11999_2010_Article_1383

download 11999_2010_Article_1383

of 5

Transcript of 11999_2010_Article_1383

  • 7/30/2019 11999_2010_Article_1383

    1/5

    S Y M P O S I U M : C O M P L I C A T I O N S O F H I P A R T H R O P L A S T Y

    Reduced Articular Surface of One-piece Cups

    A Cause of Runaway Wear and Early Failure

    William L. Griffin MD, Christopher J. Nanson MD,

    Bryan D. Springer MD, Matthew A. Davies PhD,

    Thomas K. Fehring MD

    Published online: 18 May 2010

    The Association of Bone and Joint Surgeons1 2010

    Abstract

    Background Despite the clinical success of modernmetal-on-metal articulations, concerns with wear-related

    release of metal ions persist. Evidence suggests metal ion

    release is related to the effective coverage of the head in

    the metal shell (the cups functional articular arc). A recent

    study suggests a reduced functional articular arc is asso-

    ciated with increased ion release and the arc is a function of

    component design, size, and the abduction angle.

    Questions/purposes The purposes of this study were to

    (1) measure the functional articular arc in different sizes

    of currently available one-piece metal shells from several

    different manufacturers; and (2) compare the functional

    articular arc of these one-piece metal shells with the

    1808 arc of conventional hip arthroplasty acetabular

    components.

    Methods We calculated the available articular surface arc

    for 33 one-piece metal cups using measurements of cupdepth and internal cup radius.

    Results The arc of the articular surface varied among

    manufacturers and generally decreased with decreasing

    shell diameter. The mean functional articular arc was

    160.5 3.6 (range, 151.8165.8), which was less than

    the 180 arc of a conventional acetabular component.

    Conclusions Our data show certain cup designs are at

    higher risk for failure as a result of the decreased articular

    surface arc. This, along with analysis of abduction angles,

    supports the recent findings of bearing failure with verti-

    cally placed implants. Care must be taken when implanting

    these shells to ensure they are placed in less abduction to

    avoid edge loading and the potential for early bearing

    failure.

    Introduction

    Although some of the earliest THA designs incorporated

    metal-on-metal bearings, poor understanding of the man-

    ufacturing and tribologic needs of these bearings led to

    high implant failure rates and, as a result, metal-on-poly-

    ethylene became the dominant articular couple for several

    decades. Improved metal implant manufacturing tech-

    niques with better tolerance, roundness, and metallurgy,

    however, have led to recent increases in the use of metal-

    on-metal bearings. Metal-on-metal articulations are asso-

    ciated with a very low wear rate and as a result of the

    tribology of the materials, unlike metal-on-polyethylene

    articulations, the wear rate improves as the head size

    increases [6, 7, 9, 14]. This combination of low wear and

    large head sizes with metal-on-metal bearings has promp-

    ted the resurgence of hip resurfacing arthroplasty [3, 18].

    One or more of the authors (WG, BS, TF) received funding from

    research endowments from The Winkler Fund and The Smith

    Arthritis Fund.

    This work was performed at The OrthoCarolina Hip and Knee Center

    in conjunction with The OrthoCarolina Research Institute and the

    Department of Mechanical Engineering, University of North

    CarolinaCharlotte, Charlotte, NC, USA.

    W. L. Griffin, C. J. Nanson, B. D. Springer, T. K. Fehring

    The OrthoCarolina Hip and Knee Center, The OrthoCarolina

    Research Institute, Charlotte, NC, USA

    M. A. Davies

    Department of Mechanical Engineering, University of North

    Carolina, Charlotte, Charlotte, NC, USA

    W. L. Griffin (&)

    1915 Randolph Road, Charlotte, NC 28207, USA

    e-mail: [email protected]

    123

    Clin Orthop Relat Res (2010) 468:23282332

    DOI 10.1007/s11999-010-1383-8

  • 7/30/2019 11999_2010_Article_1383

    2/5

    Moreover, as a result of the additional stability afforded by

    large femoral heads against dislocation, use of one-piece

    metal shells with large-diameter heads are now being used

    routinely in THA.

    Despite the benefits and clinical success of modern

    metal-on-metal articulations, concerns with wear-related

    release of metal ions persist. Elevated chromium and cobalt

    levels have been found in both the serum and remoteorgans of patients with metal-on-metal bearings [1, 2, 10].

