03 Reference Standards
-
Upload
dionisius-laffyanto -
Category
Documents
-
view
10 -
download
0
description
Transcript of 03 Reference Standards
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
1
Kuliah Validasi Metode Analisis
Reference StandardsFlorentinus Dika Octa Riswanto, M.Sc.
Fakultas Farmasi
Universitas Sanata Dharma
2014
1
HPLC methods rely on reference standards in order to provide accurate data.
The quality and purity of reference standards are critical and these materials should be highly purified and well characterized.
1. Reference standards from the USP/NF
2. Non-compendial primary reference standard
3. Working reference standards
Reference Standards
Kuliah Validasi Metode Analisis 2014
2
Ahuja and Dong, 2005, Handbook of pharmaceutical analysis by HPLC, p. 212.
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
2
"pure" materials by legal definition
unless a purity value other than 100%
has been assigned by the USP,
and no further characterization is
required prior to their use.
Kuliah Validasi Metode Analisis 2014
Reference standards from the USP/NF
3
If USP reference standards are not available,
a non-compendial primary reference standard may be certified for use
(commercially supplied RS obtained from a reputable commercial source)
These should be of the highest purity available and well characterized
to assure the purity, strength, identity, and quality of the material.
Methods using non-compendial reference standards
must incorporate any purity correction factor into the calculations.
It is important that any and all reference materials used
for method validation should be well documented in the validation report.
Kuliah Validasi Metode Analisis 2014
Non-compendial primary reference standard
4
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
3
Kuliah Validasi Metode Analisis 2014
5
Working reference standards are usually materials that have been characterized
and have had their purity established against a compendial reference standard.
These may be used in cases where it is more cost effective
to certify an in-house lot than to purchase USP reference materials
for routine analysis.
Kuliah Validasi Metode Analisis 2014
Working reference standards
6
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
4
1. Quantitation by External Standard
This quantitation technique is the most straightforward.
It involves the preparation of one or a series of standard solutions that approximate the concentration of the analyte.
Chromatograms of the standard solutions are obtained, and peak heights or areas are plotted as a function of concentration of the analyte.
The plot of the data should normally yield a straight line. This is especially true for pharmaceuticals of synthetic origin.
Other forms of mathematical treatment can be used but will need to be justified.
Kuliah Validasi Metode Analisis 2014
Types of quantitation
7
Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.
1. Quantitation by External Standard
There are some potential instrumental sources of error that could occur using this quantitation technique.
It is critical to have minimal variability between each independent injection, as the quantitation is based on the comparison of the sample and standard areas.
However, the current autosamplers are able to minimize this variability to less than 0.5% relative standard deviation (RSD).
Kuliah Validasi Metode Analisis 2014
Types of quantitation
8
Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
5
Kuliah Validasi Metode Analisis 2014
External Standard
9
C1
C4
C3
C2
ConcentrationArea
A1
A2
A3
A4C1 C2 C3 C4
A1
A2
A3
A4
Concentration
Peak a
rea
Calibration curve
2. Quantitation by Internal Standard
Quantitation by internal standard provides the highest precision because uncertainties introduced by sample injection are avoided.
In this quantitation technique, a known quantity of internal standard is introduced into each sample and standard solutions.
As in the external standard quantitation, chromatograms of the standard and sample solutions are integrated to determine peak heights or peak areas.
The ratio of the peak height or area of the analyte to an internal standard is determined. The ratios of the standards are plotted as a function of the concentration of the analyte. A plot of the data should normally yield a straight line.
Kuliah Validasi Metode Analisis 2014
Types of quantitation
10
Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
6
2. Quantitation by Internal Standard
Due to the presence of the internal standard, it is critical to ensure that the analyte peak be separated from the internal standard peak.
A minimum of baseline separation (resolution >1.5) of these two peaks is required to give reliable quantitation.
In addition, to quantitate the responses of internal standard accurately, the internal standard should be baseline resolved from any significant related substances and should have a peak height or area similar to that of the standard peak.
Kuliah Validasi Metode Analisis 2014
Types of quantitation
11
Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 13.
Kuliah Validasi Metode Analisis 2014
Internal Standard
12
C1
C4
C3
C2
Concentration Area
A1
A2
A3
A4C1/CIS C2 /CIS C3 /CIS C4 /CIS
A1/AIS
A2 /AIS
A3 /AIS
A4 /AIS
Concentration of target substance /
Concentration of internal standard
Are
a f
or
targ
et
substa
nce / A
rea for
inte
rnal sta
ndard
Calibration curveTarget
substance
Internal
standard
CIS
CIS
CIS
CIS
AIS
AIS
AIS
AIS
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
7
Kuliah Validasi Metode Analisis 2014
Advantages of Internal Standard Method
13
Not affected by inconsistencies in injection volume.
10 L
injected
9 L
injected
CX / CIS
AX / AIS
XIS
XIS
Same area ratio
Kuliah Validasi Metode Analisis 2014
Advantages of Internal Standard Method
14
Not affected by the pretreatment recovery rate.
100%
recovery
rate
90%
recovery
rate
CX / CIS
AX
/ A
IS
XIS
XIS
Same area ratio
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
8
It is preferable to choose an internal standard which has a similar chemical structure to the analyte in order to mimic its behaviour.
Its peak must appear relatively near that of the target substance.
It must not already be contained in the actual samples.
Its peak must be completely separated from those of other sample components.
It must be chemically stable.
It must give a similar respon to the same detector
Available in high purity
Kuliah Validasi Metode Analisis 2014
Selection Criteria for Internal Standard
15
In many instances in the pharmaceutical industry, drug products may be
manufactured in a variety of strengths (e.g., levothyroxine tablets in strengths of
50, 100, 150, 200, 500, and 750 g). To develop and validate these potency
methods, three strategies may be followed.
1. Single-Point Calibration
2. Multiple-Point Calibration
3. One Standard Calibration for Each Strength
Kuliah Validasi Metode Analisis 2014
Standard Plots for Quantitation
16
Chan, 2004, Analytical Method Validationand Instrument Performance Verification, p. 14.
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
9
A method may be developed and validated using only one standard analyteconcentration.
The standard plot generated is used to assay the complete range of tablet strengths.
This strategy should be adopted wherever possible due to the simplicity of standard preparation and minimal work for quantitation of the sample.
However, this method will require different extraction and dilution schemes of the various drug product strengths to give the same final concentration that is in the proximity of the one standard analyte concentration.
Kuliah Validasi Metode Analisis 2014
Single-Point Calibration
17
Another strategy involves two or more standard concentrations that will bracket the complete range of the drug product strengths.
In this strategy it is critical that the standard plots between the two extreme concentration ranges be linear.
Therefore, this is a valid calibration method as long as the sample solutions of different strengths are prepared within the concentration range of the calibration curve.
Its advantage is that different strengths can utilize different preparation procedures and be more flexible. Its disadvantage is that multiple weighing of standards at different concentrations may give a weighing error.
Kuliah Validasi Metode Analisis 2014
Multiple-Point Calibration
18
-
Kuliah VMA 2014 - Reference Standards 5/8/2014
10
The least favored method is to develop and validate using one standard concentration for each product strength.
This situation will arise when the analyte does not exhibit linearity within a reasonable concentration range.
Kuliah Validasi Metode Analisis 2014
One Standard Calibration for Each Strength
19
Kuliah Validasi Metode Analisis 2014
20
@pakdika