00462 PTEST DEPLIANTpag1 - Cyclohexanone · English, Italian, German, French HPa(=mBar), kPa, mmHg,...
Transcript of 00462 PTEST DEPLIANTpag1 - Cyclohexanone · English, Italian, German, French HPa(=mBar), kPa, mmHg,...
Partners in medicaldevice manufacturing
Tecnoideal’s P-TEST series is the new generation of electro-pneumatic testing equipment for leak testing of medical
devices such as blood lines, infusion and transfusion sets, bags, filters, oxygenators and many other disposable
products. The operational conditions can be programmed in a wide range of pressures and times. The P-TEST technical
features, along with the new operative interface, makes its use even simpler. Not only it can be used for validation and
qualification purposes on Q.A. and R&D Departments, but mainly in production lines for routine controls of finished
products.
P-TESTLeak Testers
TECHNICAL SPECIFICATIONS
P-TEST Plus 500 P-TEST Plus 15 P-TEST Plus 60
Range of test pressure
Type of tests
Power supply
Electrical protection
Maximum Pneumatic Supply
Pneumatic supply
Accuracy
Decay resolution
Pressure resolution
Pressurization time (T1p)
Stabilization time (T1s)
Test time (T2)
Discharge time (T3)
Serial Ports
PLC interface
Foot pedal connector
Display languages
Pressure units
Password
Memory
Test counter
Acoustic signal
Clock / Calendar
Display
Keyboard
Construction material
Protection
Operating conditions
Dimensions
Weight
Product codes
International Standards
10.0-500.0 hPa(1.00-50.00kPa/7.5-375.0
mmHg/102-5098
mmH2O/0.146-7.252PSI)
30-1500 hPa(3.00-150.00 kPa/23-1125.0
mmHg/310-15300
mmH2O/0.44-21.76 PSI)
120-6000 hPa(12.00-600.00kPa/90-4500
mmHg/1220-61180
mmH2O/1.74-87.05 PSI)
115-230V, 50-60 Hz, 40 VA
Electromagnetic interference filter and 2 1AT fuses
Leak test Leak test Leak test
5 bar 8 bar 10 bar
Dry and filtered air, max. 5µm
0.25%of the full scale
0.01 hPa 0.05 hPa 0.1 hPa
1 hPa0.2 hPa0.1 hPa
1.0 - 999 seconds
1.0 - 999 seconds
1.0 - 999 seconds
1.0 - 999 seconds
No. 1 RS232 + No. 1 RS232/485 optoisolated
PNP, isolated, 10-30 Vdc; 8-pin connector (start/stop/passed/failed)
Different tones and programmable duration, for positive and negative results
3-pin connector
English, Italian, German, French
HPa(=mBar), kPa, mmHg, mmH2O, PSI
8-digit, alfanumeric, programmable on 4 protection levels
Up to 100 test programs, with date/time and alfanumeric 16-digit code
Up to 99,999 )(for goods, rejects for leak, reject for pressure drop, total tests
To set time, date and calibration date
20 numeric (including 4 functions) keys, touch-sensitive
Graphic 128x64 pixels, blue and white, with display of 5 different fonts
Plastic and aluminium external body
Ip40
Temperature from 10 to 40° , relative room humidity max. 80%
16x20x13 cm
2.7 Kg
2006/42/CE, 2004/108/CE, 2006/95/CE
4-digit, alfanumeric, programmable on 4 protection levels
98167 98168 98169
LEAK TEST
SINGLE OBSTRUCTION TEST
MULTIPLE OBSTRUCTION TEST
(ALL MODELS)
(P-TEST ONE and NINE)
(P-TEST ONE and NINE)
P-TEST series are testing equipment for leak testing of
medical devices such as blood lines, infusion and
transfusion sets, bags, filters, oxygenators and many
other disposable products.
The device to be tested is connected to the test port of
the equipment and is pressurized at the preset value.