    Studies have also demonstrated excessive wear with one-

    piece metal shells in women with smaller diameter com-

    ponents as well as in hips in which the shell is oriented

    more vertically in the pelvis (a higher abduction angle)

    [5, 11, 13, 16, 19]. Occasionally, more dramatic local

    reactions to metal-on-metal articulations occur with a wide

    spectrum of presentations; excessive metal wear can lead to

    asymptomatic effusions, pseudotumors, and wide zones of

    soft tissue necrosis with loss of abductors and bone death

    [8, 17].

    Evidence has emerged that such elevated metal ionrelease with resurfacing hip arthroplasty is related to the

    effective coverage of the head in the metal shell or the

    cups functional articular arc. De Haan et al. [5] showed

    elevated metal ion release correlates with a reduced func-

    tional articular arc, in which the arc is a function of

    component design, component size, and the abduction

    angle of the cup. They suggested high wear is likely to

    occur in metal-on-metal resurfacing arthroplasties if the

    acetabular components are implanted steeply, likely as a

    result of a greater risk of edge loading, and this is partic-

    ularly problematic in patients with a small acetabulum and

    in low-profile acetabular components. The authors stressed

    surgeons needed to be aware of functional arc of the ace-

    tabular component they implant.

    Although De Haan et al. [5] have demonstrated the

    importance of functional articular arc in the performance of

    hip replacement components, it is still not a widely known

    concept nor are the data on functional articular arcs readily

    available from manufacturers. Moreover, as longer fol-

    lowup becomes available for one-piece metal cups, it is

    evident some patients develop metal sensitivity reactions

    attributable to increased metal wear associated with

    variables such as tolerances, material, roundness, as well

    as mechanical factors related to cup design. Because of

    the correlation with elevated wear and potential higher risk

    of failure, knowledge of how these cup characteristics

    differ among manufacturers as well as how they differ

    from conventional 1808 components is important for hip

    replacement surgeons.

    The purposes of this study were therefore (1) to measurethe functional articular arc in different sizes of currently

    available one-piece metal shells from several different

    manufacturers; and (2) to compare the functional articular

    arc of these one-piece metal shells with the 1808 arc of

    conventional hip arthroplasty acetabular components.

    Materials and Methods

    New one-piece metal-on-metal acetabular components

    were obtained from four different manufacturers in multiple

    sizes (Table 1). Eleven cups with sizes ranging from 44 to66 mm were obtained both from Smith and Nephew (Bir-

    mingham Hip Resurfacing; Smith and Nephew, Memphis,

    TN) and BIOMET (Magnum; BIOMET, Warsaw, IN). Six

    cups ranging in size from 42 to 62 mm were obtained from

    Wright Medical Technology Inc (Conserve Plus; Wright

    Medical Technology Inc, Arlington, TN), and three cups

    were obtained from Stryker Orthopaedics (Cormet; Stryker

    Orthopaedics, Mahwah, NJ). Two additional cups that are

    currently on the market (ASR; DePuy, Warsaw, IN, and

    Durom; Zimmer, Warsaw, IN) were retrieved during

    revision arthroplasty and included in the analysis. In all,

    we studied 31 new and two revision-retrieved one-piece

    metal-on-metal acetabular components.

    For each component, we calculated the arc of the

    available articular surface (Fig. 1) with trigonometry using

    measurements of cup depth (measured with a Mitutoyo

    Series 129 depth micrometer; Mitutoyo America Corpo-

    ration, Aurora, IL) and the manufacturer-supplied internal

    radius of the cup. Cup depth was defined as the distance

    from the functional rim to the internal dome, taking into

    account any cutaways or bevels at the rim. All depth

    measurements were repeated 10 times per cup and the

    Table 1. Distribution of cups studied

    Manufacturer Cup Number of cups Sizes (mm)

    Smith and Nephew Birmingham Head Resurfacing (new) 11 44, 46, 48, 50,52, 54, 56, 60, 62, 64, 66

    BIOMET Magnum (new) 11 44, 46, 48, 50,52, 54, 56, 60, 62, 64, 66

    Wright Medical Technology Inc. Conserve Plus (new) 6 42, 46, 48, 52, 56, 62

    Stryker Orthopaedics Cormet (new) 3 52, 56, 62

    DePuy ASR (retrieved) 1 48

    Zimmer Durom (retrieved) 1 44

    Volume 468, Number 9, September 2010 Reduced Articular Surface of One-piece Cups 2329

    123

  • 7/30/2019 11999_2010_Article_1383

    3/5

    average value recorded and used for articular surface

    calculations.

