09 11 2006 Chromatography

24
WenHsin Huang NDMC 1 2006/9/7 WenHsin Huang, NDMC 1 Principles of Chromatography Introduction to Analytical Separations Wen-Hsin Huang Ph. D. NDMC 2006/9/7 WenHsin Huang, NDMC 2 Principles of Chromatography Introduction to Analytical Separations Solvent Extraction What is Chromatography? A Plumbers View of Chromatography Efficiency of Separation Why Bands Spread 2006/9/7 WenHsin Huang, NDMC 3 Simple Separations Extraction transfer of solute from one phase to another phase can be gas, solid, liquid Liquid/liquid extraction 2 immiscible solvents used typically aqueous solvent and organic solvent you know water and oil dont mix organic solvents less dense than water diethyl ether, toluene, hexane organic solvents more dense than water chloroform, carbon tetrachloride, dichloromethane 2006/9/7 WenHsin Huang, NDMC 4 Solvent Extraction shake add second immiscible solvent Legend: aqueous solvent organic solvent solute particles Solute particles only have one choice of solvent Solute particles now have two choices of solvent Solute particles choose preferred solvent

description

物理生化學-09-11-2006 Chromatography-黃文鑫

Transcript of 09 11 2006 Chromatography

Page 1: 09 11 2006 Chromatography

WenHsin Huang

NDMC 1

2006/9/7 WenHsin Huang, NDMC 1

Principles ofChromatography

Introduction toAnalytical Separations

Wen-Hsin HuangPh. D.NDMC

2006/9/7 WenHsin Huang, NDMC 2

Principles of Chromatography

•Introduction to Analytical Separations–Solvent Extraction–What is Chromatography?–A Plumber’s View of Chromatography–Efficiency of Separation–Why Bands Spread

2006/9/7 WenHsin Huang, NDMC 3

Simple Separations•Extraction

–transfer of solute from one phase to another–phase can be gas, solid, liquid

•Liquid/liquid extraction–2 immiscible solvents used–typically aqueous solvent and organic solvent

•you know water and oil don’t mix

–organic solvents less dense than water•diethyl ether, toluene, hexane

–organic solvents more dense than water•chloroform, carbon tetrachloride, dichloromethane

2006/9/7 WenHsin Huang, NDMC 4

Solvent Extraction

shake

add secondimmiscible

solvent

Legend:aqueous solventorganic solventsolute particles

Solute particlesonly have one

choice of solvent

Solute particlesnow have two

choices of solvent

Solute particleschoose preferred

solvent

Page 2: 09 11 2006 Chromatography

WenHsin Huang

NDMC 2

2006/9/7 WenHsin Huang, NDMC 5

Solvent Extraction

•Like dissolves like–solute chooses solvent most like itself–polar compounds (and ionic compounds)

choose water (which is very polar)–nonpolar compounds choose nonpolar organic

solvents

2006/9/7 WenHsin Huang, NDMC 6

Phase Partitioning

If we have a solute (S) that is partitioned between twophases (1 & 2) then we can write an equilibrium expressionfor this equilibrium.

21 SS

][][

1

2

1

2

SS

A

AK

S

S

Partition Coefficient (K)

•Convention: Organic phase = phase 2

2006/9/7 WenHsin Huang, NDMC 7

Distribution Coefficient

The distribution coefficient (D) is used in place of thepartition coefficient when dealing with a species that hasmore than one chemical form.

1____2____

phaseinionconcentrattotalphaseinionconcentrattotal

D

2006/9/7 WenHsin Huang, NDMC 8

Example of Extraction with n-HexaneThe distribution coefficient for compound Z betweenn-hexane and water is 6.25. Calculate the percent Zremaining in 25.0 mL of water that was originally0.0600 M in Z after the extraction with the followingvolumes of n-hexane.

A) One 25.0-mL portion

B) Two 12.5-mL portions

C) Five 5.00-mL portions

D) Ten 2.50-mL portions

13.8%

1.90%

4.99 x 10-3%

2.49 x 10-7%

It is more efficient to do several small extractionsrather than one large extraction.