P-TEST checks the leak by measuring the pressure decay
in the test time.
The equipment can detect obstructions in medical
devices by blowing air through the part to test and
measuring if the pressure given by the flow resistance is
below a preset value.
In devices where many different pathways have to be
detected for the obstruction.
P-Test ONE and NINE can perform multiple obstruction
tests in sequence.
At the end of the multiple obstruction test the
equipment performs a leak test with the preset
conditions.
115-230V, 50-60 Hz, 40 VA
Electromagnetic interference filter and 2 1AT fuses
Leak test, multiple and single obstruction tests
Dry and filtered air, max. 5µm
1.0 - 999 seconds
1.0 - 999 seconds
1.0 - 999 seconds
1.0 - 999 seconds N.A. N.A.
PNP, isolated, 10-30 Vdc; 8-pin connector (start/stop/passed/failed)
Different tones and programmable duration, for positive and negative results
3-pin connector
Plastic and aluminium external body
Ip40
Temperature from 10 to 40° , relative room humidity max. 80%
16x20x13 cm
2006/42/CE, 2004/108/CE, 2006/95/CE
P-TEST ONE P-TEST NINE
50-1013 hPa(38-760 mmHg/0.72-14.70 PSI)
250-9000 hPa(190-6750 mmHg/3.7-130.5
PSI)
8 bar 13 bar
0.5% of the full scale 2.5% of the full scale
0.2 hPa 0.2 hPa
1 hPa1 hPa
No. 1 RS232/485 optoisolated
English
HPa(=mBar), mmHg, PSI
4-digit, alfanumeric, programmable on 4 protection levels
Up to 50,000 )(for goods, rejects, total tests
Up to 10 test programs
Alfanumeric 16 characters, LED backlight
To set calibration date
7 keys, touch-sensitive
2.0 Kg
98155 98175
Manufactured by:
Via Lea Cazzuoli, 43 - I-41037 Mirandola (MO) Italy -Tel.: 0039 0535 23653 - Fax: 0039 0535 27443
Home page: www.tecnoidealsrl.com - e-mail: [email protected]
Marketed in USA, CANADA, MEXICO
Marketed in SOUTH AMERICA by:
Begonienstraße 1 D-90542 Eckental - Brand
Tel. 0049 09126 2948 0 Fax. 0049 09126 294828
home page: www.modenplast.de - E-mail: [email protected]
São Paulo - SP - Brasil
Tel. 0055 011 37172006
home page: www.alvimed.com.br - E-mail: [email protected]
Unit C, Bulding 3, 112 Medical Village Drive, Wallace,NC 28466 - USA
Tel. 001 910 285 5003 - Fax 001 910 285 5021
Home page: www.technomedinc.com - E-mail: [email protected]
Marketed in GERMANY, POLAND, SERBIA, AUSTRIA, CZECH REPUBLIC, HUNGARY and SLOVAKIA by:
Testing of medical plastic bags requires the use of dedicated
holders to contain the plastic deformation due to the test
pressure.
The use of holders is also recommended :
to reduce the test volume;
to increase the test pressure for easier identification of the
leak.
Holders are available in two surface versions:
Flat testing surface: for testing simple bags. Surfase contact is
100% : this is indicated to test the welding perimeter.
Knurled testing surface: designed to test bags in which a leak
on the film can be easily detected because the air escape is not
blocked. Surface contact is 25%.
·
·
·
·
HOLDERS FOR BAG TEST
OPTIONALS - HOW TO ORDER
Holder 300x400 mm, flat test surface 98085
Holder 400x500 mm, flat test surface 98086
Holder 300x400 mm, knurled test surface 98172
Holder 400x500 mm, knurled test surface 98087
Foot switch for P-TEST T70239
USB-RS232 serial adapter for P-TEST ONE/Plus T89106
PTestOneSys (data extraction software) T78076
PtestPlusSys (data extraction software) T78077