    Head size versus measured articular surface arcs was

    graphed for each of the manufacturers. For parametric data,

    we used a one-sample t-test to compare the mean of a

    population to a known number. Because values for articular

    surface arc assessed in this study were normally distrib-

    uted, we used a one-sample t-test to compare the mean

    value for the 33 articular surface arcs in this study to a

    fixed value of 1808, the theoretical arc for a conventional

    acetabular implant. Owing to the small sample size, we did

    not compare differences among manufacturers. Instead, we

    assessed whether, as a group, these one piece metal cups

    have an articular surface arc that was different from 1808.

    We performed the statistical analysis using SPSS software

    (SPSS Version 8.0; SPSS Inc, Chicago, IL).

    Results

    The arc of the articular surface varied among manufac-

    turers and generally decreased with decreasing shell

    diameter (Fig. 2). The amount of articular surface coverage

    ranged from a high of 165 in the largest sizes of two of the

    manufacturers designs to 151.8 encountered in the smaller

    retrieval specimen. In general, the Conserve cup had the

    largest functional arcs across cup sizes, whereas the BHR

    and the Cormet had the smallest functional arcs across the

    range of cup sizes.

    The mean functional articular arc for the 33 cups as-

    sessed in this study was 160.5 3.6 (range, 151.8

    165.8), which was less than (p\ 0.001) the 180 arc of a

    conventional acetabular component.

    Discussion

    With the increase in the number of resurfacing and THAs

    performed with large, nonhemispheric one-piece metal

    cups, new complications associated with wear have been

    noted. These complications include runaway wear, pseu-

    dotumor formation, and acute lymphocytic vascular-

    associated lesions [8, 16, 17]. Several reports have recently

    been published regarding acetabular component position as

    it relates to serum metal ions [5, 11, 13] and cup loosening

    [4, 15]. Cups implanted with inclinations ranging from 45

    to 60 showed signs of increased edge loading, loosening,

    and serum metal ion levels. Additionally, De Haan et al.

    have demonstrated the importance of functional articular

    arc as it relates to wear complications with large metal-on-

    metal articulation [5]. De Haan et al. showed elevated

    Fig. 1 The functional articular arc (a) is a function of radius (r) and

    depth of the cup (d). The articular arc angle is a measure that is

    design-specific. The amount of coverage laterally over the head is a

    function of the abduction angle of the cup and the functional articular

    surface. Reproduced (in amended form) with permission and copy-

    right of the British Editorial Society of Bone and Joint Surgery (De

    Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K.

    Correlation between inclination of the acetabular component and

    metal ion levels in metal-on-metal hip resurfacing replacement. J

    Bone Joint Surg Br. 2008;90:12911297, Fig. 2).

    Fig. 2 Head size (mm) versus functional articular arc (degrees) is

    graphed for different manufacturer shell designs assessed in this

    study.

    2330 Griffin et al. Clinical Orthopaedics and Related Research1

    123

  • 7/30/2019 11999_2010_Article_1383

    4/5

    metal ion release correlates with a reduced functional

    articular arc, in which the arc is a function of component

    design, component size, and the abduction angle of the cup.

    The purpose of the current study, therefore, was to build on

    the work of De Haan et al. [5] and (1) to measure the

    functional articular arc in different sizes of currently

    available one-piece metal shells from several different

    manufacturers; and (2) to compare the functional articulararc of these one-piece metal shells with the 1808 arc of

    conventional hip arthroplasty acetabular components.

    We acknowledge several limitations of the current

    study. The primary limitation is that articular surface wear

    is a multifactorial process and there are many variables that

    affect it which we do not address, including bearing

    diameter and cup anteversion. The focus of this study,

    however, is functional articular arc. All else being equal,

    functional articular arc matters [5] and surgeons should be

    more aware of it. Quantification of functional articular arcs

    for cups presently on the market may help surgeons

    improve their outcomes by enabling them to understandwhich shell designs are at increased risk for edge loading,

    specifically if they are positioned suboptimally. Second, we

    did not evaluate the full range of implants for each man-

    ufacturer. Although we would have preferred to evaluate

    many more cups, we were only able to study those cups

    donated by the different manufacturers. However, all cups

    donated were evaluated and included in the present anal-

    ysis. Future analyses should include more cups from

    different manufacturers with a complete range of cup sizes.