Page 3: 09 11 2006 Chromatography

WenHsin Huang

NDMC 3

2006/9/7 WenHsin Huang, NDMC 9

Solvent Extraction (pH effects)

•Charges of acidic and basic species changewith pH–neutral species generally more soluble in

organic solvent–charged species generally more soluble in

aqueous solution

•Distribution coefficient (D) describesdistribution of species between two phases–Takes into account all forms of a compound

(i.e. H2A, HA-, A2-)–Different from partition coefficient (K)

2006/9/7 WenHsin Huang, NDMC 10

Solvent Extraction (pH effects)

•Distribution of HA between two solvents–two equilibria to consider

•Ka equilibrium•D equilibrium

aqueous HA H+ + A-Ka

organic HA H+ + A-

2006/9/7 WenHsin Huang, NDMC 11

Solvent Extraction (pH effects)

•Distribution of acids and bases between twophases

•Ionic species only found in aqueous layer

D =total conc. in phase 2 (organic)

total conc. in phase 1 (aqueous)

D =[HA]2

[HA]1 + [A-]1D =

[B]2[B]1 + [BH+]1

2006/9/7 WenHsin Huang, NDMC 12

Solvent Extraction (pH effects)•Distribution of HA between two phases

•Also have K expression for [HA]

•Substitute and rearrange to get

]HA[]A][H[

Ka ]H[]HA[K

]A[ a

1

2

]HA[]HA[

K

aK]H[]H[K

D

Eq’n

Page 4: 09 11 2006 Chromatography

WenHsin Huang

NDMC 4

2006/9/7 WenHsin Huang, NDMC 13

Solvent Extraction (pH effects)D for an acid (HA) D for a base (B)

•Distribution of acids and bases between twophases is pH-dependent

•To find D, need to know–pH of solution–Ka of acid or conjugate acid–K of acid or base

]H[KKK

Da

a

Eq’ns

aK]H[]H[K

D

2006/9/7 WenHsin Huang, NDMC 14

Solvent Extraction (pH effects)•Adjust pH to effectively extract an acid or

base (and its conjugates) into aqueous layer•Consider base B with a pKa of 9.0

–pH < 9.0, B hydrolyzes so majority of B (in theform BH+) in water layer (low D value)

2 4 6 8 10 12pH

log

D

0

-2

-4

-6

pKa

mainlyB

mainlyBH+

Fig.

Ba

a

aqaq

org KHK

KKBHB

BD

][][][

][

aq

org

B

BK

][

][

aq

aqa BH

BHK

][

]][[

2006/9/7 WenHsin Huang, NDMC 15

History of ChromatographyMikhail Tswett is credited with theinvention of chromatography.

Developed a technique that separatedvarious plant pigments (e.g.,chlorophylls and xanthophylls) bypassing solutions through glasscolumns filled with finely groundCaCO3.

The separated species appeared ascolored bands on the column.

Hence Greek chroma = color and graphein = writing

M. S. TswettRussian Scientist

(1872-1919)

2006/9/7 WenHsin Huang, NDMC 16

Tswett’s Experiment

Ground up plant leaves and addedpetroleum ether.

Filled column with chalk.

Method forgotten for many years.

Page 5: 09 11 2006 Chromatography

WenHsin Huang

NDMC 5

2006/9/7 WenHsin Huang, NDMC 17

Tswett’sExperiment

2006/9/7 WenHsin Huang, NDMC 18

What is Chromatography?

•Method to separate components in amixture based on different distributioncoefficients between the two phases

•Same principle as solvent extraction (likedissolves like), but one phase is “stationary”and one phase is “mobile”–no longer working with organic and aqueous

layers in a separatory funnel–working in a column (which is just a tube)

2006/9/7 WenHsin Huang, NDMC 19

What is Chromatography?

•Two phases: mobile and stationary•Mobile phase is solvent moving through

column–liquid (Methanol, water, buffer)–gas (He, H2, N2)

•Stationary phase fixed inside column–viscous liquid coated on inside of column–solid particles packed inside column

•Solutes have different affinities for mobilephase and stationary phase

2006/9/7 WenHsin Huang, NDMC 20

Types of Chromatography

•Adsorption Chromatography•Partition Chromatography•Ion-exchange Chromatography•Molecular Exclusion Chromatography•Affinity Chromatography

Page 6: 09 11 2006 Chromatography

WenHsin Huang

NDMC 6

2006/9/7 WenHsin Huang, NDMC 21

Adsorption ChromatographyStationary phase Mobile phaseSolid Liquid or Gas

How it separates–solute adsorbs on to

stationary phase surface

2006/9/7 WenHsin Huang, NDMC 22

Partition ChromatographyStationary phase Mobile phaseLiquid on solid support Liquid or Gas

