    1/4

    Most medical devices require authorization by the

    Food and Drug Administration (FDA) under the

    Federal Food, Drug, and Cosmetic Act (FDCA) in

    order to be legally marketed in the United States. Such author-

    ization may be clearance o a premarket notication (510(k)

    clearance), de novo classication, or approval o a premarket

    application (PMA). Each o these, in turn, may require varying

    levels o supporting bench or clinical data, depending upon the

    type o device and its intended use.

    A sound regulatory strategy requires an understanding o

    the pathway to market and data requirements applicable to

    a proposed device. Tere are several ways to obtain advance

    inormation rom FDA on these issues. Each approach has its

    strengths and weaknesses. It is worth knowing the ways inwhich FDA can be approached, the pros and cons o each type

    o approach, and how to improve communications with FDA

    in order to gain the best possible understanding o the require-

    ments that will likely pertain to your proposed device. Te

    discussion below will address the available options.

    FDAs WebsiteTe Device Centers website (www.da.gov/cdrh) is a good place

    to start the research, because it is quick, ree and ofen quite

    useul. Te 510(k) database can be used to search or potential

    predicate devices. In many cases, there will be a 510(k) sum-

    mary that can provide inormation on other predicate devices

    and the data provided to support clearance. Unortunately, this

    inormation is ofen presented in airly general terms. Nonethe-

    less it can still be helpul. For example, the 510(k) summary will

    usually provide at least an indication as to whether clinical data

    were required. In rare cases, FDA also may post its 510(k) deci-

    sion memo, which will be somewhat more detailed.

    I useul predicate devices are identied, it is possible to obtain

    the underlying 510(k) submission via a Freedom o Inorma-

    tion Act request to FDA.1 Unortunately, it is unpredictable how

    long FDA will take to respond and it may take more time than

    is practical. It is a good idea to check whether FOI Services (aprivate rm) has already obtained the 510(k) in question (www.

    oiservices.com), in which case it can be downloaded or a ee.

    by Jefrey K. Shapiro

    Mr. Shapirois a Director at the law rm of

    Hyman, Phelps & McNamara PC

    in Washington, DC.

    The Pathway to Market forYour Medical Device:A Primer on Obtaining Information from FDA

    w w w . f d l i . o r gUpdate May/June 2008

    Medical DevicesUpdate 2008, Issue 3With Permission from FDLI, www

    http://www.fdli.org/updatehttp://www.fdli.org/updatehttp://www.fdli.org/http://www.fdli.org/http://www.fdli.org/update
  • 7/30/2019 04 Update Shapiro

    2/4

    May/June 2008 UpdateFDLI

    Medical Devices

    FDAs guidance document data-

    base also should be reviewed. In some

    instances, FDA has published guidance

    that will reveal the likely pathway (e.g.,

    510(k) v. PMA) and/or the typical data

    required to support a 510(k) submissionor PMA ling. Tese documents, when

    available, usually will provide a detailed

    map o the path to market.

    I a proposed device is in Class III

    and approval has already been granted

    to other devices in the same class, FDAs

    PMA database may be useul. Every

    approved device will have a summary

    o saety and eectiveness (SSE) posted

    in the database discussing in detail the

    approved labeling or the device and the

    preclinical and clinical data submitted to

    support approval. Even i the proposed

    device has somewhat dierent technol-

    ogy or intended use (thus, potentially

    altering the data requirements), a prior

    SSE will provide a wealth o inormation

    about the road ahead.

    Informal ConsultationAfer making the most o FDAs website,

    it may be appropriate to contact FDA

    ocials to obtain more inormation. A

    useul approach is to identiy ocials

    who would likely review a 510(k) or PMA

    or the proposed device.

    In the Device Centers Oce o Device

    Evaluation (ODE), there are ve divi-

    sions, each encompassing a varying

    number o branches that review spe-

    cic types o devices. For example, the

    Division o Reproductive, Abdominal,and Radiological Devices has a Division

    Director, a Deputy Director and our

    branch chies. I your device were intend-

    ed or obstetric use (as an example), you

    might call or e-mail Colin Pollard, cur-

    rently chie o the Obstetrics/Gynecology

    Devices Branch, to initiate discussions.

    I the device is an in vitro diagnostic,

    then the Oce o In Vitro Diagnostic

    Device Evaluation and Saety (OIVD)

    is the analog o ODE. Tis oce, too, is

    organized by division. However, instead

    o searching or a branch chie, you

    would look or the relevant Associate

    Director. For example, in the Divisiono Chemistry and oxicology Devices,

    the Associate Director or Chemistry is

    currently Carol Benson.

