10576 TN43183_E 11-13C-St2-High

download 10576 TN43183_E 11-13C-St2-High

of 3

Transcript of 10576 TN43183_E 11-13C-St2-High

  • 8/9/2019 10576 TN43183_E 11-13C-St2-High

    1/3

    Thermo Scientific iCAP 7000 Series ICP-OES

    Gas Control Systems

    TechnicalN

    ote4318

    3

    Key Words

    ICP gas control, Pressure control, Mass flow control

    Introduction

    The majority of commercial Inductively Coupled

    Plasma-Optical Emission Spectrometers (ICP-OES) use aplasma torch based on the original Fassell design,

    consisting of 3 glass tubes (Figure 1).

    The three concentric quartz glass tubes of the torch are

    precisely aligned to ensure that gas flows meet the

    required velocity in a uniform fashion so that the plasma

    formation is stable. Typical gas flows within the torch

    are:

    Coolant or plasma gas 12 l/min with some applications

    requiring slightly higher or lower flows

    Auxiliary or intermediate gas 0.5 l/min with high salt

    or organic applications requiring up to 2 l/min

    Nebulizer gas 0.5 l/min depending upon the type of

    nebulizer

    Figure 1. ICP-OES torch design

    It is critical that these gas flows remain constant in order

    to achieve optimum instrument performance, resulting inlong term stability of the analytical signal. As each gas

    flow serves a different purpose, changes in their flow rates

    affect both the plasma and signal differently. The

    nebulizer gas flow affects the analytical signal, with only

    small changes in the flow rate causing relatively large

    changes in the signal; this effect is commonly referred to

    as instrument drift. For example, a 0.1% change in the

    gas flow can result in a 1% change in the analytical signal.

    The overall shape of the plasma is affected by the coolant

    gas flow, at higher flows the plasma becomes elongated

    and has a larger annular space, which affects the axial

    channel, squeezing it and reducing its diameter. Stabilityof the coolant gas flow is particularly important for axial

    plasmas where the viewing zone may only be 1-2 mm

    in diameter.

    The auxiliary gas flow affects the relative position of the

    plasma, which is particularly important for radial

    plasmas. Small changes in the position of the plasma

    affect where the radial viewing channel passes through

    the plasma which alters the signal intensities differently

    for each emission line, as higher energy lines (UV) are

    more emitted more frequently in the centre of the plasma.

    Two types of gas flow control systems are available forthe Thermo ScientificiCAP7000 Series ICP-OES.

    Coolant Gas

    Auxillary Gas

    Nebuliser Gas

  • 8/9/2019 10576 TN43183_E 11-13C-St2-High

    2/3

    2 Pressure Control System

    A pressure control gas system operates using the

    relationship between gas pressure and the corresponding

    flow through a fixed orifice, as seen in figure 2.

    Figure 2. Schematic of pressure control system

    The high pressure input is first reduced with a precision

    regulator to the operating pressure of the system.

    Coolant and auxiliary gas flows are diverted to a series of

    fixed restrictors each controlled with a solenoid valve. In

    order to achieve the required flow, combinations of valves

    are opened to allow the gas to pass through different

    restrictors. The nebulizer gas is passed to a variable, high

    precision pressure regulator. The regulator can be

    adjusted to give the desired gas flow.

    Pressure control gas systems are very robust, reliable and

    simple to operate. However, the use of fixed restrictors

    can mean a limited number of steps for the gas flow.

    Also since gas density changes with temperature, if the

    temperature of the incoming gas varies greatly the gas

    flow rate will change, affecting the long term stability.

    Pressure control of the gas flows is a cost effective design

    and ideal for application where a limited number of

    methods with similar parameters are used. If different

    nebulizer flows are used with different methods then the

    flow will need to be manually adjusted each time the

    method is changed.

    Mass Flow Control System

    Mass flow controllers, as the name suggests, operate by

    controlling the actual mass flow of the gas. A typical

    mass flow controller consists of 3 main components: a

    control valve, a flow sensor and the electronic control

    circuitry as show in figure 3. The sensor consists of a

    long, thin capillary tube with temperature dependent

    resistors on the outside. When gas starts to flow the

    temperature profile becomes asymmetric, with the

    corresponding shift in resistance being proportional to

    the gas flow rate. This changing resistance is converted toa voltage and used to operate the gas flow control valve.

    Figure 3. Schematic of a typical mass flow controller

    The main benefit of mass flow controllers is that since

    they operate using electron sensing of the gas flow they

    are not affected by changes in atmospheric pressure or

    ambient temperature, giving better long term stability.

    In addition they offer rapid response times, typically lessthan 1 second, and full computer control. The gas flow

    settings are saved in the Thermo Scientific Qtegra

    software LabBooks and are automatically set when

    analysis is performed.

    Due to this automated computer control, full mass flow

    control is ideal for complex applications where multiple,

    varying analytical methods are required.

  • 8/9/2019 10576 TN43183_E 11-13C-St2-High

    3/3

    TechnicalN

    ote4318

    3

    Conclusions

    By utilizing mass flow control for the nebulizer gas flow,

    significant benefits can be gained in both long term

    analytical stability and ease of use. By adding mass flow

    controllers to the auxiliary and coolant gas flows,

    performance will be further improved when using axial or

    radial plasma viewing. Figure 4 displays the long term

    (over 16 hours) analytical stability achieved when using

    full mass flow control.

    Figure 4. Long term signal stability using full mass flow control

    Table 1. Recommended configurations

    * Thermo Scientific iCAP 7400 ICP-OES with additional MFC

    # Thermo Scientific iCAP 7600 ICP-OES for increased sample

    throughput with Sprint Valve as standard upgrade

    TN43183_E 11/13C

    Africa-Other +27 11 570 1840Australia +61 3 9757 4300

    Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)

    400 650 5118

    Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0

    Finland +358 9 3291 0200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

    Japan +81 45 453 9100Latin America +1 561 688 8700

    Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00Russia/CIS +43 1 333 50 34 0

    Singapore +65 6289 1190South Africa +27 11 570 1840

    Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

    www.thermoscientific.com

    2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.

    All oth er tradem arks ar e the prop erty of Thermo Fisher S cientifi c and its subsidiar ies. This informa tion is pr esented as an

    example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any

    manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.

    Not all products are available in all countries. Please consult your local s ales representative for details.

    Thermo Electron ManufacturingLtd (Cambridge) isISO 9001:2008 Certified.

    Model 7200 7400 7400* 7600 #

    Nebulizer Gas Pressure Control Mass Flow Control Mass Flow Control Mass Flow Control

    Auxiliary & Control Gas Pressure Control Pressure Control Mass Flow Control Mass Flow Control

    Routine analysis

    Similar basic methods

    Routine analysis

    Varying basic methods

    Routine analysis

    Similar advanced methods

    Research analysis

    Varying advanced methods