How it separates–solute dissolves into

liquid coating

2006/9/7 WenHsin Huang, NDMC 23

Ion-Exchange ChromatographyStationary phase Mobile phaseAnions or cations covalently Liquidbound to solid stationary phase

How it separatessolute ions of opposite chargeattracted to stationary phase

2006/9/7 WenHsin Huang, NDMC 24

Molecular ExclusionChromatography

Stationary phase Mobile phasePorous gel Liquid

How it separates–small molecules

trapped in pores ofstationary phase

Also calledgel filtration,gel permeation,

orsize-exclusion

chromatography

Page 7: 09 11 2006 Chromatography

WenHsin Huang

NDMC 7

2006/9/7 WenHsin Huang, NDMC 25

Affinity Chromatography

Stationary phase Mobile phaseImmobilized molecules on Liquidliquid or solid stationary phase

How it separates–Molecules with specific

shape dock with ligands

2006/9/7 WenHsin Huang, NDMC 26

Classification Specific Method Stationary Phase Type of equilibriumLiquidchromatography(LC) (Mobilephase a liquid)

Liquid-liquid orpartition

Liquid-bondedphase

Liquid-solid, oradsorption

Ion exchange

Size exclusion ormolecular-exclusion

AffinityChromatography

Liquid adsorbed ontosolid

Organic species bondedto solid surface

Solid

Ion-exchange resin

Interstices of polymericsolid

Covalently bondedmolecule (eg antibody)

Partition between immiscibleliquids

Partition between liquid and bondedsurface

Adsorption

Ion exchange

Sieving

Selective Affinity – (eg specificProtein/antibody bind together)

Gaschromatography(GC) (Mobilephase a gas)

Gas-liquid

Gas-bondedphase

Gas-solid

Liquid adsorbed ontosolid surface

Organic species bondedto solid surface

Solid

Partition between gas and liquid

Partition between gas and bondedliquid

Adsorption

2006/9/7 WenHsin Huang, NDMC 27

Basic PrinciplesTwo phases considered:

1) Mobile Phase: solvent moving through the column.

2) Stationary Phase: stays in place inside of the column.

COLUMN“Eluent” “Eluate”

Process is called “elution”

2006/9/7 WenHsin Huang, NDMC 28

Simple Column

Fresh eluent

Initial bandof A and B

ColumnPacking

Porous Disk

Page 8: 09 11 2006 Chromatography

WenHsin Huang

NDMC 8

2006/9/7 WenHsin Huang, NDMC 29

Chromatogram

A chromatogram is agraph showing thedetector response(proportional toconcentration) assolutes emerge from achromato-graphiccolumn as a function oftime or volume.

Detector

2006/9/7 WenHsin Huang, NDMC 30

Chromatographic Terms

tm = “dead time”which isthe minimum time formobile phase to passthrough the column

tr = “retention time”which is the time requiredfor the solute (analyte) topass through the column.

t’r = “adjusted retentiontime”describes timesolutes spend in stationaryphase

mrr ttt •All solutes spend equal time in mobile phase

2006/9/7 WenHsin Huang, NDMC 31

Chromatographic Terms

= “relative retention”which is the ratio ofadjusted retention times.

>1 always

then separationbetween two components

k’= “capacity factor”

m

r

tt

mobileintimestationaryintime

k

__

__

1

2

r

r

tt

2006/9/7 WenHsin Huang, NDMC 32

A Plumber’s View…•kcapacity factor describes how long

solute retained on column

•Higher kindicates solute retained longer•If k= 0, solute unretained•If k= 1, solute spent same amount of time

in stationary phase as in mobile phase

m

mr

ttt

'k

time spent in stationary phasetime spent in mobile phase

=

Page 9: 09 11 2006 Chromatography

WenHsin Huang

NDMC 9

2006/9/7 WenHsin Huang, NDMC 33

A Plumber’s View…•kdescribes time spent in two phases•kalso describes number of moles of solute

partitioned between two phases

•K Partition coefficient comparesconcentration of solute in one phase toconcentration of solute in the other

mm

ss

VCVC

'k

m

s

VV

K'k m

s

1

2

CC

]S[]S[

K

Eq’n

Eq’n

2006/9/7 WenHsin Huang, NDMC 34

Partition Coefficient Relationship

m

r

tt

mobileinmolesstationaryinmoles

mobileintimestationaryintime

k

__

____

__

m

r

m

s

mm

ss

tt

VV

KVSVS

k

][][

1

2

1

2

1

2

KK

kk

tt

r

r

What this says is that analytes with different partitioncoefficients will have different retention times.

