2-Inlets.pdf

download 2-Inlets.pdf

of 24

Transcript of 2-Inlets.pdf

  • 7/29/2019 2-Inlets.pdf

    1/24

    Inlets

  • 7/29/2019 2-Inlets.pdf

    2/24

    2

    In This Section, You Will Learn:

    Identifying the Types of Inlets

    Identifying the Components in the Packed and Split Inlets

    Understanding the Flow Dynamics of the Split Inlet

    Calculating Split Ratio

    Identifying preventative Maintenance and Troubleshooting

  • 7/29/2019 2-Inlets.pdf

    3/24

    3

    Inlet Systems

    Purpose: To allow the insertion of a sample into the gas

    chromatograph in a repeatable, reproducible manner.

    The sample should be representative of the bulk, and

    unless specifically desired should be inserted without

    chemical change.

    EPC onlyVolatiles Interface

    EPC onlyProgrammed Temperature Vaporization

    EPC onlyCool on-columnEPC and non-EPCPurged Pack

    EPC and non-EPCSplit/Splitless

    Gas ControlInlet Type

  • 7/29/2019 2-Inlets.pdf

    4/244

    Flow System: Purged Packed Inlet

    Flow controlled mode (recommended for packed columns)

    Pressure controlled mode (recommended for capillary columns)

  • 7/29/2019 2-Inlets.pdf

    5/245

    Exploded View of Purged Packed Inlet

    Insulation

    Insulation cup

    Ferrule

    Column nut

    Insulation

    Septum nut

    Knurled nutHeater/sensor assembly

    SeptumO-ring

    PurgedPackedInlet body

    Insulation

  • 7/29/2019 2-Inlets.pdf

    6/246

    Liners/Adapters

    Inlet Fitting

    1/4-Inch Ferrule

    1/4-Inch Nut

    1/8-Inch Liner

    Front Ferrule

    Back Ferrule

    1/8-Inch Nut

    1/8-InchColumn

    Using Two Wrenches in Opposition

    to Tighten Column Fittings

    1/4-Inch MetalFerrule

    Back Ferrule

    1/4-Inch Nut

    1/4-Inch Column

    Inlet Fitting

    1/4-Inch Metal Column, 1/8-Inch Metal Column,

    Packed InletPacked Inlet

    There are a variety of liners

    and adapters for different

    columns, inlets and detectors.

  • 7/29/2019 2-Inlets.pdf

    7/247

    Split/Splitless Inlet

    Split Mode forPulsed Split

    Splitless Mode forPulsed Splitless

    Major Component AnalysesAllows for larger injection volume.

    Trace Component AnalysesAllows for larger injection volume

  • 7/29/2019 2-Inlets.pdf

    8/24

    8

    Split/Splitless Flow System

  • 7/29/2019 2-Inlets.pdf

    9/24

    9

    Split Flow Diagram: Pre-Injection

    SEAL

    (INLET)

    (SEPTUM)

    SPLIT VENT

    TOTAL FLOW

    COLUMN

    SEPTUM

    CARRIER GAS

    LINER

    50 mL/min

    2 mL/min

    0.6 mL/min

    48 mL/min

    47.4 mL/min

    PURGE VENT

    PURGE VALVE

  • 7/29/2019 2-Inlets.pdf

    10/24

    10

    Split Flow Diagram: Sample Injection

    (INLET)

    (SEPTUM)

    SPLIT VENT

    TOTAL FLOW

    COLUMN

    = CARRIER GAS

    PURGE VALVE

    PURGE VENT

    SPLIT RATIO = COLUMN + SPLIT VENT FLOW

    COLUMN FLOW

    = LIQUID SAMPLE MOLECULES

  • 7/29/2019 2-Inlets.pdf

    11/24

    11

    Split Flow Diagram: Sample Vaporization

    COLUMN

    (INLET)

    (SEPTUM)

    SPLIT VENT

    TOTAL FLOW

    = CARRIER GAS

    = SAMPLE MOLECULES

    = SOLVENT MOLECULES

    PURGE VENT

    PURGE VALVE

    Vaporization takes place whensufficient thermal energy istransferred to the sample. Thiscan occur in the carrier gas, but

    is more likely upon contact witha solid surface, such as the lineror mixing medium.

  • 7/29/2019 2-Inlets.pdf

    12/24

    12

    Split Flow Diagram: Sample/Carrier Mixing

    (INLET) PURGE VALVE

    (SEPTUM)PURGE VENT

    SPLIT VENT

    TOTAL

    FLOW

    COLUMN

    = CARRIER GAS

    = SAMPLE MOLECULES

    = SOLVENT MOLECULES

    For the concept of SPLIT RATIO tobe valid, the sample (solvent +analyte) must be mixed with thecarrier gas to give a homogeneousmixture.

    At the bottom of the injection port a

    small part of this mixture will transferto the column, while the bulk of themixture will leave the chromatographvia the SPLIT VENT.

  • 7/29/2019 2-Inlets.pdf

    13/24

    13

    Split Flow Diagram: Liner Overload

    (INLET)

    PURGEVALVE

    (SEPTUM)

    PURGE VENT

    SPLIT VENT

    TOTAL FLOW

    COLUMN

    = CARRIER GAS= SAMPLE MOLECULES

    = SOLVENT MOLECULES

    A large injection of a solvent with alarge expansion volume can causean overload of the injection portliner.

    This can result in loss of sample out

    the PURGE VENT as well ascontamination of the in-comingcarrier gas line.

