© NMISA 2010 Implementing a Shock Calibration System Using a Vibration Exciter and Pendulum...
-
Upload
rosalind-tyler -
Category
Documents
-
view
218 -
download
2
Transcript of © NMISA 2010 Implementing a Shock Calibration System Using a Vibration Exciter and Pendulum...
© NMISA 2010
Implementing a Shock Calibration System Using a Vibration Exciter and Pendulum
Presented byIan Veldman
© NMISA 2010
Presentation overview
• Introduction• System description• Measurement procedure• Uncertainty contributors• Measurement results• Conclusions
© NMISA 2010
Introduction
• Mechanical shock (shock) is the measurand of numerous industry-, testing- and scientific research applications.
• In everyday activities, the peak shock is monitored as part of safety measures.
• The NMISA, as the custodian of South Africa’s measurement standards, developed a shock calibration system to provide industry with traceability to the SI for mechanical shock.
• ISO TC108 has developed ISO 16063-22 for the shock calibration of accelerometers by comparison.
• This standard describes calibration techniques, covering shock calibration of transducers for shock levels up to 1 000 km/s2.
© NMISA 2010
System description
• Depending on the peak shock level, two different shock excitation systems are generally employed.
• For shock signals with pulse durations longer than 3 ms and peak levels below 1,5 km/s2, pendulum systems are used.
• For peak shock levels up to 100 km/s2 with pulse durations as short as 0,05 ms, Hobkinson bar systems are employed.
• The main differences in operation between the two techniques are;– for the pendulum system, the acceleration of the anvil is
measured.– for the Hobkinsonbar system, the shock wave travelling through
the bar is measured.
© NMISA 2010
System description (continue)
© NMISA 2010
System description (continue)
• This mechanical shock calibration system was developed around an anvil, suspended with four 2 m long cables to form the pendulum.
• The pendulum is accelerated by an impulse acting on the suspended anvil.
• The impulse is induced using an electro-dynamic vibration exciter, on which a steel ram with a rounded edge is mounted.
• The peak level of the shock is controlled by the peak amplitude of the electrical signal applied to the exciter.
• The shock pulse duration is varied by using rubber contact surfaces of different composites and thickness on the surface being impacted on by the exciter.
© NMISA 2010
System description (continue)
• The reference transducer with accelerometer to be calibrated is mounted in the “back to back” fashion on the opposite end of the pendulum.
• The peak output amplitudes of both the reference transducer (ûRef) and the unit under test (ûUUT) are recorded using a dual channel analogue to digital converter.
• Knowing the two peak amplitudes, the shock sensitivity of the accelerometer is determined using:
f
UUTfUUT u
uSS
ReRe ˆ
ˆ
© NMISA 2010
Measurement procedure
• The measurement is initiated by sending a shock pulse to the vibration exciter.
• The AtoD is armed for measurement by a trigger pulse being sent to the AtoD trigger channel some 10 ms prior to the shock pulse being sent to the vibration exciter.
• This “pre-trigger” initiates the 500 ms long data capturing period.• A 4th order Butterworth low pass filter is applied to the captured time
series as a smoothing filter to improve the signal to noise ratio.
© NMISA 2010
Measurement procedure– captured time series
• The first peak is the shock pulse as the pendulum is accelerated by the exciter.
• The second peak is the pendulum swinging back.
• The third peak is from the pendulum hitting the exciter on its return.
© NMISA 2010
Measurement procedure– captured time series
• The pulse width of interest is selected by the metrologist with two mouse clicks.
• Using the mouse, the operator can click and drag on an area of interest in order to zoom in on that area.
• The time difference between the two selected points is used to determine the pulse width.
• The difference in amplitude is used to calculate the zero offset.
© NMISA 2010
Measurement procedure
• Once the signal processing is initiated, the peak determination and curve fitting routines are executed.
• The curve is fitted to the top 5% of the signal only as proposed in ISO 16063-22.
• With ûRef and ûUUT known, SUUT is calculated.
• The sensitivity results are recorded in a result sheet using a Microsoft® Excel spread sheet.
© NMISA 2010
Uncertainty contributors
Mathematical model:
• SS = Shock sensitivity
• SRef = Reference accelerometer shock sensitivity
• ûRef = Reference peak voltage
• ûUUT = UUT peak voltage
• GRef = Reference conditioning amplifier gain
• GUUT = UUT conditioning amplifier gain
UUT
UUT
f
ffS G
u
u
GSS
ˆ
ˆRe
ReRe
© NMISA 2010
Uncertainty contributors
• Reference transducer sensitivity– This is the expanded uncertainty reported on the calibration
certificate for the reference accelerometer’s shock calibration.