2006/9/7 WenHsin Huang, NDMC 35

A Plumber’s View…•q fraction of solute in mobile phase

ssmm

mm

ms

m

VCVCVC

molesmolesmoles

q

m

s

mm

ss

VV

K1

1

VCVC

1

1q

'k11

q

fraction in MP

'k1'k

)q1(

fraction in SP2006/9/7 WenHsin Huang, NDMC 36

Review

m

s

CC

K Partition coefficient describesconc. of solute in SP and MP

m

s

m

mr

VV

Kt

tt'k

Capacity factor describestime solute spends (or moles ofsolute) in SP and MP

1

2

1

2

1r

2r

KK

'k'k

't't

Relative retention comparesretention times (or partitioncoefficients or capacity factors)of solute 1 and 2

Page 10: 09 11 2006 Chromatography

WenHsin Huang

NDMC 10

2006/9/7 WenHsin Huang, NDMC 37

Chromatographic Peaks•Solutes elute from a column in a Gaussian peak

shape–w width of peak at base–w1/2 width of peak at half-height

tr

timet0

h

1/2h

w1/2=2.35

w=4

averageretention time

Peak width (w) is defined as the baseline width (4).2006/9/7 WenHsin Huang, NDMC 38

Resolution of peaks

2/)ww()tt(

wt

resolution21

2r1r

av

r

As tr grows,resolutionbetween twopeaksimproves.

R=0.50 R=0.75

R=1.00 R=1.50

Resolution is a measureof how well two peaks areseparated.

2006/9/7 WenHsin Huang, NDMC 39

Resolution Example

Two solutes have retention times and widths of tr1 =235 s, w1 = 8 sec; tr2 = 250 s, w2 = 10 s.

Resolution = 1.7

2006/9/7 WenHsin Huang, NDMC 40

Plate Height (H)Bands broaden as they travel through the column.

Plate height (H) relates the amount of broadening tothe linear distance traveled.

xH

2H = plate height

x = distance traveled2 = variance of peak

Page 11: 09 11 2006 Chromatography

WenHsin Huang

NDMC 11

2006/9/7 WenHsin Huang, NDMC 41

Evaluating Separation Efficiency•Plate theory

–Breaks separation up into many discrete stages–Stages represent individual equilibria

•N (Theoretical plate) represents eachequilibrium between MP and SP

•Each time a solute moleculeenters the SP from the mobilephase is a theoretical plate

2006/9/7 WenHsin Huang, NDMC 42

Evaluating Separation Efficiency•“N”in chromatography is analogous to “n”

in liquid-liquid extractions–each extraction represents a theoretical plate

•Higher N corresponds to better separation–indicates solute molecules enter SP a higher

number of times

[S]1

[S]2

2006/9/7 WenHsin Huang, NDMC 43

Evaluating Separation Efficiency

•N is dependent on–column–analyte–retention time of analyte–width (at base or at half-height) of peak

2

2r

2

2r

wt16t

N

22/1

2r

wt55.5

N

N is dimensionless someasure tr and w (or w1/2)in same units (i.e. time)

2006/9/7 WenHsin Huang, NDMC 44

Chromatographic Peaks•Distribution of retention times for a solute•Peak shape represents differences in

behavior of solute molecules duringseparation

tr

timet0

h

1/2h

w1/2=2.35

w=4

averageretention time

Page 12: 09 11 2006 Chromatography

WenHsin Huang

NDMC 12

2006/9/7 WenHsin Huang, NDMC 45

Evaluating Separation Efficiency•Do not want solute band to spread out as it

travels through column•H (Plate height) represents the relationship

between the width of a solute band to thedistance traveled through column

•Lower H corresponds to better separation(higher efficiency)

•H also called HETP–Height Equivalent to a Theoretical Plate–One solute equilibrium between SP and MP is

one theoretical plate (N)–Equilibrium occurs in one HETP

2006/9/7 WenHsin Huang, NDMC 46

Evaluating Separation Efficiency

•H related to N and length of column (L)