    I the device is intended to be used

    in combination with drugs or biologics,

    there may be jurisdictional questions

    as to which Center (Drug, Device or

    Biologics) has primary jurisdiction.2

    Tese issues can be discussed inormally

    with the Oce o Combination Products

    (OCP) and/or the product jurisdiction

    ocers attached to each Center.3

    Inormal contacts with agency o-

    cials will yield inormation or opin-

    ions that are not legally binding upon

    the agency.4 Furthermore, when FDA

    responds to questions, the answers

    are only as valid as the inormation

    provided as the basis or the questions.

    Tus, it is critical to provide accurate

    and complete inormation to your con-tact at FDA. Despite these limitations,

    inormal contacts with FDA are usually

    the least expensive and astest way to

    elicit a great deal o helpul advice.

    PreIDE ProgramFDAs preIDE program is the agencys

    basic mechanism or interactive presub-

    mission communications with industry.

    Te current Director o ODEs Investiga-

    tional Device Exemption (IDE) Program,Captain Stephen P. Rhodes, USPHS, has

    estimated that ODE and OIVD have

    a combined average o more than 400

    meetings per year, not including tele-

    phone or e-mail interactions.5

    Te benet o this extensive outreach

    eort is to help industry with design

    testing and development plans that will

    expedite review and approval. Early

    interaction also allows FDA personnel

    to amiliarize themselves with the new

    technologies. Although there is a cost in

    time and money to both agency and in-

    dustry, the benet is generally perceived

    to exceed the cost.Tere are many possible time points

    in the premarket process where industry

    may request advice. Prior to proo o

    concept, it is possible to discuss concepts,

    testing plan (in broad outline), bench

    test methodologies, possible regulatory

    pathways, and potential combination

    product issues. ypically, FDA considers

    it premature to discuss these matters in

    a ace-to-ace meeting, so telephone and

    e-mail contact are generally oered.

    At the preclinical and clinical phases,

    FDA encourages preIDE meetings

    and teleconerences. At the preclinical

    phase, FDAs eedback may be sought on

    bench testing plans, animal study pro-

    tocols, and easibility study protocols. In

    addition, FDA may be willing to provide

    preliminary eedback on the regulatory

    pathway and guidance as to whether an

    IDE approval will be required or anyclinical study (i.e., whether the device

    qualies as signicant risk under the

    IDE regulations).6

    At the clinical phase, FDAs eedback

    may be sought on the need or urther

    bench/animal studies, the need or a pilot

    study prior to initiating the pivotal study,

    proposed study protocols, the regulatory

    pathway, and proposed indication or

    use. In addition, FDA may be willing toprovide preliminary eedback on issues

    such as whether the device qualies or

    expedited status and the need or an

    advisory panel meeting.

    o obtain a preIDE meeting, a rm

    must submit a complete pre-meeting

    package that includes a device descrip-

    tion, pre-clinical test plan/results,

    clinical/statistical test plan, and other

  • 7/30/2019 04 Update Shapiro

    3/4

    8 w w w . f d l i . o r gUpdate May/June 2008

    Medical Devices

    inormation necessary or FDA to un-

    derstand the device and the issues or

    which eedback is sought. It is essential

    to prepare a list o ocus questions or

    which FDAs eedback is sought. Te

    process o preparing this list o questions

    will help the rm determine the key out-

    standing scientic and regulatory issues.

    Te list will also guide FDAs preparation

    and may help prevent the meeting rom

    wandering o course.

    Some Divisions have their own

    checklists and guidelines as to what

    must be in the pre-meeting package,

    so it is important to inquire. Once the

    package is submitted to FDA, a project

    manager is assigned and a meeting will

    be scheduled. ypically, these meetingsare scheduled to occur 4 to 6 weeks afer

    submission. I signicant new inorma-

    tion becomes available afer the package

    is submitted (e.g., device modications,

    changes to study design), it may be nec-

    essary to reschedule the meeting. As the

    meeting approaches, it will be necessary

    to provide a proposed agenda to FDA

    and exchange lists o attendees.

    Te preIDE process is typically

    ocused on study design and testing

    issues. Yet, it may be the best op-

    portunity to obtain FDAs eedback

    on the regulatory pathway to market.

    Tis regulatory issue is intimately

    linked to study design, since studies to

    support PMA approval are generally

    more extensive than those supporting

    510(k) clearance or de novo classica-

    tion. Nonetheless, the regulatory issues

    sometimes are lost in the scienticdiscussion o study design. Tereore,

    unless it is clear that a proposed device

    will require PMA approval, it is a good

    idea to provide FDA in the preIDE

    package with a specic analysis o the

    proposed pathway to market (e.g., pro-

    posed predicate devices and substantial

    equivalence argument) and to include

    at least one ocus question asking or

    FDAs eedback on this issue.