• Peak voltage determination– Taken as the one year accuracy specification of the 12 bit AtoD
card, using the manufacturer’s online accuracy calculator.– A worse case accuracy taking the widest measurement
possibilities into account was assumed.– It was assumed that peak input will be at least 50% of the
selected input range.
© NMISA 2010
Uncertainty contributors
• Anvil motion/Transverse sensitivity– The anvil’s motion is effectively a circle with a radius equal to the
length of the suspension cables.– This results in non-linearities in the motion of the anvil.– This non-linear motion, combined with the transverse sensitivity
of the reference transducer, results in a systematic error when determining ûRef and ûUUT.
– The maximum transverse motion (aT) was measured using a tri-axial accelerometer and calculated using
yxT aaa
© NMISA 2010
Uncertainty contributors
– The error due to transverse motion in the presence of transverse sensitivity was calculated using
e*T = ST•âT•cosβT•cosφT
• e*T = transverse error
• ST = transverse sensitivity
• âT = transverse acceleration amplitude
• cosβT = angle between the transverse motion direction and direction of transverse sensitivity
• cosφT = phase shift between disturbing voltage and measuring voltage
– A worse case scenario is assumed:• the transverse motion and accelerometer transverse sensitivity is in
phase, cosβT = 1
• the disturbing voltage and measuring voltage is in phase, cosφT = 1
© NMISA 2010
Uncertainty contributors
• Pendulum/Exciter Alignment– The influence of miss-alignment between the pendulum and
vibration exciter was determined experimentally.– The angle of impact between the exciter and pendulum was
increased from 0° to 30 ° in 10° steps.– With the exciter impacting at a 30° angle, the transverse motion
did not increase.– A decrease in the peak shock level was noted, equal to the
cosine of the impact angle.
© NMISA 2010
Uncertainty contributors
• Polynomial fit error– This was calculated as the Root Sum Square (RSS) error
between the fitted values and the peak time series.
• Conditioning amplifier gain– The gain accuracy specified by the manufacturer was used as
the amplifier gain was not calibrated prior to the measurements being performed.
– Calibration of the amplifier can be considered to reduce the over all uncertainty, should it be required.
© NMISA 2010
Uncertainty contributors
• Repeatability– The calibration system is setup to take five measurements per
calibration.– Three accelerometers were calibrated in a variety of
configurations.– A worst case standard deviation of 0,1 % recorded.– A maximum of 0,5 % was used for the type “A” evaluation in the
uncertainty budget.
© NMISA 2010
Measurement results
• The system performance was evaluated using a combination of three accelerometers.
– Two laboratory standard accelerometers, VS-WStd-01 and VS-WStd-09, of the back to back type
– One general purpose accelerometer, VS-01
• All three accelerometers have been calibrated using a laser interferometer system in compliance with ISO 16063-11, method 3.
UUT ReferenceSS
(pC/g)Us(%)
VS-WStd-09 VS-WStd-01 2,037 3,5
VS-WStd-09 VS-WStd-01 2,011 3,5
VS-WStd-01 VS-WStd-09 1,234 3,5
VS-WStd-01 Laser 1,218 3,5
VS-01 VS-WStd-01 67,29 3,5
VS-01 VS-WStd-01 66,94 3,5
VS-01 VS-WStd-09 66,68 3,5
VS-01 Laser 65,80 3,5
© NMISA 2010
Measurement results
• The two laboratory standard accelerometers were calibrated against each other
– using the certified shock sensitivities
– using the sensitivity obtained with sinusoidal calibration.
• VS-01 was calibrated using:1. Using sinusoidal calibration
2. VS-WStd-09 as reference with its shock sensitivity
3. VS-WStd-01 as reference with its sinusoidal sensitivity
4. VS-WStd-01 as reference with its shock sensitivity obtained when calibrated against VS-WStd-09
© NMISA 2010
Conclusions
• A pendulum shock calibration system was successfully implemented using a vibration exciter and pendulum system with improved repeatability of the shock level.
• The effect of the misalignment of the exciter with the pendulum on the pendulum transverse motion was studied.
• It was found that the pendulum transverse motion is insensitive to the exciter/pendulum alignment.
• Calibration results obtained using three different accelerometers, traceable to shock and linear calibration primary systems, showed agreement within the estimated uncertainty of measurement.
© NMISA 2010
Thank you for your attention!