•Resolution related to N

–R is dimensionless–k2 = capacity factor for solute retained longer

on column (higher tr)

av

2

'k1'k1

4N

R

NL

H Report H in unitsof length (i.e. cm)

Eq’n

2006/9/7 WenHsin Huang, NDMC 47

Evaluating Separation Efficiency

•Increase L to increase R

•Increase to increase R

–to change , must change nature of MP and SPto change relative affinity analytes have for SP

1

2

1

2

1r

2r

KK

'k'k

't't

av

2

'k1'k1

4N

R

HL

NNL

H

Eq’n

2006/9/7 WenHsin Huang, NDMC 48

Number of Theoretical Plates (N)The number of theoretical plates (N) increases asthe efficiency of the separation increases.

2

2

2

216

rr tw

tN

or

2

2

2

21

255.5

rr tw

tN

Derived

Page 13: 09 11 2006 Chromatography

WenHsin Huang

NDMC 13

2006/9/7 WenHsin Huang, NDMC 49

Example

A solute with a retention time of 302 s has a width of11 s on a column that is 15 m long. Find both N andH for this separation.

N = 1.2 x 10-4

H = 1.2 mm

2006/9/7 WenHsin Huang, NDMC 50

Factors Affecting ResolutionThe greater the resolution, the better the separation.The more theoretical plates, the better the separation.

avek

kN1

14

R 2

NR LR 0R 10R 02 k

2006/9/7 WenHsin Huang, NDMC 51

Principles of Chromatography

•Introduction to Analytical Separations–Solvent Extraction–What is Chromatography?–A Plumber’s View of Chromatography–Efficiency of SeparationWhy Do Bands Spread

2006/9/7 WenHsin Huang, NDMC 52

Why Do Bands Broaden?

Solute invariably spreads apart as it travels(diffusions) through a chromatographiccolumn.

The observed variance is a sum of variancesfrom all the broadening mechanisms.

22iobs

Page 14: 09 11 2006 Chromatography

WenHsin Huang

NDMC 14

2006/9/7 WenHsin Huang, NDMC 53

Where Does Broadening Occur?

Outside the column:Width of injection plugMixing in detector dead volume

I

Pump

Column Detector

Injector

2006/9/7 WenHsin Huang, NDMC 54

Where Does Broadening Occur?Inside the column:

Multiple pathsLongitudinal diffusionEquilibration time

I

Pump

Column Detector

Injector

2006/9/7 WenHsin Huang, NDMC 55

Multiple Paths (A)•Solute molecules can choose many different

paths through packed columns

layer of SPcoated oninside of column

open tubularcolumn

SP particlespacked insidecolumn

packedcolumn

Compare to solutetraveling through

a straw

Compare to solutetraveling through

a bean bag2006/9/7 WenHsin Huang, NDMC 56

Multiple Paths (A)

•Open tubular columns do not provide multiple paths for soluteduring tm (no A term)

time

Band of 3 solutemolecules travelingthrough packed column

Band spreads

Occurs only in packed columns.

Smaller particles reduce the plate height.AH

Page 15: 09 11 2006 Chromatography

WenHsin Huang

NDMC 15

2006/9/7 WenHsin Huang, NDMC 57

Longitudinal Diffusion (B /ux)•S molecules travel through column in a band•[S] varies throughout band

–some solute molecules get ahead of band–some solute molecules lag behind band

start

as bandtravels

H B/ux so,increase ux (flow rate)to reduce problem

Caused by diffusion of solute in the mobile phase.

Faster flow rate means less time, thus less diffusion.

2006/9/7 WenHsin Huang, NDMC 58

Equilibration Time (Cux)•Finite time required to allow solute molecules to

equilibrate between MP and SP•If flow rate too high, MP will “leave behind”

molecules in SP (i.e., Solute in stationary remains“stuck”while solute in mobile phase moves forward.)

•Keep SP thin to decrease equilibration time

sloweq.