    We observed a wide variation of functional arcs among

    currently available manufacturers and cup sizes in which

    functional arc generally decreased as shell diameter

    decreased. In general, the Conserve cup (Wright Medical)

    had the largest functional arcs, whereas the BHR (Smith

    and Nephew) and the Cormet (Stryker) had the smallest

    functional arcs across the range of cup sizes. Of all the cups

    studied, the 44-mm ASR (DePuy) had the smallest func-

    tional articular arc. The reason for the variation among

    manufacturers is related to design. Design decisions that

    compromise hemisphericity are based on many factors,

    including retention of bone stock and reducing the risk of

    impingement in a resurfacing situation. Some cups are

    designed to be less than a hemisphere to prevent

    impingement of the native femoral neck against the shell

    edge. Some designs reduce hemisphericity by incorporat-

    ing a thickened dome to stiffen the shell and lessen the

    deformation that can occur during press-fit implantation.

    Still other designs incorporate a tapered radius at their rim

    to eliminate sharp edges and maintain fluid ingress into the

    articulation. Overall, these design differences result in

    variability in functional articular arcs. This difference

    between designs is clearly demonstrated by a recent clini-

    cal report. In a single center study of 660 metal-on-metal

    resurfacings, Langton et al. [12] reported 17 failures from

    adverse reactions to metal debris in patients implanted with

    ASR implants (3.5%) and no failures of this nature in

    patients implanted with BHR (Smith and Nephew)

    implants. Langton et al. suggest the increased failure of the

    ASR cup (DePuy) secondary to the increased generation of

    metal debris was the result of its reduced functional artic-

    ular arc as compared with that of the BHR (DePuy)component [12].

    The articular surface arcs of cups in the current study

    averaged 160.6, which was substantially less than the 180

    arcs associated with standard THA components. The

    smaller bearing surface arc demonstrated with these and

    other one-piece metal cups results in a smaller amount of

    coverage laterally over the head, which could lead to edge

    loading at much lower abduction angles, resulting in

    elimination of fluid film lubrication and increased metal

    wear. This is consistent with the reports of edge loading

    and component loosening as well as increased metal ions

    noted in subjects with inclination angles averaging 45 ormore [5, 11, 13, 15]. As previously noted, the amount of

    arc available for coverage is a function of the abduction

    angle of the cup, the functional arc of the particular cup

    design, and the size of the implanted cup. Cups with a

    smaller functional articular surface are at higher risk of

    edge loading and high wear rates. As an example, if a cup

    with a functional articular surface of 151 is implanted at

    55 abduction, it will behave like a 180 cup implanted at

    69.5 of abduction and be at risk for edge loading. A tra-

    ditionally acceptable 45 inclination angle leaves no room

    for error in these nonhemispheric cups, particularly in

    smaller sizes. This supports the findings of Ollivere et al.

    who studied the rate and mode of early failure in 463

    Birmingham hip resurfacings [16]. They reported a 3.1%

    rate of metallosis-related revision at 5 years with risk

    factors for revision including female gender and a high

    abduction angle. When combined with excessive antever-

    sion, these one-piece cups provide even less coverage of

    the weightbearing portion of the femoral head. Given this

    possibility, and the increasing number of reports of run-

    away wear and loosening as well as pseudotumor formation

    and devastating soft tissue necrosis and nerve palsies, it

    may be advisable to implant nonhemispheric resurfacing

    cups between 408 to 458 of abduction and 158 of antever-

    sion and check for impingement intraoperatively. Because

    cup position error is not well tolerated with these designs,

    an intraoperative radiograph may be advisable to ensure

    accurate position.

    Our data and analysis of abduction angles are consistent

    with recent findings of bearing failure with vertically

    placed implants. It is evident certain designs are at higher

    risk for failure as a result of the decreased articular surface

    arc. Care must be taken when implanting these shells to

    Volume 468, Number 9, September 2010 Reduced Articular Surface of One-piece Cups 2331

    123

  • 7/30/2019 11999_2010_Article_1383

    5/5

    ensure they are placed in less abduction to avoid edge

    loading and the potential for early bearing failure.

