03 Reference Standards

download 03 Reference Standards

of 10

description

farmasi

Transcript of 03 Reference Standards

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    1

    Kuliah Validasi Metode Analisis

    Reference StandardsFlorentinus Dika Octa Riswanto, M.Sc.

    Fakultas Farmasi

    Universitas Sanata Dharma

    2014

    1

    HPLC methods rely on reference standards in order to provide accurate data.

    The quality and purity of reference standards are critical and these materials should be highly purified and well characterized.

    1. Reference standards from the USP/NF

    2. Non-compendial primary reference standard

    3. Working reference standards

    Reference Standards

    Kuliah Validasi Metode Analisis 2014

    2

    Ahuja and Dong, 2005, Handbook of pharmaceutical analysis by HPLC, p. 212.

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    2

    "pure" materials by legal definition

    unless a purity value other than 100%

    has been assigned by the USP,

    and no further characterization is

    required prior to their use.

    Kuliah Validasi Metode Analisis 2014

    Reference standards from the USP/NF

    3

    If USP reference standards are not available,

    a non-compendial primary reference standard may be certified for use

    (commercially supplied RS obtained from a reputable commercial source)

    These should be of the highest purity available and well characterized

    to assure the purity, strength, identity, and quality of the material.

    Methods using non-compendial reference standards

    must incorporate any purity correction factor into the calculations.

    It is important that any and all reference materials used

    for method validation should be well documented in the validation report.

    Kuliah Validasi Metode Analisis 2014

    Non-compendial primary reference standard

    4

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    3

    Kuliah Validasi Metode Analisis 2014

    5

    Working reference standards are usually materials that have been characterized

    and have had their purity established against a compendial reference standard.

    These may be used in cases where it is more cost effective

    to certify an in-house lot than to purchase USP reference materials

    for routine analysis.

    Kuliah Validasi Metode Analisis 2014

    Working reference standards

    6

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    4

    1. Quantitation by External Standard

    This quantitation technique is the most straightforward.

    It involves the preparation of one or a series of standard solutions that approximate the concentration of the analyte.

    Chromatograms of the standard solutions are obtained, and peak heights or areas are plotted as a function of concentration of the analyte.

    The plot of the data should normally yield a straight line. This is especially true for pharmaceuticals of synthetic origin.

    Other forms of mathematical treatment can be used but will need to be justified.

    Kuliah Validasi Metode Analisis 2014

    Types of quantitation

    7

    Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.

    1. Quantitation by External Standard

    There are some potential instrumental sources of error that could occur using this quantitation technique.

    It is critical to have minimal variability between each independent injection, as the quantitation is based on the comparison of the sample and standard areas.

    However, the current autosamplers are able to minimize this variability to less than 0.5% relative standard deviation (RSD).

    Kuliah Validasi Metode Analisis 2014

    Types of quantitation

    8

    Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    5

    Kuliah Validasi Metode Analisis 2014

    External Standard

    9

    C1

    C4

    C3

    C2

    ConcentrationArea

    A1

    A2

    A3

    A4C1 C2 C3 C4

    A1

    A2

    A3

    A4

    Concentration

    Peak a

    rea

    Calibration curve

    2. Quantitation by Internal Standard

    Quantitation by internal standard provides the highest precision because uncertainties introduced by sample injection are avoided.

    In this quantitation technique, a known quantity of internal standard is introduced into each sample and standard solutions.

    As in the external standard quantitation, chromatograms of the standard and sample solutions are integrated to determine peak heights or peak areas.

    The ratio of the peak height or area of the analyte to an internal standard is determined. The ratios of the standards are plotted as a function of the concentration of the analyte. A plot of the data should normally yield a straight line.

    Kuliah Validasi Metode Analisis 2014

    Types of quantitation

    10

    Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    6

    2. Quantitation by Internal Standard

    Due to the presence of the internal standard, it is critical to ensure that the analyte peak be separated from the internal standard peak.