MPSP

Directionof travel

bandwidth

MPSP

bandwidth

H Cux so decrease ux to reduce problem

2006/9/7 WenHsin Huang, NDMC 59

Van Deemter EquationPutting the three factors together yields the VanDeemter equation that helps predict how thecolumn flow rate will affect the theoretical plateheight.

xx

CuuB

AH

MultiplePaths Longitudinal

Diffusion

EquilibrationTime

•Break equation up into A, B, and C terms

–A H independent of flow rate

–B/ux H inversely proportional to flow rate

–Cux H proportional to flow rate2006/9/7 WenHsin Huang, NDMC 60

Band Spreading

•Three terms in van Deemter equationcorrespond to three sources of bandspreading or broadening

•A term Multiple Paths term(also called Eddy Diffusion)

•B term Longitudinal Diffusion term•C term Equilibration Time term

Page 16: 09 11 2006 Chromatography

WenHsin Huang

NDMC 16

2006/9/7 WenHsin Huang, NDMC 61

van Deemter plot•Plot dependence of H on individual terms•Sum and find optimum ux to use to minimize band

spreading

H

Flowrate

Bux

Cux

A

uopt

Hmin

2006/9/7 WenHsin Huang, NDMC 62

Column Type Affects H

•Advantages–high resolution (low H)–high sensitivity–low analysis time

•Disadvantages–not useful for large-scale

analyses

•Advantages–useful for large-scale

analyses

•Disadvantages–A term in van Deemter

equation increases H

layer of SPcoated oninside of column

open tubularcolumn

SP particlespacked insidecolumn

packedcolumn

2006/9/7 WenHsin Huang, NDMC 63

Asymmetric Band Shapes

•Expect Gaussian peak shape from solute Seluting from column, independent of [S]total

•Separation based on K

•In reality, K changes as [S]total changes•If [S] is too high or too low, peak shape

deviates from Gaussian shape

m

s

CC

K

2006/9/7 WenHsin Huang, NDMC 64

Asymmetric Band Shapes•Overloaded column (high [S])

–so much S enters SP that the SP starts to lookmore like S than it looks like original SP

–most S molecules retained longer on column

Band shape

Majority of S retainedon column longer

Observed peakin chromatogram

Faster-moving (lessretained) S reachesdetector first

Page 17: 09 11 2006 Chromatography

WenHsin Huang

NDMC 17

2006/9/7 WenHsin Huang, NDMC 65

Asymmetric Band Shapes•Underloading (low [S])

–“hot sites”on SP more available when less Smolecules trying to enter SP

–some (minority) S get “stuck”on hot sites–some S retained longer on column

Band shapeObserved peakin chromatogram

Minority of S retainedon column longer

Slower-moving (moreretained) S reachesdetector last 2006/9/7 WenHsin Huang, NDMC 66

Asymmetric Band Shapes•Load less sample to reduce overloading

problem•Protect groups on SP to reduce number of

hot sites to reduce tailing problem

Cs

Cm

Constant K (slope) resultsin ideal peak shape

tailed

overloaded

Fig. 23-19

2006/9/7 WenHsin Huang, NDMC 67

Chromatographer’s TriangleRelationship

Resolution

Speed Capacity

2006/9/7 WenHsin Huang, NDMC 68

Quantitative Analysis

•In general, detectors can tell us–“Yes, something is eluting from the column.”

•Use calibration methods to determine how muchof a compound is eluting from column

Finding Peak Area•A = ½ wbaseh•A w½ h•cut-and-weigh•computer integrationconcentration (M)

Pea

kA

rea

x =-bm

Page 18: 09 11 2006 Chromatography

WenHsin Huang

NDMC 18

2006/9/7 WenHsin Huang, NDMC 69

Quantitative Analysis•Generally use an internal standard

–internal standard is a known amount of acompound different from analyte

–compare analytical signal from analyte toanalytical signal from internal standard

–not the same as method of standard additions–internal standards discussed in Ch. 4

]S[A

F]X[

A sx

Ax area of analyte peakAs area of standard peak[X] concentration of analyte[S] concentration of standardF response factor

2006/9/7 WenHsin Huang, NDMC 70

Example

When 1.06 mmol of 1-pentanol and 1.53 mmolof 1-hexanol were separated by GC, they gaverelative peak areas of 922 and 1570 units,respectively. When 0.57 mmol of pentanol wasadded to an unknown containing hexanol, therelative chromatographic peak areas were843:816 (pentanol:hexanol). How muchhexanol did the unknown contain?

