fig. 1), valvesturned after 10min ofgas flow and the pumpworked topush 3mL

    0.4M phosphate buffer solution to wash out TCP hydrolyzate for

    detection at the sensor, following the solid (blue) arrow path in

    Fig. 1 (its method of control was mentioned in Ref. [17]).

    Engineoil samples were preparedby dissolving thesamples into

    methanol in 2L/0.5 mL ratio and with the same procedure as TCP

    samples. Oil BP274 does not include TCP, Mobil Jet Oil II includes

    13%, and BP2380 includes 15%TCP [47]. BP274oil samples spiked

    with TCP in different concentrations were also prepared, sampled,

    and measured.

    3. Results and discussion

    3.1. Diffusion controlled cresol oxidation

    The oxidation process of cresol on a bare glassy carbon working

    electrode was studied with cyclic voltammetry (CV). In the CV

    curve (Fig. 2), current increased quickly at the beginning in each

    potential scan due to the charging current, followed by a flat line.

    Starting from around0.4 V, current increasedagainwhichindicates

    the onset of oxidation of cresol. The oxidation was restricted to

    Fig. 2. Cyclicvoltammetry results ofcresoloxidationon bare glassycarbon electrode. (a)Firstcycleof cyclic voltammetryof 100M cresolin 0.4M phosphatebuffer solution

    at scan rates varying from 10 to 200mV/s. scan rates increase along the arrow direction. Range of potential: 0800mV vs. Ag/AgCl/3M KCl. (b) Calibration of the relation

    between thepeakcurrent density andsquare root ofthe scan rate. Thecoefficientof determination R2 indicates the variabilityfrom linear fit. Glassy carbonworkingelectrode

    (3 mm) was usedfor all of detections and polished between uses.

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    4/9

    662 X. Yang et al. / Sensorsand ActuatorsB 177 (2013) 659667

    Fig. 3. Observationsof fouling on glassy carbon electrode from SEMand EDS. (a) SEMimage of Au sputtering coated bare electrode, (b) SEM image of fouled electrode, and

    (c) EDS results of thebare and fouledelectrodes.

    the vicinity of electrode surface and reduced the concentration

    of oxidizable cresol in this area. Once the concentration gradient

    existed between the vicinity of electrode surface and the bulk

    solution, diffusion occurred which yielded a mass flow ofcresol

    from bulk to electrode surface. In this batch mode, migration was

    reduced to negligible levels by addition ofa high concentration of

    phosphate ions and convection was avoided by preventing stirring

    and vibrations, which made diffusion the only path for mass

    transfer. We define a diffusion layer where linear concentration

    gradient exists. The thickness of the diffusion layer increased with

    time which would flatten the concentration profile. In contrast,

    the potential continued to scan positively and decreased the

    concentration of oxidizable cresol in the vicinity of electrode

    surface, which steepen the concentration profile. The slope of con-

    centration difference was intensified; when the effect ofpotential

    scan dominated first, so did the diffusion, and thus the current

    increased which yields the left shoulder of CV peak. Potential scan

    kept going positively and the concentration of oxidizable cresol

    in the vicinity decreased approaching zero. By the point when

    vincinity concentration dropped relatively slower, since it already

    approached zero, the diffusion layer thicken effect dominated

    and thus the slope of concentration difference reduced. The

    current started to decrease, meanwhile, an IV peak was created

    [48].

    Fig. 4. PEDOT:PSSmodification of glassy carbonelectrodes. (a) Amperometric results of 4 electrodesapplied with0.5 mC of charge in EDOT:PSS solution, (be)optical images

    ofthese 4 electrodes after modification with 0.5mC of charge, (fi) opticalimages of 4 electrodes after modification with 1, 1.3, 2, and 4mC of charge, (jl) SEM images of

    electrode surfaceapplied with 0.3, 0.5, and 2 mC of chargein EDOT/PSS solution.

