10576 TN43183_E 11-13C-St2-High
-
Upload
yan-rivera -
Category
Documents
-
view
214 -
download
0
Transcript of 10576 TN43183_E 11-13C-St2-High
-
8/9/2019 10576 TN43183_E 11-13C-St2-High
1/3
Thermo Scientific iCAP 7000 Series ICP-OES
Gas Control Systems
TechnicalN
ote4318
3
Key Words
ICP gas control, Pressure control, Mass flow control
Introduction
The majority of commercial Inductively Coupled
Plasma-Optical Emission Spectrometers (ICP-OES) use aplasma torch based on the original Fassell design,
consisting of 3 glass tubes (Figure 1).
The three concentric quartz glass tubes of the torch are
precisely aligned to ensure that gas flows meet the
required velocity in a uniform fashion so that the plasma
formation is stable. Typical gas flows within the torch
are:
Coolant or plasma gas 12 l/min with some applications
requiring slightly higher or lower flows
Auxiliary or intermediate gas 0.5 l/min with high salt
or organic applications requiring up to 2 l/min
Nebulizer gas 0.5 l/min depending upon the type of
nebulizer
Figure 1. ICP-OES torch design
It is critical that these gas flows remain constant in order
to achieve optimum instrument performance, resulting inlong term stability of the analytical signal. As each gas
flow serves a different purpose, changes in their flow rates
affect both the plasma and signal differently. The
nebulizer gas flow affects the analytical signal, with only
small changes in the flow rate causing relatively large
changes in the signal; this effect is commonly referred to
as instrument drift. For example, a 0.1% change in the
gas flow can result in a 1% change in the analytical signal.
The overall shape of the plasma is affected by the coolant
gas flow, at higher flows the plasma becomes elongated
and has a larger annular space, which affects the axial
channel, squeezing it and reducing its diameter. Stabilityof the coolant gas flow is particularly important for axial
plasmas where the viewing zone may only be 1-2 mm
in diameter.
The auxiliary gas flow affects the relative position of the
plasma, which is particularly important for radial
plasmas. Small changes in the position of the plasma
affect where the radial viewing channel passes through
the plasma which alters the signal intensities differently
for each emission line, as higher energy lines (UV) are
more emitted more frequently in the centre of the plasma.
Two types of gas flow control systems are available forthe Thermo ScientificiCAP7000 Series ICP-OES.
Coolant Gas
Auxillary Gas
Nebuliser Gas
-
8/9/2019 10576 TN43183_E 11-13C-St2-High
2/3
2 Pressure Control System
A pressure control gas system operates using the
relationship between gas pressure and the corresponding
flow through a fixed orifice, as seen in figure 2.
Figure 2. Schematic of pressure control system
The high pressure input is first reduced with a precision
regulator to the operating pressure of the system.
Coolant and auxiliary gas flows are diverted to a series of
fixed restrictors each controlled with a solenoid valve. In
order to achieve the required flow, combinations of valves
are opened to allow the gas to pass through different
restrictors. The nebulizer gas is passed to a variable, high
precision pressure regulator. The regulator can be
adjusted to give the desired gas flow.
Pressure control gas systems are very robust, reliable and
simple to operate. However, the use of fixed restrictors
can mean a limited number of steps for the gas flow.
Also since gas density changes with temperature, if the
temperature of the incoming gas varies greatly the gas
flow rate will change, affecting the long term stability.
Pressure control of the gas flows is a cost effective design
and ideal for application where a limited number of
methods with similar parameters are used. If different
nebulizer flows are used with different methods then the
flow will need to be manually adjusted each time the
method is changed.
Mass Flow Control System
Mass flow controllers, as the name suggests, operate by
controlling the actual mass flow of the gas. A typical
mass flow controller consists of 3 main components: a
control valve, a flow sensor and the electronic control
circuitry as show in figure 3. The sensor consists of a
long, thin capillary tube with temperature dependent
resistors on the outside. When gas starts to flow the
temperature profile becomes asymmetric, with the
corresponding shift in resistance being proportional to
the gas flow rate. This changing resistance is converted toa voltage and used to operate the gas flow control valve.
Figure 3. Schematic of a typical mass flow controller
The main benefit of mass flow controllers is that since
they operate using electron sensing of the gas flow they
are not affected by changes in atmospheric pressure or
ambient temperature, giving better long term stability.
In addition they offer rapid response times, typically lessthan 1 second, and full computer control. The gas flow
settings are saved in the Thermo Scientific Qtegra
software LabBooks and are automatically set when
analysis is performed.
Due to this automated computer control, full mass flow
control is ideal for complex applications where multiple,
varying analytical methods are required.
-
8/9/2019 10576 TN43183_E 11-13C-St2-High
3/3
TechnicalN
ote4318
3
Conclusions
By utilizing mass flow control for the nebulizer gas flow,
significant benefits can be gained in both long term
analytical stability and ease of use. By adding mass flow
controllers to the auxiliary and coolant gas flows,
performance will be further improved when using axial or
radial plasma viewing. Figure 4 displays the long term
(over 16 hours) analytical stability achieved when using
full mass flow control.
Figure 4. Long term signal stability using full mass flow control
Table 1. Recommended configurations
* Thermo Scientific iCAP 7400 ICP-OES with additional MFC
# Thermo Scientific iCAP 7600 ICP-OES for increased sample
throughput with Sprint Valve as standard upgrade
TN43183_E 11/13C
Africa-Other +27 11 570 1840Australia +61 3 9757 4300
Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)
400 650 5118
Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0
Finland +358 9 3291 0200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591
Japan +81 45 453 9100Latin America +1 561 688 8700
Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00Russia/CIS +43 1 333 50 34 0
Singapore +65 6289 1190South Africa +27 11 570 1840
Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752
www.thermoscientific.com
2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.
All oth er tradem arks ar e the prop erty of Thermo Fisher S cientifi c and its subsidiar ies. This informa tion is pr esented as an
example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any
manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.
Not all products are available in all countries. Please consult your local s ales representative for details.
Thermo Electron ManufacturingLtd (Cambridge) isISO 9001:2008 Certified.
Model 7200 7400 7400* 7600 #
Nebulizer Gas Pressure Control Mass Flow Control Mass Flow Control Mass Flow Control
Auxiliary & Control Gas Pressure Control Pressure Control Mass Flow Control Mass Flow Control
Routine analysis
Similar basic methods
Routine analysis
Varying basic methods
Routine analysis
Similar advanced methods
Research analysis
Varying advanced methods