2006/9/7 WenHsin Huang, NDMC 71

Liquid Chromatography

•Analyte–Soluble in MP (more likely than being volatile)

•Mobile phase–Liquid

•Methanol, Water, Acetonitrile, Hexane

–Plays more active role in separation•Stationary phase

–Liquid (partition) or solid (adsorption)•High Performance Liquid Chromatography(HPLC)

–High pressure used to force solvent through column (asopposed to gravity)

2006/9/7 WenHsin Huang, NDMC 72

LC Columns

•Constructed of steel or plastic•5 –30 cm length•1 –5 mm i.d.•Easily contaminated and degraded•Guard column

–Contains same SP–Dust, other particles, strongly adsorbed

solutes retained on guard column–Expendable–Extends life of analytical column

ANALYTICAL

GUARD

Page 19: 09 11 2006 Chromatography

WenHsin Huang

NDMC 19

2006/9/7 WenHsin Huang, NDMC 73

LC Columns•Almost exclusively use packed columns•Solutes move much slower through liquids

than through gases–Time needed to diffuse to SP in open tubular

column is too long

S S

2006/9/7 WenHsin Huang, NDMC 74

LC Columns•Packing particle diameter (dp) crucial

–Typical particle diameters are 3 –10 m–Decrease particle diameter to decrease H

•Provide more uniform flow (low A)•Less time needed for solute to get to particle (low C)

Flow rate (mL/min)

10m

0 2 4 6 8

H(

m)

20

40

60

10

5m

3m

Fig. 21-14

2006/9/7 WenHsin Huang, NDMC 75

LC –Adsorption Chromatography

•Solid SP –Silica gel–Pure, spherical, microporous

particles•Permeable to solvent•Very high surface area•Use when pH < 8

Silica gel with silanolgroups on surface

Deprotonated silanols(Si –O-) are “hot sites”that can lead to tailing

Si

Si

SiSi

O

O

OO

OSi SiO

SiO

O

Si

O

OHOH OH

HO

OHOH

OHOH

Microporous Silica Particles

Aggregate of Particles Sponge-like Structure

2006/9/7 WenHsin Huang, NDMC 76

LC –Adsorption Chromatography•Solvent molecules compete with solute molecules

for sites on SP

•Solute is displaced by solvent molecule

SolventSolute

MPFlow SP

Fig. 21-10

Page 20: 09 11 2006 Chromatography

WenHsin Huang

NDMC 20

2006/9/7 WenHsin Huang, NDMC 77

LC –Adsorption Chromatography

•Forcible desolvation of solute by solvent is nearlyindependent of solute identity–Dependent on solvent identity

•Eluent strength (o)–Ability of solvent to displace solute

•Elutropic series–Relative eluent strengths of common solvents–Must know about relative polarity of SP/MP

•Eluent strength increases as MP becomes morelike SP

2006/9/7 WenHsin Huang, NDMC 78

Isocratic vs. Gradient Elution

•Isocratic elution–One solvent or one constant solvent mixture used as

MP throughout entire separation•Gradient elution

–Adjust solvent mixture through separation–Adjusting eluent strength–May speed up separation process

•Table 21-2 includes o values of solvents foradsorption chromatography on silica–polarity, o

2006/9/7 WenHsin Huang, NDMC 79

Isocratic vs. Gradient Elution

Start w/ 100% Benzenethen add Acetonitrile gradient

elution

Hexane °=0.01

Acetonitrile °=0.52

isocraticelutions

Fig. 21-18&

Fig. 21-19

2006/9/7 WenHsin Huang, NDMC 80

Isocratic vs. Gradient Elution

•Previous example illustrated balance of goodresolution and reasonable retention times

•Gradient elution may help–Improve resolution–Shorten retention times (and total analysis time)–Improve peak shapes

Page 21: 09 11 2006 Chromatography

WenHsin Huang

NDMC 21

2006/9/7 WenHsin Huang, NDMC 81

LC –Partition Chromatography•Liquid SP coated on solid support

–Often, liquid SP covalently attached to surfaceof silica gel particle

Si O Si R

CH3

CH3

support liquid SP

Common polar R groups Common nonpolar R groups

(CH2)3NH2 (CH2)17CH3 (or C18)

(CH2)3CN (CH2)7CH3 (or C8)(CH2)2OCH2CH(OH)CH2OH (CH2)3C6H5 2006/9/7 WenHsin Huang, NDMC 82