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    5/9

    X. Yang et al. / Sensors and Actuators B 177 (2013) 659667 663

    Fig. 5. Amperometric results of 10M cresol on bare and modified electrodes. (a) Initial amperometric peaks obtained from successive injections of 10M cresolin 0.4M

    phosphate buffer. Two experiments have been carriedout on bare electrodeand fiveon modified electrodes, two examples of which are shown here. (b) Calibration of peak

    current vs. peaks. Note that the first peak ineachcase has been normalized to 100%. Unicell glassycarbonelectrodes with 2 mm diameter were used.

    The results of CV experiments demonstrating the change of the

    IV at different scan rates are shown in Fig. 2. 100M p-cresol

    in 0.4M phosphate buffer was used as the analyte solution, and

    the potential was scanned in the range of 00.8V vs. Ag/AgCl/3M

    KCl with scan rates of10, 20, 50, 80, 100, 120, 150, and 200 mV/s.

    Following the higher scan rate, the whole IV cycle expanded due

    to the increased charging current. The peak current also increased

    (Fig. 2a). Fig. 2b shows the calibration ofcurrent density vs square

    root of scan rate. The linear relation indicates the oxidation pro-

    cess of cresol is limited by the diffusion to electrode surface, while

    the electron transfer at electrode/solution interface is relatively

    quick [41,49]. The diffusion controlled property ofcresol oxidation

    ensures thatcurrent signalis proportional to the bulkconcentration

    inCV as well as amperometry mode.

    Lookingat thereversal scan curve in Fig. 2a,we note theabsence

    of reduction peak which indicates the process was not reversible.

    Cresol was oxidized to radicals, followed by the radical coupling

    and the formation of inert dimers or oligomers. A stable polymericlayerformed,finally, andbroughtthe electrode fouling [30]. Indeed,

    while this fouling layer could be removed by physical polishing, it

    is inert to chemicals and cannot be oxidized until a potential ofup

    to 3V is applied [50].

    3.2. Microscopy images of electrode fouling

    The formation offouling layer was studied by SEM, EDS, and

    AFM. For SEM imaging, the electrodes were sputter coated prior to

    analysis. One electrode was left polished, while the other one was

    fouled by 10 injections of 10M cresol in 0.4MPB buffer in amper-

    ometry mode. The polished electrode and fouled electrode were

    comparedin SEMimages (Fig.3aand b).Onlysputter coatedAu was

    seenon polished electrode,but additionalstructures wereobserved

    from the surface of fouled electrode. EDS shows the existence ofoxygen on thefouledelectrode, which could come from the fouling

    products of cresol (Fig. 3c). AFM images also confirm the existence

    of fouling layer on electrode surface (Results not shown). The sur-

    face profile of fouled electrodes varied in the range of 030 nm

    while the variation in the bare electrode was less than 15nm.

    3.3. Controlled PEDOT:PSS modification onglassy carbon

    electrode

    To reduce the fouling from oxidation of cresol, PEDOT:PSS

    modification was applied on glassy carbon electrode via elec-

    tropolymerization. One important parameter for modification is

    the reproducibility. To show the ability to control modifica-

    tion, 4 unicell glassy carbon electrodes (2mm ) were immersed

    in EDOT/NaPSS solution and applied 0.5 mC of charge (poten-

    tial= 0.95V in single-potential amperometry mode). Fig. 4ae

    shows the results ofthese modifications. The amperometry pro-

    files were almostthe same forthese 4 electrodesas shownin Fig.4a,

    and all of these 4 electrodes had the same color after modification

    Fig. 6. Amperometric results of injections of 0.210M cresol in 0.4M phosphate

    buffer. (a) Two representative sets of continuously amperometric results. Different

    concentrations of cresol samples were injected in random order to avoid poten-

    tial system error and by this way electrode fouling could be seen more clearly.

    (b) Calibration curve of detection of cresol with different concentrations on modi-

    fied electrode. The coefficient of determination R2 indicates the variation of linear

    fit. Error bars are marked as bars above and below current symbols. Unicell glassy

    carbon electrodes with 2 mm diameter were used.

