20080912_February1989CSBohannon
-
Upload
tripty-khanna -
Category
Documents
-
view
213 -
download
0
Transcript of 20080912_February1989CSBohannon
-
8/11/2019 20080912_February1989CSBohannon
1/3
Accuracy
of
Spring and Strain Gauge
Hand-held Dynamometers
RICHARD W BOHANNON, A WILLIAMS ANDREWS2
The accuracy of tw o spring gauge and two strain gauge hand-held dynamometers
was determined using certified weights. Each dynamometer, which had extensive prior
use, was vertically loaded with the certified weights in 5 Ib increments from 5-55 Ibs.
Analysis of variance was used to compare the actual certified weights with the weights
measured by each dynamometer. Additionally, Pearson product moment correlations
were calculated between the weights measured by e ach device. T he two spring gauge
dynamometers measured comparably as did the two strain gauge dynamometers. Only
the two strain gauge dynamometers, however, registered (measured) weights not
differing significantly from the actual weights with which they were loaded. The corre-
lations between each device s measurements were 0.98 or above. If the dynamometers
tested are indicative, strain gauge dynamometers may be more accurate than spring
gauge dynamometers after extensive use .
The imprecision of subjective manual muscle
test grades and their insensitivity to changes in
muscle strength provide clinicians with an impetus
for using more objective instruments for measur-
ing muscle strength (1). Among the instruments
that have been used extensively for this purpose
in a clinical environment are hand dynamometers,
fixed dynamometers, isokinetic dynamometers,
and hand-held dynamometers
(2).
Unlike that of
some other instruments (3-6), the accuracy of
specific hand-held dynamometers has not been
determined and reported. Before clinicians can
make informed decisions regarding the purchase
and use of instruments, they must be aware of
the instruments' accuracy. The purpose of this
brief report is to compare the accuracy of two
spring gauge and two strain gauge dynamome-
ters. We expected from previous informal calibra-
tion checks that the two strain gauge dynamom-
eters, unlike the two spring gauge dynamometers,
would be accurate. We expected the measure-
ments obtained with all the dynamometers to be
highly correlated.
Associate Professor, School of Allied Health. University of Connecti-
cut. U-101. Storrs. CT 06269.
Graduate student. Division of Physical Therapy. University of North
Carolina. Chapel Hill. NC.
0196-6011/89/1008/0323 02.00/0
THEJOURNAL
F
ORTHOPAEDIC
ND
SPORTSHYSICALHERAPY
Copyright
Q
1989 by The Orthopaedic and Sports Physical Therapy
Sect~ons f the American Physical Therapy Association
METHOD
Two different types of gauges were com-
pared. The two spring gauge dynamometers were
the same model (Spark hand-held dynamometer,
Spark Instruments and Academics, Inc, P.O. Box
5123, Coralville, IA 52241). They had a range of
0 to 60 Ibs and registered orce with a dial display.
The dial face was divided into half-pound incre-
ments. The two strain gauge dynamometers were
the same model (Chatillon model DFG-100, John
Chatillon and Sons, 83-30 Kew Gardens Road,
Kew Gardens, NY 11415-1
999 .
They had a range
of 0 to 115 Ibs and registered force with a digital
display. The display registered to 0.1 Ibs of sen-
sitivity. All four dynamometers were over 2 years
old and had been used in thousands of clinical
tests.
To measure instrument accuracy, each in-
strument was placed vertically in a vise and a
pendulum goniometer was used to assure that its
flat end piece was perpendicular to the pull of
gravity. Certified weights were then placed on the
flat end piece of each device in 5 Ib increments
from 5-55 Ibs. (Fig. 1). Each dynamometer was
reset (zeroed) between the application of each
calibration load.
The weight indicated (measured) by each
dynamometer was compared to the actual certi-
fied weight it was measuring as well as to the
weight measured by the other three dynamome-
ters. An analysis of variance (ANOVA) procedure
JOSPT February
989
SPRING AND STRAIN HAND-HELD DYNAMOMETERS
323
-
8/11/2019 20080912_February1989CSBohannon
2/3
and a Tukey HSD post hoc test were used for
this purpose. In addition, Pearson product mo-
ment correlations were calculated between each
instrument s measurements of the weights.
RESULTS
Figure
2
illustrates the measurements of the
certified weights obtained using each dynamom-
eter. A close inspection of the figure reveals that
throughout the range of weights applied, the two
spring gauges measured higher weights than the
two strain gauges. At every 5 Ib increment both
spring gauges measured at least 1 Ib high. The
mean weight of the certified weights was 30 Ibs,
whereas the mean of the weights measured by
the dynamometers were strain gauge one (30.0
Ibs), strain gauge two (30.0 Ibs), spring gauge one
(31.7 Ibs) and spring gauge two (31.8 Ibs). The
ANOVA results are reported in Table 1. They
reveal that the actual certified weights and the
weights measured by the four dynamometers
were significantly different
p