08 FMM-420

download 08 FMM-420

of 21

Transcript of 08 FMM-420

  • 7/31/2019 08 FMM-420

    1/21

    FMM-420 MODULE

    PROGRAMMING UPDATE

  • 7/31/2019 08 FMM-420

    2/21

    FMM-420 MODULE

    Before we can connect a device to our SLC, it

    makes sense to understand the operating

    characteristics of a 4-20 device

    In most cases, the device will have 3 wires

    that connect to the FMM-420 Module.

    The wires consist of:

    Negative (-)

    Sense

    Positive (+)

  • 7/31/2019 08 FMM-420

    3/21

  • 7/31/2019 08 FMM-420

    4/21

  • 7/31/2019 08 FMM-420

    5/21

  • 7/31/2019 08 FMM-420

    6/21

  • 7/31/2019 08 FMM-420

    7/21

    Select FMM 4-20

    FMM-420 MODULE

  • 7/31/2019 08 FMM-420

    8/21

    After selecting FMM-420 Type Code

    CBE-3, 4, 5, 6 and 7 are highlighted.

    We will look at specific CBE programming later in this presentation.

    FMM-420 MODULE

  • 7/31/2019 08 FMM-420

    9/21

    FMM-420 MODULE Programming

    STEP 1

    Specific Device Information

    STEP 2Device Thresholds

    STEP 3

    Device CBEs

  • 7/31/2019 08 FMM-420

    10/21

    FMM-420 MODULE

    4 mA 20 mA

    Operating Range is: 4 mA to 20 mA

    0 % 100 %

    4 mA represents

    0 % in this case

    20 mA represents

    100 % in this case

    0100

  • 7/31/2019 08 FMM-420

    11/21

    Units is the measurement unit of the device:

    - Parts Per Million (PPM)

    - Parts Per Billion (PPB)

    - etc.

    FMM-420 MODULE

    UNITS is a required field for the device. This value is used as a

    programming reference and is not used for FACP computations.

    0100

    PPM

  • 7/31/2019 08 FMM-420

    12/21

    If the measurement range is from 0 to 100, you may choose to have a

    Hysteresis percentage range of 5.

    If so, 9 would be the low end and 14 would be the high end of the

    moving percentage range as shown in the graphic below.

    4 mA 20 mA

    0 9 14 100

    5

    Hysteresis is a percentage range

    along the 4 mA to 20 mA scale

    FMM-420 MODULE

    0100

    PPM

    5

  • 7/31/2019 08 FMM-420

    13/21

    0100PPM

    5

    When you are satisfied with the Device Parameters

    Click the THRESHOLDS button:

    FMM-420 MODULE

  • 7/31/2019 08 FMM-420

    14/21

    FMM-420 MODULE

    A blank Threshold screen appears when clicking the Threshold button

    When planning your programming in the threshold screen, it is important

    to know what Values will be significant for the device you have installed,

    because the Value will correspond directly with a CBE

    Before inserting a Value, you must check the CBE box on the left.

  • 7/31/2019 08 FMM-420

    15/21

    FMM-420 MODULE

    For this example, I will check ALL the CBEs

    Once a CBE box is checked, Latching is automatically checked. If you

    do NOT want the Value to trigger a latching event, uncheck Latching.

    Keep in mind that CBE 3, 4, 5, 6 and 7 are automatically assigned to

    Thresholds 1 5.

  • 7/31/2019 08 FMM-420

    16/21

    FMM-420 MODULE

    At this point in my programming, I have decided to use all the CBEs

    and a fairly even range of 15, 30, 45, 60 and 75

    These values are what my device will read in its chamber or sensor

    Upon this value being recognized, I must decide what EVENT will

    correspond with the value

  • 7/31/2019 08 FMM-420

    17/21

    FMM-420 MODULE

    Note that I have several options for my event types.

    Since Fire Codes dont allow us to generate a Fire Alarm by

    using a GAS detection device, I will select Non-Fire Trouble

  • 7/31/2019 08 FMM-420

    18/21

    FMM-420 MODULE

    When Non-fire Trouble was selected, Trouble 1 automatically appeared

    in the Trouble Type field. However, before we can set the Trouble Type,

    we must first set our Trigger field.

    The choice Specific Level, simply looks at the SPECIFIC Level of your Value.Upper Limit, is when the Value is reached in an upward direction

    and Lower Limit is when the Value is reached in a downward direction.

    For this example, I will use Upper Limit

  • 7/31/2019 08 FMM-420

    19/21

    FMM-420 MODULE

    Once the TRIGGER Value has been set, we can select a Trouble Type

    Our choices are: Trouble 1, Trouble 2 and Device Inhibit

    If we choose Device Inhibit, when the FMM-420 is in a self-calibration

    state, a trouble will be generated at the FACP.

  • 7/31/2019 08 FMM-420

    20/21

    FMM-420 MODULE

    REMEMBER

    Without AHJ approval

    Gas Sensing Devices are NOT to be used to indicate a Fire Condition

    Z15

    Z45

    Z30

    Z75

    Z60

    For this example, I have used the Values for the Zone numbers.

  • 7/31/2019 08 FMM-420

    21/21

    Questions