INDUSTRIAL OPERATION AND VALIDATION OF A DOUBLE DOOR
AUTOCLAVE
PRESENTED BYB.KARTHIKM. Sc MICROBIOLOGY
INTRODUCTION OF AN AUTOCLAVE:-
An autoclave is a pressure chamber used to carryout industrial
processes requiring elevated temperature and pressure different
from ambient air pressure.
It is used for the sterilization of media, glass were, gloves,
gowns etc.,
It is a moist type of sterilization unit.
Normal temperature and pressure required for autoclave is
121 C and 15lbs pressure. ⁰
Aim:- To operate and validate the industrial double door autoclave.
(Pharmalab)
Objective:- Operate and Validate the industrial double door auto clave.
Dehydrated media, glass were and accessories sterilization for
analysis .
HOW DOES AN AUTOCLAVE WORKS
• Steam enters the chamber jacket, passes through an operating valve and enters the rear of the chamber behind a baffle plate. It flows forward and down through the chamber and the load, exiting at the front bottom. A pressure regulator maintains jacket and chamber pressure at a minimum of 15 lbs, the pressure required for steam to reach 121ºC (250ºF). Overpressure protection is provided by a safety valve.
WHAT IS VALIDATION ?
• Validation shows that a process (or item of equipment) does what it is claimed to do.
• Why validate? – Need to show that sterile equipment is being taken
into sterile core - do not want to contaminate production (which could lead to adulterated product)
– For micro testing (both bacteriological and Viable Environmental Monitoring), using sterile equipment helps reduce false positives.
– Need to show that decontamination cycles are effective
HOW TO VALIDATE
• The usual – URS. IQ, OQ, PQ.
– This discussion will focus on PQ.
• IQ: does equipment meet the URS requirements? Is everything that was
on the box, in the box? Is the unit installed properly. Are support
programs in place for ongoing operation of A/C?
• OQ: does the A/C operate properly? Does the unit hold temp and pressure
correctly?
• PQ: validation of autoclave cycles and loading patterns – need to show
sterilization.
• What is D value?– Refers to decimal reduction time - The time required at a
certain temperature to kill 90% of the organisms being studied. Thus after an organism is reduced by 1 D, only 10% of the original organisms remain. Dependant on microbe and initial numbers. E.g. D value of 1.5 = 1CFU .
• What is Z value?– Refers to the temperature change required to produce a 1
log reduction in D value.• What is F0?– The number of minutes to kill a specified number of
microbes with a Z value of 10oC at a temp of 121.5oC.– Often confused with the time the chamber is held at
elevated temperature and pressure and in practice is the same thing.
• Overkill– Use many more microbes than would find
on items typically autoclaved. Negates the need to test sample for bio load before running the cycle.
– Use a sterilisation time exceeding what is necessary to kill a large number of microbes. Negates the need to determine D value of microbe.
Procedure to operate an autoclave :-
Switch on auto clave
Fill the water tank with water line
Switch On the heater switch
Open the door – 1
Load the chamber with media or glass were or with accessories
Arrange the sensors as we required
Close the door- 2
login with user id and password
Select the type of recipe
Start the cycle
Set up the strip chart
Observe the temperatures on the display
After completion of cycle open the door – 2
Give print to the printer
Calculate the F0- value, D-value and Z- value
with the use of computer software
Recipes for validation of a double door autoclaveThere are mainly 5 recipes present in the validation of a double
door auto clave.
They are :-
1. VLT (Vacuum leak test)
2. Bowie – dick test cycle
3. Steam In Process (SIP)
4. Gravity test cycle
5. HPHV test cycle
Vacuum leak test:-
Vacuum leak test is performed to check any leakage in
autoclaves. This test is also called chamber integrity test. In
VLT, first vacuum is applied inside the chamber and then wait
for the vacuum to reach at set vacuum point then hold the
vacuum for required time and when hold time completed
vacuum breakage occurs. After that check the results whether
autoclave leakage is within the limits or not.
PARAMETERS:-
Chamber level test vacuum :- -0.007bar
Stabilization time :- 5min
Vacuum hold time :- 10min
Acceptable leakage:- 0.0013bar/min
Print interval :- 60sec
Frequently test :- 15 days ± 2 days
Acceptable criteria :- 0.0013 bar/Min
No. of cycles run for validation :- 3
Bowie Dick test :- This is the cycle done for the removal of contaminated air
pockets in an autoclave.
In this the chemical indicator Bowie Dick
chemical indicator papers will be used.
The Bowie Dick indicator paper will turn from yellow to
black(or) brown in colour.
Bowie – Dick test papers
Parameters :-Pre vacuum :- -0.700Bar Pre vacuum hold time :- 01 minPre pressure :- 0.300BarPre pressure hold time :- 01 minHeat up 1 :-107 C⁰Heat up 1 hold :- 01minHeat up 2 :- 112⁰Heat up 2 hold :- 01minHeat up 3 :- 115⁰Heat up 3 hold :- 01 minHeat up 4:-117⁰Heat up 4 hold :- 1minSterile hold temp :- 121.3⁰
Control band :- 000.3Sterile hold time :- 11 minSterile hold reset :- 120.6⁰ Sterile hold stop count :- 121⁰Post vacuum :- -0.700BarPost vacuum hold time :- 02 minDoor open temp :- 90Print interval :- 60 secControl temp :- min for all
SIP (Steam In Process) cycle:-
It is the cycle done for vent filter sterilization of an autoclave.