    A minimum of baseline separation (resolution >1.5) of these two peaks is required to give reliable quantitation.

    In addition, to quantitate the responses of internal standard accurately, the internal standard should be baseline resolved from any significant related substances and should have a peak height or area similar to that of the standard peak.

    Kuliah Validasi Metode Analisis 2014

    Types of quantitation

    11

    Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.

    Kuliah Validasi Metode Analisis 2014

    Internal Standard

    12

    C1

    C4

    C3

    C2

    Concentration Area

    A1

    A2

    A3

    A4C1/CIS C2 /CIS C3 /CIS C4 /CIS

    A1/AIS

    A2 /AIS

    A3 /AIS

    A4 /AIS

    Concentration of target substance /

    Concentration of internal standard

    Are

    a f

    or

    targ

    et

    substa

    nce / A

    rea for

    inte

    rnal sta

    ndard

    Calibration curveTarget

    substance

    Internal

    standard

    CIS

    CIS

    CIS

    CIS

    AIS

    AIS

    AIS

    AIS

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    7

    Kuliah Validasi Metode Analisis 2014

    Advantages of Internal Standard Method

    13

    Not affected by inconsistencies in injection volume.

    10 L

    injected

    9 L

    injected

    CX / CIS

    AX / AIS

    XIS

    XIS

    Same area ratio

    Kuliah Validasi Metode Analisis 2014

    Advantages of Internal Standard Method

    14

    Not affected by the pretreatment recovery rate.

    100%

    recovery

    rate

    90%

    recovery

    rate

    CX / CIS

    AX

    / A

    IS

    XIS

    XIS

    Same area ratio

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    8

    It is preferable to choose an internal standard which has a similar chemical structure to the analyte in order to mimic its behaviour.

    Its peak must appear relatively near that of the target substance.

    It must not already be contained in the actual samples.

    Its peak must be completely separated from those of other sample components.

    It must be chemically stable.

    It must give a similar respon to the same detector

    Available in high purity

    Kuliah Validasi Metode Analisis 2014

    Selection Criteria for Internal Standard

    15

    In many instances in the pharmaceutical industry, drug products may be

    manufactured in a variety of strengths (e.g., levothyroxine tablets in strengths of

    50, 100, 150, 200, 500, and 750 g). To develop and validate these potency

    methods, three strategies may be followed.

    1. Single-Point Calibration

    2. Multiple-Point Calibration

    3. One Standard Calibration for Each Strength

    Kuliah Validasi Metode Analisis 2014

    Standard Plots for Quantitation

    16

    Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 14.

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    9

    A method may be developed and validated using only one standard analyteconcentration.

    The standard plot generated is used to assay the complete range of tablet strengths.

    This strategy should be adopted wherever possible due to the simplicity of standard preparation and minimal work for quantitation of the sample.

    However, this method will require different extraction and dilution schemes of the various drug product strengths to give the same final concentration that is in the proximity of the one standard analyte concentration.

    Kuliah Validasi Metode Analisis 2014

    Single-Point Calibration

    17

    Another strategy involves two or more standard concentrations that will bracket the complete range of the drug product strengths.

    In this strategy it is critical that the standard plots between the two extreme concentration ranges be linear.

    Therefore, this is a valid calibration method as long as the sample solutions of different strengths are prepared within the concentration range of the calibration curve.

    Its advantage is that different strengths can utilize different preparation procedures and be more flexible. Its disadvantage is that multiple weighing of standards at different concentrations may give a weighing error.

    Kuliah Validasi Metode Analisis 2014

    Multiple-Point Calibration

    18

  • Kuliah VMA 2014 - Reference Standards 5/8/2014

    10

    The least favored method is to develop and validate using one standard concentration for each product strength.

    This situation will arise when the analyte does not exhibit linearity within a reasonable concentration range.

    Kuliah Validasi Metode Analisis 2014

    One Standard Calibration for Each Strength

    19

    Kuliah Validasi Metode Analisis 2014

    20

    [email protected]

    @pakdika