1 / 17
Deutscher WetterdienstDeutscher WetterdienstMeteorological Observatory Lindenberg Meteorological Observatory Lindenberg
Richard Assmann ObservatoryRichard Assmann Observatory
The GCOS Reference Upper Air NetworkThe GCOS Reference Upper Air Network
Holger Vömel
GRUAN Lead CenterMeteorological Observatory Lindenberg
German Weather Service
2 / 17
GRUAN MandateGRUAN Mandate
The GCOS Reference Upper-Air Network is tasked to:
1. Provide long-term high quality climate records
2. Constrain and calibrate data from more spatially-comprehensive global observing systems (including satellites and current radiosonde networks)
3. Fully characterize the properties of the atmospheric column
3 / 17
Measured quantitiesMeasured quantities
Focus on priority 1: Pressure, temperature, water vaporFocus on priority 1: Pressure, temperature, water vapor Focus on upper troposphere and stratosphereFocus on upper troposphere and stratosphere Focus on reference observations for climate researchFocus on reference observations for climate research
4 / 17
Initial NetworkInitial Network
To be expanded to about 30 to 40 stations worldwide
5 / 17
Initial NetworkInitial Network
To be expanded to about 30 to 40 stations worldwide
Tibetan Plateauis missing
6 / 17
Reference observationsReference observations
To be a referenceTo be a referenceGRUAN observations must:GRUAN observations must:
• Specify measurement uncertainty for every data point
• Relate to the definition of a unit or a transfer standard
• Be traceable (meta-data, documentation)
7 / 17
Establishing reference qualityEstablishing reference quality
8 / 17
Establishing UncertaintyEstablishing Uncertainty
• Guide to the expression of uncertainty in measurement (GUM, 1980)Evaluation of measurement data – Guide to the expression of uncertainty in
measurement(http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_E.pdf)
• CIMO guide for observationsGuide to Meteorological Instruments and Methods of Observation, World
Meteorological Organization, 7th edn., 2006(http://www.wmo.int/pages/prog/www/IMOP/publications/CIMO-Guide/CIMO Guide-7th Edition-
2006.html)
• A guide for upper-air reference measurementsPaper by Immler et al. (2010) submitted to Atmos. Meas. Techn.
Open for public comments
9 / 17
UncertaintyUncertainty
GUM concept:GUM concept:• The "true value" of a physical quantity is no longer used.
• Error is replaced by uncertainty
• Result of a measurement = a range of values
generally expressed by m ± u
m is corrected for systematic effects
u is (random) uncertainty
10 / 17
Example: Vaisala RS92 temperature
11 / 17
RS92 calibrationRS92 calibration
12 / 17
Sources of uncertaintySources of uncertainty
• Sensor calibration:Sensor calibration:Accuracy of calibration referenceAccuracy of calibration referenceAccuracy of calibration modelAccuracy of calibration model
• Sensor integration:Sensor integration:Integration into radiosondeIntegration into radiosondeTelemetry limitationsTelemetry limitations
• Sensor characterization:Sensor characterization:Time lag variation of polymer sensorTime lag variation of polymer sensorController stability of frostpoint hygrometerController stability of frostpoint hygrometerProduction variability Production variability
• External influences:External influences:Radiation errorRadiation errorBalloon/payload contaminationBalloon/payload contaminationSensor icingSensor icing
13 / 17
Sources of uncertaintySources of uncertaintyExample : TemperatureExample : Temperature
• Sensor calibration:Sensor calibration:Accuracy of calibration referenceAccuracy of calibration referenceAccuracy of calibration modelAccuracy of calibration model
• Sensor integration:Sensor integration:Integration into radiosondeIntegration into radiosondeTelemetry limitationsTelemetry limitations
• Sensor characterization:Sensor characterization:Time lag variation of polymer sensorTime lag variation of polymer sensorController stability of frostpoint hygrometerController stability of frostpoint hygrometerProduction variability Production variability
• External influences:External influences:Radiation errorRadiation errorBalloon/payload contaminationBalloon/payload contaminationSensor icingSensor icing
Vaisala RS92 daytimeVaisala RS92 daytime
0.01K (est.)0.01K (est.) 0.01K (est.)0.01K (est.)
? *? * - -
0.5 K * (strat.)0.5 K * (strat.) 0.5 K 0.5 K
(strat.)(strat.)>1.0K (temporarily)>1.0K (temporarily)
(* after correction is applied)(* after correction is applied)
14 / 17
Ground check correctionGround check correction
15 / 17
0.4
0.2
0.0
-0.2
-0.4 Sep 09 Oct 09 Nov 09 Dec 09 Jan 10 Feb 10
Lindenberg additional groundcheckLindenberg additional groundcheck
Mean = -0.11 +/- 0.06 K
ΔT
[K]
16 / 17
Determination of radiation effectDetermination of radiation effect
Vacuum chamber used for measuring the radiation effect using direct sunlight.
Radiation data of BSRN station was used.
RS-92, DFM-06 and Imet1 sensors tested.
Ventilation measured and varied during measurement.
17 / 17
18 / 17
Redundant observations Redundant observations to validate uncertainty estimateto validate uncertainty estimate
LUAMI radiosonde intercomparison 2008LUAMI radiosonde intercomparison 2008
19 / 17
SummarySummary
GRUAN is a reference network for upper air essential climate variables
To start focus on temperature and water vapor
Network is world wide (but large regions missing)
Reference observation: TraceableDocumentedQuantified uncertainties
Uncertainties: Analyze sources, synthesize best estimate, verify in redundant observations
20 / 17
Extras
21 / 17
Metrological Metrological TraceabilityTraceability
The result of a measurement can be related to the definition of a unit by an unbroken document chain of calibrations, each of which contributes to the measurement uncertainty
22 / 17
What is a reference measurement?What is a reference measurement?
A reference measurement gives: A reference measurement gives:
The best estimate for the quantity to be measured
The best estimate for the level of confidence for this measurement (i.e. uncertainty)
To be a referenceTo be a referenceGRUAN observations must includeGRUAN observations must include
the measurement uncertaintythe measurement uncertainty
23 / 17
UncertaintyUncertainty
Type A evaluation: statistical analysis
Type B evaluation:
previous measurement data
experience with or general knowledge of the behavior and properties of relevant instruments
manufacturer's specifications
data provided in calibration and other certificates
uncertainties assigned to reference data taken from handbooks
24 / 17
Redundancy and ConsistencyRedundancy and Consistency
GRUAN stations should provide redundant measurements ( → collocation issue)
Redundant measurements should be in agreement.
an essential part of operational quality assurance
Consistency:
No meaningful consistency analysis possible without uncertainties
22
2121 uukmm
25 / 17
Consistency analysisConsistency analysis
sig. diff.
sig. diff.
agree
consistent
0
1
2
3
expanded errorbars ( k=2)
Measurement A
Measerument B
22
2121 uukmm