1083ch6_3
-
Upload
scribdthakur -
Category
Documents
-
view
216 -
download
0
Transcript of 1083ch6_3
-
7/27/2019 1083ch6_3
1/3
823
6.3 Hydrometers
C. H. HOEPPNER (1982) B. G. LIPTK (1969, 1995, 2003)
Types: Laboratory or industrial, can be indicating or transmitting
Design Pressure: Generally atmospheric, but the glass variable area flowmeter designs can be used up
to 100 PSIG (6.9 bars)
Design Temperature: Generally to 200F (93C)
Materials of Construction: Laboratory units are made of glass and unbreakable plastics; industrial units are
available in similar materials as are variable area flow-meters (Section 2.27).
Ranges: Minimum span is 0.05 SG; maximum span is 0.5 SG. Spans can be selected within
the specific gravity range of 0.6 and 2.1. Minimum scale divisions are 0.0005 SG.
Inaccuracy: 1% of span
Cost: Laboratory units cost between $15 and $50. Industrial indicators start at around $500,
and transmitters at $2000 in standard materials. (Costs are higher for corrosion-
resistant materials.)
Partial List of Suppliers: ABB Inc. (www.abb.com);
Brooklyn Thermometer Co. (www.brooklynthermometer.com)
Cole-Parmer Instrument Co. (www.coleparmer.com)Elite Scientific Corp. (www.ambalayellowpages.com)
ERTCO (Ever Ready Thermometer Co.) (www.ertco.com/hydrometers)
H-B Instrument Co. (www.hbinstruments.com)
Princo Instruments Inc. (www.princoinstruments.com)
INTRODUCTION
According to Archimedes principle, when a body is immersed
in a fluid, its weight drops by the same amount as the weightof the liquid that it has displaced. Therefore, all hydrometers
will sink until they displace the same mass of liquid as their
own mass. The volume of liquid displaced is indicated by the
level on the scale and density is the ratio of the hydrometers
mass divided by the displaced volume. As the density of liquidschange with temperature, most hydrometers include a ther-
mometer for temperature compensation purposes.
DESIGN VARIATIONS
The hydrometer element consists of a weighted float with a
small-diameter indicator stem attachment at the top of the
float as shown on Figure 6.3a. The stem is graduated in any
of the density units discussed in Section 6.1. The hydrometer
has a constant-weight body, which, if immersed in fluids with
differing densities, will displace differing volumes of fluid.
Therefore, the degree of stem scale submersion is an indica-
tion of the fluids density. Readings are made at the point
where the stem emerges from the liquid. The accuracy of
DI
FO
Flow Sheet Symbol
FIG. 6.3a
Hand hydrometer.
Graduations
in Any Units
(Minimum Division
is 0.0005 SpG)
Weights
6 to 15 in.
(150 to 380 mm)
2003 by Bla Liptk
http://1083ch2_27.pdf/http://www.abb.com/http://www.brooklynthermometer.com/http://www.coleparmer.com/http://www.ambalayellowpages.com/http://www.ertco.com/hydrometershttp://www.hbinstruments.com/http://www.princoinstruments.com/http://1083ch6_1.pdf/http://www.princoinstruments.com/http://www.hbinstruments.com/http://www.ertco.com/hydrometershttp://www.ambalayellowpages.com/http://www.coleparmer.com/http://www.brooklynthermometer.com/http://www.abb.com/http://1083ch6_1.pdf/http://1083ch2_27.pdf/ -
7/27/2019 1083ch6_3
2/3
824 Density Measurement
measurement is a function of the combined effects of surface
tension, turbulence, and sample contamination.
In-Line Designs
One of the simplest in-line density indicators is illustrated inFigure 6.3b. It consists of a transparent glass tee with a hydro-
meter and a thermometer inside. Constant level is maintained
in this tee, because the process fluid sample enters from the
bottom and overflows. The sample flow rate is maintained at
less than l gal/m (3.78 l/m) to minimize the effects of velocity
and turbulence. If the process temperature varies, a thermom-
eter is added to allow for manual temperature compensation.
