2-Inlets.pdf
Transcript of 2-Inlets.pdf
-
7/29/2019 2-Inlets.pdf
1/24
Inlets
-
7/29/2019 2-Inlets.pdf
2/24
2
In This Section, You Will Learn:
Identifying the Types of Inlets
Identifying the Components in the Packed and Split Inlets
Understanding the Flow Dynamics of the Split Inlet
Calculating Split Ratio
Identifying preventative Maintenance and Troubleshooting
-
7/29/2019 2-Inlets.pdf
3/24
3
Inlet Systems
Purpose: To allow the insertion of a sample into the gas
chromatograph in a repeatable, reproducible manner.
The sample should be representative of the bulk, and
unless specifically desired should be inserted without
chemical change.
EPC onlyVolatiles Interface
EPC onlyProgrammed Temperature Vaporization
EPC onlyCool on-columnEPC and non-EPCPurged Pack
EPC and non-EPCSplit/Splitless
Gas ControlInlet Type
-
7/29/2019 2-Inlets.pdf
4/244
Flow System: Purged Packed Inlet
Flow controlled mode (recommended for packed columns)
Pressure controlled mode (recommended for capillary columns)
-
7/29/2019 2-Inlets.pdf
5/245
Exploded View of Purged Packed Inlet
Insulation
Insulation cup
Ferrule
Column nut
Insulation
Septum nut
Knurled nutHeater/sensor assembly
SeptumO-ring
PurgedPackedInlet body
Insulation
-
7/29/2019 2-Inlets.pdf
6/246
Liners/Adapters
Inlet Fitting
1/4-Inch Ferrule
1/4-Inch Nut
1/8-Inch Liner
Front Ferrule
Back Ferrule
1/8-Inch Nut
1/8-InchColumn
Using Two Wrenches in Opposition
to Tighten Column Fittings
1/4-Inch MetalFerrule
Back Ferrule
1/4-Inch Nut
1/4-Inch Column
Inlet Fitting
1/4-Inch Metal Column, 1/8-Inch Metal Column,
Packed InletPacked Inlet
There are a variety of liners
and adapters for different
columns, inlets and detectors.
-
7/29/2019 2-Inlets.pdf
7/247
Split/Splitless Inlet
Split Mode forPulsed Split
Splitless Mode forPulsed Splitless
Major Component AnalysesAllows for larger injection volume.
Trace Component AnalysesAllows for larger injection volume
-
7/29/2019 2-Inlets.pdf
8/24
8
Split/Splitless Flow System
-
7/29/2019 2-Inlets.pdf
9/24
9
Split Flow Diagram: Pre-Injection
SEAL
(INLET)
(SEPTUM)
SPLIT VENT
TOTAL FLOW
COLUMN
SEPTUM
CARRIER GAS
LINER
50 mL/min
2 mL/min
0.6 mL/min
48 mL/min
47.4 mL/min
PURGE VENT
PURGE VALVE
-
7/29/2019 2-Inlets.pdf
10/24
10
Split Flow Diagram: Sample Injection
(INLET)
(SEPTUM)
SPLIT VENT
TOTAL FLOW
COLUMN
= CARRIER GAS
PURGE VALVE
PURGE VENT
SPLIT RATIO = COLUMN + SPLIT VENT FLOW
COLUMN FLOW
= LIQUID SAMPLE MOLECULES
-
7/29/2019 2-Inlets.pdf
11/24
11
Split Flow Diagram: Sample Vaporization
COLUMN
(INLET)
(SEPTUM)
SPLIT VENT
TOTAL FLOW
= CARRIER GAS
= SAMPLE MOLECULES
= SOLVENT MOLECULES
PURGE VENT
PURGE VALVE
Vaporization takes place whensufficient thermal energy istransferred to the sample. Thiscan occur in the carrier gas, but
is more likely upon contact witha solid surface, such as the lineror mixing medium.
-
7/29/2019 2-Inlets.pdf
12/24
12
Split Flow Diagram: Sample/Carrier Mixing
(INLET) PURGE VALVE
(SEPTUM)PURGE VENT
SPLIT VENT
TOTAL
FLOW
COLUMN
= CARRIER GAS
= SAMPLE MOLECULES
= SOLVENT MOLECULES
For the concept of SPLIT RATIO tobe valid, the sample (solvent +analyte) must be mixed with thecarrier gas to give a homogeneousmixture.
At the bottom of the injection port a
small part of this mixture will transferto the column, while the bulk of themixture will leave the chromatographvia the SPLIT VENT.
-
7/29/2019 2-Inlets.pdf
13/24
13
Split Flow Diagram: Liner Overload
(INLET)
PURGEVALVE
(SEPTUM)
PURGE VENT
SPLIT VENT
TOTAL FLOW
COLUMN
= CARRIER GAS= SAMPLE MOLECULES
= SOLVENT MOLECULES
A large injection of a solvent with alarge expansion volume can causean overload of the injection portliner.
This can result in loss of sample out
the PURGE VENT as well ascontamination of the in-comingcarrier gas line.
-
7/29/2019 2-Inlets.pdf
14/24
14
Split Ratio Calculations
Split Ratio = Split Vent Flow + Column Flow
Column Flow
Split Vent Flow = (Split Ratio)(Column Flow) - Column Flow
33:1952.03.0100530
98:1972.01.0100320
196:197.52.00.5100200
SplitRatioSplit Flow
Septum
Purge
Column
FlowTotal Flow
Column
I.D. (m)
DCBA
-
7/29/2019 2-Inlets.pdf
15/24
15
Gas Saver
Gas Saver reduces the carrier gas flow from the split vent after thesample is on the column.