LC Phases•Normal-phase Chromatography

–SP polar–MP weakly polar or non polar

–More polar solvent has higher o

•More attracted to SP and able to displace solute

•Reversed-phase Chromatography–SP nonpolar or weakly polar–MP polar

–Less polar solvent has higher o

•More attracted to SP and able to displace solute

2006/9/7 WenHsin Huang, NDMC 83

LC Phases

•Normal-phase developed first•Reversed-phase is more common

–Many choices of polar solvents in which solutesare commonly soluble•Water, methanol

–Nonpolar SP less likely to have hot sites•Less peak tailing

2006/9/7 WenHsin Huang, NDMC 84

HPLC Instrumental Design

•Already covered columns•Focus now on

–Injector–Detectors

I

Pump

Column Detector

Injector

Page 22: 09 11 2006 Chromatography

WenHsin Huang

NDMC 22

2006/9/7 WenHsin Huang, NDMC 85

HPLC 6-way Injection Valve

column

Sample loop Sample loop

waste waste

Fig. 21-17

2006/9/7 WenHsin Huang, NDMC 86

Detectors

•Questions to ask when evaluating an LCdetector:–Is the detection universal?–Is the detector response linear?–What is the limit of detection (LOD)?–Is the detector useful with gradient elution?

•In other words, is the detector insensitive to solventcomposition?

2006/9/7 WenHsin Huang, NDMC 87

UV Detector (w/ flow cell)•Common HPLC detector

–Many solutes absorb UV light

Fig. 21-20

Lightsource

Eluate out

Eluate in

Detector

pathlength

Flow cell

2006/9/7 WenHsin Huang, NDMC 88

UV Detector

•Will cover during spectrophotometry–Any UV spectrophotometers discussed previously are

good LC detectors

•UV cutoff wavelengths–Below given , MP absorbs UV radiation–MP absorbance overwhelms analyte absorbance

Page 23: 09 11 2006 Chromatography

WenHsin Huang

NDMC 23

2006/9/7 WenHsin Huang, NDMC 89

UV Detector

•Applicable for UV chromophoresEssentially linear

–Will discuss deviations from linearity in CH 19

LOD ~ 0.1 ngGood detection method with gradient

elution–Use solvents that do not absorb UV radiation

2006/9/7 WenHsin Huang, NDMC 90

Fluorescence Detector

•Similar idea to UV detector•Few molecules fluoresce•Derivatize solutes with fluorescent tag

–Derivatize mixture of solutes before separation–Derivatize solutes as they elute from column

before detection•Post-column Derivatization

2006/9/7 WenHsin Huang, NDMC 91

Fluorescence Detector

Applicable to fluorophoresEssentially linear

–Will discuss deviations from linearity

LOD ~ 0.001 ngFine with gradient elution

–Use solvents that do not fluoresce

2006/9/7 WenHsin Huang, NDMC 92

Refractive Index Detector•Compare refractive indices of eluate & reference

•Cannot use with gradient elution–reference composition cannot exactly match solvent

composition during experiment

(Reference) (MP solvent & analyte = eluate)

Page 24: 09 11 2006 Chromatography

WenHsin Huang

NDMC 24

2006/9/7 WenHsin Huang, NDMC 93

Refractive Index Detector

Essentially universalLOD ~ 100 ngEssentially non-linear

–Linear range very small

Disaster with gradient elutionFinicky

–Need constant T

2006/9/7 WenHsin Huang, NDMC 94

Comparison of LC DetectorsType Approx LOD Approx

Linear rangeComments

UV-VIS 10 -10 g 10 4 For light absorbingcompounds

Fluorescence 10-14 g 10 5 For Fluorescentcompounds

MS 10-7-10-9 g 10 5 Universal detector

Electrochemicalconductometric

10-8 g/mL 10 5 Specific detector; forall ions

Electrochemicalamperometric

10- 10-10 –11 g 10 5 Specfic detector;electroactvecompounds

2006/9/7 WenHsin Huang, NDMC 95

Liquid Chromatography (LC)

Especially high-performance liquid chromatography (HPLC).

The term “high-performance” refers to the use of packed columns with very smallpacking particles (diam. 5-10 m) -giving greatly enhance resolution.

Note: several types of LC. Inaddition to partition (as describedso far) - there are also ion-exchange and size-exclusionchromatography using liquidmobile phases. We will concentrateonly on partition.