  • 7/29/2019 2-Inlets.pdf

    14/24

    14

    Split Ratio Calculations

    Split Ratio = Split Vent Flow + Column Flow

    Column Flow

    Split Vent Flow = (Split Ratio)(Column Flow) - Column Flow

    33:1952.03.0100530

    98:1972.01.0100320

    196:197.52.00.5100200

    SplitRatioSplit Flow

    Septum

    Purge

    Column

    FlowTotal Flow

    Column

    I.D. (m)

    DCBA

  • 7/29/2019 2-Inlets.pdf

    15/24

    15

    Gas Saver

    Gas Saver reduces the carrier gas flow from the split vent after thesample is on the column.

    Column head pressure and the column flow rate are maintained,while the purge/split vent flows decreases. Can be used in split and splitless modes. Gas Saver time is set to activate during the run after the sample is

    on the column.

    Split vent

    flow(mL/min)

    200 -175 -150 -

    100 -75 -50 -25 -

    -2 -1 0 1 2 3 4 5 6 7 8 9 10

    Prep Run Time

    Gas Saver

    Flow

    StartGas saver time set to 2.5 min.

    Run ends

    Gas saver flow

    RegularFlow

  • 7/29/2019 2-Inlets.pdf

    16/24

    16

    Splitless Flow Diagram

    Purge valve closed during injection (no split vent flow). At a specified time after injection, the purge valve opens to sweep remaining

    vapors out of inlet.

  • 7/29/2019 2-Inlets.pdf

    17/24

    17

    Split/Splitless Capillary Inlet Exploded View

    Insert Assembly

    Liner

    Insulation

    Washer

    Reducing Nut

    Insulation

    Ferrule

    Column Nut

    Septum retainernut

    Septum

    O-ring

    Split vent line

    Split/splitlessinlet body

    Retaining nut

    Inlet base seal

    Insulation cup

  • 7/29/2019 2-Inlets.pdf

    18/24

    18

    Ferrule:

    GRAPHITE: Good seal easilyVESPEL: Excellent seal - hard

    GRAPHITE/VESPEL: Very good seal

    with some difficulty

    Column Nut: 5181 - 8830

    5020 - 8292 Finger-tight

    5020 - 8293 530 um Finger-tight

    19258-20920 spring loaded -Vespel

    Seal: 18740 - 20885, new with 2 slits 5182-9652

    Septum: 5182 - 0739

    High-temperature

    11 mm DIAMETER

    Liner Seal:

    Viton: 5180 - 4182

    Graphite: 5180 - 41685180 - 4173

    BELOW 300 C

    SPLIT LINER ABOVE 300 CSPLITLESS LINER ABOVE 300 C

    3 - 6 mm

    APPLICATION

    DEPENDENT

    GOLD PLATED, STAINLESS STEEL MUST BE KEPTCLEAN CAN BE DEACTIVATED

    Washer obsolete

    CHANGE WITH LINER

    Capillary Injection Port Consumables

  • 7/29/2019 2-Inlets.pdf

    19/24

    19

    Most Commonly Used Liners

    #See the HP Analytical Columns and Supplies Catalog for the complete list of liners.

    4-mm id 1000-ul nominalvolume, borosilicateglass (untreated)

    silanized glass wool plug

    Split and splitless modes,especially recommended forrapid injections with the 7673

    Automatic Sampler

    19251-60540Split/Splitless

    Liners: Untreated

    Configuration Recommended Use Part No.Type

    4-mm id, nominal volume900 ul, boroscilicate

    glass (deactivated),silanized glass wool plug(deactivated)

    Split and splitlessmodes.

    Can be self-packed.

    5062-3587Packed, tapered

    Liners: Deactivated

    4-mm id, nominalvolume 900 ul,boroscilicate glass(deactivated)

    5181-3316Unpacked, tapered Split and splitlessmodes.Can be self-packed.

    !!New Deactivated Split Liner!! 5183-4647

  • 7/29/2019 2-Inlets.pdf

    20/24

    20

    Capillary Liner Maintenance

    General Purpose, Non-Deactivated LinersCLEANING PROCEDURE:1. Remove glass wool

    2. Sonicate in solvent or acid bath3. Rinse & dry4. Replace deactivated glass wool5. Deactivate Silylate*

    6. Dry*Silanizing agents available from column suppliers.

    Factory Deactivated (Pesticide Liners)CLEANING PROCEDURE:1. Remove glass wool2. Sonicate in solvent(s)3. Dry4. Care should be taken to avoid scratching the inner surface of the liner.

    #TREAT AS ULTRA CLEAN PART

  • 7/29/2019 2-Inlets.pdf

    21/24

    21

    Injection Port Seal: Maintenance

    Sonicate in Solvent Bake Dry

    Deactivate with Non-Chlorinated Reagent HMDS BSTFA BSA

    TSIM Wash with Solvents

    Inert First Wash - Toluene Alcohol Second Wash - Methanol

    Bake Dry

    #TREAT AS ULTRA CLEAN PART

  • 7/29/2019 2-Inlets.pdf

    22/24

    22

    Injection Port Problems

    Non-Representative sample Incorrect Temperature

    Blockage Contamination Leaks

    Detecting Leaks Snoop Electronic Leak Detector Pressure test Chromatographic Changes (compare shifts in retention time

    and area counts)

  • 7/29/2019 2-Inlets.pdf

    23/24

    23

    Injection Port Maintenance

    Change the septum regularly.

    Use the lowest practical temperature.

    Purge with flow. Use clean liners.

    Solvent clean the seal.

    Use clean syringes.

  • 7/29/2019 2-Inlets.pdf

    24/24

    24

    Other Inlets

    Programmed Temperature Vaporization Inlet:

    Large Volume Injection (LVI) with PTV for trace analysis of

    later eluting solutes and dirty samples. Multiple injections are made in a cool inlet and then the

    temperature is ramped.

    Inlet uses Septum-less head.

    Volatiles Interface:

    Introduce a gas sample from an external device such as the

    headspace, purge and trap, or air toxics samplers.