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    6/9

    664 X. Yang et al. / Sensorsand ActuatorsB 177 (2013) 659667

    which indicates the uniform and controllable configuration. Addi-

    tionally, 4 more electrodes were coated with 1, 1.3, 2, and 4mC

    of charge. The color of modified electrode surface changed with

    different amounts ofcharge (Fig. 4fi).

    SEM was also carried out to compare the electrode surfaces

    modified with 0.3, 0.5, and 2 m C of charge. Images from left to

    rightcorrespondto increasinglymore charge(Fig.4jl). During SEM

    imaging, the configuration was seen to be uniform over the whole

    electrode surface and so representative SEM images in small areas

    can be produced. Fig. 4jl illustrate the change in morphology andparticle size of the PEDOT layer. Cracks formed on the PEDOT:PSS

    layer prepared by applying 2 mC of charge

    3.4. Detection of cresol with modified electrodes

    The controlled PEDOT:PSS modification described above was

    used to reduce fouling effect in the amperometric detection of

    cresol. 10M cresol in 0.4M phosphate buffer was injected suc-

    cessively on both modified (left side ofFig. 5a) and un-modified

    electrodes (right side of Fig. 5a). The peak currents decayed in

    both cases after repeated exposure to cresol, but magnitude of

    the decrease was much lower in the modified electrode. For com-

    parison, the peak height was normalized to the first peak in each

    case being normalized to 100%. The successive peaks were seri-ously decayed on bare electrode but much less fouling occurred

    on modified electrode (Fig. 5b). The comparison indicates that for

    continuous on-site monitoring ofTCP, the bare electrode is not a

    good candidate. Although thoroughly polishing the electrode could

    remove the fouling layer, this would require trained personnel and

    adds complexity to sensor operation. Modified electrodes, how-

    ever, overcome this limitation.

    With modified electrodes, a series of cresol samples with con-

    centrations of 0.2, 0.5, 1, 2, 5, 7.5, and 10M were detected in

    amperometry mode 3 times at each concentration). Two represen-

    tativeexamplesofinitialamperometric results are shownin Fig.6a.All samples were injected in random order to avoid system error

    andto show foulingclearer since signal from a samplewoulddecay

    more if a higher concentration was previously injected. The cali-

    bration curve is shown in Fig.6b. A linear relationship between the

    peak current and the concentration ofcresol was obtained, con-

    firmed by the R2 being very close to unity. The error bars which

    were contributed mainly from electrode fouling were too small to

    be distinguished from the data symbols. Therefore, thedetection of

    0.210M cresol was reliable with the modified electrode.

    3.5. Detection of TCP with modified electrodes

    To show the ability of detecting TCP samples, several TCP

    samples with 10, 20, 30, 50, 100, 200, and 300p pb were gasi-fied, hydrolyzed, and detected with PEDOT:PSS modified electrode.

    Fig. 7. Detectionof TCPon bare/modified electrode. (a,left)Calibrationof detectionof TCPsamples with concentrationsof 10,20, 30, 50,100, 200, and300 ppbon PEDOT:PSS

    modified electrode, (a, right) Calibration of detection of 30, 50, 100, and 300ppb TCP on bare electrode. 10 and 20ppb were not detected and compared with modified

    electrode, (b) responsesfrom modified electrode were normalized to 100%and the error bars andsignals from bare electrodewere relatively calibrated. RSD wascalculated

    from the ratio of Stdev (standard deviation) to Avg (average). Unicell glassy carbon electrodes with 2mm diameter were used. Note that the bare electrode was polished

    after measurement of each setof 4 samples, while modified electrode was not polished through three sets of measurements.

