I N N O V A T I O N A N D T E C H N O L O G Y F O R B E T T E R P A T I E N T O U T C O M E S
ORTHOPAEDIC DRAINAGE
BLOOD MANAGEMENT
ORTHO P.A.S.DRY WASH
Quick User’s Guideline
This document is not intended to replace the ORTHO P.A.S. Dry Wash user’s manual, it is for quick reference only.
For complete ORTHO P.A.S. Dry Wash set up and procedure please refer to its User’s manual.
Close both clamps
Innovation and Technology for Better Patient Outcomes
www.eurosets.it
Remove caps (male and female)
from the bag
Remove bag and filter. The system is now working
as drainage only
42 3
disconnect the bag from the reservoir and close the connections
END OF REINFUSION
1
Turn vacuum to max, turn lever to “pause”, disconnect vacuum pump
REDON FUNCTIONING
Remove the green pouchKeep the transparent pouch
1
A - Place the Quality Sensor to the reservoir.
B - Open the cover
1
Move Ortho P.A.S. Dry Wash out of the sterile field. Remove the tran-sparent pouch. Close the clamp.
4
Turn the lever to the left (“vent” position). The yellow
indicator becomes visible
4
Turn the pump on and adjust the vacuum Close the clamp
7 7
cut the fingers connector according to the
drainage size in use
2
A - Slide the bag out from the pouch at the back of the reservoir. B - Put the red tube inside the clamp
C - Hang the bag
2
Connect the vacuum pump: holding the black ring down,
slide the pump up
5
A - Turn on Quality SensorB - Open the clamp
5
The yellow indicator disappears
(vacuum ON)
8
Turn the lever back to “suction” position.
Yellow indicator disappears
8
write down the blood collection
starting date and time
6
Press Start Procedure. The blood moves into the bag. Supernatant and fat remain in the reservoir
6
Hang Ortho P.A.S. Dry Wash to the bed.
System working
Remove Quality Sensor at the end of the procedure
9
REINFUSIONSET UP
Connect the drainage to the fingers connector
3
Close the cover
Transfer Ortho P.A.S. Dry Wash
to the sterile field.
3
9
In case of multiple reinfusion, repeat the complete sequence without disconnecting the Goccia filter, neither from the bag, nor from the Patient
Take the bag andopen the connector
10
Spikez the Goccia 40mm filter
11
Prime the Goccia 40mm filter and the reinfusion line. Reinfuse the blood*
12
*Th
e b
loo
d s
ho
uld
be
rein
fuse
d a
cco
rdin
g to
yo
ur H
osp
ital
’s Pr
oto
col
1 Holding system (hook)
2 Open/close button
3 RBC’s sensor + automatic clamp
4 Hook for transfer bag
5 Procedure key-pad
QUALITY SENSOR DESCRIPTION
A
B
C
A
B
1
2
3
4
5
A
B