    FDAs preIDE program is not itera-

    tive. Generally, FDA will grant a single

    meeting or the same topic. Tus, it is

    important to think through careully the

    timing o the meeting and its objectives.

    You should view it as your meeting

    in the sense that you are responsible or

    determining what questions need to be

    answered, or ensuring that FDA actually

    answers those questions, and or seeking

    clarication or elaboration when FDAs

    answers do not seem clear or complete.

    It is important to bring the right team

    to the meeting with all o the expertise

    necessary to discuss the ocus questions

    you have developed.

    FDAs advice at a preIDE meeting isnot legally binding and may change in

    light o technological and regulatory

    evolution. Tus, it is important to main-

    tain contact with FDA on these issues i

    your development program extends over

    several years.

    Finally, i a binding commitment is

    needed, FDA by statute oers deter-

    mination7 and agreement8 meet-

    ings. In general, the ormer is or PMA

    devices and allows a written request to

    or a determination o the type o valid

    scientic evidence that will be neces-

    sary to demonstrate that the device is

    eective or its intended use. Te latter

    generally allows a binding agreement

    between FDA and the sponsor or ap-

    plicant regarding the parameters o an

    investigational plan (including a clinical

    protocol).9 Tese two types o meetings

    are not widely used, but they may be ap-propriate in some circumstances.

    The 510(k) ProgramFDA regulates devices under a risk-based

    scheme, in which devices determined

    to be substantially equivalent to law-

    ully marketed Class I and II devices are

    cleared to market subject to postmarket

    controls. A 510(k) review is, in essence,

    a classication proceeding during which

    FDA determines whether a proposed de-

    vice is substantially equivalent (SE) or not

    substantially equivalent (NSE) to a Class I

    or II device already cleared to market.10

    FDA will nd a device SE i it has the

    same intended use and either the same

    technological characteristics as a predi-

    cate device or dierent technological

    characteristics that 1) do not raise new

    types o questions o saety and eective-

    ness, and 2) data are provided to dem-

    onstrate that the proposed device is at

    least as sae and eective as the predicate.

    As a practical matter, this review does

    examine saety and eectiveness to a de-

    gree, at least to the extent o determining

    equivalence to a device already cleared tomarket. It is not, however, a ull review o

    saety and eectiveness. As a legal matter,

    FDAs issuance o a clearance letter is

    actually an order classiying the device

    into Class I or II based upon substantial

    equivalence.

    I a proposed device is ound NSE, it

    will generally require a PMA approval or

    de novo classication. 11 Tis may happen

    i the proposed device does not have a

    predicate device, presents a new intended

    use, or has dierent technological charac-

    teristics compared to the predicate device

    and raises new questions o saety or e-

    ectiveness. FDA can make the oregoing

    determinations without a review o data.

    In addition, FDA may nd a device NSE i

    the data provided do not demonstrate that

    the device is at least as sae and eective

    as the predicate device. In this situation,

    it may be possible to generate appropriatedata to ultimately support clearance o a

    new 510(k) submission rather than move

    to the de novo or PMA pathways.

    An important implication o the

    oregoing discussion is that the 510(k)

    process itsel may be an appropriate way

    to obtain FDAs ruling on the classica-

    tion o the device, without the necessity

    to conduct a study and provide data. I

  • 7/30/2019 04 Update Shapiro

    4/4

    May/June 2008 UpdateFDLI

    Medical Devices

    a 510(k) is submitted and ound to be

    NSE or lack o data, a new 510(k) may

    be submitted afer the appropriate test-

    ing has been perormed. I the 510(k) is

    submitted and ound to be NSE or the

    other reasons mentioned above, that is

    a denitive ruling that the device is in

    Class III, requiring de novo classication

    or PMA approval.

    A 513(g) RequestAn option or obtaining written advice

    about the classication o a device is a

    513(g) request.12 It is named or Section

    513(g) o the FDCA, which provides:

    Within 60 days o the receipt

    o a written request o any person

    or inormation respecting theclass in which a device has been

    classied or the requirements

    applicable to a device under this

    Act, [FDA] shall provide such

    person a written statement o the

    classication (i any) o such de-

    vice and the requirements o this

    Act applicable to the device.

    A 513(g) request is airly simple to

    prepare. Generally speaking, it should

    include a cover letter, a complete devicedescription, an Indications or Use

    statement, and either proposed labeling

    or labeling o a similar device already on

    the market. A user ee is now required

    or each 513(g) request.13 Within 60 days,

    and sometimes sooner, FDA will provide

    a written response that is binding unless

    amended or revoked in writing.14 Te

    advice in response to a 513(g) request

    only covers the device as described in theinitial request. I the device is modied

    (or was not accurate in the rst place),

    the 513(g) response may not apply.

