1 of 26 CRM-106106-AB Oct 2012 Physician Training THE S-ICD ® SYSTEM Patient Screening Prior to...
-
Upload
cory-dawson -
Category
Documents
-
view
214 -
download
0
Transcript of 1 of 26 CRM-106106-AB Oct 2012 Physician Training THE S-ICD ® SYSTEM Patient Screening Prior to...
1 of
26
CR
M-1
0610
6-A
B O
ct 2
012
Physician TrainingTHE S-ICD® SYSTEMPatient ScreeningPrior to using these devices please review the user's manual for a complete listing of indications, contraindications, warnings, precautions, potential adverse events and directions for use.
2 of
26
CR
M-1
0610
6-A
B O
ct 2
012
© Boston Scientific 2012
Patient Screening
Physician Training Topics
3 of
26
CR
M-1
0610
6-A
B O
ct 2
012
© Boston Scientific 2012
Patient Screening
Overview: Why is ECG Screening Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3: Determining an Acceptable Vector
EXAMPLES: ECG Screening Examples
4 of
26
CR
M-1
0610
6-A
B O
ct 2
012
© Boston Scientific 2012
Overview:Why is ECG Screening Necessary?
5 of
26
CR
M-1
0610
6-A
B O
ct 2
012
Overview:Why is ECG Screening Necessary?
OVERVIEW: Why is Screening
Necessary?
STEP 1:Collecting the Surface ECG
STEP 2:Evaluating the Surface ECG
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
The purpose of pre-screening S-ICD® patients is to identify the small number of patients whose S-ICD signals may be unusually challenging for detection and discrimination
The surface ECG is representative of the subcutaneous signal; therefore, a pre-operative screening process was developed to analyze a patient’s ECG without the need to first implant the S-ICD System
The Screening process is used to analyze QRS amplitude, QRS to T-wave amplitude ratios, QRS width and morphology consistency in sinus rhythm using multiple patient postures
7 of
26
CR
M-1
0610
6-A
B O
ct 2
012
Collect a surface equivalent of the subcutaneous sensing vectors:
Collect the surface ECG in the location that represents the intended position of the implanted S-ICD® System
If a non-standard S-ICD System electrode or pulse generator placement is desired, the surface ECG electrode locations should be modified accordingly
Step 1:Collecting the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 2:Evaluating the Surface ECG
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
8 of
26
CR
M-1
0610
6-A
B O
ct 2
012
Typically, the surface ECG electrodes should be positioned as follows:
ECG Electrode LL: Placed in a lateral location, at the 5th intercostal
space along the mid-axillary line to represent the intended location of the implanted pulse generator
ECG Electrode LA: Placed 1 cm left lateral of the xiphoid midline to
represent the intended location of the proximal sensing node of the implanted electrode
ECG Electrode RA: Placed 14 cm superior to the ECG Electrode LA, to
represent the intended position of the distal sensing tip of the implanted electrode
A 14 cm guide is located at the bottom of the transparent screening tool
Step 1:Collecting the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 2:Evaluating the Surface ECG
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
9 of
26
CR
M-1
0610
6-A
B O
ct 2
012
Obtain a clean ECG: Establish a stable baseline. If a wandering
ECG is noted, steps should be taken to establish a stable baseline:
Addition of ground electrode Skin prep, etc.
Record 10 – 20 seconds of ECG in each posture
ECG settings: Leads: I, II and III Sweep speed: 25 mm/sec ECG gain: 5 – 20 mm/mV
Use the largest ECG gain that does not clip the peak of the QRS complex
To yield an acceptable signal for testing, the gain may be adjusted for each ECG lead independently
Step 1:Collecting the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 2:Evaluating the Surface ECG
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
CLIPPING OF THE ECG PEAK
10 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
Record ECG signals in at least two postures: (1) Supine (2) Standing Other postures may be collected
including: Seated, Left Lateral, Right Lateral, and Prone
If morphology changes are expected at elevated heart rates, screening ECGs can be collected during exercise testing for evaluation purposes
If the S-ICD® System is to be implanted with a concomitant pacemaker, all ventricular morphologies (paced and intrinsic, if normal conduction is expected) should be collected
Step 1:Collecting the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 2:Evaluating the Surface ECG
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG14 cm
12 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
Each surface ECG should be evaluated by analyzing all complexes in 10 seconds of QRS complexes in at least 2 different patient postures (supine, standing, sitting etc.)
If multiple morphologies are noted (e.g., bigeminy, pacing, etc.), all morphologies should be evaluated
The evaluation of each QRS complex is a three step process:
1. Select the colored profile2. Align the profile3. Evaluate the QRS complex
Step 2:Evaluating the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
STEP 2:Evaluating the Surface ECG
400 300 200 150 100 90 80 70 60 50 40 30HEART RATE (25 mm/sec) 2 x RR FROM REFERENCE ARROW
200400
600800
100012000
14 cm GUIDE (Note: For screening, ECG electrodes should not extend beyond 14 cm arrows)
400 300 200 150 100 90 80 70 60 50 40 30HEART RATE (25 mm/sec) 2 x RR FROM REFERENCE ARROW
200400
600800
100012000
14 cm GUIDE (Note: For screening, ECG electrodes should not extend beyond 14 cm arrows)
13 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
1. Select the colored profile from the Patient Screening Tool that best matches the amplitude of the QRS complex:
The peak of the QRS complex must fit within one of the Peak Zones as shown
For biphasic signals, the larger peak should be used to select the profile
ECG gains > 20 mm/mV are not permitted
2. Align the left edge of the selected colored profile with the onset of the QRS complex:
The horizontal line on the colored profile should be used as a guide for the ECG baseline
Step 2:Evaluating the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
STEP 2:Evaluating the Surface ECG
1. RECORD: Supine+Standing 25 mm/s, 2-20 mm/mV
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
Peak Z
on
es
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY that one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILE
8 cm
SIMULTANEOUS 3-LEAD ECG
1. RECORD: Supine+Standing 25 mm/s, 2-20 mm/mV
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
Peak Z
on
es
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY that one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILE
8 cm
SIMULTANEOUS 3-LEAD ECG
400 300 200 150 100 90 80 70 60 50 40 30HEART RATE (25 mm/sec) 2 x RR FROM REFERENCE ARROW
200400
600800
100012000
8 cm GUIDE
14 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
3. Evaluate each QRS complex:
Acceptable complex:Entire QRS complex and trailing T-wave are contained within the colored profile
Unacceptable complex:Some portion of the QRS complex or trailing T-wave extends outside of the colored profile
Repeat the evaluation process for all QRS complexes from all surface ECG leads in all collected postures
Step 2:Evaluating the Surface ECG
OVERVIEW:Why is Screening
Necessary?