    Acknowledgments We thank Christi Sychterz Terefenko, MS, for

    her assistance with the manuscript.

    References

    1. Bhamra M, Case C. Biological effects of metal-on-metal hip

    replacements. Proc Inst Mech Eng H. 2006;220:379384.

    2. Brown C, Fisher J, Ingham E. Biological effects of clinically

    relevant wear particles from metal-on-metal hip prostheses. Proc

    Inst Mech Eng H. 2006;220:355369.

    3. Buergi ML, Walter WL. Hip resurfacing arthroplasty: the

    Australian experience. J Arthroplasty. 2007;22(Suppl 3):6165.

    4. De Haan R, Campbell PA, Su E, De Smet K. Revision of metal-

    on-metal resurfacing arthroplasty of the hip. J Bone Joint Surg

    Br. 2008;90:11581163.

    5. De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De

    Smet K. Correlation between inclination of the acetabular com-

    ponent and metal ion levels in metal-on-metal hip resurfacing

    replacement. J Bone Joint Surg Br. 2008;90:12911297.

    6. Dowson D. Tribological principles in metal-on-metal hip joint

    design. Proc Inst Mech Eng H. 2006;220:161171.

    7. Dowson D, Jin Z-M. Metal-on-metal hip joint tribology. Proc

    Inst Mech Eng H. 2006;220:107118.

    8. Grammatopolous G, Pandit H, Kwon YM, Gundle R, McLardy-

    Smith P, Beard DJ, Murray DW, Gill HS. Hip resurfacings

    revised for inflammatory pseudotumour have a poor outcome.

    J Bone Joint Surg Br. 2009;91:10191024.

    9. Isaac G, Thompson J, Williams S, Fisher J. Metal-on-metal

    bearing surfaces: materials, manufacture, design, optimization,

    and alternatives. Proc Inst Mech Eng H. 2006;220:119133.

    10. Jacobs JJ, Skipor AK, Doorn PF, Campbell P, Schmalzried TP,

    Black J, Amstutz HC. Cobalt and chromium concentrations in

    patients with metal on metal total hip replacements. Clin Orthop

    Relat Res. 1996;329(Suppl):S256S263.

    11. Khan M, Kuiper J-H, Richardson J. The exercise related rise in

    plasma cobalt levels after metal-on-metal hip resurfacing

    arthroplasty. J Bone Joint Surg Br. 2008;90:11521157.

    12. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol

    AV. Early failure of metal-on-metal bearings in hip resurfacing

    and large-diameter total hip replacement: a consequence of

    excess wear. J Bone Joint Surg Br. 2010;92:3846.

    13. Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The

    effect of component size and orientation on the concentrations of

    metal ions after resurfacing arthroplasty of the hip. J Bone Joint

    Surg Br. 2008;90:11431151.

    14. Lee R, Essner A, Wang A. Tribological considerations in primary

    and revision metal-on-metal arthroplasty. J Bone Joint Surg Am.

    2008;90(Suppl 3):118124.

    15. Morlock M, Bishop N, Zustin J, Hahn M, Ruther W, Amling M.

    Modes of implant failure after hip resurfacing: morphological and

    wear analysis of 267 retrieval specimens. J Bone Joint Surg Am.

    2008;90(Suppl 3):8995.

    16. Ollivere B, Darrah C, Barker T, Nolan J, Porteous MJ. Early

    clinical failure of the Birmingham metal-on-metal hip resurfacing

    is associated with metallosis and soft-tissue necrosis. J Bone Joint

    Surg Br. 2009;91:10251030.

    17. Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell

    D, Gibbons C, Ostlere S, Athanasou N, Gill H, Murray D.

    Pseudotumors associated with metal-on-metal hip resurfacings.

    J Bone Joint Surg Br. 2008;90:847851.

    18. Schmalzreid TP. Why total hip resurfacing. J Arthroplasty. 2007;

    22(Suppl 3):5760.

    19. Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J.

    Tribology and wear of metal-on-metal hip prostheses: influence

    of cup angle and head position. J Bone Joint Surg Am. 2008;

    90(Suppl 3):111117.

    2332 Griffin et al. Clinical Orthopaedics and Related Research1

    123