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    7/9

    X. Yang et al. / Sensors and Actuators B 177 (2013) 659667 665

    Experiments have been carried out three times. The calibration of

    allthree experiments is shown in theleft frame ofFig. 7a. Errorbars

    indicate the variation between all three experiments. A linear rela-

    tionship was observed between current signal and concentration

    of sample in the range of 50300ppb. For comparison, the corre-

    sponding results from bare electrodes are shown in the right side

    ofFig. 7a, excluding 10 and 20ppb TCP. Although a linear relation

    wasalsoobserved, theerror bars were much biggerdue toelectrode

    fouling. Note that the bare electrode was polished after measure-

    ment of each set of 4 samples. The fouling effect would be muchmore serious if the electrode were left unpolished through the

    whole measurement, which was done for modified electrode. To

    observe the comparison clearly, the responses from modified elec-

    trodeswerenormalized to100% andthe signalsfrombare electrode

    and error bars were relatively calibrated (Fig. 7b). Comparing the

    error bars andRSD values from modifiedelectrode with those from

    bare electrode,modified electrode wasseen to have much less foul-

    ing. Indicated by these results, the sensing system with PEDOT:PSS

    modifiedelectrode is able to continuouslymonitor TCPin gasphase

    in the concentration range of50300ppb. It should be mentioned

    that the lower point of the linear range for the PEDOT:PSS modi-

    fied electrode (50 ppb) is comparable with others work. De Nola

    reported the LOD of1.4 pg/20 nL(70ppb, TCP in isohexane) using

    gas chromatography [13].

    Previously, our group reported the detection ofTCP with bare

    glassy carbon electrodes with a linear range of5300 ppb [51].

    However, the bare electrodes required thorough polishing before

    each measurement, which would make impossible its usage for

    continuously monitoring TCP on the aircraft.

    3.6. Detection of hydrolysates from engine oils

    Finally, to test the ability ofthis system to determine TCP in

    air from real samples, such as samples from jet engine oils, 2L

    of commercially available BP Turbo Oil 274, BP Turbo Oil 2380, and

    MobileJet Oil II weremixed in 0.5 mLof methanol and were usedto

    prepare the hydrolysate samples with the same procedure as TCP

    hydrolysates described above. The results ofoil hydrolysates with

    Fig. 8. Detectionof commercial engineoil andoil spikedhydrolysatesamples.All samples were measured randomly 3 times.(a) Results of hydrolysate samples from engine

    oils BP 274, Mobil, and BP 2380with modified electrode; (b) with unmodified electrode. (c) Calibration of the concentrations of TCP in the oilsamples, and the results from

    modified electrode and unmodified electrode were compared with the claimed values by manufactures. (d) Results of hydrolysate samples from oil BP 274 spiked with

    0200 ppbTCP with modified electrode;(e) with unmodifiedelectrode. (f)Comparisonof theresults between themodifiedand unmodifiedelectrode. RSD(relative standard

    deviation) was calculated the same as Fig. 7.

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    8/9

  • 8/10/2019 2013-Anti-fouling PEDOT-PSS Modification on Glassy Carbon Electrodes for Continuous_2

    9/9

    X. Yang et al. / Sensors and Actuators B 177 (2013) 659667 667

    [36] A. Malinauskas, Electrocatalysis at conducting polymers, Synthetic Metals 107(2) (1999) 7583.

    [37] H.B. Mark, et al., The electrochemistry of neurotransmitters at conductingorganic polymer electrodes: electrocatalysis and analytical applications, Bio-electrochemistry and Bioenergetics 38 (2) (1995) 229245.

    [38] S.S. Rosatto, L.T. Kubota, G. de Oliveira Neto, Biosensorfor phenolbasedon thedirect electrontransferblockingof peroxidaseimmobilisingon silicatitanium,Analytica Chimica Acta390 (13) (1999) 6572.

    [39] M.A. Heras, et al., A poly(3,4-ethylenedioxythiophene)-poly(styrenesulphonate) composite electrode coating in the electrooxidation of phenol,Electrochimica Acta50 (78) (2005) 16851691.

    [40] S. Lupu, et al., Electrochemical sensors based on platinum electrodes modified

    with hybrid inorganicorganic coatingsfor determinationof 4-nitrophenolanddopamine, Electrochimica Acta 54 (7) (2009) 19321938.

