1 OPTICAL QUALITY OF DUAL- OPTIC ACCOMMODATING INTRAOCULAR LENS Andrea Galvis, MD 1,3 a Ivan Ossma,...
-
Upload
dominic-york -
Category
Documents
-
view
215 -
download
0
Transcript of 1 OPTICAL QUALITY OF DUAL- OPTIC ACCOMMODATING INTRAOCULAR LENS Andrea Galvis, MD 1,3 a Ivan Ossma,...
1
OPTICAL QUALITY OF DUAL-OPTIC ACCOMMODATING INTRAOCULAR LENS
Andrea Galvis, MD 1,3 a
Ivan Ossma, MD MPH MSc 1,2, a
Sanjeev Kasthurirangan, PhD 4, b
1 Fundacion Valle del Lili, Cali-Colombia
2 Universidad Industrial de Santander, Bucaramanga-Colombia
3 Universidad Autonoma de Bucaramanga, Bucaramanga-Colombia
4 Abbott Medical Optics Inc, Irvine, USA
Financial Disclosure
a Consultant Abbott Medical Optics
b Scientist at Abbott Medical Optics Inc.
The Synchrony IOL is not currently FDA-approved
3
+ Compare Optical Quality Pseudophakic Eyes
+ Monofocal IOL
+ Synchrony Dual Optic Accommodating IOL
+ Parameters
+ Objective iTrace aberrometry
+ Custom Matlab Software to calculate MTF
+ Subjective Contrast Sensitivity Testing
+ Optec 6500
Study Purpose
4
Study Population
iTrace aberrometry
Cycloplegia
3 mm zone
Matlab Software to calculate MTF
Monofocal IOL : 15 subjects (30 eyes)
Synchrony IOL : 16 subjects (32 eyes)
10
Dual Optic Design DOES NOT affect optical or retinal image quality of the eye
Objective evidence (iTrace, MTF)
Subjective evidence (CSF)
Contrast sensitivity of Synchrony is equivalent to that of monofocal IOLs
Discussion
11Discussion
Effect of the slight spherical aberration on Synchrony would only yield 0.42 D of increased depth of focus
Calculations based on Rocha’s work *
The improved near vision of Synchrony patients MOST LIKELY due to true accommodation
* Rocha KM. Using Adaptive Optics Simultion to Explore the Effect of Higher Order Aberrations on Depth of Focus for the Correction of Presbyopia. Wavefront Congress 2008