Industrial Designs
The hydrometer element can be mounted inside a variable-
area flowmeter (Section 2.27) housing. In such designs, as
shown in Figure 6.3c, overflow pipes are provided to maintain
the constant level inside the glass tube. Standard accessories
include needle valves for sample flow rate control at about
15 gal/h (57 l/h) and integral thermometers.
Wetted parts are available in the same range of materials
as for variable area flowmeters. These units can withstand up
to 200F (93C) and 100 PSIG (690 kPa). Specific gravity
(SG) spans of 0.1 to 0.5 are available and can be selected
within the limits of 0.6 and 2.1 SG. Reading inaccuracy is
1% of span or the smallest division on the scale, which can
range from 0.001 to 0.005 SG.
Transmitters
The variable immersion hydrometer element mounted in a
rotameter housing can also be obtained as a transmitter for
remote readout, as illustrated in Figure 6.3d three-dimensional.
This electronic transmitter can be the servo-operated imped-
ance bridge type and can be provided with automatic temper-
ature compensation. Such compensation will convert the actual
density, which is detected to the density that it would corre-
spond to at a predetermined base temperature. Ranges, accu-
racies, and design limitations are the same as given for the
local indicator version, discussed earlier.
The stem position of the hydrometer can also be detected
optically. In this design, as the stem rises and falls it changes
the amount of light that passes to a photocell. The photocell
output is calibrated in specific gravity units.
Another transmitting hydrometer design is the capaci-
tance type. Here, a stainless steel hydrometer positions a
dielectric cup inside two insulated, concentric cylinders.
The resulting change in capacitance is proportional to den-
sity. Automatic temperature compensation is provided with
this unit so that the transmitted output signal can be refer-
enced to 60F (16C). The transmission can either be analog
or digital.
FIG. 6.3b
In-line hydrometer indicator.
FIG. 6.3c
In-line hydrometer in rotameter housing.
Outlet
Glass
Tee
Mercury
Thermometer
Hydrometer
Screen Inlet
Vent or
Equalizer
Connection
OverflowHydrometer
Inlet Outlet
Thermometer
FIG. 6.3d
Hydrometer transmitter in rotameter housing.
Vent or
Equalizer
Connection
Indicating
Hydrometer
Transmitter
Coils
Temperature
Sensor
2003 by Bla Liptk
http://1083ch2_27.pdf/http://1083ch2_27.pdf/http://1083ch2_27.pdf/ -
7/27/2019 1083ch6_3
3/3
6.3 Hydrometers 825
CONCLUSIONS
The hydrometers discussed above are basic to density mea-
surement. They are accurate, frictionless, and direct-indicating
without the need for mechanical linkages or external energy
sources. They are compatible with most corrosive fluids.
Their limitations are also multiple, because the float posi-
tion should not be affected by anything except the fluid den-sity. Effects of the velocity, friction, turbulence, and viscosity
of the process must all be minimized. In addition, material
buildup on the float cannot be tolerated, because the basis of
hydrometer operation is the assumption that the weight of
the float is constant. For these reasons, hydrometers shouldonly be considered for use on clean, nonviscous process fluids
when the sample flow rate is controlled at around 1 gal/h.
Bibliography
Cameron, D., An Instrument for Measurement of Liquid Density,Indus-
trial Electronics, March 1967.
Capano, D., The Ways and Means of Density,InTech,November 2000. Density and Specific Gravity,Measurements and Control, October1991.
Gupta, S.V., Practical Density Measurement and Hydrometry, Bristol, U.K.:
Institute of Physics, 2002.
Honeywell Instrumentation Data Sheet #10.143a, Continuous Measure-
ment and Control of Density, Morristown, NJ, latest edition.
Magaris, P., On-Line Density Measurement Is Fast and Accurate,Control
Engineering, June 1981.
November, M.H., Measuring Fluid Density and Specific Weight,Instru-
mentation Insight, Vol. 2, No. 3, August 1975.
Standard Practice for Calibration of Transmission Densitometers,ASTM
Standard,1998.
2003 by Bla Liptk