1083ch6_3

download 1083ch6_3

of 3

Transcript of 1083ch6_3

  • 7/27/2019 1083ch6_3

    1/3

    823

    6.3 Hydrometers

    C. H. HOEPPNER (1982) B. G. LIPTK (1969, 1995, 2003)

    Types: Laboratory or industrial, can be indicating or transmitting

    Design Pressure: Generally atmospheric, but the glass variable area flowmeter designs can be used up

    to 100 PSIG (6.9 bars)

    Design Temperature: Generally to 200F (93C)

    Materials of Construction: Laboratory units are made of glass and unbreakable plastics; industrial units are

    available in similar materials as are variable area flow-meters (Section 2.27).

    Ranges: Minimum span is 0.05 SG; maximum span is 0.5 SG. Spans can be selected within

    the specific gravity range of 0.6 and 2.1. Minimum scale divisions are 0.0005 SG.

    Inaccuracy: 1% of span

    Cost: Laboratory units cost between $15 and $50. Industrial indicators start at around $500,

    and transmitters at $2000 in standard materials. (Costs are higher for corrosion-

    resistant materials.)

    Partial List of Suppliers: ABB Inc. (www.abb.com);

    Brooklyn Thermometer Co. (www.brooklynthermometer.com)

    Cole-Parmer Instrument Co. (www.coleparmer.com)Elite Scientific Corp. (www.ambalayellowpages.com)

    ERTCO (Ever Ready Thermometer Co.) (www.ertco.com/hydrometers)

    H-B Instrument Co. (www.hbinstruments.com)

    Princo Instruments Inc. (www.princoinstruments.com)

    INTRODUCTION

    According to Archimedes principle, when a body is immersed

    in a fluid, its weight drops by the same amount as the weightof the liquid that it has displaced. Therefore, all hydrometers

    will sink until they displace the same mass of liquid as their

    own mass. The volume of liquid displaced is indicated by the

    level on the scale and density is the ratio of the hydrometers

    mass divided by the displaced volume. As the density of liquidschange with temperature, most hydrometers include a ther-

    mometer for temperature compensation purposes.

    DESIGN VARIATIONS

    The hydrometer element consists of a weighted float with a

    small-diameter indicator stem attachment at the top of the

    float as shown on Figure 6.3a. The stem is graduated in any

    of the density units discussed in Section 6.1. The hydrometer

    has a constant-weight body, which, if immersed in fluids with

    differing densities, will displace differing volumes of fluid.

    Therefore, the degree of stem scale submersion is an indica-

    tion of the fluids density. Readings are made at the point

    where the stem emerges from the liquid. The accuracy of

    DI

    FO

    Flow Sheet Symbol

    FIG. 6.3a

    Hand hydrometer.

    Graduations

    in Any Units

    (Minimum Division

    is 0.0005 SpG)

    Weights

    6 to 15 in.

    (150 to 380 mm)

    2003 by Bla Liptk

    http://1083ch2_27.pdf/http://www.abb.com/http://www.brooklynthermometer.com/http://www.coleparmer.com/http://www.ambalayellowpages.com/http://www.ertco.com/hydrometershttp://www.hbinstruments.com/http://www.princoinstruments.com/http://1083ch6_1.pdf/http://www.princoinstruments.com/http://www.hbinstruments.com/http://www.ertco.com/hydrometershttp://www.ambalayellowpages.com/http://www.coleparmer.com/http://www.brooklynthermometer.com/http://www.abb.com/http://1083ch6_1.pdf/http://1083ch2_27.pdf/
  • 7/27/2019 1083ch6_3

    2/3

    824 Density Measurement

    measurement is a function of the combined effects of surface

    tension, turbulence, and sample contamination.

    In-Line Designs

    One of the simplest in-line density indicators is illustrated inFigure 6.3b. It consists of a transparent glass tee with a hydro-

    meter and a thermometer inside. Constant level is maintained

    in this tee, because the process fluid sample enters from the

    bottom and overflows. The sample flow rate is maintained at

    less than l gal/m (3.78 l/m) to minimize the effects of velocity

    and turbulence. If the process temperature varies, a thermom-

    eter is added to allow for manual temperature compensation.