STEP 3: Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
STEP 2:Evaluating the Surface ECG
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD III
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG
14 cm
2. SELECT the colored profile. The largest QRS peak must be within a Peak Zone.
UNACCEPTABLELEAD
ACCEPTABLELEAD
3. VERIFY at least one lead is acceptable in all postures.
INCORRECTPROFILE
CORRECTPROFILEP
eak Zo
nes
14 cm
RA
LA
LL
LEAD III
LEAD I
1. RECORD: Supine+Standing 25 mm/s, 5-20 mm/mV
SIMULTANEOUS 3-LEAD ECG
14 cm
15 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
© Boston Scientific 2012
Step 3:Determining an Acceptable Vector
16 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
Evaluate each surface ECG lead independently for acceptance
A surface ECG lead (sense vector) is acceptable only if all of the following conditions are met:
1. All QRS complexes and morphologies from the surface ECG lead pass the QRS evaluation.
2. The morphology of the QRS is stable across postures. No significant change to the QRS is noted as a result of postural changes.
3. The surface ECG lead must be deemed acceptable in all tested postures.
A patient is considered suitable for implant if at least one surface ECG lead (sense vector) is acceptable for all tested postures
Unacceptable or marginal sense vectors may result in inappropriate or failure to deliver therapy
Step 3:Determining an Acceptable Sense Vector
OVERVIEW:Why is Screening
Necessary?
EXAMPLES:ECG Screening
Examples
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
Inconsistent Morphology
18 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Example 1:Subtle changes about the baseline…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
SUPINE STANDING
19 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Example 1:Subtle changes about the baseline…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
SUPINE STANDING
20 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Example 1:Subtle changes about the baseline…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
SUPINE STANDING
21 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Example 2:Varying notch morphology…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
22 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Example 3:Inconsistent amplitude and morphology…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
23 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
ECG Screening Examples:Inconsistent amplitude and morphology…
OVERVIEW:Why is Screening
Necessary?
STEP 1: Collecting the Surface ECG
STEP 2: Evaluating the Surface ECG
STEP 3:Determining an
Acceptable Vector
EXAMPLES:ECG Screening
Examples
25 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
Brief SummaryThe S-ICD® System from Boston Scientific CRM
Indications for Use: The S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing.
Contraindications: Unipolar pacemakers are contraindicated for use with the S-ICD System.
Warnings and Cautions: The S-ICD System contains sterile products for single use only. Do not resterilize. Handle the components of the S-ICD System with care at all times and maintain proper sterile technique. All Cameron Health implantable components are designed for use with the Cameron Health S-ICD System only. Connection of any S-ICD System components to any other ICD system will result in failure to deliver lifesaving defibrillation therapy.
General:• External defibrillation equipment should be available for immediate use during the implantation procedure and follow-up.• Placing a magnet over the SQ-RX Pulse Generator suspends arrhythmia detection and therapy response. Removing the magnet resumes arrhythmia
detection and therapy response.• Battery depletion will eventually cause the SQ-RX Pulse Generator to stop functioning. Defibrillation and excessive numbers of charging cycles
shorten the battery longevity.• The S-ICD System has not been evaluated for pediatric use.• The S-ICD System does not provide long-term bradycardia pacing, Cardiac Resynchronization Therapy (CRT) or Anti-Tachycardia Pacing (ATP).
Potential Adverse Events related to implantation of the S-ICD System may include, but are not limited to, the following:
Acceleration/induction of atrial or ventricular arrhythmia; Adverse reaction to induction testing; Allergic/adverse reaction to system or medication; Bleeding; Conductor fracture; Cyst formation; Death; Delayed therapy delivery; Discomfort or prolonged healing of incision; Electrode deformation and/or breakage; Electrode insulation failure; Erosion/extrusion; Failure to deliver therapy; Fever; Hematoma; Hemothorax; Improper electrode connection to the device; Inability to communicate with the device; Inability to defibrillate or pace; Inappropriate post-shock pacing; Inappropriate shock delivery; Infection; Keloid formation; Migration or dislodgement; Muscle stimulation; Nerve damage; Pneumothorax; Post-shock/post-pace discomfort; Premature battery depletion; Random component failures; Stroke; Subcutaneous emphysema; Surgical revision or replacement of the system; Syncope; Tissue redness, irritation, numbness or necrosis.
26 o
f 26
CR
M-1
0610
6-A
B O
ct 2
012
Abbreviated Statement (CE Mark)
All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sale by or on the order of a physician. Indications, contraindications, warnings and instructions for use can be found in the product labeling supplied with each device. Information for the use only in countries with applicable health authority product registrations.