Column head pressure and the column flow rate are maintained,while the purge/split vent flows decreases. Can be used in split and splitless modes. Gas Saver time is set to activate during the run after the sample is
on the column.
Split vent
flow(mL/min)
200 -175 -150 -
100 -75 -50 -25 -
-2 -1 0 1 2 3 4 5 6 7 8 9 10
Prep Run Time
Gas Saver
Flow
StartGas saver time set to 2.5 min.
Run ends
Gas saver flow
RegularFlow
-
7/29/2019 2-Inlets.pdf
16/24
16
Splitless Flow Diagram
Purge valve closed during injection (no split vent flow). At a specified time after injection, the purge valve opens to sweep remaining
vapors out of inlet.
-
7/29/2019 2-Inlets.pdf
17/24
17
Split/Splitless Capillary Inlet Exploded View
Insert Assembly
Liner
Insulation
Washer
Reducing Nut
Insulation
Ferrule
Column Nut
Septum retainernut
Septum
O-ring
Split vent line
Split/splitlessinlet body
Retaining nut
Inlet base seal
Insulation cup
-
7/29/2019 2-Inlets.pdf
18/24
18
Ferrule:
GRAPHITE: Good seal easilyVESPEL: Excellent seal - hard
GRAPHITE/VESPEL: Very good seal
with some difficulty
Column Nut: 5181 - 8830
5020 - 8292 Finger-tight
5020 - 8293 530 um Finger-tight
19258-20920 spring loaded -Vespel
Seal: 18740 - 20885, new with 2 slits 5182-9652
Septum: 5182 - 0739
High-temperature
11 mm DIAMETER
Liner Seal:
Viton: 5180 - 4182
Graphite: 5180 - 41685180 - 4173
BELOW 300 C
SPLIT LINER ABOVE 300 CSPLITLESS LINER ABOVE 300 C
3 - 6 mm
APPLICATION
DEPENDENT
GOLD PLATED, STAINLESS STEEL MUST BE KEPTCLEAN CAN BE DEACTIVATED
Washer obsolete
CHANGE WITH LINER
Capillary Injection Port Consumables
-
7/29/2019 2-Inlets.pdf
19/24
19
Most Commonly Used Liners
#See the HP Analytical Columns and Supplies Catalog for the complete list of liners.
4-mm id 1000-ul nominalvolume, borosilicateglass (untreated)
silanized glass wool plug
Split and splitless modes,especially recommended forrapid injections with the 7673
Automatic Sampler
19251-60540Split/Splitless
Liners: Untreated
Configuration Recommended Use Part No.Type
4-mm id, nominal volume900 ul, boroscilicate
glass (deactivated),silanized glass wool plug(deactivated)
Split and splitlessmodes.
Can be self-packed.
5062-3587Packed, tapered
Liners: Deactivated
4-mm id, nominalvolume 900 ul,boroscilicate glass(deactivated)
5181-3316Unpacked, tapered Split and splitlessmodes.Can be self-packed.
!!New Deactivated Split Liner!! 5183-4647
-
7/29/2019 2-Inlets.pdf
20/24
20
Capillary Liner Maintenance
General Purpose, Non-Deactivated LinersCLEANING PROCEDURE:1. Remove glass wool
2. Sonicate in solvent or acid bath3. Rinse & dry4. Replace deactivated glass wool5. Deactivate Silylate*
6. Dry*Silanizing agents available from column suppliers.
Factory Deactivated (Pesticide Liners)CLEANING PROCEDURE:1. Remove glass wool2. Sonicate in solvent(s)3. Dry4. Care should be taken to avoid scratching the inner surface of the liner.
#TREAT AS ULTRA CLEAN PART
-
7/29/2019 2-Inlets.pdf
21/24
21
Injection Port Seal: Maintenance
Sonicate in Solvent Bake Dry
Deactivate with Non-Chlorinated Reagent HMDS BSTFA BSA
TSIM Wash with Solvents
Inert First Wash - Toluene Alcohol Second Wash - Methanol
Bake Dry
#TREAT AS ULTRA CLEAN PART
-
7/29/2019 2-Inlets.pdf
22/24
22
Injection Port Problems
Non-Representative sample Incorrect Temperature
Blockage Contamination Leaks
Detecting Leaks Snoop Electronic Leak Detector Pressure test Chromatographic Changes (compare shifts in retention time
and area counts)
-
7/29/2019 2-Inlets.pdf
23/24
23
Injection Port Maintenance
Change the septum regularly.
Use the lowest practical temperature.
Purge with flow. Use clean liners.
Solvent clean the seal.
Use clean syringes.
-
7/29/2019 2-Inlets.pdf
24/24
24
Other Inlets
Programmed Temperature Vaporization Inlet:
Large Volume Injection (LVI) with PTV for trace analysis of
later eluting solutes and dirty samples. Multiple injections are made in a cool inlet and then the
temperature is ramped.
Inlet uses Septum-less head.
Volatiles Interface:
Introduce a gas sample from an external device such as the
headspace, purge and trap, or air toxics samplers.