    A 513(g) request can be used or a wide

    range o inquiries. Tese include whether

    a product is subject to FDA regulation at

    all, whether it is regulated as a device (ver-

    sus a cosmetic, biologic or drug), whether

    it is within an exemption rom 510(k)

    requirements, whether a 510(k) is needed

    or a modication to the device, and

    whether a device with a new technology

    or intended use must undergo a 510(k) or

    PMA pathway to market.

    Tere has sometimes been uncertainty

    about whether a particular stand alone

    sofware product is subject to medical

    device regulation. FDA has withdrawn

    its 1989 draf policy statement, FDA

    Policy or the Regulation o Computer

    Products. Tis guidance had never

    been nalized but had provided opera-

    tional policy guidance to industry and

    agency sta or many years. Some agency

    ocials have said that the principles em-

    bodied in the draf guidance still apply

    in most cases. Nonetheless, FDA has no

    current written guidance on the regula-

    tion o stand alone sofware products.

    A 513(g) request can be used to elicit

    ormal advice as to whether a particular

    stand alone sofware product will be

    regulated as a medical device and the

    requirements that will apply.

    Request for DesignationA combination product raises the

    question about which Center will have

    primary jurisdiction (i.e., Drug, Device

    or Biologic). A Request or Designa-

    tion (RFD) can be submitted to the

    OCP seeking a written ruling as to which

    center will have primary jurisdiction.

    Tis process and the requirements or an

    RFD are set orth in 21 C.F.R. Part 3. Te

    OCP must make its written jurisdictional

    determination within 60 days o ling.

    Tere is no user ee or an RFD.15

    Final ThoughtsIt is critical to obtain inormation rom

    FDA on the requirements applicable to a

    proposed device. It will typically require

    signicant time and unding resources

    to do so. Tereore, when developing a

    business and regulatory strategy, it is es-

    sential to build in this inormation gath-

    ering process. Tis is particularly true

    or novel or complex devices, or which

    it is virtually impossible to successully

    conduct a clinical study and prepare an

    appropriate 510(k) submission or PMA

    ling without consulting FDA in ad-

    vance. It is essential to develop a careul

    strategy or communicating with FDA

    using the right tools at the right time in

    the development process. Hopeully, the

    oregoing discussion will help get you

    started in ormulating a successul strat-

    egy that wi ll result in a smooth pathway

    to market or your device.

    1 5 U.S.C. 552; 21 C.F.R. Part 20.

    2 Te Device Centers ormal name is the Center or

    Devices and Radiological Health (CDRH). Te Drug

    Centers ormal name is the Center or Drug Evalu-ation and Research (CDER). Te Biologics Centers

    ormal name is the Center or Biologics Evaluation

    and Research (CBER).

    3 Currently, these are: Stephen Rhodes, CAP,

    USPHS, CDRH; Virginia Behr, CDER; and Sheryl

    Lard-Whiteord, PhD, CBER.

    4 21 C.F.R. 10.85(k).

    5 Slide presented during AdvaMed AudioConerence

    (Oct. 17, 2007).

    6 21 C.F.R. 812.3(m).

    7 FDCA, 513(a)(3)(D).

    8 FDCA, 520(g)(7).

    9 FDAs descript ion o these two types o meetings can

    be ound in Early Collaboration Meetings

    Under the FDA Modernization Act (FDAMA); FinalGuidance or Industry and or CDRH Sta.

    10 Substantia l equivalence may also be claimed to a

    pre-amendment device (on the market prior to May

    28, 1976), or which PMA applications have not been

    called. Tere are also Class I and II devices that are

    exempt rom the 510(k) requirement.

    11 In de novo classication under Section 513()(2) o

    the FDCA, FDA essentially creates a new classica-

    tion or novel low risk devices that lack a predicate

    device. Te process requires an additional 60 days,

    and usually includes the development o special regu-

    latory controls or the device. Once the classication

    exists, uture devices may seek 510(k) clearance based

    upon substantial equivalence to the de novo device.

    12 An original and one copy o the 513(g) request shouldbe mailed to the 513(g) Coordinator, Oce o Device

    Evaluation, Center or Devices and Radiological

    Health c/o Document Mail Center (HFZ-401), 9200

    Corporate Boulevard, Rockville, Maryland 20850.

    Te current contact person is Lawrence J. (Jake)

    Romanell, Program Operations Sta, ODE, CDRH.

    13 In Fiscal Year 2008, the standard ee is $2,498 and the

    small business discounted ee is $1,249.

    14 See 21 C.F.R. 10.85(e).

    15 For a detai led treatment o combination products,

    see Fox and Shapiro, Combination Products: How

    to Develop the Optimal Strategic Path or Approval

    (FDANews 2006) (www.danews.com).