    [41] E. Tamburri, et al., Growth mechanisms, morphology, and electroactivity ofPEDOT layers produced by electrochemical routes in aqueous medium, Syn-thetic Metals 159 (56) (2009) 406414.

    [42] N. Sakmeche, et al., Improvement of the electrosynthesis and physico-chemical properties of poly(3,4-ethylenedioxythiophene) using a sodiumdodecyl sulfa te micellar aqueous mediu m, Langmu ir 15 (7) (1999)25662574.

    [43] L. Pigani, et al., Electro-oxidation of chlorophenols on poly(3,4-ethylene-dioxythiophene)-poly(styrene sulphonate) composite electrode, Electrochi-mica Acta 52 (5) (2007) 19101918.

    [44] WHO, et al., Tricresyl Phosphate, World Health Organization, 1990.[45] D.R. Lide, CRCHandbookof Chemistryand Physics: A Ready-ReferenceBook of

    Chemical and Physical Data,CRC Pr I Llc, 2004.[46] CAMEO-Chemicals, Tricresyl Phosphate, 1999.[47] C. van Netten, V. Leung, Comparison of the Constituents of Two Jet Engine

    Lubricating Oils and Their Volatile Pyrolytic Degradation Products, AppliedOccupational and Environmental Hygiene 15 (2000) 277283.

    [48] A.J. Bard, L.R. Faulkner, ElectrochemicalMethods: Fundamentals and Applica-tions, 2nd ed., Wiley, New York, 2001,xxi,833 p.

    [49] X. Du, Z. W an g, E ffects of poly me ri zati on poten ti al o n the pro pe rtie sof electrosynthesized PEDOT fi lms , E le ctro ch imica Acta 48 ( 12) ( 2003)17131717.

    [50] R.C. Koile,D.C. Johnson, Electrochemical removal of phenolic films from a plat-inum anode, Analytical Chemistry 51 (6) (1979) 741744.

    [51] X.Yang,et al., Portable and remote electrochemical sensing system for detec-tion of tricresyl-phosphate ingas phase,Sensorsand Actuators B: Chemical161(1) (2012) 564569.

    Biographies

    Xiaoyun Yang received the B.S. degree in Chemistry from Tongji University,Shanghai, Chinain 2006, followedby 2-years studying for Masters degree in Physicsdepartment. Currently, he is pursuing his PhD degree in Materials Engineering atAuburn University.

    Jeffrey Kirsch received his B.S. in Materials Engineering at Auburn University,Auburn, Alabama in2011. Currently, heis pursuing hisMasters Degree in MaterialsEngineering at Auburn University.

    Dr.EricV.Olsenisthe Directorof theClinicalResearch at KeeslerAFB Mississippi. Hereceived hisPh.D.in Biological Sciences andM.S. in Microbiologyfrom Auburn Uni-versity. His research interests include PCR assay development, piezoelectric-basedbiosensor systems andbio-preservation techniques. Heis a Lt.Colonelin the UnitedStates AirForce with over 20 years of service.

    Dr.JeffreyW. Fergus received his B.S. degree in Metallurgical Engineering fromtheUniversity of Illinoisin 1985andhis Ph.D. degreein MaterialsScienceand Engineer-ing from the University of Pennsylvania in 1990. He was a post-doctoral researchassociate in theCenter for Sensor Materialsat theUniversity of Notre Dame and, in1992,joined the Materials Engineering program at Auburn University, where he iscurrently a professor. His research interests are generally in high-temperature andsolid-state chemistry of materials, including electrochemical devices (e.g. chemicalsensors andfuel cells)and thechemicalstability of materials(e.g.hightemperatureoxidation).

    Dr.AlexL. Simonianisa Professorof MaterialsEngineeringat AuburnUniversityanda Biosensing Program Directorat NSF. Hereceivedhis M.S. inPhysics from theYere-van State University (Armenia, USSR), a Ph.D. in Biophysics and a Doctor of Sciencedegree in Bioengineering from the USSR Academy of Science. His current researchinterests are primarily in the areas of bioanalytical sensors, nano-biomaterials andfunctional interfaces.