    Industrial Designs

    The hydrometer element can be mounted inside a variable-

    area flowmeter (Section 2.27) housing. In such designs, as

    shown in Figure 6.3c, overflow pipes are provided to maintain

    the constant level inside the glass tube. Standard accessories

    include needle valves for sample flow rate control at about

    15 gal/h (57 l/h) and integral thermometers.

    Wetted parts are available in the same range of materials

    as for variable area flowmeters. These units can withstand up

    to 200F (93C) and 100 PSIG (690 kPa). Specific gravity

    (SG) spans of 0.1 to 0.5 are available and can be selected

    within the limits of 0.6 and 2.1 SG. Reading inaccuracy is

    1% of span or the smallest division on the scale, which can

    range from 0.001 to 0.005 SG.

    Transmitters

    The variable immersion hydrometer element mounted in a

    rotameter housing can also be obtained as a transmitter for

    remote readout, as illustrated in Figure 6.3d three-dimensional.

    This electronic transmitter can be the servo-operated imped-

    ance bridge type and can be provided with automatic temper-

    ature compensation. Such compensation will convert the actual

    density, which is detected to the density that it would corre-

    spond to at a predetermined base temperature. Ranges, accu-

    racies, and design limitations are the same as given for the

    local indicator version, discussed earlier.

    The stem position of the hydrometer can also be detected

    optically. In this design, as the stem rises and falls it changes

    the amount of light that passes to a photocell. The photocell

    output is calibrated in specific gravity units.

    Another transmitting hydrometer design is the capaci-

    tance type. Here, a stainless steel hydrometer positions a

    dielectric cup inside two insulated, concentric cylinders.

    The resulting change in capacitance is proportional to den-

    sity. Automatic temperature compensation is provided with

    this unit so that the transmitted output signal can be refer-

    enced to 60F (16C). The transmission can either be analog

    or digital.

    FIG. 6.3b

    In-line hydrometer indicator.

    FIG. 6.3c

    In-line hydrometer in rotameter housing.

    Outlet

    Glass

    Tee

    Mercury

    Thermometer

    Hydrometer

    Screen Inlet

    Vent or

    Equalizer

    Connection

    OverflowHydrometer

    Inlet Outlet

    Thermometer

    FIG. 6.3d

    Hydrometer transmitter in rotameter housing.

    Vent or

    Equalizer

    Connection

    Indicating

    Hydrometer

    Transmitter

    Coils

    Temperature

    Sensor

    2003 by Bla Liptk

    http://1083ch2_27.pdf/http://1083ch2_27.pdf/http://1083ch2_27.pdf/
  • 7/27/2019 1083ch6_3

    3/3

    6.3 Hydrometers 825

    CONCLUSIONS

    The hydrometers discussed above are basic to density mea-

    surement. They are accurate, frictionless, and direct-indicating

    without the need for mechanical linkages or external energy

    sources. They are compatible with most corrosive fluids.

    Their limitations are also multiple, because the float posi-

    tion should not be affected by anything except the fluid den-sity. Effects of the velocity, friction, turbulence, and viscosity

    of the process must all be minimized. In addition, material

    buildup on the float cannot be tolerated, because the basis of

    hydrometer operation is the assumption that the weight of

    the float is constant. For these reasons, hydrometers shouldonly be considered for use on clean, nonviscous process fluids

    when the sample flow rate is controlled at around 1 gal/h.

    Bibliography

    Cameron, D., An Instrument for Measurement of Liquid Density,Indus-

    trial Electronics, March 1967.

    Capano, D., The Ways and Means of Density,InTech,November 2000. Density and Specific Gravity,Measurements and Control, October1991.

    Gupta, S.V., Practical Density Measurement and Hydrometry, Bristol, U.K.:

    Institute of Physics, 2002.

    Honeywell Instrumentation Data Sheet #10.143a, Continuous Measure-

    ment and Control of Density, Morristown, NJ, latest edition.

    Magaris, P., On-Line Density Measurement Is Fast and Accurate,Control

    Engineering, June 1981.

    November, M.H., Measuring Fluid Density and Specific Weight,Instru-

    mentation Insight, Vol. 2, No. 3, August 1975.

    Standard Practice for Calibration of Transmission Densitometers,ASTM

    Standard,1998.

    2003 by Bla Liptk