It is the test done once per a month
Acceptance criteria :-
For SIP of filter temperature should be more than
121 C during sterilization.⁰
Parameters :-Heat up 1 :-107 C⁰
Heat up 1 hold :- 01min
Heat up 2 :- 112 C⁰
Heat up 2 hold :- 01min
Heat up 3 :- 115 C⁰
Heat up 3 hold :- 01 min
Heat up 4:-117 C⁰
Heat up 4 hold :- 1min
Sterile hold temp :- 122 C⁰
Sterile hold time :- 15 min
Sterile hold reset :- 120.6 C⁰
Sterile hold stop count :- 121 C⁰
Sterile temp over shoot :- 130 C⁰
Post vacuum :- -0.700Bar
Post vacuum hold time :- 02 min
Door open temp :- 90
Print interval :- 60 sec
Gravity cycle :- This is the cycle used for the sterilization of the media and water.
This test is done as per our requirement.
Temperature should be in between 121 – 124 during sterilization hold time.⁰ ⁰
The steam pressure should be 1.04-1.24 bar during the sterilization hold time.
In this the chemical indicator paper and biological indicator will be used.
The biological indicator is Geobacillus sterothermophillus.
This is a spore producing bacteria which can resist up to 120 C temperature.⁰
Parameters :-Heat up 1 :-105 C⁰Heat up 1 hold :- 02minHeat up 2 :- 108⁰Heat up 2 hold :- 02minHeat up 3 :- 112⁰Heat up 3 hold :- 02 minHeat up 4:-117⁰Heat up 4 hold :- 03minSterile hold temp :- 121.3⁰Control band :- 000.3Sterile hold time :- 20 minSterile hold reset :- 120.5⁰
Sterile hold stop count :- 121⁰Sterile temp over shoot :- 124⁰Jacket control temp :- 121⁰Door open temp :- 96⁰Print interval :- 60 secControl temp :- min for all
Biological indicator for gravity cycle Geobacillus sterothermophillus
Before test After test
HPHV cycle :-
High pressure and high vacuum test cycle is used for the
sterilization of a glass were, accessories etc.,
Temperature should be in between 121 – 124 during ⁰ ⁰sterilization hold time.
The steam pressure should be 1.04-1.24 bar during the
sterilization hold time.
Parameters :-Pre vacuum :- -0.700Bar Pre vacuum hold time :- 02 minPre pressure :- 0.300BarPre pressure hold time :- 02 minHeat up 1 :-107 C⁰Heat up 1 hold :- 02minHeat up 2 :- 112 C⁰Heat up 2 hold :- 02minHeat up 3 :- 115 C⁰Heat up 3 hold :- 02 minHeat up 4:-117 C⁰Heat up 4 hold :- 1minSterile hold temp :- 121.3 C⁰
Control band :- 000.3
Sterile hold time :- 30 minSterile hold reset :- 120.6 C⁰ Sterile hold stop count :- 121 C⁰Sterile temperature over shoot :-
124 C⁰Post vacuum :- -0.700BarPost vacuum hold time :- 02 minDoor open temp :- 90Print interval :- 60 sec
Observations and Results:Vacuum leak test:
Run - I Run - II Run –III
Date 25/05/2016 25/05/2016 25/05/2016
Cycle start time 14;53;50 15:16:37 15:59:19
Vacuum hold start time
15:01:16 15:24:04 16:06:49
Vacuum hold end time
15:11:16 15:34:04 16:16:49
Cycle completion time
15:12:35 15:35:23 16:18;09
Vacuum hold start pressure
-0.723 bar -0.724 bar -0.725 bar
Vacuum hold end pressure
-0.717 -0.717 -0.717
Actual leakage 0.006bar 0.007bar 0.008bar
Bowie Dick test Run –I Run – II Run – III
Date 25/05/2016 25/05/2016 25/05/2016
Cycle start time 16:38:01 18:04:20 19:11:44
Sterile hold start time 17:18:58 18:29:25 19:36:13
Sterile hold end time 17:29:58 18:40:25 19:47:13
Cycle completion time
17:47:32 19:00:08 20:05:42
No leaks observed Bowie Dick kit turns blue colour to black colour
Steam in process (SIP) test:Run – I Run – II Run - III
Date 27/05/2016 27/05/2016 27/05/2016
Cycle start time 13:36:00 14:29:42 16:14:40
Sterile hold start time
13:45:00 14:39:05 16:24:40
Sterile hold end time
14:00:00 14:54:05 16:39:40
Cycle completion time
14:03:45 14:57:35 16:42:20
Min. Temp during hold
121.1 121.1 121.1
Max. Temp during hold
124.4 124.5 124.5
Avg. Temp during hold
123.25 123.0 123.8
Min F0 22.79 22.64 27.52
Max F0 26.25 23.41 28.66
Conclusion
After operating and validating the autoclave all the
parameters which are given by USP is satisfied and the
results are also satisfied for all the recipes and all the
cycles what we done for validating the autoclave. So we
concluded that this autoclave is working very efficiently
and sterilizes the media and accessories properly without
any contamination. So now it is used for regular usage of
sterilization.