Page 1: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

100526 Training From MS scan to SIM

Page 2: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Work Flow

• MS Scan• Use Qual anal to determine which parent ion

to look at use the most abundant ion, usually the 2+ or the 3+

• SIM for determining the “fragmentor”

Page 3: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

3. When purging, open the MS scan method that we’ll be using. Double check the solvent.

2. Purge lines, nano and Cap, right click and hit “purge”Green: standby, pink: waiting, blue: in use.When purging, the QQQ should be standby.The instrument will auto turned on when done purging.

1. Load the chip, hit “operate”

Page 4: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Always make sure before start, your chip is default at enrichmentNote that it needs a time point.

Page 5: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

4. Monitoring the total ion here!

5. Make a new folder under D drive>masshunter>data. Open a new worklist, add blank, put in blank sample to autosampler, run method MS scan 001.

Page 6: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

6. Run> run worklist, columns should turn blue~~

Page 7: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Save method!

Refers to 100526 note

Page 8: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

From MS Scan to SIM

• The sample for training contains 9 diff. peptides, ea. of them has the phosphorylated form and non-phosphorylated form (the sample is phosphatase treated!).

• We’ll be looking at the cdh1pS16 and Fin1pS148

• Use ‘proteinprospector’ to determine the m/z of duplet/triplet/+4/etc. charged ions.

Page 9: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Max charge = 4

EAIyAAPFA phosphorylated TyrEAIY[50]AAPFA add in special peptide mass

Page 10: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Most likely! Av, isotopic ave m/z, mono, the most abundant one

Page 11: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

d

1. Extract MS spectra

2. Or look for the ionsDo delete the BPC (base peak chromatogragh)

3. Write down 1.the observed m/z (661.6 & 991.7)2.The corresponding charge (+2 & +3)3.Elusion time 7.3-7.6On printed out proteinprospector sheetThis time we will be using 661.6 for Fin1.

4. 806.6 for cdh1 (+2)

Page 12: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

806.6 for cdh1 (+2)

Page 13: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Open method>SIM80~SIM180, we do 80-120 this timeChange the parameters from this pre-saved methods

We can run up to 4 diff compounds at a time, but ea. Run, the fragmentor can not be changed.

Internal standard peptide?Changed automatically, depending on how many compounds you put to run~

MS2 SIM!!!

Page 14: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Open data files, look at the peak area and pick the best frag. volt.This time, 120 is the best

Peak area: 54038

Page 15: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

From SIM to PI_Scan _CE

• Modify and save the PIScanCE methods• Usually goes from 1530, super small

peptide starts at 10, the thorough range would be 5- 35.

• CE=Collision energy

Page 16: 100526 Training From MS scan to SIM. Work Flow MS Scan Use Qual anal to determine which parent ion to look at  use the most abundant ion, usually the.

Add a script because we won’t be here when the run gets to the end.

Always run two blanks after HSA standard. Because that thing is stickyyy!

Precursor ion

Record the pressures on the log book